Teachware
Metrohm AG
CH-9101 Herisau
teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been
checked with great care, errors cannot be entirely excluded. Should you
notice any mistakes please send us your comments using the address
given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documenta-
tion.
Table of content
page
1 Overview 1
1.1 Application range __________________________________________1
The Metrohm 711 Liquino is a very versatile laboratory instrument which can
be used everywhere that liquid handling is to be simplified and automated.
This is the case in both analytical laboratories and in synthesis laboratories.
It is equally suitable for both the preparation of complicated formulations in
filling plants and also for simple delivery and pumping tasks involving toxic
or aggressive media.
Owing to the freely programmable process sequences and versatile communication interfaces (RS232 or remote connection) it can be integrated into complex automated systems to undertake demanding dosing and monitoring functions, which otherwise would only be capable of realization with a
considerably greater expenditure.
1.2 Application possibilities
The 711 Liquino can be used for the whole range of applications which fall
under the general heading of 'Liquid Handling'.
In analytical laboratories
• manual, highly accurate liquid dosing.
• complex, process-controlled dosing tasks with freely selectable specifi-
cations, e.g. a certain volume in a given period, constant dosing speed
during a preselected time interval, a certain volume with constant dosing
speed. In addition this can also be temperature-controlled, i.e. if a certain temperature limit is exceeded or undercut during the dosing process
then the dosing process can be interrupted and a predefined action triggered, e.g. switching on a cooling bath.
• automated preparation of defined solutions (e.g. standard solutions) or
mixtures of liquids (e.g. mobile phases for HPLC or IC). It is possible to
connect a balance without any problems.
• highly accurate pipetting and diluting even small volumes of liquid.
• automated sampling and sample preparation for a range of analytical
methods (e.g. titrations, pH measurements, etc.) with and without a
sample changer.
• preparation of complex formulations in automated filling plant.
In synthesis laboratories
• manual, highly accurate liquid dosing.
• complex, process-controlled dosing tasks with freely selectable specifi-
cations, e.g. a certain volume in a given period, constant dosing speed
during a preselected time interval, a certain volume with constant dosing
speed. In addition this can also be temperature-controlled, i.e. if a certain temperature limit is exceeded or undercut during the dosing process
then the dosing process can be interrupted and a predefined action triggered, e.g. switching on a cooling bath.
711 Liquino, Instructions for use
1.3 Instrument description
2
Slot for memory
Always insert the card so that
1.3 Instrument description
1.3.1 Front view
card
the upper edge of the card
with the label faces right.
711 Liquino, Instructions for use
1. Overview
3
- Fire Hazard For continued pr ot ection replace
only with the same type and
rating of fuse
Fuse:
100/115V : 315mAT
230/240V : 160mAT
Keyboard/Display
Rem ot e
RS 232
Pt 100
Dos.4
Dos.3
Dos.2
Dos.1
Made by Metrohm
Herisau Switzerland
WARNING !
Connections for
Connection for
Keyboard connection
Remote
Serial RS 232
Fuses, mains switch
Name plate
1.3.2 Rear view
4 Dosinos
Pt100 temperature sensor
interface
interface
and mains connection
711 Liquino, Instructions for use
2.1 Instrument setup
4
Follow the ins
supply. If the instrument is operated with the mains voltage incorrectly set
and/or the wrong mains fuse there may be a fire risk!
2 Installation
2.1 Instrument setup
Packaging
The 711 Liquino is supplied in packaging which provides excellent protection. This includes a shock-absorbing plastic foam lining. Please store this
packaging as it is the only way to guarantee damage-free transport of the
instrument.
Checks
Please check immediately on receipt whether the shipment is complete and
undamaged.
Location
The 711 Liquino is a robust instrument and can therefore be used even under rough conditions in laboratories and factories.
However, care must be taken that it is not exposed to a corrosive atmosphere. Regular care of the instrument is essential if it is used in rough surroundings.
2.1.1 Mains connection
Setting the mains supply
Before switching on the 711 Liquino for the first time check that the mains
voltage set on the instrument (see illustration on following page) corresponds to the local mains voltage. If this is not the case then the correct
mains voltage must be set as follows:
•Pull out mains cable
Remove the mains cable from the mains connector of the 711 Liquino.
•Remove fuse holder
Use a screwdriver to loosen the fuse holder beside the mains connector
and remove completely.
•Check and replace fuses
Remove the built-in fuses for the required mains voltage from the fuse
holder and check their specifications:
100…120 V 0.315 A (slow-blow) Metrohm no. U.600.0011
220…240 V 0.16 A (slow-blow) Metrohm no. U.600.0008
tructions given below to connect the instrument to the mains
•Insert fuses
Exchange fuses if necessary and replace them in the fuse holder.
•Set mains voltage
Remove the white fuse block from the fuse holder and reinsert it so that
the correct mains voltage is visible through the window in the fuse holder.
711 Liquino, Instructions for use
2. Installation
5
230
Mains switch
Mains connection
socket
Fuse holder with
window
•Insert fuse holder
Insert fuse holder and press it until it clicks into place.
2.2 Safety information
General
This instrument left our factory in perfect condition from a safety point of
view. The following information must be carefully observed in order to maintain this condition and to ensure hazard-free operation.
Mains connection
The mains cables supplied with the instrument are three-core and are fitted
with an earthing pin. If a different plug must be used then the yellow/green
lead must be connected to the protective earth.
WARNING!
Any break in the earth conductor inside or outside the instrument can represent a source of danger to the operator. Making a break in the earth conductor is forbidden.
Before connecting the plug to the mains supply a check must be made that
the instrument is set to the local mains voltage and that the appropriate fuses are inserted.
WARNING!
When changing a fuse and when setting a different mains voltage the mains
cable must be disconnected from the mains supply.
When replacing fuses care must be taken that voltage, current and type
have been selected correctly (see Technical data, fuses, page 162).
WARNING!
Short-circuiting fuses or the temporary use of incorrect fuses is forbidden.
Repair and maintenance:
If faults or malfunctions occur during the operation of the 711 Liquino it is
recommended that the cable connections with the peripheral units are
checked first for proper functioning (see page 9ff). It may be possible to lo-
711 Liquino, Instructions for use
2.2 Safety information
6
out.
cate the source of the fault with the aid of the diagnosis functions. If the
cause of the malfunction cannot be found please contact the Metrohm service department.
The instrument should only be opened when switched on under exceptional
circumstances. As this exposes electrically live components it should only
be carried out by an expert who is familiar with the risks involved.
The instrument must be separated from all electricity sources before it is opened. Make sure that the mains plug has been pulled
Condensers in the instrument may still be charged even when the instrument is no longer connected to the mains supply.
Changing the battery
The connecting wires of the old battery must be cut before desoldering so
that any residual charge which may still be present cannot be shortcircuited. The polarity of the new battery must be taken into account when it
is fitted.
Lithium batteries cannot be recharged and should not be short-circuited or
exposed to temperatures above 70 °C under any circumstances.
Old batteries must be disposed of in an environmentally acceptable way
and must not be burnt.
Static electricity:
The circuit contains components which could be destroyed by electrostatic
charges or whose functions could be affected by them. Work on an open
instrument should only be carried out in ESD-protected surroundings.
Suitable measures:
1. A conductive mat on the work surface connected to the earth conductor
via a resistor of approx. 1 MΩ.
2. Earthed soldering iron.
3. Persons working with components must have their wrists connected to
the earth conductor via a resistor of approx. 1 MΩ.
4. Components and assembled PCBs should only be stored and transported in conductive packaging or foam.
5. New components should be stored in their original packaging until re-
quired.
If it becomes apparent that the instrument can no longer be operated safely
then it should not be used at all.
711 Liquino, Instructions for use
2. Installation
7
The keyboard is connected to the keyboard
Dos.1
The notch on the plug must coincide with the
marking on the panel.
2.3 Setup
2.3.1 Connecting the keyboard
socket provided for it on the rear panel of the
instrument.
2.3.2 Connecting the Dosinos
The Dosinos are connected to the sockets ”DOS1” to ”DOS4” on the rear
panel of the 711 Liquino. Press each Dosino plug into the socket until it
clicks into place!
A Dosino should only be connected when the instrument is switched off!
The 711 Liquino only recognizes Dosinos which are already plugged in
when it is switched on.
2.3.3 Connecting a Pt100 temperature sensor
711 Liquino, Instructions for use
The Pt100 temperature sensor 6.1111.030 with 4-wire technology can be
used with the 711 Liquino. It is connected to the socket marked with ”Pt100”
on the rear panel of the instrument.
2.4 Memory card
8
Pipet tip
Green pilot lamp
6.1562.040
Black start
button
6.1562.050
Pipetting tubing
Connection to dosing port 1
of the Dosino 700
Connection cable to 711 Liquino (remote
connection)
Screw nipple
2.3.4 Installing the pipetting attachment
The pipetting attachment available as an option is suitable for automatic pipetting functions with a Dosino 700.
Pipet handle
To install the pipetting attachment proceed as follows:
• Loosen the screw nipple on the pipet handle and push it over the pipet
tubing.
• Guide the pipet tubing into the handle and fix the tubing with the screw
nipple.
• Mount the pipet tubing on dosing port 1 of the Dosino dosing unit.
• Insert in the 25-pin connection cable plug into the remote socket of the
Liquino. The second connection cable plug is used to connect another
peripheral unit to the Liquino.
• As soon as the pipet handle is connected the green pilot lamp lights up
to indicate that it is ready for work.
• The angle between the tubing guide and the handle can be adjusted for
strain-free work.
Important: the pipet tubing must always be clean and kink-free!
Pipetting accessories:
Pipet handle 6.1562.040
with spiral pipet tubing, 3 mL 6.1562.050
Pipet tubing, 10 mL 6.1562.100
Pipet tubing, 3 mL 6.1829.020
Pipet tubing, 0.7 mL 6.1562.010
Pipet tips for pipet tubing 0.7 mL, 10 pieces 6.1562.030
Pipet tip holder 6.2052.010
2.4 Memory card
The memory card (e.g. 6.2245.010) must be inserted in the slot on the front
panel of the 711 Liquino so that the front of the memory card (with the type
name) faces to the right. A little pressure is necessary to push the memory
card completely into the slot. Never remove the memory card from the instrument when it is being accessed by the instrument.
711 Liquino, Instructions for use
9
RS 232
Remote
Dosimats
pH Meter /
Conductometer
Titrino family
726 Titroprocessor
730 Sample changer
665
..
691
692
712
713
702
751
Seiko
Citizen
Epson
HP
IBM
other devices
e. g. thermostat
730
711 Liquino
726
700 Dosinos
Pt100
Printer
Balance
pumps
heating bath
etc.
AC
DC
731 Relay box
other
devices
Dos.1-4
772 Pump unit
683 Pump
711 Liquino
pumps
heating bath
etc.
...
776
Personal computer
2. Installation
2.5 Connections
Metrohm automated systems with the 711 Liquino
711 Liquino, Instructions for use
2.5 Connections
10
Input 0
21
START
Input 1
9
STOP
Input 2
22
HOLD / CONTINUE
Input 3
10
External ready
Input 4
23
Sample Changer ready
Input 5
11
not used
Input 6
24
PIP continue
Input 7
12
not used
Output 0
5
Ready
Output 1
18
Upper limit
Output 2
4
Lower limit
Output 3
17
EOD (End of Determination)
Output 4
3
Activate 1
Output 5
16
Activate 2
Output 6
1
External start
Output 7
2
Start Sample Changer
2.5.1 Remote connections
The remote control of instruments within a Metrohm automated system can
(apart from data communication via RS232 interfaces) be carried out in a
simple manner with the help of parallel-switched signal leads, the remote
lines (or I/O lines). Signals can be set statically or transmitted as a signal
pulse (usually approx. 200 ms duration); see SCAN and CONTROL commands, (page 91ff). The signal level is +5 volt (TTL-level) in each case.
The functions which can be used for remote control or monitoring tasks
each occupy a particular remote lead. The remote socket of a Metrohm instrument has input leads (Input0…7, for the control of certain functions, e.g.
START or STOP) and output leads (Output0…8, to display instrument conditions, e.g. 'ready' or 'End of Determination').
The remote connection is normally a 25-pin socket. Different connection cables are available for different types of instrument. The suitable remote cable
is listed in the ‘Instructions for use’ of the corresponding instrument. Special
connection cables can also be made according to the customers’ particular
requirements.
Remote lead occupancy for 711 Liquino:
Remote lead Pin Function
Output leads 8…13 cannot be used.
In the SEQ mode remote leads Input3…7 and Output4…7 can be freely
used with the commands SCAN and CONTROL (see page 91ff).
711 Liquino, Instructions for use
2. Installation
11
Note:
B
A
Titrino
C
D
230
cable 6.2141.090
Connection cables
Only Metrohm cables should be used to connect the 711 Liquino to other
instruments as these are the only ones with which interference-free data
transmission can be guaranteed.
Remote cables for the 711 Liquino always have a name on the end of the
cable to indicate the instrument with which the particular plug is intended to
be used and which connection it should be plugged in to.
Example:
Titrino B
692 / 712 / 71 3
Liquino — Titrino
Cable 6.2141.090 is used in the SMPL mode. Details can be found on page
106ff.
Control commands can also be used in an SEQ method, for example:
>SEQ sequence st ep 4 ( CONTROL command)
pat t er n * 1* * ----
starts Titrino (start pulse)
>SEQ sequence st ep 5 ( Pause command)
pause 00: 00: 01
>SEQ sequence st ep 6 ( CONTROL command)
pat t ern *0**----
wait 1 s
reset remote interface
(end pulse)
>SEQ sequence st ep 7 ( SCAN command)
pattern ****1---
wait for ready message from Titrino
711 Liquino, Instructions for use
2.5 Connections
12
711
B
A
Titrino
C
D
230
683 Pump Unit
I
0
683 Pump Unit
I
0
730
cable 6.2141.100
Liquino — Titrino — 683 Pump — 730 Sample Changer
Cable 6.2141.100 is used for the SMPL mode if aspiration and rinsing the titration vessel are to be carried out as quickly as possible (with 2 x Pump
683). The sample changer connection can be left open if no sample chang-
er is required.
The Titrino, the pumps and the sample changer can also be addressed with
the control commands of a SEQ method, for example:
Start Titrino
>SEQ sequence st ep 4 ( CONTROL command)
pat t er n * 1* * ---- starts Titrino (starts pulse)
>SEQ sequence st ep 5 ( Pause command)
pause 00: 00: 01
>SEQ sequence st ep 6 ( CONTROL command)
pat t er n * 0* * ----
wait 1 s
starts Titrino (ends pulse)
>SEQ sequence st ep 7 ( SCAN command)
pattern ****1---
wait for ready message from Titrino
Run pump 1 for 10 s
>SEQ sequence st ep 8 ( CONTROL command)
pattern ***1---- switches on 1st 683 pump
>SEQ sequence st ep 9 ( Pause command)
pause 00: 00: 10
wait 10 s
>SEQ sequence st ep 10 ( CONTROL command)
pattern ***0----
switches off 1st pump 683 again
Run pump 2 for 1 min 5 s
>SEQ sequence st ep 11 ( CONTROL command)
pat t er n * * 1* ---- switches on 2nd pump 683
>SEQ sequence st ep 12 ( Pause command)
pause 00: 01: 05
wait 1 min 5 s
>SEQ sequence st ep 13 ( CONTROL command)
pat t er n * * 0* ----
switches off 2nd pump 683 again
711 Liquino, Instructions for use
2. Installation
13
711
B
A
Titr
ino
C
D
7
30
cable 6.9921.165
Command to
730 Sample Changer
SCN:Rm: ready1 ←
CTL:Rm: START device1→
Switch sample changer to next sample
>SEQ sequence st ep 14 ( CONTROL command)
pat t er n 1* * * ---- switches sample changer 1 sample further
*)
(start pulse)
>SEQ sequence st ep 15 ( Pause command)
pause 00: 00: 01
>SEQ sequence st ep 16 ( CONTROL command)
pat t er n 0* * * ----
wait 1 s
resets remote interface
(end pulse)
>SEQ sequence st ep 17 ( SCAN command)
pattern ***1*---
wait for ready message from sample
changer
711 Liquino – 730 Sample Changer – Titrino
730 Sample Changer with Titrino and 711 Liquino for automated pipetting of
different samples on the sample changer into a special beaker on the sample rack.
The 711 Liquino can be used in any automated system for pipetting tasks
(PIP mode). In this case the 730 Sample Changer controls the working process. The 711 Liquino is used in the PIP mode. The determination of the
samples is carried out by the Titrino which, like the Liquino, is started from
the sample changer.
This arrangement of instruments requires a special remote cable
(6.9921.165). Other special cables for similar instrument combinations are
available on request.
*)
For this function the sample changer must be appropriately programmed with Scan and Control com-
mands!
711 Liquino, Instructions for use
2.5 Connections
14
>st ar t sequence
3 Lift: 1: shift mm
1 Lift: 1: shift mm
711
731
772
cable
6.2125.100
Dosinos
heating bath
mains cable
Example of a sample changer method:
1 CTL: Rm: INIT
>sampl e sequence
1 CTL: Rm: START devi ce1
2 SCN: Rm: r eady1
4 CTL: Rm: *********1****
5 Wai t 1 s
6 CTL: Rm: *********0****
7 Mov e: 1: sampl e
8 Li f t : 1: wor k mm
9 CTL: Rm: *********1****
10 Wai t 1 s
11 CTL: Rm: *********0****
12 SCN: Rm: r eady1
13 Mov e: 1: spec. 1
14 Li f t : 1: wor k mm
15 CTL: Rm: *********1****
16 Wai t 1 s
17 CTL: Rm: *********0****
18 SCN: Rm: end1
19 St i r : 1: ON s
20 CTL: Rm: START devi ce2
2 1 SCN: Rm: e nd2
22 St i r : 1: OFF s
23 Pump: 1. 1: 12 s
24 Pump: 1. * : 10 s
25 Pump: 1. 1: 12 s
28 NOP
>f i nal sequence
2 NOP
← Initialize remote interface
← Start Liquino (PIP mode)
← Wait for pipetting step
← Lift out of sample vessel
← Switch pipetting further (corresponds to
<ENTER>)
— separating bubble is aspirated —
← Reset remote interface
← Position sample beaker
← Immerse pipet tubing in sample
← Switch pipetting further (corresponds to
<ENTER>)
← Reset remote interface
← Wait for next pipetting step
← Move to titrating beaker
← Pipet tubing in titrating beaker
← Switch pipetting further (corresponds to
<ENTER>)
← Reset remote interface
← Wait for end of PIP mode
← Stirrer on
← Start Titrino
← Wait for end of determination
← Switch off stirrer
← Aspirate titration beaker
← Rinse titration beaker
← Aspirate titration beaker
Liquino with 772 Pump Unit as delivery or aspiration pump and a heating
bath.
With the aid of a Metrohm 731 Relay Box the output leads of the 711 Liquino
can be monitored and, according to their line conditions, two low-voltage
d.c. voltage outputs and two mains voltage outputs can be switched.
In this way the heating bath can be switched automatically via the temperature monitoring function. If the measured temperature falls below a certain
value the bath is heated up.
A connected pump (e.g. the Metrohm 772 Pump Unit) can be selectively
switched on or off during the method procedure.
711 Liquino, Instructions for use
2. Installation
15
731 Relay Box settings:
Connect 772 Pump Unit to DC1 (or DC2)
Output voltage: 18 or 24 volt
Remote address selection disk DC1: 7 (=Output 7)
Connect heating bath to AC1 (or AC2,
mains voltage)
Remote address selection disk AC1: 2 (=Output 2)
711 Liquino settings
Temperature monitoring
XDOS mode (contained in SEQ mode):
Par amet er s
>XDOS moni t or i ng
t emperat ur e: on
low limit xx°C
upper limit ...
action: ...
al ar m pi ns: under
Control commands for the pump
SEQ mode
Switch on pump:
command: CONTROL
i nt er f ace: Remot e
pat t er n 1* * * ---- (Output 7)
(switch off with
0***----)
711 Liquino, Instructions for use
2.5 Connections
16
cable 6.2134.040 (9-pin/9-pin)
or 6.2134.050 (9
printer cables: see page 17f
711
2.5.2 Serial connections
A wide range of possible connections can be made at the RS232 interface.
As well as all Metrohm instruments which are equipped with the Metrohm
remote control language, a printer (requirement: serial interface or parallel/serial converter) or a personal computer (PC) can also be connected.
Any instrument from another manufacturer which is equipped with a serial
RS232 interface can also be connected.
-pin/25-pin)
A requirement for correct data transmission is that the transmission parameters are correctly set; these must coincide with the interface settings of the
connected instrument (see following page).
Control commands (examples):
>SEQ sequence st ep 4 (Control command)
st r i ng &M; $G
starts Metrohm instrument
>SEQ sequence st ep 6 (Control command)
st r i ng &M; $S
stops Metrohm instrument
Input data query (example):
>SEQ sequence st ep 5 (Scan command)
st r i ng * R"
wait for ready message from
Metrohm instrument
The following section provides information about the settings and cables
necessary to connect a printer.
711 Liquino, Instructions for use
2. Installation
17
The Liquino must be switched off before a printer is connected to the RS232
interface!
Printer
Cable
RS232 settings
Settings on printer
IBM Proprinter
6.2125.050
+
6.2125.010
baud rate: 9600
data bit: 8
stop bit: 1
Parity: none
handshake: Hardware
send to: IBM
see printer manual
Seiko
DPU-411
6.2125.020
+
6.2125.010
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Seiko
DIP switch settings:
DIP01 DIP02
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6
Depending on the set dialog language, the
adjustable 7-bit ASCII character set of the
printer will be set automatically to the national character set.
2.5.3 Connecting a printer
Printers with the following printer drivers can be connected:
IBM IBM Proprinter and printers with IBM emulation
Epson Epson printers and printers with Epson emulation
Seiko Seiko printer DPU-411 or DPU-414
Citizen Citizen printer IDP560 RS
HP HP printers and printers with HP PCL3 emulation
If you wish to connect a different printer then ensure that this can emulate
one of the printer modes supported by the 711 Liquino.
Most printers with a serial interface are connected with cable 6.2134.050.
Printers with a parallel interface require a serial/parallel converter (e.g.
2.145.0300) and cable 6.2125.020 + adapter cable 6.2125.010.
The interface parameters are entered in the configuration menu under
‘>RS232 settings’. The printer type must also be entered in the configura-
tion menu under ‘>Peripheral units’ and ‘send to:’
If you use an A4 printer and the graphics do not cover the whole width of the
page then you can proceed as follows:
When the instrument is switched on keep the <CONFIG> key pressed
down. In this way you access the setup menu. Press the <> key and
<ENTER>. You can now alter the relative width and height of the graphics.
In each case press <ENTER> after the entry.
The following table provides information about the connection of several selected printers.
711 Liquino, Instructions for use
2.5 Connections
18
Printer
Cable
RS232 settings
Settings on printer
Seiko
DPU-414
6.2125.130
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Seiko
Recommended DIP switch settings:
Dip SW-1 Dip SW-2 Dip SW-3
1 OFF ON ON
2 ON OFF ON
3 ON ON ON
4 OFF ON ON
5 ON ON OFF
6 OFF ON ON
7 ON OFF ON
8 ON OFF ON
Depending on the set dialog language, the
adjustable 7-bit ASCII character set of the
printer will be set automatically to the national character set.
Citizen
IDP562-RS
6.2125.050
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Citizen
DIP switch settings:
1 2 3 4 5 6 7 8 9
10
ON
SSW1
The adjustable 7-bit ASCII character set of
the printer can only be set to the national
character set by switching DIP switches 4
and 5 in the printer:
4 5 Character set
OFF OFF USA
ON ON United Kingdom
ON OFF France
OFF ON Germany
No special Spanish character set is available (it is best to select French).
Epson
with 6-pin
round plug
6.2125.040
+
6.2125.010
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Epson
DIP switch settings:
SW1 SW2
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6 7 8
Epson
with additional
serial interface
#8148
6.2134.050
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Epson
DIP switch settings at interface:
SW1 SW2
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6
Epson LX-300
6.2134.050
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: Epson
see printer manual
HP Deskjet
with serial interface
6.2134.050
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: HP
DIP switch settings:
A B
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6 7 8
HP Laserjet
with serial interface
6.2134.050
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: HP
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hardware
send to: HP
see printer manual *
* The printer must be set to the PC-8 character set. PCL3 command:
esc(10U
711 Liquino, Instructions for use
2. Installation
19
Balance
Cable
Sartorius MP8, MC1
6.2134.060
Mettler AB, AG (LC-RS25)
supplied with balance
Mettler AM, PM
from Mettler: ME 33995: green wire to Pin 2, brown wire to
Pin 3, white wire to Pin 7, yellow wire to Pin 20 of the 25-pin
plug
+ 6.2125.010 25-pin/9-pin adapter
Mettler interface 016
cable supplied with interface 016: red wire to Pin 3, white
wire to Pin 7 of the 25-pin plug
+ 6.2125.010 25-pin/9-pin adapter
Mettler interface 011 or 012
6.2125.020 + 6.2125.010
Mettler AT
from Mettler: ME 33995: green wire to Pin 2, brown wire to
Pin 3, white wire to Pin 7, yellow wire to Pin 20 of the 25-pin
plug
+ 6.2125.010 25-pin/9-pin adapter
AND models ER-60, 120, 180, 182
models FR-200, 300
models FX-200, 300, 320
with RS232 interface (OP-03)
6.2125.020 + 6.2125.010
Precisa, balances with RS232C interface
6.2125.080 + 6.2125.010
711
6.2125.010
printer cable
balance cable
6.2125.030
2.5.4 Connecting a balance
The following balances can be connected to the RS232 output of the
711 Liquino:
The balance type must be preselected on the 711 Liquino with the
<CONFIG> key under ‘>Peripheral units’.
The weight will be transmitted as a number with up to 6 places, together
with sign and decimal point. Units and control characters sent by the balance will not be transmitted.
For the simultaneous connection of printer and balance the adapter cable
6.2125.010 + with junction plug 6.2125.030 are required.
If the balance only works with 7 bit and the printer only works with 8 bit then
the balance parity must be set to "space" while the 711 Liquino and printer
run with 8 bit, "no parity".
711 Liquino, Instructions for use
3.1 The Liquino concept
20
3 Introduction
3.1 The Liquino concept
The 711 Liquino is designed so that, together with one or more 700
Dosinos, it can automate a wide range of tasks in the laboratory. Everything
which falls under the general heading of 'liquid handling', i.e. the transport of
liquids in any form, belongs to its range of applications. Its functions can be
divided up into three steps.
• Simple manual dosing and dispensing with up to four dosing devices.
• Automated procedures for frequently reoccurring and defined tasks. The
preset modes in the instrument are used for this. When they are appropriately set they can be stored as methods in the Liquino itself or on a
memory card and can be reloaded at any time.
• Complex process sequences, which can be made up from several indi-
vidual methods or modes with additional auxiliary commands as required. In this way a large number of instruments can be controlled for
use in demanding analytical or synthesis units.
The practical 700 Dosinos are used as dosing instruments; they can be
conveniently operated with the functions and commands of the 711 Liquino.
Special functions (PREP, EMPTY) are available for preparatory work (i.e.
bubble-free filling and rinsing of the tubing) and emptying dosing units
when a reagent needs to be changed. In order to carry out these procedures in as comfortable and reagent-saving manner as possible you can
define different types of dosing unit. As the length and diameter of the tubing used can be defined the 711 Liquino is able to calculate and then use
the optimal volumes for the above-mentioned functions.
With the help of the special GLP mode the 711 Liquino can be used to validate all your dosing units, i.e. to determine their accuracy. In this way GLPconform work can be ensured and also documented.
The dosing procedures in the GLP mode have been optimized and realized
for the 700 Dosino in such a way that meaningful characteristic values can
be determined; these allow detailed information about the accuracy of the
dosing cylinder to be obtained.
You can adapt the necessary limits for the assessment of the reliability of
the dosing accuracy yourself to suit your requirements. The limits recommended by Metrohm are set as standard. We recommend that you use
these limits.
711 Liquino, Instructions for use
3 Introduction
21
CONFIG
A
The dialog language, date and time can be set in the configuration menu.
J
ENTER
YES
• and then <> and <ENTER>.
DOS NR
SELECT
P
•
• With <ENTER> the entry ‘dialog: english’ can be accepted.
CLEAR
Q
• The number keys are used to enter the current date in the format year-
lete the entry. (The date is “year 2000 safe“!)
•
• Enter the current time and confirm it with <ENTER>.
QUIT
NO
•
3.2 Tutorial
This tutorial will introduce you step by step to the operation of the Metrohm 711
Liquino. It is arranged so that you can work through it from the beginning to the
end and in this way become familiar with the most important functions of the instrument. You will be guided through each section step by step so that you will
be able to carry out the simplest settings. You can also carry out further fine adjustments. The necessary information for this can be found in section 4.
3.2.1 Basic configuration
The most important basic settings are carried out as follows:
•Press <CONFIG>
Di s pl ay : conf i gur at i on
>per i pher al uni t s
Di s pl ay : >auxi l i ar i es
di al og: engl i sh
The colon in this menu entry indicates that parameters can be selected
from a list here. Press the <SELECT> key several times in order to
view the different selection possibilities and become familiar with this
operating mode.
Di s pl ay : >auxi l i ar i es
di al og: deut sch
Di s pl ay : >auxi l i ar i es
dat e: 90- 01- 01
month-date. If a mistake is made then <CLEAR> can be used to de-
The entry is accepted with <ENTER>.
Di s pl ay : >auxi l i ar i es
t i me: 01: 00: 00
Di s pl ay : >auxi l i ar i es
device l abel ********
Press <QUIT> 2x in order to exit the configuration menu.
711 Liquino, Instructions for use
3.2 Tutorial
22
Except for the "Sampling dosing unit" in the SMPL mode the standard port
occupancy for all modes is as follows:
Port 1: dosing unit output
Port 2 (bottle): refilling opening
CONFIG
A
• Press <CONFIG> and then 3x <>.
ENTER
YES
•
DOS NR
SELECT
P
•One of the ten configurable dosing units can be selected with the
Di s pl ay : >dosi ng uni t s 1
3.2.2Defining dosing units
In order for the complete tubing system of a Dosino to be fully filled with the
reagent solution without any air bubbles the 711 Liquino must know the dimensions of the connected tubing. For this reason a dosing unit can be de-
*)
fined in the configuration menu and then selected for the method.
Di s pl ay : conf i gur at i on
>dosi ng uni t s
Access the submenu ‘dosing units’ with <ENTER>.
Di s pl ay : > dosi ng uni t s
dos. uni t : D1
<SELECT> key. Press <ENTER> to access the settings for the corresponding dosing unit. The index for the dosing unit is located in the top
right-hand corner of the display.
Di s pl ay : >dosi ng uni t s 1
I D D1
• Enter an identification text for the dosing unit or leave the preset entry
as it is. Text input is explained on page 63.
•Accept the entry with <ENTER>.
Di s pl ay : >dosi ng uni t s 1
l engt h t ube1 400 mm
• Now enter the length of tubing at Dosino port 1 (for tandem dosing
including the tubing length up to the T-piece) and accept the entry with
<ENTER>.
Di s pl ay : >dosi ng uni t s 1
di am. t ube1 2. 0 mm
• The inner diameter of the tubing at Dosino port 1 can be entered here
and confirmed with <ENTER>.
Di s pl ay : >dosi ng uni t s 1
l engt h t ube2 250 mm
• Now enter the length of the tubing at Dosino port 2 (bottle) and accept
the entry with <ENTER>.
*)
In a single method ‘standard’ can also be selected for the dosing unit used. For the modes XDOS, CONT
and GLP ‘standard’ means: Port1 (dosing port): L = 40cm, D = 2mm
Port2 (bottle): L = 25cm, D = 2mm.
For PIP ‘standard’ means: Port1 (pipetting tubing): L = 150cm, D = 2mm
Port2 (bottle): L = 25cm, D = 2mm.
For SMPL ‘standard’ means: Port1 (sample tubing): L = 40cm, D = 2mm
Port2 (measuring cell): L = 16cm, D = 2mm
Port3 (rinsing agent): L = 50cm, D = 2mm.
711 Liquino, Instructions for use
3 Introduction
23
di am. t ube2 2. 0 mm
QUIT
NO
• The dosing unit has now been defined. Press <QUIT> 2x to exit the
• The inner diameter of the tubing at Dosino port 2 can be entered here
and confirmed with <ENTER>.
Di s pl ay : >dosi ng uni t s 1
l engt h t ube3 0 mm
• As no tubing is connected to Dosino ports 3 and 4 the tubing lengths
must be set to 0 mm. Check this for port 3, press <ENTER> and
check the value for port 4 (the diameter is not requested if the length
has been set to 0 mm). Press <ENTER> again.
Di s pl ay : >dosi ng uni t s 2
I D D2
configuration menu.
711 Liquino, Instructions for use
3.2 Tutorial
24
MANUAL
MAN OFF
D
DOS NR
SELECT
P
• Now use the <DOS NR / SELECT> key to select the connected dosing
PREP
4
L
•
DOSING
7
G
START
Z
• If you now press the <START> key and hold it down then dosing will take
3.2.3 Dosing and dispensing
This is where you learn the simplest functions of the Liquino: manual dosing
and dispensing.
Fit a reagent bottle with a dosing unit and a Dosino and connect the Dosino
to the Liquino. Select, e.g. connection socket 'Dos. 1' or 'Dos. 2' for this.
Now switch the 711 Liquino briefly off and then on again. In this way the instrument recognizes the connected Dosino.
Details about setting up dosing units can be found in section 6.1.2, page
143.
Now immerse the dosing tip, which must be connected to Dosino Port 1 by
tubing, in a rinsing beaker.
• Switch the Liquino to the manual operating mode. Press the <MANUAL>
key.
Di spl ay: manual r eady 0. 000 mL
device. On the 711 Liquino keyboard there are four LEDs. If you have connected the Dosino to socket '
SELECT> key once so that the LED '
Dos. 2' then press the <DOS NR /
DOS 1' lights up.
<1>
• Prepare the dosing unit by pressing the <PREP> key.
The PREP function is used to rinse the tubing and fill the dosing cylinder
free from bubbles.
Di s p l ay: wast e por t :
tip(1) or flask(2)?
• Now select whether the rinsing volume of the reagent solution should be
discharged via the dosing tip (connected to Dosino Port 1) or returned to
the storage bottle (via filling Port 2 of the Dosino). Press <
1>.
Di spl ay: manual
PREP r unni ng. . .
• The tubing is now automatically rinsed with the required volume of
reagent. Excess reagent solution is discharged into the rinsing beaker.
Wait until the Liquino reports 'ready'.
Now exchange the rinsing beaker for the vessel into which you wish to
carry out dosing.
•Press the <DOSING> key.
Di spl ay: man u a l DOSI NG
r eady 0. 000 mL
place when the Dosino has switched the stopcock. You will notice that the
dosing speed increases slowly. If you release the <START> key and then
press it down again it will again start dosing at a slow speed which is
gradually increased.
This behavior can be very useful when you need to dose accurately, e.g.
to fill a volumetric flask up to the mark.
711 Liquino, Instructions for use
Loading...
+ 150 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.