Philips Intellivue MP60, IntelliVue 70 User manual

IntelliVue MP60/70
Service Guide
IntelliVue Patient Monitor
MP60/70
Patient Monitoring
Part Number M8000-9301E
*M8000-9301E*
1 Introduction 13
Who Should Use This Guide 13 How to Use This Guide 13 Abbreviations 13 Responsibility of the Manufacturer 14 Passwords 14 Warnings and Cautions 14
2 Theory of Operation 15
Integrated Monitor Theory of Operation 15
System Boundaries 15 Hardware Building Blocks 17
IntelliVue MP60 17 IntelliVue MP70 18 Optional Hardware 19 Compatible Devices 20 Power Supply 21 CPU Boards 22 I/O Boards 22
Data Flow 23
Data Acquisition 23 Data Provider System Service 24 Persistent Data Storage System Service 24 Display and User Interface Service 24 Data Output 24 Monitor Applications 24 Internal LAN (Measurement Server Link) 25 Philips Clinical Network 26
How does the Support Tool Work with the Monitor 26 Monitor Software Block Diagram 26 Block Diagram Legend 28
3 Testing and Maintenance 33
Concepts 33 Test Reporting 33 Recommended Frequency 34 Tests Recommended When Performing... 35
Installation 35 Repair 35
3
Preventive Maintenance 35 Performance Verifications 35 Upgrades 36
Te s t s 36
Visual Test 36 Power On Test 36 NBP Tests 36
NBP Accuracy Test 36 NBP Leakage Test 37 NBP Linearity Test 38 Valve Test 38
Microstream CO2 Performance Test 38
Barometric Pressure Check and Calibration 39 Leakage Check 39 Pump Check 40 Flow Rate Check and Calibration 41 Noise Check 41 CO
Gas Measurement Calibration Check 41
2
Calibration Verification 42 Reset Time Counters 42
Temperature Accuracy 43 ECG/Resp Performance Test 43
ECG Performance 43 Respiration Performance 44
Invasive Pressure Performance Test 44 SpO2 Performance Test 44 Cardiac Output Performance Test 44
Service Tool Procedure, Version 1 45 Service Tool Procedure, Version 2 45
BIS Performance Test 45
PIC/DSC Test 45
Nurse Call Relay Performance Test 45
Phone Jack Type Connector Test (Traditional Nurse Call) 46 Multi-Port Nurse Call Connector Test (Flexible Nurse Call) 46
ECG Sync Performance Test 47 VueLink Tests using VueLink Test Module 47
Tes t P ro ce d u r e 4 8
Safety Testing 48
Warnings, Cautions, and Safety Precautions 49 Safety Test Procedures 49
Touchscreen Calibration 52 Disabling/Enabling Touch Operation 52
4 Troubleshooting 55
Introduction 55 How To Use This Section 55
4
Who Should Perform Repairs 55 Replacement Level Supported 55 Software Revision Check 56 Software Compatibility Matrix 56 Obtaining Replacement Parts 56 Troubleshooting Guide 56
Checks for Obvious Problems 57 Checks Before Opening the Instrument 57
Checks with the Instrument switched Off 57 Checks with the Instrument Switched On, AC connected 57 Initial Instrument Boot Phase 58
Troubleshooting Tables 59
How to use the Troubleshooting tables 59 Boot Phase Failures 60 Integrated Display is blank 62 Integrated Touch Display not functioning 63 External Display is blank 64 External Touch Display not functioning 65 General Monitor INOP Messages 65 Remote Alarm Device 67 Remote Extension Device 67 Speed Point 67 Keyboard/Mouse not functioning 68 Network related problems 69 Wireless Network 70 Multi-Measurement Server 71 MSL-related problems 71 Alarm Issues 73 Alarm Lamps 73 Alarm Tones 74 Alarm Behavior 74 Individual Parameter INOPS 74 Flexible Module Server 75 Integrated Module Slots 75 Printer 76 Recorder 77 MIB / RS232 78 Flexible Nurse Call Relay 79 Troubleshooting the ECG OUT 79 Data Flow Marker In and ECG Wave 79
Status Log 80 List of Error Codes 82 Troubleshooting with the Support Tool 82 Troubleshooting the Individual Measurements or Applications 82
5
5 Repair and Disassembly 83
Tools Required 83
Minimal Monitor Disassembly 83
Disconnecting the SpeedPoint 83 Removing the I/O Boards 84 Removing the ECG Out board if no MSL2 board is plugged 85 Removing the Integrated Module Slot, the Measurement Server Mount or blank covers 86 Separating the front and back half of the monitor 87 Removing Power Switch board 88 Removing the Backlights 89
Further Disassembly 89
Exchanging the Bezel (with or without Touchscreen) 89 Exchanging the LCD Assembly 91 Removing Power Supply 92 Removing the Speaker 93 Removing the ECG Out Board with an MSL2 Board plugged 94 Removing the Video Board 95 Removing the Main Board 96
Flexible Module Server (FMS) Disassembly 98
Removing the Handle and the Measurement Server Mount 98
Plug-in Modules 103
Plug-In Module Disassembly 103 tcpO2/tcpCO2 Calibration Chamber Kit 104 Recorder Module Paper 105
Disassembly Procedures for the Measurement Server Extension 106
Removing the Front Cover 106 Removing the Extension Bottom Cover 107 Removing the CO2 Scrubber 108 Removing the Pump 109
Refit Procedures for the Measurement Server Extension 109
Refitting the CO2 Scrubber 110 Refitting the Pump 110 Refitting the Extension Bottom Cover 110 Refitting the Front Cover 111 General Reassembly/Refitting Comments 111 Following Reassembly 111
6 Parts 113
MP60/MP70 Parts 114
Exchange Parts 114 Replacement Parts 115
Flexible Module Server Parts 117
Exchange and Replacement Parts 117
Multi-Measurement Server Parts 119
MMS Part Numbers - Front Bezel 120
6
MMS Exchange Part Numbers ­Software Revision A.05.xx and Lower
121
MMS Exchange Part Numbers ­Software Revision A.10.xx
122
Measurement Server Extension Parts (M3015A and M3016A) 124
Exchange Parts List 125
Plug-in Modules Part Numbers 126
Part Number Table 126
Exchange Modules, Table 1 126 Exchange Modules, Table 2 127
Plug-In Modules Replaceable Parts 128
Single-Width Plug-In Module 128 Double-Width Plug-In Module 129 Plug-in Module Replaceable Parts 129 Plug-In Module Language Specific Front Housing Kits (incl. Silicone Buttons, Frames & Bezels),
Table 1 129 Plug-In Module Language Specific Front Housing Kits (incl. Silicone Buttons, Frames & Bezels),
Table 2 130 Plug-In Module Specific Bezels 130
BIS Module Replaceable Parts 131
BIS Module Components 131
tcpO2/tcpCO2 Module Accessories 131
External Display Part Numbers 132 SpeedPoint Part Numbers 134 Remote Alarm Device Part Numbers 134 Remote Extension Device Part Numbers 135
135
7 Installation Instructions 137
Unpacking the Equipment 137 Initial Inspection 138
Mechanical Inspection 138 Electrical Inspection 138 Claims For Damage and Repackaging 138
Claims for Damage 138 Repackaging for Shipment or Storage 138
Installing the Monitor (M8005A or M8007A) 138
Mounting Instructions 139
Assembling Mounts 139
Connections 139 Installing Interface Boards 141 Connection of Devices via the MIB/RS232 Interface 141
Installing Remote Devices 141
Mounting the Remote Display (M8031A) 141
Connections 142
Mounting the 17” Remote Display (M8033A) 142
7
Connections 142
Flexible Module Server and/or Multi-Measurement Server 143
Attaching the MMS to a Mount 143 Detaching the Measurement Server from a Mount 143 Positioning the Measurement Server on a Clamp Mount 143 Mounting the MMS Mount to the FMS (M8048A) 144 Mounting the Remote Extension Device to the FMS 145 Mounting the BIS Module to the FMS 145 Mounting the FMS 147 Connections 147 MSL Cable Termination 147
Remote Alarm Devices 150
Mounting 150 Connections 150
Remote Extension Device 151
Mounting 151 Connections 152 Cabling 152
PS/2 Keyboard/Mouse 153
Philips Clinical Network (Wired) 153 Philips Clinical Network (Wireless) 153 Nurse Call Relay 153
Connections 153
ECG Out Functionality 154
Connections 154
Configuration Tasks 154
Setting Altitude and Line Frequency 155 Configuring the Equipment Label 155
8 Site Preparation 157
Introduction 157
Site Planning 157 Roles & Responsibilities 157
Site Preparation Responsibilities 157 Procedures for Local Staff 158 Procedures for Philips Personnel 160
Monitor M8005A and M8007A Site Requirements 160
Space Requirements 160 Environmental Requirements 160
Te m p e r at u r e 16 0 Humidity 161 Altitude 161
Electrical and Safety Requirements (Customer or Philips) 161
Safety Requirements 161 Electrical Requirements 161
8
Remote Device Site Requirements 161
Connecting Non-Medical Devices 162 Multi-Measurement Server M3001A or Flexible Module Server M8048A 163
Space Requirements Multi-Measurement Server M3001A 163 Space Requirements Flexible Module Server M8048A 163 Environmental Requirements Multi-Measurement Server M3001A 163 Environmental Requirements Flexible Module Server M8048A 163 Cabling Options and Conduit Size Requirements 164 Mounting 164
Remote Displays (M8031A) 165
Space Requirements 165 Environmental Requirements 165 Electrical and Safety Requirements 165
Remote Displays - M8033A 166
Space Requirements 166 Environmental Requirements 166 Electrical and Safety Requirements 166 Cabling Options and Conduit Size Requirements 166 Touch Cable 167
Remote Alarm Devices 167
Space Requirements 167 Mounting 167 Cabling Options and Conduit Size Requirements 167
Remote Extension Device 168
Space Requirements 168 Mounting 168 Cabling Options and Conduit Size Requirements 168 Input Devices 169
Local Printer 169
Philips Medical LAN 169 MIB Interface 170 Nurse Call Relay Interface 171 ECG Out Interface 171
9 Anesthetic Gas Module 173
Introduction 173
Description 173 Product Structure 173 Physical Specifications 173 Environmental Specifications 174 Performance Specifications 174
CO2 Measurement 175 AWRR derived from CO2 Waveform 175 N2O Measurement 175 O2 Measurement 175 Anesthetic Agent Measurement 175
9
Alarm Ranges 176 Alarm Delay 176 Apnea Alarm 176 INOP Alarms 176
General Measurement Principles 177 Theory of Operation 177
Main PC Board 178 Power Supply 179 Pneumatic System 179 Pump 180 Wate rt rap 180 Sample Flow Through the Pneumatic Path 181
Agent Identification Assembly 181
Measurement Principle 182
O2 Sensor 182
Specifications 182 Measurement Principle 182
Infrared Measurement Assembly 183
Installation and Patient Safety 184
Physical Installation 184 Environment 185 Label Sheet 185 Making Connections to the AGM 185 Sample Gas Connections to the Gas Exhaust 186
Returning the Gas Sample 186 Setting Up the Gas Return 187 Removing the Gas Sample 188
Setup and Configuration Procedures 188
Altitude Configuration 188 Connect Sample Input Tubing 188
Preventive Maintenance (PM) Tasks 188 Post-Installation Checks 189 Safety Requirements Compliance and Considerations 189
Explanation of Symbols Used 189 Power Supply Requirements 190 Grounding the System 190 Equipotential Grounding 191 Combining Equipment 191
Checking and Calibrating the Anesthetic Gas Module 191
Access Service Functions of the M1026A Anesthetic Gas Module 191 When and how to check the Philips M1026A Anesthetic Gas Module 193 Equipment required for checking 193 Checks and adjustments 194
Performance Leakage Check 194 Performance Diagnostic Check 195 Performance Flowrate Check 195
10
Total Flowrate Check and Adjustment in Purge Mode 195 Measurement Path Flowrate Check and Adjustment 196 Total Flowrate Check in Normal Mode 198
Zero Calibration 198 Barometric Pressure Check and Calibration 199 Span Calibration Check 200
Disposal of Empty Calibration Gas Cylinder 202
Maintaining the Anesthetic Gas Module 203
Preventive Maintenance (PM) Tasks 203 Cleaning 203 Replace PM Parts 204
Internal Nafion Tubing with Bacterial Filters and manifold Seals 204
Room-Air Filter 205 Pump Filter 206 Performance Checks 207
Other factors to maximize uptime or reduce cost of ownership: 207
Troubleshooting the Anesthetic Gas Module 207
Compatibility Criteria for the AGM and the IntelliVue Monitors 208 Flow Charts for Communication and Measurement Type Problems 208 Hardware Related Troubleshooting Strategy 213 INOPs 214 Calibration Checks 216
Calibration Checks Troubleshooting Table 217
Diagnostic Checks 218
Problem Solving Hierarchy 219 Pneumatic System Diagnostic Checks 220 O2 Assembly Diagnostic Checks 220 Optical Path Disgnostic Checks 223 IR Measurement Assembly Diagnostic Checks 224 Agent ID Assmebly Diagnostic Checks 225 Power Supply Diagnostic Checks 226 Operating Temperature Diagnostic Checks 227
Test Points, Connectors and Jumpers 227
Tes t Po in ts 2 27 Connectors 228 Jumpers 228
Repairing the Anesthetic Gas Module 230
Introduction 230 The Top Cover 232
Removal 232 Replacement 232
Lifting the IR Measurement Mounting Bracket 234
Removal 235 Replacement 235
Infrared Measurement Assembly Head 237
11
Transferring NVRAM Data to a Replacement Head 237
Sample Cell 241
Removal 241 Replacement 241
Solenoid Valve #1 245
Removal 245 Replacement 245
Power Supply Unit 247
Removal 247 Replacement 247
Main PC Board 248
Removal 248 Replacement 249
O2 Sensor 250
Removal 250 Replacement 251
Agent Identification Head 253
Removal 253 Replacement 254
Pump 255
Removal 255 Replacement 255
Fan 256
Removal 256 Replacement 256
Solenoid Valve #2 258
Removal 258 Replacement 258
Top Co ve r P C B o ar d 259
Removal 259 Replacement 259
Watertrap Manifold and Protector 261
Removal 261 Replacement 261
Power Fuses 262
Removal 262 Replacement 262
Test and Inspection Matrix 262 When to Perform Test Blocks 266 Safety Test Appendix 267
Parts List 269
Calibration Equipment 274
12
This Service Guide contains technical details for the IntelliVue MP60 and MP70 Patient Monitor, the Multi- Measurement Server (MMS), the Flexible Module Server (FMS) and the Measurement Server Extensions.
This guide provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanation of the product architecture or technical implementation. It offers enough information on the functions and operations of the monitoring systems so that engineers who repair them are better able to understand how they work.
It covers the physiological measurements that the products provide, the Measurement Server that acquires those measurements, and the monitoring system that displays them.

Who Should Use This Guide

1

1Introduction

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How to Use This Guide

This guide is divided into eight sections. Navigate through the table of contents at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Abbreviations

Abbreviations used throughout this guide are:
Name Abbreviation
IntelliVue MP60/MP70 Patient Monitor the monitor
Flexible Module Server FMS
Multi-Measurement Server MMS
Measurement Server Link MSL
Medical Information Bus MIB
Anesthetic Gas Module AGM
13
1 Introduction Responsibility of the Manufacturer

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, reliability and performance of the equipment if:
• assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the instructions for use.
To ensure safety, use only those Philips parts and accessories specified for use with the monitor. If non­Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medizinsysteme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Passwords

In order to access different modes within the monitor a password may be required. The passwords are listed below.
Monitoring Mode: No password required
Configuration Mode: 71034
Demo Mode: 14432
Service Mode: 1345
Consult the configuration guide before making any changes to the monitor configuration.

Warnings and Cautions

In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.
14
•A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.

2Theory of Operation

Integrated Monitor Theory of Operation

The IntelliVue Patient Monitor:
• displays real-time data
• controls the attached measurement servers
• alarms in the case of patient or equipment problems
• offers limited data storage and retrieval (trending)
• interfaces to the Philips Clinical Network and other equipment
A monitor with just a single integrated measurement server can be connected to additional building blocks to form a monitoring system with a large number of measurements, additional interface capabilities and multiple slave displays. These elements cooperate as one single integrated real-time measurement system.
2

System Boundaries

The following diagram discusses specific boundaries within the overall system with respect to their openness and real-time requirements:
15
2 Theory of Operation Integrated Monitor Theory of Operation
Philips Clinical Network
Measurement LAN
combines components of one patient monitor; real time requirements across all interconnected elements
Philips Clinical Network (wired LAN)
connects multiple patient monitors, information centers, application servers; closed system, only Philips qualified products (tested and with regulatory approval) are connected, Philips is responsible for guaranteed real-time functionality and performance
Philips Clinical Network (wireless)
like Philips Clinical Network (wired) LAN, however due to current wireless technologies available it has reduced bandwidth, longer latencies, reduced functionality
Hospital LAN, Internet
Standard Network, not under Philips control, no guaranteed service, no real-time requirements
16
Integrated Monitor Theory of Operation 2 Theory of Operation

Hardware Building Blocks

The following hardware building blocks make up the monitoring system:
IntelliVue MP60
The MP60 monitor:
• integrates the display and processing unit into a single package
• uses a 15” TFT XGA Color display
• uses the Philips SpeedPoint as primary input device; computer devices such as mice, trackball, and keyboard can be added optionally
• has an optional recorder
• supports the Flexible Module Server (FMS)
Building Blocks:
17
2 Theory of Operation Integrated Monitor Theory of Operation
|| I/F To Local Printer
Power Supply
I/F
PS/2 To SpeedPoint
Boards
LCD
Assembly
Main Board
MIB To AGM
LCD
Adapter
MSL
I/F
Video I/F
Board
To Ext. Display
ECG Out
IntelliVue MP70
The MP70 monitor:
• integrates the display and processing unit into a single package,
• uses a 15” TFT XGA Color display
• uses the Philips Touchscreen as primary input device, whereas the Philips SpeedPoint and computer devices such as mice, trackball, and keyboard can be added optionally
18
• has an optional recorder
• supports the Flexible Module Server (FMS)
Integrated Monitor Theory of Operation 2 Theory of Operation
es
Building Blocks:
|| I/F To Local Printer
Power Supply
I/F
PS/2 To External input devic
Boards
LCD
Assembly
Touch
Panel
MIB To AGM
Main Board
LCD
Adapter
Touch
Controller
Optional Hardware
A measurement server mount and /or an integrated module slot can be ordered optionally. If the monitor is ordered with the wireless LAN option an external wireless transmitter is required. For further details regarding the wireless network please refer to the M3185A Philips Clinical Network documentation.
Video I/F
Board
MSL
I/F
To Ext. Display
19
2 Theory of Operation Integrated Monitor Theory of Operation
Measurement Server Mount
Integrated Module Slot
Compatible Devices
Figure 1 M8048A Flexible Module Server (FMS)
20
Integrated Monitor Theory of Operation 2 Theory of Operation
Figure 2 M3001A Multi-Measurement Server (MMS)
Power Supply
AC/DC 48V /120W
Isolating DC/DC Converter
Backlight DC/DC Converter
DC/DC
DC/DC
DC Bus
Integrated Module Rack
DC/DC Converter
DC/DC
DC/DC
DC/DC
56 V
FMS MMS, MMS-EXT
12 V
3.3 V
5 V
10 V AC
)
12 V
)
unreg
5 V
60 V
5 V, 60V
HIF, LED's,
Front End Modules
Backlight
CPU
I/F boards
wireless
Front End
Modules
Figure 3 Power Supply Architecture
21
2 Theory of Operation Integrated Monitor Theory of Operation
The AC/DC converter transforms the AC power coming from the power plug into 48 V/120W DC source and isolates the monitoring system from the AC power mains.The 48V is distributed via power bus and supplies power to all the components of the system: The 56 V DC power needed for the FMS, MMS and measurement server extension is created by an isolating DC/DC converter. The power needed for the backlights is converted to 12V DC by the backlight DC/DC converter. The CPU is supplied with 3.3 V and 5 V DC power. The transformation is performed in two steps: The first DC/ DC converter is a power regulator which reduces the variations caused by load changes on the 48V power bus. The second DC/DC converter converts the power to the needed voltage. Interface boards require a power of 10V AC. The HIF board and the LEDs are supplied with 12V DC unregulated power. The integrated module slot requires a 5 V supply for the modules slots and uses the 48V and another DC/DC converter to create 60 V in order to supply power for the modules.
CPU Boards
The CPU boards have an MPC860 50 MHz processor that provides a number of on-chip, configurable interfaces. An array of 12 fast UARTS with configurable protocol options are implemented in an ASIC (along with other system functions such as independent watchdogs etc.), providing interfacing capabilities to measurement modules and I/O boards. The serial interfaces can easily be electrically isolated. The main board contains additional video hardware.
I/O Boards
Flexible Module Server
CPU
Bank of I/Os
Modules
The CPUs provide two LAN interfaces to interconnect CPUs (via the Internal LAN) or to connect to the Philips Clinical Network.
The CPU capabilities are identical. Different loading options are coded on serial EEPROMs to support the automatic configuration of the operating system at boot time.
Interfaces to the monitor are implemented via I/O boards. The location of these boards is restricted by general rules. The I/O slot designations diagram and the I/O matrix which outline the I/O board placement rules can be found in the Installation Instructions section.
IntelliVue Patient Monitor Multi-Measurement Server
CPU Video
Bank of I/Os
CPU
Bank of I/Os
Measurement
Interfaces
Acquisition
22
The following is a list of Interface (I/O) boards which may be present in your monitor, depending on your purchased configuration:
•MSL
•Video (analog)
• Philips Clinical Network (LAN wired or wireless)
Integrated Monitor Theory of Operation 2 Theory of Operation
•PS/2
• MIB/RS232
•Flexible Nurse Call
•Parallel printer
• Remote devices (Remote Alarm Device, Remote Extension Device)
The specifications for the above listed interfaces can be found in the technical data sheet for the monitor and in the Specifications chapter of the Instructions for Use.

Data Flow

The following diagram shows how data is passed through the monitoring system. The individual stages of data flow are explained below.
Display
and User
Interface
Data
Acquisition
Data
Provider Service
Applications
Data Acquisition
Monitoring data (for example patient measurement data in the form of waves, numerics and alerts) is acquired from a variety of sources:
•Measurement Servers
The Measurement Servers connected to the internal LAN convert patient signals to digital data and apply measurement algorithms to analyze the signals.
• External measurement devices
Data can be also acquired from devices connected to interface boards of the monitor. Software modules dedicated to such specific devices convert the data received from an external device to the format used internally. This applies to parameter modules and the Anesthetic Gas Module
• Server systems on the Philips Clinical Network
To enable networked applications such as the other bed overview, data can be acquired from server systems attached to the Philips Clinical Network, for example a Philips Information Center
Persistent
Data
Storage
Data
Output
23
2 Theory of Operation Integrated Monitor Theory of Operation
Data Provider System Service
All data that is acquired from measurement servers or external measurement devices is temporarily stored by a dedicated data provider system service. All monitor applications use this central service to access the data in a consistent and synchronized way rather than talking to the interfaces directly.
This service makes the applications independent of the actual type of data acquisition device.
The amount of data stored in the data provider system service varies for the different data types. for example several seconds of wave forms and the full set of current numerical values are temorarily stored in RAM.
Persistent Data Storage System Service
Some applications require storage of data over longer periods of time. They can use the persistent data storage system service. Dependent on the application requirements, this service can store data either in battery backed-up (buffered) memory or in flash memory. The buffered memory will lose its contents if the monitor is without power (not connected to mains) for an extended period of time. The flash memory does not lose its contents.
The trend application for example stores vital signs data in a combination of flash memory and buffered memory, while the system configuration information (profiles) is kept purely in flash memory.
Display and User Interface Service
Applications can use high level commands to display monitoring data or status and command windows on the internal LCD panel. These commands are interpreted by the display manager application. This application controls the dedicated video hardware which includes video memory and a special ASIC.
User input is acquired from a variety of input devices, for example the SpeedPoint, the touchscreen or other standard input devices (keyboard, mouse) which may be attached to I/O boards. The system software makes sure that the user input is directed to the application which has the operating focus.
Data Output
The monitoring system is very flexible and customizable regarding its data output devices. Built-in devices (for example LAN, alarm lamps, speaker, video) provide the basic output capabilities.
These capabilities can be enhanced by adding additional I/O boards, as required in the specific end­user setup. The additional I/O boards typically provide data to externally attached devices, for example to printers, RS232 based data collection devices, nurse call systems etc.
The monitor can identify I/O boards by means of a serial EEPROM device that stores type and version information. The operating system detects the I/O boards and automatically connects them with the associated (interface driver) application. For some multi- purpose cards it is necessary to configure the card for a particular purpose first (for example the dual MIB/RS232 card can support external touch display, data import, data export).
Monitor Applications
The monitor applications provide additional system functionality over the basic measurement and monitoring capabilities. This includes for example trending, report generating, event storage or derived measurements.
24
Integrated Monitor Theory of Operation 2 Theory of Operation
In general, the monitor applications use the data provider system service to access the measurement data. Application interfaces to the other system services allow the application to visualize data, to store data over extended periods of time or to output data to other devices.
Internal LAN (Measurement Server Link)
All components of the monitoring system (including measurement servers and CPUs in the monitor) communicate using an IEEE802.3/ Ethernet LAN in the Measurement Server Link (MSL). This network is used to distribute data between the components, for example:
• Digitized patient signals including wave data, numerical data and status information (typically from the measurement server to a display unit)
• Control data representing user interactions (typically from the display unit to a measurement server)
• Shared data structures, for example representing patient demographical data and global configuration items
The internal LAN allows plug and play configuration of the monitoring system. The system automatically detects plugging or unplugging of measurement servers and configures the system accordingly.
The components on the internal LAN are time- synchronized to keep signal data consistent in the system. Dedicated hardware support for synchronization eliminates any latency of the network driver software.
The integrated LAN provides deterministic bandwidth allocation/ reservation mechanisms so that the real-time characteristic of signal data and control data exchange is guaranteed. This applies to the data flow from the measurement server to the monitor (for example measurement signal data) and the data flow from the monitor to a measurement server (for example to feed data to a recorder module).
Integrated communication hubs in the monitor and the FMS allow flexible cabling options (star topology, daisy chaining of servers).
MDSE Internal LAN
MDSE
Internal
LAN
MDSE
Internal
LAN
25
2 Theory of Operation Integrated Monitor Theory of Operation
Philips Clinical Network
The monitoring system may be connected to the Philips Clinical Network, for example to provide central monitoring capabilities or other network services. This connection may be through a normal wired connection or through a wireless connection.
The monitor supports the connection of an external off-the-shelf wireless adapter. This allows a simple field upgrade as well as a technology upgrade in the future. Switching between wired and wireless networks is automatically triggered by the plugging or unplugging of the network cable.
The Philips Clinical Network protocols function very similarly to the protocols used on the internal LAN.
After configuration, the monitoring system sends the digitized patient signals including wave data, numerical data and status information onto the network. Control data representing user interactions can be exchanged between the monitoring system and a central station bi-directionally.
Additional protocols are supported for networked applications, for example for the other bed overview function, which allows viewing of monitoring data from other patients on the network.
For plug and play operation, the monitoring system uses the standard BootP protocol to automatically acquire a network address.

How does the Support Tool Work with the Monitor

The support tool is a Windows application typically installed on the laptop of a customer engineer or a biomedical engineer working in the customer’s own service department.
The purpose of the support tool is to upgrade, configure and diagnose all monitoring components (modules, measurement servers, and monitors) in the system over the network.
The service protocol developed for this purpose uses a raw access to the devices without the need for IP addresses etc. over a standard customer network installation, so that even defective devices can be upgraded as long as the few kBytes of initial boot code are working. The boot code itself can also be upgraded using the same protocol.
The tool allows access to internal service information and to serial numbers. It can be remote­controlled, for example via a dial-up connection from a response center, provided the proper infrastructure is in place.
For details see the Instructions for Use for the Support Tool.

Monitor Software Block Diagram

Figure 4 shows the functional block diagram for the monitoring system. A legend explaining terms and diagram elements follows. The information below varies depending on the purchased monitor options.
26
Integrated Monitor Theory of Operation 2 Theory of Operation
Philips Clinical
Network
LAN
MDSE
Video Out
Indicators
Color LCD
Display
Applications
System Services
Real Time Operating System
Visual
LEDs
Audio
Indicators
Loudspeaker Touch
Input Devices (including PS/2)
Trim Knob
Interfaces
Interface Managers
Record Alarm Trend HiRes ADT
Reports
Calc Param
Events
ECG-Out Marker-In
ECG-Out Marker-In
MDSE
LAN
LAN
RS-422
M3001A Multi­Measurement Server
12-lead ECG/Resp, NBP, SpO2,
Press/Temp
M3015/16A Measurement Server
Extension
CO2, Press/Temp
M1006B Press
Plug-In Modules
M1012A C.O.
LAN RS-422
M1018A tcPO2/CO2
M1029A Temp
RS-422
Flexible Module Server
M1032A VueLink
M1116B Recorder
Figure 4 IntelliVue Patient Monitoring System Functional Block Diagram
27
2 Theory of Operation Integrated Monitor Theory of Operation

Block Diagram Legend

Functional Block Description
Services
Operating System The Operating System (OS) provides a layer of isolation between
the specific hardware implementation and the application software. The OS performs system checks and allocates resources to ensure safe operation when the system is first started. This includes internal self-tests on several hardware modules and configuration checks for validity of configuration with the operating software. During normal operation, the OS continues to run checks on system integrity. If error conditions are detected the OS will halt monitoring operations and inform the operator about the error condition.
System Services The System Services provide generic common system services.
In particular: It uses a real-time clock component to track time. It synchronizes to network time sources and verifies the accuracy of the system time information. It is also responsible for managing persistent user configuration data for all Measurement Servers, Flexible Module Servers and IntelliVue Patient Monitoring System software modules. User configuration data is stored in a non­volatile read/write storage device
Applications
Reports The Reports Service retrieves current and stored physiological
data and status data to format reports for printing paper documentation. The following reports are supported:
•Vital Signs Report
• Graphical Trend Report
• Event Review Report
•Event Episode Report
• ECG Report (12 Lead/Multi-Lead)
• Cardiac Output Report
• Calculations Report (Hemodynamic/Oxygenation/ Ventilation)
• Calculations Review Report
•Wedge Report
•Test Report
• Other reports (e.g. Loops, Review Applications, Drug report)
The Reports service generates report data which can be printed on a local or a central printer.
28
Integrated Monitor Theory of Operation 2 Theory of Operation
Functional Block Description
Record The Record Service retrieves current and stored physiological
data and status data to format a continuous strip recording. A recording can be triggered manually by the operator or automatically by an alarm condition. The Record Service uses the services of the Recorder Interface to control an M1116B Recorder in the FMS. The Record Service can also send data to a central recorder.
Alarm The Alarm Service contains logic that prioritizes alarm conditions
that are generated either by the Measurement Servers, Flexible Module Server, or by IntelliVue Patient Monitoring System software modules. Visual alarm signals (messages) are displayed at the top of the IntelliVue Patient Monitoring System display and alarm sounds are generated by a loudspeaker. Alarm conditions may be generated when a physiological parameter exceeds preselected alarm limits or when a physiological parameter or any other software module reports an inoperative status (technical alarm, for example, the ECG leads may have fallen off the patient). The Alarm service manages the alarm inactivation states, for example suspension of alarms, silencing of alarms, and alarm reminder. Alarm signals may also be configured as latching (alarm signals are issued until they are acknowledged by the operator, even when the alarm condition is no longer true). The Alarm service controls the visual alarm signals (alarm lamps).
Trend The Trend service stores the sample values of physiological data
and status data with a resolution of 12 seconds, 1 minute or 5 minutes for a period of up to 48 hours. The data is kept in battery buffered read/write storage and flash memory devices to be preserved across power failures. The stored data is protected via consistency checks and checksums. When a new patient is admitted, the trend database erases all data of the previous patient.
HiRes The OxyCRG (Oxygen CardioRespiroGram) service derives a
high-resolution trend graph from the Beat-to-Beat Heart Rate, SpO
or tcpO2, and Respiration physiological data. The
2
OxyCRG is specialized for neonatal applications, allowing the opeartor to identify sudden drops in Heart Rate (Bradycardia) and SpO
or tcpO2 (Desaturations), and supporting the operator
2
in visualizing Apnea situations.
ADT The ADT (Admit/Discharge/Transmit) service maintains the
patient demographics information. The operator may admit a new patient, discharge the old patient and enter or modify the patient demographics. The ADT service also supports the transport of a patient (trend database) with the M3001A Multi­Measurement Server. The ADT service controls the deletion of old patient data, the upload of trend data from the M3001A and the switching back of all settings to user defaults. It also synchronizes patient information with a central station on the network.
29
2 Theory of Operation Integrated Monitor Theory of Operation
Functional Block Description
Calc Param The Calc Param (Calculated Parameters) service accesses current,
stored and manually entered physiological data as input to calculation formulas. With these formulas, derived hemodynamic, oxygenation and ventilation variables are computed. The calculation results, including the input parameters, are stored for later review using the Trend service.
Interface Managers
MDSE The MDSE (Medical Data Service Element) Interface Manager is
responsible for the exchange of real-time data between the IntelliVue Patient Monitoring System display unit and the Measurement Servers and Flexible Module Server as well as between the IntelliVue Patient Monitoring System display unit and other devices attached to the network. MDSE establishes and maintains a data communication link between the devices. It provides configuration information about the remote device to applications in the local device and it allows the exchange of measurement data and status information between the devices.
Printer The Printer Interface Manager provides a high level interface to a
printer. It provides means to:
• establish a connection to the printer
• transfer data to the printer
• get status of the printer
• close connection to the printer
The Printer Interface Manager also supervises the connection to the printer and whether the printer accepts data (for example paper out). The Printer Interface Manager notifies the operator in such cases.
30
Loading...
+ 246 hidden pages