SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called “Mindray”)
owns the intellectual property rights to this Mindray product and this manual. This manual may refer
to information protected by copyright, trademark, or patents, and does not convey any license under
the intellectual property rights of Mindray or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of
the information in this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other
derivative work of this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
is the trademark of Mindray. All other trademarks that appear in this manual are used
only for informational or editorial purposes. They are the property of their respective owners.
Applicability
This service manual is intended as a guide for technically qualified personnel during service
procedures. This service manual describes the product according to the most complete
configuration; some of the content may not apply to the specific product you are servicing. If you
have any questions, please contact the Mindray Customer Service Department (contact information
is below). Do not attempt to service this equipment unless this service manual has been consulted
and is understood. Failure to do so may result in personal injury or product damage.
Responsibility of Mindray
Contents of this manual are subject to change without prior notice. Please check with the Mindray
Customer Service Department for any updates or changes to this manual.
All information contained in this manual is believed to be correct as of the date of its publication.
Mindray shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Mindray shall not be responsible for the effects on safety, reliability, and performance of this product
if:
Installation operations, expansions, changes, modifications and repairs of this product are
conducted by personnel not authorized by Mindray;
The electrical installation of the relevant room does not comply with the applicable national and
local requirements;
The product is not used in accordance with the instructions for use.
III
Page 12
Fax:
+86 755 26582680
Warranty Statements
Mindray warrants that components within the ultrasound system under warranty will be free from
defects in workmanship and materials for the amount of time specified under Mindray’s then-current
warranty policy (please check with the Mindray Customer Service Department for the applicable
warranty period for each system). Under this warranty, Mindray will repair or replace (at Mindray’s
option) any defective component at no charge for materials according to Mindray’s then-current
warranty policy. This warranty does not cover consumable items such as, but not limited to,
traveling carrying case, acoustic gel, paper, disposable or one-off materials, and sampling
materials.
Recommended preventative maintenance, as prescribed in the Service Manual, is the responsibility
of the user, and is not covered by this warranty.
Mindray will not be liable for any incidental, special, or consequential loss, damage, or expense
directly or indirectly arising from the use of its products. Liability under this warranty and the buyer’s
exclusive remedy under this warranty is limited to servicing or replacing the affected products, at
Mindray’s option, at the factory or at an authorized Distributor, for any product which shall under
normal use and service appear to Mindray to have been defective in material or workmanship.
No agent, employee, or representative of Mindray has any authority to bind Mindray to any
affirmation, representation, or warranty concerning its products, and any affirmation, representation,
or warranty made by any agent, employee, or representative shall not be enforceable by buyer or
user.
THIS WARRANTY IS EXPRESSLY IN LIEU OF, AND MINDRAY EXPRESSLY DISCLAIMS, ANY
OTHER EXPRESS OR IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF
NON-INFRINGEMENT, MERCHANTABILITY, OR FITNESS FOR A PARTICULAR PURPOSE, AND
OF ANY OTHER OBLIGATION ON THE PART OF MINDRAY.
Damage to any product or parts through misuse, neglect, accident, or by affixing any non-standard
accessory attachments or by any customer modification voids this warranty.
Mindray makes no warranty whatever in regard to trade accessories, such being subject to the
warranty of their respective manufacturers.
A condition of this warranty is that the equipment or any accessories which are claimed to be
defective be returned, when authorized, to the appropriate Mindray affiliate. Please contact the
Mindray Customer Service Department for appropriate details for your region.
The following marks are used for describing keys on the control panel, menu items, buttons on
dialog boxes and other basic operations in the manual:
<Button>: the angle bracket for enclosing the button’s name refers to the buttons on the
keyboard, rotation knobs, switches or controls.
[Menu item or key]: the square bracket for enclosing menu item or key refers to the menu
items or the keys on dialog boxes.
Click [Menu item or key]: move the cursor to the menu item or the key on the dialog box,
and then press <Set>. Or, click other optional keys on touch screen.
[Menu item]-[Sub-menu item]: select sub-menu item based on the operation path
.
V
Page 14
Page 15
Indicates a potentially hazardous situation that, if not avoided, may result in
Symbol
Position
1 Safety Precautions
This chapter describes important issues related to safety precautions, as well as the labels and
icons on the ultrasound machine.
1.1 Meaning of Signal Words
In this service manual, the signal words DANGER, WARNING, CAUTION and NOTE
are used regarding safety and other important instructions. The signal words and their meanings
are defined as follows. Please be aware of the meaning of the signal words before reading this
manual.
Signal word Description
DANGER
WARNING
CAUTION
NOTE
Description Important information that helps you to use the system more effectively.
Indicates an imminently hazardous situation that, if not avoided, will result
in death or serious injury.
Indicates a potentially hazardous situation that, if not avoided, could result
in death or serious injury.
minor or moderate injury.
Indicates a potentially hazardous situation that, if not avoided, may result in
property damage.
1.2 Symbols
The following tables provide location and information of the safety symbols and warning labels,
please read carefully.
1.2.1 Meaning of Safety Symbols
Description
Type-BF applied part
Note: The ultrasound probes connected to
this system are type-BF applied parts.
The ECG module connected to this system
is Type-BF applied part.
Right side of main unit
Caution
Main unit
Safety Precautions 1-1
Page 16
Read the manual carefully before
On the upper left
No. Symbol
Description
8 USB port
1.2.2 Warning Labels
No. Warning Labels Description Label Position
1.
The following labels are available
when the system works with the
mobile trolley.
2.
3
1.2.3 General Symbols
using the system.
a. Do not place the system on a
sloped surface. Otherwise the
system may slide, resulting in
personal injury or the system
malfunction. Two persons are
required to move the system over
a sloped surface.
b. Do not sit on the system.
c. DO NOT push the system when
the casters are locked.
Non-ionizing radiation On the bottom part
corner of the
control panel
On the left part of
the trolley panel
of the ultrasound
system
This symbols used in the device are listed in the following table. Meanings are:
1
2
3
4
5
6
7
9
10
11
Type-BF applied part
To avoid safety accidents, refer to relevant content in the manual.
Dangerous voltage
No user serviceable parts (applied to the power adapter)
Indoor, dry location use only (applied to the power adapter)
Power button
Probe sockets
S-VIDEO signal port; VIDEO signal port
AC (Alternating current)
Battery Status Indicator
1-2 Safety Precautions
Page 17
No. Symbol
Description
DANGER:
12
13
14
15
16
17
18
19
20
21
22
Pencil probe port (reserved)
IO extend port
unlocked symbol
locked symbol
Battery installation position indicator
Connects serial port devices
ECG function
Connects a display monitor or projector
Audio signal
Microphone input jack
Remote control port
23
24
25
26
27
Product serial number
Manufacture date
This product is provided with a CE marking in accordance with the regulations
stated in Council Directive 93 / 42 / EEC concerning Medical Devices. The number
adjacent to the CE marking (0123) is the number of the EU-notified body certified for
meeting the requirements of the Directive.
The environment-protective application period of the system is 20 years period.
Standby
1.3 Safety Precautions
Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the probes.
Do not operate this system in an atmosphere containing flammable
or explosive gases such as anesthetic gases, oxygen, and hydrogen
or explosive fluid such as ethanol because an explosion may occur.
Safety Precautions 1-3
Page 18
1.3.1 Electric Safety
WARNING:
4.
In maintenance or assembly/disassembly, make sure other
1.
Connect the power plug of this system and power plugs of the
peripherals to wall receptacles that meet the ratings indicated
on the rating nameplate. Using a multifunctional receptacle may
affect the system grounding performance, and cause the
leakage current to exceed safety requirements. Use the power
cord accompanied with the system provided by Mindray.
Disconnect the AC power before you clean or uninstall the
2.
ultrasound machine, otherwise, electric shock may result.
3.
When using peripherals not powered by the auxiliary output of the
ultrasound system, or using peripherals other than permitted by
Mindray, make sure the overall leakage current of peripherals and
the ultrasound system meets the requirement of the local medical
device electrical regulation (like enclosure leakage current should
be no more than 500uA of IEC 60601-1-1), and the responsibility is
held by the user.
CAUTION:
cables are connected well before the battery connecting cable is
connected, otherwise the system may be damaged due to
hot-plug.
5.
Do not use this system simultaneously with equipment such as
an electrosurgical unit, high-frequency therapy equipment, or a
defibrillator, etc.; otherwise electric shock may result.
6.
This system is not water-proof. If any water is sprayed on or into
the system, electric shock may result.
1.
DO NOT connect or disconnect the system’s power cord or its
accessories (e.g., a printer or a recorder) without turning OFF
the power first. This may damage the system and its
accessories or cause electric shock.
2.
Avoid electromagnetic radiation when perform performance test
on the ultrasound system.
3.
In an electrostatic sensitive environment, don’t touch the device
directly. Please wear electrostatic protecting gloves if
necessary.
4.
You should use the ECG leads provided with the ECG module.
Otherwise it may result in electric shock.
of the trolley is 240 VA.
5. Maximum output power
1-4 Safety Precautions
Page 19
1.3.2 Mechanical Safety
2.
Do not subject the transducers to knocks or drops. Use of a
connector of the power adapter in case of damage.
device disassemble is allowed.
tem performance and safety, only Mindray engineers or
3.
Only technical professionals from Mindray or engineers authorized by Mindray
to the operator’s manual.
1.
WARNING:
When moving the system, you should first power off the
system, fold the LCD display, disconnect the system from other
devices (including probes) and disconnect the system from the
power supply.
CAUTION:
defective probe may cause electric shock to the patient.
1.
Do not expose the system to excessive vibration (during the
transportation) to avoid device dropping, collision, or
mechanical damage.
2.
When you place the system on the mobile trolley and move them
together, you must secure all objects on the mobile trolley to
prevent them from falling. Otherwise you should separate the
system from the mobile trolley and move them individually.
When you have to move the system with the mobile trolley
upward or downward the stairs, you must separate them first
and then move them individually.
3.
Do not move the ultrasound system if the HDD indicator is
green, sudden shake may cause the HDD in damage.4 When
moving the trolley with mounted system, please take care of the
1.3.3 Personnel Safety
The user is not allowed to open the covers and panel of the system, neither
Note:
1.
To ensure the sys
2.
engineers authorized by Mindray can perform maintenance.
after training can perform maintenance.
1.3.4 Others
Note:
For detailed operation and other information about the ultrasound system, please refer
Safety Precautions 1-5
Page 20
Page 21
2 Product Specifications
2.1 Introduction
2.1.1 Intended Use
The diagnostic ultrasound system is intended for use in clinical ultrasonic diagnosis.
2.1.2 System Appearance
Product Specifications 2-1
Page 22
No.
Name
Function
1 Monitor Displays the images and parameters during the scan.
2 Control Panel Operator-system interface or control
3 Handle Used for carrying the system
Power input port
4
5
6 USB portConnects USB devices (2 ports)
7 ECG port Connects ECG lead or DC-IN cable
8
9 Probe locking lever
10 Pencil probe port Connects to a pencil probe.
HDMI port Outputs
Probe port
Connects the power adapter
high definition multimedia signals
Connects a probe to the main unit; or connects a probe
extend module
Locks or unlocks the probe connected with the main unit
:locked symbol
:unlocked symbol
2.1.2.1 Extend Modules
Extend modules to the system are: probe extend module, video/audio extend module.
Probe extend module
No.NameFunction
<1> Connector
Connects to the probe port of the main unit, and
extends the probe port to three.
<2> Probe port To extend ports for connecting probes
2-2 Product Specifications
Page 23
No.
Symbol Function
Note: 1. Perform the plug and unplug of the probe module. To reduce the wastage of probe
module, the operation should be conducted under frozen status of the system.
2. If you use the probe extend module to connect a probe, the image quality may be
degraded.
Audio/Video extend module
Panel introduction
<1>
<2>
<3>
Indicator
Used for connecting to main unit. The green
indicator is on.
Used for audio signal output.
Used for separate video output.
<4>
VGA
VGA signal output
Connection
Connect the V/A extend module to the main unit via an HDMI port. As shown in the following figure.
Product Specifications 2-3
Page 24
2.1.2.2 Control Panel
2-4 Product Specifications
Page 25
TDI Functional
Switching key: Press to change the
Press to enter/exit application
Confirm key
Confirm key (right
Angle adjustment and
Adjust the gain of the image in various
Press to enter/ exit the Cine Review
No.NameEnglish NameDescription
1. B mode button B Press to enter B mode
2. P mode button P Press to enter Power mode.
3. C mode button C Press to enter the Color mode.
4. CW mode button CW Press to enter CW mode.
5. PW mode button PW Press to enter the PW mode.
6.
buttons
TDI Press to enter TDI mode
7. M mode button M Press to enter the M mode.
8. Update button Update
currently active window.
Start or end capturing the image in iScape
9. Measure button Measure
measurement
10. Caliper button Caliper Press to enter/exit general measurement
11. Trackball / To move the cursor.
12.
(left<Set> key)
/
Press to confirm key
13.
<Set> key)
14.
quick angle knob
/
Angle Adjust the angle
15. Gain and iTouch knob iTouch
16. Cine button Cine
modes
Press to enter iTouch mode
status.
Product Specifications 2-5
Page 26
No. Name
English Name
Description
clear off the comments or
Press to select the soft menu items
Refer to the subsequent contents for
Press to select the soft menu items
displayed on the bottom of the screen.
Refer to the subsequent contents for
Press to select the soft menu items
displayed on the bottom of the screen.
Refer to the subsequent contents for
Soft menu page up
Press to turn the soft menu pages
Soft menu display
Press to switch the mode for the soft
Press to select the soft menu items
displayed on the bottom of the screen.
Press to select the soft menu items
displayed on the bottom of the screen.
Refer to the subsequent contents for
Press to select the soft menu items
displayed on the bottom of the screen.
Refer to the subsequent contents for
enter the Patient information
Probe/exam mode
17. Clear button Clear
Press to
measurement calipers on the screen.
18. Freeze button Freeze Press to freeze or defreeze the image.
19. Power button / Power button
Soft menu adjustment
20.
button 1
/
displayed on the bottom of the screen.
specific functions.
Soft menu adjustment
21.
button 2
/
specific functions.
Soft menu adjustment
22.
button 3
/
specific functions.
23.
button
Soft menu page down
24.
button
/
upward.
/ Press to turn soft menu pages downward.
25.
mode switch button
Soft menu adjustment
26.
button 4
/
/
menu.
Refer to the subsequent contents for
specific functions.
Soft menu adjustment
27.
button 5
/
specific functions.
Soft menu adjustment
28.
button 6
/
specific functions.
Patient information
29.
button
30.
switch button
Patient
Probe/Exam Press to switch probe and exam mode
Press to
screen.
31. Image review button Review Press to review the stored images.
32. End exam button End Exam Press to end an exam.
33. User-defined key 3 P3 User-defined key
34. User-defined key 2 P2 User-defined key
35. Body mark button Body Mark Press to enter/ exit the body mark mode.
2-6 Product Specifications
Page 27
No. Name
English Name
Description
Baseline position
adjustment button
and auxiliary
Scale adjustment
button and auxiliary
TGC and functional
split window
Press to enter the Dual mode from
Image zoom and
Depth adjustment
Focus position
adjustment button
and functional
Save static image
Standby status
Hard disk read
DO NOT move the machine when the
hard disk may be damaged by sudden
36. Cursor button Cursor Press to show/hide the cursor
37.
Baseline Adjust the baseline parameter.
interface
38.
Scale Adjust scale parameter.
interface
Steer/Invert
39.
adjustment button
40.
interface
41. Single-window button Single
Dual-
42.
button
Steer/Invert Adjust steer/invert parameter.
/ Move to adjust depth gain compensation.
Press to enter active window in Dual or
Quad mode.
another mode;
Dual
Press to switch between the two windows
in the Dual mode.
43.
spot/pan zoom switch
Zoom Press to enter or exit zoom status.
knob.
44.
button and fictional
Depth Adjust depth
interface
45.
Focus Position Adjust the focus position.
interface
46.
button
Save1
Press to save the image; user-defined
key.
47. User-defined key 1 P1 User-defined key
Standby indicator
48.
indicator
/
Standby: blinking in orange
Other status: light off
Hard disk status indicator
The indicator blinks in green when hard
disk is running.
49.
indicator
/
The indicator is off on the other status.
indicator blinking in green. Otherwise the
shake.
Product Specifications 2-7
Page 28
No. Name
English Name
Description
Patient data
Activate the Home function: return to start
Battery status indicator.
Charging status:
It illuminates in orange when batteries are
charging;
It illuminates in green when batteries are
charged to full capacity;
50. Battery indicator /
Discharging status:
It illuminates in green when batteries are
charged fairly;
It illuminates in orange when batteries run
out.
AC indicator
51. AC power indicator /
The indicator is green at AC supply.
The indicator is off when batteries (AC
does not supply the power) supplied.
52. Esc Esc Cancel the operation or exit.
53. On-line help Help
Press to open or close the accompanying
help documents.
54.
management system
iStation
Press to enter or exit patient information
system.
55. Report Report Open/ close the exam report.
56. User-defined key 5-8 P5~P8
User-defined keys, functions of which can
be defined in preset.
57. Stress echo Stress Echo Press to enter Stress Echo mode
58. Physiological signal Physio Press to enter or exit ECG.
59. Setup Setup Press to enter/ exit setup.
60. Biopsy Biopsy Press to enter biopsy
61. Set home Set Home Set home of comments
Enter Quad mode in Non-Quad mode;
62. Quad-window display Quad
Press to switch among interfaces in Quad
mode.
iZoom (Full-screen
63.
Zooming)
iZoom Enter/switch/exit full-screen zoom status.
64. Arrow Arrow Enter/exit the arrow comment status.
65. Home Home
66. Comments Comment Enter/exit the textual comment status.
2-8 Product Specifications
position of comment.
Page 29
No. Name
English Name
Description
1.
→
End
Turn up the volume
/ Moves the cursor one letter each time; or,
67. Direction-control keys
select the ambient one in a selectable
area.
68. User-defined key 4 P4
User-defined keys, functions of which can
be defined in preset.
69. Alphanumeric keys / Enter characters.
For functions of undefined buttons or keys, the user can define it on your own. Refer to Operator manual for details about function setting.
Functions of key combination
The system supports multi-language input; you can use the key combinations. The key
combinations include <Shift>, <Alt Gr>, <Ctrl> and some alphabet keys.
<Shift> combination key
<Shift> + key: input the upper left letter of the key.
For the alphabet keys (<A>~<Z>), press <Shift>+key to input the letter of different case
with the current state.
[Alt Gr] combination key
Combined with other letter keys, [Alt Gr] can be used for entering other languages.
Press [Alt Gr], and meanwhile press a letter key, the letter on the upper-right corner of the
key can be entered.
[Ctrl] combination key
In iStation or Review screen, use <Ctrl> and <Set> to select more than one patient.
The combination of <Ctrl> and <Space> can switch the input between Chinese and
English.
Combination key used together with [Fn] key
For those combination keys, press <Fn>+key to use the functions indicated with a frame on
the key.
No.Fn+ NameFunction
2. ←Home Turn down the volume
3. ↑Pg Up Increase the brightness of the LCD display.
4. ↓Pg Dn Decrease the brightness of the LCD display.
Product Specifications 2-9
Page 30
2.1.3 Trolley Appearance
2-10 Product Specifications
Page 31
<6>
Probe extend module
Used to extend ports for connecting probes (optional)
<11>
Audio/video extender
Used to switch audio/video (optional)
No.NameFunction
<1> Probe cable hook Manage probe cable
<2> Ultrasound main unit Ultrasound system
<3> Ascending/descending switch Press to adjust the height of operation panel
<4> Anti-theft setting Used to fix the ultrasound system to the trolley.
<5> Printer bracket Used to place the printer
<7> Storage tray Used to place the adaptor, gel and other stuff.
<8> USB portConnects USB devices (2 ports)
<9> Probe holder Used to place probes temporarily
Used for securing or moving the system
<10> Caster
Step on the second footswitch to lock the trolley. Step
on the first footswitch to unlock the trolley.
<12> Lifting column Used to adjust the height of operation panel
Product Specifications 2-11
Page 32
No. Name
Function
No.
Name
Function
<13> Moveable hook Manage probe cable
<14> Storage box Used to place report or other stuff.
<15> Power supply panel
Used to offer output/input port, equipotential terminal
for power supply.
Battery indicator:
It illuminates in orange when batteries are
charging;
It illuminates in green when batteries are in
the charge.
<16> Indicator
It illuminates in green when batteries are
charged fairly;
It illuminates in orange when batteries run
out.
AC power indicator:
The indicator is green at AC supply.
The indicator is off when batteries (AC does not supply
the power) supplied.
2.1.3.1 Power Supply Panel
1
Equipotential
terminal
2 Power inlet AC power inlet
3
Power supply
outlet
Used for equipotential connection, balancing the protective earth
potentials between the device and other electrical equipment.
Supply power for optional peripheral devices (e.g. DVR)
The symbol represents the equipotential terminal that is used for balancing the protective
earth potentials between the system and other electrical equipment.
2-12 Product Specifications
Page 33
WARNING:
Be sure to connect the equipotential wire before inserting the
sure to pull out the power
When you connect another device to this system, you should
use the equipotential wire to connect each of equipotential
arth cable before turning ON the system.
Disconnect the earth cable after turning OFF the system.
DO NOT connect this system to outlets with the same circuit
ON, the circuit breakers and fuses of the building’s supply
1.
Graph/text printer
HP Officejet 7000 wide format
This device complies with IEC60601-1-2:2014, and its RF emission
In a domestic
the system with Class B peripheral devices; otherwise RF
interference may result and the customer or the user must take
1.
power plug into the receptacle; be
plug from the receptacle before disconnecting the equipotential
wire; otherwise electric shock may result.
2.
terminals; otherwise electric shock may result.
3. Connect the e
Otherwise electric shock may result.
4.
breakers and fuses that control the current to devices such as
life-support systems. If this system malfunctions and generates
over current, or when there is an instantaneous current at power
circuit may be tripped.
2.1.4 Peripherals Supported
The peripheral devices supported by the system. The information is shown as below:
No.ItemModel
MITSUBISHI P95DW-N
2. Black/white video printer
3. Digital color video printer
4. 1D/2DBarcode reader
5. DVR
6. External DVD R/W drive ASUS DVD SDRW-08D2S-U
NOTE: printer UP-D898MD and UP-X898MD should be used in compatible with 1.00.7 2015-1-27
OS version or above versions. You can also select “DVR: 897” in the DIGITAL -> - DRIVER path on
the printer (use UPD897 driver) if OS is not updated.
SONY UP-D897
SONY UP-D898MD
SONY UP-X898MD
SONY UP-D25MD
SYMBOL DS4308
SYMBOL LS2208
JADAK HS-1M
JADAK HS-1R
meets the requirements of CISPR11 Class B.
WARNING:
environment, the customer or the user should guarantee to connect
adequate measures accordingly.
Product Specifications 2-13
Page 34
Voltage of
Voltage
2.2 Specifications
2.2.1 External Dimensions and Weight
Size: (362±5) X (390±5) X (59±3) mm
Weight (built-in battery assembly): < 6.5KG
2.2.2 Electric Specifications
2.2.2.1 AC IN
Main unit
Voltage 100-240V~ (for adapter)
Frequency 50/60Hz (for adapter)
Output
power
Trolley
power supply
Frequency of
power supply
Output power 500VA
2.0 A max (for adaptor)
100-240V~
50/60Hz
2.2.2.2 Battery
Main unit battery:
14.8V
Battery
capacity
Trolley battery
Voltage 14.8V
Battery
capacity
5800mAh (one battery) x 2
6600mAh (one battery) x 2
2-14 Product Specifications
Page 35
Storage and Transportation
Do not use this system in the conditions other than those
2.2.3 Environment Specifications
Operational Conditions
Conditions
Ambient
temperature
Relative
humidity
Atmospheric
pressure
WARNING:
0℃-40℃-20℃-55℃
30%-85% (no condensation) 20%-95% (no condensation)
700hPa-1060hPa 700hPa-1060hPa
specified.
2.2.4 Monitor Specifications
Working
voltage
Monitor size 15.6 inches; 16: 9
Resolution 1920×1080
Visual angle ≥160
12V
Product Specifications 2-15
Page 36
Page 37
Note:
WARNING:
3 System Installation
3.1 Installation Preparations
Do not install the machine in the following locations:
Locations near heat generators
Locations with high humidity
Locations with flammable gases
3.1.1 Electrical Requirements
3.1.1.1 Requirements of Regulator
See Chapter 2.2.2 for power supply specifications. Due to the difference of the power supply
stability of different districts, please advise the user to adopt a regulator of good quality and
performance such as an on-line UPS.
3.1.1.2 Grounding Requirements
The power cord of the system is a three-wire cable. The grounding terminal should be connected
with a power grounding cable to ensure that protective grounding works normally. Make sure that
the protective grounding works normally.
DO NOT connect this system to outlets with the same circuit
breakers and fuses that control the current of devices such as
life-support systems. If this system malfunctions and generates an
over-current, or when there is an instantaneous current at power
ON, the circuit breakers and fuses of the building’s supply circuit
may be tripped.
3.1.1.3 EMI Limitation
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transient in the air wiring. They also generate EMI. Possible EMI sources
should be identified before the unit is installed.
These sources include: medical lasers, scanners, monitors, cauterizing guns and so on. Besides,
other devices that may result in high frequency electromagnetic interference such as mobile phone,
radio transceiver and wireless remote control toys are not allowed to be presented or used in the
room. Turn off those devices to make sure the ultrasound system can work in a normal way.
System Installation 3-1
Page 38
3.1.2 Installation Condition
3.1.2.1 Space Requirements
Place the system with the necessary accessories at a proper position for convenient use.
Place the system in a room with good ventilation or having an air conditioning unit.
Maintain a generous – free air flowing space around the back and both sides of the system; failure
may result due to increased rise in system operating temperature.
A combination lighting system in the room (dim/bright) is recommended.
Except the receptacle dedicated for the ultrasound system, at least 3-4 spare receptacles on the
wall are available for the other medical devices and peripheral devices.
Power outlet and place for any external peripheral are within 2 m of each other with peripheral
within 1 m of the unit to connect cables.
3.1.2.2 Network Environment
Both wireless and wired LAN functions are supported by this ultrasound device.
Data transmission is allowed between different departments or areas without network cable.
Network can be automatically connected after disconnection in case that the device is required to
be moved, wireless transmission task can be recovered after the network resumed to normal
condition. Confirm the network devices and network conditions before the installation.
General information: default gateway IP address, and the other routers related information.
DICOM application information: DICOM server name, DICOM port, channels, and IP address.
3.1.3 Installation Confirmation
Please confirm the following items before installation:
The video format of installation area or country.
The language of installation area or country.
Power frequency of installation area or country.
The universal obstetrics formula and other measurement formula of installation area or country.
The preset values of installation area or country that are different from the default values.
The doctor’s operation habits.
The items above prior to the installation training, and do the system settings according to the
universal setting of installed region or country.
3.2 Unpacking
Unpacking tool: scissor
Installation duration: 2 people, 10 minutes.
3.2.1 Unpacking Process
3.2.1.1 Trolley Unpacking
1. Use the scissor to cut off four white tapes around the external package as follows:
3-2 System Installation
Page 39
2. Take out the wooden cover.
3. Unpacking
Remove the box upwards if the space is commodious enough;
If the space is not commodious enough, please follow the instructions below:
Press the middle of plastic clasp on the one side of the box as shown below;
Pull out the plastic clasp;
System Installation 3-3
Page 40
Take out the plastic clasp out of the box (5 plastic clasps);
Take off the plastic clasps to open the box.
4. Tak e out the foam of M9’s top plate.
5. Take out the auxiliary box.
3-4 System Installation
Page 41
6. Take out foam lifting of the trolley.
7. Tak e out the fixing board
8. Turn the wooden cover to a slope.
9. Push the trolley down over the wooden slope.
System Installation 3-5
Page 42
3.2.1.2 Unpack Main Unit
1. Cut off four tapes of the external package, open the external box, and take out the auxiliary box
and fixing foam;
2. Take out the trolley case.
3. Unlock the clasp to open the trolley case.
4. Remove the protective pad, and take out the main unit and adaptor.
5. Remove the protective pad and take out other attachments
3-6 System Installation
Page 43
6. Take out the probe.
WARNING:
The probes, cables and connectors are in proper operating order
Probe
3.2.2 Check
1. After unpacking, check the objects in the container with the package list to see if anything is in
short supply or is wrong.
2. Inspect and make sure there is no damage to the machine, no indentation, no cracks. If there is,
please contact Mindray Customer Service Department.
3.3 Installation of Whole Device
3.3.1 Connecting Power Cable
Connect the connector of the power adapter to the adapter port in the system. Use a three-wire
cable to connect the adapter with the external power supply.
3.3.2 Connecting ECG
Connect the ECG cable to the corresponding lead interface on ECG of the ultrasound device.
3.3.3 Connecting a Ultrasound Probe
and free from surface defects, cracks and peeling. Using a
defective probe may cause electric shock.
System Installation 3-7
Page 44
WARNING:
Keep the cable end of the probe to the right side of the system, and insert the connector into the
port of the system, and then press in fully. See the figure below.
Toggle the locking lever to the top position.
Place the probe properly to avoid being treaded or wrapping with other devices. DO NOT allow the
probe head to hang free.
3.4 Installing Peripherals
Please see Chapter 2.1.4 for the device model that the system supports.
3.4.1 Connecting the Footswitch
The system supports footswitch of USB port type. As shown in the following figure.
You can set the functions of the footswitch in the [Key Config] page. Refer to Operator’s Manual for
footswitch setup.
3.4.2 Connecting/Removing a USB Memory Device
DO NOT directly remove a USB memory device; otherwise, the
USB device and/or the ultrasound system may be damaged.
3-8 System Installation
Page 45
When connecting a USB memory device to the ultrasound system via a USB port, You can hear
a sound if it is connected successfully. You can see the icon
imaging screen.
To remove the USB memory device: Click the icon
below. Select the memory device to be removed. Click [OK] to remove the USB memory device.
There will be sound feedback when removing the USB memory device.
. Press <Set>, as shown in the figure
in the lower right corner of the
3.4.3 Graph/Text Printer
Connecting a local printer
Note: Printers listed in Chapter 2.1.4 have drivers installed already.
As shown in the figure below, a graph / text printer has a power cord and data cable. The power
cord shall be directly plugged into a well-grounded outlet.
1. Connect the data cable to the USB port on the ultrasound system.
2. Power the system and the printer on.
Installing the printer driver
Perform the following steps to install the printer driver as needed:
1. Connect the data cable to the USB port on the ultrasound system.
2. Power the system and the printer on.
3. Put the installation optical disk of the printer driver into the external DVD R/W drive.
Enter [Setup]→[Maintenance]→[Setup]→[Enter Windows], input the password. For details
about obtaining password, please refer to "6.4 Enter Windows".
The system enters the following page:
System Installation 3-9
Page 46
4. Double click the CD driver (HP8100 is given as an example). The installation is starting.
5. Complete the operation according to the prompts on the screen. Click [Finish] to end the
installation.
6. Click [OK] to save the settings.
Print
Both report and image can be printed on a graph/text printer.
1. To set the default report printer and its attribute:
(1) Enter "[Setup]→[Print Preset]" screen.
(2) Select the "Report Print" column in the Service Type list.
(3) In the “Property” frame, select printer from the driver list next to “Printer” in the lower
screen and set the items as needed.
(4) Click [Save] after you have finished setting.
2. Report print
In Report screen, select [Print] to print the report.
If you want to use a shortcut key for report print, you need to define the user-defined key in
“Key Configuration”. For details, see Operator manual.
Please refer to the accompanying manuals of the printers for more details.
3-10 System Installation
Page 47
3.4.4 Video Printer
The system support video printers, consist of the B/W digital printers and color digital printers.
Power cable
port
1. Position the printer in the proper place.
2. Plug the printer power cord into an appropriate outlet.
3. Use a USB cable to connect between the system's USB port and the printer's USB port.
4. Load a paper roll, and turn on the system and printer.
5. See section “3.4.3 Graph/Text Printer” for the driver installation procedure (printer drivers listed
in chapter “2.1.4 Peripherals Supported” are installed already).
Add a print service:
1. Open the “[Setup] → [Print Preset]” screen.
2. Click [Add Service] to enter the page.
3. Select the service type and enter the service name manually.
4. Click [OK] to return to the page.
5. Select the target printer from the drop-down list in the “Property” box and set other printing
properties.
6. Click [Save] to complete.
System Installation 3-11
Page 48
1.
LED indicator
Green light is on if scan is successful.
1.
Barcode reader port
3.4.5 Barcode Reader
The system supports barcode reader to read the patient information (ID).
Note: If the letter case displayed on the dialog box is inversed after scanning the
barcode, please press the <Caps Lock> button on the control panel to toggle
between upper and lower cases, and then scan the barcode again.
The reader does not support Multilanguage decoding.
3.4.5.1 1-D Barcode Reader
1. The appearance of barcode reader. Each part of the barcode reader: LED indicator, scan
window and scan trigger button.
2. Plug connecting terminal of the cable to the port of the barcode reader. Ensure the contact
3. Connect USB terminal of the connecting cable to the USB port of main unit.
Red light is on if the scan fails.
2. Scan window Receive barcode.
3. Scan trigger button Press to perform the scan.
works well.
3-12 System Installation
Page 49
2.
Connecting terminal of connecting cable
4. Press scan trigger button to receive barcode when ultrasound device is running (without
installing driving program). For more operation details, see relevant barcode reader manual.
3.4.5.2 2-D Barcode Reader
Install the connecting cable
1. The appearance of barcode reader. Each part of the barcode reader: LED indicator, scan
window and scan trigger button.
1. LED Green: A barcode was successfully decoded.
Red: A data transmission error or reader malfunction occurred.
2. Scan window Scan the barcode.
3. Trigger Press to decode
2. Plug connecting terminal of the cable to the port of the barcode reader. Ensure the contact
works well.
3. Connect USB terminal of the connecting cable to the USB port of main unit.
System Installation 3-13
Page 50
4. Press scan trigger button to receive barcode when ultrasound device is running (without
installing driving program). For more operation details, see relevant barcode reader manual.
Disconnect the connecting cable
1. Using the tip of a screwdriver or some other tools with a sharp head, depress the cable’s
modular connector clip.
2. Carefully slide out the cable.
3.4.5.3 JADAK Barcode Reader
Supported Barcode Reader Model
The Ultrasound System supports the following barcode readers: HS-1M JDK-2413 and HS-1R
JDK-2601.
Operating System Version
If you need to use the JADAK barcode reader with the Ultrasound System, ensure that the
operating system version is as follows:
PC module (SOM5892-3517-M-8G) 2.0.11 or later versions
PC module (SOM-6897C3-T02A1E) 3.3.13 or later versions
If the version is lower than the requirements, upgrade the operating system.
3-14 System Installation
Page 51
For version upgrading, contact the Mindray service engineers. Select [Setup]→[About]→[About
Details]→[Operating System Version] to check the operating system of your device.
Setting Up the JADAK Barcode Reader (Taking HS-1R as an Example)
1. Turn off the power to the Ultrasound System;
2. Connect the appropriate interface cable to the barcode reader;
3. Plug the other end of the cable into a free USB port on the Ultrasound system;
4. Once the imager has been fully connected, power on the Ultrasound System.
HS-1M JDK-2413 Configuration
The barcode reader is configured through scanning the barcode. Ensure that the barcode reader is
properly connected to the Ultrasound System before scanning. After the barcode is successfully
scanned, the barcode reader buzzes, and the green indicator is On.
Configuration before use
1. Scan the following 1-D barcode to configure the barcode reader:
2. Scan the following 2-D barcode to enable the suffix:
3. Scan the following 2-D barcode to set the suffix to Enter:
Reset the barcode reader
1. If the barcode reader is in malfunction, scan the following 1-D barcode to reset to default
settings.
2. Follow the configuration steps above to reconfigure the barcode reader.
HS-1R JDK-2601 Configuration
The barcode reader is configured through scanning the barcode. Ensure that the barcode reader is
properly connected to the Ultrasound System before scanning. After the barcode is successfully
scanned, the barcode reader buzzes, and the green indicator is On.
Configuration before use
1. Scan the following 1-D barcode to configure the barcode reader:
System Installation 3-15
Page 52
Notes:
Users can customize the JADAK barcode reader based on
specific requirements or contact the Mindray service
2. Scan the following 1-D barcode to enable the suffix:
For scanning Barcode
For scanning RFID
engineers. This configuration guide is applicable for the
Ultrasound System only.
Reset the barcode reader
1. If the barcode reader is in malfunction, scan the following 1-D barcode to reset to default
settings.
2. Follow the following steps to reconfigure the barcode reader:
a) Scan the following 1-D barcode to configure the barcode reader:
b) Scan the following 1-D barcode to enable the suffix:
For scanning Barcode
For scanning RFID
c) Scan the following 1-D barcode to disable the prefix:
For scanning Barcode
For scanning RFID
3-16 System Installation
Page 53
WARNING
3.5 Wired Network Connection
Connect the USB plug of the USB to LAN Adapter to the USB port on the machine, and insert the
network cable into the LAN port of the adapter to get a wired network connection.
The LAN port of the adapter CAN ONLY be used to connect to wired network.
Device interconnection using the adapter through LAN port is forbidden;
otherwise, electric shock may result. If you intend to connect an external device,
please make sure it meets the requirement of STANDARD GB9706.15 or
IEC60601-1 chapter 16.
Note: the ultrasound system supports the following usb2lan devices for network connection:
highspeed usb ethernet control /
Anker (USB3.0 to Gigabit Ethernet Adapter) A7611
J5create (USB3.0 to Gigabit Ethernet Adapter) JUE130
uGreen (USB3.0 to Gigabit Ethernet Adapter) 20258
In case of interfering the network communication, only one usb2lan device should be connected to
the system each time
Manufacturer Model
3.6 System Configuration
3.6.1 Power-on Running
Connect the connecting terminal of the power adapter to the adapter port in the system. Use a
three-wire cable to connect the adapter with the external power supply. Ensure the connection of
ultrasound and optional device works well.
3.6.2 Enter Doppler
After the initialization process (about 1 minute), the system enters Doppler interface, as shown
below:
System Installation 3-17
Page 54
3.6.3 System Preset
1. Press [Setup]. Setup menu appears.
2. Click [System Preset] to enter System Preset interface.
3-18 System Installation
Page 55
To set the relevant information about fetal gestational age, fetal growth curve
The following settings can be performed on the System Preset interface.
Preset InterfaceDescription
Region
General
Image To set some general parameters in imaging modes.
Application
OB
Key Config To assign functions to footswitch and the user-defined keys.
Audio Control To assign functions to the speech commands.
Access Control To set the user account control relevant information.
Scan Code To set relevant information of barcode.
To set the hospital name, language, time zone, system time format, system
date format and system date/time.
To set patient information, exam setup, patient management, image storage,
display, system dormancy, etc.
To set the measure caliper, measure setting, follicle method, comment setting,
etc.
and fetal weight, etc.
3.6.4 Print Preset
Press <Setup> and click [Print Preset] to set video printer, graph/text printer parameters (set the
parameters according to the printer, and select the relevant printer services).
System Installation 3-19
Page 56
Print service setting
Add Service: click to add print service.
Remove Service: click to delete the selected print service.
Rename Service: click to rename the selected print service.
Default print service: click to set the selected print service as the default one.
Set Service Property: to preset the service property, such as paper size, paper
orientation, etc. set the print service according to the type. The print service include:
report print, digital image print and analog image print.
Image Settings
Click [Image Setting] to enter the page, you can set the brightness, contrast and saturation of
image printing, or you can use the default values.
3.6.5 Network Preset
3.6.5.1 iStorage Preset
Select [Setup]-[Network Preset]-[iStorage] to enter iStorage interface (see the figure below):
3-20 System Installation
Page 57
Service Name
The name of the iStorage service.
Name Description
IP Address IP address of the iStorage service device.
Port Port for transmitting.
Connect Click to verify connection.
Add Click to add the Network service to the service list.
Update To save the changed parameters.
Delete Click to delete the selected service from the service list.
Add an iStorage service
1. Set the network server properties as described above.
2. Click [Add] to add the service to the Service list.
Modify a network service
1. Select the service to be updated in the service list.
2. You can see properties in the Configure Service area.
3. Modify the parameters and click [Update] to update the setting.
3.6.5.2 MedTouch/MedSight Preset
You can set environment for MedTouch/MedSight here and then use the MedTouch/MedSight
function by mobile phone or tablet computers. See MedTouch/MedSight manual for details.
3.6.5.3 Q-Path
Overview
You can use the ultrasound system to check data on browser directly. After you have ordered
storage service of a network website service, you can check data using the website, authorized
System Installation 3-21
Page 58
account and password (provided by the service vendor).You can open the browser (Q-View) to
review previously sent DICOM data.
Q-path is a network server provided by Telexy Healthcare Inc. for digital image storage. Q-View is a
client viewing tool for the server. Telexy Healthcare developed technology and a command structure
that allows any Q-view enabled ultrasound system to access Q-path directly from the ultrasound
system using a single control. The primary purpose for Q-view is to provide remote access to
Q-path from the ultrasound system to complete the exam report on the ultrasound system and
submit for QA,
To access Q-Path on the M9 or M8 series, the user just opens the Q-View first and then enters the
URL, user account and password provided by Telexy Healthcare, and the M9 or M8 series software
system will call Q-View tool then for user Q-Path application.
For details, please contact Q-Path service provider.
Q-Path Basic Procedure
Set related setting: check the check box by "Enable Q-Path" in the path: “[Setup]→[Network
Preset]→ [Q-Path]”, as shown in the following:
Figure 1
The preset procedures are as follows:
1. Select [Enable Q-Path].
2. Enter the server address in the field box of "Server Address".
3. Select user type: Personal User or Default User.
Personal User: the personal user needs to enter the user name and password in
every-time login.
Default User: after the default user enters the user name and password in the field box of
the "User Name" and "Password", and click [Save], no login is required to access the Q-Path
server later.
4. Select an appropriate item from the drop-down list of "Available Items"
5. Select an exam mode in the left "Exam Mode" column.
6. Select a worksheet in the right "Worksheets" column.
3-22 System Installation
Page 59
assword On
Password On End
Sets whether to input the worksheet password after
7. Click [Save] to exit, and the system will shut down.
Parameter Description
Sets the sub URLs of "QVIew full" and "QView lite".
Advanced
The sub URL is set by default. Users can modify the
sub URL and click [Apply] to exit the "QView sub URL
setting" window.
Enable Direct Report
Worksheet Only
Sets whether to open the Q-Path server through
pressing the <Report> key.
Sets whether to directly enter the Worksheet interface
after opening the Q-Path server.
Sets whether to display the Signature field box in a
worksheet.
Press <Report>→[WorkSheet] or press <Review> →
P
Worksheet
[Report] → [WorkSheet], enter the worksheet
password in the field box, and click [OK].
Users can query the corresponding worksheet by
searching the worksheet password in the Q-Path
server.
Exam
ending an exam.
Password Visible Sets whether the password is visible.
Import
Imports a user-defined worksheet template from the
USB storage (downloaded from the Q-Path server).
Backup Backs up worksheets to the USB storage.
Restore
Restores the backup worksheet template from the
USB storage to the ultrasound system.
Delete Deletes a worksheet template.
Restore Factory Restores the worksheet template to the default state.
The operating procedures are as follows:
1. Set the DICOM storage server.
2. Send stored images or worksheet reports from iStation/Review/thumbnail area to the Q-Path
server.
3. Press the user-defined key for Q-Path or press <Report> to open the Q-View browser.
4. Log in to the Q-Path server through the Q-View browser to check the stored images and
worksheet reports.
5. Click [Esc] to exit the Q-View browser.
Note: the following prompt will display every time you enter Q-Path after powering on the machine.
Select "Yes" to enter the website service.
System Installation 3-23
Page 60
Figure 2
Press <ESC> or the user-defined key to exit the function.
Select “Enable User Account Login” and you are able to login using single user account.
NOTE: If you use the function to connect to websites other than Q-Path applications (website
setting in “[Setup]→[Network Preset]→ [Q-Path]”), the system will prompt the following
information:
Q-Path Functions using Q-view
When Q-Path is accessed from the ultrasound system using Q-view, access is restricted to the
active exam and only certain functions are available.
NOTE: What functions you can acquire by using Q-View is only related to what you have purchased
from Q-Path service provider. Please contact Q-Path service provider for details.
The primary purpose of accessing Q-path from an ultrasound machine using Q-view is to complete
the exam report template and submit the exam for review or quality assurance, eliminating the
necessity to log in to Q-path from a separate workstation.
Q-path Functions Available on the Ultrasound System:
Note: Only credentialed physicians are authorized to approve or sign a report.
Submit Exam for QA.
3.6.6 Network Configure
1. Press <Cursor> to show the cursor, click in the bottom bar to open the wireless network
manager.
2. Roll the trackball and press <Set> to select the target network, click [Connect] to connect to the
network.
When connecting an encrypted network, enter the password in the box first.You can select to
hide password characters or not.
3. The system tries to connect and the wireless manager icon turns into
or after successful connection.
4. Click [Refresh] to refresh the "Wireless Network Connection" list.
3-24 System Installation
. The icon turns into
Page 61
IP Config
Note:
When the system background is processing network task (DICOM sending for example),
Note:
If the IP address displays as 0.0.0.0, this means that the network is abnormal. The
reason for the failure may be disconnection or the system cannot obtain the IP address.
please do not enter network setting to change the IP, otherwise the background task may
fail. You can check if there are tasks undergoing in the task manager.
IP config is used for setting local network parameters, which is also applied to DICOM connection.
1. In Wireless network manager screen, click [IP Config] to open the page:
If “DHCP” is selected, the IP address will be automatically obtained from the DNS server.
If “Static” is selected (using a static IP address), enter the IP address.
IP address of the system should be in the same network segment with the server.
Subnet Mask: set different network segment.
Gateway: set the gateway IP.
2. Click [OK] to save current setting. Click [Cancel] to exit.
EAP Network (for Windows 7 system)
EAP (Extensible Authentication Protocol) is an authentification frame protocol that supports multiple
authentification methods. It is mainly used for network access authentification, and is applicable for
wireless enterprise security authentification.
1. In Wireless Network Connection screen, click [Eap Network] to open the page:
System Installation 3-25
Page 62
Import certificate: Click [Certificate Manage] to enter Certificates
page, click [Import…] to import root certification in “Trusted
Root Certification Authorities” page, then click [Import…] to
import personal certification in “Personal” page, and set Eap
network password.
Set Eap network: Click [Manage Wireless Network] to set.
Click [Add]->[Manually create a network profile] to set.
3-26 System Installation
Page 63
Network name: input Eap nerwork name;
Security type: WPA2-Enterprise;
Encryption type: AES;
Security key: keep blank;
Select “Start this connection automatically” and “Connect even if the network is not
broadcasting”.
Click [Next] to finish the network setting.
Click [Close] to exit.
Select Eap network, then press right <Set> key, and select [Properties] menu.
System Installation 3-27
Page 64
Enter “Security” page, select [Microsoft: Smart Cart or other certificate] from “Choose a
network authentication method” drop-down list.
Click [Settings], select trusted root certification from “Trusted Root Certification
Authorities” drop-down list, click [OK].
3-28 System Installation
Page 65
Click [Advanced Settings] to set in “Security” page. Select “Specify authentication
mode”, and select “user or computer authentication” from the drop-down list. Then click
[OK], close the setting page.
Select Eap network in the "Wireless Network Connection" list, click [Connect] to connect to the
network.
EAP Network (For windows 10 system)
Import certificate
1. In Wireless Network Connection screen, tap [EapNetwork] to open the Eap Network
Configpage:
2. Click [Certificate Manage] to enter Certificates page, click [Import…] to import root certification
in “Trusted Root Certification Authorities” page, then click [Import…] to import personal
certification in “Personal” page, and set Eap network password.
System Installation 3-29
Page 66
Set Eap network:
1. Login the system with the account of Service. Enter [Setup]→[Maintenance]→[Setup], and
click [Enable ms-setting].
2. Save the settings. The system is automatically powered off.
3. Power on the system. Enter [Setup]→[Maintenance]→[Setup]→[Enter Windows], input the
password. For details about obtaining password, please refer to “6.4 Enter Windows”.
4. Open Startup menu on the Windows interface. Select [Settings]→[Network & Internet]→[Wi-Fi]
→[Manage known networks]→[Add a new network], and set EAP network.
Network name: input Eap nerwork name;
Security type: WPA2-Enterprise;
Encryption type: AES;
Select “Connect automatically” and “Connect even if the network is not broadcasting”.
5. Click wireless network logo on the lower right corner of the Windows screen. Select the desired
EAP network, and click [Connect].
If the EAP network cannot be connected, operate according to the prompts on the screen
6. Double-click the tool “BackToDoppler” on the desktop to return to the Doppler interface.
7. Click [Disable ms-setting] on the maintenance screen. Save the settings. The system is
automatically powered off.
8. Power on the system. Select the desired EAP network on the wireless network connection
menu, and click [Connect].
To remove the added EAP network, perform steps 1-3. Open Startup menu on the Windows
interface. Select [Settings]→[Network & Internet]→[Wi-Fi] →[Manage known networks], select
the desired EAP network, and click [Forget].
3.6.7 DICOM/HL7 Preset
Note:
1. Click [DICOM Preset] to enter the DICOM Preset interface. Set DICOM service attributes first:
enter AE Title (use Application Entity title for ultrasound system), terminal and package size.
Only if DICOM basic option is configured, [DICOM Preset] is available.
2. Server setting
1) Enter the device name and the IP address;
2) Verify connection after entering the correct IP address by clicking [Ping].
3-30 System Installation
Page 67
3.
If the currently entered name has already existed, the system will pop up:
3) Click [Add] to add the server to the list if the connection works normally.
The following is an example:
Note:
2.DICOM communication port should be the same with the one on the server.
3. Click [DICOM Service].
1. AE Title should be the same with the SCU AE Title preset in the server
(PACS/RIS/HIS).
“The server name exists!”
When the system is configured with DICOM basic function module, and installed DICOM Worklist,
storage commitment, and Query/ Retrieve modules, the corresponding preset settings can be found
in DICOM Service screen.
System Installation 3-31
Page 68
Note:
The DICOM Service Setting is used to set properties of DICOM services. See the basic volume for
details.
Only if DICOM basic option is configured, Worklist page is available.
3.6.8 Security
Click [Security] on the Setup menu to enter the security-setting screen.
1. Drive Encryption/Secure Data Wipe
Encrypt the patient data stored in the hard disk. The system provides two encryption methods:
Factory Default and User Define.
Factory Default: the system is in factory state by default.
User Define: add a user-defined password.
1) Select [User Define]. If the patient data are already stored in the hard disk, the system prompts
the following message: (if no patient data are stored in the hard disk, perform steps 5 to 6
directly)
2) Click [OK] to return to the Security screen, click [Wipe], and the system prompts the following
message:
3-32 System Installation
Page 69
McAfee can also be installed after installation of Windows
3) Click [OK], and the system prompts the following message:
4) Click [Yes] to wipe the patient data.
5) Select [User Define] and click [Confirm].
6) Input the password and click [Confirm] to finish the password setting.
Notes 1. If you want switch to Factory Default, perform steps 1 to 6 again. The
password is the same as that of the User Define.
2. When you set password, multi-language and Chinese characters are not
supported.
2. Anti-Virus
The system provides two anti-virus software: Windows Defender and McAfee. They can effectively
prevent the ultrasound system from being attacked by virus, spyware, or other malware.
If the McAfee software is installed, the system displays “McAfee is installed”; if not, the system
displays “McAfee is not installed”. The McAfee software is an option. If you want to buy McAfee,
contact Mindray service engineers.
Notes: 1. McAfee cannot be uninstalled after successful installation.
2. After installing the McAfee, disable the Fast Startup function to
avoid McAfee failure.
3.
Defender.
4. If McAfee is improperly installed due to power-off, shut-down,
closing of cmd.exe, or any other abnormal operation during the
installation, please contact the Mindray service engineers.
3. Transmission Encryption
After accessing the network, click [VPN Config] to enter the "VPN Config" interface.
System Installation 3-33
Page 70
Item Description
Server IP
No driver: click [SetupDriver] to enter the
"TAP -Windows 9.21.2 Setup" interface, and do as
instructed.
Ready: the VPN is ready for use.
Status
Advance: VPN Advance Configuration
Connected: VPN is successfully connected.
Disconnected: VPN is disconnected.
Error: error connection.
Group
/
User Name
Password
Hide characters The password is displayed as *.
Connect/ Disconnect Connect or disconnect VPN.
3-34 System Installation
Page 71
Item
Description
Advance
Enters the "VPN Advance Config" interface.
Reset: if the system does not respond after you
click [Config], click [Reset].
Config: enters the "OpenConnect-GUI VPN client"
interface. For details about the settings, please
refer to the TAP manual.
Note: after exiting the "VPN Advance Config"
interface, you need to reboot the system;
otherwise, you cannot connect VPN normally.
Close Close the "VPN Config" interface.
Note: if the system is installed with McAfee, software like VPN that is provided by the third party will
be blocked. If users want to use VPN, select [Setup] → [Maintenance] → [Setup] → [Enter
Windows], and input password to enter Windows. Open E:\VPNDrivers\tap-windows.exe to start
installation.
3.6.9 System Information Verification
Product configuration, version number, hardware board and relevant driver information appear on
the system information interface. Confirm the relevant information on this interface.
1. Click [About] to enter the interface.
2. All details about the hardware board appear on [About Detail] interface.
1.
Note:
Confirm the system information before and after the software maintenance.
System Installation 3-35
Page 72
2. If required, the system reminds the user of saving the current system
information.
3-36 System Installation
Page 73
Master Board
Probe
Socket
Probe Board
TRA
FPGA
LCD Module
TR64 Board
AFE
AFE
64 Tx
(65~128)
CW
Module
TRB
FPGA
AFE
AFE
64 Tx
(1~64)
DSP
FPGA
System
monitoring
SM
(ARM)
COME Module
PHV Power Board
Control Panel
SSD
WiFi
LCD
Monitor
ECG Board
AC Adapter
Battery
Battery
AC
IN
DC
IN
POUT
128ch
POUT
64ch
Control
Power
4D&TEE
Signal
Control & Data Bus
XCVR
Control & Data Bus
XCVR
Power
Control
Signal
Clock
PowerLVDS
I2C
PCIE X4
PCIE X1
SATA
USB
Power
Power
Power
button
and
power
indicator
signal
Speaker
Power
Serial
port
DC-DC Module
4D&TEE
Communication
and control
USB 3.0
X2
HDMI
Pencil
Probe
Light Sensor
CLOCK
Fan
I2C
Power
supply
mgmt.
EC
(ARM)
PHV
(ARM)
Frontend
Backend
Power Supply
…
…
…
…
4 Product Principle
4.1 General Structure of Main Unit’s
Hardware System
Figure
system is shown in figure above. M9 system hardware consists of each following part:
Main board
DC-DC circuit, which offers power supply, charges or discharges the battery.
Functional circuit of back-end offers the functions of stereo audio, video audio, IO
Signal processing FPGA. It processes the beamforming signal and occupies the
4-1 Schematic diagram of system hardwareThe general structure of M9’s hardware
interface, etc.
functional controls except for transmitting and receiving functions.
Product Principle 4-1
Page 74
TRA FPGA is the FPGA which controls the transmission and the receiving.
65-128 transmitting/receiving channels.
Control interfaces related to probes.
4D&TEE functional circuit implements functions of 4D drive circuit and TEE
temperature measure, etc.
Power supply management (EC) implements functions of the power supply
management and control of starting/shutting down the main unit.
System monitoring (SM) implements the power supply monitoring and fan monitoring
Probe board connects the main unit with the probe.
PHV power supply board produces the programmed high-voltage needed in the
transmission.
COME module, also named as PC module, runs at Doppler platform.
As a main display unit, LCD module displays the image and inducts the environment
brightness.
ECG board offers ECG and respiratory signal acquisition and achieves the triggering
function with main unit.
4.2 Main Board
Main board includes: power supply, back-end and front-end.
4-2 Product Principle
Page 75
4.2.1 Power Supply
12VBUS
(
12
A)
Charger
Buck
Charger
2.5A Chg
4.6A Dsg
2.5A Chg
4.6A Dsg
19.0V/7.9A
AC/BAT
DCDC
DCDC
3V3STB(100mA)
5VSTB(500mA)
(240mA)
PWR_12V_EN_N
5VSTB_EN
DCDC
DCDC
DCDC
……
1V
1.2V
1.5V
5V
2V
3.3V
……
FPGA
COME module
(
PC module)
Gama
calibration
SSD
Wireless
net
adapter
USB
HDMI
LCD
Audio
circuit
Speaker
Circuit
control of
ultrasound
function
Control for
starting/shutting down
the device
Control
panel
Power
button
Power
supply
control
Button,
trackball
System
monitoring
SATA
PCIE
USB
HDMI
USB
HDA
LVDS
PCIE
LVDS
Figure 4-2 Principle diagram of power system
Function description:
It provides the power to start the device and the power when the main unit is in standby.
Battery charging or discharging circuit.
4.2.2 Backend
The backend is integrated with the circuit of the CPU module and main board.
Figure
4-3 Principle diagram of power system
Product Principle 4-3
Page 76
Function description:
USB ports
Provides two external USB 3.0 ports
Control panel port
Control panel port
Circuit Unit Function
The control for
Starting and
shutting down the
Takes charge of starting and shutting down the device and
controls the indicator for starting and shutting down the
device
device
System monitoring Takes charge of battery management, fan control, voltage
detection, battery indicator control,
voltage detection and temperature detection
Wireless net
WIFI and Bluetooth
adapter
SATA port SSD, mSATA type
HDMI Provides an HDMI port and supports 1920*1080 resolutions
(HD) output.
Audio Encodes and decodes the audio. Increases the power.
Signal process
Connects CPU module with the ultrasound circuit
FPGA
LCD port Connects to LCD monitor screen, displays parameter board
screen and detects of environment brightness.
Other circuits
related to
ultrasound functions
Transmitting/receiving control
Probe control
4D, TEE probe control (reserved function, to be
released)
PHV control
ECG control
4-4 Product Principle
Page 77
Signal
process
FPGA
Main board transmitting and receiving
High-
voltage
pulse
output
circuit
Low-
voltage
drive
circuit
ATGC
High-
voltage
isolation
AFE clip
x8
Clock
TRA
FPGA
Clock
circuit
64ch
RX,
64ch
64ch
64ch
CW
receiving
circuit
Connect to
probe via
TR 64
board
Control TR
64 board
Pencil
probe
receiving
switch
Pencil
probe
control
data
TRB
FPGA
65-128
channel
data
Transmitting
control
I/Qsignal
TX,64ch
PEN_RX
C
o
n
n
e
c
t
o
r
t
o
b
o
a
r
d
C
o
n
n
e
c
t
o
r
t
o
b
o
a
r
d
4.2.3 Front-end of Main Board
Fig 4-4 Systematical diagram of front-end of main board
The main functions of main board’s front-end unit:
Control FPGA (TRA, implements the control of 65-128 channels transmitting and receiving,
ATGC adjustment and CW receiving control).
Clock circuit
Transmitting circuit (transmits the pulse with the achievement of 5 electric levels)
High-voltage isolation
AFE (receives echo signal)
ATGC circuit (time gain compensation)
CW receiving circuit (CW signal receiving)
Product Principle 4-5
Page 78
4.3 TR64 Board
Main board
TRA
FPGA
Signal
process
FPGA
TR64 board
High-
voltage
pulse
output
circuit
Drive
circuit
High-
voltage
isolation
AFE clip
x8
Clock offered
by CPU
TRB
FPGA
Clock
circuit
65-128
RX,
64ch
64ch
64ch
CPU ATGC
Pencil
probe
transmitting
switch
Control
Data
Probe
board
Main
board
1-64
CW
transmitting
power
supply
Probe
control
SPI
TX,64ch
Transmitting
control
PEN_TX
C
T
B
C
T
B
C
T
B
Control signal
65-128
Fig
4-5 Principle diagram of TR64 board
The main functions of TR64 board:
Control FPGA (TRB, implements the control of 1-64 channels transmitting and receiving,
ATGC adjustment and CW receiving control).
Clock circuit provides the clock for transmitting/receiving chip.
Transmitting circuit (transmits the pulse with the achievement of 5 electric levels and CW)
High-voltage isolation
AFE (receives echo signal)
ATGC circuit (time gain compensation)
Probe port function (Probe recognition signal, power supply control of the probe)
4.4 Probe Board
Supports 192/128 array probe, phased array probe, bi-plane probe, and other nominal
probes, etc.
Implements the connection between the main unit and the probe.
4-6 Product Principle
Page 79
PHV power supply board
LDO
Flyback
circuit
Descending
circuit
Ascending
circuit
Socket
(board
to
board)
Main
board
MCU
P12V
LC
filtering
LC
filtering
LC
filtering
12VIN
12VIN
11V5VCC
CW_VIN1
CW_VIN2
CW_PWR_P
CW_PWR_N
AP100V
AN100V
PHV1P
PHV1N
PHV2P
PHV2N
UART_TX
UART_RX
UART_TX
UART_RX
CW_PWR_P
CW_PWR_N
AP100V
AN100V
PHV1P
PHV1N
PHV2P
PHV2N
DAC_CW
Over
current
and
over
voltage
signal
Over
current
and over
voltage
signal
Sampling
voltage
DAC_SDI
Power
supply
communication
&control
Implements the switch of 192 array probe within the probe. Probe board only supports
th
channel.
128
4.5 PHV Power Supply Board
Figure 4-6 Diagram of PHV power supply board
Function description:
PHV power supply board is the board that provides the power to the transmitting circuit.
AP100V, AN100V, PHV1P, PHV1N, PHV2P, PHV2N, CW_PWR_P and CW_PWR_N are
all outputs.
PHV output goes into the power supply by producing the circuit via LDO.
PHV input goes into the main board after being filtered.
4.6 ECG Unit
ECG module is designed for monitoring ECG signals and displaying ECG waveforms, which
serves as reference of ultrasound images. This module can synchronize with 2D images and
color flow images in real-time display. With R wave detected, ECG signals, after being
Product Principle 4-7
Page 80
Main Board
ECG leads
DC IN
Socket
port
Magnify Filtering
samp
ling
Heart rate
test
Serial port
ECG Board
Pwr
L
T
B
C
L
T
B
C
magnified, filtered and sampled, are sent to DSC nodule via series port UART. Triggered ECG
signals that are detected are sent to control system for control. Then, it starts scan transmitting.
ECG transducer can be connected externally in this module design, which is also compatible
with other monitoring products manufactured by Mindray.ECG unit structure is shown as below:
Figure 4-7 Principle diagram of ECG board
4-8 Product Principle
Page 81
Control Panel
Main Board
System status
indicator
Power button
TGC Board
Function
button
Button
scan
LED
backlight
drive circuit
LED backlight
Buzzer
Buzzer drive
circuit
PWM generation
circuit
FPGA
12
V
、
5V、
5V
_STB
MCU
SPI
Communication&control
Power supply
QWERTY
keyboard
L
T
B
C
Encoder
Button
Encoder
info
process
Rotate
Trackball
module
USB
USB
Right
button
of
trackball
Left
button
of
trackball
ADC
circuit
Screen closed m agnetic
detection switch
4.7 Control Panel
Control panel unit includes two boards, as shown below:
No. Board
Name
1 Control
Panel
2 TGC
Board
Description on Board Function
LCD monitor folding/unfolding detection, power button, trackball, standard
keys on keyboard, user-defined functional keys, backlight control, status
indicator of the device, encoder, buzzer, TCG detection, etc.
TCG adjustments in 8 segments
The functional diagram of the control panel unit is shown above. The main functions are described
2. MCU module implements USB communication, TGC detection, etc. FPGA and PSOC
communicate via SPI port: PSOC obtains detection result of FPGA via SPI port and sends the
result to the host by USB. PSOC obtains control information from the host via USB and sends it
to FPGA via SPI port. FPGA performs the control operations to corresponding units after
parsing control orders.
Product Principle 4-9
Page 82
Test board
of
environment
brightness
Parameter
board
+12V+3.3V
LVDS
3.3V
I2C
Backlight
control
3.3V
I2C
Main board socket
15.6"
1920*1080
LCD Monitor
LTBC
LTBC
LTBC
LTBC
3. TGC board implements the fix of TCG’s sliding rheostat.
4. Trackball sub-module communicates with the host via USB.
5. Sub-module of QWERTY keyboard adopts “thin film + silicon keys + structure keys” to
implement the standard keyboard.
6. Buzzer sub-unit adopts PWM waveforms that FPGA outputs to control the length and sound of
volume and frequency.
4.8 Main Display Unit
Fig 4-8 Principle diagram of monitor
The monitor unit consists of LCD screen, connecting wires, parameter board, environmental light
detecting board, etc.
Function description:
To make the display effect of LCD screen equal to that on the parameter board, parameter
board stores color temperature and gamma calibration data. It is necessary to update the
parameters in control panel for first time machine assembly or control panel/LCD screen
replacements.
Signal processing FPGA calibrates the video signal that CPU board outputs based on the
information from parameter board, and then drives LCD display screen. Brightness and power
supply of LCD display screen are also controlled by this FPGA.
Environmental light detecting board detects the brightness level of environment. To achieve
better display effect, main unit adjusts LCD brightness level according to the environment.
4-10 Product Principle
Page 83
Probe Extend Board
C probe socket
260pin
B probe socket
260pin
A probe socket
260pin
POUT signal switch
(relay set)
CPLD
Probe signal
4D&TEE signal
Probe control
circuit
Probe control signal
12V/5.7V
/+100V/-100V
128 channel
Transmitting/receiving signal
260PIN
probe
connector
Probe Board
Communication
&control
Power supply
Ultrasound signal
4D&TEE signal
switch
(relay set)
4.9 Probe Extension Board
Figure 4-9 Hardware diagram of probe extension board
Function description:
Supports 192/128 array probe, 4D probe, phased array probe, bi-plane probe, and other
nominal probes, etc.
Implements the switch of 192 array probe within the probe. Probe board only supports 128
channels.
Probe board contains three 260pin Mini probe sockets.
Supports the retrieving of three probes’ IDs and the switch among three probe socket. ID
recognition and the circuit of probe switch are independent to each other.
Product Principle 4-11
Page 84
VGA socket
Audio
socket
S-
Video
socket
FPGA
Video cutting, video
zooming, color space
change, sampling
HDMI
signal
receiving
Audio
encoder
S-Video
Video
encoder
HDMI
socket
Main Unit
(M9)
HDMI
socket
Communication
&control
Power supply
HDMI
5V
DDR2
LED
Power LED
VGA
Video encoder
Digital audio
signal
S-Video
VGA
Audio
Audio/
video transfer
module
4.10 Audio/Video Transfer Module
Figure 4-10 Function of audio/video extend module
Audio/video transfer board (or module) adopts non-standard HDMI input port. Main unit outputs
1920×1080p@60Hz HDMI 1.4a signal as input signal of the extender. The module supports line-out
video signal of analog stereo audio, xxxx×xxxx@60Hz analog RGB video signal, NTSC, S-Video
signal with format of PAL.
M9 audio/video transfer module includes: power supply, HDMI signal receiving, digital video signal
processing, and audio/video code.
Figure 4-11 Diagram of audio/video extend module
4-12 Product Principle
Page 85
4.11 USB HUB Board
USB main
port
USB sub-portUSB sub-port
USB sub-
port
USB sub-port
Repeater
/
Controller
Fig 4-12 Principle diagram of USB HUB board
The main functions of USB HUB board:
Compatible with the USB 2.0 standard;
One USB port can be subdivided for four USB ports (there are two ports in current products.)
Detect the plug and unplug of USB device via main port.
Charge the device connecting to USB via main port.
4.12 Trolley Power Supply
Trolley power supply provides the power to main unit system. AC-DC power supply provides the
power to main unit system when there is AC input. If there is no AC input, trolley power supply
provides the power to main unit system. Trolley power supply includes connecting board of trolley
power supply, AC-DC power supply, management board of trolley battery and two pieces of
chargeable lithium-ion batteries. The connection of each part is shown below:
Product Principle 4-13
Page 86
Connecting
board of trolley
power supply
Mgmt board of
trolley power
supply
AC-DC power
supply
Battery
A
Battery
B
J1
J1
J2
J1
J2
J3
J4
Fan
J5
J6
Indicator
signal
Power supply
and its status
indicator
J3
J2
J4
J3
AC OUT
Fan
power
supply
AC IN
AC IN
Relay
control
signal
EDC_19V
Power
supply and
signal of
battery A
Power
supply and
signal of
battery B
Figure 4-13 Diagram of related boards of trolley power supply
4.12.1 Connection Board of Trolley Power Supply
Main functions of connection board of trolley power supply:
AC input transfer
Auxiliary output control. Auxiliary output is enabled after starting main unit. Auxiliary output
disconnects when main unit is in standby or shuts down.
4.12.2 AC_DC Power Supply
AC_DC power supply implements power supply isolation and voltage transfer. It outputs 19 V DC
transferred from AC input. In terms of 150 W of main unit power and 50 W charging power, the
maximum power of AC_DC power supply is 200 W.
4.12.3 Management Board of Trolley Battery
The main functions of management board of trolley battery:
Implements AC power supply and auto switch of battery supply. AC power supply has the
priority over others.
Battery charge and battery management;
AC status and battery status indication;
Works with main unit system and recognizes the status of main unit system. Implements
4-14 Product Principle
auxiliary output control and low battery-consumption when the trolley is in standby.
Page 87
CPU module
Control for
starting/shutting
down device
Control
panel
S
3#S4#
POWER
_OK
Power
button
DC-DC
5VSTB
12V
Enabling
power
supply
3V3_STB
Function
circuit
Multi-
channel
power
supply
Principle diagram of management board of trolley battery is shown below:
Figure 4-14 Diagram of management board of trolley battery
4.13 System Power-on Control
Figure 4-15 System power-on diagram
Description of related controlling signals:
No. Controlling signal Description Comme
nts
1 PWR_BTN_N,
PWR_BTN#
Pulse signal that power-on button of control panel
produces passes to CPU board through FPGA,
and is used for starting the device.
2 S3# Output by CPU board, effectively represents that
CPU system is in standby (the electrical level of
S4# is high) and keeps 5VSTB powered on when
it is in standby via FPGA.
3 S4# Output by CPU board, effectively represents that
CPU system is in dormancy.
4 S5# The signal is not used currently
Product Principle 4-15
Page 88
Connect to AC
Press power button
(control panel)
Receive valid
PWR_BTN_N
(power supply mgmt
FPGA)
Power supply mgmt
sends valid PWRBTN#
signal
S3# and S4#invalid
(COME module
sends)
DC-DC 12V power-on
PWR_OK valid
(power supply mgmt
sends)
CPU start
(COME module)
DC-DC board sends
power supply of other
circuits
5 PWR_OK# Sent out by power management FPGA to CPU
board, indicates that the 12V is powered on.
Power supply of main unit/battery enables the start of device.
Power supply produces 5VSTB and 3.3VSTB as the AC inputs.
Unplug AC when shutting down the device. Power supply cuts off 5VSTB output, but only
keeps the output of 3.3VSTB. Only with power button pressed again, it’s re-powered on.
Unplug AC when the device is in standby. Batteries, for standby usage, provide the output of
5VSBT and 3.3VSTB.
The process of power-on is shown below:
4-16 Product Principle
Figure 4-16 Diagram of system power-on
Page 89
Probe connecting
board 1PCBA
Main board
LCD
Monitor
COME module
SSD
mSATA port
WiFi module
Battery
Battery
TR64 board
Probe board
ECG board
ECG
connecting
board
2*2*
110
mSA
TA
Mini
PCIE
2*78
2*60
6*40
Light
sensor
board
LCD
parameter
board
Wire
Power supply adapter
Audio/video transfer
module
USB HUB board
1*8
1*8
J6
J8
J17
J20
J22
J10
J24
J18
J1
J9
J12
J23
J5J4
J21
J1
J18
J2J1
J3
J3
Control panel
TGC board
Trackball
2116 small keyboard
Board to
board
socket
Board to
wire
socket
Board
Power
supply wire
material
Signal wire
material
Wire to
wire
socket
Reservation
design
Buzzer
Speaker
assembly
J15
J1
J1J1
J11
J2
J5
J6
J4
J3
J1
J2
J5
J6
Connecting board of trolley power supply
Trolley battery
management board
AC-DC module
Battery
Battery
Trolley LED
board
AC output
AC input
1*11
1*11
J2
J3
J10
J4
J5
J6
J7
J1
J1
J2
J4
J3
Probe extend module
2*40
2*
40
2*20
Probe connecting
board 2PCBA
M7 probe extender
4DConnecting board
PCBA
A probe
connecting
board PCB
M9 probe
extender
B connecting
probe
260
PIN
260
PIN
Probe extension
wire
Main unit
Trolley
PHV power supply
board
2*20
J7
J1
Fan assembly
4.14 Internal Connection Diagram of Main Unit
and Trolley
Figure
4-17 Internal connection diagram of main unit and trolley
4.15 Details in Main Unit’s Power-on
Start-up procedure of main unit and the performance of power supply & display in various steps are
shown below:
Fig
Product Principle 4-17
Page 90
Press Power
Button
EC ARM receives
signal
Effective
power-on
Exit, without
starting
N
Y
Enabling
CPU_PWR_BTN
COME module
receives signal
COME module
sends S3, and
starts power-on
flow
EC ARM receives
S3 signal
EC ARM enables
power supply of
main unit
Power supply
power-on
1Và……
EC ARM delay
DSP FPGA(PCIE
device)start
configuration
DSP FPGA(PCIE
device) complete
configuration
EC ARM sends
PWR_12V_OK
order
COME module
receives
PWR_12V_OK
COME module
,
start BIOS
DSP FPGA(PCIE
device) in non-
reset status
Control panel
power-on
5V 12V
Control panel
FPGA
configuration
Control panel
PSOC BOOT
Initialize control
panel
(white backlight
is on)
Control panel
ready
No sign on
monitor
Initialize display.
Turn off
backlight. Do not
output LVDS
signal
No sign on
screen
DSP FPGA sends
LVDS data to
LCD
Open DSP FPGA
Open LCD
backlight, adjust
brightness
Backlight of
monitor
brightens. No
display
Correct image
displays on
monitor(BIOS
boot screen)
COME module,
output boot
screen
Work status (below
Power
Button),green,blinks
Work status
indicator (below
Power Button),
green, lighting
COME module
,
complete BIOS
Complete OS
loading
Start loading OS
Display changes
of resolution
LOGO
(MINDRAY)
displays on
monitor
Adjust final
resolution
Enter OS
desktop
Image from
desktop displays
on monitor
No sign on
monitor
No sign on
monitor
Start Doppler
Doppler start
screen displays
on monitor
Hard disk
indicator,
green, blinks
AC is activeà
Active AC
indicator,green,
lighting
Battery supply
à
battery status
indicator,
green,lighting
4-18 Product Principle
Page 91
5 Checking Performance and
Functions
5.1 Description
The chapter describes checking methods to main functions and performance. The methods are
only for reference.
5.2 Checking System Status
5.2.1 Running Status
1. Power on/off normal (duration time is normal), no abnormal sounds or occasion occur
during normal operation.
2. After ultrasound system gets started, the fan starts working, and no abnormal noise is
heard when the fan is working.
3. Check whether product configurations and software versions are normal via [About]
interface.
4. Check whether contrast and brightness of the monitor are normal.
5. Check whether time and date are valid and correct.
6. Check whether all status indicators are normal.
7. Check all log records together with the users to confirm whether there are any
abnormalities.
5.2.2 Working Condition
Check the ambient temperature and humidity. The measurement related to security features is
sensitive to humidity. If the insulation feature of the system deteriorates due to the increase of
system service time or system malfunctions, the fluctuation range of measurement results are likely
to increase with the humidity increasing.
Checking Performance and Functions 5-1
Page 92
5.3 General Check
Check the control panel
Check the monitor
Check peripherals
Check I/O port
Check ECG
Press <Freeze> to enter frozen status and
Press <Measure> to enter measurement
status. Measure the vertical distance and
horizontal distance and perform other
. The
the trackball goes is same with that of the
5.3.1 Check Flow
5.3.2 Check Content
5.3.2.1 Check Control Panel
1. Check all buttons, keys and knobs
Follow the direction: left to right, and up to down.
2. Trackball function check
ProcedureChecking criteria
move the trackball.
trackball operations.
All keys and knobs are effective.
The move of the trackball is smooth
response to the cursor is rapid. The direction
cursor.
5-2 Checking Performance and Functions
Page 93
5.3.2.2 Check Monitor
nt named as
“Service”, select
[Test Main
Monitor] to check the monitor
[General]. Click
2 The adjoining dark dots are no more than 3
pairs, and there are no adjoining dark dots in
ProcedureChecking criteria
Monitor brightness adjustment
Monitor contrast adjustment
Monitor maintenance
Log on with the accou
[Setup]-[Maintenance]-
functions.
Height adjustment:
Auto adjustment: select [Setup]-[System
Preset]-[General]. Click [Brightness Auto
Adjustment] from [About].
Manual adjustment: press <Fn>+<
the brightness. Press <Fn>+<
brightness.
Select [Setup]-[System Preset]-
[+] from [Contrast] to increase the contrast. Click
[-] to decrease the contrast.
The monitor display works well after testing each
function. Bad points are:
1 Light dot is 0; blinking dot defect is 0.
image area.
> to increase
> to decrease the
3. 3 or 3 successive dark dots are no more than 0
pair.
The dark dots are no more than 7 and those in the
image area are no more than 2
5 The distance between bad dots is no less than
5mm.
Note: the image area refers to the area enclosed
by the rectangle with black/white background.
Checking Performance and Functions 5-3
Page 94
Procedure
Checking criteria
between footswitch and USB port. Check
[Key Config] and corresponding
color print; left
same with that on the screen of main
unit. The characters and images are
complete; the images are stable; no
5.3.2.3 Checking Peripherals
Footswitch: confirm the normal connection
the configuration status of footswitch under
implementation. (E.g. right key-image
freeze; middle keykey-white/black print).
Video printer: confirm the connection
between video printer and ultrasound
device works well and check the
implementation of each function.
Graph/text printer: confirm the connection
between graph/text printer and ultrasound
device works well and check the
implementation of each function.
Barcode reader: scan any piece of barcode
when the system is under running.
5.3.2.4 Checking ECG Module
Trigger the freeze key of the footswitch (right key).
Image freeze menu and freeze menu appear. The
image is unfrozen if the key is triggered again.
Trigger the print key of the footswitch (middle key).
Start color print.
Trigger the print key of the footswitch (left key). Start
black/white print.
Press the Print key which is setup well already to start
the print. The image has no defect or degradation.
Switch video output terminal. Then repeat the
previous steps.
Press the Print key which is setup well already to start
the print. The image has no defect or degradation.
The barcode information displays on the image
interface. The information is correct compared with
the data information of barcode.
ProcedureChecking criteria
Perform the following tests after the
system is configured with ECG
module:
Connects ECG lead.
Move the cursor to image menu.
Click [Physio] to enter its interface.
ECG activation, ECG waveform and heart icon appear on
the right corner of the interface.
The parameters for [Scan speed], [ECG gain] and [ECG
position] can be adjusted in real-time.
Review ECG signals.
5.3.2.5 Checking I/O Interface
Procedure Checking criteria
Checking I/O interface
Verify video/audio interface, USB port, etc.
VGA interface
Connect external VGA/LCD monitor (the resolution
should be one of them: 1920*1080,1024*768,
1280*1024,1280*720).
Other USB ports
The information on VGA/LCD monitor is
color cast; no shaking and no blinking.
Network connection and communication
work well.
USB storage and read work well.
5-4 Checking Performance and Functions
Page 95
Check image mode
Check basic
measurement
Check cine mode
Check probe usage
Image and imaging
management
Record and save the
check
Adjustment
Parameter Item
5.4 Functions Checking
Note: The chapter lists the system checking items with complete configurations and
describes them in details. If the items are not configured, the relevant tests can
be ignored.
5.4.1 Checking Flow
5.4.2 Checking Content
5.4.2.1 Imaging Mode
B-mode
In B mode scan, the image parameter area in the upper left corner of the screen displays the
Parameters that can be adjusted to optimize the B Mode image are indicated in the following.
Control panel and
image soft menu
adjustment
real-time parameter values as follows:
Items F D G FR DR iClear iBeam iTouch
Meaning Frequency Depth Gain Frame
Rate
Frequency (image quality), gain, depth, TGC, acoustic power, focus, scan
range, steer, ExFov, line density, dynamic range, iClear, persistence,
rotation/invert, iBeam, auto piecing, gray map, tint map, TSI, iTouch,
HScale, dual live, LGC, Echo, Boost.
B
Dynamic
Range
Display when the function is
activated.
Checking Performance and Functions 5-5
Page 96
Parameter Adjustment
Procedure
Press <B> button.
Enter B mode image. B mode interface appears.
me value of frequency is displayed in the image
parameter area in the upper right corner of the screen
Press <Focus Position> to
is displayed on the right side of the
on and move the trackball to adjust the scan
Steer
Dynamic Range
Rotate the knob clockwise to increase dynamic range; rotate the
represents iClear is disabled, and the bigger the value is the
Checking criteria
Frequency adjustment
B image soft menu-[Image
quality]
Gain adjustment G
Rotate <Gain> button
Depth adjustment D
Press <Depth> button.
TGC adjustment
Adjust 8 sliders on the control
panel
Acoustic power adjustment
B image soft menu-[Acoustic
power].
Focus
B image soft menu-[Focus
number]
adjust focus position.
The real-ti
(fundamental wave-F, Harmonic frequency-H). Values of frequency
vary depending upon the probe types.
Gain increases with rotating the knob clockwise;
Gain decreases with rotating the knob anticlockwise;
The depth of the image changes accordingly. Depth range varies
depending upon the probe types.
Push the slider right to increase the gain. The brightness of the
area becomes brighter.
Push the slider right to decrease the gain. The brightness of the
area becomes darker.
About 1.5s after the adjustment is finished, the TGC curve
disappears.
The system offers 34 values to adjust the acoustic power.
Acoustic power (AP) is displayed in real time in the upper part of
the screen.
Focus position/number adjustment
The focus position icon
image.
Scan range and FOV position
B image soft menu-[Scan
range].
Press <Steer> arrow button.
ExFov
B image soft menu-[ExFov]
Line Density
B image soft menu-[Line
Density].
B image soft menu-[Dynamic
Range].
iClear
B image soft menu-[iClear].
Image display adjustment
[Scan range] is
range.
To steer the beam the probe transmits.
Click [ExFov] on the soft menu to enable/disable the function.
The function determines the quality and information of the image.
Levels of line density: UH/ H/ M/ L.
knob counterclockwise to decrease dynamic range.
The adjusting range of parameter is 30-180 dB in increments of 5
dB.
The system provides 7 levels of iClear effects adjustment, Off
stronger the effect becomes.
5-6 Checking Performance and Functions
Page 97
B image soft
The system provides 4 values of iBeam in B mode. iBeam is
the same probe type, depth, invert status, rotation status and
Click [iTouch] on the image menu to adjust the gain in iTouch
Persistence
menu-[Persistence].
Rotation/Invert
B image soft menu-[L/R
Flip]/[U/D Flip].
iBeam
B image soft menu-[iBeam].
Auto Merge
B image soft menu-[Auto
Merge].
Gray Map
B image soft menu-[Gray Map]
Tint Map
B image soft menu-[Tint Map]
The system provides 7 level of persistence. The bigger the value
is the stronger the effect becomes.
To invert the image horizontally or vertically.
Image can be rotated by the angle of 0°, 90°, 180° and 270°.
When the image is rotated in the angle of 90° or 270°, the depth
scale is displayed on the upper part of the screen.
The “M” mark indicates the direction of the image; the M mark is
located on the top of the imaging area by default.
disabled when it is off.
In the Dual-split mode, when the images of the two windows have
magnification factor, the system will merge the two images so as
to extend the field of vision.
Turn on or off the function through the [Auto Merge] item in the
soft menu;
Adjust the gray from [Gray Map] on the soft menu;
There are 8 different maps available.
Select the tint map from [Tint Map] on the soft menu.
Turn on or off the tint map from [Tint Map] on the soft menu.
TSI
B image soft menu-[TSI].
iTouch
Press <iTouch> button
HScale
B image soft menu-[HScale]
Dual live
B image soft menu-[Dual live].
Select TSI from [TSI] on the soft menu.
The system provided 4 ways of optimization for specific tissues:
general, muscle, fluid and fat.
Press <iTouch> on the control panel to enter the iTouch status, the
symbol of which will be displayed in the image parameter area of
the screen.
status among -12 through 12dB.
Long press <iTouch> to exit iTouch mode.
Click [HScale] on the menu to display or hide the scale (HScale).
Enable [Dual Live] on soft menu, and dual-split window of images
are displayed on the screen.
Two pages of adjustable parameters are displayed on the touch
screen as well; where, shared parameters and left window
parameters are displayed on the B(L) page, while right window
parameters are displayed on the B(R) page.
Switch the display mode via <Switch Mode> and adjust the image
parameters.
Checking Performance and Functions 5-7
Page 98
Frequency
Color
Color Wall
Pulse
Gain, Depth, TGC, Steer, iTouch
In addition, the system provides several preset parameters for
LGC
B image soft menu-[LGC].
Images corresponding to four groups of parameters are displayed
on the soft menu (from left to right).
Click [LGC1-5] to adjust the parameters. To each segment of
image area, the system offers 5 values to adjust the gain. The
gain increases with the value becoming bigger.
imaging.
Echo Boost
[Echo Boost] is enabled when it is on in B mode.(Highlighted) the
B image soft menu-[Echo
system is in “Echo Boost” status.
Boost].
M mode
In M mode scan, the image parameter area on the right side of the screen displays the
real-time parameter values as follows:
Parameter Item F D G V DR
Meaning Frequency Depth M Gain M Speed M Dynamic Range
Parameters that can be adjusted to optimize the M Mode image are indicated in the following.
Adjustment Parameter Item
Control panel and
soft menu
adjustment
Gain, TGC, Depth, Focus Position, Display Format, Line Correlation, Tint
Map, Scan Speed, Edge Enhance, Dynamic Range, Gray Map.
Color mode
In Color mode scan, the image parameter area on the right side of the screen displays the
real-time parameter values as follows:
Parameter
F G WF PRF
Item
Meaning
Gain
Filter
Repetition
Frequency
Parameters that can be adjusted to optimize the Color mode image are indicated in the