Metrohm 709IC User Manual

709 IC Pump
IC Pump709
FLOW
P
min
P
max
P
actual
R/S
METROHM Ltd. CH-9101 Herisau
Instructions for Use
8.709.1033
CH-9101 Herisau/Switzerland Phone ++41 71 353 85 85 Fax ++41 71 353 89 01
CompuServe 100031,3703 Internet http://www.metrohm.com E-Mail sales@metrohm.ch
709 IC Pump
IC Pump709
FLOW
P
min
P
max
P
actual
R/S
SELECT EXT. PURGE
8.709.1033 Instructions for Use
16.10.2003 / chs
Table of contents
Table of contents
1 Introduction ..................................................................................... 1
1.1 Instrument description.................................................................... 1
1.2 Parts and controls ........................................................................... 2
1.3 Information about the Instructions for Use.................................. 4
1.3.1 Organization........................................................................... 4
1.3.2 Notation and pictograms ....................................................... 5
1.4 Safety notes...................................................................................... 6
1.4.1 Electrical safety ......................................................................6
1.4.2 General safety rules ............................................................... 6
2 Installation........................................................................................ 7
2.1 Setting up the 709 IC Pump .......................................................... 7
2.1.1 Packaging .............................................................................. 7
2.1.2 Check ..................................................................................... 7
2.1.3 Location.................................................................................. 7
2.1.4 Arrangement of the instruments............................................. 7
2.2 Mounting the pump head ...............................................................8
2.3 Attaching the accessories.............................................................. 9
2.3.1 Capillaries............................................................................... 9
2.3.2 PEEK connectors ...................................................................9
2.3.3 Steel connectors .................................................................. 10
2.3.4 Pulsation dampener ............................................................. 10
2.3.5 Filter unit PEEK..................................................................... 11
2.3.6 Filter unit Manufit.................................................................. 12
2.4 Connecting the tubing ..................................................................13
2.4.1 Connection to injection valve with PEEK capillaries ............ 13
2.4.2 Connection to injection valve with steel capillaries .............. 14
2.4.3 Connection to eluent container ............................................ 15
2.5 Electrical connection ....................................................................15
709 IC Pump
2.6 Mains connection ..........................................................................16
2.6.1 Setting the mains voltage..................................................... 16
2.6.2 Fuses.................................................................................... 16
2.6.3 Mains cable and mains connection ..................................... 17
2.6.4 On/off switching of the instrument ....................................... 17
2.7 Deaerating the pump..................................................................... 18
I
Table of contents
3 Operation ........................................................................................ 19
3.1 Sequence of operations ............................................................... 19
3.2 Flussrate......................................................................................... 20
3.3 Flow correction.............................................................................. 20
3.4 Pressure limit values for safety shutdown ................................ 21
3.4.1 Minimum pressure limit........................................................ 21
3.4.2 Maximum pressure limit....................................................... 22
3.5 Switching the pump drive on and off ......................................... 22
3.6 On/off switching of the external control .................................... 23
3.7 Inquiry of the current pressure ................................................... 23
3.8 Overview of the key functions..................................................... 23
4 Notes – Maintenance – Faults...................................... 25
4.1 Practical notes............................................................................... 25
4.1.1 Protection against foreign particles ..................................... 25
4.1.2 Pulsation dampener............................................................. 25
4.1.3 Eluents ................................................................................. 25
4.2 Maintenance and servicing.......................................................... 26
4.2.1 General information ............................................................. 26
4.2.2 Maintenance work at the pump head .................................. 27
4.3 Faults and malfunctions............................................................... 31
4.4 Diagnosis........................................................................................ 32
4.4.1 General information ............................................................. 32
4.4.2 Entry of diagnostics with automatic display test.................. 33
4.4.3 Keyboard test....................................................................... 33
4.4.4 RS232 Test........................................................................... 34
4.4.5 Initialize and test RAM.......................................................... 35
4.5 Validation / GLP............................................................................. 36
5 Interfaces ....................................................................................... 37
5.1 RS232 interface ............................................................................. 37
5.1.1 General rules for remote control .......................................... 37
5.1.2 Call-up of objects................................................................. 38
5.1.3 Trigger.................................................................................. 39
5.1.4 Status messages ................................................................. 40
5.1.5 Error messages.................................................................... 41
5.1.6 RS232 error rectification....................................................... 41
5.1.7 Remote control commands ................................................. 42
5.1.8 Data transmission protocol.................................................. 43
709 IC Pump
II
Table of contents
5.1.9 Handshake ........................................................................... 43
5.1.10 Pin assignment................................................................... 44
5.1.11 Setting the baud rate.......................................................... 45
5.2 Control interface............................................................................ 46
5.2.1 Functions.............................................................................. 46
5.2.2 Settings for the control interface ..........................................46
5.2.3 Control inputs and outputs................................................... 47
6 Appendix ..........................................................................................49
6.1 Technical specifications............................................................... 49
6.2 Scope of delivery...........................................................................52
6.3 Optional accessories .................................................................... 54
6.4 Warranty and conformity .............................................................. 57
6.4.1 Warranty ............................................................................... 57
6.4.2 EU Declaration of conformity ...............................................58
6.4.3 Certificate of conformity and system validation ................... 59
6.5 Index................................................................................................ 60
List of figures
Fig. 1: IC system with 732 IC Detector, 733 IC Separation Center
and 766 IC Sample Processor .............................................................. 1
Fig. 2
: Front and rear of 709 IC Pump .............................................................2
Fig. 3
: Connectors for capillaries ..................................................................... 9
Fig. 4
: Filter unit PEEK (6.2821.100) .............................................................. 11
Fig. 5
: Filter unit Manufit (6.2821.000) ........................................................... 12
Fig. 6
: Connection to injection valve with PEEK capillaries ........................... 13
709 IC Pump
Fig. 7
: Connection to injection valve with steel capillaries............................. 14
Fig. 8
: Connection at 732 IC Detector............................................................ 15
Fig. 9
: Components of the pump head.......................................................... 28
Fig. 10
Fig. 11
: Replacement of the piston seal 67 ..................................................... 28
: Components of inlet valve 68 and outlet valve 69.............................. 30
III
Table of contents
List of numbered parts and controls
1 Display .......................................... 2
2 LED "R/S"....................................... 2
3 Key <R/S> ................................... 2
4 Key <SELECT> ........................... 2
5 Key < ↓ >..................................... 2
6 Key < ↑ >..................................... 2
7 LED "Ext." ...................................... 2
8 Key <EXT.> ................................. 2
9 Key <PURGE> ............................ 2
10 LED "FLOW" .................................. 2
11 LED "P 12 LED "P 13 LED "P
"....................................... 2
min
" ...................................... 2
max
" .................................... 2
actual
14 Connection capillary ..................... 2
15 Aspirating capillary........................ 2
16 Inlet valve screw connection......... 2
17 Connection capillary ..................... 2
18 Pump head .............................. 2,28
19 Fastening screws .......................... 2
20 Outlet capillary .............................. 2
21 Outlet valve screw connection ...... 2
22 Connection capillary ................ 2,13
23 Connection.................................... 2
24 Purge valve ................................... 2
25 Screw ............................................ 2
26 Screw ............................................ 2
27 Housing cover............................... 2
28 Fuse data ...................................... 2
29 Mains switch ................................. 2
30 Fuse cover .................................... 2
31 Mains connection plug ................. 2
32 Mains voltage selectors ................ 2
33 Control interface............................ 2
34 Recorder output socket ................ 2
35 Sliding switch ................................ 2
36 Sliding switch ................................ 2
37 RS232 interface............................. 2
38 Current Loop interface .................. 2
39 Manufacturing number.................. 2
40 Instrument data ............................. 2
41 Ferrule........................................... 9
42 Pressure screw ............................. 9
43 Capillary ......................... 9,11,13,14
44 Compression fitting .................. 9,11
45 Connector ................................... 11
46 Housing for filter unit ................... 11
47 Connector ................................... 11
48 Inlet capillary .......................... 12,14
49 Manufit pressure screw............... 12
50 Counterpart end .......................... 12
51 PTFE gasket ................................ 12
52 Steel meshes .............................. 12
53 Steel mesh holding end .............. 12
54 Manufit housing .......................... 12
55 Outlet capillary ....................... 12,14
56 Filter unit PEEK ...................... 11,13
57 Pulsation dampener ............... 13,14
58 Filter unit Manufit.................... 12,14
59 Coupling...................................... 14
60 Screw .......................................... 28
61 Piston .......................................... 28
62 Spring retainer............................. 28
63 Spring.......................................... 28
64 Piston cartridge ........................... 28
65 Piston guide sleeve..................... 28
66 Sapphire supporting ring ............ 28
67 Piston guide sleeve ..................... 28
68 Piston seal................................... 28
69 Inlet valve ............................... 28,30
70 Outlet valve ............................ 28,30
71 Screw holder ............................... 28
72 Special tool ................................. 28
73 Special tool ................................. 28
74 Valve housing.............................. 30
75 Sealing ring ................................. 30
76 Sleeve.......................................... 30
77 Sapphire sleeve .......................... 30
78 Sapphire sphere.......................... 30
79 Ceramic holder............................ 30
80 Seal .............................................30
709 IC Pump
IV
1.1 Instrument description

1 Introduction

1.1 Instrument description

The 709 IC Pump is a serial dual piston pump especially developed for ion chromatography and optimized for use with Metrohm ion chroma­tographic instruments (e.g. 732 IC Detector and 733 IC Separation Cen­ter). The 709 IC Pump operates with minimal residual pulsation and has an excellent flow constancy. In addition to the 2.709.0010 Standard version with parts made of sapphire, ruby, ceramics, Kel-F, Teflon or stainless steel, the totally metal free 2.709.0110 version is available.
All functions can be set simply at the front of the unit. The current and entered data are indicated on the LED display. To assure optimum op­erational reliability, both lower and upper pressure limit values can be set. If such a limit is reached, the pump switches itself off automatically. An additional purge valve allows rapid removal of air from the pump system.
709 IC Pump
Fig. 1
: IC system with 732 IC Detector, 733 IC Separation Center
and 766 IC Sample Processor
1
1 Introduction

1.2 Parts and controls

3 1 13 24
2 12 11 10 23 22 21 20
IC Pump709
FLOW
P
min
P
max
P
actual
R/S
SELECT EXT. PURGE
45
28
Made by Metrohm Herisau Switzerland
26 2627
FUSE 230 V: 0,25 AT 115 V: 0,5 AT
S
6 8 9 7 14 15 1918171625 25
29
30 3940
WARNING - Risk of olectric shock - disconnect equ ipment from supply before opening.
230V
CONTROL
REC.
Type 1.709. f=50-60 Hz S=50 VA
TIME
CONST.
HIGH
LOW
CURRENT
LOOP
RS 232
CURRENT
LOOP
RS 232
31 32 33 34 38
Fig. 2
: Front and rear of 709 IC Pump
709 IC Pump
2
35 36 37
1.2 Parts and controls
1 Display (LED)
2 LED "R/S"
Lights up when pump drive switched on
3 Key <R/S> (Run/Stop)
On/off switching of the pump drive
4 Key <SELECT>
Switching display 1 to "Flow", "P "P
max
" or "P
actual
"
5 Key < ↓ >
Continuous decrement of the dis­played digital values
6 Key < ↑ >
Continuous increment of the displayed digital values
7 LED "Ext."
Lights up when external control is switched on
8 Key <EXT.>
On/off switching of external control via RS232 or current loop interface
min
",
18 Pump head (6.2824.040/6.2824.100)
19 Fastening screws for pump head 18
20 Outlet capillary of purge valve
21 Outlet valve screw connection
22 Connection capillary to 733 IC
Separation Center
23 Attachment for connection capillary to
733 IC Separation Center
24 Purge valve
25 Screw for housing cover
26 Screw for housing cover
27 Housing cover
28 Fuse data
29 Mains switch
Switch for on/off switching of the unit: I = ON 0 = OFF
30 Fuse cover
Changing the fuses, see section 2.6.2
9 Key <PURGE>
Deaeration of pump
10 LED "FLOW"
Lights up when flow rate is shown in display 1
11 LED "P
min
"
Lights up when lower pressure limit value for automatic shutoff of the pump is shown in display 1
12 LED "P
max
"
Lights up when upper pressure limit value for automatic shutoff of the pump is shown in display 1
13 LED "P
actual
"
Lights up when current pressure is shown in display 1
14 Connection capillary between purge
valve and pump head
15 Aspirating capillary
31 Mains connection plug
Mains connection see section 2.6
32 Mains voltage selector
115V: 100...120 V ± 10% 230V: 220...240 V ± 10%
33 Control interface
see section 5.2
34 Recorder output socket
Recorder output: 1 mV/MPa
35 Sliding switch "Time Constant"
Time constant for the recorder output: HIGH ≈ 1 s; LOW ≈ 20 ms
36 Sliding switch for external control:
RS232 / Current Loop
37 RS232 interface
38 "Current Loop" interface
(details on request)
39 Manufacturing number
16 Inlet valve screw connection
17 Connection capillary
in pump head (integral)
709 IC Pump
40 Instrument data
Type number, mains frequency, power consumption
3
1 Introduction

1.3 Information about the Instructions for Use

Please read through these Instructions for Use carefully before operating the 709 IC Pump. The Instructions for Use contain informa­tion and warnings to which the user must pay attention in order to assure safe operation of the instrument.

1.3.1 Organization

These 8.709.1033 Instructions for Use for the 709 IC Pump provide a comprehensive overview of the installation, startup procedure, opera­tion, fault rectification and technical specifications of this instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
Description of the instrument, parts and controls, safety notes
Section 2 Installation
Mounting the pump head, electrical connections, tubing connections, mains connection
Section 3 Operation
Detailed description of operation, explanation of all key functions
Section 4 Notes – Maintenance – Faults
Practical notes, maintenance, fault rectification, diagnosis, validation
Section 5 Interfaces
Description of RS232 interface, remote control language, control interface
Section 6 Appendix
Technical data, standard equipment, options, warranty, declaration of conformity, index
To find the information you require about the instrument please use ei­ther the Table of contents or the Index at the back.
As a supplement to the Instructions for Use, the Metrohm Monograph
8.732.2003 "Ion chromatography" is also supplied. This provides an introduction to the theoretical fundamentals and general information on separating columns and sample pretreatment. You will find detailed in­formation on the separating columns available from Metrohm and on special IC applications in the relevant "Application Bulletins", which are available on request free of charge from your Metrohm agency.
709 IC Pump
4
1.3 Information about the Instructions for Use

1.3.2 Notation and pictograms

The following notations and pictograms (symbols) are used in these In­structions for Use:
<SELECT> Key
"0.0" Entry value
25 Part or control
(see section 1.2)
Hazard This symbol draws attention to a possible danger to life or injury if the associated directions are not followed correctly.
Warning This symbol draws attention to possible damage to instru­ments or instrument parts if the associated directions are not followed correctly.
Caution This symbol marks important information. Read these direc­tions before continuing.
Comment This symbol marks additional information and tips.
709 IC Pump
5
1 Introduction

1.4 Safety notes

1.4.1 Electrical safety

While electrical safety in the handling of the 709 IC Pump is assured in the context of the specifications IEC 1010-1 (protection class 1, degree of protection IP40), the following points should be noted:
Mains connection
Set the mains voltage and check the mains fuse and mains connection in accordance with the instructions in section 2.6.
Opening the 709 IC Pump
To avoid all danger of coming into contact with live components do not open the instrument or remove any parts when the 709 IC Pump is connected to the power supply. Always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection plug 31 !
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 709 IC Pump, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instru­ment or a radiator) to eliminate any static charges which exist.

1.4.2 General safety rules

Handling of solvents
Check the pump tubing and all input and output leads periodically for possible leaks. Follow the relevant instructions regarding the handling of flammable and/or toxic solvents and their disposal.
709 IC Pump
6

2.1 Setting up the 709 IC Pump

2 Installation

2.1 Setting up the 709 IC Pump

2.1.1 Packaging

The 709 IC Pump is supplied together with the separately packed accessories in special packagings containing shock-absorbing foam linings designed to provide excellent protection. The actual instrument is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they assure transport of the instruments free from damage.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 6.2). In the case of transport damage, see instructions in section 6.4.1 "Warranty".

2.1.3 Location

Position the 709 IC Pump in the laboratory at a location convenient for operation, free from vibrations and protected against a corrosive at­mosphere and contamination by chemicals.
To avoid disturbing temperature influences on the insulated column compartment, the pump and eluent reservoir must be protected against direct sunlight.

2.1.4 Arrangement of the instruments

In one-channel operation, the 709 IC Pump, 733 IC Separation Center and 732 IC Detector are best stacked on top of one another in this or­der.
709 IC Pump
In two-channel operation (733.0X20 IC Separation Center), the optimum arrangement (1, 2 or 3 towers) depends on the laboratory space avail­able. However, the 709 IC Pumps should be set up at the very bottom and the 732 IC Detectors at the very top.
7
2 Installation

2.2 Mounting the pump head

To ensure that the pump drive is not damaged during transport a black plastic block is mounted in place of the pump head on the right side of the 709 IC Pump front panel. This transport safeguard block has to be removed after undoing the four screws.
To avoid any damage to the pump head, the protective block must be remounted if the pump has to be moved any great distance.
Next, the 6.2824.040 Pump Head 18 (6.2824.040 or 6.2824.100) with in­tegrated piston units is unpacked and fixed in place of the transport safeguard block with the four hexagon screws 19 as shown in Fig. 2.
To ensure that the pump head is not positioned wrongly, the holes at the rear for the clamping bolts have different depths, i.e. 1 clamping bolt is longer than the rest. The deepest hole must naturally accom­modate the longest bolt. If this is not the case, the pump will not function properly.
The connection capillary 14 fastened to purge valve 24 is screwed onto pump head 18 as shown in Fig. 2.
The 2.709.0110 IC Pump MF (metal-free version) should only be
operated in the pressure range of 0 higher pressure, the 6.2824.100 PEEK pump head can be damaged.
So make sure that the maximum shutoff pressure always set to
25.0 MPa (see section 3.4.2).
25.0 MPa (0…250 bar). With
P
of the pump is
max
709 IC Pump
8

2.3 Attaching the accessories

2.3 Attaching the accessories

2.3.1 Capillaries

Some of the connections under high pressure between the 709 IC Pump and the IC system must be set up by the user. For the pressure range 025 MPa (0250 bar) the 6.1831.010 PEEK capillary
1
(i.d. = 0.25 mm, e.d. = of the 709 IC Pump can be used for both pump versions. For the pres­sure range 2550 MPa (250500 bar), which is permissible only for the not (i.d. = 0.25 mm, e.d. =
With PEEK capillaries the connection is made preferably with the
6.2744.010 PEEK connectors (see section 2.3.2), with steel capillaries, the 6.2620.000 and 6.2620.010 steel connectors should be used (see section 2.3.3).
metal-free 2.709.0010 version, the 6.2620.020 Steel capillary
Capillaries fitted with new connectors must have a perfectly flat cut surface. For PEEK capillaries it is best to use the 6.2621.080 Capil-
lary tubing cutter, for steel capillaries the 6.2621.040 Capillary tubing cutter.
/16", length = 3 m) supplied in the accessories
1
/16", length = 3 m) must be used.
4241 43 44
: Connectors for capillaries
Fig. 3

2.3.2 PEEK connectors

41 Ferrule (6.2620.010)
42 Pressure screw (6.2620.000)
43 Capillary
6.2620.020 Steel capillary or
6.1831.010 PEEK capillary
44 Compression fitting
(6.2744.010)
709 IC Pump
For the connection of 6.1831.010 PEEK capillaries or of 6.1822.010 PTFE Microcapillaries (i.d. = 0.3 mm), the supplied 6.2744.010 PEEK Compression fittings are used. It is best to proceed as follows:
1 Mount compression fitting
Slide a compression fitting 44 (6.2744.010) over the end of the capillary 43 to be fastened as shown in Fig. 3.
9
2 Installation
2 Insert capillary in connection
Push capillary end in the corresponding connection as far as it will go (to avoid dead volume).
3 Tighten compression fitting
Tighten compression fitting 44 by hand (never use tools).

2.3.3 Steel connectors

For the connection of steel capillaries, the steel connectors 6.2620.010 Ferrule and 6.2620.000 Pressure screw available as an option are
used. Proceed as follows:
1 Mount connectors
Slide a pressure screw 42 (6.2620.000) and a ferrule 41 (6.2620.010) over the end of the capillary 43 to be fastened as shown in Fig. 3.
2 Insert capillary in connection
Push capillary end into the corresponding connection as far as it will go (to avoid dead volume).
3 Tighten pressure screw
Tighten pressure screw 42 with the open-end spanner 1/4" (6.2621.010) supplied.

2.3.4 Pulsation dampener

To protect the column material against pressure drops caused by the injector, the use of a pulsation dampener connected between the pump and the injection valve of the 733 IC Separation Center is recom­mended. The optional 6.2620.150 Pulsation dampener MF is very well suited to this purpose (see section 6.3).
The metal-free 6.2620.150 Pulsation dampener is supplied fully assem­bled and has two connections for capillaries, for which the connectors supplied or two 6.2744.010 PEEK compression fittings can be used. The flow direction is arbitrary. The pulsation dampener is positioned in the interior of the 733 IC Separation Center on the base below the injec­tion valve (see 732/733 Instructions for Use).
709 IC Pump
10
2.3 Attaching the accessories

2.3.5 Filter unit PEEK

The 6.2821.100 Filter unit PEEK supplied (see Fig. 4) serves to avoid contamination of the piston seals by abrasive particles and can be used in the pressure range 025 MPa (0250 bar). The filter unit consists of the housing 46 and the two connectors 45 (with filter) and 47 (without filter) to be screwed into the housing 46. For the connection of capillar­ies 43, PEEK compression fittings 44 (6.2744.010) must be used.
New connectors 45 with filter are available as an option with the order­ing number 6.2821.110 (10 pieces).
For the connection of the filter unit, please note the flow direction arrow printed on the housing.
43 44 45 46 47 44 43
Fig. 4
: Filter unit PEEK (6.2821.100)
43 Capillary
6.1831.010 PEEK capillary
44 Compression fitting (6.2744.010) 47 Connector without filter
45 Connector with filter (6.2824.110)
Part of 6.2824.100 Filter unit
46 Housing for filter unit
Part of 6.2824.100 Filter unit
Part of 6.2824.100 Filter unit
709 IC Pump
11
2 Installation

2.3.6 Filter unit Manufit

The 6.2821.000 Filter unit Manufit available as an option (see sec­tion 6.3) serves to avoid contamination of the piston seals by abrasive
particles and can be used in the pressure range 050 MPa (0500 bar) together with steel capillaries. It is installed as follows (see Fig. 5):
1 Prepare Manufit housing
Insert outlet capillary 55 with steel mesh holding end 53 into
Manufit housing 54.
Insert the 4 steel meshes 52 provided into the steel mesh
holding end 53.
Press the PTFE gasket 51 into the steel mesh holding end
53.
2 Prepare Manufit pressure screw
Insert inlet capillary 48 with counterpart end 50 into Manufit
pressure screw 49.
3 Assembly
Fit the two capillary end pieces 50 and 53 together.
Screw Manufit pressure screw 49 and Manufit housing 54
firmly together.
To replace contaminated steel meshes, proceed in the reverse order.
48
IC Pump 709
48 Inlet capillary
49
49 Manufit pressure screw
50
51
52
50 Counterpart end
51 PTFE gasket (6.2821.010)
52 4 Steel meshes (6.2821.020)
53
53 Steel mesh holding end
54 Manufit housing
54
IC Separation
55
Fig. 5
: Filter unit Manufit (6.2821.000)
709 IC Pump
12
Center 733
55 Outlet capillary

2.4 Connecting the tubing

2.4 Connecting the tubing

2.4.1 Connection to injection valve with PEEK capillaries

For the pressure range 025 MPa (0250 bar), it is recommended to use 6.1831.010 PEEK capillaries, a 6.2620.150 Pulsation dampener (see section 2.3.4), and a Filter unit PEEK (see section 2.3.5) to connect the 709 IC Pump to the injection valve of the 733 IC Separation Center. Proceed as follows:
1 Connection to 709 IC Pump
Cut connection capillary 22 (6.1831.010 PEEK capillary) to
the required length and equip it with connectors.
Attach connection capillary 22 to connection 23 of the 709 IC
Pump (see Fig. 2).
2 Connection of filter unit PEEK
Attach the other end of connection capillary 22 at the connec-
tor 45 (with filter) of the filter unit 56 (see Fig. 6).
Connect a PEEK capillary 43 cut to the required length and
equipped with connectors to the connector 47 of the filter unit
56.
3 Installation of the capillary in the IC Separation
Center
Procedure, see 732/733 Instructions for Use, section 2.6.5.
4 Connection of the pulsation dampener
Procedure, see 732/733 Instructions for Use, section 2.6.5.
5 Connection to the injection valve
Procedure, see 732/733 Instructions for Use, section 2.6.5.
22 56 43
709
Fig. 6
: Connection to injection valve with PEEK capillaries
57 43
733
22 Connection capillary
6.1831.010 PEEK capillary
43 PEEK capillary (6.1831.010) 57 Pulsation dampener (6.2620.150)
709 IC Pump
56 Filter unit PEEK (6.2821.100)
13
2 Installation

2.4.2 Connection to injection valve with steel capillaries

For the pressure range 2550 MPa (250…500 bar), it is recommended to use 6.2620.020 steel capillaries, a 6.2620.150 Pulsation dampener (see section 2.3.4), and a Filter unit Manufit (see section 2.3.6) to con­nect the not metal-free 733 IC Separation Center. Proceed as follows:
1 Connection of filter unit Manufit
Attach inlet capillary 48 of the filter unit Manufit 58 to connec-
tion 23 of the 709 IC Pump (see Fig. 2).
Attach outlet capillary 55 of the filter unit Manufit 58 using a
coupling 59 to a steel capillary 43 cut to the required length and equipped with connectors (see Fig. 7).
2 Installation of the capillary in the IC Separation
Center
Procedure, see 732/733 Instructions for Use, section 2.6.5.
2.709.0010 IC Pump to the injection valve of the
43 Capillary
6.2620.020 Steel capillary
3 Connection of the pulsation dampener
Procedure, see 732/733 Instructions for Use, section 2.6.5.
4 Connection to the injection valve
Procedure, see 732/733 Instructions for Use, section 2.6.5.
45 59 43
709
Fig. 7
: Connection to injection valve with steel capillaries
57 4358 52
733
57 Pulsation dampener (6.2620.150)
48 Inlet capillary
58 Filter unit Manufit (6.2824.000)
of filter unit Manufit 58
55 Outlet capillary
59 Coupling (6.2620.060)
of filter unit Manufit 58
709 IC Pump
14
2.5 Electrical connection

2.4.3 Connection to eluent container

The 6.1834.000 Aspirating tubing (i.d. = 1.5 mm, e.d. = 2.5 mm, length = 1.2 m) supplied is slid over the aspirating capillary 15. The
6.2821.090 Aspirating filter supplied is screwed onto the other end of the tubing. The end of the tubing with the aspirating filter is then intro­duced in the eluent container.
Only degassed (with N
, He or vacuum) and microfiltered (0.45 µm
2
filter) eluents may be used!
It must be ensured that the eluent used is freely miscible with any residual solvent in the pump head (the pump head is filled with isopropanol or methanol/water in the factory). If this is not the case, the pump must first be rinsed with a solvent that is miscible with both the previous and subsequent eluent (e.g. acetone).
If you use the optional available 6.5324.000 Bottle rack, the
6.1834.010 Aspirating tubing (accessory of the bottle rack, length =
2.5 m) must be installed instead of the 6.1834.000 Aspirating tubing. The end of the tubing with the screwed-on 6.2821.090 aspirating filter is then fixed in the eluent vessel according to the instructions on the leaflet supplied with the bottle rack.

2.5 Electrical connection

The connection of the 709 IC Pump at the 732 IC Detector is made acc. to Fig. 8 using the 6.2125.060 cable or another RS cable specified as a "null modem" cable. To ensure proper functioning of the communi­cation between the 732 IC Detector and 709 IC Pump, the sliding switch 36 on the IC pump must be set to "RS 232" and the external control switched on with <EXT.> key 8 (see section 3.6).
732
709
Fig. 8
: Connection at 732 IC Detector
6.2125.060
709 IC Pump
15
2 Installation

2.6 Mains connection

Follow the instructions below for connecting to the power supply. If the instrument is operated with a mains voltage set wrongly and/or wrong mains fuse, there is a danger of fire!

2.6.1 Setting the mains voltage

Before switching on the 709 IC Pump for the first time, check whether the mains voltage set on the instrument (can be read in the mains volt­age selector 32) matches the local mains voltage. If this is not the case, you must switch the mains voltage by pushing the mains voltage selec­tor 32 with a screwdriver:
Position of the mains voltage selector 32:
230V
115V
230V: 220240 V ± 10%
115V: 110120 V ± 10%

2.6.2 Fuses

The 709 IC Pump contains two fuses built in as standard, either of type
0.25 AT (for 230 V: 0.25 A, slow-blow, Metrohm ordering number U.600.0010) or 0.5 AT (for 110 V: 0.5 A, slow-blow, Metrohm ordering number U.600.0013). Ensure that the two fuses correspond to the fuse type specified for the local mains voltage (see fuse data 28). To change wrong or blown fuses, proceed as follows:
Ensure that the instrument is never put into operation with fuses of another type, otherwise there is danger of fire!
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 31 of the 709 IC Pump.
2 Remove fuse cover
Using a screwdriver, lever out fuse cover 30 forwards until it opens.
3 Check and change fuses if necessary
Carefully take the fuses installed for the desired mains voltage out of the fuse holder and check its specifications:
100120 V 0.5 A (slow-blow) Metrohm No. U.600.0013 220240 V 0.25 A (slow-blow) Metrohm No. U.600.0010
Change fuse if necessary and reinsert in fuse holder.
709 IC Pump
16
2.6 Mains connection
4 Install fuse holder
Reinsert fuse holders in the instrument (the arrows printed on the holders must point in the same direction as the arrows on the inside of fuse cover 30).
5 Install fuse cover
Push in fuse cover 30 firmly until it clicks into place.
6 Connect mains cable
Plug mains cable into mains connection plug 31.

2.6.3 Mains cable and mains connection

Mains cable
The instrument is supplied with one of three mains cables:
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a dif­ferent plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it a hazard!
Mains connection
Plug the mains cable into mains connection plug 31 (see Fig. 2).

2.6.4 On/off switching of the instrument

The 709 IC Pump is switched on and off using mains switch 29. When the instrument is switched on, all segments of display 1 and all LEDs light up for a few seconds. The instrument is then in the standby mode ("ready" status).
709 IC Pump
17
2 Installation

2.7 Deaerating the pump

The purge valve 24 is opened by turning anticlockwise and the pump can now be deaerated by pressing the <PURGE> key 9.
The solvent bottle should be located somewhat above the pump, but at least at the same level as the pump head 18 (never below the pump). When the <PURGE> key is pressed the pump operates at the maxi­mum flow rate (5 mL/min).
9
For safety reasons the <PURGE> key mode of the pump, i.e. the <PURGE> key is disabled when the pump is running in normal operation.
Should the pump not prime by itself, the 6.2816.020 syringe supplied is used to siphon off solvent via the outlet capillary 20 with opened purge valve through the pump head until there are no more bubbles in the solvent entering the syringe. The pump can then again be purged briefly by pressing the <PURGE> key 9.
functions only in the stop
On completion of this procedure, the purge valve 24 is closed by turn­ing clockwise. The pump is now ready for operation.
709 IC Pump
18

3.1 Sequence of operations

3 Operation

3.1 Sequence of operations
Installation
1. Setting up the pump Section 2.1
2. Mounting the pump head Section 2.2
3. Attaching the accessories Section 2.3
4. Tubing connections Section 2.4
5. Electrical connection Section 2.5
6. Mains connection Section 2.6
Startup procedure
1. Switching on the pump Section 2.6
2. Deaerating the pump Section 2.7
Settings
1. Setting the flow rate Section 3.2
2. Flow correction (option) Section 3.3
3. Setting the pressure limit values Section 3.4
Switching on the pump drive
Switching off the pump drive
Section 3.5
Section 3.5
Standby mode
709 IC Pump
Switching off the pump
Section 2.6
19
3 Operation

3.2 Flow rate

The flow rate can be set in the range of 0.055.00 mL/min with the pump drive switched on or off as follows:
1 Select "FLOW"
Press <SELECT> key 4 until the "FLOW" LED 10 lights up.
2 Set flow rate
Press < > key 5 or < > key 6 until the desired flow rate (in mL/min) appears in display 1.
3 Store flow rate
Release < > key 5 or < > key 6: the value shown in display 1 at the time of release will be stored.

3.3 Flow correction

Since the difference between displayed and actual flow rate can be maximum 3 % owing to unavoidable tolerances in the manufacture of mechanical parts, there is a possibility to adjust the 709 IC Pump after installation of a new pump head. For this, a correction factor of 0.9…1.1 can be entered in the 709 IC Pump.
The correction factor is determined by measurement of the actual flow rate with the aid of a measuring cylinder:
Correction factor =
In the entry of the correction factor, proceed as follows:
1 Select flow correction
Press <SELECT> key 4 for ca. 3 s until all LEDs (10...13) go out and the correction factor appears in display 1.
2 Set flow correction
Press < > key 5 or < > key 6 until the desired flow correc­tion appears in display 1.
Displayed flow rate (in mL/min)
Measured flow rate (in mL/min)
3 Store flow correction
Release < > key 5 or < > key 6: the value shown in display 1 at the time of release will be stored.
4 Return to standby mode
Press <SELECT> key 4 briefly, the pump is again in the standby mode.
709 IC Pump
20

3.4 Pressure limit values for safety shutdown

3.4 Pressure limit values for safety shutdown

3.4.1 Minimum pressure limit

The minimum shutoff pressure of the pump can be set in steps of
0.1 MPa from 0.150 MPa with the pump drive switched on or off. Pro­ceed as follows:
1 Select "P
Press <SELECT> key 4 until the "P
min
"
" LED 11 lights up.
min
2 Set shutoff pressure
Press < > key 5 or < > key 6 until the desired minimum shutoff pressure (in MPa) appears in display 1.
3 Store shutoff pressure
Release < > key 5 or < > key 6: the value shown in display 1 at the time of release will be stored.
Entry of the value "0.0" switches off the lower pressure limit monitoring. The set limit value should lie far enough below the particular operating pressure. If the pump pressure falls below this preset lower limit during operation, the microprocessor checks whether this violation persists for several revolutions of the cam or whether it was a temporary drop in pressure due to an air bubble or a brief leak in the valves. If the cause is a temporary loss of pressure, the pump is not switched off. In the case of a pressure drop due to leaks or interrupted inflow of the eluent, the pump drive is switched off and disabled.
If the pump has been shut off by the minimum pressure monitoring, the LED 11 flashes, the shutoff pressure can be requested in the display
" (see section 3.7). The safety shutdown is not active during the
"P
actual
first two minutes after the pump drive has been switched on with the <R/S> key 3.
709 IC Pump
Pressing the <R/S> key 3 re-enables the pump drive, the LED 11 no longer flashes and the pump is in the standby mode. Pressing the <R/S> key 3 again puts the pump back into operation.
21
3 Operation

3.4.2 Maximum pressure limit

The maximum shutoff pressure of the pump can be set in steps of
0.1 MPa from 0.150 MPa with the pump drive switched on or off. Pro­ceed as follows:
1 Select "P
Press <SELECT> key 4 until the "P
max
"
" LED 12 lights up.
max
2 Set shutoff pressure
Press < > key 5 or < > key 6 until the desired shutoff pressure (in MPa) appears in display 1.
3 Store shutoff pressure
Release < > key 5 or < > key 6: the value shown in display 1 at the time of release will be stored.
The set limit value should lie between 5.0 and 10.0 MPa above the par­ticular operating pressure or the maximum admissible operating pres­sure of the column. If the pump exceeds the preset limit value during operation, the pump drive is switched off within 1 pump cycle and dis­abled. If the pump has been switched off by the maximum pressure limit monitoring, the LED 12 flashes, the shutoff pressure can be re­quested in the display "P
" (see section. 3.7).
actual
Pressing the <R/S> key 3 re-enables the pump drive, LED 12 no longer flashes and the pump is in the standby mode. Pressing the <R/S> key 3 again puts the pump back into operation.
The 2.709.0110 IC Pump MF (metal-free version) should only be
operated in the pressure range of 0 higher pressure, the 6.2824.100 PEEK pump head can be damaged.
So make sure that the maximum shutoff pressure
always set to
25.0 MPa.
25.0 MPa (0…250 bar). With
P
max

3.5 Switching the pump drive on and off

Switching on Pressing the <R/S> key 3 switches on the pump drive.
While the pump drive is running, LED 2 above the <R/S> key 3 lights up.
Switching off To switch off the pump drive, press the <R/S> key 3
again. LED 2 goes out.
of the pump is
709 IC Pump
22

3.6 On/off switching of the external control

3.6 On/off switching of the external control
Switching on Pressing the <EXT.> key 8 switches on the external
control. The interface (RS232, see section 5.1 or "Cur- rent Loop", details on request) is selected with sliding switch 36. With active external control the LED 7 above the <EXT.> key 8 lights up. All keys with the exception of <SELECT> 4 (selection of display function) are blocked.
Switching off To switch off the external control, the <EXT.> key 8 is
pressed again. The LED 7 goes out and all functions are once more accessible manually via the keypad.

3.7 Inquiry of the current pressure

The current pressure of the pump can be queried at any time when the pump drive is switched on. Proceed as follows:
1 Select "P
Press <SELECT> key 4 until the "P
actual
"
actual
2 Read pressure
The current pressure (in MPa) is shown in display 1.

3.8 Overview of the key functions

On/off switching of the pump drive
R/S
When the pump drive is switched on, the LED above the key is on.
" LED 13 lights up.
709 IC Pump
SELECT
Selecting the display function
FLOW Current flow rate in mL/min
Minimum shutoff pressure for automatic
P
min
Range: 0.01...5.00 mL/min
shutdown of the pump drive Range: 0.1... 50 MPa (1...500 bar) (P
min
< P
max
)
0.0: no shutoff
23
3 Operation
Maximum shutoff pressure for automatic
P
max
shutdown of the pump drive Range: 0.1...50 MPa (1...500 bar) (P
P
Current pressure
actual
max
> P
min
)
NX Correction factor for flow rate (see
section 3.2). The <SELECT> key must be pressed for ca. 3 s. Range: 0.90 ... 1.10
Input key for continuous lowering of the value selec­ted with the <SELECT> key. When the key is released the current value is stored.
Pressing this key briefly changes the displayed value by one digit, if the key is pressed longer the value is changed continuously and increasingly faster.
PURGE
EXT.
Input key for continuous raising of the value selected with the <SELECT> key. When the key is released the current value is stored.
Pressing this key briefly changes the displayed value by one digit, if the key is pressed longer the value is changed continuously and increasingly faster.
On/off switching of the external control of the pump via RS232 or current loop interface.
On/off switching of the maximum delivery rate (5 mL/min) to purge the pump.
This key is active only in the standby mode.
709 IC Pump
24
4.1 Practical notes

4 Notes – Maintenance – Faults

4.1 Practical notes

4.1.1 Protection against foreign particles

To protect the column against foreign particles which could have an adverse influence on the separation efficiency, we advise you to subject both the eluents and all samples to microfiltration (0.45 µm filter) and to siphon the eluent through the 6.2821.090 Aspirating filter.
To avoid contamination by abrasive particles arising from piston seals of the 709 IC Pump, it is advantageous to install an in-line filter be­tween the pump and the 733 IC Separation Center. We recommend to use either the 6.2821.100 Filter unit PEEK for the standard operation with PEEK capillaries (see section 2.3.5) or the 6.2821.000 Filter unit Manufit for the operation in the pressure range > 25.0 MPa (> 250 bar) with steel capillaries (option, see section 2.3.6).

4.1.2 Pulsation dampener

To protect the column material against pressure shocks caused by in­jection we recommend to use a pulsation dampener between the 709 IC Pump and the 733 IC Separation Center. The optional 6.2620.150 Pulsation dampener MF (see section 2.3.4) is eminently suitable for this purpose.

4.1.3 Eluents

Treatment
For the preparation of the eluents one should use chemicals of a purity degree of at least "p.a.". For dilution please use only high purity water.
Fresh eluents should always be microfiltered (0.45 µm filter) and de-
gassed (with N stirred with a magnetic stirrer, particularly when the recycling proce-
dure is employed or when alkaline eluents are used. For alkaline elu­ents and eluents with low buffering capacity one should preferably use
absorbers.
CO
2
, He or vacuum). The eluent should be continuously
2
709 IC Pump
The supply vessel containing the eluent must be closed as tightly as possible to avoid excessive evaporation. This is primarily important with eluents containing organic solvents (e.g. acetone), the evaporation of which can lead to drifts in the long term. If work is performed in a very sensitive range, even if one drop of condensate falls back in the eluent this can cause a noticeable change in the background conductivity.
25
4 Notes – Maintenance – Faults
Precipitates
Salt crystals between the piston and the seal are the cause of abrasive particles, which can enter the eluent. These lead to contaminated valves, pressure rise and in extreme cases to scratched pistons. It is thus essential to ensure that no precipitates can appear, e.g. when the eluent is changed. Solutions used in direct succession must there­fore be miscible. If the system has to be rinsed with an organic solution, several solvents with increasing or decreasing lipophilic character may possibly have to be used (e.g. water acetone chloroform).

4.2 Maintenance and servicing

4.2.1 General information

Care
The 709 IC Pump requires proper care and attention. Excessive con­tamination of the instrument could possibly lead to malfunctions and a shorter service life of the inherently rugged mechanical and electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is especially important to protect the plug connections at the rear of the instrument (particular the mains plug) against contamination.
Although constructional measures have been designed to virtually eliminate such a situation, should corrosive media penetrate the interior of the instrument the mains plug of the 709 IC Pump must be immediately disconnected to prevent extensive damage to the instrument electronics. Inform Metrohm service if your instrument has been damaged in such a way.
The instrument must not be opened by untrained personnel. Please comply with the safety notes in section 1.4.1.
Maintenance by Metrohm service
Maintenance of the 709 IC Pump is best done as part of an annual ser­vice performed by specialists from the Metrohm company. If work is frequently performed with caustic and corrosive chemicals, it may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice on the maintenance and servicing of all Metrohm instruments.
709 IC Pump
26
4.2 Maintenance and servicing

4.2.2 Maintenance work at the pump head

In many cases, an unstable baseline (pulsation, flow fluctuations) can be traced to contaminated valves or faulty, leaky piston seals. For cleaning contaminated valves and/or replacement of wear parts such as pistons, piston seals and valves, proceed as follows:
1 Detach pump head
Unscrew connection capillary 14 from the pump head 18 with the aid of the ¼" wrench (see Fig. 2). Then remove the pump head 18 from the pump housing by loosening the 4 hexagon screws 19 using the 6.2621.030 hexagon key. The main piston is on the left (when viewed from front), the auxiliary piston on the right.
2 Disassemble pump head
Strip down pump head 18 in accordance with Fig. 9. Main and auxiliary pistons are identical with the following exceptions:
The spring 63 of the auxiliary piston (right piston) is more
powerful (longer) than that of the main piston (left piston).
Inlet and outlet valve are not present in the secondary
cylinder.
61
To prevent the piston
64
, the screw 60 must be undone very carefully by hand.
suddenly jumping out of the piston cartridge
3 Cleaning/replacement of piston 61
Pistons contaminated by abrasive particles or deposits are cleaned with scouring powder and rinsed free of any particles with dist. water. Relatively badly contaminated or scratched pistons must be replaced (spare part: 6.2824.070 Zircon piston).
4 Replacement of piston seal 68
To remove damaged piston seals 68 the special tool 72 is used. This is screwed into the seal 68, which can then be pulled out (see Fig.10A).
72
When the special tool completely destroyed!
If you use only aqueous eluents the 6.2741.000 piston seal can be replaced by the 6.2741.010 PE piston seal available as an option.
is screwed into the piston seal 68 the latter is
709 IC Pump
27
4 Notes – Maintenance – Faults
71
18
70
60 61 62 63 64 65 66 68 69
67
70
: Components of the pump head
Fig. 9
A
18
72
73
68
68
B
73
C
72
Fig. 10
: Replacement of the piston seal
68
18 Pump head (6.2824.040/6.2824.100) 67 Piston guide sleeve (4.709.4370)
60 Screw for piston cartridge 64 68 Piston seal (6.2741.000/6.2741.020)
61 Piston with piston shaft (6.2824.070 zircon
piston or 6.2824.000 sapphire piston)
62 Spring retainer 70 Outlet valve (6.2824.010/6.2824.080)
63 Spring (6.2824.050) for main piston or
Spring (6.2824.060) for auxiliary piston
64 Piston cartridge (4.709.0760) 72 Special tool (6.2617.010) to remove the
65 Piston guide sleeve (4.709.4380) 73 Special tool (6.2617.010) to install the
66 Sapphire supporting ring (6.2824.030)
709 IC Pump
28
69 Inlet valve (6.2824.020/6.2824.090)
71 Screw holder for valve
piston seal 68
piston seal 68
4.2 Maintenance and servicing
To install a new piston seal
68 the special tool 73 is used:
First the new seal is inserted firmly in the recess of tool
73 by
hand (see Fig. 10B). The seal spring must be located on the outside.
The tool
head with the aid of tool
73 together with the seal is then inserted in the pump
18 and the seal pressed into the pump head recess
72 (see Fig. 10C).
The seal surface in the pump head contact with tool)!
18
must not be damaged (avoid
5 Cleaning/replacement of inlet valve 69 and outlet
valve 70
Different valves are used depending on the pump version. In the
2.709.0010 IC Pump, the 6.2824.020 Inlet valve and the 6.2824.010 Outlet valve are built-in, in the metal-free 2.709.0110 IC Pump, the
6.2824.090 Inlet valve and the 6.2824.080 Outlet valve are built-in..
Contaminated or blocked valves are cleaned by rinsing with dist. water, RBS solution or acetone. The rinsing effect can be reinforced by brief treatment in an ultrasonic bath (max. 20 s; if longer the sapphire sphere of the valve can be damaged).
If this does not have the desired effect, the valves can be disassembled as shown in Fig. 11. The valve components are pushed out with the aid of a syringe needle inserted through the upper opening in the valve housing
74. The individual compo-
nents are rinsed with dist. water and/or acetone, and the sap­phire sphere cleaned with a paper towel. The valve is then reassembled in accordance with Fig. 11. The components of the inlet and outlet valves are identical, they are distinguished only by the positioning of the sapphire sleeve holder
79 (see Fig. 11).
77 and the ceramic
Valves that fail to function faultlessly after such cleaning must be replaced.
In the reinstallation of the inlet valve
69 or the outlet valve 70 on
no account must the two outwardly identical valves be inter­changed. To determine which valve is which, note that the liquid flows through the pump head from the bottom up. The flow direction of the valves can be checked simply by blowing through the clean valve. Both valves are installed with the black face in the direction of the pump head (see Fig. 9).
709 IC Pump
29
4 Notes – Maintenance – Faults
69
If by mistake an inlet valve an extreme pressure buildup occurs within the working cylinder, which is not detected by the pressure transducer and will destroy the piston
68
seal
6 Mounting the pump head
Inlet valveOutlet valve
6.2824.020/ 6.2824.010/
6.2824.090 6.2824.080
!
The components of the pump head 18 are reassembled as shown in Fig. 9. The screw be tightened only by hand. On the other hand, the two valve screw holders then remounted on the pump (see section 3.2).
71 are tightened with a wrench. The pump head is
is installed instead of the outlet valve 70,
60 and the piston cartridge 64 may
74
75
76
77
78
79
80
Fig. 11
: Components of inlet valve
outlet valve
70
74
75
76
79
78
77
80
69 and
74 Valve housing
75 Sealing ring (black)
76 Sleeve
77 Sapphire sleeve
The bright side must point towards the sapphire sphere
78 Sapphire sphere
79 Ceramic holder for sapphire
sphere The large recess must point in the direction of the sapphire sphere
80 Seal
The larger opening must point outwards
709 IC Pump
30

4.3 Faults and malfunctions

4.3 Faults and malfunctions
If difficulties appear with the IC system during analyses, their causes are best investigated in the order separating column 709 IC Pump
eluent 732/733 IC System (see 732/733 Instructions for Use, section 5.3.2). Several of the malfunctions which may appear with the
709 IC Pump are listed in the following table with details of possible causes and countermeasures.
Malfunction Cause Rectification
Dark or undefined
display
Pump does not
respond to key­stroke
Pump drive is
switched on auto­matically after switching on the instrument without keystroke
Baseline with high noise level, pulsation
Drift of the baseline Thermal equilibrium
Battery of Zero Power
RAM discharged
Contaminated pump
values
Faulty piston seals
Quality of the pump
does not suffice for the selected sensitivity
not yet reached
Leak in system
Evaporation of organic
solvent in eluent
Inform Metrohm
service
Clean the valves (see
section 4.2.2)
Replace the piston
seals (see sec- tion 4.2.2)
Use pulsation
dampener, use more powerful pump or lower the sensitivity
Condition system with
heating switched on
Check connections
and make leakproof
Ensure better closure
of eluent supply vessel
709 IC Pump
Considerable pressure drop
Considerable pressure rise
Error message "E01" in the display
Error message "E02" in the display
Error message "E03" or "E05" in the display
Leak in system Check connections
and make leakproof
Contamination of the
filter in the 6.2821.000 Filter unit Manufit
Contamination of the
filter in the 6.2821.100 Filter unit PEEK
RS232 receive error Inform Metrohm
ROM test error Inform Metrohm
RAM test error Inform Metrohm
Clean or replace
6.2821.020 Steel mesh(es)
Replace 6.2821.110
connector with filter
service
service
service
31
4 Notes – Maintenance – Faults

4.4 Diagnosis

4.4.1 General information

The 709 IC Pump is a very precise and reliable instrument. Thanks to its rugged construction it is virtually impossible for external mechanical or electrical influences to have an adverse effect on its functions.
Although the occasional fault in the instrument can not be excluded completely, it is certainly much more likely that malfunctions are caused by wrong operation or handling or through improper connections and operation with non-Metrohm instruments.
It is thus advisable in each case to isolate the fault with the rapid and easy to perform diagnostic tests. The customer thus need not call Metrohm service until there is a true fault in the instrument. In addition, with the aid of the numbering in the diagnostic program he can provide the service engineer with much more accurate information.
In inquiries always quote the manufacturing number 39 of the 709 IC Pump (see Fig. 2), program version (see section 4.4.2) and specify possible error messages.
Procedure
The diagnostic steps must be performed in sequence and compared with the reactions of the 709 IC Pump (indented). In the "yes" case, con­tinue with the next instruction.
If the instrument does not show the expected reaction ("no" case), the appropriate diagnostic step must be repeated to exclude an operating error. With repeated wrong reactions, however, there is a strong possi­bility that a malfunction exists.
Equipment needed
3.496.8480 Test plug (Necessary only if external required functions (RS 232) should also be checked).
709 IC Pump
32
4.4 Diagnosis

4.4.2 Entry of diagnostics with automatic display test

1. Power off.
2. Disconnect all external connections (cables at rear) except mains cable.
3. Press and hold <R/S> key 3 (keep pressed) and power on.
The display 1 shows for about 3 s the switch-on test pat­tern and all 6 LEDs are switched on for visually checking.
8.8.8.8.8
The display 1 shows for about 3 s the number of the in­stalled program version.
7 0 9.1 0
4. Release <R/S> key 3.

4.4.3 Keyboard test

1. Press <R/S> 3.
2. Press <SELECT> 4.
3. Press <
4. Press <
5. Press <EXT> 8.
6. Press <PURGE> 9.
1 1
1 2
> 5.
1 3
> 6.
1 4
1 5
709 IC Pump
1 6
2 x x
(x x) = not relevant
33
4 Notes – Maintenance – Faults

4.4.4 RS232 Test

This test is meaningful only if the 709 IC Pump is used interconnected with other instruments via RS 232 connection. In addition, a
3.496.8480 Test plug normally used in the re­pair service is required for this test. However, this plug can also be purchased by customers under the above number.
For the sake of completeness, the procedure is described here.
(If a diagnostic test of the RS232 interface is not required, continue with section 4.4.5)
1. Power off.
2. Insert 3.496.8480 plug in port "RS232" 37.
3. Sliding switch 36 on rear panel to setting "RS232".
4. Press and hold <SELECT> key 4 (keep pressed) and power on.
2
5. Release <SELECT> key 4.
Connections in the
3.496.8480 plug
TxD
RxD
DCD
DTR
DSR
RTS
CTS
2 . .
2 0 0
The test runs automatically and checks several functions in succession. If one of these functions is faulty, the display "2 0 0" does not appear directly but some value is shown briefly. If, e.g., no test plug is plugged in, the following appears:
2 . .
2 0 8
2 0 0
6. Power off.
7. Remove test plug.
709 IC Pump
34
4.4 Diagnosis

4.4.5 Initialize and test RAM

On the odd occasion large disturbing signals (e.g. mains spikes, light­ning, etc.) can have an adverse effect on the processor functions and hence lead to a system crash. After such a crash the RAM area must be initialized. Although the basic instrument data remain stored, the RAM initialization should be performed only when necessary since the stored
, P
user data (flow rate, P
1. Power off.
min
etc.) are cleared as a result.
max
2. Press and hold <
3
3. Release <
> key 5.
3
RAM is tested and initialized.
> key 5 (keep pressed) and power on.
3 0 0
8.8.8.8.8
2.5 x
4. The lost data of the user must now be reentered.
The "FLOW" LED lights up.
709 IC Pump
35
4 Notes – Maintenance – Faults

4.5 Validation / GLP

The requirements of GLP (Good Laboratory Practice) include a peri- odic check of analytical measuring instruments with regard to their re­producibility and accuracy using Standard Operating Procedures, SOP).
The 709 IC Pump as part of the complete ion chromatography system, whose most important components also include separating column, 732 IC Detector, 733 IC Separation Center, and evaluation system, must be incorporated in its comprehensive validation.
Testing of the electronic and mechanical function groups of Metrohm instruments can and should be performed as part of a regular service by trained personnel of the manufacturing company (see section 4.2.1). All Metrohm instruments are equipped with start-up-test routines which check for perfect functioning of the relevant assemblies when the in­strument is switched on. If no error message is displayed, it may be as­sumed the instrument is operating without faults.
The Metrohm company also supplies its instruments with an integrated diagnostic program (see section 4.4) which, in the case of possible malfunctions or faulty behavior, allows the user to check the functioning of certain assemblies and localize the fault. Diagnostic programs can also be integrated in a validation procedure.
709 IC Pump
36
5.1 RS232 interface

5 Interfaces

5.1 RS232 interface

5.1.1 General rules for remote control

The 709 IC Pump is equipped with the comprehensive Metrohm remote control language, which allows full control over the instrument via an RS232 interface, i.e. the 709 IC Pump can receive data from an external device or send data to an external device. The 709 IC Pump sends
and LF as the terminator of a requested data block. In contrast,
2×C
R
and LF are used as the terminator of a data line. On receipt of data
C
R
from an external device, this must always close its commands with C and L as a delimiter between the individual commands.
The data are grouped logically and readily understandable. For exam­ple, to select the flow rate the command
. If more than one command is sent on a line, ‘;’ must be used
F
R
&Parameter.A.Flow"1.5"
must be sent with entry of the boldface characters sufficing, in other words
&Pa.A.F"1.5"
All quantities of the 709 IC Pump are collected in groups. The entries for the parameters, for example, are located in the group
&Parameter
The ‘Parameter’ group contains sub-groups, e.g. for setting the pa­rameter for the Metrohm pump head
&Parameter.A
This subgroup in turn contains the individual inquiries for the settings, e.g. the inquiry regarding the flow rate
&Parameter.A.Flow
or regarding the setting of the minimum shutoff pressure
&Parameter.A.Pmin
The data have a hierarchical structure (tree structure). The quantities which appear in this tree are called objects in what follows. The flow rate is that object which is called up with the command
709 IC Pump
&Parameter.A.Flow
Once you are at the desired location in the tree, you can request the value of the object:
&Parameter.A.Flow $Q Q for Query
37
5 Interfaces
The inquiry ‘$Q’ initiates the output of the value on the instrument, in other words the value output is triggered. Entries which begin with the character ‘
$’ always trigger something and are hence referred to as
triggers in what follows.
However, values of objects can not only be requested, they can also be modified. Values are always inputted in inverted commas, e.g.
&Parameter.A.Flow"1.5"

5.1.2 Call-up of objects

A section from the object tree is shown below:
Node 3
Node 2
Node 1
Flow NX
A
Parameter
M
Pump
Node 0
&
Root
The following rules apply to the call-up of objects:
Rules Examples
The root of the tree is designated by ‘&’.
For the call-up of an object, the nodes (levels) of the tree are marked by a point (.).
For the call-up of the objects, as many characters as necessary to allow unambiguous allocation of the object suffice. If the call-up is not unambiguous, the first object in the series is identified.
Call up of the minimum shutoff pressure:
&Parameter.A.Pmin
or
&Pa.A.Pmi
Uppercase and lowercase letters can be used. &PA.A.PMI or &pa.a.pmi
A value can be assigned to an object. Values are marked at both their beginning and end by inverted commas ("). They can contain maximum 8 ASCII characters. Numeric values can contain up to 3 digits and a decimal point. Numbers with more than 3 digits are not accepted. With numbers <1, leading zeros must be entered.
709 IC Pump
38
Entry of "1.5" for the flow rate:
&Parameter.A.Flow
Correct numeric entries: "38.5", "0.97"
Incorrect numeric entries: "1,5" or "+3" or ".13"
"1.5"
5.1 RS232 interface
Rules Examples
Until a new object is called up, the old object remains in force.
New objects can be addressed relative to the old object:
A leading point leads one node forwards in the tree.
More than one leading point leads one node backwards
in the tree. n nodes backwards require n+1 leading points.
To return to the root, enter a leading '&'. Jump from the node 'M' via the

5.1.3 Trigger

Triggers initiate an action at the 709 IC Pump, e.g. starting of a mode or sending of data. Triggers are marked by the introducer
The following triggers are possible:
From the root to the node 'A':
&Pa.A
Forwards from the node 'A' to 'Flow': .F
Jump from Flow onto the node
'A' and selection of a new object 'NX' at this node: ..NX
root into the node 'Pump': &Pu
'$'.
$G Go Starts processes, e.g. starting the pump
$S Stop Stops processes
$Q Query Used to request all information from the current
node in the tree forwards up to and including the values
$Q.P Path Used to request the path from the root of the tree
up to the current node
$Q.H Highest
Index
$Q.N"i" Name Used to request the name of the daughter node
$D Detail-Info Used to request detailed status information
Used to request the number of daughter nodes of the current node
with index i, i = 1...n
The triggers
'$G' and '$S' are linked to particular objects, see Over-
view table in section 5.1.5.
All other triggers can always be used at all locations in the data tree.
Examples Inquiry of the value of the pressure:
:
&Info.Actual.Pressure $Q
Start of the pump: &Pump.Start $G Inquiry of the detailed status: $D
709 IC Pump
39
5 Interfaces

5.1.4 Status messages

To ensure appropriate control of an external control unit is possible, it must also be possible to request status conditions which provide infor­mation on the status of the 709 IC Pump. The output of a status mes­sage is triggered by the trigger global status, the detailed status and eventual error messages, e.g.
$S.P.S;E32'. The global status informs on the activity of the process,
' while the detailed status conditions show the exact activity within the process.
The following global status conditions are possible:
$G Go The pump is running.
$R Ready The pump is ready.
$S Stop The pump has been stopped by "unnatural means".
Detailed status conditions
Status conditions of the global $G:
'$D'. Status messages consist of the
$G.P.S The pump has been started with normal start.
Status conditions of the global $R:
$R.P.S The pump is ready, inactive.
Status conditions of the global $S:
The instrument gives the status from which it has been
stopped, e.g. $S.P.S;E32;
709 IC Pump
40
5.1 RS232 interface

5.1.5 Error messages

Error messages ‘EXXX’ are added to the status messages and sepa­rated from them by the sign " sible:
Error Meaning Exit/Corrective action
E29 Wrong value or no value allowed. Send correct value or call up new
E30 Wrong value or no value allowed. Send correct value or call up new
;". The following error messages are pos-
object.
object.
E31 Wrong trigger, this trigger is not allowed
or carrying-out of action not possible.
E32 Pump stopped by automatic pressure
shutdown (P
E39 RS232 receive error; overflow of internal
receive buffer (>82 characters).
min
or P
violated).
max

5.1.6 RS232 error rectification

Problem Questions for corrective action
No data transmission occurs and an error message appears in the display of the 709 IC Pump.
E01: Transmission error.
Is the cable used properly wired and plugged in? Is the printer switched on and set to ”on-line”?
E01: Receive error. Are the RS232 data transmission
parameters of the third-party device and the pump the same?
Send correct trigger (exception: $D) or call up new object.
First press <EXT.> key, then <R/S> key.
First press <EXT.> key, then <R/S> key.
The received characters are garbled.
709 IC Pump
Are the RS232 data transmission parameters of the third-party
device and the pump the same?
Does the baud rate of the two devices have the same setting?
41
5 Interfaces

5.1.7 Remote control commands

Object Meaning Entry range
&
Root
├.Pump
│ ├.Start │ ├.Soft │ └.Identifier
Pump Start the pump $G, $S Soft start the pump (within 5 s) $G, $S Instrument identification up to 8 ASCII characters
├.Parameter
│ ├.A │ │ ├.Flow │ │ ├.Pmax │ │ ├.Pmin │ │ └.NX │ ├.P │ │ ├.Flow │ │ ├.Pmax │ │ ├.Pmin │ │ └.NX │ ├.M │ │ ├.Flow │ │ ├.Prnax │ │ ├.Pmin │ │ └.NX │ └.R ├.Flow ├.Pmax ├.Pmin └.NX
Parameters Metrohm pump head Flow rate 0.01…5.00 mL/min Upper shutoff pressure P Lower shutoff pressure 0.1…P Correction factor 0.90...1.10 Parameter for prep. head Flow rate read only Upper shutoff pressure read only Lower shutoff pressure read only Correction factor read only Parameter for micro head Flow rate read only Upper shutoff pressure read only Lower shutoff pressure read only Correction factor read only Parameter for reserve head Flow rate 0.01...5.00 mL/min Upper shutoff pressure 0.150.0 MPa Lower shutoff pressure 0.150.0 MPa Correction factor 0.90…1.10
50.0 MPa
min
max
MPa
├.Info
│ ├.Head │ └.Actual ├.Pressure ├.Flow ├.Pmax ├.Pmin ├.NX ├.MaxPmax ├.MinPmax ├.MaxPmin └.MinPmin
Current configuration Pump head Current data Pressure Flow rate Upper shutoff pressure Lower shutoff pressure Correction factor Maximum possible upper shutoff pressure Minimum possible upper shutoff pressure Maximum possible lower shutoff pressure Minimum possible lower shutoff pressure
└.Setup
.Trace .Autoinfo .P .R .S .E .Pressure
Instrument settings Message if values or paths change ON,OFF Automatic message on changes: when power switched on ON,OFF when instrument in "ready" status ON,OFF when instrument in "stop" status ON,OFF for error message ON,OFF on every change of measured pressure value ON,OFF
709 IC Pump
42
5.1 RS232 interface

5.1.8 Data transmission protocol

The RS232 interface of the 709 IC Pump is configured as DTE (Data Terminal Equipment) with the following technical specifications:
Standard Data interface in accordance with EIA standard
RS 232C (DIN 66020, page 1)
Baud rate 9600 (300, 600, 1200, 2400, 4800 can be set
with jumpers on main board, see section 4.1.8)
Data Bits 8
Stop Bits 1
XON/XOFF off
Control characters C
Max. line length 80 characters + C
DEC 13 HEX 0D
R
DEC 10 HEX 0A
L
F
R LF
Cable For interconnections of the 709 IC Pump with
non-Metrohm units, only a shielded data cable (e.g. Metrohm D.104.0201) may be used. The cable shielding must be faultlessly earthed at both units (pay attention to current loops; always use star-head earthing). Only connec­tors with adequate shielding may be used (e.g. Metrohm K.210.0001 with K.210.9004).
Cable length max. ca. 15 m
As only the baud rate of the RS232 parameters of the IC pump can be set by hardware, all other RS232 parameters of the third-party device must be matched to those of the pump.

5.1.9 Handshake

The handshake outputs (DTR, RTS) are set by the 709 IC Pump, the handshake inputs (CTS, DSR, DCD), on the other hand, are not tested. In practical operation with third-party devices, this has the following consequences:
XON/XOFF must be switched off in the third-party device.
With inquiries of the third-party device to the pump, the re-
sponse of the pump must be waited for before the next inquiry can be started.
709 IC Pump
43
5 Interfaces

5.1.10 Pin assignment

RS232C interface external
Transmitted Data (TxD) If no data transmission takes place, the line is maintained in the ”ON” condition. Data are sent only when CTS and DSR are in the ”ON” condition and DCD is in the ”OFF” condition.
Pin 2
Transmitted Data
Received Data (RxD) Data are received only when DCD is ”ON”.
Request to Send (RTS) ON condition: 709 IC Pump is ready to send data.
Clear to Send (CTS) ON condition: Remote station is ready to receive data.
Data Set Ready (DSR) ON condition: The transmission line is connected.
Signal Ground (GND)
Data Carrier Detect (DCD)
ON condition: The received signal level is within the tolerance range (remote station is ready to send data).
Pin 3
Received Data
Pin 4
Request to Send
Pin 5
Clear to Send
Pin 6
Data Set Ready
Pin 7
Signal Ground
Pin 8
Data Carrier Detect
Data Terminal Ready (DTR) ON condition: 709 IC Pump is ready to receive data.
Protective earth
Direct connection from cable connector to protective earth of the device.
Contact arrangement at the connector (female) for socket "RS 232" (male)
1 13
Polarity assignment of the signals
Data lines (TxD, RxD) Voltage negative (<–3 V): signal status ”ONE” Voltage positive (>+3 V): signal status ”ZERO”
Control or message lines (CTS, DSR, DCD, RTS, DTR) Voltage negative (<–3 V): OFF status Voltage positive (>+3 V): ON status
In the transition region from +3 V to –3 V, the signal status is undefined.
Driver 14C88 to EIA RS 232C specification
Connection cables to external devices must have an appropriate 25-pin connector (female). Metrohm offers connector sockets (K.210.9004) and the associated housing (K.210.0001) as an option for the preparation of such cables.
14 25
Receiver 14C89 to EIA RS 232C specification
No liability whatsoever will be accepted for damage arising from the improper connection of devices.
Pin 20
Data Terminal Ready
709 IC Pump
44
5.1 RS232 interface
2P2P2

5.1.11 Setting the baud rate

The baud rate of the RS232 interface of the 709 IC Pump is set to 9600 as standard. If you wish to set a lower baud rate, proceed as follows:
The electronic circuitry of the 709 IC Pump is sensitive to electrostatic charges. You should thus take the appropriate measures (earthing, e.g. by touching metal) before opening the instrument.
1 Disconnect cables
Disconnect mains cable from mains connection plug 31 of the 709 IC Pump, also unplug all other cables connected to the rear of the instrument.
2 Remove housing cover
Unscrew the two fastening screws 25 at the bottom of the instrument and the two fastening screws 26 at the rear of the instrument, remove housing cover 27.
3 Set baud rate
Change over blue jumpers at positions P21, P22 and P23 of the main board (on right side of instrument when viewed from back) using tweezers or flat-nosed pliers to match the baud rate you wish:
321
P21
P22
P23
321
P21
P22
P23
321
P21
P22
P23
321
321
P21
P22
P23
9600 4800 2400 1200 600 300
4 Replace housing cover
Replace housing cover 27 and fasten with fastening screws 25 and 26.
5 Plug in cables
Connect mains cable to mains connection plug 31 and also plug in all other cables.
P21
P22
P23
321
P
709 IC Pump
45
5 Interfaces
g

5.2 Control interface

5.2.1 Functions

The 709 IC Pump has the following control inputs and outputs, which are accessible via the control interface:
Inputs Reduce flow rate (possible via contact or TTL signal)
Stop pump (possible via contact or TTL signal)
Outputs Signal when pump has been stopped or
Signal when pump is running
You will find detailed information on these functions in section 5.2.3, possible changeover of the jumpers is described in section 5.2.2.

5.2.2 Settings for the control interface

If you wish to change the default settings for the control inputs and out­puts (see section 5.2.3), proceed as follows:
The electronic circuitry of the 709 IC Pump is sensitive to electrostatic charges. You should thus take the appropriate measures (earthing, e.g. by touching metal) before opening the instrument.
1 Disconnect cables
Disconnect mains cable from mains connection plug 31 of the 709 IC Pump, also unplug all other cables connected to the rear of the instrument.
2 Remove housing cover
Unscrew the two fastening screws 25 at the bottom of the instrument and the two fastening screws 26 at the rear of the instrument, remove housing cover 27.
3 Change settings
Change over blue jumpers at positions P12, P27, P28, P29 and P30 of the main board (on left side of instrument when viewed from back) using tweezers or flat-nosed pliers to match the settings you wish (settings, see section 5.2.3):
P13 P12
1 2 3
P27 P28 P29 P30
Default settin
s
4 Replace housing cover
Replace housing cover 27 and fasten with fastening screws 25 and 26.
5 Plug in cables
Connect mains cable to mains connection plug 31 and also plug in all other cables.
709 IC Pump
46
5.2 Control interface

5.2.3 Control inputs and outputs

709 IC Pump external Function
Inputs
+5V
1k
P27.3
1
P27.2 P27.1
Reduced flow rate (Reduced Flow) with contact
(default settings)
Contact open: Flow rate 100%
P28.3
Contact closed: Flow rate 10%
With reduced flow, the "FLOW" LED 10
0V
P28.2 P28.1
2
flashes.
+5V
1k
P27.3
P27.2 P27.1
1
TTL
Reduced flow rate (Reduced Flow) with TTL logic signal
(Position of jumpers P27 and P28
P28.3
must be changed, see
Signal Low (L): Flow rate 100%
section 5.2.2)
Signal High (H): Flow rate 10%
With reduced flow, the "FLOW" LED 10
P28.2 P28.1
2
0V
flashes.
+5V
1k
P29.3
7
P29.2 P29.1
Stop pump with contact
(default setting)
Contact closed: Pump is
stopped
P30.3
Following a stop, the pump must be restarted manually with the <R/S>
0V
P30.2 P30.1
8
key
.
+5V
1k
P29.3
7
Stop pump with TTL logic signal
P29.2 P29.1
TTL
(Position of jumpers P29 and P30 must be changed, see
section 5.2.2)
Signal High (H): Pump is
P30.3
stopped
Following a stop, the pump must be restarted manually with the <R/S>
P30.2 P30.1
8
0V
key
.
709 IC Pump
47
5 Interfaces
709 IC Pump external Function
Outputs
3
P12.3
P12.2 P12.1
P12.3
P12.2 P12.1
Pump stopped
(default setting)
Contact open: Pump stopped Contact closed: Pump running
4
3
4
5
6
Pump stopped
(Position of jumper P12 must be changed, see
Contact open: Pump stopped Contact closed: Pump running
Display of suction phase
This function can not be used with the 709 IC Pump.
section 5.2.2)
9
Contact arrangement at connector (male) for socket "Control" (female)
5 1
96
No liability whatsoever will be accepted for damage arising from the improper connection of devices.
Connection not in use
709 IC Pump
48
6.1 Technical specifications

6 Appendix

6.1 Technical specifications

Type Serial dual piston with two valves
Pump capacity
Flow range 0.05...5.00 mL/min
Flow accuracy < 3.5 % (at 150 bar and 1 mL/min)
Flow constancy < 0.5 % (after unit warmed up)
Reproducibility of eluent flow
Pressure
Pressure range 2.709.0010: 0...50.0 MPa (0...500
Resolution for pressure indicator
Pressure measurement Piezoresistive measurement principle
Accuracy of pressure measurement
Residual pulsation 2.709.0010: < 7 %
Recorder output for pressure indication
± 0.1 %
bar)
2.709.0110 (metal-free): 0...30.0 MPa (0...300 bar)
0.1 MPa (1 bar)
Response time: 3 ms Measurement volume: ca. 50 µL
< 3 %
2.709.0110 (metal-free): < 10 % (measured at pump head at 2 mL/min and 100 bar)
1 mV/MPa (1 mV/10 bar) Time constant switchable: HIGH = 20 ms
LOW = 1 s
709 IC Pump
Safety shutdown
Function Automatic shutdown when upper and lower
pressure limits violated
Maximum pressure limit Adjustable between 0.1...50.0 MPa (1...500 bar)
Response time: 1 pump cycle
Minimum pressure limit Adjustable between 0.1 ... 50.0 MPa (1...500 bar),
inactive at 0 bar Response time: 3 pump cycles The cutoff function first becomes active 2 min after the pump flow has been switched on with the <R/S> key.
Pump head
Pump head volumes Main piston: 40 µL
Priming piston: 20 µL
49
6 Appendix
Pump displacement volumes Main piston: 28.5 µL
Priming piston: 14.25 µL
Length of stroke Main piston: 3.6 mm
Priming piston: 1.8 mm
Display
Type 5 digit LED display each with 7 segments
6 LED indicators
Displayed parameters Flow rate (in mL/min)
Pressure (in MPa)
Function keys
Number 6
Identification by LED indicators
Programming
Keyboard Flow rate
Maximum and minimum pressure limit values Correction factor for pump head
Remote control Flow rate
Maximum and minimum pressure limit values Correction factor for pump head Start/Stop
Remote control
Interfaces RS232 (syntax Metrohm, see section 5.1)
Current Loop (details on request) Control (see section 5.2)
Storage of entries
Type Non-volatile (RAM)
Duration max. 10 years (lithium battery)
RS232 interface
Standard EIA RS232C, CCITT V.24, ISO 2110, DIN 66020
Plug D-Sub 25 pins
Mode Full duplex
(simultaneous sending and receiving)
Parameters Baud rate: 300,600,1200,2400,4800,9600
(adjustable by hardware) Handshake: no Handshake (see section 5.1.9) Parity: none Stop bits: 1 Data bits: 8
Mains connection
Voltage 115 V: 100...120 V ± 10 %
230 V: 220...240 V ± 10 % switchable
Frequency 50...60 Hz
709 IC Pump
50
6.1 Technical specifications
Power consumption ca. 50 VA
Fuses Diameter = 5 mm, length = 20 mm
100120 V: 0.5 A (slow-blow) 220240 V: 0.25 A (slow-blow)
Safety specifications
Construction / Testing According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions This Instructions for Use include information and
warnings which must be heeded by the user to assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference Standards met:
EN55011 (class B), EN55022 (class B), EN50081-1/2
Immunity to interference Standards met:
IEC801-2/IEC1000-4-2 (class 2), IEC801-3/ IEC1000-4-3 (class 3), IEC801-4/IEC1000-4-4 (class 3), IEC801-5/IEC1000-4-5 (class 2/3), IEC801-6/IEC1000-4-6 (class 3), EN50082-2
Ambient temperature
Nominal operating range +5+45°C
(at 585 % atmospheric humidity)
Storage, transport –40+70°C
Diagnosis
Self-diagnostic test Automatic self-diagnosis when instrument
switched on
User diagnostic test Built-in diagnostic program (see section 4.4)
Materials (in contact with solvent)
2.709.0010 V4A stainless steel, sapphire, ruby, ceramic, PCTFE, PTFE
2.709.0110 (metal-free) Sapphire, ruby, ceramic, PCTFE, PTFE
Housing
Material of cover Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, FCH-free
Material of base Steel, enamelled
709 IC Pump
Width 255 mm
Height 130 mm
Depth 381 mm
Weight 10 kg (incl. accessories)
51
6 Appendix

6.2 Scope of delivery

Subject to changes ! All dimensions are given in mm.
The 709 IC Pump is available in the two following versions:
2.709.0010 709 IC Pump (standard version)
2.709.0110 709 IC Pump (metal-free version)
The two instruments include the following parts:
Quant. Order No. Description
2.709.0010 2.709.0110
1 1 6.1831.010 PEEK capillary
Length = 3 m
1 1 6.1834.000 Aspirating tubing
made of PTFE, with connector for
6.2821.090 Aspirating filter
Length = 1.2 m
For the connection 709 IC Pump – eluent container
1 1 6.2115.040 Connection cable
For the connection 709 IC Pump – recorder
For plotting curves delivery vs time.
1 1 6.2122.0X0 Mains cable
according to customer's specification:
Cable socket
Type IEC 320/C 13 Type SEV 12 (CH…) ............................... 6.2122.020
Type IEC 320/C 13 Type CEE (7), VII (D…) ........................... 6.2122.040
Type CEE (22), V Type NEMA 5-15 (USA…) ...................... 6.2122.070
Cable plug
1.5 m
0.25
1
16"
2.51.5
1 1 6.2617.010 Special tool
For removing the piston seal of the pump head
35 50
1 1 6.2621.000 Adjustable spanner
max.
20
150
709 IC Pump
52
M3.5
6.2 Scope of delivery
Quant. Order No. Description
2.709.0010 2.709.0110
1 1 6.2621.030 Hexagon key 4 mm
For mounting the pump head of the 709 IC Pump.
1 1 6.2744.010 PEEK compression fitting
For the connection of 6.1831.010 PEEK capillaries or 6.1822.010 PTFE microcapillaries Set of 5
1 1 6.2816.020 Syringe
made of PP, volume = 10 mL
with Luer connector
1 1 6.2816.040 Tubule
with PTFE tubing and Luer connection
For 6.2816.020 syringe
73
102
120
29.5
26
1 1 6.2821.090 Aspirating filter
Pore dimension 20 µm
For 6.1834.000 Aspirating tubing.
Set of 5
1 1 6.2821.100 Filter unit PEEK 2 µm
To avoid contamination due to abrasive particles of piston seals
Spare part: 6.2821.110 Filter
1 - 6.2824.040 Pump head (standard version)
Complete, with fixation screws
- 1 6.2824.100 Pump head (metal-free)
Complete, with fixation screws
1 1 8.709.1033 Instructions for Use (English)
for 709 IC Pump
9
16
35
25.5
709 IC Pump
53
6 Appendix

6.3 Optional accessories

Order No. Description
6.1834.010 Aspirating tubing
made of PTFE, with connector for 6.2821.090 Aspirating filter
Length = 2.5 m
For the connection 709 IC Pump – eluent vessel of 6.5324.000 Bottle rack
6.2620.000 Pressure screw
made of steel For the connection of 6.2620.020 steel capillaries, set of 5
10.8
2.51.5
6.2620.010 Ferrule
made of steel For the connection of 6.2620.020 steel capillaries, set of 5
6.2620.020 Steel capillary
Length = 3 m
6.2620.150 Pulsation dampener MF
Metal-free pulsation dampener for reduction of pulsation and preservation of separating columns.
6.2621.040 Capillary tubing cutter for 6.2620.020 steel capillaries
6.2621.080 Capillary tubing cutter
for 6.1831.010 PEEK capillaries and
6.1822.010 PTFE micro capillaries
incl. 5 additional blades
3.7
3.9
52
76
55
118
6.2741.000 Piston seal
Spare part for 6.2824.040 and 6.2824.100 pump heads
6.5
709 IC Pump
54
3
6.3 Optional accessories
Order No. Description
6.2741.010 Piston seal PE
Spare part for 6.2824.040 and 6.2824.100 pump heads (only suitable for aqueous eluents)
6.2744.070 PEEK compression fitting (short)
Spare part for 6.2824.100 pump head Set of 5
6.2821.000 Filter unit Manufit
To avoid contamination due to abrasive particles of piston seals.
For the pressure range 050 MPa with steel capillaries.
6.2821.110 Filter for Filter unit PEEK
Spare part for 6.2821.100 Filter unit PEEK Set of 10
6.2824.070 Zircon piston
Spare part for 6.2824.040 and
6.2824.100 pump heads
6.2824.010 Outlet valve
Spare part for 6.2824.040 pump head
6.5
350 35031
11
65.5
21
3
13.5
3
8
6.2824.020 Inlet valve
Spare part for 6.2824.040 pump head
6.2824.030 Sapphire supporting ring
Spare part for 6.2824.040 and
6.2824.100 pump heads
6.2824.050 Spring for main piston
Spare part for 6.2824.040 and
6.2824.100 pump heads
6.2824.060 Spring for auxiliary piston
Spare part for 6.2824.040 and
6.2824.100 pump heads
16
16
3.2
6.3
8
1.3
11
30
11
47
709 IC Pump
55
6 Appendix
Order No. Description
6.2824.080 Outlet valve (metal-free)
Spare part for 6.2824.100 pump head
6.2824.090 Inlet valve (metal-free)
Spare part for 6.2824.100 pump head
6.5324.000 Bottle rack
For fixing 3 supply bottles for elent, regenera­tion and rinsing solution, incl. accessories (bottles, siphon, etc.)
8
16
8
16
709 IC Pump
56

6.4 Warranty and conformity

6.4 Warranty and conformity

6.4.1 Warranty

The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error which occur within 12 months from the day of delivery. In this case, the defects will be recti­fied in our workshops free of charge. Transport costs are to be paid by the customer.
For day and night operation, the warranty is limited to 6 months.
Glass breakage in the case of electrodes or other parts is not covered by the warranty. Checks which are not a result of material or manufac­turing faults are also charged during the warranty period. For parts of outside manufacture insofar as these constitute an appreciable part of our instrument, the warranty stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifica­tions in the instruction manual are authoritative.
Concerning defects in material, construction or design as well as the absence of guaranteed features, the orderer has no rights or claims ex­cept those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if the goods show signs of transport damage after unpacking, the carrier must be informed immediately and a written damage report demanded. lack of an official damage report releases Metrohm from any liability to pay compensation.
If any instruments and parts have to be returned, the original packaging should be used if at all possible. This applies above all to instruments, electrodes, burette cylinders and PTFE pistons. Before embedment in wood shavings or similar material, the parts must be packed in a dust­proof package (for instruments, use of a plastic bag is imperative). If open assemblies are enclosed in the scope of delivery that are sensi­tive to electromagnetic voltages (e.g. data interfaces etc.) these must be returned in the associated original protective packaging (e.g. con­ductive protective bag). (Exception: assemblies with built-in voltage source belong in a non-conductive protective packaging).
No warranty responsibility whatsoever will be accepted by Metrohm for damage which arises as a result of non-compliance with these instruc­tions.
709 IC Pump
57
6 Appendix

6.4.2 EU Declaration of conformity

EU Declaration of Conformity
The Metrohm AG company, Herisau, Switzerland hereby certifies that the instru­ment:
709 IC Pump
meets the requirements of EC Directives 89/336/EWG and 73/23/EWG.
Source of the specifications
EN 50081-1/2 Electromagnetic compatibility, basic specification
Emitted Interference
EN 50082-2 Electromagnetic compatibility, basic specification
Interference Immunity
EN 61010 Safety requirements for electrical laboratory measurement
and control equipment
Description of the instrument
Dual piston pump for use in Ion Chromatography (IC) and
High Performance Liquid Chromatography (HPLC).
Herisau, December 4, 1995
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
709 IC Pump
58
6.4 Warranty and conformity

6.4.3 Certificate of conformity and system validation

Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appli­ances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
Name of commodity: 709 IC Pump
System software: Stored in ROMs
Name of manufacturer: Metrohm Ltd., Herisau, Switzerland
Principal technical information: Voltages: 100…120, 220…240 V Frequency: 5060 Hz
This Metrohm instrument has been built and has undergone final type testing according to the standards:
IEC801-2/IEC1000-4-2 (class 2), IEC801-3/ IEC1000-4-3 (class 3), IEC801-4/IEC1000-4-4 (class 3), IEC801-5/IEC1000-4-5 (class 2/3), IEC801-6/IEC1000-4-6 (class 3), EN50082-2, EN55011 (class B), EN55022 (class B), EN50081-1/2 — Electromagnetic compatibility
IEC1010, EN61010, UL3101-1 — Security specifications
It has also been certified by the Swiss Electrotechnical Association (SEV), which is member of the International Certification Body (CB/IEC).
The technical specifications are documented in the instruction manual.
The system software, stored in Read Only Memories (ROMs) has been validated in connection with standard operating procedures in respect to functionality and performance. The features of the system software are documented in the instruc­tion manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servicing.
Herisau, December 4, 1995
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
709 IC Pump
59
6 Appendix

6.5 Index

A
Abrasive particles.................................. 26
Adjustable spanner (6.2621.000) ......... 52
Ambient temperature ............................ 51
Appendix ............................................... 49
Arrangement of the instruments ............. 7
Aspirating capillary 15
Figure .................................................. 3
Mounting the suction tubing............. 15
Aspirating filter (6.2821.090)
Mounting ........................................... 15
Ordering designation ........................ 53
Practical notes ..................................25
Aspirating tubing (6.1834.000)
Mounting ........................................... 15
Ordering designation ........................ 52
Aspirating tubing (6.1834.010)
Mounting ........................................... 15
Ordering designation ........................ 54
Attaching the accessories ......................9
Attachment 23
Figure .................................................. 3
B
Baud rate.......................................... 43,45
Bottle rack 6.5324.000
Mounting ........................................... 15
Ordering designation ........................ 56
C
Cable (6.2115.040) ...............................52
Cable (6.2125.060) ...............................15
Call-up of objects.................................. 38
Capillaries ............................................... 9
Capillary 43
Figure ................................... 9,11,13,14
Capillary tubing
cutter (6.2621.040)......................... 9,54
Capillary tubing
cutter (6.2621.080)......................... 9,54
Care....................................................... 26
Caution....................................................5
Ceramic holder 78
Figure ................................................ 30
Certificate of conformity
and system validation ....................... 59
Comment ................................................5
Compression fitting 44
Figure ............................................. 9,11
Mounting ............................................. 9
Ordering designation ........................ 53
Connecting the tubing ..........................13
Connection 23
Install connection capillary 22 ..........13
Install filter unit Manufit 58 ................ 14
Connection at 732 IC Detector ............. 15
Connection capillary 14
Detaching ......................................... 27
Figure.................................................. 3
Fixing at the pump head..................... 8
Connection capillary 17
Figure.................................................. 3
Connection capillary 22
Figure............................................. 3,13
Mounting........................................... 13
Connector 45
Connection ....................................... 13
Figure................................................ 11
Mounting........................................... 11
Ordering designation........................ 55
Connector 47
Connection ....................................... 13
Figure................................................ 11
Mounting........................................... 11
Connectors for capillaries....................... 9
Control characters ................................ 43
Control inputs ....................................... 47
Control interface 33
Figure.................................................. 3
Functions ..........................................46
Settings ............................................. 46
Control outputs ..................................... 48
Correction factor ................................... 20
Counterpart end 50
Figure................................................ 12
Mounting........................................... 12
Coupling 59
Figure................................................ 14
Mounting........................................... 14
Current Loop interface 38
Figure.................................................. 3
Switching .......................................... 23
Current pressure................................... 23
D
Data transmission protocol .................. 43
Deaerating the pump............................ 18
Degassing of the eluent........................ 25
Degree of protection............................... 6
Diagnosis .............................................. 32
Dimensions ........................................... 51
Display 1
Figure.................................................. 3
Technical specifications ................... 50
Test ................................................... 33
Drift........................................................ 25
E
Earthing.............................................. 6,17
Electrical connection ............................ 15
Electrical safety....................................... 6
Electromagnetic compatibility .............. 51
Eluent container .................................... 15
Eluents ..................................................25
EMC ......................................................51
Emitted interference.............................. 51
Error messages ............................... 32,41
EU Declaration of conformity ............... 58
Evaporation .......................................... 25
External control
Switching off ..................................... 23
Switching on ..................................... 23
F
Fastening screws 19
Detach pump head .......................... 27
Figure.................................................. 3
Mounting the pump head................... 8
Faults .......................................... 25,31,32
Ferrule 41
Figure.................................................. 9
Mounting........................................... 10
Ordering designation ....................... 54
Filter unit Manufit 58
Connection ....................................... 14
Figure........................................... 12,14
Mounting........................................... 12
Ordering designation ....................... 55
Filter unit PEEK 56
Connection ....................................... 13
Figure........................................... 11,13
Mounting........................................... 11
Ordering designation ....................... 53
Flow accuracy ...................................... 49
Flow constancy..................................... 49
Flow correction ..................................... 20
Flow range............................................ 49
Flow rate ............................................... 20
Front........................................................ 2
Fuse cover 30
Figure.................................................. 3
Open................................................. 16
Fuse data 28
Exchange of fuses............................ 16
Figure.................................................. 3
Fuses
Exchange.......................................... 16
Technical specifications ................... 51
G
General information
on maintenance................................ 26
General safety rules................................ 6
GLP....................................................... 36
H
Handling of solvents............................... 6
Handshake ........................................... 43
Hazard .................................................... 5
Hexagon key (6.2621.030) .............. 27,53
Housing ................................................ 51
Housing 46 for filter unit
Figure................................................ 11
Mounting........................................... 11
Housing cover 27
Figure.................................................. 3
Remove ....................................... 45,46
709 IC Pump
60
6.5 Index
I
Immunity to interference....................... 51
Information about the
Instructions for Use .............................4
Inlet capillary 48
Connection ....................................... 14
Figure........................................... 12,14
Mounting........................................... 12
Inlet valve 68
Cleaning ........................................... 29
Components..................................... 30
Figure........................................... 28,30
Ordering designations................. 55,56
Replacement .................................... 29
Inlet valve screw connection 16
Figure...................................................3
Inquiry of the current pressure ............. 23
Installation................................................7
Installing the filter unit Manufit.............. 12
Installing the filter unit PEEK................. 11
Instructions for Use (8.709.1033)...... 4,53
Instrument data 40
Figure...................................................3
Instrument description.............................1
Interfaces .............................................. 37
Introduction..............................................1
K
Key < > 5
Figure...................................................3
Function............................................ 24
Key < > 6
Figure...................................................3
Function............................................ 24
Key <EXT.> 8
Figure...................................................3
Function............................................ 24
Switching on/off external control...... 23
Key <PURGE> 9
Figure...................................................3
Function............................................ 24
Key <R/S> 3
Figure...................................................3
Function............................................ 23
Switch pump drive on/off ................. 22
Key <SELECT> 4
Display of actual pressure................ 23
Display of flow rate ........................... 20
Display of P
Display of P
Figure...................................................3
Flow correction ................................. 20
Function............................................ 23
Key functions........................................ 23
Keyboard test ....................................... 33
................................. 22
max
.................................. 21
min
L
Leaks .......................................................6
LED "Ext." 7
Display of external control................ 23
Figure...................................................3
LED "FLOW" 10
Display of flow rate ........................... 20
Figure...................................................3
LED "P
Display of actual pressure ................23
Figure ..................................................3
LED "P
Display of P
Figure ..................................................3
LED "P
Display of P
Figure ..................................................3
LED "R/S" 2
Display of pump drive .......................22
Figure ..................................................3
Location................................................... 7
actual
max
min
" 13
" 12
..................................22
max
" 11
...................................21
min
M
Mains cable
Ordering designation ........................52
Plug ...................................................17
Mains connection....................................6
Procedure..........................................16
Technical specifications.................... 50
Mains connection plug 31
Figure ..................................................3
Mains connection..............................17
Mains frequency.................................... 50
Mains switch 29
Figure ..................................................3
On/off switching of
the instrument ...................................17
Mains voltage
Setting ...............................................16
Technical specifications.................... 50
Mains voltage selector 32
Figure ..................................................3
Setting the mains voltage .................16
Maintenance..........................................26
Malfunctions .....................................31,32
Manufacturing number 39
Figure ..................................................3
Inquiries.............................................32
Manufit filter unit ...... see Filter unit Manufit
Manufit housing 54
Figure ................................................12
Mounting ...........................................12
Manufit pressure screw 49
Figure ................................................12
Mounting ...........................................12
Materials ................................................51
Maximum pressure limit ...................22,49
Metrohm service...............................26,32
Microfiltration.........................................25
Minimum pressure limit....................21,49
Mounting the pump head .......................8
N
Noise .....................................................31
Notation ...................................................5
Notes .....................................................25
O
Object....................................................37
On/off switching of the external control 23
Open-end spanner (6.2621.010) ..........10
Operation ..............................................19
Optional accessories ............................ 54
Ordering designations ..........................52
Organization............................................ 4
Outlet capillary 20
Deaerating the pump........................18
Figure .................................................. 3
Outlet capillary 55
Connection........................................ 14
Figure ........................................... 12,14
Mounting ........................................... 12
Outlet valve 69
Cleaning ............................................ 29
Components .....................................30
Figure ........................................... 28,30
Ordering designations ................. 55,56
Replacement.....................................29
Outlet valve screw connection 21
Figure .................................................. 3
P
Packaging ...............................................7
Parts and controls...................................2
PEEK capillary (6.1831.010) .............. 9,52
PEEK compression fitting (6.2744.070)
Ordering designation ........................ 55
PEEK connectors .................................... 9
Pictograms .............................................. 5
Piston 61
Cleaning ............................................ 27
Figure ................................................ 28
Replacement.....................................27
Piston cartridge 64
Figure ................................................ 28
Tighten ..............................................30
Piston guide sleeve 65
Figure ................................................ 28
Piston seal 67
Figure ................................................ 28
Ordering designations ................. 54,55
Replacement.....................................27
Power consumption .............................. 50
Practical notes ......................................25
Precipitates ...........................................26
Pressure ................................................ 49
Pressure limit values
Maximum pressure limit.................... 22
Minimum pressure limit..................... 21
Pressure measurement......................... 49
Pressure range.............................. 8,22,49
Pressure screw 42
Figure .................................................. 9
Mounting ........................................... 10
Ordering designation ........................ 54
Program version.................................... 33
Protection against foreign particles ...... 25
Protection class....................................... 6
PTFE gasket 51
Figure ................................................ 12
Mounting ........................................... 12
PTFE Microcapillary (6.1822.010)........... 9
Pulsation................................................ 31
Pulsation dampener 57
Connection.............................. 10,13,14
Figure ........................................... 13,14
Ordering designation ........................ 54
Practical notes ..................................25
Pump capacity ...................................... 49
709 IC Pump
61
6 Appendix
Pump head 18
Components .....................................28
Disassemble .....................................27
Figure ............................................. 3,28
Maintenance .....................................27
Mounting ........................................ 8,30
Ordering designation ........................ 53
Technical specifications ................... 49
Pump stopped ......................................48
Purge valve 24
Connection to the pump head ........... 8
Deaerating the pump........................18
Figure .................................................. 3
Switching on/off ................................24
R
RAM initialization................................... 35
RAM test................................................ 35
Rear......................................................... 2
Recorder output socket 34
Figure .................................................. 3
Technical data................................... 49
Recycling............................................... 25
Reduced flow rate................................. 47
Remote control...................................... 50
Remote control commands .................. 42
Remote control language .....................37
Remote control via RS232 .................... 37
Residual pulsation ................................49
RS232 interface 37
Baud rate...........................................43
Data transmission protocol .............. 43
Error messages................................. 41
Error rectification............................... 41
Figure .................................................. 3
Handshake........................................ 43
Pin assignment .................................44
Remote control commands.............. 42
Status messages ..............................40
Switching........................................... 23
Technical specifications ................... 50
Trigger............................................... 39
RS232 Test............................................ 34
Rules for remote control language ....... 38
S
Safety notes ............................................6
Safety rules.............................................. 6
Safety shutdown ..............................21,49
Safety specifications ............................. 51
Sapphire sleeve 76
Figure ................................................ 30
Sapphire sphere 77
Figure ................................................ 30
Sapphire supporting ring66
Figure ................................................ 28
Ordering designation ........................ 55
Scope of delivery ..................................52
Screw 25
Figure .................................................. 3
Remove housing cover................45,46
Screw 26
Figure .................................................. 3
Remove housing cover................45,46
Screw 60
Figure................................................ 28
Tighten .............................................. 30
Undo .................................................27
Screw holder 70
Figure................................................ 28
Seal 79
Figure................................................ 30
Sealing ring 74
Figure................................................ 30
Sequence of operations ....................... 19
Service .................................................. 36
Servicing ............................................... 26
Setting the baud rate ............................ 45
Setting the flow rate .............................. 20
Setting up the pump............................... 7
Shutoff pressure
.......................see Pressure limit values
Sleeve 75
Figure................................................ 30
Sliding switch 35
Figure.................................................. 3
Sliding switch 36
Figure.................................................. 3
Switching .......................................... 23
SOP....................................................... 36
Special tool 71
Figure................................................ 28
Ordering designation........................ 52
Remove piston seal 67..................... 27
Special tool 72
Figure................................................ 28
Install piston seal 67......................... 29
Ordering designation........................ 52
Spilled chemicals.................................. 26
Spring 63
Figure................................................ 28
Ordering designations................. 55,56
Spring retainer 62
Figure................................................ 28
Srew holder 70
Tighten .............................................. 30
Standard Operating Procedures .......... 36
Static charges......................................... 6
Status messages .................................. 40
Steel capillary (6.2620.020) .................... 9
Steel capillary (6.2620.020) .................. 54
Steel connectors................................... 10
Steel mesh holding end 53
Figure................................................ 12
Mounting........................................... 12
Steel meshes 52
Figure................................................ 12
Mounting........................................... 12
Stirring of the eluent.............................. 25
Stop pump ............................................ 47
Storage ................................................. 51
Sunlight ................................................... 7
Switching on/off the instrument............ 17
Switching the pump drive on/off .......... 22
Syringe (6.2816.020) ............................ 18
Ordering designation........................ 53
T
Technical specifications ....................... 49
Test plug (3.496.8480) .................... 32,34
Transport ................................................ 7
Transport damage................................ 57
Transport safeguard block ..................... 8
Treatment of eluents............................. 25
Trigger.............................................. 38,39
Tubing connections
Connection to eluent container ........ 15
Connection to injection valve ...... 13,14
Tubule (6.2816.040)
Ordering designation ....................... 53
V
Validation .............................................. 36
Valve housing 73
Figure................................................ 30
W
Warning................................................... 5
Warranty................................................ 57
Z
Zircon piston (6.2824.070) .............. 27,55
709 IC Pump
62
Loading...