Installation plan G 7883 CD As per: 01.12.2012 Page 1
Technical Data Sheet
Laboratory Glassware Washer
G 7883 CD
Legend:
Highlighted circled abbreviations: Connection required
EN
Broken line circled abbreviations: Connection optional or required depending on model version
Electrical
connection
1. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 400/50
9.7 3 × 16 5 × 2.5
1.7 Country variations:
2. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
3. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
Installation plan G 7883 CD As per: 01.12.2012 Page 2
4. Standard voltage (at delivery) Connected load Fuse rating (integrated in plug) Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
3N AC 400/50
9.2 3 × 15/16 5 × 2.5
1.7
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 230/50
6.7 1 × 30 3 × 4
2.1
5. Standard voltage (at delivery) Connected load Fuse rating (integrated in plug) Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
AC 230/50
6.7 1 × 30 3 × 4
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3N AC 400/50
9.2 3 × 15/16 5 × 2.5
1.7
6. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
3N AC 400 - 415/50
9.2 - 9.9 3 × 15 5 × 2.5
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
Convertibl
e
V/Hz kW A mm² m
AC 230 - 240/50
6.7 - 7.2 1 × 30 3 × 4
2.1
7. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
V/Hz kW A mm² m
AC 230 - 240/50
6.7 - 7.2 1 × 30 3 × 4
2.1
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
Convertibl
e
V/Hz kW A mm² m
3N AC 400 - 415/50
9.2 - 9.9 3 × 15 5 × 2.5
2.1
8. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
V/Hz kW A AWG m
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
Convertibl
e
V/Hz kW A AWG m
AC 208/60
6.0 2 × 30 3 × 10
1.9
Installation plan G 7883 CD As per: 01.12.2012 Page 3
9. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
V/Hz kW A AWG m
AC 208/60
6.0 2 × 30 3 × 10
1.9
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
Convertibl
e
V/Hz kW A AWG m
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
10. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/50
5.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 200/50
5.7 2 × 30 3 × 4
2.1
11. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/60
5.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
AC 200/60
5.7 2 × 30 3 × 4
2.1
12. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
13. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Convertible
V/Hz kW A mm² m
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
Installation plan G 7883 CD As per: 01.12.2012 Page 4
It is recommended to make electrical connection via a plug and socket so that electrical safety checks can be carried out easily. The socket must be accessible after machine installation. With a fixed installation, this must be via a main switch to be provided on site, which must totally isolate the machine from the mains with a contact gap of at least 3 mm. The fitting of an earth leakage circuit breaker is strongly recommended. If necessary, an equipotential bond with good contact connection must be provided in accordance with all appropriate national and local regulations.
Cold water
Two cold water connections are required with machines with steam condenser (DK). If necessary connect the second cold water hose via the Y-piece provided
Max. temperature Max. water hardness Min. flow pressure Max. pressure Flow rate On-site threaded union
Connection hose length - Disinfector (supplied) Connection hose length - Steam condenser (supplied with DK variant)
Machine is DVGW certified. Direct connection to mains water supply without special safety stopcock permitted.
Water connection Australia: To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
°C °dH kPa kPa l/min inch
mm mm
20 60 250 1000 8 ¾ male thread (USA: 11.5 NH) 1500 1500
Hot water
Max. temperature Max. water hardness Min. flow pressure Max. pressure Flow rate On-site threaded union
Connection hose length (supplied) Machine is DVGW certified. Direct connection to mains water
supply without special safety stopcock permitted. Water connection Australia:
To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
°C °dH kPa kPa l/min inch
mm
70 60 250 1000 8 ¾ male thread (USA: 11.5 NH) 1500
Demineralised water
Standard version: Min. flow pressure Min. flow pressure with extended water intake Max. pressure Flow rate Variant with demineralised water pump (aqua destillata pump) (ADP): Min. flow pressure Max. pressure On-site threaded union
Connection hose length (supplied) Water connection Australia:
To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
kPa kPa kPa l/min
kPa kPa inch
mm
250 100 1000 8
5 30 ¾ male thread (USA: 11.5 NH) 1500
Installation plan G 7883 CD As per: 01.12.2012 Page 5