Mindray Z6, Z6T, Z6Vet, Z6S, Z5 Service Manual

...
Z6/Z6T/Z6S/Z6W/Z6Vet
Z5/Z5BW/Z5T/Z5Vet
Diagnostic Ultrasound System
Service Manual
Revision 15.0

Table of Content

Table of Content................................................................................................................... i
Revision History .................................................................................................................. I
Intellectual Property Statement ......................................................................................... II
Applicable for...................................................................................................................... II
Statement ............................................................................................................................ II
Responsibility on the Manufacturer Party ........................................................................ II
Customer Service Department .......................................................................................... III
1 Preface....................................................................................................................... 1-1
1.1 Meaning of Signal Words ................................................................................................. 1-1
1.2 Meaning of Symbols ........................................................................................................ 1-1
1.2.1 Meaning of Safety Symbols ...................................................................................... 1-1
1.2.2 Warning Labels ........................................................................................................ 1-2
1.2.3 General Symbols ................................................................................................ ..... 1-2
1.3 Safety Precautions .......................................................................................................... 1-3
1.3.1 Electric safety .......................................................................................................... 1-4
1.3.2 Mechanical Safety .................................................................................................... 1-4
1.3.3 Personnel Safety ...................................................................................................... 1-5
1.3.4 Other ....................................................................................................................... 1-5
2 Product Specifications ............................................................................................. 2-1
2.1 Overview ......................................................................................................................... 2-1
2.1.1 Intended Use ................................ ................................................................ ........... 2-1
2.1.2 Introduction of Each Unit .......................................................................................... 2-1
2.1.3 Peripherals Supported ................................ ............................................................. 2-7
2.2 Specifications .................................................................................................................. 2-8
2.2.1 Dimensions and Weight ........................................................................................... 2-8
2.2.2 Electrical Specifications ............................................................................................ 2-8
2.2.3 Environmental Conditions......................................................................................... 2-8
2.2.4 Monitor Specification ................................................................................................ 2-8
3 System Installation ................................................................................................... 3-1
3.1 Preparations for Installation ............................................................................................. 3-1
3.1.1 Electrical Requirements ........................................................................................... 3-1
3.1.2 Installation Condition ................................................................................................ 3-2
3.2 Unpacking ....................................................................................................................... 3-2
3.2.1 Unpacking ................................................................................................................ 3-3
3.2.2 Checking.................................................................................................................. 3-5
3.3 Installation of Main Unit ................................................................................................... 3-5
3.3.1 Installing Battery ...................................................................................................... 3-5
3.3.2 Control Panel Adjusting ............................................................................................ 3-6
3.3.3 Display Adjusting ...................................................................................................... 3-6
3.3.4 Installing Probe Holder ............................................................................................. 3-7
3.3.5 Connecting a Probe ................................................................................................. 3-7
3.4 Installing Peripherals ....................................................................................................... 3-8
3.4.2 Video Printer Installation........................................................................................... 3-8
3.4.3 Installing a Graph / Text Printer ................................................................................ 3-9
3.4.4 Installing External DVD-R/W .................................................................................... 3-9
3.4.5 Installing Barcode Scanner ..................................................................................... 3-10
3.5 System Configuration ..................................................................................................... 3-11
3.5.1 Power ON / OFF ..................................................................................................... 3-11
3.5.2 Enter Doppler .......................................................................................................... 3-11
3.5.3 System Preset ....................................................................................................... 3-12
3.5.4 Print Preset ............................................................................................................ 3-14
3.5.5 Network Preset ...................................................................................................... 3-16
3.5.6 System Information ................................................................................................ 3-18
4 Hardware Principle ................................ ................................ ................................... 4-1
4.1 General Structure of Hardware System ............................................................................ 4-1
4.2 Main Unit ......................................................................................................................... 4-2
4.2.1 Probe Board............................................................................................................. 4-3
4.2.2 Main board ............................................................................................................... 4-4
4.2.3 IO Broad .................................................................................................................. 4-7
4.2.4 Ultrasound System Monitor ...................................................................................... 4-8
4.2.5 Ultrasound System Indicator .................................................................................... 4-9
4.2.6 Display ................................................................................................................... 4-10
4.2.7 Control Panel .......................................................................................................... 4-11
4.3 Power System ............................................................................................................... 4-12
4.3.1 Power Output of the Power Supply module and Supporting Function Distribution ... 4-12
4.3.2 System Power-on Control ...................................................................................... 4-13
5 Function and Performance Checking Method ........................................................ 5-1
5.1 Instruction ................................ ................................................................ ....................... 5-1
5.2 Checking System Status .................................................................................................. 5-1
5.2.1 System Running Status ................................ ............................................................ 5-1
5.2.2 System Running Status ................................ ............................................................ 5-1
5.3 General exam.................................................................................................................. 5-2
5.3.1 Check Flow .............................................................................................................. 5-2
5.3.2 Checking Content ................................ ................................................................ .... 5-2
5.4 Function Checks.............................................................................................................. 5-4
5.4.1 Check Flow .............................................................................................................. 5-4
5.4.2 Checking Content ................................ ................................................................ .... 5-5
5.5 Performance Test ........................................................................................................... 5-11
5.5.1 Test Process ........................................................................................................... 5-11
5.5.2 Test Content............................................................................................................ 5-11
6 Software Upgrade and Maintenance ........................................................................ 6-1
6.1 Enter the Maintenance Window.......................................................................................... 6-1
6.2 System Software Installation/ Restoration ........................................................................ 6-2
6.2.1 Restore the Operating System and Doppler ............................................................. 6-2
6.2.2 Doppler Restoration ................................................................................................. 6-6
6.3 Installation of Optional Devices ...................................................................................... 6-10
6.4 Data Backup and Storage ............................................................................................... 6-11
6.4.1 Manage Settings ..................................................................................................... 6-11
6.4.2 Patient Data Backup and Restore............................................................................ 6-11
6.5 Software Maintenance ................................................................................................... 6-12
6.5.1 Product Configuration ............................................................................................ 6-12
6.5.2 Log Maintenance ................................................................................................... 6-13
6.6 Display Parameter Setting ............................................................................................. 6-14
ii
6.7 HDD Partition ................................ ................................................................ ................ 6-15
7 Structure and Assembly/Disassembly .................................................................... 7-1
7.1 Structure of the Complete System ................................................................................... 7-1
7.2 Field Replaceable Unit .................................................................................................... 7-2
7.3 Preparations ................................................................................................................... 7-11
7.3.1 Tools Required ........................................................................................................ 7-11
7.3.2 Engineers Required ................................................................................................ 7-11
7.3.3 Assembly/Disassembly Required ............................................................................ 7-11
7.4 Assembly/Disassembly ................................................................................................... 7-11
7.4.1 Battery Connecting Board ...................................................................................... 7-12
7.4.2 Power Supply Module ............................................................................................ 7-14
7.4.3 IO Broad ................................................................................................................ 7-17
7.4.4 Probe Board........................................................................................................... 7-18
7.4.5 Main Board and CPU Module ................................................................................. 7-19
7.4.6 Top Cover Assembly of Keyboard ........................................................................... 7-21
7.4.7 Display Assembly ................................................................................................... 7-25
7.4.8 Hard Disk ............................................................................................................... 7-29
7.4.9 Speaker ................................................................................................................. 7-30
8 System Diagnosis and Support ................................................................ ............... 8-1
8.1 General Status Indicator .................................................................................................. 8-1
8.1.1 Status Indicators of the Control Panel ....................................................................... 8-1
8.1.2 Status Indicator of the Power Supply on the IO Board ............................................... 8-2
8.1.3 Status of whole machine .......................................................................................... 8-2
8.2 Starting Process of the Whole System ................................................................ ............. 8-3
8.2.1 Start Process of Complete System ........................................................................... 8-3
8.2.2 Start-up Process of BIOS ......................................................................................... 8-4
8.2.3 Start-up of Linux ....................................................................................................... 8-5
8.2.4 Start-up of Doppler ................................................................ ................................... 8-6
8.3 Alarming and Abnormal Information ................................................................................. 8-8
8.3.1 Turning on the System Configuration File is Abnormal .............................................. 8-9
8.3.2 The voltage of system power is abnormal ................................................................. 8-9
8.3.3 Temperature Alarming .............................................................................................. 8-9
8.3.4 Fan Alarming .......................................................................................................... 8-10
8.3.5 Battery Alarming..................................................................................................... 8-10
8.3.6 PHV Related Alarming............................................................................................. 8-11
9 Care and Maintenance .............................................................................................. 9-1
9.1 Overview ......................................................................................................................... 9-1
9.1.1 Tools, Measurement Devices and Consumables....................................................... 9-1
9.1.2 Care and Maintenance Items ................................................................................... 9-2
9.2 Cleaning .......................................................................................................................... 9-3
9.2.1 Clean the System ..................................................................................................... 9-3
9.2.2 Content .................................................................................................................... 9-3
9.2.3 Clean the Peripherals ............................................................................................... 9-6
9.3 Checking ......................................................................................................................... 9-6
9.3.1 General check .......................................................................................................... 9-6
9.3.2 System Function Check ........................................................................................... 9-7
9.3.3 Peripherals and Options Check ................................................................................ 9-8
9.3.4 Mechanical Safety Inspection ................................................................................... 9-8
9.3.5 Electrical Safety Inspection ...................................................................................... 9-9
iii
10 Troubleshooting of Regular Malfunctions............................................................. 10-1
10.1 System cannot be powered on ...................................................................................... 10-1
10.1.1 Module or Board Related ....................................................................................... 10-1
10.1.2 Key Points Supporting Troubleshooting .................................................................. 10-1
10.1.3 Troubleshooting ..................................................................................................... 10-1
10.2 System Cannot Start up Normally .................................................................................. 10-2
10.2.1 Module or Board Related ....................................................................................... 10-2
10.2.2 Key Points Supporting Troubleshooting .................................................................. 10-2
10.2.3 Troubleshooting ..................................................................................................... 10-2
10.3 Image Fault ................................................................................................................... 10-3
10.3.1 Module or Board Related ....................................................................................... 10-3
10.3.2 Key Points Supporting Troubleshooting .................................................................. 10-4
10.3.3 Troubleshooting ..................................................................................................... 10-4
10.4 Probe Socket System Malfunction ................................................................................. 10-4
10.4.1 Module or Board Related ....................................................................................... 10-4
10.4.2 Key Points Supporting Troubleshooting .................................................................. 10-5
10.4.3 Troubleshooting ..................................................................................................... 10-5
10.5 IO Interface System ................................ ....................................................................... 10-5
10.5.1 Module or Board Related ....................................................................................... 10-5
10.5.2 Key Points Supporting Troubleshooting .................................................................. 10-6
10.5.3 Troubleshooting ..................................................................................................... 10-6
10.6 Control Panel ................................................................................................ ................ 10-6
10.6.1 Module or Board Related ....................................................................................... 10-6
10.6.2 Key Points Supporting Troubleshooting .................................................................. 10-7
10.6.3 Troubleshooting ..................................................................................................... 10-7
10.7 LCD Display ................................................................................................ .................. 10-8
10.7.1 Module or Board Related ....................................................................................... 10-8
10.7.2 Key Points Supporting Troubleshooting .................................................................. 10-8
10.7.3 Troubleshooting ..................................................................................................... 10-8
Appendix A ELECTRICAL SAFETY INSPECTION....................................................A-1
Appendix B Phantom Usage Illustration..................................................................B-1
iv

Revision History

Revision
Date
Reason for Change
1.0
2012.03.20
Initial release
2.0
2011.09.01
1 A new model is added to the cover page. Z5/Z5BW/Z5T/Z5Vet; 2 4.1 Hardware system overall structure: add hardware differences
between Z5 series and Z6 series systems;
3 Add principle description of main board and probe board of Z5
series to 4.2;
4 Add material information of main board and probe board of Z5
series to 7.2 field replaceable unit; 5 Increase disassembly steps of speaker in 7.4.9; 6 Update Appendix A electrical safety inspection form and limit value
requirement.
3.0
2013.04.17
Add FRU part number for HDD of different regions/types (CE/FDA/human/vet)
4.0
2013.6.24
Add “The attentions to the assembly/disassembly, otherwise the hard disk will be damaged to Chapter 7.4.8.
5.0
2015.1.14
Section 7.2, add new main board FRU number and compatibility description.
6.0
2015.6.10
Section 7.2, add new HDD FRU numbers for Z5 Vet FDA and Z5 FDA.
7.0
2015.7
Section 2.1.3, add printer SONY UP-D898MD and SONY UP-X898MD
8.0
2015.8
Section 7.2, change HDD and Main board FRU numbers.
9.0
2015.1.14
Change the FRU number of the display assembly in chapter 7.2
10.0
2016.4.10
Add HDD number in chapter 7.2.
11.0
2016.8.9
Update the labels in 1.2.2.
12.0
2016.9.23
Update FRU number of main board, HDD and PC module due to COME module change matched with BayTrail.
13.0
2016.11.23
Update the log of Alarming and Abnormal Information in chapter 8.3
14.0
2018.8
Add part numbers of AC-DC board and DC-DC board in chapter 7
15.0
2018.9
Correct effective date of version 13.0 in the cover document.
Mindray may revise this publication from time to time without written notice.
© 2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All Rights Reserved.
I

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may referring to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
, , , , , BeneView, WATO,
BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners.

Applicable for

This service manual is applicable for the service engineers, authorized service personnel and service representatives of this ultrasound system.

Statement

This service manual describes the product according to the most complete configuration; some of the content may not apply to the product you are responsible for. If you have any questions, please contact Mindray Customer Service Department.
Do not attempt to service this equipment unless this service manual has been consulted and is understood. Failure to do so may result in personnel injury or product damage.

Responsibility on the Manufacturer Party

Mindray is responsible for the effects on safety, reliability and performance of this product, only if: All installation operations, expansions, changes, modifications and repairs of this product are
conducted by Mindray authorized personnel;
The electrical installation of the relevant room complies with the applicable national and local
requirements;
The product is used in accordance with the instructions for use.
II
Mindray's obligation or liability under this warranty does not include any transportation or other
WARNING:
It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
Manufacturer:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address:
Mindray Building,Keji 12th Road South,High-tech industrial park,Nanshan,Shenzhen 518057,P.R.China
Website:
www.mindray.com
E-mail Address:
service@mindray.com
Tel:
+86 755 81888998
Fax:
+86 755 26582680
charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
Any Mindray product which has been subjected to misuse, negligence or accident; Any Mindray product from which Mindray's original serial number tag or product identification
markings have been altered or removed;
Any products of any other manufacturers.

Customer Service Department

III

1 Preface

Signal word
Meaning
DANGER
Indicates death or serious injury may occur imminently in this hazardous situation if not avoided.
WARNING
Indicates death or serious injury may occur potentially in this hazardous situation if not avoided.
CAUTION
Indicates minor or moderate injury may occur potentially in this hazardous situation if not avoided.
NOTE
Indicates property damage may occur potentially in this hazardous situation if not avoided.
Symbol
Meaning
Location
Type-BF applied part The ultrasound probes connected to this system are type-BF
applied parts. The ECG module connected to this system is Type-BF applied part.
Above the IO panel
This chapter describes important issues related to safety precautions, as well as the labels and icons on the ultrasound machine.

1.1 Meaning of Signal Words

In this operator’s manual, the signal words DANGER, WARNING, CAUTION and
NOTE are used regarding safety and other important instructions. The signal words and their
meanings are defined as follows. Please understand their meanings clearly before reading this manual.

1.2 Meaning of Symbols

The meaning and location of the safety symbols and warning labels on the ultrasound machine are described in the following tables, please read them carefully before using the system.

1.2.1 Meaning of Safety Symbols

Preface 1-1
Caution
On the rear panel
Dangerous voltage
No.
Warning Labels
Meaning
1.
Please carefully read this manual before use device.
2.
The following labels are available when the system works with the mobile trolley.
a. Do not place the device on a sloped surface. Otherwise the device may slide, resulting in personal injury or the device malfunction. Two persons are required to move the device over a sloped surface.
b. Do not sit on the device. c. DO NOT push the device. When the casters are locked.
Symbol
Meaning
Location
Equipotentiality
Power panel
Power button
Upper right corner on the control
panel
Network port
IO panel
USB port

1.2.2 Warning Labels

1.2.3 General Symbols

This system uses the symbols listed in the following table, and their meanings are explained as well.
1-2 Preface
Video output Remote port
VGA port
VGA signal output
AC indicator
Lower left corner on the control
panel
Battery indicator
Standby indicator
Lower right corner on the control
panel
Hard disk indicator
A
Probe connector A
Rear panel
B
Probe connector B
This product is provided with a CE marking in accordance with the regulations stated in Council Directive 93 / 42 / EEC concerning Medical Devices. The number adjacent to the CE marking (0123) is the number of the EU-notified body certified for meeting the requirements of the Directive.
DANGER
Do not operate this system in an atmosphere containing flammable or explosive gases such as anesthetic gases, oxygen, and hydrogen or explosive fluid such as ethanol because an explosion may occur.

1.3 Safety Precautions

Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the probes.
Preface 1-3

1.3.1 Electric safety

WARNING:
1.
Do connect the power plug of this system and power plugs of the peripherals to wall receptacles that meet the ratings indicated on the rating nameplate. Using a multifunctional receptacle may affect the system grounding performance, and cause the leakage current to exceed safety requirements.
2.
Do not use any cables other than the cables provided with the device by Mindray.
3.
Use the cable provided with this system to connect the printer. Other cables may result in electric shock.
4.
Disconnect the AC power before you clean or uninstall the ultrasound machine, otherwise, electric shock may result.
5.
Do not use this system simultaneously with equipment such as an electrosurgical unit, high-frequency therapy equipment, or a defibrillator, etc.; otherwise electric shock may result.
6.
This system is not water-proof. If any water is sprayed on or into the system, electric shock may result.
CAUTION:
1.
DO NOT connect or disconnect the systems power cord or its accessories (e.g., a printer or a recorder) without turning OFF the power first. This may damage the system and its accessories or cause electric shock.
2.
Avoid electromagnetic radiation when perform performance test on the ultrasound system.
3.
In an electrostatic sensitive environment, don’t touch the
device directly. Please wear electrostatic protecting gloves if necessary.
4.
You should use the ECG leads provided with the ECG module. Otherwise it may result in electric shock.
WARNING:
1.
When moving the system, you should first power off the system, fold the LCD display, disconnect the system from other devices (including probes) and disconnect the system from the power supply.
2.
Do not subject the probes to knocks or drops. Use of a defective probe may cause electric shock to the patient.
CAUTION:
1.
Do not expose the system to excessive vibration (during the transportation) to avoid device dropping, collision, or mechanical damage.

1.3.2 Mechanical Safety

1-4 Preface
2.
When you place the system on the mobile trolley and move them together, you must secure all objects on the mobile trolley to prevent them from falling. Otherwise you should separate the system from the mobile trolley and move them individually. When you have to move the system with the mobile trolley upward or downward the stairs, you must separate them first and then move them individually.
3.
Do not move the ultrasound system if the HDD indicator is green, sudden shake may cause the HDD in damage.
NOTE:
1.
The user is not allowed to open the covers and panel of the system, neither device disassemble is allowed.
2.
To ensure the system performance and safety, only Mindray engineers or engineers authorized by Mindray can perform maintenance.
3.
Only technical professionals from Mindray or engineers authorized by Mindray after training can perform maintenance.
NOTE:
For detailed operation and other information about the ultrasound system, please refer to the operator’s manual.

1.3.3 Personnel Safety

1.3.4 Other

Preface 1-5

2 Product Specifications

2.1 Overview

2.1.1 Intended Use

Z5 series and Z6 series are diagnostic ultrasound system, which are intended for use in clinical ultrasonic diagnosis.
Z5 series include: Z5/Z5BW/Z5T/Z5Vet; Z5 series include: Z6/Z6T/Z6S/Z6W/Z6Vet.

2.1.2 Introduction of Each Unit

Rear view:
Product Specifications 2-1
Left side:
2-2 Product Specifications
No.
Name
Function
1
Probe holder
Sets the probe
2
Display
Displays the image and parameters
3
Control Panel
Man-machine interface for operation control, for details, please refer to 2.1.2.1 Control Panel”.
4
Handle
For lifting the machine
5
IO Panel
Signal input / output panel, for details, please refer to “2.1.2.2 IO Panel”.
6
Power supply panel
Signal input / output panel, for details, please refer to “2.1.2.3 Power Supply Panel”.
7
Probe sockets
Sockets connecting probes and the main unit
8
Battery cover
Covering battery
9
USB ports
For connecting USB device (2 ports)
2.1.2.1 Control Panel
Product Specifications 2-3
2-4 Product Specifications
No.
Name
Description
Function
1
/
Power button
Off: when system is turned off; Green: when system is turned on by pressing this
button; Orange: when system is in standby.
2
Esc
Exit
Press to exit the current status to the previous status.
3
Help
/
Press to open or close the accompanying help documents.
4
Patient
Patient Information
Press to open/ exit patient information screen.
5
Probe
Probe switch
Press to switch Probe and Exam Type
6
Review
/
Press to review the stored images.
7
Report
/
Press to open or close the diagnosis reports.
8
End Exam
/
Press to end an exam.
9
iStation
/
Press to enter or exit the patient information management system.
10
F1
User-defined key
You can assign a function to the key.
11
Biopsy
/
Press to show or hide the biopsy guide line.
12
Setup
/
Press to open/close the setup menu.
13
Del / Press to delete the comment, etc.
14
/
Alphanumeric keys
Same as on PC
15
Dual
Dual-split screen
Press to enter Dual mode from non-Dual mode; Press to switch between windows in Dual mode.
16
Arrow
/
Press to enter or exit the arrow comment status.
17
Cine
Cine Review
Press to enter or exit the Cine Review status.
18
/
Direction key
To adjust display brightness or contrast when pressing with <Fn> key.
19
Steer
Press to open slant function for linear probe
20
Body Mark
/ Press to enter or exit the Body Mark status.
21
Comment
/
Press to enter or exit the character comment status.
22
Clear
/
Press to clear the comments or measurement calipers on the screen.
23
Scale
/
Press to adjust the Scale value
24
Baseline
/
Press to adjust the baseline value
25
Nav.Rot
/
Multifunction knob
26
/
/
Press to implement the different function with “Scale/Baseline/Nav.Rot” three keys.
No.
Name
Description
Function
27
Cursor
/
Press to show the cursor.
28
Back
/
Press to return to the previous operation or to delete the last project
29
/
Trackball
Roll the trackball to change the cursor position.
30
Set / Press to confirm an operation.
31
PW / Press to enter the PW mode
32
Color
/
Press to enter the color mode
33
M / Press to enter the M mode.
34
B / Press to enter the B mode
35
Measure
/
Press to enter/ exit Application Measurement
36
Update
/
Measurement status: press to switch between the fixed and active end of the caliper;
Multi-imaging mode: press to change the currently active window.
37
Caliper
/
Press to enter/ exit Measurement
38
Gain/ iTouch
/
Rotate: to adjust B or M gain Press: to enter/ exit iTouch
39
TGC
/
Move to adjust time gain compensation.
40
Focus Freq./THI
/
Press: to switch between Focus and Freq./THI; Rotate: to adjust corresponding parameter
41
Depth Zoom
/
Press: to switch between Depth and Zoom; Rotate: to adjust corresponding parameter
42
Save 1
/
Press to save, user-defined key
43
Save 2
/
Press to save, user-defined key
44
Freeze
/
Press to freeze or unfreeze the image.
45
Print
/
Press to print: user-defined key.
46
/
Indicator 1
AC indicator
AC supply: light green; Battery supply: light off.
47
/
Indicator 2
Battery status indicator
Charging: light in orange Full: light in green Discharge (electricity >20%): light in green Discharge (electricity <20%): blinking in orange Discharge (electricity <5%): blinking in orange
rapidly Non-charge/ discharge: light off
Product Specifications 2-5
No.
Name
Description
Function
48
/
Indicator 3
Standby indicator
Standby: blinking in orange Other status: light off
49
/
Indicator 4
HDD status indicator
Read/ write: blinking in green Other status: light off
NOTE: DO NOT move the machine when the indicator blinking in green. Otherwise the HDD may be damaged by sudden shake.
<2> <3> <4>
<1>
<5> <6> <7>
VGA
12V 5V
3.3V A
<8>
No.
Symbol
Function
1
Network port
2
USB ports
3
4
Separate video output, connecting video printer or monitor
5
Remote control port
6
VGA
VGA signal output
7
Composite video output
8
/
Power indicator
2.1.2.2 IO Panel
The IO panel is on the back of the main system.
2-6 Product Specifications
2.1.2.3 Power Supply Panel
<1>
<2>
100-240~ 50/60Hz 1.5-0.8A
No.
Name
Function
1
Power inlet
AC power inlet
2
Equipotential terminal
Used for equipotential connection, that balances the protective earth potentials between the system and other electrical equipment.
Name
Model
Footswitch
971-SWNOM (2-pedal, USB port) SP-997-350 (3-pedal, USB port)
Black / white video printer
Analog: SONY UP-897MD, MITSUBISHI P93W-z Digital: SONY UP-D898MD, SONY UP-X898MD
Color video printer
Analog: SONY UP-20, MITSUBISHI CP910E Digital: HP Photosmart plus B210A
Graph/ text Printer
HP series Tested: HP deskjet 1280, HP Laserjet CM1015, HP officejet 6000, HP officejet J3608 all-in-one
External DVD R/W
SE-S224Q
Barcode scanner
SYMBOL LS2208-SR
Wireless network card
EDUP EP-MS150N11N series
NOTE:
If the ultrasound system can not recognize the SONY UP-X898MD and SONY UP-D898MD printers automatically, you may need to change the settings on the printer: push <PUSH ENTER> to enter the main menu and select [DIGITAL]->[DRIVER], and select [897].

2.1.3 Peripherals Supported

Product Specifications 2-7

2.2 Specifications

Voltage
100-240V
Current
1.5-0.8A
Frequency
50/60Hz±3Hz
Operating conditions
Storage and transportation conditions
Ambient temperature
0℃~40
-20℃~55
Relative humidity
30%~85% (no condensation)
30%~95% (no condensation)
Atmospheric pressure
700hPa1060hPa
700hPa1060hPa
Warning
Do not use this system in the conditions other than those specified
Voltage
12V
Dimension
15 inch
Resolution
1024×768
Adjustable angle
30 degree

2.2.1 Dimensions and Weight

Dimensions Fold378mmH×190mmD×415mmW Unfold396mmH×476mmD×415mmW Net weight≤8.8Kgincluding batteryno probe holder

2.2.2 Electrical Specifications

2.2.3 Environmental Conditions

2.2.4 Monitor Specification

2-8 Product Specifications

3 System Installation

NOTE:
Do not install the machine in the following locations: Locations near heat generators;
Locations of high humidity; Locations with flammable gases.
WARNING:
DO NOT connect this system to outlets with the same circuit breakers and fuses that control the current of devices such as life-support systems. If this system malfunctions and generates an over current, or when there is an instantaneous current at
power ON, the circuit breakers and fuses of the building’s
supply circuit may be tripped.

3.1 Preparations for Installation

3.1.1 Electrical Requirements

3.1.1.1 Requirement of Regulated Power Supply
Requirement of power supply is referring to 2.2.2.Due to the difference of the power supply stability of different districts, please advise the user to adopt a regulator of good quality and performance such as an on-line UPS.
3.1.1.2 Grounding Requirements
The power cord of the system is a three-wire cable, the protective grounding terminal of which is connected with the grounding phase of the power supply. Please ensure that the grounding protection of the power supply works normally.
3.1.1.3 EMI Limitation
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI.
Possible EMI sources should be identified before the unit is installed. Electrical and electronic equipment may produce EMI unintentionally as the result of defect.
These sources include: medical lasers, scanners, monitors, cauterizing guns and so on. Besides, other devices that may result in high frequency electromagnetic interference such as mobile phone, radio transceiver and wireless remote control toys are not allowed to be presented or used in the room. Turn off those devices to make sure the ultrasound system can work in a normal way.
System Installation 3-1

3.1.2 Installation Condition

3.1.2.1 Space Requirements
Place the system with the necessary accessories at a proper position for convenient use.
1. Place the system in a room with good ventilation or having an air conditioning unit.
2. Leave at least 20cm clearance around the system to ensure effective cooling.
3. A combination lighting system in the room (dim/bright) is recommended.
4. Except the receptacle dedicated for the ultrasound system, at least 3-4 spare receptacles on the wall are available for the other medical devices and peripheral devices.
5. Power outlet and place for any external peripheral are within 2 m of each other with peripheral within 1 m of the unit to connect cables.
3.1.2.2 Network Environment
Both wireless and wired LAN are supported by this ultrasound system.
Data transmission is allowed between different departments or areas without network cable. Network can be automatically connected after disconnection in case that the device is required to be moved, wireless transmission task can be recovered after the network resumed to normal condition. Confirm the network devices and network conditions before the installation.
1. General information: default gateway IP address, and the other routers related information.
2. DICOM application information: DICOM server name, DICOM port, channels, and IP address.
3.1.2.3 Confirmation before Installation
Please confirm the following items before installation:
1. The video format used in the region or country where the system is installed.
2. The language used in the region or country where the system is installed.
3. The power voltage and frequency used in the region or country where the system is installed.
4. Obstetric formulae and other measurement formulae used in the region or country where the system is installed.
5. Other settings to be used in the region or country where the system is installed but different from the factory settings.
6. The doctor’s habits when using the system.
Please confirm the items above prior to the installation training, and do the system settings
according to the universal setting of installed region or country.

3.2 Unpacking

Tools Required: None. 3-2 System Installation
Installation duration: 1 person, 0.5 hour.

3.2.1 Unpacking

1. Use the scissor to clip the 2 rubber belts as follows:
2. Use a knife to open the tapes:
3. Take out the operation manual:
4. Take out the probe box 1:
System Installation 3-3
5. Take out the accessory box:
6. Take out the probe box2, as follows (the box is below the accessory box):
7. Hold the handle tightly and pull the machine with the surrounding foam:
8. Remove the surrounding protective foam to take out the main system.
3-4 System Installation

3.2.2 Checking

1. After unpacking, check the objects in the container with the package list to see if anything is in short supply or is wrong.
2. Inspect and make sure there is no damage to the machine, no indentation, no cracks.

3.3 Installation of Main Unit

3.3.1 Installing Battery

As shown in the following figure.
1. Open the battery cover and put the battery near the bay (Note that the battery shall be put
in correct direction).
2. Push the battery into the battery bay , and push to the right side as until it’s tightly
locked.
3. Close the battery cover.
Figure 3-1 Installing Battery
System Installation 3-5

3.3.2 Control Panel Adjusting

Press the lock buttons on the both sides of the control panel as illustrated in the following
picture, open the control panel.
Figure 3-2 Control Panel Adjustment

3.3.3 Display Adjusting

1. Open the display as described in 3.3.2
2. Put the finger into the bottom of the display and pull the display to tilt the display (30 degrees max.)
3. Ear off the screen protective film.
Figure 3-3 Open and tilt the display
Press <Fn>+ directional keys to adjust brightness and contrast. Brightness Adjustment:
refer to the brightness control keys; press <Fn>+ to decrease the
brightness; and press <Fn>+ to increase the brightness.
3-6 System Installation
Contrast adjustment:
NOTE:
On the monitor, the brightness adjustment comes before contrast. After readjusting the monitor’s contrast and brightness, adjust all preset and peripheral settings.
, refer to the contrast control keys; press <Fn>+ to decrease the contrast;
and press <Fn>+ to increase the contrast. Restore the factory default settings Open the page via “Setup <F10>→General” and restore the brightness/contrast to factory
default setting.

3.3.4 Installing Probe Holder

Fix the probe holder hanger into the groove in the right side and the rear-right side of the main
unit, and push downwards to confirm the installation.
Figure 3-4 Installing the Probe Holder

3.3.5 Connecting a Probe

1. Check the probe connector, if is not open, Turn the locking lever anticlockwise to open it.
2. Keep the cable toward probe holder and insert the connector into the port.
3. Turn the locking lever 90° clockwise in the horizontal position to lock the probe connector.
4. Place the probe properly to avoid being trod or wrapping with other devices. DO NOT allow the
probe head to be hung freely.
5. To disconnect the probe, turn the locking lever 90° anticlockwise and pull the probe straight out.
System Installation 3-7
NOTE:
Before inserting the connector into the probe port, inspect the connector pin. If the pin is bent, do not use the probe until it has been inspected / repaired / replaced.

3.4 Installing Peripherals

For the models of the supported peripherals, please refer to “2.1.3 Supported Peripherals”.

3.4.1.1 Footswitch Installation

1. Connecting: Take 971-SWNOM as an example: insert the USB connector to the system
available USB ports (on the left and rear of the machine).then the footswitch can be used directly.
2. For settings of footswitch, please refer to 3.5.3 system preset.
Figure 3-5 Footswitch connection

3.4.2 Video Printer Installation

Analog video printer:
1. Connect one end of the signal line to the Video In interface of the printer, and the other to the video output port in the ultrasound system IO panel.
2. Connect the Remote control line to the Remote interface in the ultrasound system IO panel.
3. Insert the power cord to a power supply receptacle that is well grounded.
3-8 System Installation
Figure 3-6 Installing Analog Video Printer
Digital Video Printer
1. Connect one terminal of the data cable of the video printer to the USB port of the
ultrasound system and the other terminal to the video input port of the video printer;
2. Insert the power cord to a power supply receptacle that is well grounded.
3. Please refer to 3.5.4 for the driver installation. And you need not install the driver of the
printer listed in section “2.1.3”.

3.4.3 Installing a Graph / Text Printer

A graph / text printer is configured with power cord and data cable, and is connected to the
system via the USB port. The connection method is described as follows:
1. Connect the data cable to the USB port in the ultrasound system.
2. Plug the power plug into a power supply receptacle that is well grounded.
3. Please refer to 3.5.4 for the driver installation. And you need not install the driver of the printer listed in section “2.1.3”.

3.4.4 Installing External DVD-R/W

1. Connect the USB cable connector of the external DVD recorder to the USB port in the ultrasound system;
System Installation 3-9
2. Connect the power supply adapter plug of the DVD recorder to an appropriate receptacle that is well grounded.
Figure 3-7 Installing External DVD-R/W

3.4.5 Installing Barcode Scanner

The system supports barcode reader to read the patient information (ID).
1. For structure of the scanner, see the figure below. The important parts are: LED indicator, scanning surface, and the switch.
2. Connect the cable to the port on the scanner.
3. Connect the other end of the cable to the USB port on the ultrasound system.
3-10 System Installation
4. When the ultrasound system is working, information scanning can be performed by pressing the switch on the scanner. For detailed operations, please refer to the operator’s manual of the scanner.
5. Fix the scanner on the bracket (see the figure below) to avoid accidental falling.
2D scanner 1D scanner

3.5 System Configuration

3.5.1 Power ON / OFF

Connect the system power cord to the AC power, and make sure the ultrasound system and
other optional devices are correctly connected.
When the AC indicator is green, you can turn on the power button (located at the upper right
corner of the system) to initiate the system. After being normally rebooted, it will display image interface. Or press the power button directly when the battery is of sufficient capacity.

3.5.2 Enter Doppler

After the system is powered on after initiation (about 30S), it enters Doppler directly:
System Installation 3-11

3.5.3 System Preset

1. Press the <Setup> key to enter the [Setup] menu.
2. Click <System Preset> to enter the screen as follows:
Region
3-12 System Installation
In the Region page, set the system language, date format, date, time and hospital related information, etc.
General
Click <System Preset> to enter:
In this page, set the time of standby, brightness/contrast and color temperature of display,etc.
Key Config
System Installation 3-13
1) Function of keyboard keys F1 and the footswitch keys (left, mid, right) are user-defined.
2) Key brightness, key volume, trackball backlight and trackball sensitivity can be adjusted. Image Preset Click [Image Preset] to enter:
General image parameters can be set in this page.

3.5.4 Print Preset

1. Press <Setup> and click [Print Preset] to set video printer, graph/text printer parameters:
3-14 System Installation
2. After connecting the local printer, Click “Printer Driver”, the system will display the printer name and status (Ready) automatically which already installed printer driver successfully.
3. Return “Printer Service” page, Select the corresponding service from the printer list and increase the service.
The system integrates drivers of HP printers, after HP printers are connected, drivers will be installed automatically (about 10s).If auto installation fails, icon will display on the right lower corner of the screen to warn you that manual installation is necessary. The driver installation method is as follows:
System Installation 3-15
a) Download the ppd file from HP official website (contact R&D engineer if necessary), and copy
NOTE:
1.
Before connect the network printer, make sure the ultrasound system and the printer are in the same network domain, and the network is working normally.
2.
The IP address and the server name should be valid, e.g. \\10.2.40.123 or
\\5-HP, otherwise, the system will fail to connect.
3.
If the server has set accessing limitation, the system will prompt a dialogue box to identify the user. Enter the correct user name, domain name and password, and then click [OK].
the ppd file to the storage device (USB disk as an example).
b) Connect the U disk to the USB port nearside the control panel of ultrasound system, click the
icon to pop up the screen, select the U disk to run the ppd file and finish the installation.
Add network printer
1. In “Printer Driver” screen, click [Add Network Printer] to pop up the screen, enter the necessary
information (IP address, shared printer name, server name, domain name and password).
2. After successful connection, the newly added network printer name will be shown in the printer
driver list.

3.5.5 Network Preset

3.5.5.1 Network Preset
Open “[Setup]→[Network Preset]” to enter the screen.
1) Please select the network type according the actually status, Select “DHCP”, click [OK].
2) Or, select “Static”, and input the IP address, subnet mask and gateway, then click [OK].
3-16 System Installation
Name
Description
Current Network Adapter
To select the network connection mode
DHCP
/ Static
If “DHCP” is selected, IP address will be automatically obtained
from DNS server; if “Static” is selected (using static IP address), you need to enter the IP address.
IP Address
IP address of the system should be at the same network segment with the server IP.
Subnet Mask
Used to set different network segment.
Gateway
Used to set the gateway IP.
3.5.5.2 DICOM Preset
NOTE:
Only if DICOM basic option is configured, [DICOM Local], [DICOM Server], [DICOM Service] are available.
NOTE:
1.
AE Title should be the same with the SCU AE Title preset in the server (PACS/RIS/HIS).
2.
DICOM communication port in the ultrasound system should be the same with the one in the server.
1. Click [DICOM Preset] to open the DICOM preset screen.
DICOM Setting
Enter the AE Title of the ultrasound system, port and PDU according to the actual situation, and
then click [OK] to exit the screen.
DICOM Server Setting
1) Enter the device name and the IP address.
2) You can ping other machines to verify connection after entering the correct IP address. Also you can check the connection of the already added server in the list.
System Installation 3-17
3) If connection is successful, click [Add] to add the service to the Service list.
NOTE:
If the currently entered name has already existed, the system will pop up: “The server
name exists!” Click [OK] to enter another name.
NOTE:
DICOM Work list can be configured only after DICOM Basic is configured, and if DICOM Work list function is not configured, the “Work list” page is not accessible.
2. Click [DICOM Service] to open the DICOM Service screen.
Only when the system is configured with DICOM basic function module, and installed DICOM Work list, MPPS, DICOM Structured Reporting and Query/ Retrieve modules, can the corresponding preset settings be found in DICOM Service screen.
The DICOM Service Setting is used to set properties of DICOM services as Storage, Print, Work list, MPPS, Storage Commitment and Query/ Retrieve.

3.5.6 System Information

In System Information screen, it displays the product configuration, software version, hardware
& boards, and driver related information. You can check the product information here.
1. Press the <Setup> and click [System Info].
3-18 System Installation
NOTE:
1.
Be sure to confirm the system information before and after the software maintenance.
2.
Ask the user to save the current system information if necessary.
2. Click [About Detail] to check the detailed board information.
3. Click [Save] to export the information in the format of “*.TXT”; click [Exit] to return to the Setup menu.
System Installation 3-19

4 Hardware Principle

DC-DC power
board
AC IN
Battery
connecting
board
battery
AC-DC power
board
Main board
Probe board
I/O board
Fan
Fan
Display
Control panel
Power supply
Communication
& control
Ultrasound signal
CPU
module
Discharge/
charge

4.1 General Structure of Hardware System

Figure 4-1 Schematic Diagram of System
As a portable black-and-white ultrasonic productthe system supports two probe slots. The
difference between Z5 series and Z6 series is as follows.
Hardware Principle 4-1
Different
Items
Z6 Series
Z5 Series
Remarks
Main board
64 transmitting/ receiving channels
32 transmitting/ receiving channels
Using the same PCB, different BOM Distinguishing system software by the broad ID
Probe Board
Supportting128-array, 192-array and bi-planar probes by using 156-pin probe port.
Supporting 80-array probe by using 96-pin probe port.
Probe boards of two series products are totally different, but the consitent interface definition of the main board.
The detailed structure of hardware system is in the figure above, and they can be divided into 4
modules:
Main unit module Power supply module Monitor module Control panel module
The detailed description of 4 modules will be in the following:

4.2 Main Unit

The main unit mainly involves 3 boards:
Probe board Main board IO board
4-2 Hardware Principle

4.2.1 Probe Board

High-voltage
switch
Relay
CPLD
Socket A (156 pin)
Relay
Driver circuit
Socket B (156 pin)
POUTA[128:1] POUTB[128:1]
A socket control signalNote 1
RELAY_EN
PRB_A_VDD_EN , PRB_A_VCC_EN , HV_PWR_SW1
BOARD_ID[7:0]
3.3V/5V/+95V/-95V
Power ON/OFF
Control circuit
POUT[128:1]
driver
LC filter
Driver
driver
B socket control signal Note2
PRB_B_VDD_EN,PRB_B_VCC_EN,HV_PWR_S W2
3.3V/5V/+95V/-95V
Power ON/OFF
Control circuit
BOARD ID
CTRL_PROBE_A_245_OE_N
Probe board of Z6 series
+5V +95V
-95V
RELAY_PWR_EN
Note:
1.A socket control signal: PROBE_A_SPI_CLK/CLK PROBE_A_SPI_CS/nLE PROBE_A_SPI_DIN/DATA0 PROBE_A_SPI_DOUT/DP_SWITCH PROBE_A_SPI_nWP/DATA1 PROBE_A_SPI_SEL
Probe board control signal
CTRL_PROBE_B_245_OE_N
PRB_A_ON
VDD
A socket control signal
VDD
PRB_B_ON
B socket control signal
Driver
High-voltage switch signal
Driver
driver
LDO
D+3V3
A+5V A+95V A-95V
+5V
D+3V3
2.B socket control signal: PROBE_B_SPI_CLK/CLK PROBE_B_SPI_CS/nLE PROBE_B_SPI_DIN/DATA0
PROBE_B_SPI_DOUT/DP_SWITCH PROBE_B_SPI_nWP/DATA1 PROBE_B_SPI_SEL
4.2.1.1 Probe Board of Z6 Series
The function of Z6 series probe board describes as following:
The ultrasound signal is transferred from 64-channel to 128-array by the high-voltage
Probe information could be obtained by probe board via reading probe online signal and
The probe board supplies power to corresponding Flash of saving probe ID when reading
Support two 156-array probes (A and B), and only one probe could be chosen by relay
Support 192-array probe and bi-planar probe. Support 192-array probe and bi-planar probe to work normally by 5v output of the probe
The main board supplies ±95v output to 192-array probe and bi-planar probe to work
Figure 6-3 Principle Diagram of Z5 Series Probe Board
switch on the probe board.
probe ID.
probe ID, and Flash power shuts off after getting probe ID.
switching on the probe board.
broad.
normally.
Hardware Principle 4-3
4.2.1.2 Probe Board of Z5 Series
High-voltage
switch
Relay
CPLD
Transducer
socket A
(96 pin)
Relay drive
circuit
Transducer
socket B
(96 pin)
POUTA[80:1]
POUTB[80:1]
RELAY_EN
BOARD_ID[7:0]
POUT[80:1]
driver
BOARD ID
Probe board of Z5 series
RELAY_PWR_EN
PRB_B_ID[6:1]
PRB_A_ID[6:1]
VDD
Driver
High-voltage switch control signal
+5V +95V
-95V
Probe board control signal
LC
LDO
D+3V3
A+5V A+95V A-95V
PRB_B_ON
VDD
driver
driver
Driver
driver
PRB_A_ON
OE_N
PRB_A_ID[6:1]
OE_N
PRB_A_ID[6:1]
PRB_B_ON
PRB_A_ON
Figure 6-3 Principle Diagram of Z5 Series’ Probe Board
The function of Z5 series probe board describes as following:
The ultrasound signal are transferred from 32-physical channel to 80-array by the
high-voltage switch on probe board.
Probe information could be obtained by probe board via reading probe online signal and
probe ID.
Support two 80-pin probes, and only one probe could be chosen to work by relay switching
on probe board.

4.2.2 Main board

The main board could be divided into two parts: front-end of the main board and back-end of the
main board.
4-4 Hardware Principle
4.2.2.1 Front-end of main board
Main board
HVISO
32ch
BF FPGA
Control
signal
DATA
DSP
FPGA
High-voltage control signal
IQ
DDR2
UPLOAD
DDR2
PCIE X1
CPLD control signal
Rx9~
16
Rx49~64
or
Tx25~32
Rx1~8
...
LVDS*8
Tx9~16
Tx49~64
or
Tx25~32
Tx1~8
Probe board
PCIE
PHY
Communication
control bus
Control
signal
...
High-
voltage
pulse output circuit
Drive
circuit
Tx
TGC
8-channel digital
LNA VGA ADC
Rx
8-channel analog
Control signal
Control signal
CPU
module
Clock circuit
Clock
Clock
Figure 4-4 Principle Diagram of Front-end of the Main Board
As shown in the figure above, ultrasound front-end internal structure consists of ultrasound receiving module(Rx1~Rx64), ultrasound transmission module(Tx1~Tx64), high-voltage isolation moduleHVISO, BF FPGA module, DSP FPGA module ,PCIE_PHY module and so on.
Ultrasound receiving moduleRx1~Rx32, ultrasound transmission moduleTx1~Tx32, high-voltage isolation moduleHVISOof Z5 series are half of Z6 series.
Function describes as following:
Support 32/64-channel high-voltage isolationHVISOto protect ultrasound receiving
moduleRxto avoid ultrasound transmission high-voltage contamination.
32/64-channel transmission wave forms according to scanning time sequence and control
parameters.
32/64-channel transmission wave forms to 32/64-channel high-voltage transmission wave
by driver.
Echo receiving LNA is to enlarge fixed gain. By the control of TGC, echo receiving carries out variable gain amplification by depth. 32/64-channel analog echo could be switched to 32/64-channel digital signal by ADC. BF FPGA combines 64-channel echo signal to output to DSP FPGA, and DSP FPGA is
sent to signal processing to a single image which is uploaded to CPU module by PCIE_PHY, then CPU module performs post processing.
PCI_PHY implements PCIE interface switching. BF FPGA can communicate with the CPU module through transferring DSP FPGA.
Hardware Principle 4-5
The CPU module communicates with the probe board via BF FPGA.
IO board
PCIE PHY
PCIE X1
HDD
VGAUSB
Video S-Video Remote
Ethernet
Multifunction
FPGA
LPC
LVDS
USB
CPU Module
SATA
+12V
LVDS
Display
SMBUS
Control signal of the inverter
Video
Encoder
Control
USB
Lateral
USB
System monitor curcuit
Power Supply Module
ARM control signal
HDD Indicator
Control signal of parameters board
Control panel
Main Board
Status indicator
5V
12V
3.3V
POWER
12V
5V
5VSTB
Status indicator
Power_BTN_N
Closed status indicator on the control panel
ARM
Power
Management
FPGA
Control signal
Power on/off signal
Data
Audio
Codec
Audio
Power
amplifier
Speaker
4.2.2.2 Back-end of Main Board
Figure 4-5 The Schematic Diagram of Back-end of Main Board
Back-end of main board is the core of back-end ultrasound system, which mainly includes back-end of main board, CPU module, IO interface board, monitor assembly and control panel. Mainly implements display interface, user's operation interface, power module controlling interface and all kinds of Peripherals Supported interface.
Also provides system monitoring, running status of the monitor system, indicator of system status and running status of the indicating system via LED.
Function describes as following:
Back-end of main board connects to front-end of main board via PCIE PHY.
The USB on the lateral of Main board is the output of the CPU module.
The relevant signal in HDD of the main board is output of the CPU module.
Back-end of the main board MUL FPGA implements display ports, it also supports display
assembly, video and s-video interface of IO board.
Multifunction FPGA on the back-end of the main board is to implement the REMOTE port
4-6 Hardware Principle
on the IO panel.
The CPU module implements interfaces including USB, Ethernet and VGA on the IO
board.
System monitor circuit connects Multifunction FPGA though smbus.
Multifunction FPGA communicates with ARM by serial port to control transmission
high-voltage, and also implements ARM update by serial port.
FPGA on the power supply module implements the machine power on/off by the switch
Main Board
Video S-Video Remote
Multifunction FPGA
Video Encoder
Control
Data
VGAUSB Ethernet
CPU module
System Monitor
Circuit
Status Indicator
signal.
Standby indicator, AC present indicator and working status indicator on the control panel
are output by FPGA.
The control panel implements communication with CPU module by USB.
CPU output the audio signal to the speaker on the front cover of main unit, via audio coder
and audio amplifier.

4.2.3 IO Broad

Figure 4-6 Principle Diagram of IO Board
IO board is connected to the main board via connectors in order to implement the user interface function and indicate if the power is normal.
IO board supports Video and S-Video, and the two channels of signal get by Video
Encoder driven by Multifunction FPGA.
IO board supports REMOTE signal, which is obtained by Multifunction FPGA on the main
board.
USB signal, Ethernet signal and VGA signal on the IO board are obtained by CPU module
output.
Status indicators of IO board respectively refers to12V, 5V and 3.3V powers are normally
working.
Hardware Principle 4-7

4.2.4 Ultrasound System Monitor

CPU
module
Multifunction
FPGA
System monitor
curcuit
LPC
SMBUS
Temperature
sensor
Main unit front-
end FPGA
temperature
D12V
D-12V
D-5V D2.5V D1.8V D1.5V D1.2V
Voltage of
clock
battery
Back-end of main board
CPU module
temperature
Main unit fan
Power supply
module fan
Power supply
module
Figure 4-7 Principle Diagram of Ultrasound System Monitor
Ultrasound system monitor is mainly to monitor every voltage of interior machine and
temperature.
Function: On the back-end of the main board, a special monitor IC is used to monitor signal voltages
and clock battery power that are provided to the main board by the main unit box power supply module.
The system temperature is detected in three positions: BF FPGA, CPU module and power
supply module.
Monitoring chips control the main unit fan and power supply module fan, and then read the
fan rotating, which send to Multifunction FPGA by smbus, and to PC module by LPC bus.
4-8 Hardware Principle
CPU
Module
Main board
Power status
indicator
Power-on status
indicator
Battery status
indicator
AC supply status
indicator
Standby status
indicator
Power
management
FPGA
HDD indicator
driver
5VSTB_CPU
Control panel
+12V
+5V
3.3V
ARM
driver
Power supply
module
IO Board
driver
driver
driver

4.2.5 Ultrasound System Indicator

Figure 4-8 Principle Diagram of Ultrasound System Indicator
Function describes as following:
There are four power status indicators on the IO panel: three are used for indicators of
+12V, +5V and 3.3V output on the DC-DC power board, the other one is used for reserved indicator.
There are five indicators on the control panel, in which power-on status indicator is below
the power button.
AC power-on status indicator control is output by the power management FPGA on the
power module.
The hardware indicator control is output by CPU. The status indicator of battery are controlled by the battery management ARM on the
power module which driven by 5VSTB on the main board.
The status indicators of power-on and standby are both generated by the battery
management FPGA in the power module and driven by 5VSTB on the main board.
Hardware Principle 4-9

4.2.6 Display

LVDS3.3V
Display
High-voltage
LCD
screen
Parameter
board
I2C
3.3V
Inverter board
ON/OFF
12V
Brightness
control
Main board
IO board
12V 3.3V
Multifunction FPGA
LVDS
I2C
triode triode
Control signal
of switch
Brightness
PWM
ON/
OFF
Brightness
control
Figure 4-9 Principle Diagram of Display
Monitor mainly consists of the inverter board, LCD and the parameter board. Function describes as following: The inverter board produces high voltage to illuminate backlight of LCD which is adjusted
by signal control switch and brightness of the IO board output.
Color temperature and Gamma correction data memory in parameter board used matches
LCD so that the corresponding display effect shall be consistent.
Multifunction FPGA generates switch control signal which drives triode to control inverter to
open or shut off high-voltage output.
Multifunctional FPGA generates PWM signal, which drives triode to control output
high-voltage of the inverter to adjust the brightness of the display.
Multifunctional FPGA outputs LVDS digital signal of display, which is sent to screen directly. Multifunctional FPGA communicates with parameter via 12C signal and supports to adjust
parameters of display.
4-10 Hardware Principle

4.2.7 Control Panel

5 status
indicators
TGC board
Trackball
Functional
button
Single-
encoder
board
Key
scan
LED
backlight
driver curcuit
LED
backlight
USB
Interface
chip
ADC curcuit
Buzzer
Buzzer driver curcuit
Main
control
Trackball
scan
TGC scan
PWM generating
curcuit
FPGA
Control panel PCBA
Main
board
USB
+12V
+5V
Double­encoder
board
Figure 4-10 Principle Diagram of Control Panel
The control panel PCBA is the core of the control panel which comprises control panel PCBA,
trackball, TGC board, encoder board and buzzer.
Function describes as following:
FPGA is regarded the control panel as a main controller , key, single/dual-encoder,
trackball and TGC.ect are basically scanned by FPGA; Key LED backlight and buzzers are both controlled by FPGA, and only when FPGA is normal can key LED backlight be on.
Five indicators on the control panel are directly drived by interior of main unit. The control panel communicates with the CPU module by USB ports.
Hardware Principle 4-11

4.3 Power System

AC Input 100-240V~,50/60Hz
Battery
14.8V6600mAh
DC Output 19V/3.95A
Power supply module
Power
management
FPGA
DC-DC power board
AC-DC power board
PWR_BTN_N
Main board
Control
panel
System monitor module
Temperature signal
CPU
module
ARM
Multifunction
FPGA
ARM control signal
POWER
I2C
PHV
AD DA
Closed signal of the control panel
AC in place
Power on/off control signal
Status indicator
control signal
12V5V
battery in place
Fan
Fan monitor
12V
Display
12V
Battery
connecting
board
I2c
battery in place
14.8V 6600mAh
Status indicator
Restart automatically
No.
Power
description
Support circuit module or function
Remarks
1
+12V
CPU module, Display, Fan, Control panel
2
-12V
No use
3
+5V
HDD, Control panel, Front-end of main board, Probe board
4
-5V
Front-end of main board
5
5Vst
CPU module
6
VDD
Front-end of main board, Back-end of main board, Display
Figure 4-11 Block Diagram of Power System
The power supply configuration of the system describes as following: The power supply module can be divided into AC-DC power board, DC-DC power board,
battery and battery connecting board.
High-voltage programmable ARM and power management FPGA are on the DC-DC power
board.
Power output by the power supply module is transmitted to main board, and then it is output to
control panel and display by the main board.
Battery is connected to the DC-DC board via the battery connecting board and connecting
cable of BTB sequentially.

4.3.1 Power Output of the Power Supply module and Supporting Function Distribution

4-12 Hardware Principle
7
1V2
Back-end of main board (FPGA core voltage)
8
1V8A
Front-end of main board (receiving chip)
9
1V8
Back-end of main board ( video encoder, FPGA IO voltage , DDR2)
10
2V5
Back-end of main board ( video encoder, FPGA IO voltage)
11
+95V
Probe board (high-voltage switch)
12
-95V
Probe board (high-voltage switch)
13
+PHV
Transmitting high-voltage
14
-PHV
Transmitting high-voltage

4.3.2 System Power-on Control

CPU
module
PWR_BTN_N
D12V
SUS_S3_N
SUS_S4_N
CPU_PWR_BTN_N
PWR_OK_N
5VSTB
Power
manag
ement FPGA
AC-DC
Power
board
12V
AC in place signal
DC-DC power board
Main board
Control
panel
PWR_12V_EN
PWR_5VSTB_EN
PWR_5VSTB_OK_N
PWR_12V_OK_N
AC_IN_STATUS_N
No
Controlling signal
Description
Remarks
1
PWR_BTN_N
The pulse generated by the power button on the control panel is sent to CPU module through power management FPGA, to power on the system.
2
SUS_S3_N
CPU module output effectively represents that CPU system has been in the standby status, when 5VSTB which is controlled by power management FPGA is in the charging status.
3
SUS_S4_N
CPU module output effectively represents that CPU system has been in the dormancy status.
4
PWR_OK_N
The output of power supply module is transmitted to the CPU module by power management FPGA, which represents the +12V output of power supply module has been powered on.
Figure 6-12 Diagram of System Power-on Supported Circuit
Related controlling signal comment:
5
CPU_PWR_BTN_N
Power management FPGA is transmitted to CPU, which notifies CPU power on or power off.
6 PWR_12V_EN
The output of power management FPGA is transmitted to the power supply module, which is used to control other power of the DC-DC power board except 5VSTB and 3V3STB to be power-on..
7
PWR_5VSTB_EN
The output of power management FPGA notifies power to provide 5V standby power.
8 PWR_12V_OK_N
Power status signal; indicate 12V power has been provided steadily.
9
PWR_5VSTB_OK_N
Power status signal; indicate 5VSTB power has been provided steadily.
When battery and AC supply, the power supply module always maintains the 3V3STB output,
and supports power management FPGA of the power module to be standby status.
No matter the system is on when AC supply, Power management FPGA will always control
DC-DC power board to output 5VSTB via PWR_5VSTB_EN_N However, when powered by battery, power management FPGA could control DC-DC power board to output 5VSTB though PWR_5VSTB_EN_N if only users turn on the system via power button of the control panel.
The detailed flow of power-on is shown as follows:
4-14 Hardware Principle
Pressing power button
(Control panel)
Is AC in place?
(Ppower management
FPGA)
NO
Receiving PWR_BTN_N effective
(Power management FPGA)
PWR_5VSTB_EN_N is valid (Power management FPGA)
5VSTB start-up
(Power supply module)
PWR_5VSTB_OK_N is valid
(Power supply module)
CPU_PWR_BTN_N is valid
(Power management FPGA)
SUS_S3_N and SUS_S4_N
is invalid(CPU module)
PWR_12V_EN is valid
(Power management FPGA)
PWR_12V_OK_N is valid
(Power supply module)
CPU start-up
(CPU module)
12V and +5V start-up
(DC-DC power board)
PWR_OK_N is valid
(Power management FPGA)
YES
Indicating 12V output of power module is normal
Indicating 5VSTB_CPU output of power module f is normal
Figure 4-2 System Power-on
5 Function and Performance
Checking Method

5.1 Instruction

The chapter supplies the method to verify main function and performance of product. This is only used for reference, not preventive execution.
Function checking and testing of this part shall be carried out by Mindray service engineers and the user together.

5.2 Checking System Status

5.2.1 System Running Status

1. The ultrasound system can be turned on or off normally, working normally, no any abnormal noise and phenomenon appear when the system is running.
2. The fan starts to work as soon as the system is turned on, no any abnormal noise appear in process
3. Check basic system information such as product configuration, options, software version, etc. Confirm all the information is normal.
4. Check the preset of the monitor, such as contrast and brightness.
5. Check if the time and date are correct, if not, please preset them again.
6. Check the log together with the user to confirm any abnormality or occasional abnormalities when the system is running.

5.2.2 System Running Status

Check the ambient temperature and humidity. The measurements related to safety features are particularly sensitive to humidity. If the insulation feature of the system deteriorates due to the increase of system service time or system malfunctions, the fluctuation range of measurement results are likely to increase with the increase in humidity.
Function and Performance Checking Method 5-1

5.3 General exam

Control panel
Display
DVD-R/W
Peripheral devices
I/O ports
Procedure
Standard
Check the functions of all keys and knobs Follow the direction: left to right, and up to down.
All keys and knobs are effective.
Function checking of the trackball Press the <Freeze> key to enter the Freeze
status.
Press <Measure> to enter into measure
status, do vertical and horizontal measurement, or do other trackball operations.
The trackball can be rotated easily and smoothly, the cursor responds sensitively, the rotation direction is the same as the direction of the cursor.
Procedure
Standard
Adjust LCD brightness Adjust LCD contrast Display color temperature
adjustment
Click “Cold” or “Warm” in the
[General] page in System Preset screen.
Press “<Fn>+< >”, the brightness increases; and press
“<Fn>+< >”, the brightness decreases.
Press “<Fn>+< >”, the contrast increases; and press
“<Fn>+< >”, the contrast decreases.
The LCD color temperature changes correspondingly. Click each functional button, the LCD responds correctly,
the standard is as follows

5.3.1 Check Flow

5.3.2 Checking Content

5.3.2.1 Check Control Panel
5.3.2.2 Check the Display
5-2 Function and Performance Checking Method
LCD maintenance
Log in as “Service”, click
[Maintenance] in the Setup menu, then click [Monitor Test] to go for the monitor test.
1 .Light-spot: 0; flash point: 0.
2. The adjoining dark dots are no more than 3 pairs, and there is no adjoining dark dot in image area.
3.There is no adjoining dark dots of 3 or more than 3
4.The dark dots are no more than 7 and those in the image area are no more than 2
5. The distance between bad dots is no less than 5mm. Note: image area refers to the red the background is
black/right.
5.3.2.3 Check DVD-R/W
Procedure
Standard
DVD-R/W Drive
Check if connection is normal Press <Eject> button on the drive Perform DVD reading and writing.
The DVD-R/W driver can be ejected smoothly. Data reading and writing performs normally
without abnormal noise.
Procedure
Standard
Footswitch:
Connect the footswitch connection, and check the functions of footswitch according to the functions listed in Key Config.
Press the freeze key (the right key), image is
frozen, the freeze menu is displayed; press the key again, image is unfrozen.
Press the key (middle key), color printing starts Press the print key (left key),B/W printing starts
Video Printer:
Check if the video printer and ultrasound system are correctly connected. Then check the function of each key.
Press <Print> key, the printer begins to work, no
image print deficiency or degradation.
Switch video output port Press <Print> key, the printer begins to work, no
image print deficiency or degradation.
Graph/ text Printer:
Check if the graph/text printer and ultrasound system are correctly connected. Then check the function of each key.
Press <Print> key, the printer begins to print, no print deficiency or degradation.
bar Code scanner:
Perform code bar scanning when the ultrasound system is running normally.
The bar code is correctly displayed on the screen.
Procedure
Standard
Checking IO ports: Video/audio interface, USB
The contents displayed on the VGA/LCD are the same as
5.3.2.4 Check Peripheral Device
5.3.2.5 Check IO Ports
Function and Performance Checking Method 5-3
port have been verified, remains are:
VGA port Connect VGA display
Network port USB port
those displayed on the ultrasound system display, no character and image loss, no color difference, no fluttering and flicking.
Smooth communication in both network and connection. USB port data storage/accessing are normal.

5.4 Function Checks

NOTE:
A complete function inspection is described here, do the checking according to the actual system configuration.
Check each imaging
mode
Check the measurements
Check in cine mode
Check the probe
application
Image & video manage
Record and save the
exam

5.4.1 Check Flow

5-4 Function and Performance Checking Method

5.4.2 Checking Content

Display
F
D
G
FR
DR
Image Parameters
Frequency
Depth
Gain
Frame Rate
B Dynamic Range
Adjustment
Items
Control Panel
Gain, Depth, TGC, iTouch, Focus, Frequency
Menu
Frequency, Acoustic power Line Density,ExFOV ,iClear, H Scale, Focus Position, Focus Number, Persistence, iBeam, Dynamic Range, TSI, Gray Map,Tint Map,L/R Flip, U/D Flip, Gray Invert ,, , , Lithotrity,FOV, iTouch, Biopsy Kit.
Procedure
Standard
Press <B>
Enter B mode, and B image displays
Gain adjustment. Rotate <Gain/iTouch>
Rotate clockwise to increase Rotate anticlockwise to decrease The adjusting range is 0-100.The real-time value will be displayed in the
image parameter area in the upper right corner of the screen.
Depth Adjustment Press <Depth/zoom> to
light on the Depth indicator, rotate the knob
Rotate clockwise to increase Rotate anticlockwise to decrease The adjustable depth values vary depending upon the probe types.
TGC adjustment Adjust through the
8-segment toggles
Push the toggle to the right to increase the corresponding area brightness
Push the toggle to the left to decrease the corresponding area brightness
About 1.5s after the adjustment is finished, the TGC curve disappears.
5.4.2.1 B Mode
In B Mode scanning, the image parameter area in the right corner of the screen will display the
real-time parameter values as follows:
Parameters that can be adjusted to optimize the B Mode image are indicated in the following.
1. Control Panel
Function and Performance Checking Method 5-5
2. Menu
Procedure
Standard
Acoustic power B mode menu
[A.power]
Click [A.power] and rotate the multifunctional knob to adjust. AP value is displayed in real time in the upper part of the screen. You should perform exams according to actual situation and follow the
ALARA Principle.
Focus B mode menu
[Focus Number]
Click [Focus Number] and rotate the multifunctional knob to adjust (B mode image has max. 4 focus )
The focus position symbol is displayed on the right side of the image.
FOV (Field of View) B mode menu->
[FOV]
Click [FOV] and rotate the multifunctional knob to adjust
Dynamic Range B mode menu->
[Dynamic Range]
Click [Dynamic Range]; rotate the multifunctional knob to adjust, the real time value will be displayed on the image parameter area in the upper right of the screen. As the dynamic range increases, the darker the image and the contrast, as well as the noise may increase.
Line Density
Click [Line Density], and rotate the multifunction knob to adjust the
Focus Position Adjustment
Press <Focus/Freq./THI.> to light on the Focus indicator, rotate the knob
The focus position will change in correspondence with the knob rotates.
Frequency adjustment Press
<Focus/Freq./THI.> to light on the Freq./THI indicator, rotate the knob
Rotate clockwise to increase Rotate anticlockwise to decrease “FR” will be changed relevantly.
iTouch Press <Gain/iTouch>
iTouch sign will be displayed in the image area to activate image effect auto optimization.
Click [iTouch] on the image menu to adjust gain value in iTouch mode. Click [iTouch Brightness] on the image menu to adjust brightness in
iTouch mode.
Image Magnification Press <Depth/zoom> to
light on the Zoom indicator, rotate the knob
Rotate clockwise to zoom in the image (max. magnification factor is 10); rotate anticlockwise to zoom out the image.
Roll the trackball to change position of the magnified image. Press <Zoom> in zoom status to exit the mode, the current window
returns to the display before zoom.
Press <B>, and then press
Enter dual mode, and the right image is activated  Press again to switch between the windows
Press <B>, and then press
Press 4 times, and 4 images will be displayed on the screen with only
one image activated at one time
Press again to switch among the windows
Press <B>
To enter single mode in multiple window mode, or to exit from other modes.
5-6 Function and Performance Checking Method
B mode menu-> [Line Density]
parameter. The higher the line density, the higher the resolution, and the lower the frame
rate.
iClear B mode menu->
[iClear]
Click [iClear] to adjust, off represents no iClear is turned on, and the bigger the value the stronger the effect. The bigger the value the more clearly the profile of the image.
B mode menu-> [U/D Flip]/[L/R Flip]
B mode menu-> [U/D Flip]/[L/R Flip] Click [Rotate] to adjust.
5.4.2.2 M Mode
Display
V
IP 6
DR
G
F
D
Image Parameters
M Speed
M IP
M Dynamic Range
M Gain
M Frequency
Depth
Adjustment
Items
Control Panel
Frequency, Gain, TGC, Depth
Menu
Frequency, Acoustic Power Focus Position, Edge Enhance, M Soften, Speed, Dynamic Range, Gray Map, Tint Map, Display Format.
Procedure
Standard
Press <M>
Press <M> on the control panel, and roll the trackball to adjust the sampling line.
Then press [M] again.
Press [M] on the control panel again to enter M Mode, then you can observe the tissue motion along with anatomical images of B Mode.
Press <Update>
To switch between the active B image and frozen B image.
Tips:
a) Adjustment of the depth, focus position or TGC to the B Mode image will lead to
corresponding changes in M Mode image.
b) For details of other control panel adjusting parameters, please refer to descriptions in B
mode
In M Mode scanning, the image parameter area in the right corner of the screen will display the
real-time parameter values as follows:
Parameters that can be adjusted to optimize the M Mode image are indicated in the following.
1. Control Panel
2. Menu
Function and Performance Checking Method 5-7
Procedure
Standard
Speed B mode menu-> [Speed]
Click [Speed], and rotate the multifunction knob to adjust the parameter.
The lower the value the faster the refreshing.
Display Format M mode menu-> [Display Format]
There are 4 formats available for image display: L/R, V1:1, V1:2, Full.
Tips:
a) During M Mode scanning, frequency and acoustic power of the probeare synchronous with
that of B Mode.
b) Refer to B mode for more details.
5.4.2.3 Color Mode
Display
F G WF
PRF
Image Parameters
Frequency
Color
Gain
Color Wall Filter
Pulse Repetition Frequency
Adjustment
Items
Control Panel
Color GainFrequency、Depth、Steer, Scale
Menu
FrequencyAcoustic Power, Persistence, Line Density,Packet Size, Priority , B/C Align, Smooth, Wall Filter,Scale, Baseline,Invert,Color Map
Dual Live
In Color Mode scanning, the image parameter area in the right corner of the screen will display
the real-time parameter values as follows:
Parameters that can be adjusted to optimize the Color Mode image are indicated in the
following.
5-8 Function and Performance Checking Method
Display
F
G
WF
PRF
SVD
SV
Angle
Image Parameters
Frequency
PW/CW
Gain
Wall
Filter
Pulse
Repetition
Frequency
Sample Volume
Depth
Sample Volume
Size
Angle
Adjustment
Items
Control Panel
GainFrequency
Menu
Frequency, Acoustic Power ,Scale, Wall Filter, Speed, Duplex/Triplex, SV, Quick Angle, Angle, Invert, Baseline, Trace Area, T/F Res, Audio, Auto Calc, Auto Calc Cycle, Auto Calc Para, HPRF, Dynamic Range, Gray Map, Tint Map, Display Format
Procedure
Standard
Press [Freeze] key to freeze an image, and
the [Cine] key indicator lights on. The system automatically enters the manual cine status.(Precondition: set “Status after Freeze” to be “Cine”)
Press iStation, click [Review] or press
<Review> to open the cine file
Cine Review status
To enter cine auto cine review status.
Roll the trackball
Manual cine review
Click image menu->[Auto Play]
Activate auto cine play status
Manually review the images until the frame which
you want to set it as start point, and then click [Set
First Frame] to set a start mark.
Set first frame:
Manually review the images until the frame which
you want to set it as end point, and then click [Set
End Frame] to set an end mark.
Set end frame Click [Auto Play]
The cine is played within the setting region.
Then press the [Cine] key again.
Cine review stops
Press the <Freeze> key to defreeze the
image.
<Freeze> backlight is off, the image
returns to the scanning process and exits
5.4.2.4 PW Mode
In PW Mode scanning, the image parameter area in the right corner of the screen will display
the real-time parameter values as follows:
Parameters that can be adjusted to optimize the PW Mode image are indicated in the following.
5.4.2.5 Cine Review
Function and Performance Checking Method 5-9
Press <ESC> or <Cine>
cine review.
The images are still frozen but the system
exits cine review.
5.4.2.6 Measurement
Procedure
Standard
In B mode
Press <Measure>: Press <Caliper> key
Enter the application measurement mode. Enter the general measurement mode.
Click any 1-2 measurement items, the measurement results will be displayed below the image
Press <ESC> or the same key again.
Exits measurement.
Perform similar operations in other modes
Application measurement functions are related to certain application software packages.
Procedure
Standard
Press <Save1/2> in image scanning process (preset already).
The image will be saved to corresponding patient database and the saving icon will be displayed in the right of the screen.
Press <Review> Click [Exit] on the Review
screen; or press <Review> or <ESC>.
To enter Review: To exit Review:
Press <iStation> to enter Patient information management (iStation page)
Saved images and information for the patient can be checked here, and you can:
Backup/ Restore Send to (DICOM or network storage)
5.4.2.7 Patient Information Management
5-10 Function and Performance Checking Method

5.5 Performance Test

Resolution
Detection depth
Geometric Position
Accuracy
Blackout area test
Record the exam
images
NOTE:
The image used here is only for reference, stick to the image effect in the real situation.

5.5.1 Test Process

5.5.2 Test Content

Requirements:
1. Display: set the contrast and brightness at the clinical application value (or the default status)
2. Operation environment: dark room, simulating the clinical application environment.
3. Scanning techniques: contact the probe with the acoustic window of the phantom, no spacing nor pressing.
Tips:
For the testing phantoms, please refer to Appendix B.
5.5.2.1 Resolution
transverse resolution Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the probe with the scan surface, making the transverse resolution testing targets to be displayed around the midline of the image.
2. Adjust the focus point focuses at the position where the transverse resolution testing targets are displayed.
3. Adjust parameters like gain, dynamic range, TGC, making the background tissue unseen, just displaying the target image clearly.
Function and Performance Checking Method 5-11
4. In condition that the transverse resolution testing targets are horizontally displayed, record the minimal distance of two targets that can be clearly recognized.
5. Repeat the operation above for the transverse resolution testing targets at other depths.
As shown in figure below.
Axial resolution Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the probe with the scan surface, making the longitudinal resolution testing targets to be displayed around the midline of the image.
2. Adjust the focus point focuses at the position where the longitudinal resolution testing targets are displayed.
3. Adjust parameters like gain, dynamic range, TGC, making the background tissue unseen, just displaying the target image clearly.
4. Record the minimal distance of two longitudinal resolution testing targets that can be clearly recognized.
5. Repeat the operation above for the longitudinal resolution testing targets at other depths.
5-12 Function and Performance Checking Method
NOTE:
1.
When using the convex probe, keep the transverse resolution testing targets to be displayed near the midline.
2.
When using a linear probe with steer function, do not turn on the steer function when perform the transverse resolution test.
3.
Zoom in the region where the targets located if necessary.
4.
The diameter of the target point at a certain depth is equal to the transverse resolution at the depth.
NOTE:
1.
Increasing the gain will also increase the noise, and echo may be covered.
2.
When using a linear probe, please completely contact the probe with the scan surface, no side clearance is allowed.
3.
When using a convex or phased-array probe, make the axis targets to be displayed at the middle of the scanning image.
4.
When system is not frozen, the fast field target information may be similar to that of the noise, do not use this target.
5.5.2.2 Maximum Depth
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the probe with the scan surface
2. Set the system display depth according to the expected maximum available depth of the probe in use.
3. Adjust the focus point to the deepest, and AP at the maximum value.
4. Set gain, contrast, TGC at a greater value, but no halation nor defocus is allowed.
5. Record the depth of the furthest target (the target can be seen clearly).
Function and Performance Checking Method 5-13
As shown in figure below.
NOTE:
1.
The measurement caliper should be positioned at the upper edge of the target, not the middle nor the lower edge.
2.
The scanning plane should be vertical to the target line, that means the scanning plane is parallel with the cross-section of the phantom
5.5.2.3 Geometric positioning accuracy
Longitudinal geometric positioning accuracy Test Step:
1. Do adjustments as the way in testing the maximum depth.
2. Record the distance by 20mm each segment on the longitudinal targets line using the measurement caliper;
3. Select the value with the greatest error (to 20mm), calculate the accuracy using the formula below
As shown in figure below.
5-14 Function and Performance Checking Method
NOTE:
1.
When using a linear probe, record the transverse distance by segment.
2.
When using a convex probe, all transverse targets should be displayed integrally in an image.
3.
The measure caliper should be posited at the upper side or lower side of the target center.
Transverse geometric positioning accuracy Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the probe with the scan surface
2. Adjust the depth, making the transverse targets to be displayed in the image.
3. Adjust the focus point to be posited beside the transverse targets (the standard is not clear)
4. Adjust parameters like gain, TGC, making each transverse targets to be clearly displayed.
5. Record the distance by 20mm each segment on the transverse targets line by using the measurement caliper
6. Select the value with the greatest error (to 20mm), calculate the accuracy by using the formula below
As shown in figure below.
Function and Performance Checking Method 5-15
NOTE:
1.
When using a linear probe, please completely contact the probe with the scan surface, no side clearance is allowed.
2.
For convex probe, the targets in the blackout area should be positioned on the midline of the scanning plane.
5.5.2.4 Blackout Area
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the probe with the scan surface
2. Adjust the depth at a lower value, and set the focus at the nearest place to the scan surface.
3. Decrease the value of parameters like AP, Gain until the background noise just can be seen.
4. Record the smallest depth of the target that can be seen clearly, that value is the blackout area value.
As shown in figure below.
5-16 Function and Performance Checking Method
6 Software Upgrade and
NOTE:
Log on the system with the identity of Service before perform system maintenance.
Maintenance
WARNING: DO NOT directly remove a USB memory device; otherwise, the USB
memory device and / or the system may be damaged.

6.1 Enter the Maintenance Window

To log on the system:
1. When access control function has not been activated: press “Ctrl+/” to show the Login dialogue box, and then select the Service as the user name.
2. When access control function has been activated already: press “Ctrl+/” when the Login dialogue box is displayed, and then select the Service as the user name.
3. Press <Setup> key to open the Setup menu, click [Maintenance] and then select the target items to perform the maintenance respectively.
Software Upgrade and Maintenance 6-1
WARNING:
1.
Back up the patient data and preset data before restoring the system.
2.
Do not cut off, shut down or restart the system during system restoring.

6.2 System Software Installation/ Restoration

6.2.1 Restore the Operating System and Doppler

6.2.1.1 Preparation before Restoring
1. Back up preset data and patient data.
2. USB disk requirement:
a) Volume of USB disk should be more than 2G. b) The system supports only FAT or FAT32 format USB disk. c) If there are many drives in USB disk, installation may fail. d) Disconnect the extra USB devices to the machine.
6.2.1.2 Method
The operation system (“OS” for short) adopts Linux kernel and uses USB disk to install. You
can install the ultrasound software to the hardware when installing the operation system.
1. Copy files of the system restoration package (including operation system and Doppler) to USB disk root, and make sure that all the files of Doppler restoration package are within the Doppler file.
6-2 Software Upgrade and Maintenance
2. Insert the USB disk to the ultrasound machine.
Option
Display
Note
1
1.Install system with keeping
user’s data
Start the OS restoring directly, user data will be kept when finished.
2
2. Install system without
keeping user’s data (format
all disk)
The HDD will be formatted and all data will be erased at first, then start the OS restoring.
3
3.Human version converts to vet version, or vet version converts to human version
It is only used for implementing the conversion between human and vet version and all data will be erased (only applied for service restoration package)
q
quit
Exit, The system does nothing and restarts.
NOTE
1.
The restoration package must be copied to USB disk root, or else it may cause to failure of restoration;
2.
Please check all the files in the restoration package when copying, or else copy the package again.
3. Power on the machine, press the button to start, and press <F1> immediately, and when the "Mindray eRescue” displays on the screen, stop pressing F1 to enter system installation program.
4. Select the installation operation to enter. Enter the corresponding method (1, 2,3 or q) and press <Enter>.
5. Screen prompts:
If you enter q, the screen displays as below:
If you enter 1, the screen displays as below:
Software Upgrade and Maintenance 6-3
Note
Please select the method “3” to install the operation system only when human version
converts to vet version or vet version converts to human version.
If you enter 2 and press <Enter>, the system will ask the user to confirm if to delete all data,
enter “Y” and the system performs formatting and then goes to installation.
If you enter N here, the installation is stopped, and the system remains unchanged.
If you enter 3 and press <Enter>, the system will implement the conversion between human
and vet version.
The system will check if the conversion is passed before installing. Only when the service restoration package type named “Doppler” in the U disk which is different from the model (for example, human model with vet restoration package or vet model with human restoration package), the conversion can be passed and installation will continue.
If conversion is passed, the system will ask user to confirm again.
6-4 Software Upgrade and Maintenance
Other steps are the same as method “2”. When enter 3 and press <Enter>, if the model type is the same as service restoration package (for
example, human model with human restoration package or vet model with vet restoration package), the conversion will be failed .System installation will stop without any change.
6. After the system has been installed successfully, the following information will display on the
screen, press <Enter> to restart the system.
7. After the OS is restored, Doppler restoration starts automatically. The process of restoring refer
to 6.2.2 Doppler restoration (from the 5th to the 8th step)
6.2.1.3 File Processing during Restoration
During the OS installation procedure, the machine will back up the key file in the HDD
automatically, see below:
Select the choice 1:
1. Back up the option key file of the C drive to D/mnt/bkcfg (temporary catalogue)
2. Format the C drive and restore the OS.
3. Copy the option Key file to the previous catalogue in C drive, and the temp file will be deleted.
Select the choice 2:
1. Back up the option key file of the C drive to U/bkcfg (temporary catalogue)
2. Format the whole HDD and restore the OS.
3. Copy the option Key file to the previous catalogue in C drive, and the temporary file in U disk will be deleted.
In the OS with previous configuration data, if the ultrasound system is restored by package without configuration data, the product configuration and the option state will take effect as the configuration data in OS.
Software Upgrade and Maintenance 6-5
6.2.1.4 Software Introduction
NOTE
Please check all the files in the restoration package before starting restoring, or else copy the package again.
NOTE:
1.
Before executing the restoration package, please back up the patient data as well as the user-defined preset data.
2.
When using the restoration package, key catalog in the HDD will remain the same, and after restoration, the option key can take effect directly.
Operating System restoration package includes the files as follow:
Doppler: Doppler software restoration data will be placed here. If the data exists, ultrasound software will be restored after the operation system is restored; otherwise, only the OS will be installed.
Sys: operation system data is placed here. Mrcs.sh: installation program. Version.txt: version file.

6.2.2 Doppler Restoration

6.2.2.1 Using Package in Doppler
1. Copy the Doppler restoration package to USB disk root (USB disk requirement is the same as
in6.2.1.1)
2. Insert the USB disk to USB interface in the ultrasound system, the USB icon displays on
the right corner of the screen.
3. Press <Setup> to enter [Setup] menu, Click [Maintenance] to enter the [Other] page, Select
[Recover] in the Maintenance menu and press <Set>.
4. Switch to the USB catalog in the popped up dialogue box, and select boot_doppler.sh file in the
package and click [OK].
6-6 Software Upgrade and Maintenance
NOTE:
The old user preset data may be not compatible with the restored system, if
this happen, the operator must select “Y” when the system asks whether to delete
user preset data or not, otherwise, the system can’t be restarted normally.
5. The system then will ask to restart the system. Click [OK] to enter restore procedure, click
[Cancel] will cancel the operation and returns to Doppler.
6. During the course, when the restoration is almost done, the following prompts will display:
Enter “Y” or “y” to delete the data in D:\z5\Preset\Enter "N" and the system will not delete data.
7. After the whole system is restored, if there is only green shutdown prompts displayed, the
restore process is normal, and press <Enter> to quit.
Software Upgrade and Maintenance 6-7
NOTE:
If red prompts displays, please check the failed items.
No.
Regulation Type
Applicable for
1.
CE
European region
2.
FDA
North America
3.
SFDA
China
Check if every item is restored successfully, press <Enter> according to the prompts, and the
machine shut down automatically.
8. After restoration and restart, please check software version information in the path:
<Setup>-[About]
6.2.2.2 product_info.ini Description
product_info.ini is used for configuring the regulation type (SFDA, FDA, CE).
If OS and Doppler are restored together, the regulation taken effect will be the default type in the package after the restoration finished. If the regulation type is not the same as the area the system to be used, please change the regulation type as the procedures below:
1. Open product_info.ini in notepad or WordPad in the package.
2. If the default regulation type is not the same as the area that the system to be used. Change
the regulation type, e.g., change “SFDA” into “FDA”, save it and then exit.
6-8 Software Upgrade and Maintenance
NOTE:
After the Doppler is restored, product configuration data will be overwritten by general configuration data in Doppler restore package, if necessary, please configure the product manually.
NOTE:
Just change the contents in the red box, otherwise may cause an error.
Type
File
Suffix
Process
Destination
HDD
data
APD
Analyze the update file; replace the old data with the new data according to the update item.
C:\z5\app
EXD
C:\z5\exe
VDD
C:\z5\video
GUD
C:\z5\gui
IMD
C:\z5\Image
PST
C:\z5\Preset
PCD
C:\z5\config,
C:\usr\share\splashy\themes\default\po
weron.png and poweroff.png
Demo
Data
Replace the old demo with the new demo in the resume package.
E:\z5\Demo
Driver
DRD
Decompress the update file to the
temporary directory in the hard disk,
then install the driver,
Install the drivers
C:\z5\driver
FPGA
FPD
Decompress the update files to the temporary directory of the hard disk, and write data in their configured flash through the corresponding drive program. Delete the temporary directory when finished.
Configuration flash chip in the main
board in corresponding to logic.
Control
Panel
KUG
Decompress the update files and cache the keyboard data to the buffer, inspect keyboard communication and erase the flash.
Control Panel
Power
supply
monitor
ARM
Decompress the update file, save ARM data to the temporary directory in the hard disk, inspect the ARM communication and erase the flash, and then delete the temporary directory in the hard disk after the installation finished.
Power supply panel
6.2.2.3 Data Processing during Restoration
Data structure and location is as the following after Doppler package restoring.
Table 6-1 data processing list
Software Upgrade and Maintenance 6-9
6.2.2.4 Doppler Software Introduction
NOTE:
1.
You can only install one key at one time.
2.
The option uninstall function is only accessible to internal user, for service
engineer, you can log on the system as “Service” first, and then conduct
uninstall function. You can uninstall one key or more at one time.
3.
If select Trial, you need to input the password. Please contact service engineers.
Doppler restoration includes files as follow:

6.3 Installation of Optional Devices

1. Copy the optional devices to the U disk, then insert it to the USB port of the ultrasound system;
2. Enter [Preset], click [Maintenance] to enter [Option] menu to select the option module to be installed.
3. Select the corresponding key file in the prompted “Load File” dialogue box and click [OK].
4. After the Key file is successfully installed, you can see the option module in the “Installed” status. Corresponding functions are available after return from preset.
6-10 Software Upgrade and Maintenance

6.4 Data Backup and Storage

6.4.1 Manage Settings

Press <Setup> to open the Setup menu, click [Other] to open the Manage Setting interface. System preset can be performed here. Image parameters can be exported, imported or restored to factory.
6.4.1.1 Back up the Preset Data
1. On Manage Settings page, Click [Export] to open the [Export Data] dialogue box.
2. Select the path to save data.
3. Click [OK], a progress bar will appear and the preset data of the selected items will be exported to the specified path.
6.4.1.2 Restore the Preset Data
1. On Manage Settings page, Click [Import] to open the [Import Data] dialogue box.
2. Select the path of the preset data.
3. Click [OK], a progress bar will appear and the preset data will be imported to the specified path. Note: If selecting [Load factory]it will restore the system to the factory setting, but the
[Region],[Admin] , [Network Preset] and [DICOM Preset] cannot be restored.

6.4.2 Patient Data Backup and Restore

6.4.2.1 Patient Data Backup
1. Press [iStation] on the control panel to open the iStation dialogue box.
2. Click [Select All] to select all the data or select the target data one by one.
3. Select the data, click [Backup] to pop up the Backup patient Record dialogue box, select the target storage device (recorder or USB disk), click [Backup], the data will be backed up.
6.4.2.2 Restore Patient Data
1. Press [iStation] on the control panel to open the iStation dialogue box;
Software Upgrade and Maintenance 6-11
2. Select the drive which contains the patient data, click [Select All] to select all the data or select
NOTE:
[Config] is available on the Maintenance menu only if the operator logged on the system as Service.
NOTE:
After Doppler restoration, please do product configuration manually according to
the practical situation.
the target data one by one, and click [Restore] to restore the patient data from the current drive to the patient database.

6.5 Software Maintenance

6.5.1 Product Configuration

1. Make the product configuration filePCF file (generated by the production line or already archived PCF file) according to specific requirements and copy to the USB disk.
2. Connect the USB disk to the machine, click [Config] and load the file in the popped up dialogue box.
3. Select correct file (PCF format), and click [OK]. After the operation is completed, the successful prompt appears. Restart the machine according to the prompt.
Specific configuration: a) Model configuration b) Customized product logo, hospital logo, screen saver/hibernation icon, active/inactive
icon of dual B mode, system information, boot up/shut down graphics.
6-12 Software Upgrade and Maintenance

6.5.2 Log Maintenance

NOTE:
The log can be exported to the external USB storage device only, make sure the connection between U disk and ultrasound is normal before the exporting.
6.5.2.1 Export the Log
1. Connect the USB disk.
2. Click [Export Log] on the Maintenance menu.
3. Select the path in the Browse page to save the log, and click [OK].
4. When the log is exported, the system prompts “Export succeed! ” click [OK] to return to the Maintenance menu.
6.5.2.2 Log Uploading
1. Click [Upload Log] on the Maintenance menu, and select the log type in the sub-menu
2. Click [Upload] in the popping up dialogue, the system performs log uploading automatically. After the uploading is finished, the system will prompt “Upload succeed
Software Upgrade and Maintenance 6-13
NOTE:
1.
Before uploading the log, make sure the ultrasound system has been connected
to network, otherwise, the system may warn “could not connect to server”
2.
The server (smtp.163.com) is already specified by the system, the user doesn’t have to select it.
NOTE:
In normal condition, when changing display assembly or main board assembly, the
hardware will update the display data automatically. If synchronization fails, please load display parameters manually to assure image effect.
If uploading is failure or can’t find the server, the system will prompt information as below.

6.6 Display Parameter Setting

1. Log in as Service user.
2. Enter [Maintenance] menu in Setup menu. Then select [Test Main Monitor] to enter.
6-14 Software Upgrade and Maintenance
Blocks(G)
system
Notes
10
Linux
C:
263
Linux
D:
38
Linux
E:
3. Click the upper icon in the screen (prompts “Update” if you put the cursor on it), the update step
is carried out, if fails, please send to R&D center.
Figure 6-1 Main Monitor Maintenance Screen
4. After update is successful, restart the machine after power off to make the data effective.

6.7 HDD Partition

The storage capacity of the system HDD is 320G or more, and the partition information is as
follows:
Software Upgrade and Maintenance 6-15
Catalog
Data
Illustration
C:\z5 \Image
POD
Probe data (supported probes in the system) ColorSpectrum
map data, color scale and colorize
\gui
Font
font file
Skin
Skin file
pinyinmap
pinyin data
Word
input setting data
LayoutInfo
Layout configuration file
version.txt
Version file
\app
bodymark
pictogram
DCM
structured report related data file and DICOM viewing software
Report
report template
obd.bin
OBD data file
version.txt
Version file
Comment
Comment file
AnatomyImage
Anatomical images
Measurement
Measurement library
\exe
main
Doppler program and related boot configuration file and plug-ins, remote desktop server
nls
Multilanguage string file
videoplay
Video play exe file
Sound
Cine saving and hint sound file
version.txt
Version file
\video Picture
Display testing picture
version.txt
Version file
\preset factory
Factory data
version.txt
Version file
\config
Boot_logo.png
Doppler boot-up graphics
Hospital logo.bmp
Hospital logo
manu_logo.png
Manufacturer logo
Real_Image.bmp
Active icon
Non_Real_Image.bmp
Inactive icon
Dorm_logo.bmp
Standby graphics
Product.PCF
Product configuration file
\Key
Factory.key
Factory key file
1. C drive
6-16 Software Upgrade and Maintenance
Catalog
Data
Illustration
D:\z5
\gui
\word
User-defined word library
\log
\DcmLog
DICOM log
\Monitor
Monitor log
\Operation
Operation log
\SystemStat
Power on/off log
\selftest
Self test log
appmon_window_monitor.txt
Appmon monitor log
message.txt
Warning message
PeriLog.txt
Peripheral log
commentlog.txt
Comment log
burn_cd_msg.txt
Burn log
error.txt
System error log
\PATIENTDATA
\
Main patient database path
\Preset
\Current
User preset data
\temporary
\
Temporary file
\PatientBack
\
Patient back up data
Catalog
Data
Illustration
E:z5 Demo
Ivision default DEMO path
Patient Backup
Patient data back up
demofile.txt
Ivision default DEMO path
2. D drive
3. E drive
Software Upgrade and Maintenance 6-17
No.
Name
No.
Name
1
Holder
5
Null turning axis of keyboard
2
Rear cover assembly of main unit
6
Silicon cap of keyboard
3
Assembly of main unit rack
7
Top cover assembly of keyboard
4
Display assembly
8
Bottom cover assembly of keyboard
7 Structure and
Assembly/Disassembly

7.1 Structure of the Complete System

Figure 7-1 Overall Exploded View
Structure and Assembly/Disassembly 7-1
No .
Classificatio
n
Description
Photo
Order Number
Model
Compatibility
Assembly/
Disassembly
1.
Main unit
related
1112 Main board(FRU)
801-1112-00001-00
Z6 Series
Refer to7.4.5
2.
1112 Main Board Assembly(FRU EAF328-F)
115-031629-01
This board should be used with software of version 02.01.00 Rev7960) and later.
3.
1112 Main Board Assembly(FRU Matched with BayTrail PC module)
115-031629-02
This board should be used with software of version 03.00.00 Rev8742) and later.

7.2 Field Replaceable Unit

7-2 Structure and Assembly/Disassembly
4.
1114 Main Board(FRU)
115-015654-00
Z5 Series
Refer to7.4.5
5.
1114 Main Board Assembly(FRU EAF328-F)
115-015654-02
This board should be used with software of version 02.01.00 Rev7959) and later.
6.
1114 Main Board Assembly(FRU Matched with BayTrail PC module)
115-015654-03
This board should be used with software of
03.00.00 (Rev8856) version and later
7.
CPU Module
801-1150-00002-00
Z5&Z6 Series
Refer to
7.4.5
8.
BayTrail PC Module (1112/1114 FRU)
115-043600-00
Z5&Z6 Series
Z5: be used with software of 03.00.00 (Rev8856) version and later Z6: be used with software of version 03.00.00 Rev8742) and later.
Structure and Assembly/Disassembly 7-3
9.
HDD
115-018452-01
Z6 (FDA) /
Refer to7.4.8
115-018452-02
Matched with software version of 03.00.00 (Rev8742) and later.
115-018453-01
Z6Vet (FDA) /
115-018453-02
Matched with software version of 03.00.00 (Rev8742) and later.
115-018454-01
Z6 (CE) /
115-018454-02
Matched with software version of 03.00.00 (Rev8742) and later.
115-018455-01
Z6Vet (CE) /
115-018455-02
Matched with software version of 03.00.00 (Rev8742) and later.
7-4 Structure and Assembly/Disassembly
10.
HDD
115-018456-01
Z5 (CE)
/
Refer to
7.4.8
115-018456-02
Matched with software version of
03.00.00 (Rev8856) and later.
115-033819-01
Z5 (FDA)
/
115-033819-02
Matched with software version of
03.00.00 (Rev8856) and later.
115-018457-01
Z5BW (CE)
/
115-018457-02
Matched with software version of
03.00.00 (Rev8856) and later.
115-018459-01
Z5Vet (CE)
/
115-018459-02
Matched with software version of
03.00.00 (Rev8856) and later.
115-033820-01
Z5Vet (FDA)
/
115-033820-02
Matched with software version of
03.00.00 (Rev8856) and later.
115-039358-00
DP-50 (color) [CE]
/
115-039358-01
Matched with software version of
03.00.00 (Rev8856) and later.
115-018458-01
Z5T (CE)
/
115-018458-02
Matched with software version of
03.00.00 (Rev8856) and later.
Structure and Assembly/Disassembly 7-5
11.
Battery connecting board
801-1150-00004-
00
Z5&Z6 Series
Refer to7.4.1
12.
IO board
801-1150-00005-
00
Z5&Z6 Series
Refer to7.4.3
13.
1150 probe board
801-1151-00006-
00
Z5&Z6 Series
Refer to7.4.4
14.
1151 probe board
801-1151-00002-
00
Z5&Z6 Series
Refer to7.4.4
7-6 Structure and Assembly/Disassembly
Loading...