To avoid the risk of accidents or
damage to the machine it is
essential to read these instructions
before it is installed, commissioned
and used for the first time.
en - GB, AU, NZ
M.-Nr. 07 776 620
Contents
Warning and safety instructions .....................................4
Symbols on the appliance .........................................8
Intended Use .....................................................9
Guide to the machine .............................................10
This appliance complies with all
relevant legal safety requirements.
Please note that incorrect use can
lead to personal injury and damage
to property.
To avoid the risk of accidents and
damage to the appliance, please
read these instructions carefully
before starting to use it. Keep these
instructions in a safe place for
reference, and pass them on to any
future user.
Correct application
This washer-disinfector is designed
~
for use with the applications described
in these Operating Instructions only.
Alterations or conversions to the
machine, or using it for purposes other
than those for which it was designed,
are not permitted and could be
dangerous.
This machine must only be used for
cleaning and disinfecting instruments
or medical devices if the manufacturer
has stated that they are suitable for
machine processing. Manufacturer's
cleaning and maintenance instructions
for instruments etc. must also be
observed.
Miele cannot be held liable for damage
caused by improper or incorrect use or
operation of the machine.
This machine is intended for indoor
~
use only.
This machine is not designed for
~
use in mobile installations unless a risk
assessment of the installation has been
carried out by a suitably qualified
engineer.
notes to avoid injury and damage.
This machine should be
~
commissioned and then maintained by
a Miele authorised and trained service
technician only. A Miele service
contract is recommended to ensure
compliance with GLP Guidelines and
the Medical Devices Directive.
Unauthorised repairs could be
dangerous, for which the manufacturer
cannot be held liable.
Do not install the machine in an area
~
where there is any danger of explosion
or of freezing conditions.
The electrical safety of this machine
~
can only be guaranteed if connected to
a correctly installed earthing system on
site. It is most important that this basic
safety requirement is present and
tested regularly, and where there is any
doubt the on-site electrical system
should be inspected by a qualified
electrician.
The manufacturer cannot be held liable
for damage or injury caused by the lack
of or inadequacy of an effective
earthing system (e.g. electric shock).
A damaged or leaking machine
~
could be dangerous and compromise
your safety. Disconnect the machine
from the mains immediately and call the
Miele Service Department.
Personnel operating the appliance
~
should be trained regularly. Untrained
personnel must not be allowed access
to the machine or its controls.
4
Warning and safety instructions
Take care when handling chemical
~
agents such as cleaning agent,
neutralising agent, rinsing agent etc.
These may contain irritant or corrosive
ingredients.
Please follow the manufacturer's safety
instructions. Wear protective gloves
and goggles. With all chemical agents,
the manufacturer's safety instructions
and safety data sheets must be
observed.
The machine is designed for
~
operation with water and recommended
additive chemical agents only. Organic
solvents and flammable liquid agents
must not be used in this machine.
These could cause an explosion,
damage rubber and plastic
components in the machine, and cause
liquids to leak out of the machine.
The water in the cabinet must not be
~
used as drinking water.
Be careful when sorting items with
~
sharp pointed ends and positioning
them in the machine that you do not
hurt yourself or create a danger for
others. Sharp knives etc. should be
placed in baskets with the pointed ends
facing downwards.
When using this machine in the
~
higher temperature ranges, be
especially careful not to scald or burn
yourself or come into contact with
irritant substances when opening the
door. Where disinfecting agents are
used there is a danger of inhaling toxic
fumes.
Where there is a risk of toxic or
~
chemical substances occuring in the
suds solution (e.g. aldehyde in the
disinfecting agent), it is essential to
regularly check door seals and make
sure that the steam condensor is
functioning correctly.
Opening the machine door during a
programme interruption carries
particular risks in such circumstances.
Should personnel accidentally come
~
into contact with toxic vapours or
chemical agents, follow the emergency
instructions given in the manufacturer's
safety data sheets.
Mobile units, modules, inserts and
~
the load must be allowed to cool down
before they are unloaded. Any water
remaining in containers could still be
very hot. Empty them into the wash
cabinet before taking them out.
After using the hot air drying unit,
~
open the door to allow the everything in
the cabinet from the load itself to the
mobile units, modules and inserts to
cool down.
Steam heating is permissible up to a
~
pressure of 1000 kPa (10 bar). This
corresponds to a water steam
temperature of 179°C.
Never clean the machine or near
~
vicinity with a water hose or a pressure
washer.
The machine must be disconnected
~
from the mains electricity supply before
any maintenance or repair work is
carried out. Do not reconnect it until the
maintenance or repair work has been
successfully completed.
5
Warning and safety instructions
The following points should be
observed to assist in maintaining
quality standards when processing
medical devices, in order to protect
patients, and to avoid damage to the
loads being cleaned.
If the machine is being used for
~
disinfection in accordance with official
regulations on the control of epidemics,
the steam condenser and its
connections to and from the wash
cabinet must be cleaned and
disinfected whenever any repairs are
carried out or parts replaced.
If it is necessary to interrupt a
~
programme in exceptional
circumstances, this may only be done
by authorised personnel.
The standard of cleaning and
~
disinfection in the disinfection
programmes must be routinely
confirmed by the user. The process
should be validated on a regular basis,
and checked against documented
control results. Chemical disinfection
procedures should also be validated
using chemical or bio indicators.
For thermal disinfection, use
~
temperatures and temperature holding
times to achieve the required infection
prophylaxis in accordance with current
health and safety regulations.
Disinfection programmes which use
~
a chemical disinfecting agent at
moderate temperatures such as 65 °C
or lower are not recognised by §18 IfSG
for disinfection. They should only be
used for items which cannot withstand
the higher temperatures used by
thermal disinfection. The range of
effective disinfection is based on claims
made by the producer of the
disinfecting agent. Their instructions on
handling, use and effectiveness must
be observed. The use of chemical
disinfection procedures is the
responsibility of the operator.
Theatre (OP) shoes should only be
~
cleaned and disinfected in a machine
installed specifically for this purpose.
Only use chemical agents
~
formulated for special processes and
approved by Miele for use with this
machine. Use of unsuitable agents
could adversely affect the components
of the machine. In the event of any
damage or material deterioration
please contact your Miele application
specialist.
Pre-treatments with cleaning or
~
disinfecting agents can create foam, as
can certain types of soiling and
chemical agents. Foam can have an
adverse effect on the disinfection and
cleaning result.
Foam must not be able to escape
~
from the wash cabinet. It would hinder
the correct functioning of the machine.
Check the process used regularly to
~
monitor foaming levels.
6
Warning and safety instructions
To avoid the risk of damage to the
~
machine and any accessories used
with it caused by chemical agents,
soiling and any reaction between the
two please read the notes in "Chemical
Processes and Technology".
Where a chemical agent is
~
recommended on technical application
grounds (e.g. a cleaning agent), this
does not imply that the manufacturer of
the machine accepts liability for the
effect of the chemical on the items
being cleaned.
Please be aware that changes in
formulation, storage conditions etc.
which may not be publicised by the
chemical manufacturer, can have a
negative effect on the cleaning result.
When using cleaning agents and
~
specialised products it is essential that
the manufacturer's instructions are
followed. Chemicals must only be used
for the purpose they are designed for
and in the situation specified, to the
exclusion of other chemicals, to avoid
such dangers as chemical reactions
and material damage.
In critical applications where very
~
stringent requirements have to be met,
it is strongly recommended that all the
relevant factors for the process, such
as chemical agents, water quality etc.
are discussed with the Miele
Application Technology specialists.
If the cleaning result is subject to
~
particularly stringent requirements, a
regular quality control test should be
carried out by the user to ensure that
required standards of cleanliness are
being achieved.
Mobile units, modules and inserts
~
should only be used for the purpose
they are designed for.
Hollow instruments must be thoroughly
cleaned, internally and externally.
Empty any containers or utensils
~
before arranging them in the machine.
Do not allow any remains of acids or
~
solvents, and in particular hydrochloric
acid or chloride solutions, to get into
the wash cabinet. Similarly avoid any
materials with a corrosive effect. The
presence in compounds of any solvents
should be minimal (especially those in
hazard class A1).
Ensure that solutions or steam
~
containing hydrochloric acid do not
come into contact with the steel outer
casing of the machine, to avoid any
corrosion damage.
Please follow the advice on
~
installation in these instructions and the
separate Installation Instructions.
Using accessories
Only use genuine Miele original
~
accessories with this machine. Consult
Miele on the type and application of
such equipment.
Only use Miele mobile units,
~
modules and inserts in this machine.
Using accessories made by other
manufacturers, or making modifications
to Miele accessories, can result in
unsatisfactory cleaning and disinfecting
results, for which Miele cannot be held
liable. Any resultant damage would also
invalidate the machine guarantee.
7
Warning and safety instructions
Only use chemical agents which
~
have been approved by their
manufacturer for use in the application
you are using. The chemical agent
manufacturer is responsible for any
negative influences on the material the
load is made from and for any damage
they may cause to the machine.
Symbols on the appliance
Warning:
Observe the operating
instructions.
Warning:
Danger of electric shock.
Disposing of your old machine
Please note that the machine may
~
have contamination from blood or other
bodily fluids in it and must be
decontaminated before disposal.
For environmental and safety reasons
ensure the machine is completely
drained of any residual water, chemical
residues and cleaning agent. Observe
safety regulations and wear safety
goggles and gloves.
Make the door lock inoperable, so that
children cannot accidentally shut
themselves in. Then make appropriate
arrangements for its safe disposal.
For tank system machines ensure that
any water is emptied out of the tank.
The manufacturer cannot be held
liable for damage caused by
non-compliance with these Warning
and Safety instructions.
8
Intended Use
This washer-disinfector can be used to
clean, rinse, disinfect (thermally and
chemically) and dry a wide range of
reusable medical devices.
Follow the manufacturer's instructions
(according to EN ISO 17664) on how to
process their items by machine.
Examples of application areas:
surgical instruments,
–
minimally invasive instruments,
–
anaesthetic and intensive care
–
instruments,
– baby bottles and teats,
– stainless steel and aluminium
containers (including anodised
aluminium),
– operating theatre shoes.
Processing medical devices by
machine achieves reproducible results,
and should be used in preference to
processing them by hand.
Where disinfection is required for the
protection of staff and/or patients, a
thermal disinfection programme, e.g.
the DESIN vario TD programme, should
be selected.
The parameters for thermal disinfection
according to EN ISO 15883-1 (A
0
concept) vary depending on the
disinfection standard required. They
are 80 °C (+ 5 °C, - 0 °C) with a 10 min
holding time (A
600) or 90 °C (+ 5 °C,
0
- 0 °C) with a 5 min holding time
(A
3000). An A0value of 3000 is
0
suitable for deactivating the HBV virus.
National health and safety regulations
regarding disinfection must also be
observed.
The cleaning programme as well as any
chemical agents must be chosen
according to the type of soiling and
load being processed.
The agents used for processing the
products should be selected to suit the
level of cleaning required as well as the
method of analysis used.
The cleaning result must ensure that
the instruments can be disinfected
correctly, that no residues are left
behind, that subsequent sterilisation
can be carried out and that the
instruments can be used again safely.
Medical devices are best processed
using the DESIN vario TD programme,
or the ORTHOVARIO or OXIVARIO
programme where applicable.
The use of a suitable carrier (mobile
unit, module, insert etc.) is important to
ensure the adequate cleaning of the
load. Examples are given in the section
"Areas of application".
The washer-disinfector is programmed
to carry out the final rinse with mains
water or with processed water of a
quality to suit the application (e.g. aqua
destillata, purified water, de-ionised
water, demineralised water).
This machine complies with EN ISO
15883 for validation purposes.
9
Guide to the machine
Unclean side
a Handle
b Profitronic electronic controls (see
the Programming manual for more
information)
c Controls
10
d Master switch (with emergency
cut-off function for service work)
e Drop-down door (closed)
f Service panel
Unclean side
Guide to the machine
a Salt container connection point
(for the water softener)
b Filter combination
c Drop-down door (open)
d Containers for dispensing systems
DOS 1 / DOS 3
(DOS 2 / DOS 4 optional)
11
Guide to the machine
Clean side (G 7824 only )
a Handle
b Door release
c Drop-down door (closed)
12
d Printer (optional)
G 7823: on the unclean side
e Service panel
Controls
Guide to the machine
a Display
with screen saver;
i.e. background lighting switches off automatically after
15 mins. Press any button to switch the display back on
again.
Faults occurring during operation are shown in the display.
They are listed in a chart in the Programming manual.
b On/Off button (I-0)
c Cursor button: left 1
Moves the cursor to the left:
– to the previous menu point
– to the previous parameter
– to the previous input position
13
Guide to the machine
d Cursor button: right 2
Moves the cursor to the right:
– to the next menu point
– to the next parameter
– to the next input position
e Minus button 3
– Selects programmes from position 24 upwards
– Scrolls back page by page in menus
– Is used for entering numbers and letters
– Alters pre-settings e.g. service parameters
f Plus button 4
– Selects programmes from position 24 upwards
– Scrolls forward page by page in menus
– Is used for entering numbers and letters
– Alters pre-settings e.g. service parameters
g Door switch 5
hStart button 6
– Starts programmes
– Activates input mode
– Confirms activated values and settings
– Confirms menu points for entry into the relevant
sub-menu
14
iStop button 7
– Cancels a programme
– Exits from input screen without saving
– Exits from menu screen
j Service interface 8
kProgramme selector
Selects programme places 1-23
Water softener softener (Optional)
In order to achieve good cleaning
results, the washer-disinfector needs to
operate with soft water. Hard water
results in the build-up of calcium
deposits on instruments, etc, and in the
washer-disinfector.
Mains water with a hardness level
higher than 0.7 mmol/l (4° d – German
scale) needs to be softened. This takes
place automatically in the integrated
water softener.
The water softener requires
–
reactivation salt.
The washer-disinfector must be
–
programmed to correspond to the
water hardness in your area.
– Your local water authority will be able
to advise you on the water hardness
in your area.
The water softener is set ex-works for a
hardness level of 3.4 mmol/l or 19°d.
If your water hardness differs from
this (even if it is below 0.7 mmol/l or
4°d) you will need to programme the
level into your machine.
Where the water hardness fluctuates
e.g. between 1.4 - 3.1 mmol/l (8 - 17 °d
German scale) always programme the
washer-disinfector to the higher value
(3.1 mmol/l in this example).
For future servicing it is useful to make
a note of your water hardness level.
Enter your water hardness level here:
°d or mmol/l
Setting the water softener
When the machine is first
commissioned the Miele service
technician has to set the machine for
your local water hardness level (see
Programme manual,
Operation/Reactivation).
Reactivation display
After a certain number of cycles the
message REACTIVATION will appear in
the display to warn you that the water
softener is depleted and cannot supply
any more softened water. As soon as
the programme has finished
reactivation salt will need to be
replenished.
If this cannot be done immediately, and
further cleaning cycles have been
carried out, the reactivation process will
need to be carried out twice in
succession.
The built-in water softener has
settings from 1 °d - 60 °d (0.2 -
10.8 mmol/l).
15
Water softener softener (Optional)
Reactivating the water softener
Please only use special coarse grained
reactivation salt with granules of
approx. 1-4 mm. Do not use other types
of salt, e.g. table salt, agricultural or
gritting salt. These could contain
components which are insoluble in
water which could result in damage to
the water softener. If in doubt consult
the Miele Professional Department.
The salt container holds approx. 2 kg of
salt.
If only fine grained reactivation salt
is available please consult the
MIELE Professional Sales office for
advice.
Salt with granules ü 4 mm must not
be used in this machine.
Inadvertently filling the salt
,
reservoir with cleaning agent can
cause serious damage to the water
softener and damage the filter insert,
and cause pressure to build-up in
the salt container. To avoid the risk
of injury please exercise caution
when removing the salt container as
there may be irritant alkaline
solutions present.
Before filling the salt container make
sure that you have picked up the
right packet of reactivation salt.
16
Water softener softener (Optional)
Filling the salt container
^ Unscrew the filter insert from the salt
container and remove.
^ Fill the salt container with reactivation
salt and screw the filter insert back in
place.
Position the salt container as follows
^ Remove any mobile units from the
cabinet.
^ Unscrew the plastic cap situated in
the top right-hand side of the cabinet.
,
There will be a small amount of
residual water in the cap. Take care
as it may be hot from the previous
programme.
^
Screw the salt container firmly onto
the socket.
17
Water softener softener (Optional)
Run the reactivation programme
Close the door.
^
Open the stopcocks.
^
Select the REACTIVATION
^
programme.
The programme will run automatically.
Then:
Switch off the machine.
^
Carefully unscrew the salt container,
^
making sure any water pressure still
remaining can depressurise. Do not
use force!
If the container will not come off
manually please contact the Miele
Service Department.
^ Replace the plastic cap.
^ Push the mobile unit back into the
cabinet.
The water pressure (flow pressure at
the take off point) must be at least
100 kPa (1 bar).
If it is below100 kPa (1 bar) or if it
fluctuates a lot, the water softener
will not work properly. After
reactivation, salt remains may still be
found in the salt container.
To use up salt remains and to rinse
out the water softener, the
REACTIVATION programme must be
carried out again.
^ Turn off the stopcocks.
^ Empty the salt container. Do not
empty it into the wash cabinet.
^
Rinse the salt container and filter
insert with clean water.
18
Automatic mobile unit recognition (Optional)
In operating level C the automatic
mobile unit recognition system (AWK)
allocates programme places from 1 15 to mobile units with the correct
coding.
The unit coding (on the mobile unit) and
the programme place with the
corresponding programme (in the
electronic control unit) have to match
each other.
To do this:
Each mobile unit must be coded
^
before being used for the first time
(see "Coding the mobile unit"), and
^ The programme for which the mobile
unit is coded must be allocated to the
correct programme place.
See "System function - Selector switch
organisation" in the Programming
manual for information on how to
change programme positions.
Before starting a programme it is
,
absolutely essential that you check
that the programme shown in the
display is the correct one for the
mobile unit being used.
Otherwise inadequate cleaning or
disinfection could be the result.
Please make sure, that the places
assigned for the programmes using
Automatic mobile unit recognition
are not changed around arbitrarily.
19
Automatic mobile unit recognition (Optional)
Mobile unit coding
The automatic mobile unit recognition
feature assigns a programme place to
a mobile unit. The mobile units must be
coded with a magnetic strip (via a Bit
combination).
In operating level C, the only
programme available for a coded
mobile unit is the one assigned to the
corresponding programme place.
After a coded mobile unit has been put
into the machine and the door closed,
the automatic mobile unit recognition
system will select the allocated
programme.
Make sure that there are no
,
small metallic objects or instrument
parts attracted to the magnetic strip,
in particular to the underside of it.
Any metallic objects on the strip can
result in the coding being incorrectly
read.
Coding is effected through 5 Bits:
Bits 1 to 4 define the mobile unit
–
code,
– Bit 5 serves as a control (Parity Bit).
Bit 6 controls the volume of inflowing
water and the circulation pump for
mobile units with side coupling (WAK).
Its settings cannot be altered.
Mobile units without side coupling are
coded with tracks which do not contain
Bit 6.
Press 6 to start the programme.
20
Automatic mobile unit recognition (Optional)
Mobile unit coding strip
15 different codes can be set. They are
assigned to programme places 1 to 15.
Under "System function, Selector switch
organisation" the matching
programmes have to be put into the
first 15 programme places.
Programme
place
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Bit1Bit 2Bit 3Bit 4Bit 5
0
0
0
I
0
I
0
I
0
I
0
I
0
I
0
I
0
I
0
0
0
I
0
I
I
0
I
0
I
I
I
I
0
0
0
0
0
I
0
I
I
0
I
0
I
I
I
I
Parity
Bit
0
0
I
0
I
0
0
0
I
0
0
0
0
0
I
0
I
I
0
I
0
I
I
I
0
I
I
I
I
I
0
I
Setting mobile unit coding
To set or alter the coding of a mobile
unit with automatic recognition (AWK),
proceed as follows:
^ Unscrew the track with AWK (using
an allen key) and remove from the
retainer.
The coded total must be an even
number.
The settings on Bit 6 (WAK) are not
included in this calculation.
If the coded total gives an odd number,
the message CHECK MOBILE UNIT
RECOGNITION appears.
If the mobile unit code equals 0, the
message NO MOBILE UNIT
RECOGNITION appears. In neither
case can the programme be started,
The mobile unit recognition function
must be re-set.
^
Remove the magnetic strip from the
track.
21
Automatic mobile unit recognition (Optional)
Place the track in the holder in the
Set the programme place coding.
^
^
Put the magnetic strip back in the
track.
^
mobile unit and screw firmly in place.
Bit 6 (WAK) is not a component part
of the modifiable magnetic strip.
Make sure that you code mobile
units with side connection with a
track where Bit 6 is set to I.
Mobile units without side coupling
must be coded with a track without
Bit 6.
The magnetic strips on G 7823 /
G 7824 / G 7825 / G 7826 must have
grey magnets.
The magnetic strips on G 7827 /
G 7828 must have black magnets.
The magnetic strip must be placed
in the track such that the Bit coding
set out in the chart is visible through
the round windows in the track.
22
Areas of application
This washer-disinfector can be fitted
with a variety of mobile units, which can
be equipped with a variety of modules
and inserts depending on the type and
shape of items requiring cleaning and
disinfection.
Select mobile units, modules and
inserts which are appropriate for the
application.
Notes on the individual areas of
application and examples of loading
are given on the following pages.
Before starting a programme you
should carry out a visual check on
the following:
– Is everything correctly
loaded/connected for cleaning?
– Are the spray arms clean and do
they rotate freely?
–
Are the filters clean? Remove any
coarse soiling and clean them
properly if necessary.
At the end of each programme:
Carry out a visual check of the load
–
for cleanliness.
Check that all hollow shafted
–
instruments are still securely located
on their jets.
Any hollow instruments that have
,
become disconnected from their
adapters during processing must be
re-processed.
Check that the lumen of hollow
–
instruments are free of obstruction.
– Check that jets and connectors are
securely held in position in the
baskets or inserts.
Protein test
Cleaning results should be subjected to
periodic protein tests, e.g. with the
Miele test kit.
–
Is the adapter connecting the water
supply to the spray arms/jets
correctly connected?
–
Are all chemical containers
sufficiently filled?
23
Areas of application
Loading the machine
Arrange the load so that water can
–
access all surfaces. This ensures
that it gets properly cleaned.
Do not place items to be cleaned
–
inside other pieces where they may
be concealed.
Hollow instruments must be
–
thoroughly cleaned, internally and
externally.
Ensure that instruments with long
–
narrow hollow sections can be
flushed through properly before
placing them in inserts or connecting
them to jets.
– Hollow vessels should be inverted
and placed in the correct mobile
units, modules and inserts to ensure
that water can flow in and out of
them unrestricted.
A cover can be used to reduce the
risk of glass breakage during the
wash process.
–
Lightweight items should be secured
with a cover net (e.g. an A 6) and
small items placed in a mesh tray to
prevent them blocking the spray
arms or being attracted by the
magnetic strip on the automatic
mobile unit recognition system.
–
Deep-sided items should be placed
at an angle to make sure water runs
off them freely.
–
Tall, narrow pieces should be placed
in the centre of the baskets. This
ensures good water coverage.
Mobile units with an adapter must
–
engage correctly.
Modules must be correctly
–
connected in the mobile unit.
The spray arms must not be blocked
–
by items which are too tall or which
hang down in their path. If in doubt,
test for free movement by manually
rotating the spray arms.
It is advisable to use only instruments
–
made of special application steel
which are not susceptible to
corrosion.
– Instruments which cannot withstand
high temperatures should be
chemically disinfected.
Disposable instruments must not
,
be put into the machine for
processing.
Preparing the load
Empty all containers before loading into
the machine (paying particular attention
to regulations regarding infectious
diseases and epidemics).
,
Ensure that acids and residual
solvents, especially hydrochloric
acid or chlorides, cannot get into the
wash cabinet.
Unloading instruments
Instruments should be unloaded dry
ready for further processing.
24
Areas of application
Surgical instruments (OP)
Surgical instruments should be stored
for as short a time as possible before
machine processing.
The OXIVARIO or ORTHOVARIO
processes should be used for
instruments where there is a long delay
between the time they are used and the
time they can be reprocessed. See
"Special processes" at the end of this
booklet.
Disinfection of surgical instruments and
of those used for minimally invasive
surgery should preferably take place
thermally.
Fully demineralised water should be
used for the final rinse when possible to
ensure no marks are left on the load
and to avoid corrosion. If the water
used contains more than 100 mg
chloride/litre there is danger of
corrosion.
When cleaning narrow lumen
instruments e.g. those used for
minimally invasive surgery,an
intensive internal cleaning result is
imperative. The Vario-TD and
OXIVARIO programmes are the only
ones which offer a thorough enough
level of cleaning for this. It is essential
that instruments are loaded as directed
and that the cleaning agent used is
suitable for the programme and for the
sensitive instruments being processed.
The final rinse must be carried out
using fully demineralised water with a
conductivity level of ~15 ZS/cm
(microsiemens per centimeter).
Instruments with very narrow lumen
must be manually pre-rinsed where
necessary. Follow the instrument
manufacturer's instructions on how
to handle them.
OP-containers can be disinfected
thermally in the CONTAINER
programme, or chemically. If anodised
aluminium containers are used the final
rinse must be carried out with fully
demineralised water. These containers
must not be processed with an alkaline
cleaning agent in a 93 °C programme
with 10 minutes holding time
(according to § 18 IfSG).
Mobile units for surgical instruments
and containers are supplied with
their own operating instructions.
25
Areas of application
Ophthalmology
Opthalmic instruments should be
cleaned and disinfected in a special
mobile injector unit.
The top level is fitted with various
connections for hollow instruments, e.g.
rinsing and suction hand pieces, and
cannulae.
Instruments are secured to the hose
connectors in the mobile injector unit
using the silicone holders and
connectors in the rack.
The lower level of the mobile injector
unit is designed to take E 441/1 inserts
or E 142 mesh trays for processing
solid instruments.
,
Plastic fibre cover nets must not
be used in washer-disinfectors
which are used to process hollow
ophthalmic instruments with narrow
lumen.
The mobile injector unit for
ophthalmic instruments is supplied
with its own operating instructions.
26
Anaesthetic instruments (AN)
Anaesthetic instruments should be
thermally disinfected using the
DES-VAR-TD-AN programme.
If this is not possible, they can be
processed using a chemical
disinfection programme. The water
level in the machine should be
increased when using a chemical
disinfection programme. To establish
the effectiveness of disinfection, consult
the manufacturer of the disinfecting
agent being used.
If the process is not to be followed
by sterilisation, the load should be
dried completely to avoid the
development of water-borne
bacteria.
A sufficient drying time is, therefore,
absolutely essential.
Areas of application
Mobile units for anaesthetic
instruments are supplied with their
own operating instructions.
27
Areas of application
Baby bottles (BC)
Baby bottles can be processed in
E 135 containers, wide-necked teats in
an E 364 insert and screw connection
teats in an E 458 insert.
Ensure the level marker on the
–
bottles is machine-washable.
If there is a delay of 4 hours or more
–
before bottles can be washed, fill
them with water to prevent residues
from drying on.
If the process is not to be followed
by sterilisation, the load should be
dried completely to avoid the
development of water-borne
bacteria.
A sufficient drying time is therefore
absolutely essential.
Inserts for baby bottles and teats are
supplied with their own operating
instructions.
Loading in an E 555
Bottom level with3xE135
28
Top level with2xE135
E 750 frame
^ Place the frame centrally in the top
level of the E 555 mobile unit.
^ Hook one E 135 container onto each
side of the frame.
The angled positioning of the
containers ensures that the jets of water
reach right into the bottles.
Areas of application
The two blocked corners in the
inserts should be located over the
positions marked with an X on the
drawing. The jets of water cannot
get into the bottles properly in these
positions to clean them.
29
Areas of application
Operating theatre shoes (OS)
Polyurethane operating theatre shoes,
and/or insoles should be cleaned and
chemically disinfected at 60 °C. These
can only be thermally disinfected (using
programme SHOE-TD-75/2) if the
manufacturer states that they are made
of a suitably heat-resistant material.
To establish the effectiveness of
chemical disinfecting agents for
disinfection consult the manufacturer.
Theatre shoes should be cleaned
and disinfected in a machine
installed specifically for this
purpose only.
Moble unit E 550 with a suitable insert
e. g. E 730, can be used for this
purpose.
A large amount of fluff can build up
in the machine when cleaning
theatre shoes. The filters in the wash
cabinet should therefore be
checked regularly, and cleaned
when necessary. See "Cleaning and
care".
30
Transfer trolley for loading and
unloading the machine
When a disinfection process is
,
being carried out according to
nationally laid down hygiene
standards (e.g. § 18 IfSG), the
contaminated surfaces of the trolley
must be disinfected, after the
machine has been loaded, using a
spray disinfectant in accordance
with national health and safety
regulations and § 18 IfSG.
Disinfecting agents listed by DGHM
(German institute for microbiology
and hygiene) are also suitable.
The height of the transfer trolley can be
adjusted using the feet.
The height should be set so that the
open machine door is held underneath
the side catches on the transfer trolley.
If the machine is not fitted on a plinth
the position of the foot pedal on the
transfer trolley may need to be
changed.
Areas of application
^ Unscrew foot pedal a.
^ Loosen counter nut b on adjusting
screw c.
^ Screw adjusting screw c further into
the foot pedal, so that the stopping
point is reached earlier.
^ Re-secure the adjusting screw.
^ Screw the foot pedal back on.
31
Areas of application
Hang the mobile unit into the locating
^
slots on the transfer trolley using both
hooks.
^ To lift the mobile unit step down on
the transfer trolley foot pedal.
^ Push the transfer trolley under the
open door on the washer-disinfector
as far as it will go. The mobile unit will
then sit over the door.
^ Secure the brakes on the wheels.
32
^
To lower the mobile unit step right
down on the foot pedal.
^
At the end of the programme push
the transfer trolley up to the
washer-disinfector making sure that
the door of the machine is held
underneath the side catches on the
transfer trolley.
^
Then pull the mobile unit out as far as
it will go onto the open door, so that it
can be lifted up and taken away by
the transfer trolley.
Chemical processes and technology
General notes
ProblemHow to resolve it
If seals, hoses and plastics in the
washer-disinfector suffer damage they
will not be water-tight and will not
function correctly.
Heavy foaming during a programme
affects cleaning and rinsing results.
Foam escaping from the wash cabinet
can cause damage to the
washer-disinfector.
Cleaning processes cannot be
regulated and validated where there
has been a build-up of foam..
Corrosion to stainless steel in the wash
cabinet and to accessories can give
them a different appearance:
– Rust (red marks / discolouration),
– black marks / discolouration,
– white marks / discolouration (etched
surface).
Corrosive pitting can lead to the
washer-disinfector not being
water-tight. Depending on application
corrosion can influence cleaning and
rinsing results (laboratory analysis) or
cause corrosion to stainless steel items
in the cabinet.
Establish the cause of the damage
–
and rectify it.
See information regarding "Chemical
agents“, "Soiling“ and "Reaction
between chemical agents and soiling“.
Establish the cause of the foam and
–
rectify it.
Check the process used regularly to
–
monitor foaming levels.
See information regarding "Chemical
agents“, "Soiling“ and "Reaction
between chemical agents and soiling“.
– Establish the cause of the corrosion
and rectify it.
See information regarding "Chemical
agents“, "Soiling“ and "Reaction
between chemical agents and soiling“.
33
Chemical processes and technology
Chemical agents
ProblemHow to resolve it
The ingredients in chemical agents
have a strong influence on the longevity
and functionality (throughput) of the
dispensing system. The dispensing
system (hoses and pumps) should be
set up for a particular type of chemical
agent.
General types:
Alkaline to pH neutral products,
–
acidic to pH neutral products,
–
hydrogen peroxide.
–
Chemical agents can damage
elastomers and plastics in the
washer-disinfector and accessories.
The following oxidising chemical agents
can damage elastomers (hoses and
seals) and plastics in the
washer-disinfector:
–
Nitric acid,
–
peracetic acid,
–
products containing active chlorine.
Use Miele approved chemical agents
–
in this machine.
Carry out a regular visual check of
–
the dispensing system for any
damage.
Regularly check the flow rate of the
–
dispensing system.
– Use Miele approved chemical agents
in this machine.
– Carry out a regular visual check of
any accessible elastomers and
plastics for damage.
The wash temperature should be
limited to:
– 50 °C Nitric acid,
– 35 °C with peracetic acid,
–
80 °C with products containing
active chlorine.
The following chemical agents can
release large amounts of oxygen:
–
Hydrogen peroxide,
–
peracetic acid.
34
–
Only use approved processes such
as OXIVARIO or OXIVARIO PLUS.
–
The wash temperature must be less
than 70 °C when using hydrogen
peroxide.
–
Please contact Miele Service or your
Miele Application Professional for
advice.
Chemical processes and technology
Chemical agents
ProblemHow to resolve it
The following chemical agents can
cause large amounts of foam to build
up:
Cleaning agents and rinsing agents
–
containing tensides,
emulsifiers.
–
Foam can occur:
in the programme block in which the
–
chemical agent is dispensed,
in the following programme block if it
–
has been spilt,
– in the following programme with
rinsing agent if it has been spilt.
De-foaming agents, especially silicone
based ones can cause the following:
– Deposits to build up in the cabinet,
– deposits to build up on the load,
– damage to elastomers and plastics
in the washer-disinfector,
– damage to certain plastics (e.g.
polycarbonate and plexiglass) in the
load being processed.
Process parameters in the wash
–
programme, such as dispensing
temperature, dosage concentration
etc. must be set to ensure the whole
process is foam free or very low
foaming.
Please observe chemical agent
–
manufacturer's instructions.
– De-foaming agents should be used
in exceptional cases only, for
instance when absolutely essential
for the process.
– The wash cabinet and accessories
should be periodically cleaned
without a load and without
de-foaming agent using the
ORGANICA programme.
–
Please contact Miele Service or your
Miele Application Professional for
advice.
35
Chemical processes and technology
Soiling
ProblemHow to resolve it
The following substances can damage
elastomers (hoses and seals) and
plastics in the washer-disinfector:
Oil, wax, aromatic and unsaturated
–
hydrocarbons,
emollients,
–
cosmetics, hygiene and care
–
products such as creams (analytical
applications).
The following substances can lead to a
heavy build-up of foam during washing
and rinsing:
– Some disinfecting agents and
dishwashing detergents,
– reagents for analysis e.g. for
microtiter plates,
–
cosmetics, hygiene and care
products such as shampoos and
creams (analytical applications),
–
active foaming agents such as
tensides.
Refit the washer-disinfector with oil
–
resistant elastomeres.
Depending on usage wipe the lower
–
door seal on the washer-disinfector
periodically with a lint-free cloth or
sponge.
Clean the wash cabinet and
accessories without a load using the
ORGANICA programme.
Prepare the load
–
using the "OIL" programme (where
this is available) or use a special
programme that dispenses
emulsifiers.
– Thoroughly rinse items in water
beforehand.
– Select a cleaning programme with at
least one short pre-rinse in cold or
hot water.
– Depending on application use
de-foaming agents that do not
contain silicone oils.
The following substances cause
corrosion to stainless steel in the wash
cabinet on accessories:
–
Hydrochloric acid,
–
other substances containing
chlorides such as sodium chloride
etc.,
–
concentrated sulphuric acid,
–
chromic acid,
–
Particles of iron and swarf.
36
–
Thoroughly rinse items in water
beforehand.
–
Place items drip dried into mobile
units, baskets and inserts and then
place these in the wash cabinet.
Chemical processes and technology
Reaction between chemical agents and soiling
ProblemHow to resolve it
Natural oils and fats can be emulsified
with alkaline chemical agents. This can
lead to a heavy build-up of foam.
Soiling containing high protein levels
such as blood can cause a heavy
build-up of foam when processed with
alkaline chemical agents.
Non-precious metals such aluminium,
magnesium and zinc can release
hydrogen when processed with very
acidic or alkaline chemical agents
(oxyhydrogen reaction).
Where available use the "OEL“
–
programme.
This special programme dispenses
–
emulsifiers (pH neutral) in the
pre-rinse.
Depending on application use
–
de-foaming agents that do not
contain silicone oils.
Select a cleaning programme with at
–
least one short pre-rinse in cold
water.
– Please observe chemical agent
manufacturer's instructions.
37
Dispensing liquid chemical agents
Only use agents formulated
,
specifically for use in
washer-disinfectors and make sure
you follow the manufacturer’s
instructions.
This machine is fitted with 2 dispensers
as standard:
Dispensing system DOS 1 (blue) to
–
dispense liquid cleaning agent. It
can dispense up to 120 ml/min.
– Dispensing system DOS 3 (red) to
dispense acid media such as
neutralising agent or rinsing agent. It
can dispense up to 20 ml/min.
Additional optional DOS modules:
– Dispensing system DOS 2 (white) to
dispense acid media such as
neutralising agent / conditioning
agent. It can dispense up to
20 ml/min.
–
Dispensing system DOS 4 (green) for
low-foaming disinfecting agents
suitable for machine use, or for an
additional cleaning agent. It can
dispense up to 120 ml/min.
The correct amount of liquid agent etc.
required for the application chosen will
then be dispensed through these
dispensing systems.
Further information regarding the
,
OXIVARIO and ORTHOVARIO
programmes and connection of the
H
solution container can be
2O2
found at the end of this booklet
under "Special processes OXIVARIO and ORTHOVARIO".
OXIVARIO kit:
This washer-disinfector can be set up
or retro-fitted to use the OXIVARIO
process by adding an additional
dispensing pump and a special buffer
tank for hydrogen peroxide (H
2O2
solution).
The H
solution hose is colour coded
2O2
black.
The DOS 2 dispenser is used for the
H
solution.
2O2
ORTHOVARIO kit:
This washer-disinfector can be
retro-fitted to use the ORTHOVARIO
process by adding a special
dispensing pump to the DOS 4
dispenser in the OXIVARIO system.
–
Dispensing system DOS 4 (green) to
dispense a special tenside cleaning
agent. It can dispense up to
105 ml/min.
If the ORTHOVARIO process is used
chemical disinfecting agents cannot
be used in the DOS 4 dispenser.
38
Dispensing liquid chemical agents
Preparing the DOS dispensers
The agents can be filled into 5 l plastic
containers, which are colour coded for
the relevant dispensing system.
When first commissioning, or when the
message FILL DOS 1 CONTAINER
and/or FILL DOS 2, 3 or 4 CONTAINER
flashes in the display after switching on
or at the end of a programme, replace
the storage containers with new ones,
or refill them with the relevant agent.
If the message CHECK DISPENSING
SYSTEM 1 and/or CHECK DISPENSING
SYSTEM 2, 3 or 4 flashes in the display,
check the storage containers,
dispensing tubes and hoses. Then
replace the storage containers with new
ones, or refill them with the relevant
agent.
The programme will stop automatically.
When first commissioning or when
the message CHECK DISPENSING
SYSTEM... appears, start the
DOS-FILL Programme to vent the
dispensing system (see "Venting the
dispensing system").
Fill the containers with the relevant
agent
Switch the machine off using the
^
master switch.
^ Lift the service panel up and then
drop it forwards and release it from
its lower brackets.
^ Pull open the drawer to the right of
the service panel to access the liquid
supply containers.
^
Take the storage containers out,
open them and fill with the agent
required, (observe the colour code).
,
Take great care when handling
liquid agents and additives. These
may contain irritant or corrosive
ingredients.
Please follow the manufacturer’s
safety instructions.
Wear protective gloves and
goggles.
^
Set the service panel onto the lower
bracket and then raise it upwards
and pressing against the
washer-disinfector lower it into
position.
39
Dispensing liquid chemical agents
Insert the siphon tube into the
^
container and screw on securely
(observe the colour code).
Once the storage containers have been
filled the relevant message goes out.
Please remember to refill containers
in good time. Do not let them get
empty. It is best to refill them as
soon as the FILL DOS 1, 2, 3 or 4
message appears in the display.
If a container is not being used, the
level query for the unused
dispensing system can be turned off
to avoid the error message (see
"Machine functions, container
query..." in the Programming Manual
supplied with the washer-disinfector).
40
Dispensing liquid chemical agents
Venting the dispensing system
Before the washer-disinfector is first
commissioned, and later on if one or
more containers has been allowed to
run dry, the dispensing system(s) for
liquid media will need to be vented. To
do this:
The DOS-FILL Service Programmes
are allocated to programme places
58-61 ex works, but can be
re-allocated to other programme
places, if preferred.
Dispensing system
maintenance
To ensure trouble-free operation, the
following regular maintenance should
be carried out by a Miele approved
service technician.
Every 12-18 months
Replace the hoses in the dispensing
system(s).
External Dispensing systems
If external pumps are to be used for
dispensing liquid agents, please inform
the Miele Service Department. The
instructions given in the Programming
Manual under "Machine functions" will
need to be observed.
41
Operation
Master switch
The master switch with "emergency
cut-off function" is set to I-ON.
Switching on
Open the stopcocks (if turned off).
^
Press the I-0 button.
^
In operating levels A, B and D the most
recently selected PROGRAMME NAME
appears in the display and in operating
level C AUTOMATIC MOBILE UNIT
RECOGNITION is displayed.
Door lock
The door can only be opened if:
– the electricity supply to the machine
is switched on,
– the red (emergency cut out) master
switch is in the I-ON position,
– the I-0 On-Off button is pressed in,
and
–
a cleaning or disinfection programme
is not in progress.
To open the door
Press the 5 door button, and holding
^
onto the door grip open the door.
To open the door on the Clean side
(through-feed model), the following
conditions must also be met:
DESIN (disinfection) programmes
–
have been correctly completed,
or
the door locking function is active
–
(see "Machine function/Door
Interlock" in the Programming
manual).
Do not touch the heating
,
elements under the wide area filter
during or directly after the end of a
programme. You could burn
yourself. They remain hot for some
time after the programme has
finished.
If the boiler has been programmed
to BOILER READY, be particularly
careful of hot water and steam when
opening the door. Danger of burning
or scalding. The inlet to the wash
cabinet is located, as viewed from
the unclean side, underneath the
basket runner on the left hand side.
42
To close the door
Lift the door upwards and push until it
clicks shut.
Operation
Operating levels
There are four operating levels
available on this machine:
A = Fixed and free access
programmes
B = Free programme selection
C = AWK – Automatic mobile unit
recognition (programme selection
via mobile unit coding) - optional -
D = Programming / free programme
selection / change code
(see Programming Manual)
To change operating level:
^ Press 1 and 2 at the same time.
Operating levelsABCDareshown
in the display.
^ Select the operating level you want
using the 1 or 2 buttons.
^ Press 6 to confirm selection.
^ Enter code when requested by the
display.
The code is set ex-works to ü0000<.
To enter the code:
– Press the start button 6,
[0000] will appear.
– Enter numbers using the 4 and
3 buttons.
– Select number position with the
1 or 2 button.
– To confirm code press 6.
If you enter the wrong code:
FALSE CODE, ENTER AGAIN will
appear in the display.
Close the door.
^
Selecting or changing your own
code
The factory default code can be
changed. See "System functions" in
the Programming Manual.
Code 1 for levels ABC
Code 2 for levels ABCD
43
Operation
Starting a programme
See the Programme charts in the
Programming Manual supplied with
the machine for detailed information
and important notes on the standard
Miele programmes.
When processing medical
,
devices, any changes made to
programmes or dispensing systems
must be documented in a log book
kept with the machine in
accordance with the Medical
Devices directive . The machine's
cleaning and disinfecting standards
must also be re-validated.
A. Fixed programme
Set up fixed programme(s) once in
operating levels B or D and for free
access in operating level A (see
"System functions, Programmes under
A free access" in the Programming
Manual).
^
Select operating level A.
^
Check in the display that the
programme shown is the one
required.
If several fixed programmes have
been made freely accessible, select
the one required using the
programme selector.
^
Press the Start button 6.
B. Free programme selection
Select operating level B.
^
In operating level B there are three
ways of selecting a programme.
1. Programme places1-23canbe
selected using the Programme selector.
Turn the programme selector to the
^
required programme.
The programme name will appear in
the display.
2. Programmes above place 24 are
selected using the 4 and 3 keys.
^ Turn the programme selector to 24.
^ Press 4 (scrolls forwards) until the
required programme is shown.
^ Press 3 (scrolls back) until the
required programme is shown.
3. The PROGRAMME SURVEY menu
lists all stored programmes. A
programme can be selected from this
menu. To do so:
^
Select Programme Survey with 1
and confirm with 6.
^
Select a programme using 1 or 2.
^
Press 6 to confirm selection.
This exits the Programme Survey, and
the selected programme is shown in
the display.
After selecting one of the three options
above:
^
Press the start button 6. The
programme proceeds.
44
For further information on programme
selection see "Operating Level B" in the
Programming Manual.
Operation
C. AWK – Automatic mobile unit
recognition
Select operating level C.
^
Push the coded mobile unit into
^
place.
Make sure that there are no
,
small metallic objects or instrument
parts attracted to the magnetic strip,
in particular to the underside of it.
Any metallic objects on the strip can
result in the coding being incorrectly
read.
^ Close the door.
Before starting a programme it is
,
absolutely essential to check that
the programme shown in the display
is the correct on for the mobile unit
being used.
Otherwise inadequate cleaning or
disinfection could be the result.
Please make sure, that the places
assigned for the programmes using
Automatic mobile unit recognition
are not changed around arbitrarily.
^
Press the Start button 6.
Programme sequence
The programme will start automatically
as soon as the Start button has been
pressed. It is finished when
PROGRAMME-END appears in the
display and the background lighting
flashes (press any button to switch off
the flashing lights).
See "System functions" in the
Programming Manual for further
information on switching off the flashing
signal.
Detailed information on programme
sequences is given in the appendix
of the Programming Manual.
The display background lighting
goes out automatically after approx.
15 mins. To bring it back on again
press one of the buttons.
,
On machines with a built-in
printer you must not change the
colour ribbon cartridge or the paper
roll whilst a programme is in
progress.
45
Operation
Switching off
Press and release the I-0 button.
^
Turn off the stopcocks.
^
Cancelling a programme
A programme can only be cancelled
in operating level B or D.
When testing to EN ISO 15883-1 or
HTM 2030 standards, the
programme must not be interrupted
or cancelled.
In operating levels B or D
Press the Stop button 7.
^
The programme is interrupted.
CANCEL or üCONTINUE< will appear
in the display.
With established
,
decontamination programmes any
contaminated water must be
decontaminated using a disinfecting
agent before it is discharged into the
sewerage system . The door on the
unclean side of the machine can be
opened to do this.
46
^
Use cursor 1 to select üCANCEL<,
the ü < cursors will start flashing.
^
Press the Start button 6. The
programme is cancelled and the
water pumped away.
DRAIN WATER appears in the display.
^
After the water has been pumped
away the required programme can
be selected and started again.
Interrupting a programme
A programme can only be
interrupted in operating level B or D.
If you absolutely have to open the door,
e.g. because the load is obviously
unstable (conscious intervention):
After interrupting a DESIN
disinfection programme, if you wish
to continue it please check the
message in the display at the end of
the programme. If the message
PROCESS PARAMETERS NOT MET
is displayed, it means the door was
opened after the disinfection part of
the programme and the DESIN
parameters were not fulfilled. If
necessary the programme must be
repeated.
In operating levels B or D
^ Press the Stop button 7.
Operation
Washer-disinfectors used as
medical devices (in accordance with
EN ISO 15883) should always be set
to WITHOUT DOOR INTERLOCK
NO.
Caution. Water and items in the
,
machine may be hot. Danger of
burning or scalding.
Where a chemical disinfection
programme has been used, be
aware that steam may contain high
quantites of disinfecting agent.
^ Rearrange the load. Follow infection
control regulations and wear
protective gloves.
^ Close the door.
^ Press the Start button 6.
The programme will continue.
The programme is interrupted.
CANCEL or üCONTINUE< will appear
in the display.
^
Open the door.
On barrier machines please note: If
the machine is set to WITHOUT DOOR
INTERLOCK - YES the doors on either
the clean or unclean side can be
opened. If it is set to WITHOUT DOOR
INTERLOCK - NO only the door on the
unclean side can be opened at this
stage.
47
Data transfer
Each machine is supplied witha5m
long interface cable for data transfer
between the Profitronic system and an
external report printer or PC.
The interface cable is rolled up
inside the machine and must only be
connected by an authorised Service
Dealer or the Miele Service Dept.
The serial interface is RS 232
compatible.
For the interface configurations see
"PC/Printer Function" in the
Programming manual.
Various printers can be used as
external printers:
– Epson-compatible
(contact Miele for a list of suitable
models).
– HP Laserjet.
Pin configuration in the 9-pole sub-D
connector on the interface cable:
5GND (base)
3TXD (transmit)
2RXD (receive)
1-4-6(linked)
7-8(linked)
A standard null-modem or laplink cable
can be connected.
The extension cable to the printer/PC
must not exceed 10 m in length.
Please note the following when
connecting a printer or PC:
Only use an industry-standard PC or
–
printer (e.g. in Germany TÜV, or VDE
approved).
The size of the printer or PC
–
must be taken into account when
installing the machine.
The report printer settings are
described in detail in the Programming
manual under "PC/Printer function".
48
Maintenance
Periodic checks
This washer-disinfector should be
inspected in accordance with local and
national safety regulations after every
1000 operating hours, or every 6
months by a Miele approved service
technician.
This maintenance will cover the
following:
Electrical safety tests
–
Door mechanism and door seal
–
Any screw connections and
–
connectors in the wash cabinet
– Water inlet and drainage
– Internal and external dispensing
systems
– Spray arms
– Filter combination
– Sump including drain pump and
non-return valve
The following operational tests will be
carried out within the framework of the
maintenance:
A programme will be run as a test
–
run
Thermo electrical measurements will
–
be taken
Seals will be tested for water
–
tightness
All relevant measuring systems will
–
be safety tested including error
message displays.
–
Steam condenser
–
All mobile units, modules and inserts
and where applicable:
–
The drying unit
–
Any printer connected to the
machine
49
Maintenance
Process validation
The standard of cleaning and
disinfection in the disinfection
programmes must be confirmed by the
user as a routine matter.
Safety checks must be carried out to
the machine every 3 months by a Miele
approved service technician in
accordance with EN ISO 15883 and
HTM 2030.
Routine checks
Before the start of each working day the
user must carry out a number of routine
checks. A check list is supplied with the
machine for this purpose.
The following need to be inspected:
All filters in the wash cabinet
–
The spray arms in the machine and
–
in any mobile units or modules
The wash cabinet and the door seal
–
Mobile units, modules and inserts
–
50
Maintenance
Cleaning the filters in the wash
cabinet
The filters in the base of the wash
cabinet are designed to prevent coarse
soiling getting into the circulation
system.
A build-up of coarse soiling can cause
the filters to clog up. They should,
therefore, be checked daily and
cleaned if necessary.
This machine must not be used
,
without all the filters in place.
Watch out for glass splinters,
,
needles etc. which could cause
injury.
Cleaning the coarse filter
^ Press the two lugs together, remove
and clean the coarse filter.
^ Put the clean filter back in position
and press until it clicks in place.
51
Maintenance
Cleaning the coarse, flat and
micro-fine filters
Remove the coarse filter.
^
Remove the fine filter (if fitted) from
^
between the coarse and the
micro-fine filters.
^ To unscrew the micro-fine filter, take
hold of the two lugs and turn twice in
an anti-clockwise direction.
Then remove together with the flat
^
filter.
^ Clean the filters.
^ Replace the filters by carrying out the
above steps in the reverse order.
Ensure that the flat filter sits flat in the
base of the wash cabinet.
52
Cleaning the spray arms
Waste particles can lodge in the spray
arm jets.
It is therefore important to inspect the
spray arms regularly.
Use a sharp pointed object to push
^
particles into the spray arm jets, and
rinse well under running water.
Remove the spray arms as follows:
Take any mobile units out of the wash
^
cabinet.
Maintenance
Spray arms in the wash cabinet:
^
Loosen the securing clip on the spray
arm connection and then pull the
spray arm upwards or downwards.
^ Remove the seal and the bearing and
clean them.
^ Refit the bearing and the seal.
Spray arms in mobile units/modules:
^ Turn the nut (left hand thread)
together with the spray arm
clockwise and then pull the spray
arm downwards and off.
^
After cleaning the spray arms screw
them back into position.
^
After replacing the spray arms, check
that they rotate freely.
53
Maintenance
Miele offer a range of branded cleaning
products. These are available from your
Miele retailer, the Spare Parts
Department, or via the internet on
www.miele-shop.com.
Cleaning the control panel
The control panel should only be
^
cleaned using a damp cloth or with a
proprietary cleaning agent for glass
or plastic surfaces.
An approved and listed disinfecting
agent can be used to wipe surfaces.
Do not use abrasive cleaners or
,
all-purpose cleaners.
Because of their chemical
composition they could cause
serious damage to plastic
components.
Cleaning the front of the
machine
Never clean the machine or near
,
vicinity with a water hose or a
pressure washer.
Cleaning the wash cabinet
The wash cabinet is generally
self-cleaning.
However, should a build-up of deposits
occur in the cabinet please contact the
Miele Service Department for advice.
Cleaning the door seals
The door seals should be cleaned
regularly with a damp cloth to remove
any soiling.
Seals which are no longer tight or which
have suffered damage must be
replaced with new ones by the Service
Department.
^ The front should be cleaned using a
damp cloth and a little washing-up
liquid, or with a non-abrasive
proprietary cleaning agent designed
for use on stainless steel.
^
To help prevent fingermarks etc. a
stainless steel conditioning agent
can also be used.
,
Do not use any cleaning agents
containing ammonia or thinners as
these can damage the surface
material.
54
Maintenance
Mobile units, modules and
inserts
Mobile units, modules and inserts
should be checked daily to make sure
they are functioning correctly. The
washer-disinfector is supplied with a
check list.
The following need to be inspected:
Check that rollers on mobile units/
–
inserts are free of hindrance and that
they are secure.
Check that the mobile unit connector
–
is at the correct height and screwed
on correctly.
– Check that the locking caps in the
module connectors of modular
system mobile units are working
properly.
– Check that jets, sleeves and hose
adapters are securely held in
position in mobile units/inserts.
and where applicable:
Make sure that the spray arms rotate
–
freely.
Make sure the spray arm jets are free
–
of any blockages. See "Cleaning the
spray arms".
Make sure that there are no small
–
metallic objects sticking to the
magnetic strip on a mobile unit.
Make sure the magnetic strip on
–
mobile units with automatic mobile
unit recognition is correctly screwed
into position.
–
Check that wash liquor can flow
unhindered through all jets, sleeves
and hose adapters.
–
Make sure caps and closures on
sleeves are correctly located.
55
Maintenance
Printer (optional)
Renewing the paper roll
Renew the paper roll when the red
indicator lights up in the printer.
To do this:
Open up the front flap on the printer
^
and then open it downwards.
Take the empty roll together with its
^
spindle from the roller holder, fit the
new roll on the spindle and put back
in place.
^ Guide the paper upwards onto the
paper feed roller (slit behind the
colour cassette), ensuring that the
green paper feed button is kept
pressed in until the paper emerges
above the cassette.
^ Guide through the slit in the front flap,
then close the front flap.
Replacement paper rolls (58 mm wide,
outer diameter approx. 50 mm) can be
purchased from the Miele Spare Parts
Department or from an office stationery
supplier.
Renewing the colour cassette
Open up the front flap on the printer
^
and then open it downwards.
Pull the colour cassette (above the
^
paper roll) forward and out of its
holder, then put in a new one. Ensure
the paper still passes between the
colour ribbon and the cassette.
Turn the little wheel for manual colour
^
ribbon feed on the right in a
clockwise direction until the ribbon is
taut.
Cut the paper at an angle and guide
^
through the slit in the front flap then
close the front flap.
Replacement colour ribbon cassettes
can be bought from the Miele Spare
Parts dept or from an office stationery
supplier.
56
Repairs should only be carried
,
out by a suitably qualified and
trained Miele technician in
accordance with local and national
safety regulations. Unauthorised or
incorrect repairs could cause
personal injury or damage the
machine.
After sales service
If, having followed the advice in the
operating instruction manual and in the
programming manual, you are still
unable to resolve a problem please call
the Miele Service Department (see
back cover for contact details).
When contacting Miele, please quote
^
the model and serial number of your
machine
To avoid unnecessary service call-outs,
check that the fault has not been
caused by incorrect operation.
An overview of all error messages
that can appear in the display are
given in the Programming manual
under "Messages".
These are located on the data plate
(see "Electrical connection").
57
Converting the type of heating
Converting from steam to electric
heating or from electric to steam
If your machine is convertible, you can
use the STEAMüüELECTRO or
ELECTROüü STEAM service
programme to switch from one type of
heating method to the other.
Select STEAMüüELECTRO or
^
ELECTROüüSTEAM (see Operation /
B. Free programme selection).
Press the Start button 6.
^
When the Service programme has
ended, a message will appear in the
display.
^ To üCONTINUE< press 6.
58
All electrical work must be carried
out by a suitably qualified and
competent person in accordance
with local and national safety
regulations (BS 7671 in the UK).
Connection should be made via a
–
suitable isolator with an on-off switch
which is easily accessible for
servicing after the machine has been
installed.
An electrical safety test must be
–
carried out after installation and after
any servicing work. It is therefore
essential that the on-off switch is
easily accessible.
– When switched off there must be an
all-pole contact gap of at least 3 mm
in the isolator switch.
Electrical connection
The machine must only be operated
with the voltage, frequency and fusing
shown on the data plate.
Further notes on electrical
connection are given on the
Installation diagram supplied with
the machine.
The data plate showing relevant test
marks (VDE, DVGW etc.) is located on
the cover located behind the service
panel on the unclean side of the
machine.
The wiring diagram is supplied with
the machine.
WARNING
– For extra safety it is advisable to
protect the machine with a suitable
residual current device (RCD).
–
Check that the phase rotation is
correct when connecting the
machine. Failure to do so could have
an adverse effect on the circulation
pump and impair the wash quality.
–
Equipotential bonding should be
carried out.
–
For technical data see data plate or
wiring diagram supplied.
THIS APPLIANCE MUST BE
EARTHED
59
Plumbing
This machine should be
,
connected to the mains water
supply and to the drainage system
in accordance with local and
national water regulations. Please
refer to the installation diagram
supplied with the machine.
If the water supply has a high iron
–
content there is a danger of
corrosion occurring on items being
processed in the machine, as well as
in the machine itself.
If the chloride content of the water
exceeds 100 mg/l the risk of
corrosion to items being processed
in the machine will be further
increased.
60
Technical data
Height1536 mm (incl. 100 mm plinth and lid)
1928 mm (incl. 100 mm plinth and MAV)
Width90 cm
Depth
Depth with door open
Weight (incl. plinth, DKWK and
MAV):
Weight in operationApprox. 500 kg
VoltageSee data plate
Connected loadSee data plate
Fuse ratingSee data plate
Compressed air connection600 kPa (6.0 bar) required for steam operation
Steam connection (with electric
TA):
Water pressure (flow rate)100 - 1000 kPa overpressure
Water connection:
Cold, hot and de-ionised (AD)
water
Ambient temperature5 °C to 40 °C
Relative humidity:
maximum
decreasing proportionately to
AltitudeMax. 1500 m
Soiling level
(according to IEC/EN 61010-1)
Protection category (according to
IEC 60529)
Noise level in dB (A),
Sound level LpA during cleaning
and drying phases
Test marksVDE, radio interference suppression
! markMDD-Guidelines 93/42/EWG, Class IIb
Manufacturer's addressMiele & Cie. KG, Carl-Miele-Str. 29,
#
If installed at altitudes above 1500 m the boiling point of the suds solution will be lower.
Disinfecting temperature parameters should be lowered and the holding time increased
(A0value). This must be done by a Miele authorised service technician.
77 cm
133.7 cm
Approx. 350 kg
250 - 1000 kPa / 140 - 180 °C
up to max. 70 °C
80 % for temperatures to 31 °C
50 % for temperatures to 40 °C
#
P2
IP20 (Dust permeation)
<70
33332 Gütersloh, Germany
61
Optional extras
Optional extras
Drying unit (TA)
–
Steam condenser (DK)
–
Water softener
–
Report printer (PRT)
–
Dispensing systems DOS 2 and
–
DOS 4
OXIVARIO kit
–
ORTHOVARIO kit (this has to be set
–
up by an approved Miele service
technician)
62
Special processes - OXIVARIO and ORTHOVARIO
Description of the machine
This washer disinfector can be set up
or retro-fitted to use the OXIVARIO
process by adding two additional
dispenser pumps together with a buffer
tank for hydrogen peroxide solution
H
. The DOS 2 dispenser is used for
2O2
the H
To use the ORTHOVARIO process the
DOS 4 dispensing system has to be
replaced by a dispenser pump for a
tenside based cleaning agent.
The OXIVARIO, OXIVARIO PLUS and
the ORTHOVARIO programmes are
available for these requirements. The
machine is supplied with these
programmes allocated to programme
places 51 to 53.
2O2
solution.
The OXIVARIO process releases active
oxygen under alkaline conditions. The
cleaning agent used must be tenside
free and have a pH value of between
11 and 11.5.
The ORTHOVARIO process uses a
special tenside based cleaning agent
in the first cleaning phase and then
releases active oxygen in the second
phase with a pH value of between 10.0
and 11.0 and a slightly higher
temperature of 65 °C.
63
Special processes - OXIVARIO and ORTHOVARIO
Areas of application
The OXIVARIO procedure has an
alkaline main wash making it
particularly suitable for the processing
of surgical instruments where existing
procedures are not satisfactory.
It is particularly suitable for instruments
such as those used in high frequency
surgery, orthopaedic surgery as well as
for instruments which have dried out
because of the length of time between
using them and reprocessing them. It is
also suitable for antiseptic
circumstances.
The cleaning process is gentle enough
for instruments used in minimally
invasive surgery, including fibre optics,
where the manufacturer has declared
that they are suitable for processing
with an alkaline cleaning agent.
It is not suitable for anodised
aluminium.
Be aware that items made from alloys
containing titanium such as some
implants, do not always state the
material's compatibility. Processing
such items can result in colour
changes, rendering any coding on
them ineffective. If in doubt please
consult the manufacturer of the item.
The ORTHOVARIO procedure is
suitable for processing aluminium
instruments which are sensitive to
alkalines. It is particularly suitable for
orthopaedic instruments including
motor systems.
Because of its oxidising effect it
must not be used for titanium
implants, and in particular colour
colded ones.
Due to the high efficacy of the
,
Oxivario and Orthovario processes it
is important to lubricate joints and
contact parts on instruments pre
sterilisation. Please refer to the
instrument manufacturer for
guidance.
64
Special processes - OXIVARIO and ORTHOVARIO
Warning and Safety
instructions
These warning and safety
,
instructions are in addition to those
given at the beginning of this
booklet.
H2O2solution must only be used in
~
the special containers and adapters
provided by the Ecolab and Dr. Weigert
companies.
Important: observe the
~
manufacturer's safety instructions as
given on their safety data sheet.
Be especially careful when handling
~
H
solution. It is an irritant chemical.
2O2
Observe all current safety instructions
pertaining to its use.
Wear protective gloves and goggles.
Empty containers must be disposed
~
of in accordance with the
manufacturer's instructions.
H2O2solution must not be mixed
~
with other chemicals. This could cause
a serious chemical reaction, e. g. the
release of harmful vapours or gases.
Only use tenside cleaning agents
~
supplied by Ecolab or Dr. Weigert.
Connecting the H2O2solution
container
The connection hose for the H2O
solution container is identified by a
black label. The connection hose is
supplied without a safety cap as the
container for hydrogen peroxide will
vary depending on supplier.
Connect the supplier's safety cap to
^
the connection hose (black).
Then connect this to the H
^
solution container.
^ Then start the
programme.
Unlike with other chemical agents the
H
solution container must be fully
2O2
emptied before it is exchanged for a
new one.
Do not replace the H
container with a new one until the
DOS 2 CONTAINER
displayed. After connecting the new
container you must run the
programme.
If the
CHECK DISPENSING SYSTEM 2
message appears you should check
the container as well as the dispensing
system. The programme will stop
automatically.
DOS2-FILL
solution
2O2
message is
2
2O2
FILL
DOS2-FILL
65
Disposal of your old machine
Electrical and electronic machines
often contain materials which, if
handled or disposed of incorrectly,
could be potentially hazardous to
human health and to the environment.
They are, however, essential for the
correct functioning of your or machine.
Please do not therefore dispose of it
with your general waste.
Please dispose of it at your local
community waste collection / recycling
centre or contact your dealer for
advice. Ensure that it presents no
danger to children while being stored
for disposal.
66
67
Alteration rights reserved / Publication date 11.01.2010
0366
M.-Nr. 07 776 620 / 00