Ohaus 3000 User Manual

0 (0)
Defender™ 3000 Series Base
Instruction Manual
Compliance
Disposal
I
n conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
Defender 3000 Series Base
EN-1
INTRODUCTION
This manual covers installation, and maintenance instructions for the Ohaus Defender 3000 Series Base. Please read this manual completely before installation and operation.
TM
SAFETY PRECAUTIONS
For safe and dependable operation of this product, please comply with the following precautions:
Operate the base only under ambient conditions specified in these instructions
Ensure that the load cell cable does not pose an obstruction or tripping hazard
Do not operate the base in hazardous environments or unstable locations
Do not drop loads on the base
Do not lift the base by the top frame; always lift from the bottom frame when moving the base
Service should only be performed by authorized personnel
INSTALLATION
Unpacking
Unpack and inspect the product to make sure that all components have been included. The package includes the following:
Base Weighing Pan Warranty Card Instruction Manual
When purchased as a complete Defender 3000 Series Scale, will also include:
3000 Series Indicator Column Assembly
Assembly
Weighing Pan
Place the weighing pan securely over the rubber load pads on the top frame of the base.
Wiring Connections
Connect the load cell cable to an indicator using the wiring codes in Table 1 below. Refer to the indicator manual for information about load cell connections, setup and calibration.
Note: When purchased as a Defender Series Scale, the base is already pre-wired to the 3000 Series Indicator.
TABLE 1. SIX WIRE LOAD CELL CONNECTION.
FUNCTION WIRE COLOR
+ Excitation Green
- Excitation Black + Signal Red
- Signal White + Sense Blue
- Sense Brown Shield Yellow
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat.
Level the base by adjusting the four leveling feet until the bubble in the level indicator (located at the rear of the base) is centered. A wrench may be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts up against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
CORRECT INCORRECT
Loading...
+ 5 hidden pages