EN-1Valor 2000 Series Scales
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the
Valor 2000 Series Scales. Please read the manual completely before using the
scale.
TM
1.1 Safety Precautions
Please follow these safety precautions:
• Verify that the AC Adapter input voltage matches the local AC power supply.
• Do not immerse the scale in water or other liquids.
• Do not operate the scale i
n hostile environments.
• Do not drop loads on the platform.
• Do not place the scale upside down on the pan.
• Service should only be performed by authorized personnel.
• Disconnect the scale from the power suppl
y when cleaning.
1.2 Controls
Figure 1-1. Controls.
Figure 1-2. Rear Display.
EN-2 Valor 2000 Series Scales
TABLE 1-1. CONTROL FUNCTIONS.
Button
Primary Function
(Short Press)
ON/ 0/T
Turns scale ON.
If scale is On:
Sets zero.
Performs a tare operation.
UNITS
Changes the weighing Unit.
Secondary
Function
(Long Press)
Off
Turns scale off.
Menu
Enter the User menu.
Menu Function
(Short Press)
Yes
Accepts the current setting on
the dis
play.
No
Advances to the next menu or
menu item.
Rejects the current setting on
the dis
play and advances to
the next available setting.
2. INSTALLATION
2.1 Package Contents
Scale
Pan
Power Adapter
Instruction Manual
Warranty Card
2.2 Location
Use the scale on a firm, steady surface. Avoid locations with excessive air current,
vibrations, heat sources, or rapid temperature changes.
2.3 Leveling the Scale
Adjust the leveling feet so the bubble is centered in the circle.
EN-3Valor 2000 Series Scales
Level Bubble
Adjustable Feet
Figure 2-1. Leveling the Scale.
IMPORTANT: To relieve any internal pressure and allow proper airflow
inside the scale, unscrew the two rear leveling feet slightly, then level
the scale accordingly. When cleaning the scale, the rear leveling feet
must be screwed in tightly to avoid entry of water inside the scale.
2.4 Power
2.4.1 AC Power
The AC adapter (included) may be used to power the scale when battery power is
not available. Remove the rubber seal in the back. Connect the AC adapter plug to
the input jack. Then plug the AC adapter into a properly grounded power outlet.
NOTE: To prevent water ingress during use, insert rubber Seal into
the input jack recess.
Seal
Figure 2-2. Power Connection and Sealing
.
EN-4 Valor 2000 Series Scales
2.4.2 Battery Power
The scale can be operated on the internal rechargeable battery when AC power is not
available. The scale will automatically switch to battery operation if there is a power
failure or the power cord is removed.
Before using the scale for the first time, the battery should be fully
char
ged for up to 12 hours. During charging, the battery Indicator
displays the charge level (see Table 2-1), and the scale can be operated
during charging. The battery is protected against over charging and the
scale can remain connected to the AC power line.
TABLE 2-1. BATTERY CHARGE STATUS.
Charge Status Indication Light Charge level
Off Fully Charged
Flashing Green 70 – 100 %
Flashing Yellow 40 – 70 %
Flashing Red 0 – 40 %
The scale can operate for approximately 60 hours on a fully charged battery.
During battery operation, the battery Indicator displays the battery status (see Table
2-2).
TABLE 2-2. BATTERY CAPACITY STATUS.
Charge Status Indication Light Capacity Remaining
Solid Green 70 – 100 %
Solid Yellow 40 – 70 %
Solid Red 10 – 40 %
Flashing Red 0 – 10 %
Flashing Red, Lo bAt Requires recharge;
Accuracy not guaranteed
NOTE:
• Charging the scale must only be performed in a dry environment.
• Install the rubber seal securely to ensure the scale is prote
cted when
operating in a wet environment.
CAUTION: Batt
ery is to be replaced only by an authorized Ohaus service
dealer.
Risk of explosion can occur if replaced with the wrong type or connected
improperly.
Dispose of the lead acid battery according to local laws and regulations.
EN-5Valor 2000 Series Scales
2.5 Initial Calibration
When the scale is operated for the first time, a Span calibration is recommended to
ensure accurate weighing results. Before performing the calibration, be sure to have
the appropriate calibration weight.
Refer to section 4.2 for Span calibration procedures.
3. OPERATION
3.1 Turning Scale On/Off
To turn the scale on, press the ON/ 0/T Off button. The scale performs a display test,
momentarily displays the model/software version, and then enters the active
weighing mode.
To turn the scale off, press and hold ON/ 0/T Off button until OFF is displayed.
3.2 Zero Operation
Press the ON/ 0/T Off button to zero the weight display. The scale must be stable to
accept zero operation.
3.3 Manual Tare
Place a container on the weighing pan, then press ON/ 0/T Off. The display will
show a net value of “0”. To clear the Tare value, press
ON/ 0/T Off with the pan
empty.
3.4 Changing Units of Measure
Press the UNITS/Menu button to display the next available measuring unit. Only
measuring units enabled in the Unit Menu will be displayed (refer to Section 4.4).
4. SETTINGS
The Menu allows the user to customize scale settings.
4.1 Menu Navigation
Press and hold UNITS/Menu button until MENU is displayed, then release it.
Press Yes to enter the displayed menu or press No to advance to the next menu.
When viewing the menu setting, press Yes to accept the setting, or press No to
change it, then
press
Yes to accept the setting and move to the next menu. When
End is dis
played, press
Yes to exit the menu.
EN-6 Valor 2000 Series Scales
4.2 CAL Menu
SPAN Perform Span calibration Yes or No.
LIN Perform Linearity calibration Yes or No.
NOTE:
• Ensure the appropriate calibration masses are available before beginning
calibration (see Table 4-1).
• Ensure that the scale is level and stable during the e
ntire calibration
process.
• Calibration is unavailable with LFT set to On.
• Allow the scale to warm up for approxim
ately 5 minutes after stabilizing
to room temperature.
• To abort calibration, power off the scale.
TABLE 4-1. CALIBRATION WEIGHTS.
Model Span Calibration weight (kg)
User selectable
Linearity Calibration weight (kg)
Fixed
V21PW3 1, 2, 3 2, 3
V21PW6 2, 4, 6 4, 6
V21PW15 5, 10, 15 10, 15
4.2.1 Span Calibration
Span Calibration uses two points to adjust the scale. The first point is the zero value
where there is no weight on the scale. The second point is the Span value where a
calibration mass is placed on the scale.
When SPAN is displayed, press the Yes
button to access
the Span Calibration menu item.
The display flashes 0.
With no weight on the scale, press the Yes b
utton to
establish the zero point.
The display shows --C-- while the zero point is
establis
hed.
EN-7Valor 2000 Series Scales
The display flashes the span calibr
ation point. Place the
specified weight on the scale and press the Yes button.
To choose a different span poi
nt, repeatedly press the No
button to increment the selections. When the desired
value is displayed, place the specified weight on the scale
and press the Yes button.
The display shows --C-- while
the span point is
established.
If span calibration was successful, dONE is displayed
an
d the scale exits to the active weighing mode and
displays the actual weight value.
4.2.2 Linearity Calibration
Linearity calibration uses 3 calibration points. The first calibration point is
established with no weight on the scale. The second calibration point is established
at approximately half capacity. The third calibration point is established at capacity.
The Linearity calibration points are fixed and cannot be altered by the user during the
calibration procedure.
When LIN is displayed, press the Yes button to access the
Linearity Calibration menu item.
The display flashes 0.
With no weight on the scale, press the Yes b
utton to
establish the zero point.
The display shows --C-- while the zero point is
establis
hed.
The display flashes the mid calibr
ation point. Place the
specified weight on the scale and press the Yes button.
The display shows --C-- while the mid point is
established.
EN-8 Valor 2000 Series Scales
The display flashes the full c
alibration point. Place the
specified weight on the scale and press the Yes button.
The display shows --C-- while the full calibration point is
established.
If linearity calibration was successful, dONE is displayed
an
d the scale exits to the active weighing mode and
displays the actual weight value.
4.3 RES Menu
RESOLUTION
Set the resolution to LO (low) or HI (high).
4.4 UNIT Menu
KILOGRAM UNIT
Set kilograms ON or OFF. When kg is lit, press the Yes
butt
on to access the
ON or OFF setting, then press No to
choose ON or
OFF, then
Yes to accept the setting.
GRAM UNIT
Set grams ON or OFF. When g is lit, press the Yes b
utton
to access the
ON or OFF setting, then press No to choose
ON or OFF, then
Yes to accept the setting.
POUND UNIT
Set pounds ON or OFF. When lb is lit, press the Ye
s
button to access the
ON or OFF setting, then press No to
choose ON or
OFF, then
Yes to accept the setting.
OUNCE UNIT
Set ounces ON or OFF. When oz is lit, press t
he Yes
button to access the ON or OFF setting, then press No to
choose ON or OFF, then Yes to accept the setting.
EN-9Valor 2000 Series Scales
4.5 A OFF Menu
AUTO SHUT OFF
Set the automatic shut off to 5, 10, 15 minutes (shut off
after 5 minutes
, etc.) or OFF (scale stays on).
4.6 FILTR Menu
FILTERING LEVEL
Set the filter level to LO (low), MEd (medium), or HI
(hi
gh).
4.7 AZT Menu
AUTO ZERO TRACKING RANGE
Set the automatic zero tracking range to OFF, 0.5d, 1d,
3d (
divisions).
4.8 LFT Menu
LEGAL FOR TRADE (LFT) MODE
Set the legal for trade status to ON or OFF.
4.9 END
End menu item.
5. LEGAL FOR TRADE
5.1 Settings
Enter the menu to verify the settings and perform a calibration as explained in
Section 4.2. Set the LFT menu to ON.
NOTE: Whe
n LFT is set to ON, the following menu settings cannot be
accessed/changed: Calibration, Resolution, Unit, AZT, LFT.
To regain access to the locked menu settings, remove the small cover on
the bottom of t
he scale. Then during power up press and hold the
recessed lock switch momentarily until zero is displayed.
EN-10 Valor 2000 Series Scales
Figure 5-1. Lock Switch.
Lock Switch
When accessing the bottom of the scale, avoid placing the scale upside
down on the pan or sub-platform. Place the scale on its side.
5.2 Verification
Before this product can be used in a trade approved application, it must be inspected
in accordance with local weights and measures regulations. It is the responsibility
of the purchaser to ensure that all pertinent legal requirements are met. Please
contact your local weights and measures office for further details.
5.3 Sealing
The weights and measures official can apply a wire or paper security seal as shown
below.
SEAL
SEAL
Figure 5-2. Sealing.
6. MAINTENANCE
CAUTION: to avoid possible water entry, before cleaning the scale:
Turn off the scale, remove the AC
adapter and install the rubber seal securely.
Screw in the two rear leveling feet tightly. W
hen the scale is put back into
operation, remember to loosen the two rear feet to allow proper airflow inside the
scale.
EN-11Valor 2000 Series Scales
6.1 Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if
necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean
the h
ousing or control panels.
6.2 Troubleshooting
The following table lists common problems and possible causes and remedies. If
the problem persists, contact Ohaus or your authorized dealer.
TABLE 6-1. TROU
BLESHOOTING.
Symptom Possible Cause(s) Remedy
Scale will not turn on No power to scale
Battery power used up
Verify connections and voltage.
Connect power and charge the battery
Poor accuracy Improper calibration • Unstable
environment • Internal pressure
build-up
Perform calibration • Move scale to
suitable location • Relieve internal
pressure by loosening rear leveling feet
Unable to calibrate LFT Menu set to ON
Unstable environment
Incorrect calibration mass
Set LFT menu to OFF (see section 5.1)
Move the scale to suitable location
Use correct calibration mass
Cannot display weight in
desired weighing unit
Weighing unit not enabled Enter menu and enable unit
Cannot change menu
settings
LFT Menu set to ON Set LFT menu to OFF (see section 5.1)
Scale displays Lo bAt Battery is discharged Connect power and charge the battery
Battery fails to charge
fully
Battery is defective Have battery replaced by authorized
service dealer
Err 1 Weight on pan exceeds capacity Remove weight from the pan
Err 2 Underload limit exceeded • Pan
was removed prior to power on
Install pan and re-zero
Err 9, Err 21 Factory data corrupted Contact Ohaus or an authorized service
agent
EN-12 Valor 2000 Series Scales
6.3 Service Information
If the troubleshooting section does not resolve or describe your problem, contact
your authorized Ohaus service agent. For service assistance or technical support in
the United States call toll-free 1-800-526-0659 between 8.00 AM and 5:00 PM
EST. An Ohaus product service specialist will be available to provide assistance.
Outside the USA, please visit our web site,
www.ohaus.com to locate the Ohaus
office nearest you.
7. TECHNICAL DATA
The technical data is valid under the following ambient conditions:
Ambient temperature: 0°C to 40°C
Relative humidity: 10% to 95% relative humidity, non-condensing
Height above sea level: Up to 2000 m
Operability is assured at ambient temper
atures between 0°C and 40°C
7.1 Specifications
TABLE 7-1. SPECIFICATIONS.
Model V21PW3 V21PW6 V21PW15
Capacity x Readability
(max x d non-approved)
3 kg x 0.0005 kg
3000 g x 0.5 g
6 lb x 0.001 lb
96 oz x 0.02 oz
6 kg x 0.001 kg
6000 g x 1 g
15 lb x 0.002 lb
240 oz x 0.05 oz
15 kg x 0.002 kg
15000 g x 2 g
30 lb x 0.005 lb
480 oz x 0.1 oz
Maximum displayed resolution 1:6000 1:6000 1:7500
Capacity x Readability
(max x e approved)
3 kg x 0.001 kg
3000 g x 1 g
6 lb x 0.002 lb
96 oz x 0.05 oz
6 kg x 0.002 kg
6000 g x 2 g
15 lb x 0.005 lb
240 oz x 0.1 oz
15 kg x 0.005 kg
15000 g x 5 g
30 lb x 0.01 lb
480 oz x 0.2 oz
Approved resolution 1:3000 1:3000 1:3000
Repeatability +0.001 kg +0.002 kg +0.005 kg
Linearity +0.001 kg +0.002 kg +0.005 kg
Weighing Units kg, g, lb, oz
Tare range To capacity by subtraction
Stabilization Time
2 seconds
Power Requirements
9-12 VDC 0.7A AC Adapter (supplied)
Internal rechargeable sealed lead-acid battery
Display
2x red LED (front and rear)
6-digit 7-segment, 15mm / 0.6 in characters
Keypad 2-button mechanical keys
Ingress Protection IP65
Pan Size (mm / in) 233 x 184 / 9.2 x 7.2
Typical Battery life 60 hours with 12 hour recharge time
Operating temperature range 0°C to 40°C
Net Weight (kg / lb) 3.5 / 7.8
Shipping Weight (kg / lb) 4.3/ 9.3
EN-13Valor 2000 Series Scales
7.2. Drawings and Dimensions
249 mm
(9.8 in.)
229 mm
(9.0 in.)
233 mm
(9.2 in.)
184 mm
(7.2 in.)
142 mm
(5.6 in.)
Figure 7-1. Valor 2000 Series Overall Dimensions.
EN-14 Valor 2000 Series Scales
7.3. Compliance
Compliance to the following standards is indicated by the corresponding mark on the product
.
Mark Standard
This product conforms to the EMC directive 2
004/108/EC and the
Low Voltage Directive 2006/95/EC. The complete declaration of
Conformity is available from Ohaus Corporation
AS/NZS4251.1 Emission; AS/NZS4252.1 Immunity
NSF/ANSI 169 – 2005
NSF/ANSI /3-A 14159-1 - 2002
Important notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture
bear one of the preceding mark on the packing label and
the green ‘M’ (metrology) sticker on the descriptive plate.
They may be put into service immediately.
Weighing Instruments to be verified in two stages have no
green ‘M’ (metrol
ogy) on the descriptive plate and bear
one of the preceding identification mark on the packing
label. The second stage of the initial verification must be
carried out by the approved service organization of the
authorized representative within the EC or by the national
weight & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturers work. It
comprises all tests according to the adopted European standard EN 45501:1992,
paragraph 8.2.2.
If national regulations limit the validity period of the verification, the
user of the weighing
instrument must strictly observe the re-verification period and inform the respective W+M
authorities
.
EN-15Valor 2000 Series Scales
Disposal
In conformance with the European Directive 2002/96 EC on Waste Electrical
and Electronic Equipment (WEEE) this device may not be disposed of in
domestic waste. This also applies to countries outside the EU, per their
specific requirements.
Please dispose of this product in accordance with local regulations at the
collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the
distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional
use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
FCC Note
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit d
ifferent from that to which the receiver
is connected.
• Consult the dealer or an experienced radio/TV technician for help.
This Class B digital apparatus complies with
Canadian ICES-003.
Cet appareil numérique de la classe B est conforme à la norme NMB-00
3 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau
Veritus Quality International (BVQI), confirming that the Ohaus quality management system is
compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation,
USA, was re-registered to the ISO 9001:2000 standard.
EN-16 Valor 2000 Series Scales
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from
the d
ate of delivery through the duration of the warranty period. During the warranty
period Ohaus will repair, or, at its option, replace any component(s) that proves to
be defective at no charge, provided that the product is returned, freight prepaid, to
Ohaus. This warranty does not apply if the product has been damaged by accident
or misuse, exposed to radioactive or corrosive materials, has foreign material
penetrating to the inside of the product, or as a result of service or modification by
other than Ohaus. In lieu of a properly returned warranty registration card, the
warranty period shall begin on the date of shipment to the authorized dealer. No
other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation
shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please
cont
act Ohaus or your local Ohaus dealer for further details.
ES-1Básculas serie Valor 2000
1. INTRODUCCIÓN
Este manual contiene instrucciones de instalación, operación y mantenimiento para
las básculas de la serie Valor 2000. Por favor lea completamente el manual antes
de usar la báscula.
TM
1.1 Precauciones de seguridad
Por favor siga estas precauciones de seguridad:
• Verifique que el voltaje de entrada del adaptador de
corriente alterna (CA) coincida con el
voltaje del suministro de eléctrico CA local.
• No sumerja la báscula en agua u otros líquidos.
• No opere la báscula en ambientes hostiles.
• No deje caer objetos en la plataforma.
• No coloque la báscula al revés sobre el plato de pesaje.
• El mantenimiento debe ser reali
zado solamente por personal autorizado.
• Desconecte la báscula de la fuente de alimentación cuando la esté limpiando.
1.2 Controles
Figura 1-1. Controles.
Figura 1-2. Pantalla posterior.
ES-2 Básculas serie Valor 2000
TABLA 1-1. FUNCIONES DE CONTROL.
Botón
Función primaria
(Pulsación corta)
ON/ 0/T
Enciende la báscula.
Si la báscula está encendida:
Establece cero.
Realiza una operación de
tara.
UNITS (UNIDADES)
Cambia la unidad de peso.
Función
secundaria
(Pulsación larga)
Off
Apaga la báscula.
Menu
Accede al menú de usuario.
Función del
menú:
(Pulsación corta)
Yes
Acepta la configuración que
está act
ualmente en la
pantalla.
No
Avanza al siguiente menú o
ítem del menú.
Rechaza la configuración que
está act
ualmente en la
pantalla y avanza a la
siguiente configuración
disponible.
2. INSTALACIÓN
2.1 Contenido del paquete
Báscula
Plato de pesaje
Adaptador de corriente
Manual de instrucciones
• Tarjeta de garantía
2.2 Ubicación
Use la báscula en una superficie firme y segura. Evite lugares con corrientes de aire
excesivas, vibraciones, fuentes de calor o cambios bruscos de temperatura.
2.3 Nivelación de la báscula
Ajuste las patas niveladoras para que la burbuja esté centrada en el círculo.
ES-3Básculas serie Valor 2000
Burbuja de nivelación
Patas ajustables
Figura 2-1. Nivelación de la báscula.
IMPORTANTE: Para liberar cualquier presión interna y permitir el flujo de
aire adecuado dentro de la báscula, destornille ligeramente las dos
patas de nivelación posteriores y nivele la báscula según sea necesario.
Cuando limpie la báscula, las patas de nivelación posteriores deben
atornillarse con fuerza para evitar que ingrese agua en la báscula.
2.4 Alimentación
2.4.1 Corriente CA
El adaptador de corriente (incluido) puede usarse para alimentar la báscula cuando
la energía de la batería no esté disponible. Retire el sello de goma de la parte
trasera.
Conecte el adaptador de corriente alterna al enchufe de entrada. Después
co
necte el adaptador de corriente a una salida de energía con conexión a tierra.
NOTA: Para evitar la entrada de agua durante su uso, coloque un
sello de g
oma en la ranura del enchufe de entrada.
Sello
Figura 2-2. Conexión de corriente y sello.
ES-4 Básculas serie Valor 2000
2.4.2 Alimentación por corriente de batería
La báscula puede ser usada con la batería interna recargable cuando la energía de la
corriente CA no esté disponible. La báscula cambiará automáticamente a la operación por
batería su hay una falla de corriente o si el cable de corriente es removido.
Antes de usar la báscula por primera vez, la batería debe ser cargada
completamente hasta por 12 horas. Durante la carga, el indicador de batería
muestra
el nivel de carga (consulte la tabla 2-1). La báscula puede ser usada
mientras se está cargando.
La batería está protegida contra sobre cargas y la
báscula pued
e permanecer conectada al cable de corriente CA.
TABLA 2-1. ESTADO DE CARGA DE LA BATERÍA.
Luz indicadora de estado de
carga
Nivel de carga
Off Cargada
completamente
Verde parpadeando 70 – 100 %
Amarillo parpadeando 40 – 70 %
Rojo parpadeando 0 – 40 %
La báscula puede operar durante aproximadamente 60 horas con la batería
completamente cargada.
Durante el funcionamiento con batería, el indicador muestra el estado de la batería
(con
sulte la tabla 2-2).
TABLA 2-2. ESTADO DE CAPACIDAD DE LA BATERÍA.
Luz indicadora de estado de
carga
Capacidad restante
Verde fijo 70 – 100 %
Amarillo fijo 40 – 70 %
Rojo fijo 10 – 40 %
Rojo parpadeando 0 – 10 %
Rojo parpadeando, Lo bAt Requiere carga; Exactitud
no garantizada
NOTA:
• La carga de la báscula solo debe realizarse en un ambiente seco.
• Instale el sello de goma correctamente para
asegurar que la báscula esté protegida cuando
se use en un ambiente húmedo.
PRECAUCIÓN: La batería solo debe ser reemplazada por un distribuidor de Ohaus
autorizado.
Puede producirse un riesgo de explosión si se reemplaza con un tipo equivocado
o si se conecta de forma inapropiada.
La eliminación de la batería de ácido conductor debe hacerse de acuerdo a las
leyes y regulaciones locales.
ES-5Básculas serie Valor 2000
2.5 Calibración inicial
Cuando la báscula es utilizada por primera vez, se recomienda hacer una calibración de
rango para asegurar resultados de pesaje exactos. Antes de realizar la calibración, asegúrese
de tener el peso de calibración apropiado.
Diríjase a la sección 4.2 para los procedimientos de calibración de rango.
3. OPERACIÓN
3.1 Apagado y encendido de la báscula
Para encender la báscula, presione el botón ON/ 0/T Off. La báscula realiza una
prueba de pantalla, muestra momentáneamente el modelo y la versión del software,
después entra al modo de pesaje activo.
Para apagar la báscula, presione el botón ON/ 0/T Off hasta que se muestre OFF.
3.2 Operación de puesta a cero
Presione el botón ON/ 0/T Off para establecer cero en la pantalla de peso. La
báscula debe estar estable para aceptar la operación de cero.
3.3 Tara manual
Coloque un recipiente en la bandeja de pesaje, después presione ON/ 0/T Off. La
pantalla mostrará un valor neto de “0”. Para borrar el valor de tara, presione ON/
0/T Off con la bandeja vacía.
3.4 Cambio de unidades de medida
Presione el botón UNITS/Menu para mostrar la siguiente unidad de medida
disponible. Solo se mostrarán las unidades habilitadas en el menú de unidades
(consulte l
a sección 4.4).
4. CONFIGURACIONES
El menú permite al usuario personalizar las configuraciones de la báscula.
4.1 Menú de navegación
Mantenga presionado el botón UNITS/Menu hasta que se muestre MENU, después suéltelo.
Presione Ye
s (Sí) para entrar al menú mostrado o presione No para avanzar al siguiente
menú.
Cuando consulte la configuración del menú, presione Yes para aceptar la
configuración, o presione No para cambiarla, después presione Yes para aceptar la
configuración y moverse al siguiente menú. Cuando se muestre End, presione Yes
para salir del menú.
ES-6 Básculas serie Valor 2000
4.2 Menú de calibración (CAL Menu)
SPAN Realizar calibración de rango Si o No (Yes o No).
SPAN Realizar calibración de linealidad Si o No (Yes o
No).
NOTA:
• Asegúrese de tener disponibles las masas
de calibración correctas antes de
empezar
la calibración (cons
ulte la tabla 4-1).
• Asegúrese de que la báscula esté nivelada y estable durante todo el pr
oceso de
calibración.
• La calibración no está disponible cuando LFT está activado (On).
• Permita que la báscula se caliente durante aproximadamente 5 minutos después
de estabilizarse
a la temperatura ambiente.
• Para cancelar la calibración, apague la báscula.
TABLA 4-1. PESOS DE CALIBRACIÓN.
Modelo Peso de calibración de rango
(kg)
Seleccionable por el usuario
Peso de calibración de linealidad
(kg)
Fijo
V21PW3 1, 2, 3 2, 3
V21PW6 2, 4, 6 4, 6
V21PW15 5, 10, 15 10, 15
4.2.1 Calibración de rango
La calibración de rango usa dos puntos para ajustar la báscula. El primer punto es
el valor cero donde no hay peso en la báscula. El segundo punto es el valor de
r
ango donde se coloca una masa de calibración en la báscula.
Cuando no se muestre SPAN, presione el botón Yes para
acceder al ítem del menú de c
alibración de rango.
La pantalla destella 0.
Presione el botón Yes si
n peso en la báscula para
establecer el punto cero.
La pantalla muestra --C-- mientras se establece el punto
cero.
ES-7Básculas serie Valor 2000
En la pantalla destella el p
unto de calibración de rango.
Coloque el peso especificado en la báscula y luego
pr
esione el botón Yes.
Para elegir un punto de rango diferente, presione el botón
No varias veces para incrementar las selecciones.
Cuando se muestre el valor deseado, coloque el peso
especificado en la báscula y luego presione el botón Yes.
La pantalla muestra --C-- mientras se establece el punto
de rango.
Si la calibración de rango fue exitosa, aparece dONE y la
báscula sale al modo de pesaje activo y muestra el valor
del peso actual.
4.2.2 Calibración de linealidad
La calibración de linealidad usa 3 puntos de calibración. El primer punto de
calibración se establece sin peso en la báscula. El segundo punto de calibración se
establece con aproximadamente la mitad de la capacidad. El tercer punto de
calibración se establece con toda la capacidad.
Los puntos de calibración de
lineali
dad son fijos y no pueden ser modificados por el usuario durante el
procedimiento de calibración.
Cuando se muestre LIN, presione el botón Yes para
acceder al ítem del menú de calibración de linealidad.
La pantalla destella 0.
Presione el botón Yes sin peso en la báscula para
establecer el punto cero.
La pantalla muestra --C-- mientras se establece el punto
cero.
La pantalla destella el punto medio de calibración.
Coloque el peso especificado en la báscula y presione el
botón Yes
.
La pantalla muestra --C-- mientras se establece el punto
medio.
ES-8 Básculas serie Valor 2000
La pantalla destella el punto máximo de calibración. Coloque el
peso especificado en la báscula y presione el botón Yes.
La pantalla muestra --C-- mientras se establece el punto
máximo.
Si la calibración de linealidad fué exitosa, aparece dONE y la
báscula sale al
modo de pesaje activo y muestra el valor
del peso actual.
4.3 Menú de resolución (RES Menu)
RESOLUCIÓN
Establece la resolución en LO (baja) o HI (alta)
4.4 Menú de unidades (UNIT Menu)
UNIDAD DE KILOGRAMO
Establece los kilogramos en ON (encendido) u OFF (apagado).
Cuando kg está encendido, presione el botón Yes para acceder
a la configuració
n ON u OFF, después presione No para elegir
ON u OFF, finalmente presione
Yes para aceptar la
configuración.
UNIDAD DE GRAMO
Establece los gramos en ON (encendido) u OFF (apagado).
Cuando g está encendido, presione el botón Yes para ingresar a
la configuración
ON u OFF, después presione No para elegir ON
u OFF, finalmente presione
Yes para aceptar la configuración.
UNIDAD DE LIBRA
Establece las libras en ON (encendido) u OFF (apagado).
Cuando lb está encendido, presione el botón Yes para ingresar
a la configuració
n ON u OFF, después presione No para elegir
ON u OFF, finalmente presione Yes para aceptar la
configuración.
UNIDAD DE ONZA
Establece las onzas en ON (encendido) u OFF (apagado).
Cuando oz está encendido, presione el botón
Yes para
ingresar a la configuración ON u OFF, después presione No
para elegir ON u OFF, finalmente presione Yes para aceptar la
configuración.
Loading...
+ 58 hidden pages