Miele G 7883 CD Installation diagram

Page 1
Installatietekening
Plano de instalación
Asennusohje
CDN
Plan d`installation
Plano de instalação
Installasjonsplan
Pianta di installazione
Σχέδιο εγκατάστασης
Installation plan
G 7883 CD
Materialnr.
/
Mat. no. 9180800
Modification index
/
Version
00
Change no.
/
Alteration number
XX
Datum Zeichnung
/
Drawing Date
01.06.2005
Datum Legende
/
Legend Date
01.06.2003
Freigabe
/
Approval
Name:
Schröder
Name:
Götza
Installation plan G 7883 CD As per: 01.12.2012 Page 1
Page 2
Technical Data Sheet
Laboratory Glassware Washer
G 7883 CD
Legend:
Highlighted circled abbreviations: Connection required
EN
Broken line circled abbreviations: Connection optional or required depending on model version
Electrical
connection
1. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 400/50
9.7 3 × 16 5 × 2.5
1.7 Country variations:
2. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
3. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
Installation plan G 7883 CD As per: 01.12.2012 Page 2
Page 3
4. Standard voltage (at delivery) Connected load Fuse rating (integrated in plug) Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
3N AC 400/50
9.2 3 × 15/16 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 230/50
6.7 1 × 30 3 × 4
2.1
5. Standard voltage (at delivery) Connected load Fuse rating (integrated in plug) Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
AC 230/50
6.7 1 × 30 3 × 4
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3N AC 400/50
9.2 3 × 15/16 5 × 2.5
1.7
6. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
3N AC 400 - 415/50
9.2 - 9.9 3 × 15 5 × 2.5
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
Convertibl
e
V/Hz kW A mm² m
AC 230 - 240/50
6.7 - 7.2 1 × 30 3 × 4
2.1
7. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
AC 230 - 240/50
6.7 - 7.2 1 × 30 3 × 4
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
Convertibl
e
V/Hz kW A mm² m
3N AC 400 - 415/50
9.2 - 9.9 3 × 15 5 × 2.5
2.1
8. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
V/Hz kW A AWG m
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
Convertibl
e
V/Hz kW A AWG m
AC 208/60
6.0 2 × 30 3 × 10
1.9
Installation plan G 7883 CD As per: 01.12.2012 Page 3
Page 4
9. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
V/Hz kW A AWG m
AC 208/60
6.0 2 × 30 3 × 10
1.9
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
Convertibl
e
V/Hz kW A AWG m
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
10. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/50
5.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 200/50
5.7 2 × 30 3 × 4
2.1
11. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/60
5.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 200/60
5.7 2 × 30 3 × 4
2.1
12. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
13. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
Installation plan G 7883 CD As per: 01.12.2012 Page 4
Page 5
It is recommended to make electrical connection via a plug and socket so that electrical safety checks can be carried out easily. The socket must be accessible after machine installation. With a fixed installation, this must be via a main switch to be provided on site, which must totally isolate the machine from the mains with a contact gap of at least 3 mm. The fitting of an earth leakage circuit breaker is strongly recommended. If necessary, an equipotential bond with good contact connection must be provided in accordance with all appropriate national and local regulations.
Cold water
Two cold water connections are required with machines with steam condenser (DK). If necessary connect the second cold water hose via the Y-piece provided
Max. temperature Max. water hardness Min. flow pressure Max. pressure Flow rate On-site threaded union
Connection hose length - Disinfector (supplied) Connection hose length - Steam condenser (supplied with DK variant)
Machine is DVGW certified. Direct connection to mains water supply without special safety stopcock permitted.
Water connection Australia: To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
°C °dH kPa kPa l/min inch
mm mm
20 60 250 1000 8 ¾ male thread (USA: 11.5 NH) 1500 1500
Hot water
Max. temperature Max. water hardness Min. flow pressure Max. pressure Flow rate On-site threaded union
Connection hose length (supplied) Machine is DVGW certified. Direct connection to mains water
supply without special safety stopcock permitted. Water connection Australia:
To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
°C °dH kPa kPa l/min inch
mm
70 60 250 1000 8 ¾ male thread (USA: 11.5 NH) 1500
Demineralised water
Standard version: Min. flow pressure Min. flow pressure with extended water intake Max. pressure Flow rate Variant with demineralised water pump (aqua destillata pump) (ADP): Min. flow pressure Max. pressure On-site threaded union
Connection hose length (supplied) Water connection Australia:
To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
kPa kPa kPa l/min
kPa kPa inch
mm
250 100 1000 8
5 30 ¾ male thread (USA: 11.5 NH) 1500
Installation plan G 7883 CD As per: 01.12.2012 Page 5
Page 6
Drain water
Two drain hoses with DK machine variants Drain hose (inner dia. × thickness × l)
Drain pump (supplied) Steam condenser (supplied with DK variant) Max. drain pump head height from bottom edge of machine Max. brief flow rate per hose Hose connector on site for drain hose (outer dia. × l) Drain pump Steam condenser
mm mm m l/min
mm mm
22 × 6 × 1500 22 × 6 × 1500
1.0 50
22 × 30 22 × 30
Cabinet vent (optional)
Required on variants without DK (steam condenser) Vent connection, 2 (oval), for hose diameter (Connection hoses not supplied as standard) Use genuine Miele connection hoses.
Temperature - Average / Brief max. Relative air humidity - Average / Brief max.
I. Connection to an external vent system with fan: Flow rate per machine Vent coupling (with open air inlet) to be provided on site.
II. Vent direct to atmosphere without fan: Flow rate - Drying unit Max. permitted duct pressure loss Do not combine vent ducts! Ducts must be laid separately!
Condensate must not flow back to the machine! The vent duct must first rise vertically and then slope downwards away from the machine. A condensate trap should be provided at the lowest point.
An adequate supply of fresh air must be available at the installation site to replace that extracted by the air vent.
mm
°C/°C %/%
m³/h
m³/h Pa
25
70/95 80/100
70
55 10
Machine data
Height without lid Height with lid (optional) Width Depth Depth with removable side panels (standard delivery) Net weight Floor load in operation Min. installation location access width including transport pallet Min. installation location access height including transport pallet
mm mm mm mm mm kg N mm
mm
820 850 900 600 700 120 2000 800
1100
When using plinths (UC 30-90/60-78, UC 30-90/70-78) these must be secured to the floor. It is essential to fasten the machine to the plinth. The electrical, water, drain and exhaust connections must be made in accordance with all appropriate local and national safety regulations and applicable standards. During installation it is essential to follow the installation instructions provided! Modification rights reserved! Dimensions in mm.
Installation plan G 7883 CD As per: 01.12.2012 Page 6
Page 7
Caractéristiques techniques
Laveur pour verrerie de laboratoire
G 7883 CD
Légende :
Sigles entourés d’un cercle : raccordement nécessaire
FR
Sigles entourés d’un cercle en pointillés : Raccordement optionnel ou en fonction de l’appareil
Raccordement
électrique
1. Tension (état départ d’usine) Valeurs de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07) RN-F) sans fiche
V/Hz kW A mm² m
3N AC 400/50 9,7 3 × 16 5 × 2,5 1,7 Il en découle pour les autres pays :
2. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3N AC 400/50 7,3 3 × 16 5 × 2,5 1,7
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
3 AC 230/50 7,3 3 × 20 5 × 2,5 1,7
3. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3 AC 230/50 7,3 3 × 20 5 × 2,5 1,7
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
3 N AC 400/50 7,3 3 × 16 5 × 2,5 1,7
Installation plan G 7883 CD As per: 01.12.2012 Page 7
Page 8
4. Tension (état départ usine) Valeur de raccordement Fusibles (intégrés dans la fiche) Diamètre min. câble Longueur câble de raccordement avec fiche
V/Hz kW A mm² m
3N AC 400/50 9,2 3 × 15/16 5 × 2,5 1,7
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
AC 230/50 6,7 1 × 30 3 × 4 2,1
5. Tension (état départ usine) Valeur de raccordement Fusibles (intégrés dans la fiche) Diamètre min. câble Longueur câble de raccordement avec fiche
V/Hz kW A mm² m
AC 230/50 6,7 1 × 30 3 × 4 2,1
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
3N AC 400/50 9,2 3 × 15/16 5 × 2,5 1,7
6. Tension (état départ usine) Valeur de raccordement Fusibles (intégrés dans la fiche) Diamètre min. câble Longueur câble de raccordement avec fiche
V/Hz kW A mm² m
3N AC 400-415/50 9,2-9,9 3 × 15 5 × 2,5 2,1
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement avec fiche
modifiables
V/Hz kW A mm² m
AC 230-240/50 6,7-7,2 1 × 30 3 × 4 2,1
7. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement avec fiche
V/Hz kW A mm² m
AC 230-240/50 6,7-7,2 1 × 30 3 × 4 2,1
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement avec fiche
modifiables
V/Hz kW A mm² m
3N AC 400-415/50 9,2-9,9 3 × 15 5 × 2,5 2,1
8. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement sans fiche
V/Hz kW A AWG m
3 AC 208/60 6,0 3 × 20 4 × 12 1,9
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement sans fiche
modifiables
V/Hz kW A AWG m
AC 208/60 6,0 2 × 30 3 × 10 1,9
Installation plan G 7883 CD As per: 01.12.2012 Page 8
Page 9
9. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement sans fiche
V/Hz kW A AWG m
AC 208/60 6,0 2 × 30 3 × 10 1,9
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement sans fiche
modifiables
V/Hz kW A AWG m
3 AC 208/60 6,0 3 × 20 4 × 12 1,9
10. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3 AC 200/50 5,7 3 × 20 5 × 2,5 2,3
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
AC 200/50 5,7 2 × 30 3 × 4 2,1
11. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3 AC 200/60 5,7 3 × 20 5 × 2,5 2,3
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
AC 200/60 5,7 2 × 30 3 × 4 2,1
12. Tension (état départ usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3N AC 380/60 6,7 3 × 16 5 × 2,5 2,3
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
3 AC 220/60 6,7 3 × 20 5 × 2,5 2,3
13. Tension (état départ d’usine) Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
V/Hz kW A mm² m
3 AC 220/60 6,7 3 × 20 5 × 2,5 2,3
Tension Valeur de raccordement Fusibles Diamètre min. câble Longueur câble de raccordement (H05(07)RN-F) sans fiche
modifiables
V/Hz kW A mm² m
3N AC 380/60 6,7 3 × 16 5 × 2,5 2,3
Installation plan G 7883 CD As per: 01.12.2012 Page 9
Page 10
Il est conseillé de raccorder l’appareil par un dispositif de
prises afin que les contrôles de sécurité puissent être réalisés plus facilement. Les prises doivent être accessibles
une fois l’appareil installé.
En cas de raccordement fixe, l’utilisateur devra prévoir un
interrupteur principal avec séparation du réseau sur tous les pôles et une largeur d’ouverture de contact min. de 3 mm.
Pour plus de sécurité, il est fortement conseillé d’installer un
disjoncteur de courant à défaut en amont Le cas échéant il faudra également installer une liaison équipotentielle conforme aux normes nationales et disposant d’une bonne liaison de contact.
Eau froide
Deux raccordements à l’eau froide sont nécessaires avec
condensateur de vapeur (DK).
Raccorde le cas échéant le deuxième tuyau d’eau froide par
le "Y" joint. Température max.
Dureté d’eau max. autorisée Pression min. Pression max. Débit Raccord fileté à prévoir par l’utilisateur
Longueur tuyau de raccordement (fourni) Longueur tuyau de raccordement condensateur de vapeur (fourni, modèle DK)
Laveur certifié DVGW. Raccordement direct sans armature de sécurité possible.
°C °dH kPa kPa l/min pouces
mm mm
20 60 250 1000 8 ¾ externe (USA: 11,5 NH)
1500 1500
Eau chaude
Température max. Dureté d’eau max. autorisée Pression min. Pression max. Débit Raccord fileté à prévoir par l’utilisateur
Longueur tuyau de raccordement (fourni) Laveur certifié DVGW.
Raccordement direct sans armature de sécurité possible.
°C °dH kPa kPa l/min pouces
mm
70 60 250 1000 8 ¾ externe (USA: 11,5 NH)
1500
Eau déminéralisée
Modèle standard : Pression min. Pression max.en cas de prolongation de temps d’entrée d’eau Pression max. Débit Variante avec pompe de circulation pour eau dém. (ADP): Pression min. Pression max. Raccordement fileté à prévoir par l’utilisateur
Tuyau de raccordement (fourni)
kPa kPa
kPa l/min
kPa kPa pouces
mm
250 100
1000 8
5 30 ¾ außen (USA: 11,5 NH)
1500
Vidange
Deux tuyaux de vidange fournis avec modèle DK. Tuyau de vidange (di × s × l)
pompe de vidange (fournie) condensateur de vapeur (fourni avec le modèle DK) hauteur de refoulement pompe de vidange à partir de rebord inférieur de l’appareil max.
Débit par tuyau de vidange momentanément max.
Embout à olive à prévoir par l’utilisateur pour tuyau de
vidange (da × l) Pompe de vidange
Condensateur vapeur
mm mm m
l/min
mm mm
22 × 6 × 1500 22 × 6 × 1500 1,0
50
22 × 30 22 × 30
Installation plan G 7883 CD As per: 01.12.2012 Page 10
Page 11
Aération (optionnel)
Nécessaire pour le modèle sans raccordement pour condensateur de vapeur. Deux raccords par appareil (ovals) pour un diamètre de tuyau de : (les tuyaux de raccordements ne sont pas fournis). Pour le raccordement utiliser des tuyaux Miele.
température moyenne / sur une courte période max. : taux d’humidité de l’air moyen relatif / sur une période courte max.
I. Raccordement à un dispositif d’aération externe avec ventilateur : Débit dispositif d’aération externe : Hotte (ouverture air additionnel) à prévoir par l’utilisateur
II. Aération vers l’extérieur sans ventilateur : Débit dispositif de séchage Perte de pression max. autorisée Ne pas réunir les conduits d’aération de différents appareils ! Evacuer l’air appareil par appareil.
Eviter un retour de condensation vers l’appareil. Poser le
conduit d’aération en déclivité et vider l’eau à l’endroit le plus
bas.
Il faut aérer le local dans lequel se trouve l’appareil en
fonction de la quantité d’air évacuée..
mm
°C/°C %/%
m³/h
m³/h Pa
25
70/95 80/100
70
55 10 Données appareil
Hauteur d’encastrement Hauteur, tôle de protection incl. Largeur Profondeur Profondeur avec bandeaux latéraux démontables (état départ usine) Poids net Charge au sol en fonctionnement
Largeur d’encombrement min. y compris palette de transport Hauteur d’encombrement min. y compris palette de transport
mm mm mm mm mm
kg N mm mm
820 850 900 600 700
120 2000 800 1100
En cas d’utilisation des châssis (UC 30-90/60-78, UC 30-90/70-78) il faut veiller à une bonne liaison au sol. Fixation de l’appareil obligatoire.
Les installations ne doivent être effectuées que par des techniciens concessionnés en conformité avec les normes et les prescriptions de sécurité applicables ! Suivez les indications de la notice d’installation pour l’installation ! Sous réserve de modifications ! Dimensions en mm
Installation plan G 7883 CD As per: 01.12.2012 Page 11