For this Operator’s Manual, the issue date is April, 2014.
I
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual. This
manual may refer to information protected by copyright or patents and does not convey any
license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution,
rental, adaptation, translation or any other derivative work of this manual in any manner
whatsoever without the written permission of Mindray is strictly forbidden.
,,and SynoVent are the trademarks, registered or
otherwise, of Mindray in China and other countries. All other trademarks that appear in this
manual are used only for informational or editorial purposes. They are the property of their
respective owners.
II
Responsibility on the Manufacturer
Party
Contents of this manual are subject to change without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements;and
the product is used in accordance with the instructions for use.
WARNING
It is important for the hospital or organization that employs this equipment to
carry out a reasonable service/maintenance plan. Neglect of this may result in
machine breakdown or personal injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Address: Eiffestraβe 80, 20537 Hamburg, GERMANY
Tel: 0049-40-2513175
Fax: 0049-40-255726
V
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a prerequisite for
proper product performance and correct operation and ensures patient and operator safety.
This manual is based on the maximum configuration and therefore some contents may not
apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the equipment
so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practices and terminology as required for monitoring of
critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the
setup or data displayed on your ventillator.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
[ ] is used to enclose screen texts.
→ is used to indicate operational procedures.
Password
A password is required to access different modes within the ventilator machine.
Indicates an imminent hazard that, if not avoided, will result in death or serious
injury.
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
minor personal injury or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the
most from your product.
1-1
1.1.1 Dangers
There are no dangers that refer to the product in general. Specific “Danger” statements may
be given in the respective sections of this manual.
1.1.2 Warnings
WARNING
The ventilator must only be operated and used by authorized medical personnel
well trained in the use of this product. It must be operated strictly following the
Operator’s Manual.
Before putting the system into operation, the operator must verify that the
equipment, connecting cables and accessories are in correct working order and
operating condition.
To avoid the risk of electric shock, this equipment must only be connected to
supply mains with protective earth.
Use AC power source before the batteries are depleted.
To avoid explosion hazard, do not use the equipment in the presence of flammable
anesthetic agent, vapors or liquids.
Do not place the ventilator adjacent to any barrier, which can prevent cold air
from flowing, resulting in equipment overheat.
Do not open the equipment housings. All servicing and future upgrades must be
carried out by the personnel trained and authorized by us only.
Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Remember that alarm settings should be customized according to different patient
situations and always keeping the patient under close surveillance is the most
reliable way for safe patient monitoring.
The physiological parameters and alarm messages displayed on the screen of the
equipment are for doctor’s reference only and cannot be directly used as the basis
for clinical treatment.
Dispose of the package material, observing the applicable waste control regulations
and keeping it out of children’s reach.
All staff should be aware that disassembling or cleaning some parts of the
ventilator can cause risk of infection.
Maintenance mode should be used only when the equipment is not connected to a
patient.
1-2
WARNING
Positive pressure breathing may be accompanied by some side effects such as
barotrauma, hypoventilation, hyperventilation etc.
Using the ventilator in the vicinity of high-frequency electrosurgery units,
defibrillators or short-wave therapy equipment may impair correct functioning of
the ventilator and endanger the patient.
Do not use antistatic or conductive masks or breathing hoses. They can cause burns
if they are used near high frequency electrosurgical equipment.
Do not use the ventilator in a hyperbaric chamber to avoid potential fire hazard
due to an oxygen-enriched environment.
If the equipment internal monitoring system malfunctions, an alternative plan
must be available to ensure adequate level of monitoring. The operator of the
ventilator must be responsible for proper patient ventilation and safety under all
circumstances.
As required by the relevant rules and regulations, oxygen concentration should be
monitored when the equipment is used on the patient. If your ventilator is not
configured with such monitoring function or this function is turned off, use a
monitor which complies with the relevant international rules and regulations for
oxygen concentration monitoring.
When auxiliary electrical outlets are configured, the voltage and current
specifications of the devices connected to the electrical outlets must be within the
permissible ranges for those of the electrical outlets. When the protection
grounding is defective, connection of equipment to the auxiliary electrical outlet
may increase the patient leakage current to values exceeding the allowable limits.
When the auxiliary electrical outlet does not work normally, check if the
corresponding fuse is burned.
All analog or digital products connected to this system must be certified passing the
specified IEC standards (such as IEC 60950 for data processing equipment and
IEC 60601-1 for medical electrical equipment). All configurations shall comply
with the valid version of IEC 60601-1-1. The personnel who are responsible for
connecting the optional equipment to the I/O signal port shall be responsible for
medical system configuration and system compliance with IEC 60601-1-1 as well.
Do not touch the patient when connecting the peripheral equipment via the I/O
signal ports or replacing the oxygen cell to prevent patient leakage current from
exceeding the requirements specified by the standard.
This product must be operated by doctors, respiration therapist or other specially
trained and authorized personnel. Anyone unauthorized or untrained must not
perform any operation on it.
This equipment is not suitable for use in an MRI environment.
1-3
WARNING
When the ventilator input system fails or has faults, please contact us immediately
for specified personnel to service the ventilator.
Use the humidifiers with a CE mark or recommended by us only.
The ventilator cannot use He and O2 mixed gas.
Do not move the ventilator before removing the support arm from it, in order to
avoid the ventilator getting tilted during the movement.
Nebulization or humidification can increase the resistance of breathing system
filters and that you need to monitor the filter frequently for increased resistance
and blockage.
The ventilation accuracy can be affected by the gas added by use of a nebulizer.
The ventilator shall not be used with nitric oxide.
The ventilator shall not be used with helium or mixtures with Helium.
For non-invasive ventilation, the exhaled volume of the patient can differ from the
measured exhaled volume due to leaks around the mask.
Check if the alarm limit settings are appropriate before taking measurement.
The mains plug is used to isolate the ventilator circuits electrically from the
SUPPLY MAINS,not to position the ventilator so that it is difficult to operate the
plug.
No modification of this equipment is allowed.
Do not touch the patient when connecting the peripheral equipment via the I/O
signal ports or replacing the oxygen cell to prevent patient leakage current from
exceeding the requirements specified by the standard.
RJ45 port is used for software upgrades by the personnel trained and authorized
by us only. Do not connect it to other devices or internet.
1.1.3 Cautions
CAUTION
The ventilator must be inspected and serviced regularly by trained service
personnel.
To ensure patient safety, always prepare pulmotor for use.
Always have a special person attend and monitor the operation of the equipment
once the ventilator is connected to the patient.
During the operation of the ventilator, do not disassemble the expiration valve and
expiratory flow sensor, which, however, can be disassembled in standby mode.
1-4
CAUTION
To ensure patient safety, use only parts and accessories specified in this manual.
At the end of its service life, the equipment, as well as its accessories, must be
disposed of in compliance with the guidelines regulating the disposal of such
products.
Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason make sure that all external devices
operated in the vicinity of the equipment comply with the relevant EMC
requirements. Mobile phone, X-ray equipment or MRI devices are a possible
source of interference as they may emit higher levels of electromagnetic radiation.
This system operates correctly at the electrical interference levels identified in this
manual. Higher levels can cause nuisance alarms that may stop mechanical
ventilation. Pay attention to false alarms caused by high-intensity electrical fields.
Before connecting the equipment to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the
equipment’s label or specified in this manual.
Always install or carry the equipment properly to avoid damage caused by drop,
impact, strong vibration or other mechanical force.
The ventilator or its part should be positioned so that is easy to view.
The ventilator is intended to be used in the patient environment.
Additional MULTIPLE SOCKET- OUTLET or extension cord shall not be
connected to the system.
When pushing the ventilator over the obstacles such as threshold, force the
ventilator downwards to avoid getting tilted.
1-5
1.1.4 Notes
NOTE
Put the equipment in a location where you can easily see the screen and access the
operating controls.
Keep this manual close to the equipment so that it can be obtained conveniently
when needed.
The software was developed in compliance with IEC 60601-1-4. The possibility of
hazards arising from software errors is minimized.
This manual describes all features and options. Your equipment may not have all
of them.
Humidifier and nebulizer are independent devices which shall be purchased
separately if necessary. When using the humidifier and nebulizer, refer to their use
methods specified by their manufacturers.
1-6
1.2 Equipment Symbols
Alternating current
Equipotential
RS-232 port
O2↑button
Video input/output
connection
Network connection
Air supply connection
Pneumatic outlet
Battery
Fuse
CO2/calibration connection
O2 sensor connector
VGA output connection
USB port
Oxygen supply connection
Flow sensor
Expiratory port
Unlock
Compressor status indicator
Compressed air outlet (of
the compressor)
Manufacture date
Serial number
Refer to instruction
manual/booklet
Inspiratory port
Nebulizer connection
Nurse call connection
Central pipeline gas supply
inlet (of the compressor)
Manufacturer
Caution
General warning sign
1-7
IP21
Type BF applied part.
Defibrillation-proof
protection against electric
shock.
IP21
Degrees of protection
provided by enclosure (IP
Code)
The following definition of the WEEE label applies to EU member states only.
This symbol indicates that this product should not be treated as household
waste. By ensuring that this product is disposed of correctly, you will help
prevent bringing potential negative consequences to the environment and
human health. For more detailed information with regard to returning and
recycling this product, please consult the distributor from whom you purchased
it.
* For system products, this label may be attached to the main unit only.
The product bears CE mark indicating its conformity with the provisions of the
Council Directive 93/42/EEC concerning medical devices and fulfils the
essential requirements of Annex I of this directive.
No pushing
Protective earth (ground)
1-8
2 The Basics
2.1 System Description
2.1.1 Intended Use
The Ventilator is intended to provide ventilation assistance and breath support for adult,
pediatric and infant patients with respiratory insufficiency or respiratory failure in the
hospital or other medical institutions. Ventilation may be delivered via mask or tracheotomy.
This product must be operated by doctors, respiration therapist or other specially trained and
authorized personnel. Anyone unauthorized or untrained must not perform any operation on
it.
2.1.2 Contraindications
There are no absolute contraindications for this product. However, for some patients who
suffer from special diseases, special ventilation is required or treatment has to be carried out
before mechanical ventilation. Otherwise, hazards may be resulted.
2.1.3 Components
The ventilator consists of a main unit (including pneumatic circuit, electronic system,
mechanical structure, software, display, CO2 module), cart, support arm, air compressor, and
breathing hoses (refer to chapter 12 Accessories).
Connect the patient to the ventilator via the patient breathing circuit.
The applied part of the ventilator is breathing masks.
2-1
2.2 Equipment Appearance
2.2.1 Front View
8
7
6
11
5
4
3
2
9
10
12
1
2-2
1. Caster and brake
The four casters of the ventilator have brakes.
2. Compressor
3. Inspiratory water trap
Collects condensed water in the hose.
4. Expiratory water trap
Collects condensed water in the hose.
5 Test lung
6. Expiratory hose
7. Inspiratory hose
8. Support arm
Supports the patient’s breathing hoses.
9. Display
10. Expiratory filter
Prevents water and bacteria inside the patient hoses from entering the ventilator’s
internal pneumatic circuit.
11. Water trap at the expiratory port
12. Inspiratory filter
Prevents water and bacteria inside the patient hoses from entering the ventilator’s
internal pneumatic circuit.
2-3
2.2.2 Rear View
13
5
2
3
4
1. RS-232 port
Connects to the medical-grade external device via RS-232 protocol to implement the
communication between the ventilator and external device.
910
8
6
7
1
11
12
2. Oxygen supply connection (with filter water trap)
3. Air supply connection (with filter water trap)
4. CO2/calibration connection
One multiplex connector for calibrating inspiratory and expiratory flows and supplying
power for the external CO2 analyzer.
5. Fan
6. AC mains inlet
7. VGA connection
The ventilator provides a D-Sub 15, female video output connector, which connects to
an external display and outputs VGA visual signals same to the primary display. This
connection allows for interfacing to an externally located 24 bits, 800 x 600, SVGA
monitor, which should be a medical grade monitor.
2-4
8. USB port
9. Network connection
One multiplex connector for network and software online upgrade.
10. Nurse call connection
Connects to the hospital’s calling system and outputs nurse call signals when an alarm
occurs.
11. Inspiratory dust filter
12. Fan filter (filter at air intake vent)
13. Equipotential stud / lug
Provides a ground point. Eliminates the ground potential difference between different
devices to ensure safety.
2.2.3 Air Compressor
The air compressor has standby function. In the standby mode, the compressor starts to
deliver compressed air to the ventilator automatically if the hospital central pipeline gas
supply fails. The compressor stops delivering compressed air automatically when the central
pipeline gas supply returns to normal.
1 2
1. Power indicator
4
3
56 7
8
The power indicator is lit when the compressor is connected to power supply and the
power switch is turned on.
2. Status indicator
The status indicator is lit when the central pipeline gas supply is applied.
3. Alarm indicator
The alarm indicator is lit when the compressor internal temperature is abnormally high.
In this case, the compressor may shut off at any time and stop delivering gas.
2-5
4. Pressure gauge
The pressure gauge indicates the air pressure at the compressed air outlet.
5. Compressed air outlet
6. Central pipeline gas supply inlet
7. Power switch
8. Mains power inlet (with fixing pressure plate)
9. Air intake vent (with dust filter)
10. Hourmeter
The hourmeter indicates the accumulated running time of the compressor (not including
the accumulated running time when the central pipeline gas supply is applied)
NOTE
Burn-in is required for the compressor before delivery. The reading indicated by
the compressor hourmeter shall be less than 150 hours at the time of delivery.
2-6
3 Installations and Connections
WARNING
Do not use antistatic or conductive masks or breathing hoses. They can cause burns
if they are used near high frequency electrosurgical equipment.
To ensure optimum performance of the ventilator, re-do system check each time
when accessories or components like hose, humidifier, and filter are replaced.
Adding accessories or other components to the ventilator can increase system
inspiratory and expiratory resistance.
3.1 Connect the Power Supply
3.1.1 Connect the System Power Supply
A
B
C
3-1
A. AC power cord B. AC Power cord retainer C. Screw
1. Plug the AC power cord into the AC power outlet.
2. Place the AC power cord retainer above the power outlet and align the retainer with the
screw holes.
3. Tighten the two screws.
3.1.2 Connect the Main Unit Power Supply
A. Main unit power cord B. Main unit power cord retainer C. Screw
1. Plug the main unit power cord into the power outlet.
2. Place the main unit power cord retainer above the power outlet and align the retainer
with the screw holes.
3. Tighten the two screws.
3-2
Loading...
+ 162 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.