EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Symbols
Symbols
The meaning(s) of the symbol(s) shown on the package with the components,
the back cover of this instruction manual and/or this instrument are as follows:
Refer to instructions.
Endoscope
TYPE BF applied part
Manufacturer
Authorised representative in the European Community
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
1
Important Information — Please Read Before Use
Important Information — Please Read
Before Use
Intended use
These instruments have been designed to be used with an Olympus video
system center, light source, documentation equipment, video monitor,
endo-therapy accessories (such as a biopsy forceps) and other ancillary
equipment.
Use the GIF-XP160, GIF-160, GIF-Q160, GIF-1TQ160, GIF-XTQ160 for
endoscopy and endoscopic surgery within the upper digestive tract (including
the esophagus, stomach and duodenum).
Use the CF-Q160L/I, CF-Q160AL/I, PCF-160AL/I for endoscopy and endoscopic
surgery within the lower digestive tract (including the anus, rectum, sigmoid
colon, colon and ileocecal valve).
Use the CF-Q160S for endoscopy and endoscopic surgery within the lower
digestive tract (including the anus, rectum and sigmoid colon).
Do not use these instruments for any purpose other than their intended uses.
Applicability of endoscopy and endoscopic treatment
If there is an official standard on the applicability of endoscopy and endoscopic
treatment that is defined by the hospital’s administration or other official
institutions such as academic societies on endoscopy, follow that standard.
Before starting endoscopy and endoscopic treatment, thoroughly evaluate its
properties, purposes, effects, and possible risk (their natures, extent and
probability). Perform endoscopy and endoscopic treatment only when its
potential benefits are greater than its risks.
Fully explain to the patient the potential benefits and risks of the endoscopy and
endoscopic treatment as well as any examination/treatment methods that can be
performed in its place, and perform the endoscopy and endoscopic treatment
only after obtaining the consent of the patient.
Even after starting the endoscopy and endoscopic treatment, continue to
evaluate the potential benefits and risks, and immediately stop the
endoscopy/treatment and take proper measures if the risks to the patient
become greater than the potential benefits.
2
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Instruction manual
This instruction manual contains essential information on using this instrument
safely and effectively. Before use, thoroughly review this manual and the
manuals of all equipment which will be used during the procedure and use the
equipment as instructed.
Note that the complete instruction manual set for this endoscope consists of this
manual and the “REPROCESSING MANUAL” whose cover lists the model of
your endoscope. It also accompanied the endoscope at shipment.
Keep this and all related instruction manuals in a safe, accessible location.
If you have any questions or comments about any information in this manual,
please contact Olympus.
User qualifications
Important Information — Please Read Before Use
The operator of this instrument must be a physician or medical personnel under
the supervision of a physician and must have received sufficient training in
clinical endoscopic technique. This manual, therefore, does not explain or
discuss clinical endoscopic procedures. For details on the clinical endoscopic
procedures, the physician and operator are requested to form judgments from
their viewpoints as specialists.
Instrument compatibility
Refer to the “System chart” in the Appendix to confirm that this instrument is
compatible with the ancillary equipment being used. Using incompatible
equipment can result in patient or operator injury and/or equipment damage.
This instrument complies with EMC standard for medical electrical equipment;
edition 2 (IEC 60601-1-2: 2001). However, when connected with an instrument
that complies with EMC standard for medical electrical equipment; edition 1
(IEC 60601-1-2: 1993), the whole system complies with edition 1.
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
3
Important Information — Please Read Before Use
Reprocessing before the first use/reprocessing and
storage after use
This instrument was not cleaned, disinfected or sterilized before shipment.
Before using this instrument for the first time, reprocess it according to the
instructions given in the endoscope’s companion manual, the “REPROCESSING
MANUAL” whose cover lists the model of your endoscope.
After using this instrument, reprocess and store it according to the instructions
given in the endoscope’s companion reprocessing manual. Improper and/or
incomplete reprocessing or storage can present an infection control risk, cause
equipment damage or reduce performance.
Spare equipment
Be sure to prepare another endoscope to avoid that the examination will be
interrupted due to equipment failure or malfunction.
Repair and modification
This instrument does not contain any user-serviceable parts. Do not
disassemble, modify or attempt to repair it; patient or operator injury and/or
equipment damage can result. This instrument is to be repaired by Olympus
technicians only.
Signal words
The following signal words are used throughout this manual:
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury. It may also
be used to alert against unsafe practices or potential
equipment damage.
4
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Warnings and cautions
Follow the warnings and cautions given below when handling this instrument.
This information is to be supplemented by the warnings and cautions given in
each chapter.
•After using this instrument, reprocess and store it according
Important Information — Please Read Before Use
Indicates additional helpful information.
to the instructions given in the endoscope’s companion
reprocessing manual. Using improperly or incompletely
reprocessed or stored instruments may cause patient
cross-contamination and/or infection.
•Do not strike, bend, hit, pull, twist, or drop the endoscope’s
distal end, insertion tube, bending section, control section,
universal cord, or endoscope connector of the endoscope
with excessive force. The endoscope may be damaged and
could cause patient injury, burns, bleeding and/or
perforations. It could also cause parts of the endoscope to fall
off inside the patient.
•Never perform angulation control forcibly or abruptly. Never
forcefully pull, twist or rotate the angulated bending section.
Patient injury, bleeding and/or perforation can result. It may
also become impossible to straighten the bending section
during an examination.
•Never insert or withdraw the endoscope’s insertion tube while
the bending section is locked in position. Patient injury can
result.
•Do not touch the light guide of the endoscope connector
immediately after removing it from the light source because it
is extremely hot. Operator or patient burns can result.
•Never perform flexibility adjustment, operate the bending
section, feed air or perform suction, insert or withdraw the
endoscope’s insertion tube without viewing the endoscopic
image. Never use endo-therapy accessories without viewing
the endoscopic image. Patient injury can result.
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
5
Important Information — Please Read Before Use
•Never perform flexibility adjustment, operate the bending
•Regardless of the flexibility of the endoscope’s insertion tube,
•Do not pull the universal cord during an examination. The
•Do not coil the insertion tube or universal cord into a diameter
section, feed air or perform suction, insert or withdraw the
endoscope’s insertion tube while the image is frozen. Never
use endo-therapy accessories while the image is frozen.
Patient injury can result.
never insert or withdraw it with excessive force. Otherwise,
patient injury could result.
endoscope connector will be pulled out from the output
socket of the light source and the endoscopic image will not
be visible.
of less than 12 cm. Equipment damage can result.
•Do not touch the electrical contacts inside the electrical
connector. CCD damage may result.
•Do not apply shock to the distal end of the insertion tube,
particularly the objective lens surface at the distal end. Visual
abnormalities may result.
•Do not twist or bend the bending section with your hands.
Equipment damage may result.
•Do not squeeze the bending section forcefully. The covering
of the bending section may stretch or break and cause water
leaks.
•Turn the video system center OFF before connecting or
disconnecting the videoscope cable from the electrical
connector on the endoscope. Turn the switch ON or OFF
only when the videoscope cable is connected to both the
video system center and electrical connector on the
endoscope. Failure to do so can result in equipment damage,
including destruction of the CCD.
•The endoscope’s remote switches cannot be removed from
the control section. Pressing, pulling or twisting them with
excessive force can break the switches and/or may cause
water leaks.
6
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Important Information — Please Read Before Use
•If remote switch 1 does not return to the OFF position after
being pressed strongly from the side, gently pull the switch
upwards to return it to the OFF position.
•Do not hit or bend the electrical contacts on the endoscope
connector. The connection to the light source may be
impaired and faulty contact can result.
•Do not attempt to bend the endoscope’s insertion tube with
excessive force. Otherwise, the insertion tube may be
damaged.
•Do not attempt to bend the endoscope’s insertion tube with
excessive force unless flexibility is at the stiffest position.
Otherwise, the insertion tube may be damaged.
•The endoscope contains a memory chip that stores
information about the endoscope and communicates this
information to the CV-160. Although the memory chip is
durable, damage will prevent data from being backed up on
it. When data are lost or damaged, contact Olympus.
•Electromagnetic interference may occur on this instrument
near equipment marked with the following symbol or other
portable and mobile RF (Radio Frequency) communications
equipment such as cellular phones. If electromagnetic
interference occurs, mitigation measures may be necessary,
such as reorienting or relocating this instrument, or shielding
the location.
Examples of inappropriate handling
Details on clinical endoscopic technique are the responsibility of trained
specialists. Patient safety in endoscopic examinations and endoscopic treatment
can be ensured through appropriate handling by the physician and the medical
facility. Examples of inappropriate handling are given below;
•Over-insufflating the lumen may cause patient pain and/or perforation.
•Applying prolonged suction with the distal end in contact with the
mucosal surface may cause bleeding or suction lesions.
•Retroflexing the endoscope within the esophagus or duodenal bulb may
cause mucosal trauma or impaction of the endoscope (for GIF models
only).
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7
Important Information — Please Read Before Use
•Inserting, withdrawing and using endo-therapy accessories without a
clear endoscopic image may cause burns or perforation.
•Inserting or withdrawing the endoscope, feeding air, applying suction or
operating the bending section without a clear endoscopic image may
cause patient injury.
8
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Chapter 1 Checking the Package Contents
Chapter 1Checking the Package
Contents
Match all items in the package with the components shown below. Inspect each
item for damage. If the instrument is damaged, a component is missing or you
have any questions, do not use the instrument; immediately contact Olympus.
This instrument was not disinfected or sterilized before shipment.
Before using this instrument for the first time, reprocess it according to the
instructions given in the endoscope’s companion manual, the “REPROCESSING
MANUAL” whose cover lists the model of your endoscope.
Channel cleaning brush
(BW-20T)
Endoscope
Channel cleaning brush
(BW-17K for GIF-XTQ160 only)
Injection tube (MH-946)
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
9
Chapter 1 Checking the Package Contents
Channel-opening cleaning
brush (MH-507)
Mouthpiece
(MA-474, MB-142 for
GIF-XP160, 1 pc. each)
Suction valve
(MH-443, 2 pcs.)
Suction cleaning adapter
(MH-856)
AW channel cleaning
adapter (MH-948)
Mouthpiece (MB-142 for GIF-160,
GIF-Q160, GIF-1TQ160,
GIF-XTQ160, 2 pcs.)
Air/water valve
(MH-438, 2 pcs.)
Auxiliary water tube
(MAJ-855 for endoscopes with
auxiliary water feeding only)
Water-resistant cap
(MH-553)
Biopsy valve
(MB-358, 10 pcs.)
Channel plug (MH-944)
Forceps Suction Plug (T-Plug)
(MH-405 for GIF-XTQ160 only)
Instruction Manual
(Forceps Suction Plug,
for GIF-XTQ160 only)
10
Operation manual
Instructions (Leaflet type, for
CF-Q160AL/I, PCF-160AL/I only)
Reprocessing manual
Instructions
(Leaflet type, for
GIF-XTQ160 only)
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Chapter 1 Checking the Package Contents
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
11
Chapter 2 Instrument Nomenclature and Specifications
Chapter 2Instrument Nomenclature
and Specifications
2.1Nomenclature
GIF-XP160, GIF-160, GIF-Q160
Universal cord
5. Electrical connector
Air pipe
Light guide
1. Suction connector
2. S-cord connector mount
3. Air supply connector
3. Water supply connector
Product name and serial number
Electrical contacts
4. Endoscope connector
12
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7. UP/DOWN angulation lock
6. UP/DOWN angulation control knob
16. RIGHT/LEFT angulation
control knob
15. RIGHT/LEFT angulation lock
Chapter 2 Instrument Nomenclature and Specifications
9. Air/water valve (MH-438)
8. Suction valve (MH-443)
Control
section
Suction cylinder
Grip section
14. Color code
Boot
Air/water cylinder
Biopsy valve (MB-358)
10. Instrument channel port
11. Insertion tube limit mark
Working length
Distal end
13. Remote switches
Top view
Insertion tube
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
12. Bending section
13
Chapter 2 Instrument Nomenclature and Specifications
GIF-1TQ160, CF-Q160L/I, CF-Q160S
Universal cord
1. Suction connector
5. Electrical connector
Air pipe
Light guide
17. Auxiliary water inlet
Auxiliary water inlet cap
(MAJ-215)
2. S-cord connector mount
3. Air supply connector
3. Water supply connector
Product name and serial number
Electrical contacts
14
4. Endoscope connector
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7. UP/DOWN angulation lock
6. UP/DOWN angulation control knob
16. RIGHT/LEFT angulation
control knob
15. RIGHT/LEFT angulation lock
Chapter 2 Instrument Nomenclature and Specifications
9. Air/water valve (MH-438)
8. Suction valve (MH-443)
Control
section
Suction cylinder
Grip section
14. Color code
Boot
13. Remote switches
Air/water cylinder
Biopsy valve (MB-358)
10. Instrument channel port
11. Insertion tube limit mark
Working length
Distal end
12. Bending section
Top view
Insertion tube
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
15
Chapter 2 Instrument Nomenclature and Specifications
GIF-XTQ160
Universal cord
1. Suction connector
5. Electrical connector
Air pipe
Light guide
17. Auxiliary water inlet
Auxiliary water inlet cap
(MAJ-215)
2. S-cord connector mount
3. Air supply connector
3. Water supply connector
Product name and serial number
Electrical contacts
16
4. Endoscope connector
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7. UP/DOWN angulation lock
6. UP/DOWN angulation control knob
16. RIGHT/LEFT angulation
control knob
15. RIGHT/LEFT angulation lock
Chapter 2 Instrument Nomenclature and Specifications
9. Air/water valve (MH-438)
8. Suction valve (MH-443)
Suction cylinder
Air/water cylinder
Control
section
Grip section
14. Color code
Boot
13. Remote switches
Biopsy valve
(MB-358)
Forceps Suction Plug
(T-Plug) (MH-405)
10. Instrument channel port
11. Insertion tube limit mark
Working length
Distal end
12. Bending section
Top view
Insertion tube
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
17
Chapter 2 Instrument Nomenclature and Specifications
CF-Q160AL/I
Universal cord
1. Suction connector
5. Electrical connector
Air pipe
Light guide
17. Auxiliary water inlet
Auxiliary water inlet cap
(MAJ-215)
2. S-cord connector mount
3. Air supply connector
3. Water supply connector
Product name and serial number
Electrical contacts
18
4. Endoscope connector
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7. UP/DOWN angulation lock
6. UP/DOWN angulation control knob
16. RIGHT/LEFT
angulation control
knob
15. RIGHT/LEFT angulation lock
Chapter 2 Instrument Nomenclature and Specifications
8. Suction valve (MH-443)
9. Air/water valve (MH-438)
Suction cylinder
Control
section
Grip section
14. Color code
Mark
18. Flexibility adjustment ring
Air/water cylinder
Biopsy valve (MB-358)
Mark
Boot
11. Insertion tube limit mark
Working length
10. Instrument channel port
Insertion tube
Top view
13. Remote switches
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Distal end
12. Bending section
19
Chapter 2 Instrument Nomenclature and Specifications
PCF-160AL/I
Universal cord
5. Electrical connector
Air pipe
Light guide
1. Suction connector
2. S-cord connector mount
3. Air supply connector
3. Water supply connector
Product name and serial number
Electrical contacts
4. Endoscope connector
20
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
7. UP/DOWN angulation lock
6. UP/DOWN angulation control knob
16. RIGHT/LEFT
angulation control
knob
15. RIGHT/LEFT angulation lock
Chapter 2 Instrument Nomenclature and Specifications
8. Suction valve (MH-443)
9. Air/water valve (MH-438)
Suction cylinder
Control
section
Grip section
14. Color code
Mark
18. Flexibility adjustment ring
Air/water cylinder
Biopsy valve (MB-358)
Mark
Boot
11. Insertion tube limit mark
Working length
10. Instrument channel port
Insertion tube
Top view
13. Remote switches
12. Bending section
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Distal end
21
Chapter 2 Instrument Nomenclature and Specifications
2.2Endoscope functions
1. Suction connector
This connector connects the endoscope to the suction tube of the suction
pump.
2. S-cord connector mount
This mount connects the endoscope with the Olympus electrosurgical unit
via the S-cord. The S-cord conducts leakage current from the endoscope to
the electrosurgical unit. To connect the S-cord, refer to the instruction
manual for the electrosurgical unit.
3. Water supply connector and air supply connector
These connectors connect the endoscope to the water container via the
water container tube, to supply water to the distal end of the endoscope.
4. Endoscope connector
This connector connects the endoscope to the output socket of the light
source and transmits light from the light source to the endoscope.
5. Electrical connector
This connector connects the endoscope to the video system center via the
videoscope cable. The endoscope contains a memory chip that stores
information about the endoscope and communicates this information to the
video system center CV-160. For more details, refer to the instruction
manual of the CV-160.
6. UP/DOWN angulation control knob
When this knob is turned in the “U” direction, the bending section moves
UP; when the knob is turned in the “D” direction, the bending section
moves DOWN.
7. UP/DOWN angulation lock
Moving this lock in the “F” direction frees angulation. Moving the lock in
the opposite direction locks the bending section at any desired position.
8. Suction valve (MH-443)
This valve is depressed to activate suction. The valve is used to remove any
fluid, debris, flatus or air from the patient.
9. Air/water valve (MH-438)
The hole in this valve is covered to insufflate air and the valve is depressed
to feed water for lens washing. It also can be used to feed air to remove any
fluid or debris adhering to the objective lens.
22
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Chapter 2 Instrument Nomenclature and Specifications
10. Instrument channel port
The instrument channel port functions as:
− channel for the insertion of endo-therapy accessories
− suction channel
− fluid feed channel (from a syringe via the biopsy valve)
11. Insertion tube limit mark
This mark shows the maximum point to which the endoscope may be
inserted into the patient’s body.
12. Bending section
This section moves the distal end of the endoscope when the UP/DOWN
and RIGHT/LEFT angulation control knobs are operated.
13. Remote switches 1 to 4
The functions of the remote switches 1 to 4 can be selected on the video
system center. When selecting the functions, refer to the instruction manual
for the video system center.
14. Color code
This code is used to quickly determine the compatibility of endo-therapy
accessories. The endoscope can be used with endo-therapy accessories
that have the same color code.
• Blue:GIF-XP160
• Yellow:GIF-160, GIF-Q160, PCF-160AL/I
• Orange:GIF-1TQ160, GIF-XTQ160, CF-Q160L/I/S,
CF-Q160AL/I
15. RIGHT/LEFT angulation lock
Turning this lock in the “F” direction frees angulation. Turning the lock in
the opposite direction locks the bending section at any desired position.
16. RIGHT/LEFT angulation control knob
When this knob is turned in the “R” direction, the bending section moves
RIGHT; when the knob is turned in the “L” direction, the bending section
moves LEFT.
17. Auxiliary water inlet (for endoscopes with auxiliary water feeding only)
This inlet is connected to the auxiliary water tube. Feed water from this inlet
through the auxiliary water channel when necessary, (e.g. when blood
adheres to mucosa in the patient’s body cavity). When the auxiliary water
inlet is not being used, make sure that it is covered by the auxiliary water
inlet cap.
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
23
Chapter 2 Instrument Nomenclature and Specifications
18. Flexibility adjustment ring (for CF-Q160AL/I, PCF-160AL/I only)
Turn this ring to adjust the flexibility of the insertion tube.
When the “z” mark on the ring is aligned with the “” mark at the bottom of
the grip section, the insertion tube has the softest condition. To decrease the
flexibility, turn the ring so that the numbers are aligned with the “” mark
(“3” corresponds to the stiffest condition). In the section between “z” and
“3”, insertion tube flexibility can be changed gradually regardless of the
positions of other index markings (“1” and “2”).
2.3Specifications
Operating environment
Operating
environment
Ambient temperature 10 – 40°C (50 – 104°F)
Relative humidity30 – 85%
Atmospheric
pressure
700 – 1060 hPa
(0.7 – 1.1 kgf/cm
(10.2 – 15.4 psia)
2
)
24
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Specifications
Endoscope functions
ModelGIF-XP160
Chapter 2 Instrument Nomenclature and Specifications
Optical
system
Insertion tubeDistal end outer
Field of view120°
Direction of viewForward viewing
Depth of field3 – 100 mm
diameter
Distal end enlarged1. Air/water nozzle
Insertion tube outer
diameter
Working length1030 mm
ø5.9mm
2. Light guide lens
3. Objective lens
4. Instrument channel outlet
UP
3.
RIGHT
2.
DOWN
ø5.9mm
1.
LEFT
4.
Instrument
channel
Air flow rate
Bending
section
Total length1345 mm
Channel inner
diameter
Minimum visible
distance
Direction from which
endo-therapy
accessories enter
and exit the
endoscopic image
Note: Standard when CLV-160 (high air
pressure) is used.
Angulation rangeUP 180°, DOWN 90°,
3 mm from the distal end
RIGHT 100°, LEFT 100°
ø2mm
25 cm
3
/s
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
25
Chapter 2 Instrument Nomenclature and Specifications
ModelGIF-160GIF-Q160
Optical
system
Insertion tubeDistal end outer
Field of view140°
Direction of viewForward viewing
Depth of field3 – 100 mm
diameter
Distal end enlarged1. Air/water nozzle
ø8.6mmø9.8mm
2. Light guide lens
3. Objective lens
4. Instrument channel outlet
UP
1.
2.
Instrument
channel
Air flow rate
RIGHT
3.
DOWN
Insertion tube outer
diameter
Working length1030 mm
Channel inner
diameter
Minimum visible
distance
Direction from which
endo-therapy
accessories enter
and exit the
endoscopic image
ø8.6mmø9.5mm
ø2.8mm
3 mm from the distal end
3
25 cm
/s
Note: Standard when CLV-160 (high air
pressure) is used.
LEFT
4.
26
Bending
section
Total length1345 mm
Angulation rangeUP 210°, DOWN 90°,
RIGHT 100°, LEFT 100°
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Loading...
+ 72 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.