•Before using this product, read and understand the instructions.
•Save these instructions for future reference.
•All work must be performed by qualified personnel trained in the proper application, installation, and
maintenance of equipment and/or systems in accordance with all applicable codes and ordinances.
•Failure to follow the instructions set forth in this publication could result in property damage, personal injury,
or death from fire and/or explosion, or other hazards that may be associated with this type of equipment.
!!
!!
!
WARNINGWARNING
WARNINGWARNING
WARNING
DescriptionPage Number
LCR-II System Components4
LCR-II Specifications5
Lap Pad Operation6-8
Lap Pad Overview6
Navigating Top-Level Menus7
Secondary Data Screens8
Lap Pad Calibration & Set-Up9-26
Before Getting Started9
Removing the LCR-II Switchplate9
Security10
General Setup10-13
System Calibration14-16
Product & Shift Information17-18
Product Calibration19-23
Multi-Point Calibration24-26
LCR-II Menu Flow Charts38-39
Weights & Measures Requirements40
LCR-II Troubleshooting Guide40-44
Appendix A - Compensation Types &
Parameters45
Appendix B - VT100 Emulation Codes46
Appendix C - Error Messages47-50
Appendix D - Lap Pad Menu Flow
Charts51-57
Appendix E - W & M Prover Guide58-59
Publication Updates and Translations
The most current English versions of all Liquid Controls publications are available on our website, www.lcmeter.com.
It is the responsibility of the Local Distributor to provide the most current version of LC Manuals, Instructions, and
Specification Sheets in the required language of the country, or the language of the end user to which the product s are
shipping . If there are questions about the language of any LC Manuals, Instructions, or Specification Sheets, please
contact your Local Distributor.
Power, input and output (I/O) wiring must be in
accordance with the area classification for which it is
used (Class I, Div 2). For North America, inst allations
must be per the U. S. National Electrical Code, NFPA
70, or the Canadian Electrical Code in order to maintain
Class I, Division 2 ratings. This may require using
connections or other adaptations in accordance with
the requirements of the authority having jurisdiction.
WARNING: Explosion Hazard - Substitution of
components may impair suitability for Class I, Division
2 applications.
WARNING: Explosion Hazard - When in hazardous
locations, turn power OFF before replacing or wiring
modules.
WARNING: Explosion Hazard - Do NOT disconnect
equipment unless power has been switched OFF or
the area is known to be Non-Hazardous.
2
Software License Agreement
Read this license carefully. You agree that by using the itemized software package, you have agreed to the software
license terms and conditions. This agreement constitutes complete and entire agreement between you and Liquid
Controls with respect to this product.
1.Liquid Controls hereby grants to Licensee a non
exclusive license to use SR260 (hereinafter referred to
as “Licensed Software”)
2.Under the License granted herein, Licensee may use
the itemized machine readable (executable code) copy
of the Software, including any subsequent updates
which may be provided. Licensee shall not, without
Liquid Controls’ prior written consent, (a) rent, lease,
lend, sublease or otherwise transfer the materials
hereunder; (b) remove or obscure proprietary or
copyright notices which may be set forth on the
Licensed Software; or (c) alter, decompile, or
disassemble the program.
3.One (1) copy of the Licensed Software, including any
software distributed on disks may be made for backup
purposes only. No other copies may be made or used
without the written consent of Liquid Controls.
4.Title. No title to ownership of any Licensed Software
is transferred to the Licensee.
5.Upgrades. License upgrades may become available
for the Licensed Software. Any cost associated with
such upgrades will solely be determined by Liquid
Controls.
6.Warranty. Liquid Controls makes and licensee
receives no warranty express or implied and there are
expressly excluded all warranties of merchantability
and fitness for a particular purpose.
8.Termination. Liquid Controls may terminate this
software license granted hereunder and require return
of the Licensed Software if Licensee fails to comply
with these license terms and conditions.
9.Licensee acknowledges that it has read this
agreement, understands it, and agrees to be bound by
its terms, and further agrees that this is the complete
and exclusive statement of the agreement between
Liquid Controls and Licensee, which supersedes and
merges all prior proposals, understandings, and all
other agreements, oral or written, between the parties
relating to this agreement. This agreement may not be
modified or altered except by written instrument duly
executed by both parties.
10. This Agreement and performance hereunder shall be
construed and interpreted under the laws of the State
of Illinois.
11. If any provision of this agreement is invalid under
any applicable statute or rule of law, it is to that extent
to be deemed omitted.
12. Licensee may not assign or sublicense, without the
prior written consent of Liquid Controls, its rights,
duties, or obligations under this Agreement to any
person or entity in whole or in part.
13. The waiver or failure of Liquid Controls to exercise in
any respect any right provided herein shall not be
deemed a waiver of any further right hereunder.
7.Limitation of Liability. Licensee shall have the sole
responsibility for adequate protection and backup of its
data in connection with the Licensed Software. In no
event shall Liquid Controls be liable for (a) special,
indirect or consequential damages; (b) any damages
whatsoever resulting from loss of use, data, or profits,
product, inaccurate input or work delays, or any direct
property damage arising out of or in connection with
this agreement or the use or performance of the
Licensed Software.
3
LCR-II System Components
Communication and
Programming Devices
Lap PadHand Held ComputerLaptop Computer
Pump & Print
4
LCR-II Overview
This manual provides instructions for the initial setup, the calibration, and the operation of LectroCount LCR-II software,
also known as SR260 firmware. The Liquid Controls LectroCount LCR-II is a microprocessor-based electronic meter
register available in a Division 1 or a Division 2 enclosure. Its primary functions consists of configuring the meter
system to the properties of the liquids to be measured, interfacing with the electronic components of the meter system
components (and external components such as pumps, injectors, and shutdown devices), and performing Weights &
Measures approved custody transfer actions.
The LCR-II performs well in mobile and fixed installations. It can control a meter system as a stand-alone unit or used
as a slave to a host controller such as a process controller or an in-cab data management system.
The LectroCount LCR-II features include:
•Full Weights & Measures Approvals
•Class I, Division 2, Group C, D safety approvals.
•NEMA 4X enclosure with hinged display p anel featuring single connection latch bolt, Weights & Measures sealable
•(2) ¾" and (5) ½" conduit connections
•Operation with most flowmeters. LCR-II accepts input from PD, Amplified Turbine, Electromagnetic and Mass
flowmeters.
•Multi-Point Calibration for improved meter accuracy over the full flow range
•Multiple Product Calibration
•Optional Electronic Temperature Volume Compensation (ETVC)
•Large character, backlit LCD display
•10 digit totalizer
•Two control buttons on front panel provide:
•"SELECT" for accessing internal menus
•"INCREASE" for adjusting internal programmable data
•Push button presetting, rate of flow indication, temperature indication, product selection and more.
LectroCount LCR-II Division 1
Enclosure
LectroCount LCR-II Division 2
Enclosure
5
Lap Pad Operation
LAP PAD OVERVIEW
This section of the manual provides instructions for the
initial set-up, calibration and operation of your LCR-II
Electronic Register using the Liquid Controls Lap Pad
data entry device. This section should first be scanned
to become familiar with the components of the system,
the layout and format of the document, the basic menu
(screens) and data entry required. This section should
then be carefully followed as the system is being set up
and calibrated. (Refer to the Installation Manual for
instructions on connecting the Lap Pad for use with the
LCR-II).
Following is a brief overview of the top level Program
Menu screens.
DELIVERY & PRESET
This menu includes up to seven data screens for setting preset quantities, product prices, taxes and
other relevant delivery information.
PRODUCT & SHIFT INFORMATION
This menu includes four secondary data screens for viewing product types and basic parameters
pertaining to shift information.
GENERAL SET-UP
This menu includes five secondary data screens for setting the internal clock and calendar , initializing
sale and ticket number counters, setting the no-flow timer and defining data to be printed on tickets.
SYSTEM CALIBRATION
This menu includes five secondary screens for entering Meter ID, printer selections, units of measure,
rounding and truncating.
PRODUCT CALIBRATION
This menu includes eight secondary screens for assigning unique codes to products; defining product
types such as gasoline, diesel fuel, LPG , etc.; and compensation parameters such as base temperature
for temperature compensated deliveries.
DIAGNOSTICS
This menu includes four secondary screens for viewing system parameters such as software version,
supply voltage, and pulser diagnostic information.
SECURITY
This menu includes one secondary screen for user password and system security.
6
Lap Pad Operation
NAVIGATING TOP-LEVEL MENUS
When the LCR-II is powered up with a Lap Pad attached, the Lap Pad will briefly display a copyright message for two
seconds followed by the initial top-level menu.
DELIVERY & PRESET
Press Enter or Scroll with ARROW keys.
Press the
Continue pressing the
Press the
Selecting a secondary menu is accomplished by pressing the ENTER key.
Pressing the
ii
i key to navigate to the next top-level menu.
ii
ii
i key to scroll through the rest of the top-level menus (as seen below).
ii
hh
h key to navigate in the opposite direction.
hh
ii
hh
i,
h, M1 and M# keys allows for navigation of the secondary level screens.
ii
hh
DELIVERY & PRESET
Press Enter or Scroll with ARROW keys.
PRODUCT & SHIFT INFORMATION
Press Enter or Scroll with ARROW keys.
GENERAL SETUP
Press Enter or Scroll with ARROW keys.
SYSTEM CALIBRATION
Press Enter or Scroll with ARROW keys.
PRODUCT CALIBRATION
Press Enter or Scroll with ARROW keys.
DIAGNOSTICS
Press Enter or Scroll with ARROW keys.
SECURITY
Press Enter or Scroll with ARROW keys.
7
Lap Pad Operation
SECONDARY DATA SCREENS
Each of the top-level menus has a secondary level consisting of one or more screens. Each screen is divided into one
or more fields where data may be entered or information displayed. Access to data entry is dependent on the following:
DATA FIELD TYPE:
Some fields are for display only and data entry is not necessary (e.g. DIAGNOSTICS and FLOW
RATE).
PASSWORD PROTECTION:
The LCR-II has a user-definable password that allows the system to be locked. This feature authorizes
a supervisor to limit access to certain fields. An operator can still view all data fields whether the unit
is locked or unlocked. Data entry in general system set-up functions (e. g. NO-FLOW TIMER and
PRESET TYPE) are still accessible if the system is locked.
SELECTOR SWITCH:
The multi-position Selector Switch on the front of the LCR-II™ has a CALIBRATION position that is
accessed by removing a Weights & Measures sealable plate. This allows data entry in the metrologically
significant fields (e. g. COMPENSATION TYPE, TEMPERATURE PULSES/UNIT and others).
When a secondary menu is entered by pressing the ENTER key, the cursor will be positioned in the top line of the first
field. Pressing
ii
i will cause the cursor to move to the next field.
ii
Repeatedly pressing
secondary menu screen will be displayed. The cursor will be in the first field. Af ter the last field of the last screen, the
cursor returns to the top line of the first field of the first screen.
NOTE: Pressing the "M#" key scrolls forward to the next secondary menu. Pressing "M1" returns the display to the toplevel menu from any secondary level menu.
The cursor indicates if data can be entered or changed in the field below it. If the cursor is flashing, pressing ENTER
will move the cursor to the second line in the field and data may be entered or changed. In some cases, data is entered
using the keypad. In other cases, data is entered by selecting entries using the
of the field indicates that a scroll list is used for the field. In fields that have a "?", a "YES" or "NO" response is selected.
Information that is entered through the Lap Pad or selected from a menu of choices is stored in the LCR-II by pressing
ENTER. The cursor also moves back to the top line of that menu screen.
Menu maps of the screens accessible in the system are located in Appendix D on pages 50-57.
The next section covers Calibration & Set-up using the Lap Pad.
ii
i will move the cursor through the fields from left to right. After the right most field, the next
ii
ihih
ih keys. A "^" Character in the top line
ihih
8
Lap Pad Calibration & Set-Up
OPERATOROperato can change or enter data.
PASSWORDSystem must be unlocked.
NOT EDITBLERead only (can not be changed by operator).
WEIGHTS & MEASURES
Switch must be in the CALIBRATION position to enter or
change data.
FACTORYRead only (can not be changed by operator).
BEFORE GETTING STARTED
•Make sure that the Lap Pad or other data entry device is connected to the LCR-II register as described in the LCRII Installation Manual.
•When proving the LCR-II system, follow the pre-test and inspection procedures established by Weights & Measures
authorities. The primary indicating and recording element on a vehicle-mounted LCR-II is the 6-digit liquid crystal
display.
•Weights & Measures inspectors are responsible for determining if the truck metering and recording elements of the
system are within tolerance. To make this determination, the system should be tested under normal conditions.
•The LCR-II is pre-programmed with common values for many of the programmable parameters. These values
should be checked to ensure that they fit the specific requirements of the given installation. Detailed instructions
for checking and changing these values follow later in this publication.
•Changes cannot be made to metrologically significant data without first removing the Switchplate and its associated
Weights & Measures seal. The security level for each of the programming screens is defined in the following table:
NOTE: This table applies to the lower line of the Lap Pad display only. The upper line cannot be edited.
STEP 1: REMOVE THE LCR-II
SWITCHPLATE
•Remove the lead seal and wire assembly from the
LCR-II Switchplate (if previously sealed by Weights
& Measures authorities).
•Remove the four screws that secure the Switch Plate
over the red Selector Switch located on the front of
the LCR-II, and remove the plate.
•Rotate the Selector Switch counter clockwise to the
6 o'clock (CALIBRATION) position to permit access
to all programming fields (except FACTORY).
9
Lap Pad Calibration & Set-Up
USER KEY
0001
SECURITY^
UNLOCKED
Security - Screen 1
General Setup - Screen 1
TIME HH:MM:SS
9:05:24
DATE
10/6/2002
DATE FORMAT^
MM/DD/YY
STEP 2: SECURITY
Press the M1 Key to access a top-level menu. Press
gain access to the secondary menu.
USER KEY (Password)
Use this field to enter the owner/office password. To enter a password, press ENTER to drop into the
data entry field. Enter a password using up to 10 alphanumeric characters. Press ENTER. This
password will be required in the future to unlock the system in order to gain access to secured menu
and data entry fields. This password will also be needed to change the time using the LCR-II control
buttons. If it is desired to have access to time adjustment, the USER KEY must be numeric only and
contain 5 digits.
ii
Press
i and the cursor will move to:
ii
SECURITY^
This field is used to lock the system. Press ENTER to drop to the bottom field. Use i to scroll between
LOCKED and UNLOCKED. Press ENTER to select the desired option. While the system is LOCKED,the operator is unable to change system data other than GROSS and NET PRESETS, PRODUCT
CODE, PRODUCT NAME, and NO-FLOW TIMER.
ii
i to scroll to the SECURITY menu. Press ENTER to
ii
NOTE: The LCR-II is shipped LOCKED from the factory.
Press M1 to return to the top-level menu. Press
STEP 3: GENERAL SETUP
The fields in this menu are not specific to any of the four possible product calibrations, therefore the
Selector Switch does not have to be in the CALIBRATION position to allow data entry in many of these
fields and changes can be made without breaking Weights & Measures seals.
Press ENTER and the cursor will move to:
DATE FORMAT^
This field is used to determine in which format the date will be displayed and printed: month first or day
first. Press ENTER and the cursor will drop to the second line. Now you can use the arrow keys to
select the format, either month first (MM/DD/YY) or day first (DD/MM/YY). When the choice is displayed,
press ENTER and the cursor will move back to the top line.
ii
Press
i and the cursor will move to:
ii
DATE MM/DD/YY (or DD/MM/YY depending on the previous field).
This field is used to set the LCR-II internal calendar. The LCR-II updates its calendar and will print the
correct date on delivery tickets. Press ENTER and the cursor will drop to the bottom line. Enter the
current date on the keypad using only numbers. Press ENTER and the cursor will move back to the top
line.
ii
i to scroll to GENERAL SETUP.
ii
10
STEP 3: GENERAL SETUP CONTINUED
UNIT ID
123456
NO-FLOW TIMER
180
SALE#
25
TICKET#
1
General Setup - Screen 2
ii
Press
i and the cursor will move to:
ii
TIME HH:MM:SS
This field is used to set the LCR-II internal clock. The time, like the date, is updated by the LCR-II and
printed on the delivery tickets. Press ENTER and the cursor will drop to the bottom line. Enter the
current hour, minutes, and seconds. Use military time (e.g. press 13:01:15 for 1:01:15 PM).
HINT: To synchronize the clock to the second, press ENTER when the appropriate second is reached.
Press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
SALE #
This field is used to track the number of transactions that the LCR-II processes. The LCR-II will increment
the field each time a delivery is made, wrapping around to 000000 after 999999. Press ENTER and the
cursor will drop to the bottom line. Key in the starting SALE #, up to 6 digits. Press ENTER and the
cursor will move to the top line.
Lap Pad Calibration & Set-Up
ii
Press
i and the cursor will move to:
ii
TICKET #
This is similar to the sale number. It will increment every time that the LCR-II print s a ticket. If multiple
or duplicate tickets are used for transactions, the ticket number will be incremented more than the sale
number. NOTE: If a TICKET # of 0 is entered, the TICKET # will not print on the delivery ticket and the
TICKET # will not increment. Press ENTER and the cursor will drop to the bottom line. Key in the
starting TICKET #, up to 6 digits. Press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
UNIT ID
This is a number that can be used to identify the driver, location, or truck that the LCR-II is associated
with. Press ENTER and the cursor will drop to the bottom line. Key in your UNIT ID, up to 10 characters.
Press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
NO-FLOW TIMER
The NO-FLOW TIMER is an internal timer in the LCR-II that starts running when the LCR-II senses
that there is no longer any product moving through the meter. If the timer count s up to its set point, the
LCR-II will assume that the delivery is over, and print a ticket. The timer can be set to count up to 3600
seconds before printing the ticket. This feature can be deactivated by entering 0 seconds, allowing
multiple tanks to be filled at a location. The timer helps to ensure that deliveries are not split between
authorized and unauthorized locations. NO-FLOW TIMER is not active in prover mode. If the value is
set to 0 or any value greater than 180, "Multiple Deliveries at One Site" will be printed on the delivery
ticket.
NOTE: Internal timer is activated once the flow delivery has started and a minimum amount of flow (1
gallon or 5 liters) is registered and then is stopped.
11
Lap Pad Calibration & Set-Up
PULSE OUTPUT EDGE^
RISING
PRESET TYPE^
CLEAR
PRESET^
GROSS
General Setup - Screen 3
STEP 3: GENERAL SETUP CONTINUED
NO-FLOW TIMER CONTINUED
Press ENTER and the cursor will drop to the bottom line. Enter the desired NO-FLOW TIMER value
inseconds. Press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
PRESET^
Available preset options include NET, BOTH, GROSS, and NONE. Selecting NONE will disable
presetting. Presetting NET, GROSS, or BOTH is allowed. Press ENTER and the cursor will drop to the
bottom line. Use
press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
ii
i to scroll through the available preset options. When the desired option is displayed,
ii
PRESET TYPE^
There are four preset types:
CLEAR
RETAIN
MULTIPLE
INVENTORY
Press ENTER and the cursor will drop to the bottom line. Use
When the desired choice is displayed, press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
PULSE OUTPUT EDGE^
This feature allows you to synchronize the calibrated pulse output waveform with the requirements of
an external system that uses that LCR-II signal. There are two choices: RISING or FALLING. (Some
counters increment on rising, others on falling pulse edges. Refer to equipment technical manual to
determine the specific requirements of your connected equipment). Press ENTER and the cursor will
drop to the bottom line. Use
press ENTER and the cursor will move to the top line.
Resets the preset values to 0 after the current delivery is ended.
Maintains the preset values to be used again on the next delivery.
Allows more than one preset to be run before a delivery ticket is printed. Printing must be
initiated by a PRINT command.
Maintains the remaining preset amount between deliveries (i. e. indicates the remaining volume
in the vehicle tank).
ii
i to scroll through the four options.
ii
ii
i to scroll between the two choices. When the desired choice is displayed,
ii
ii
Press
i and the cursor will move to:
ii
12
STEP 3: GENERAL SETUP CONTINUED
TICKET HEADER LINEHDR^
1
General Setup - Screen 4
General Setup - Screen 5
PRINT GROSS & PARAM?
VOL CORRECTED MSG?
YESYES
HDR^
This field is used to select the line number of the ticket header that will be edited in the following field.
Press ENTER and the cursor will drop to the bottom line. Use the
from 1 to 12. When the desired choice is displayed, press the ENTER key and the cursor will move to
the top line.
ii
Press
i and the cursor will move to:
ii
TICKET HEADER LINE
This feature is useful when using blank (as opposed to pre-printed) tickets. The field allows the fuel
dealer to enter the company name, address, telephone number, etc., which will be printed on each
delivery ticket. This field (35-character max.) allows you to enter or edit ticket header labels. The
Header Line will scroll in tandem with the HDR # selected in the previous field. Any lines that are left
blank will not be printed on the ticket.
Lap Pad Calibration & Set-up
ihih
ih keys to scroll through the numbers
ihih
NOTE: Header Lines 11 and 12 can only be edited in the CALIBRATION mode. TICKET HEADER
LINE 11 only prints when AUX OUT 1 is ON or ON DURING DELIVERY as shown on the Lap Pad
Screen. TICKET HEADER LINE 12 only prints when AUX OUT 2 is ON or ON DURING DELIVERY.
After data has been entered, press ENTER.
ii
Press
i and the cursor will move to:
ii
PRINT GROSS & PARAM?
This allows one to choose whether to have the gross volume and compensation parameter printed on
the ticket if the product is temperature compensated. Press ENTER and the cursor will drop to the
bottom line. Use
press the ENTER key and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
VOL CORRECTED MSG?
This field allows you to print a message on the delivery ticket indicating that delivery volume has been
corrected to a base temperature (defined later in SYSTEM CALIBRATION). The user has the option
(YES or NO) to print the message. Press ENTER and the cursor will drop to the bottom line. Use
scroll between the two choices, YES or NO. When the desired choice is displayed, press the ENTER
key and the cursor will move to the top line. Press M1 to return to the top of the GENERAL SETUP
menu.
ii
i to scroll between the two choices, YES or NO. When the desired choice is displayed,
ii
ii
i to
ii
13
Lap Pad Calibration & Set-up
System Calibrat ion - S creen 1
METER ID
12345678 90YESEP SON New FontB
PRINTERTICKE T?
RESIDUAL^
UNITS^
GALONS
System Calibration - Screen 2
ROUND
DECIMAL^
TENTHS
RATE BASE^
PER MINUTE
STEP 4: SYSTEM CALIBRATION
From the top-level menu, use
METER ID
This number allows you to uniquely identify an LCR-II/Meter combination. If the LCR-II is part of a
multiple-meter system, it is important that this number be unique. It is recommended that the meter
serial number be entered here. This number will print on the calibration/diagnostic ticket. Press
ENTER and the cursor will drop tot he bottom line. Key in 1 to 10 characters. Press ENTER and the
cursor will move back to the top line.
ii
Press
i and the cursor will move to:
ii
TICKET?
This field allows you to choose whether or not a ticket will be required for each delivery . Most W eights
& Measures governed truck applications will require a ticket. If YES is chosen, the LCR-II will not allow
deliveries to be made unless a ticket is in the printer, the printer is operational and the previous delivery
ticket has been printed in its entirety. Press ENTER and the cursor will drop to the bottom line. Use
to scroll between the two choices, YES or NO. When the desired choice is displayed, press ENTER.
The cursor will move to the top line. When NO is selected, tickets are not mandatory but will print if a
printer is connected and available to print a ticket.
ii
i to scroll to SYSTEM CALIBRATION. Press ENTER and the cursor will move to:
ii
ii
i
ii
ii
Press
i and the cursor will move to:
ii
PRINTER^
This field allows you to select a printer type. The six choices are EPSON NewFontB (select for use
with EPSON 200 Roll and EPSON 220 Roll). EPSON NewFontA (select for use with EPSON TMT88iii), EPSON OldFontA (select for use with EPSON 290 Slip and EPSON 295 Slip), EPSON OldFontB
(select for use with EPSON 300 Roll), OKIDATA ML184T (select for use with Okidata ML184T),BLASTER (select for use with Cognitive Solutions Thermal Printer). Press ENTER and the cursor will
drop to the bottom line. Use
press ENTER and the cursor will move back to the top line.
ii
Press
i to move to:
ii
UNITS^
This field is used to choose the units of flow measurement. The choices are GALLONS, LITRES,
CUBIC M, LBS (pounds), KGS (kilograms), BARRELS, and OTHER. Press ENTER and the cursor
will drop to the bottom line. Use the
displayed, press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
RATE BASE^
Use this field to select the time unit for measurement on the flow rate display . The choices are rate PER
HOUR, PER MINUTE and PER SECOND. Press ENTER and the cursor will drop to the bottom line.
ii
Use
i to scroll between the three choices. When the desired choice is displayed, press ENTER and
ii
the cursor will move to the top line.
ii
i to scroll between the choices. When the desired choice is displayed,
ii
ihih
ih keys to scroll through the choices. When the desired choice is
ihih
ii
Press
i and the cursor will move to:
ii
14
STEP 4: SYSTEM CALIBRATION CONTINUED
TEMP
25.53
System Calibration - Screen 3
RTD OFS
2.817
RTD SLP
0.024724
T TEMP^
DEG. F
OFFSET
0.00
DECIMAL^
This field allows you to choose between whole units, tenths and hundredths of a unit. Press ENTER
and the cursor will drop to the bottom line. Use
When the desired choice is displayed press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
RESIDUAL^
Volumes less than the least significant digit can be rounded into that digit or truncated (thrown away)
depending on the selection mode. Press ENTER and the cursor will drop to the bottom line. Press
scroll between ROUND and TRUNCA TE. When the desired choice is displayed press ENTER and the
cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
TEMP
This field displays the current temperature reading from the RTD temperature probe and allows one to
enter a corrected value from a Weights & Measures thermometer . If the LCR-II is not equipped with the
optional Electronic Temperature Volume Compensation (ETVC) Kit, this field will read "--".
Lap Pad Calibration & Set-up
ii
i to scroll between WHOLE, TENTHS, and HUNDR.
ii
ii
i to
ii
To make a correction:
ii
Press
i and the cursor will move to:
ii
OFFSET
The offset represents the difference between the official Weights & Measures reading and the LCR-II
RTD reading. The offset is automatically calculated if an entry is made in the previous field.
If a Weights & Measures thermometer reading was not entered in the previous field, subtract the value
listed in the TEMP field from the Weights & Measures measurement. Key in the difference in the of fset
field. If the number is negative, use "*" on the Lap Pad to change the sign. Adjustment is limited to
±0.3°C (±0.54°F). Any adjustment larger than ±0.3°C (±0.54°F) generates a " RANGE ERROR" message
and requires replacement of the RTD probe.
ii
Press
i and the cursor will move to:
ii
Run enough product through the meter to allow the temperature to stabilize. Compare the
TEMP reading with the current Weights & Measures thermometer reading. If the readings are
different, the Weights & Measures reading should be entered. Press ENTER and the cursor
will drop to the bottom line. Enter the new value. The new value will overwrite the previous one
and the OFFSET field will be recalculated by the LCR-II. Offset adjustments are limited to
±0.3°C (±0.54°F). Adjustments greater than this limit require replacement of the RTD probe.
Press ENTER and the cursor will move to the top line.
15
Lap Pad Calibration & Set-up
System Calibration - Screen 4
FLOW DIR^
-->
LAST CALIBRATED
7/11/2002 11:44
LCR #
250
System Calibration - Screen 5
CONFIG EVENT
1
CALIB EVENT
19
CALIB#
34
STEP 4: SYSTEM CALIBRATION CONTINUED
T UNIT^
This field allows you to choose the temperature units - Degrees Fahrenheit or Degrees Celsius. Press
ENTER and the cursor will drop to the bottom line. Use
and DEG. C. When the desired choice is displayed, press ENTER and the cursor will move to the top
line.
ii
Press
i and the cursor will move to:
ii
RTD SLP
Factory calibration setting only. Press
RTD OFS
Factory calibration setting only. Press
FLOW DIR^
This field allows you to match the LCR-II flow direction to the meter's pulse output. If the counter runs
backwards, this arrow should be changed. The choices are
drop to the bottom line. Use
press ENTER.
ii
i to scroll between the two choices. When the desired choice is displayed,
ii
ii
i and the cursor will move to:
ii
ii
i and the cursor will move to:
ii
ii
i to scroll between the two choices, DEG. F
ii
ff
gg
f or
g. Press ENTER and the cursor will
ff
gg
ii
Press
i and the cursor will move to:
ii
LAST CALIBRATED
This field contains the date and time of the last calibration. It cannot be edited.
ii
Press
i and the cursor will move to:
ii
LCR #
This field contains the node address for RS485 communication. Enter a number between 1 and 250.
The default value is 250.
ii
Press
i and the cursor will move to:
ii
CALIB #
For metrology use only. This is the number of times the calibration switch position has been entered.
DISPLAY ONLY.
CALIB EVENT
For metrology use only . This is the number of times the calibration has been changed. DISPLA YONLY .
CONFIG EVENT
For metrology use only. This is the number of times the configuration has been changed. DISPLAY
ONLY.
Press the M1 key to return to the top of the SYSTEM CALIBRATION menu. Then press
SHIFT INFORMATION.
16
ii
i to scroll to PRODUCT &
ii
STEP 5: PRODUCT & SHIFT INFORMATION
Product & Shift - Screen 1
DLVRY
22
SHIFT START
7/01/2002 1:44:00
PROD TYPE
DISTILLATE
#^
1
Product & Shift - Screen 2
UNITS
GALLONS
SHIFT NET
0.0
SHIFT GROSS
1647.6
#^
1
Press ENTER and the cursor will move to:
#^ (Product Number)
This field is used to select one of four possible product types/calibrations. Press ENTER and the
cursor will drop to the bottom line. Use
is selected for which corresponding PROD TYPE^ information will be entered, press ENTER and the
cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
PROD TYPE^
This field will scroll along with the #^(Product Number) that is shown in the first field. The product type
AMMONIA, AVIATION, DISTILLATE, GASOLINE, METHANOL, LPG, LUBE OIL or "blank" can be
selected. Press ENTER and the cursor will drop to the bottom line. Press
available options. When the desired choice is displayed, press ENTER and the cursor will move to the
top line.
Lap Pad Calibration & Set-up
ii
i to scroll through the numbers 1-4. When the product number
ii
ii
i to scroll through the
ii
ii
Press
i and the cursor will move to:
ii
SHIFT START
The starting time and date for the current shift are shown here. DISPLAY ONLY.
DLVRY
This field shows the number of deliveries that have been made since the last time CLR SHIFT was set
to YES or a shift ticket was printed. DISPLAY ONLY.
ii
Press
i and the cursor will move to:
ii
#^ (Product Number)
This field is used to select one of four possible product types/calibrations that can be set up in the LCRII. At least one calibration must be set up to allow deliveries.
SHIFT GROSS
This field displays the gross number of units that have been delivered through the meter since the last
time the CLR SHIFT field was set to YES or a shift ticket was printed. SHIFT GROSS is the sum total
of the delivery ticket gross quantities during the current shift. DISPLAY ONLY.
ii
Press
i and the cursor will move to:
ii
SHIFT NET
This field displays the net (temperature compensated) number of units that have been delivered through
the meter since the last time CLR SHIFT field was set to YES or a shift ticket was printed. SHIFT NET
is the sum total of the delivery ticket net quantities during the current shift. DISPLAY ONLY.
17
Lap Pad Calibration & Set-up
Product & Shift - Screen 3
UNITS
GALLONS
NET TOTAL
0.0
GROSS TOTAL
6193.8
#^
1
Product & Shift - Screen 4
CLR SHIFT?
NO
STEP 5: PRODUCT & SHIFT INFORMATION CONTINUED
UNITS^
From this field, a unit of measure is selected. Choose GALLONS, LITRES, CUBIC M, LBS (pounds),
KGS (kilograms), BARRELS, or OTHER.
ii
Press
i and the cursor will move to:
ii
#^ (Product Number)
This field is used to select one of four possible product types/calibrations. At least one calibration must
be set up in the LCR-II to allow deliveries.
GROSS TOTAL
This field shows the total gross volume that has gone through the register since the field was last
initialized. This value can be initialized to any positive number if the Selector Switch is in the
CALIBRATION position. This is a "live" totalizer and will count regardless of a delivery being active.
ii
Press
i and the cursor will move to:
ii
NET TOTAL
This field shows the total net volume that has gone through the register since the field was last initialized.
This value can be initialized to any positive value if the Selector Switch is in the CALIBRA TION position.
This is a "live" totalizer and will count regardless of a delivery being active.
ii
Press
i and the cursor will move to:
ii
UNITS^
From this field, a unit of measure is selected. Choose from GALLONS, LITRES, CUBIC M, LBS
(pounds), KGS (kilograms), BARRELS, or OTHER. If more than one product is entered, the above
listed procedures should be repeated for each.
ii
Press
i and the cursor will move to:
ii
CLR SHIFT?
This field is used to print a SHIFT TICKET and clear the SHIFT GROSS, SHIFT NET, and DLVRY
fields. If YES is selected, the shift fields will be reset to 0 after the shift ticket is printed. After the ticket
is printed, this field will revert to NO. If TICKET? is set to NO, shift totals can be cleared without
printing a shift ticket. Press ENTER and the cursor will drop to the bottom line.
Press the
return to the top line.
ii
i key to scroll between YES and NO. When the desired entry is displayed, press ENTER and the cursor will
ii
18
Lap Pad Calibration & Set-up
Product Calibration - Screen 1
PRODUCT TYPE
DISTILLATE
PRODUCT NAME
Product Name
CODE
12345
#^
1
Product Calibration - Screen 2
BASE TEMP
15.0
COMP PARAM
653
COMPENSATION TYPE
TABLE 54B
STEP 6: PRODUCT CALIBRATION
Press M1 to return to the top-level menu. Press i to scroll to PRODUCT CALIBRATION.
There are four product calibrations that can be set up in the LCR-II. The products are calibrated one at a time. Press
ENTER and the cursor will move to:
#^(Product Number)
This number will correspond to one of the four product types/calibrations that can be set up in the LCRII. At least one calibration must be set up to allow deliveries. Press ENTER and the cursor will drop to
the bottom line. Use i to scroll between the four choices, 1 to 4. Press the ENTER key when the
desired choice is displayed.
ii
Press
i and the cursor will move to:
ii
CODE
A five character alphanumeric code can be entered here. The code will correspond to one of the
products within this product type. This code can also be entered in the DELIVERY & PRESET menu.
Press ENTER and the cursor will drop to the bottom line. Enter a code that will aid the operator in
identifying the product or one that has been specified for use by a host computer system. Press ENTER
and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
PRODUCT NAME
The PRODUCT NAME contains a product description (up to 18 alphanumeric characters) that will print
on the delivery ticket. NOTE: Make sure that the PRODUCT NAME matches up with the PROD
TYPE^, #^, and CODE that are also entered on this screen. This name can also be entered in the
DELIVERY & PRESET menu. Press ENTER and the cursor will drop to the bottom line. Enter theappropriate PRODUCT NAME. Press ENTER and the cursor will move to the top line.
ii
Press
i and the cursor will move to:
ii
PROD TYPE^
This field will scroll along with the #^(Product Number) that is shown in the first field. The product type
AMMONIA, AVIATION, DISTILLATE, GASOLINE, METHANOL, LPG, LUBE OIL and "blank" canalso be set up in the PRODUCT & SHIFT INFORMATION menu.
ii
Press
i and the cursor will move to:
ii
COMPENSATION TYPE^
Select the type of Temperature Volume Compensation to be used for this specific product number. No
deliveries are permitted if the LCR-II is configured for temperature compensation and the temperature
circuit has failed or the measured temperature is out of the specified range (see Table in Appendix A).
19
Lap Pad Calibration & Set-up
Product Calibration - Screen 3
UNITS
GALLONS
PROVER QTY
0.000
PULSE/UNIT
2222.000000
#^
1
Product Calibration - Screen 4
GALLONS
UNITS
----------
REMAINING
PRESET
0.0
GROSS QTY
0.0
STEP 6: PRODUCT CALIBRATION CONTINUED
COMPENSATION TYPE^ CONTINUED
Scroll through the choices: Linear F, Linear C, API Table 24, API T able 54, API Table 6B, API Table
54B, API Table 54C, API Table 54D, NH3 or NONE. If "NONE" is selected, deliveries will be in gross
quantities only. Net Quantities will read "0" on the screen. Refer to the Compensation Types and
Parameters table (Appendix A) when choosing the type of compensation to be used.
COMP PARAM
This field contains the coefficient of expansion, the standard density, API gravity, or the specific gravity
that will be used with the COMPENSA TION TYPE selected in the previous field. Enter a number based
on the COMPENSATION TYPE and the product that is being metered. Refer to the Compensation
Types and Parameters table found in Appendix A.
BASE TEMP
This field is used to set the base temperature for temperature compensated deliveries when using
either the Linear C or Linear F Compensation type. It defaults to either 60°F or 15°C when using
either the API tables or NH3. Enter a number. Press ENTER to return to the top line.
ii
Press
i and the cursor will move to
ii
#^ (Product Number)
This number will correspond to one of the four product types/calibrations that can be set up in the LCRII. At least one calibration must be set up to allow deliveries.
PULSES/UNIT
This field is the number of pulse edges that the LCR-II counts per unit of measure. (Only Gross V olume
applies). This number is used to scale the PROVER QTY (Gross).
PROVER QTY
This field displays the metered volume. When this value is overwritten with the actual PROVER QTY,
a new PULSES/UNIT (k-Factor) will be recalculated automatically. (Only for Calibration use.) If this is
the initial calibration for this meter, first enter the rest of the PRODUCT CALIBRATION data before
entering this field. Once the other PRODUCT CALIBRATION parameters are entered, press START
and fill up a reliable prover. As the prover is filled, this field will increment, displaying a volume based
on the existing PULSES/UNIT parameter. After the prover has been filled, press PRINT and enter the
exact prover reading in place of the reading displayed. When the new reading is entered, the PULSES/UNIT parameter will be recalculated by the LCR-II, "zeroing" out the error. Refill the prover. Check to
ensure that the prover and LCR-II gross volumes are within tolerance limits. If not, enter the new
prover volume, and retest.
ii
Press
i and the cursor will move to:
ii
GROSS QTY
This field displays the gross delivery total for the product being delivered. DISPLAY ONLY.
20
STEP 6: PRODUCT CALIBRATION CONTINUED
Product Calibration - Screen 5
TEMP
UNITS^
GALLONS
23.4
NET QTY
0.0
GROSS QTY
0.0
Product Calibration - Screen 6
AUX UNIT^
LITRES
AUT QTY
0.0
AUX MULT
0.000
S1 CLOSE
0.0
PRESET
The Gross Preset amount is entered here. When this amount is reached the valve outputs are turned
off.
REMAINING
The number of units remaining to be delivered on the current GROSS DELIVERY preset is displayed.
UNITS^
Choose the units of measure to be displayed. The choices are GALLONS, LITRES, CUBIC M, LBS
(pounds), KGS (kilograms), BARRELS, or OTHER. Use the hi keys to scroll through the choices.
Press ENTER to make a selection.
ii
Press
i and the cursor will move to:
ii
GROSS QTY
This field displays the gross delivery total for the product being delivered.
Lap Pad Calibration & Set-up
NET QTY
This field will show the NET QTY of the delivery in progress, or the last delivery if the meter isn't
running. The net quantity is calculated by multiplying the gross quantity by the calculated Volume
Correction Factor (VCF). If no compensation is active, it is set to zero. The VCF calculation is based in
part on the temperature reading from the RTD and the offset that was calculated/entered in the SYSTEM
CALIBRA TION menu.
UNITS^
Choose the units of measure to be displayed. The choices are GALLONS, LITRES, CUBIC M, LBS
(pounds), KGS (kilograms), BARRELS, or OTHER. Use the
Press ENTER to make a selection.
TEMP
This displays the actual product temperature if the optional ETVC Kit is installed.
ii
Press
i and the cursor will move to:
ii
S1 CLOSE
This number is used to set the dwell period if the LCR-II is controlling a two-stage valve or presetting
with a one-stage valve. This represents the number of measurement units (e.g. GALLONS, LITRES)
that are remaining on a preset delivery before the valve is placed in a bypass, trickle, or dwell mode. An
S1 CLOSE value is needed for each of the four product numbers. Enter a number between 0 and 500.
hihi
hi keys to scroll through the choices.
hihi
The number entered here will match up with the #^(Product Number) that was selected in the first
field of this menu.
21
Lap Pad Calibration & Set-up
PT^
1
Product Calibration - Screen 7
RATE
0.00
%ERROR
0.000
RATE BASE^
PER MINUTE
UNITS^
GALLONS
x 100
(Prover Qty - Meter Qty)
Prover Quantity
%Error =
STEP 6: PRODUCT CALIBRATION CONTINUED
AUX MULT
This field is used to convert the quantity delivered, (e.g. GALLONS, LITRES) to an alternate volume or
mass unit such as CUBIC METERS, LBS, etc. The user must furnish an applicable conversion factor.
If AUX MUL T is set to 0, it will not print on the delivery ticket. Example: To convert GALLONS to LBS,
AUX MULT is calculated as follows:
AUX MULT= (SpGr) x (8.345 Ibs/gal)
(Where SpGr is the specific gravity.)
AUX QTY
This field is used to display the volume delivered in terms of the alternate volume or mass unit.
AUX UNIT^
This field is used to define the auxiliary unit of measure that will be printed on the delivery ticket, i.e.,
GALLONS, LITRES, CUBIC M, LBS, KGS, BARRELS, or OTHER.
ii
Press
i and the cursor will move to:
ii
PT^
This field is used to calibrate the meter/register at up to 10 different flow rates for each product type/
calibration.
RATE
This field displays actual flow rate during delivery. Enter the actual flow rate for multi-point calibration.
UNITS^
Choose the units of measure to be displayed. The choices are GALLONS, LITRES, CUBIC M, LBS
(pounds), KGS (kilograms), BARRELS, or OTHER. Use the hi keys to scroll through the choices.
Press ENTER to make a selection.
RATE BASE^
Choose the time unit for measurement on the flow rate display. The choices are units PER HOUR,
PER MINUTE and PER SECOND. Use the
make a selection.
%ERROR
This field contains the %ERROR difference against the base number.
ii
Press
i and the cursor will move to:
ii
hihi
hi keys to scroll through the choices. Press ENTER to
hihi
22
STEP 6: PRODUCT CALIBRATION CONTINUED
PT^
1
Product Calibration - Screen 8
LINEAR^
SETUP
%ERROR
0.000
UNITS^
GALLONS
PROVER QTY
0.000
x 100
(Prover Qty - Meter Qty)
Prover Quantity
%Error =
All calibration proving runs are measured in
gross quantity
PT^
This field is used to calibrate the meter/register at up to 10 different flow rates for each product type/
calibration.
PROVER QTY
This field needs to contain the exact volume in the prover tank tested at actual flow rate. When an entry
is made, the actual %ERROR (delivery) against the base number will be recalculated automatically.
UNITS^
Choose the units of measure to be displayed. The choices are GALLONS, LITRES, CUBIC M, LBS
(pounds), KGS (kilograms), BARRELS, or OTHER. Use the
Press ENTER to make a selection.
%ERROR
This field contains the %ERROR difference against the base number.
Lap Pad Calibration & Set-up
hihi
hi keys to scroll through the choices.
hihi
LINEAR^
Select SETUP for single-point calibration or when setting up multi-point calibration. Select APPLIED to
activate multi-point calibration.
NOTE: See pages 23-24 for optional multi-point calibration instructions. If your LCR-II is set up for
single-point calibration, this concludes the Initial Calibration and Set-up procedures. Proceed to Page
25 for BASIC OPERATION instructions.
The next section covers Multi-Point Calibration.
23
Lap Pad Calibration & Set-up
PT^
1
0.00GALLONSPER MINUTE0.000
Product Calibration - Screen 7
RATEUNITS^RATE BASE^%ERROR
All calibration proving runs are measured in gross quantity
MULTI-POINT CALIBRATION
Multi-point calibration allows a flowmeter to be used over a wide range of flow rates without experiencing a loss in
measurement accuracy dueA to the shape of the meter curve (usually at low flow rates). Af ter determining the inaccuracy
of the meter at a variety of flow rates and inputting that information into the LCR-II multi-point calibration table, accuracy
corrections are then continuously made by the LCR-II based on the measured flow rate. There are two ways to obtain
the information necessary to load the multi-point calibration table:
1. Manufacturer's data supplied with the meter can be entered directly.
2. The meter can be field calibrated by proving at various flow rates.
1. MANUFACTURER'S DATA TECHNIQUE
Step 1:
Go to Screen 7 of PRODUCT CALIBRA TION. From the manufacturer's data sheet showing the accuracy
at various flow rates, enter into PT^1 the flow RATE and the delivery %ERROR corresponding to the
maximum flow rate of the meter.
Step 2:
Select up to nine additional points from the manufacturer's data sheet to characterize the remainder of
the meter's accuracy curve and enter this data into PT^2, PT^3, etc., respectively, in descending order
by rate. (While it is not necessary to enter the selected points in a specific order , it is desirable to do so
because the LCR-II will re-sort the data and arrange it in descending order by rate when it is APPLIED.)
T o enter negative %ERROR values, first enter the number , then press the "*" key to display the "-" sign.
Errors that can occur during multi-point table data entry include:
DUPLICATE FLOW RATE ERROR
The same flow rate value was entered twice.
RANGE ERROR
A %ERROR outside the ±3.0% limit was entered.
POINT SELECTION TIPS:
•Because the LCR-II uses a piece-wise linear interpolation method of correction, it is desirable to
use more data points along segments of the accuracy curve that have the greatest curvature
(typically in the low flow rate region).
•Where the meter curve already approximates a straight line (in the high flow rate region), use
fewer of the available points.
Step 3:
Set the LINEAR^ field to APPLIED and enter it. This puts the linearizing algorithm into effect. If the
error message ADJACENT POINTS OUT OF 0.25% RANGE is displayed, APPLIED will not take
effect and the points will need to be reselected to ensure that no two points adjacent to each other are
more than 0.25% apart in %ERROR.
Step 4:
Field prove the meter at the normal high flow rate of the system. Adjust the PULSES/UNIT in Screen 3
of the PRODUCT CALIBRA TION menu by entering the actual PROVER QTY (see Page 19) to obtain
accuracy within tolerance limits.
Refill the prover at the same flow rate as Step 4 to verify that the meter remains within tolerance
limits. Assuming that the dat a provided by the meter manufacturer is correct, and that proper technique
was used for selecting and entering the multi-point table, it should be possible to fill the prover at any
flow rate within the linearized range and obtain an error near zero.
2. FIELD PROVING TECHNIQUE
When the meter's accuracy is not provided by the manufacturer, points to enter into the multi-point calibration table
need to be determined by field proving at various flow rates.
Step 1:
Start a proving run of the meter while viewing PRODUCT CALIBRATION - Screen 3 at the normal high
flow rate of the system while the LINEAR^ field is in the SETUP mode. With the prover full, press
PRINT and adjust the PULSES/UNIT by entering the actual PROVER QTY (see Page 22).
Step 2:
Step 3:
Step 4:
Go to of PRODUCT CALIBRATION - Screen 7. With PT^ set to 1 and LINEAR^ still set to SETUP,
start a new prover run at the same flow rate as Step 1. During prover filling, note the maximum RATE
that is displayed. When the prover is filled, press PRINT. Advance to PRODUCT CALIBRATION -
Screen 8. Enter the volume delivered to the prover in PROVER QTY. The %ERROR field will then
display the calculated error for that rate. This should be a very small value since Step 1 "zeroed" the
meter at the same rate. Return to PRODUCT CALIBRA TION - Screen 7 and enter the maximum flow
rate observed during prover filling in RATE for PT^ 1.
With PT^ set to the next point and LINEAR^ still set to SETUP, start a prover run at a different flow
rate. During prover filling, note the maximum RATE that is displayed. When the prover is filled, press
PRINT. Advance to PRODUCT CALIBRATION - Screen 8. Enter the volume delivered to the prover in
PROVER QTY. The %ERROR field will then display the calculated error for that rate. Return to
PRODUCT CALIBRA TION - Screen 7 and enter the maximum flow rate observed during prover filling
in RATE for that PT^.
Continue proving at other flow rates using additional PT^ numbers following the same procedure as
Step 3. A minimum of three points is recommended (high, medium and low flow) to obtain a good fit to
the curve. All ten points can be used to obtain optimum results. Any unused points should be left at
RATE=0 and %ERROR=0.
25
Lap Pad Calibration & Set-up
#^
1
Delivery & Preset - Screen 1
CODE
PRODUCT NAME
PRODUCT TYPE^
DISTILLATE
PRODUCT NAME
12345
MULTI-POINT CALIBRATION CONTINUED
Step 5:
Set the LINEAR^ field to APPLIED and enter it. This puts the linearizing algorithm into effect. The
LCR-II will re-sort the multi-point table and display the points in descending order by RA TE. Do not be
alarmed if the data changes in a given PT^. If the error message ADJACENT POINTS OUT OF 0.25%RANGE is displayed, APPLIED will not take effect and more points will need to be added to ensure
that no two points adjacent to each other are more than 0.25% apart in %ERROR.
Lap Pad Basic Operation
PRESET DELIVERY WITH A LAP PAD
If a terminal or Lap Pad is attached to the LCR-II serial port, it can be used to preset the delivery volume. The product
number is selected in the first screen of the DELIVERY & PRESET menu and the GROSS or NET PRESET is entered
in the third or fourth screen. (NOTE: The NET PRESET screen will only be displayed if the product being delivered is
temperature compensated). If the Selector Switch has been left in the RUN position, a delivery will begin when the
START key is pressed on the Lap Pad or terminal. Product will then be delivered until NET or GROSS PRESET amount
is reached, at which point the delivery will be halted.
A delivery ticket will be printed:
•As soon as the preset volume is reached and if CLEAR or RETAIN has been selected
•If the PRINT key is pressed
•If the NO-FLOW TIMER times out
If both NET and GROSS PRESETS are non-zero, the delivery will end when the lower of the two is reached.
If the preset type is set to MULTIPLE, the valve will close when the preset amount is reached. The operator can
continue that delivery, making additional "drops" of the same quantity or preset amounts. The preset amount can be
changed between drops. The delivery ends when the PRINT key is pressed or the Selector Switch is placed in the
PRINT position. The delivery ticket will show the total GROSS and/or NET volume delivered. If the NO-FLOW TIMER
is not set to 0, the delivery will end if the timer times out between drops.
Preset may also be controlled directly from the LCR-II. This feature is described on Page 34.
The DELIVERY & PRESET menu group is the only one needed to make preset deliveries.
#^ (PRODUCT)
This field is used to select the product type that will be delivered. Scroll through numbers 1-4 and
select the product.
CODE
The code associated with the product name and type is displayed here and can be changed.
PRODUCT NAME
The product name associated with the product code and type is displayed here and can be changed.
26
PRESET DELIVERY WITH A LAP PAD CONTINUED
Delivery & Preset - Screen 2
UNITS
GALLONS
NET COUNT
0.0
GROSS COUNT
0.0
GALLONS
Delivery & Preset - Screen 3
UNITS
NET COUNT
REMAINING
----------
PRESET
0.0
0.0
0.0
0.0
----------
GALLONS
GROSS COUNT
PRESET
REMAINING
UNITS
Delivery & Preset - Screen 3
PRODUCT TYPE^
This field displays the product type previously selected in CALIBRATION.
ii
Press
i to advance to:
ii
GROSS COUNT
This field displays a running total of the gross number of units for the current delivery. DISPLAY
ONLY.
NET COUNT
This field displays a running total of the net number of units for the current delivery. DISPLAY ONLY.
UNITS^
This field displays delivery units selected during CALIBRATION: GALLONS, LITRES, CUBIC M,
LBS, KGS, BARRELS, and OTHER.
Lap Pad Basic Operation
ii
Press
i to advance to:
ii
ii
Press
i to advance to:
ii
(NOTE: This screen displays only if "Preset" is set to NET or BOTH in General Setup - Screen 3 and the
current product is temperature compensated).
NET COUNT
This field displays a running total of the net number of units for the current delivery. DISPLAY ONLY.
PRESET
The NET PRESET amount can be entered here. A preset option of NET or BOTH must have been
selected in GENERAL SETUP.
REMAINING
This field displays the number of units remaining for the current delivery. DISPLAY ONLY.
UNITS^
This field displays delivery units selected during CALIBRA TION: GALLONS, LITRES, CUBIC M, LBS,
KGS, BARRELS, and OTHER.
(NOTE: This screen displays only if "Preset" is set to GROSS or BOTH in General Setup - Screen 3 and the
current product is temperature compensated).
GROSS COUNT
This field displays a running total of the gross number of units for the current delivery. DISPLAY ONLY.
27
Lap Pad Basic Operation
Delivery & Preset - Screen 6
TOTAL
0.000
TAX
0.000
SUBTOTAL
0.000
Delivery & Preset - Screen 7
AUX OUT 2^
FLOW DIRECTION
AUX OUT 1^
ON DURING DELIVERY
Delivery & Preset - Screen 5
PRICE/UNIT
0.000
%TAX
0.000
TAX/UNIT
0.000
PRESET DELIVERY WITH A LAP PAD CONTINUED
PRESET
The PRESET amount can be entered here. A preset of GROSS or BOTH must have been selected in
GENERAL SETUP.
REMAINING
This field displays the number of units remaining for the current delivery. DISPLAY ONLY.
UNITS^
This field displays delivery units selected during CALIBRA TION: GALLONS, LITRES, CUBIC METERS,
LBS, KGS, BARRELS, and OTHER.
ii
Press
i to advance to:
ii
PRICE/UNIT
Enter value.
TAX/UNIT
Enter unit tax if applicable.
ii
Press
i to advance to:
ii
ii
Press
i to advance to:
ii
%TAX
Enter % Tax rate if applicable.
SUBTOTAL
Price per unit times quantity . DISPLAY ONLY.
T AX
Total of all taxes for delivery. DISPLAY ONLY.
TOT AL
Sum of SUBTOTAL plus T AX. DISPLAY ONLY.
AUX OUT 1^
This field activates optional outputs, e. g. pump control, additive injection, alarms, etc. Choices are
OFF, ON, ON DURING DELIVERY or MONITOR FLOWRATE. MONITOR FLOWRATE becomes
active at the beginning of each delivery. If the flowrate meets or exceeds 40 units/time, the output
deactivates. If the flowrate does not meet or exceed 40 units/time, the output remains active.
AUX OUT 2^
This field can be programmed to indicate FLOW DIRECTION or it can be programmed similar to the
AUX OUT 1^ functions of OFF, ON, or ON DURING DELIVERY.
28
Diagnostics/Factory Calibration
Diagnostics - Screen 1
SUPPLY VOLTAGE
13.5
DIAGNOSTIC MESSAGES^
TVC ACTIVE
Diagnostics - Screen 2
TICKET
ST200v1.05
LANGUAGE
SL200v1.04
SOFTWARE
SR210v1.00
Diagnostics - Screen 3
RATE BASE^
PER MINUTE
UNITS^
GALLONS
RATE
0.0
GROSS COUNT
0
DIAGNOSTICS
Most of the DIAGNOSTICS fields are for DISPLA Y ONL Y. They contain information that can be useful in troubleshooting.
DIAGNOSTIC MESSAGES^
This field allows an operator to check pending diagnostic messages. Press ENTER and the cursor will
ii
i to scroll through the list of messages. The last message
ii
ii
Press
i to advance to:
ii
drop to the second line of the screen. Press
is "END OF LIST".
SUPPLY VOLTAGE
The field shows the current supply voltage. The correct operating voltage is +9 to +28 VDC. DISPLAY
ONL Y. NOTE: Voltage above +17 VDC will not indicate correctly.
ii
Press
i to advance to:
ii
SOFTWARE
The software number is shown here. This number is critical for any troubleshooting discussions with the
Liquid Controls factory service personnel. DISPLAY ONLY.
LANGUAGE
The software number designating language selection (English, S panish, etc.) is displayed here. DISPLA Y
ONLY.
TICKET
The software number designating the ticket format is displayed here. DISPLAY ONLY.
GROSS COUNT
This field displays a running total of the gross number of units for the current delivery. DISPLAY ONLY .
RA TE
This field displays the actual flow rate. DISPLAY ONLY.
UNITS
This field displays the units of measure for flow total.
RATE BASE
ii
Press
i to advance to:
ii
This field displays the time base for flow rate.
29
Diagnostics/Factory Calibration
Diagnostics - Screen 4
PULSR FAULTS
0
R 1 00 .0
- 1 05 .2
R A W A DC
7 9 6
R 1 28 .6
2 8 84 .7
R T D OF S
2 . 60 5
R T D SL P
0 . 02 47 5
F a ct o ry C a li b ra ti o n - S c r e e n 1
7200015NO
Factory Calibration - Screen 2
LCR RESET^
USER KEY
SERIAL ID
DIAGNOSTICS CONTINUED
PULSR FAULTS
This field displays the number of quadrature pulser faults detected during the last delivery. DISPLAY
ONL Y.
FACTORY CALIBRATION (ONLY ACCESSIBLE BY LC FACTORY PERSONNEL)
R100.0
Used by qualified personnel to calibrate the temperature probe.
R128.6
Used by qualified personnel to calibrate the temperature probe.
ii
Press
i to advance to:
ii
RAW ADC
This is the raw value being returned by the A to D converter used by the temperature probe and can be
used to assist qualified personnel in troubleshooting.
RTD SLP
The RTD Slope is displayed here. This value can assist qualified personnel in troubleshooting.
RTD OFS
The RTD Offset is displayed here. This value can assist qualified personnel in troubleshooting.
SERIAL ID
The LCR-II printed circuit board serial number is shown here.
USER KEY
This is where the owner/office password is entered or viewed. When the system is locked, the operator
will be prevented from changing field entries with the exception of GROSS and NET PRESET, PRODUCTCODE, PRODUCT DESCRIPTION and NO-FLOW TIMER. If the password is lost or forgotten, it can
be recovered by the factory in this screen.
LCR RESET
This is a factory-entered value that will assist qualified personnel in clearing out memory problems,
should they arise.
NOTE: A complete flow chart of the Lap Pad Menu structure appears in Appendix D.
30
Recalibration/Proving without a Lap Pad
000.000
SINGLE POINT RECALIBRATION/PROVING
The addition of two control buttons on the front panel of the LCR-II allows for single point recalibration or proving
without the need of a Lap Pad or other data entry device. This calibration is limited to recalculating the k-Factor or
single point recalibration for up to four different product types. These two buttons can be used to select different
display values as well as enter product selections and preset values. These buttons are labeled SELECT and INCREASE.
If a menu item can be edited, the INCREASE button will be active and can be used to increase a value or scroll through
a list of options. The SELECT button will be used to scroll through consecutive digits across the display (from left to
right).
STEP 1: REMOVING THE LCR-II SWITCHPLATE FOR INITIAL SET-UP
•Remove the lead seal and wire assembly from the LCR-II Switchplate (if previously sealed by Weights & Measures
authorities).
•Remove the four screws securing the nameplate over the red Selector Switch located on the front of the LCR-II and
remove the plate.
•Rotate the Selector Switch counter-clockwise to the 6 o'clock (CALIBRA TION/PROVER) position to permit access
to the LCR-II calibration menu.
STEP 2: CALIBRATION/PROVER Position
When the switch is rotated to the CALIBRATION/PROVER Position, the initial display is Prover
Quantity.
Calibration - Screen 1, Prover Quantity
The units of measure are specified on the face label of the LCR-II. If a delivery is NOT active, press INCREASE to start
a new delivery. If a delivery is active, press INCREASE to stop the delivery and advance to Calibration - Screen 5,
Editable Prover Quantity (described on Page 31).
31
Recalibration/Proving without a Lap Pad
Prod 1
F 70.0
23.8 ºF
101.323
SINGLE POINT RECALIBRATION/PROVING CONTINUED
Press SELECT to advance to:
Calibration - Screen 1, Product Number
The current active Product Number is displayed. If a delivery is not active, the digit will be flashing, indicating
that it can be changed. Press INCREASE to scroll through the options. These options are labeled Prod 1,Prod 2, Prod 3, and Prod 4. When the desired option has been chosen, press SELECT. This will advance the
display to Calibration - Screen 3, Flow Rate.
NOTE: If the display is NOT flashing, it cannot be changed. Press INCREASE to stop delivery and advance to
Calibration - Screen 5, Editable Prover Quantity.
Calibration - Screen 3, Flow Rate
This displays the current flow rate through the meter. The display will update every ½ second. The unit of
measure is not displayed on the LCD, but appears directly below the display on the face of the LCR-II. This
screen is DISPLA Y ONL Y. If a delivery is active, press INCREASE to stop delivery and advance to Calibration
- Screen 5, Editable Prover Quantity.
Press SELECT to advance to:
Calibration - Screen 4, Temperature
This display shows the current temperature of fluid moving though the meter. Values will be shown in degrees
Celsius or Fahrenheit. The display updates every ½ second. DISPLAY ONLY. If a delivery is active, pressINCREASE to stop delivery and advance to Calibration - Screen 5, Editable Prover Quantity.
Press SELECT to return to Calibration - Screen 1, Prover Quantity.
Calibration - Screen 5, Editable Prover Quantity
The Prover Quantity is displayed with the most significant digit flashing. The most significant digit appears in
the left most position of the display . If the two most significant digit s are 98 or 99, a leading 0 will be displayed.
To change the value of this digit, press INCREASE. Each time INCREASE is pressed, the digit increases by a
value of 1. Once the number 9 has been reached, pressing INCREASE again will roll the digit back to 0. Once
the desired number is chosen, press SELECT to move one digit to the right.
32
Recalibration/Proving without a Lap Pad
101.663
SINGLE POINT RECALIBRATION/PROVING
CONTINUED
Repeat the process above for each digit position until
the value for each has been selected. When the least
significant digit is entered, the k-Factor is recalculated
and the display returns to Calibration - Screen 1.
NOTE: If the new value entered is out of range for the
LCR-II, "Error" will appear in the display. If this occurs,
press SELECT to return to the Prover Quantity display.
Repeat the process by beginning a new prover run.
When this process is complete, turn the Selector Switch
clockwise out of the Calibration Position and replace
the Switch plate.
NOTE: Step-by-step instructions on conducting a
recalibration/proving run appear in Appendix E.
CALIBRATION MENU
CALIBRATION - SCREEN 1
101.323
SELECT
CALIBRATION - SCREEN 2
Prod 1
CALIBRATION - SCREEN 3
F 70.0
INCREASE
“Current
Delivery
Active”
SELECT
CALIBRATION - SCREEN 4
27.2 ºF
SELECT
CALIBRATION - SCREEN 5
101.663
SELECT
ERROR SCREEN
SELECT
Error
If an unacceptable
value is entered.
33
Basic Operation without a Lap Pad
2345.6
F 70.0
DELIVERY & SELECTIONS MENU
Once the LCR-II has been calibrated, a Lap Pad is not required for daily use of the register. There are two push buttons
located on the LCR-II. These can be used to select different display values as well as enter product selections and
preset values. These buttons are labeled SELECT and INCREASE.
Values displayed or entered depend on the position of the red Selector Switch. The switch plate is divided into two
categories, DELIVERY and SELECTIONS. When the Selector Switch is set to RUN on the DELIVER Y side, it is set toFlowrate/T emp on the SELECTIONS side. By pressing the SELECT button, the menu on the SELECTIONS side is
accessed.
NOTE: Although unlabeled, each switch position displays the delivery total as its first display option.
If a menu item can be edited, the INCREASE button will be active and can be used to increase a value or select an item
from a list of options. The SELECT button will be used to advance to the next display or scroll through consecutive
digits across the display (from left to right).
The LCR-II display characters are seven segments. An all segments display appears as follows:
All Segments Display
RUN Position (Flowrate/Temp.)
A typical delivery is initiated by moving the Selector Switch to the RUN position. The LCR-II will perform a series of
diagnostics tests, reset the counter and activate the valve control output(s), opening the security valve. The control
output will remain active until the Selector Switch is moved to the STOP or PRINT position, the NO-FLOW TIMER
"times out", or a preset delivery volume is reached. A minimum amount of product flow must be detected (1 gallon or
5 litres) before the NO-FLOW TIMER starts counting. Delivery Total is the initial display.
Delivery Total Display
This display reflects the total for the current delivery. The display updates every 15.625 ms.
Press SELECT once to display Flowrate.
Flowrate Display
This displays the current flow rate through the meter. The display will update every ½ second. The unit of
measure appears directly below the display on the face of the LCR-II. After 5 seconds the display will return to
Delivery Total. This screen is DISPLAY ONLY.
34
Basic Operation without a Lap Pad
23.8 ºF
Prod 1
00200.0
00200.0
00000.0
DELIVERY & SELECTIONS MENU CONTINUED
RUN Position (Flowrate/Temp.) Continued
Press SELECT a second time (within 5 seconds) to display Temperature.
Temperature Display
This display shows the current temperature of fluid moving though the meter. Values will be shown in degrees
Celsius or Fahrenheit, depending on the initial set-up selection. The display updates every ½ second. After 5
seconds, the display will return to Delivery Total. This screen is DISPLAY ONLY.
Pressing SELECT a third time will return the display to Delivery Total. The INCREASE button has no function in the
RUN position.
STOP Position (Product #/Preset)
The STOP Position will cause the valve control output to close the security valve. This provides a means to pause the
delivery . The valve will remain closed until the switch is returned to the RUN position. If the switch is returned to theRUN position and a ticket has not been printed, the amount for the delivery will NOT be reset. The counter will continue
to display the accumulated total. If a NO-FLOW TIMER timeout is reached, the ticket will be printed, ending this
delivery cycle. Delivery Total is the initial display. The display will update every 15.625 ms.
Press SELECT to advance to Product Number.
The current active Product Number is displayed. If a delivery is not active, the digit will be flashing, indicating
that it can be changed. If the display is NOT flashing, it cannot be changed. Press INCREASE to scroll through
the options. These options are Prod 1, Prod 2, Prod 3, or Prod 4, depending on the number of products
programmed during calibration. When the desired option has been chosen, press SELECT.
The display will advance to Preset.
The preset amount is displayed with the most significant digit flashing. This will be the digit to the far left of the
display . To change the value of this digit, press INCREASE. Each time INCREASE is pressed, the digit
increases by a value of 1. Once the number 9 has been reached, pressing INCREASE again will roll the digit
back to 0. Once the desired number is chosen, press SELECT to move one digit to the right.
Repeat the steps above to change the value of each digit. Press SELECT to advance through the digits of the
display until the desired value is reached.
Once the final digit is entered, press SELECT to return to the Delivery Total. Or, if the preset number is as
desired, switch to RUN to begin delivery.
35
Basic Operation without a Lap Pad
H12345
11:44:00
P00000
Prod 1
L6789.0
DELIVERY & SELECTIONS MENU CONTINUED
PRINT Position (Product #/Totalizer)
Turning the Selector Switch to the PRINT Position will cause the valve control outputs to turn off, closing the security
valve. A ticket will be printed. Once the PRINT position has been entered, the delivery cannot be resumed. If tickets
are required, the system will NOT allow another delivery to begin until the current ticket is printed. Once the ticket is
printed, another delivery will be permitted when the Selector Switch is turned to the RUN position. Delivery T otal is the
initial display. The display will update every 15.625 ms.
Press SELECT to advance the display to Product Number:
The current active Product Number is displayed. If a delivery is not active, the digit will be flashing, indicating
that it can be changed. If the display is NOT flashing, it cannot be changed. Press INCREASE to scroll through
the options. These options are Prod 1, Prod 2, Prod 3, or Prod 4, depending on the number of products
programmed during calibration. When the desired option has been chosen, press SELECT.
The display will advance to:
This displays the five most significant digits of the GROSS or NET grand totalizer. Press INCREASE to return
to the Product Number display. Press SELECT to advance to:
This displays the five least significant digits of the GROSS or NET grand totalizer. The grand totalizer is a
combination of these two displays. The total in this example is "123,456,789.0" units. Press INCREASE to
return to the previous totalizer display. Press SELECT to advance to Time of Day:
If the unit is locked (set by the Lap Pad), a password needs to be entered to change the Time of Day. If the unit
is unlocked, pressing INCREASE will cause the first digit on the left of the display to begin flashing on and off.
Use INCREASE to adjust each digit of the time. The time is entered in military format, with 14:00:00 indicating
2:00 PM. When the final digit is programmed, press SELECT to return to Delivery Total. If a password is
required, the next screen on Page 36 will be displayed (by pressing INCREASE) before the time can be
changed.
Beginning with the digit to the immediate right of "P", use INCREASE to adjust it from 0-9. Once this digit is
selected, press SELECT to move one digit to the right. Once the complete, correct password has been
entered, pressing SELECT will allow for the adjustment of the Time of Day.
NOTE: If the unit is password protected as programmed by the Lap Pad, the password must be numeric only,
and contain exactly 5 digits.
36
Basic Operation without a Lap Pad
Prod 1
L6789.0
H12345
DELIVERY & SELECTIONS MENU CONTINUED
PRINT Position (Product #/Totalizer) Continued
A shif t ticket is printed when the Selector Switch is placed in the SHIFT PRINT Position for more than 2 seconds and a
printer is loaded with paper and available to print the ticket. (NOTE: If the Selector Switch is turned to SHIFT PRINT
and then out of SHIFT PRINT in less than 2 seconds, a Diagnostic/Calibration Information ticket is printed). DeliveryTotal is the initial display. The display will update every 15.625 ms.
Press SELECT to advance to Product Number Display:
The current active Product Number is displayed. If a delivery is not active, the digit will be flashing, indicating
that it can be changed. If the display is NOT flashing, it cannot be changed. Press INCREASE to scroll through
the options. These options are Prod 1, Prod 2, Prod 3, or Prod 4, depending on the number of products
programmed during calibration. When the desired option has been chosen, press SELECT.
The display will advance to:
This displays the five most significant digits of the GROSS or NET Shift totalizer. Press INCREASE to return
to the Product Number display. Press SELECT to advance to:
This displays the five least significant digits of the GROSS or NET Shift totalizer . The shift totalizer is a combination
of these two screens. The total in this example is “543,216,789.0” units. Press INCREASE to return to the
previous totalizer display. Press SELECT to return to Delivery Total.
Printing Diagnostic/Calibration Tickets
If a delivery was terminated due to a pulser failure, temperature probe failure or other fault, the last line on the delivery
ticket will read, "*FAULT: PRINT DIAGNOSTICS TICKET*".
A Diagnostic Ticket is printed when the Selector Switch is turned to the SHIFT PRINT Position for less than two
seconds. This ticket contains the same information found on the calibration ticket that prints when the Calibration
Mode is exited, but does not increment the calibration number or require a Weights & Measures seal to be broken.
This allows the operator to spot check the calibration of the register and/or diagnose problems with the unit. The ticket
will show any problems on the last line to assist in troubleshooting the problem.
37
LCR-II Menu Flow Charts
Selector SwitchSTOP Position of Selector Switch
DELIVERY TOTAL
22345.6
SELECT
PRODUCT NUMBER
RUN Position of Selector Switch
DELIVERY TOTAL
22345.6
SELECT
FLOWRATE
F 70.0
SELECT
TEMPERATURE
SELECT
62.9 ºF
Prod 1
SELECT
PRODUCT NUMBER
Prod 2
SELECT
PRODUCT NUMBER
Prod 3
SELECT
PRODUCT NUMBER
Prod 4
INCREASE
INCREASE
INCREASE
INCREASE
38
SELECT
00200.0
SELECT: Step through digits.
INCREASE: Increment each position 0-9.
SELECT
PRESET
LCR-II Menu Flow Charts
PRINT Position of Selector SwitchSHIFT Position of Selector Switch
DELIVERY TOTAL
DELIVERY TOTAL
22345.6
22345.6
SELECT
PRODUCT NUMBER
Prod 1
INCREASE
SELECT
Prod 4
GROSS OR NET TOTALIZER
H12345
SELECT
GROSS OR NET TOTALIZER
L6789.0
PRODUCT NUMBER
INCREASE
INCREASE
(Scroll Through
Products 1-4)
INCREASE
SELECT
PRODUCT NUMBER
Prod 1
SELECT
GROSS OR NET TOTALIZER
H12345
SELECT
GROSS OR NET TOTALIZER
INCREASE
INCREASE
PRODUCT NUMBER
(Scroll Through
Products 1-4)
Prod 4
INCREASE
INCREASE
SELECT
(When the last
digit is entered)
SELECT
TIME OF DAY
12:00:00
INCREASE
INCREASE
(If System is
“Locked”)
(Adjust each digit
0-9)
PASSWORD
P00000
INCREASE
SELECT
(After Password
is entered)
L6789.0
39
LCR-II Troubleshooting Guide
To comply with Weights & Measures requirements, it is necessary to start and stop each delivery with a fully packed
hose. Normally , this will be the case. However, there are times where the hose is not fully packed (e. g. after a preset
delivery). As such, the hose must be packed and the register zeroed prior to making the next delivery.
NOTE: This procedure will not work for an empty or dry hose. If more than 1 gallon or 5 litres of liquid is
required to pack the hose, a delivery ticket must be printed.
To pack the hose at the start of a delivery:
1. Turn the Selector Switch to RUN. DO NOT OPEN THE HOSE NOZZLE. The display will reset and the
solenoid will open the valve, packing the hose.
2. When the display has stopped incrementing, the hose is packed.
3. Then the Selector Switch to STOP. This will de-energize the solenoid and close the valve.
4. Turn the Selector Switch back to RUN. The display will reset to zero, and the solenoid will energize again
and open the valve.
5. Open the hose nozzle and begin the delivery.
6. The Selector Switch can only be turned to STOP one time per delivery, at the start of the delivery, to pack
the hose.
7. To end a delivery with a packed hose, complete the delivery by closing the hose nozzle prior to turning the
Selector Switch to STOP. If making a preset delivery where the solenoid closes the valve prior to the hose
nozzle being closed, the above noted procedure to pack the hose must be conducted on the next delivery .
LCR-II Troubleshooting Guide
The LCR-II troubleshooting guide covers most common problem situations. Call your local Liquid Controls Service
Center or the Liquid Controls Electronic Customer Service department for any problems that are not covered in this
guide.
NOTICE
It may be necessary to break Weights & Measure seals
to perform certain troubleshooting steps in this guide.
Contact your local supervisor or local Weights &
Measures authorities for information before breaking
any seals.
Serious Injury or Death may Result - Fire, Explosion and
Electrical Shock Hazard.
Only trained and authorized personnel should perform
troubleshooting. Use extreme caution when
troubleshooting and taking all electrical measurements.
Be sure that equipment is in a well ventilated area and
that hazardous or flammable vapors are not and will not
be present during testing. If equipment is located in a
hazardous environment, it may be necessary to remove
the equipment and perform troubleshooting in a safe area.
!!
! DANGER
!!
40
LCR-II Troubleshooting Guide
1. With the key in the accessory position, check the battery
voltage to the circuit board at J6. Use terminal #11 as positive
and terminal #12 as DC ground. While the LCR-II will powerup at +9VDC, it is recommended that the input be at least
+12.6VDC. Ref
2.
Check the 7.5A, in-line fuse (Part # 70985) for continuity. It is
located in the accessory power line. Replace if necessary.
1. For safety reasons, remove the 7.5A, in-line fuse from the
accessory power cable.
2. Remove the 3-pin power connector J6. Inspect for stray wire
stands and visible shorts.
3. Inspect the full length of the power cable. Look for damaged
insulation, which may cause shorting between the cable and
ground potential (e. g. truck frame or chassis). If the power
cable (Part # 81512) is damaged, it must be replaced.
4. Replace the 7.5A fuse (Part # 70985) and re-install the J6
connector.
5. If the 7.5A fuse blows again, replace the LCR-II circuit board.
6. If the 7.5A blows after the LCR-II circuit board has been
replaced, then replace the power cable (Part # 81512).
Unit will not
power up or
there is no
display.
Inadequate supply voltage.
+9 to 28VDC is required for
operation.
Unit blows 7.5A
fuse.
+12VDC Battery line is
shorted to ground
PROBLEMPROBABLE CAUSESOLUTION
NOTE: Check for proper operating voltages before changing the circuit board. If the circuit board needs to be changed,
be sure to remove all power to the LCR-II.
5. If an error occurs during a delivery, the message "*FAULT: PRINT DIAGNOSTICS TICKET*" will appear on the
delivery ticket. The diagnostic ticket will contain an error message that can be useful in troubleshooting. Print adiagnostic ticket by moving the red Selector Switch to "SHIFT PRINT" for less than two seconds. Examine the
diagnostic ticket to make sure all set-up fields are accurate, e.g. pulses per unit volume, temperature coefficient
and base temperature.
6. NEVER remove a terminal block or jumper with the power on.
7. NEVER install a terminal block or jumper with the power on.
8. NEVER force a terminal block into its location.
9. NEVER exchange or reposition terminal blocks on the circuit board.
10. In case of a major problem such as a burned or water-damaged circuit board, evaluate possible causes before
replacing it and turning the power back on.
11. Isolate the problem before changing the circuit board.
12. Return faulty circuit boards with the proper forms, concisely completed.
There may be several probable causes for a system malfunction. Listed on the next four pages are several probable
causes to aid in returning the system to operation as quickly as possible. The list is not all-inclusive and should only be
used as a guide.
41
LCR-II Troubleshooting Guide
Power to LCR-II interrupted
during delivery.
1. Check accessory power cable for damage. Ensure that the
power, common and ground wires (11, 12 and 13) on power
connector J6 are secure. J6 is located on the rear of the
circuit board.
Static discharge.
2. Turn on all truck accessories (headlamps, 2-way radio,
heater, etc.). Engage the hose reel and monitor the DV
voltage at J6 using terminal #11 as positive and terminal #12
as ground. If the voltage drops below +10VDC, the truck
electrical system may not
3. Verify proper grounding of the LCR-II. Refer to the LCR
Installation Manual for proper grounding procedures.
Epson printer
release light
flashes
Low voltage to the Epson
printer.
1. Check the battery voltage for a minimum of +12.6VDC.
2. Under extreme cold conditions, the printer may not operate.
Warm up the cab of the vehicle.
3. If the release light continues to flash, replace the Epson 295
Printer (Part # E49001).
No power
indicator lights to
the Epson
printer.
No power to the Epson printer.1.
Verify that the power switch is in the ON position. This switch
is located on the left-hand side of the Epson 295 printer.
2. Check the printer power cable (part # 825001) to ensure that
it is seated properly. If the problem persists, replace the
power cable followed by the Epson printer (Part # E49001).
1. Check the RTD Probe and terminal block for continuity.
2. Remove terminal block J14 from the circuit board. On the
terminal block, measure and record the resistance between
the following pins:
PIN # PIN# READING
21 to 22 100 ? ± 20?
20 to 23 100? ±20?
3. If the readings are not within the above tolerances, replace
the RTD probe (Part # 71130).
“Temperature
Error” appears on
diagnostic ticket
or on the Lap
Pad.
Open or shorted circuit between
the RTD probe and LCR-II.
PROBLEMPROBABLE CAUSESOLUTION
“Power Failure”
appears on Lap
Pad or diagnostic
ticket.
42
LCR-II Troubleshooting Guide
Pulser failure.2. If the LCR-II display counter increments, this may indicate a
mechanical problem. Contact your local Liquid Controls
Service Center or the Liquid Controls Service Department for
assistance.
3.
If the LCR-II display counter does not increment, see Pulser
Failure
.
Missing pulse counts.1. Check the pulser output.
Excessive reversals.
2. Open the LCR-II cover by loosening the single screw that
holds the cover closed. Using a reliable multimeter, measure
the following DC voltages on terminal block J8, while J8 is
connected to the circuit board. Use terminal #37 as ground
reference.
TERMINAL # VOLTAGE
J8 #32 +5 VDC
J8 #33 0
or
+5VDC
J8 #34 0
or
+5VDC
3. If the following voltages are observed, check for loose pulser
wiring connections. If no loose connections are found,
replace the encoder harness.
(Part # 81584 – for Series E2611X and E2612X)
(Part # 81412 – for Series E2605X and E2606X) or
4. Insert a ticket into the Epson printer. Begin a delivery. With
product flowing through the meter, measure the following DC
voltages on the terminal block.
TERMINAL # VOLTAGE
J8 #32 0 VDC
J8 #33 +1-3VDC
J8 #34 +1-3VDC
5. If the voltages shown below are observed, ensure that the
pulser shaft is rotating. Look for any mechanical problems
that may cause either the meter or the pulser shaft to lock up.
If the Lap Pad continues to indicate a pulser failure, replace
the pulser.
(Part # 81582 for Series E2611X and E2612X)
(Part # 81170 for Series E2605X and E2606X) or
(Part # 81159 for POD pulser)
TERMINAL # VOLTAGE
J8 #32 +5 VDC
J8 #33 0
or
+5VDC
J8 #34 0 or +5VDC
Pulser shaft is not turning with
product flow.
Manually spin the pulser shaft and monitor the LCR-II display.1.Product flow
does not register
on Lap Pad or
LCR-II display
“Pulser Failure”
appears on the
diagnostic ticket
or on the Lap
Pad display.
NOTE: Pulser faults generally
occur in a high vibration
environment. Liquid Controls is
not responsible for pulser
failures caused by excessive
system vibrations.
SOLUTIONPROBABLE CAUSEPROBLEM
43
LCR-II Troubleshooting Guide
1. Insert a delivery ticket into the Epson printer and start a
delivery.
Foreign debris in system.2.
Move the selector switch from RUN to STOP and back to
RUN. Listen for an audible clicking sound from the solenoids.
3. If there is an audible click from the solenoid (but still no flow),
this may be an indication of a mechanical problem with the
main valve or its associated components. Contact your local
Liquid Controls Service Center or the Liquid Controls Service
Depar
4. If there is NO audible click from the solenoid, check the
voltage to Pin #s 14, 15, 16 and 17 of terminal block J13.
While still in RUN
mode, use a reliable multimeter to measure
the following DC voltages on the circuit board. Use J8 pin
5. If the above voltages are correct, this may be an indication of
a problem with the valve or its associated components.
Contact your local Liquid Controls Service Center or the
Liquid Controls Service Department for assistance.
6.
If the above voltage are NOT correct, replace the following
components with known working units until the problem is
resolved:
Solenoids* Varies depending on system.
LCR-II Circuit Board* Part # 81547-1 for series E3611X and
E3612X
*Contact your local Liquid Controls Service Center or the
Liquid Controls Service Department for assistance.
Valve will not
open.
Solenoids are inactive or
inoperative.
SOLUTIONPROBABLE CAUSEPROBLEM
44
Appendix A - Compensation Types & Parameters
General
LinearLinear0 to 0.003 ºC15-90N/A+100
General
LinearLinear0 to 0.005ºF60-130N/A+212
LPG
USA
LPG
Europe &
Cana da
Re fin ed
Petroleu m
Pro d u cts
Europe &
Cana da
Re fin ed
Petroleu m
Pro d u cts
USA
General
API Table 54CCoefficient
0.000486 to
0.001674
ºC15-50-40+95
Lube Oil
Europe &
Cana da
Ammonia
Cana da
+40
-50-40+95
15-30-30NH
3
N/AN/A
60
Density
kg/m3
API Table 54D800 to 1164ºC15
ºC
+95
-50-40+200API Table 6BAPI Gravity0 to 85ºF
API Table 54B653.0 to 1075.0ºC15-50-40
0.5 to 0.600ºC15-46-46+60
T
m ax
API Table 24
Specific
Gravity
0.5 to 0.550ºF60-50-50+140
Scale
ºC or ºF
Density
kg/m3
Density
kg/L
T
base
T
m in
T
hold
ProductV CF Type
Param eter
Coefficient
Range
API Table 54
The table below shows the supported methods of calculating the Volume Correction Factor (VCF). The method is
independently selectable for each product in Product Calibration. The table below shows the valid temperature
ranges for each table. The hold temperature, T
> T
. This is done to allow compensation in very cold environments where further extrapolation of equations is not
min
allowed.
, is the temperature that VCF is held at for temperatures < T
hold
hold
and
45
Appendix B - Operation Using VT 100 Compatible Terminal
OperationLap Pad KeystrokePC Terminal Emulation
Return to Top Level MenuM1<control L>
Move to Next Menu Item
<control D>
Move to Previous Menu Item
<control U>
Select ItemENTERENTER
Submit Data and ReturnENTER (after data entry)ENTER (after data entry)
Move to Next Field
<control D>
Move to Previous Field
<control U>
Clear Field/Start Data EntryClearBackspace
StartStart<control B>
StopStop<control S>
PrintPrint<control P>
Move to Next ScreenM#<control N>
If the LCR-II is operated using a VT 100 compatible terminal (instead of the LC Lap Pad), refer to the following chart for
the terminal keystrokes that correspond to the Lap Pad keystrokes. VT 100 communication parameters that must be
observed include:
•Baud Rate (9600)
•Data Bits (8)
•Parity (None)
•Stop Bits (1)
•Flow Control (None).
46
Appendix C - Error Messages
The LCR-II was designed to perform diagnostic analysis of problems as they occur. If an error should occur, one of the
following messages may be displayed on the Lap Pad or printed on a Diagnostic Ticket (see troubleshooting section).
The list below explains the error message and any possible corrective action. The LCR-II was designed to handle most
foreseeable problems on its own. Consequently, if the suggestions listed are not successful, the assistance of a
trained authorized Liquid Controls service provider is recommended.
RUN TIME MESSAGE LIST
These messages may be displayed on the Lap Pad if an error occurs while the system is running.
FLASH CRC FAILURE
The CRC stored with an LCR-II data block indicates the data or CRC has been corrupted. A "REBUILD"
operation could possibly fix the problem. If not, try "CLEAR ALL". If that does not work, hardware may
be the problem and the unit should be returned to Liquid Controls for more analysis. NOTE: Before
performing a "CLEAR ALL" operation, print a calibration ticket.
FLASH WRITE FAILURE
A failure was detected while a data block was being written to the flash memory. See the resolution
suggestion under FLASH CTC FAILURE.
FLASH NOT INITIALIZED
This error indicates that a "CLEAR ALL" or "REBUILD" has been attempted without the Selector
Switch in CALIBRATION position.
RANGE ERROR
The data entered is not within the legal range limits for the field being changed.
DUPLICATE FLOW RATE ERROR
This message only appears when setting up multi-point calibration. When calibrating a multi-point
meter, the linearization point s must all have unique flow rates. This message is displayed when a flow
rate being entered is already used on another point.
ADJACENT POINTS OUT OF 0.25% RANGE
This message only appears when setting up multi-point calibration. When calibrating a multi-point
meter, adjacent linearization points must be within a 0.25% correction of each other. To avoid this
error, select an additional flow rate and correction factor between the two points. More than one
additional point may be required. Adjacent flow rates entered for linearization cannot have their percent
error fields further than 0.25% apart.
COUNTER TEST IN PROGRESS…
A delivery has been started and the counter test is in progress. When the test is complete, flow will
start. This is not an error condition.
START ENABLED - SWITCH TO RUN TO BEGIN DELIVERY
The START button on the Lap Pad was pressed but the Selector Switch was not in the RUN position.
When the switch is put in the RUN position, the delivery will begin.
47
Appendix C - Error Messages
DIAGNOSTIC MESSAGE LIST
If an error occurs during a delivery , the message "*F AUL T : PRINT DIAGNOSTICS TICKET*" will appear on the delivery
ticket. The appropriate message below is then printed on the diagnostic ticket if an error has occurred. A diagnostic
ticket is printed by loading a ticket into the printer and then moving the Selector Switch to the SHIFT PRINT position for
less than two seconds.
PREVIOUS DELIVERY TICKET PENDING
A new delivery was attempted but the previous delivery ticket was not printed in its entirety. The PRINT
button should be pressed on the Lap Pad, or the Selector Switch should be placed in the PRINT
position. Once the previous delivery ticket is printed, a new delivery can begin.
ROM CHECKSUM ERROR
This message indicates that the program memory space of the LCR-II has been corrupted. If this error
occurs, the unit must be re-flashed with the control software before any deliveries can be made.
TEMPERATURE ERROR
This message appears if the temperature circuit returns an error or if the temperature calibration data
is in error. Check the temperature probe and it s connections. A broken temperature probe wire is the
most common cause of this error. See the troubleshooting section.
VCF DOMAIN ERROR
This message indicates that the calculated temperature falls outside the allowable range for the current
compensation parameter . V erify that the correct product compensation type (see PRODUCT SET -UP)
is being used.
PULSER FAILURE
This message appears if the number of pulser faults exceeds the allowable amount. The number of
allowable faults is determined by an equation and cannot be changed. A maximum of 5 times the
number of pulses required for the least significant displayed digit OR 0.1% of the pulses generated for
the current delivery are allowed. Pulser faults generally occur in high vibration environments or when
the flow rate is highly throttled. Liquid Controls is not responsible for pulser failures caused by excessive
system vibrations. See the troubleshooting section.
METER CALIB ERROR
A product that has not been calibrated has been selected for delivery . Select a valid calibrated product.
PRESET STOP
This message appears if the delivery has stopped due to a GROSS or NET PRESET amount being
reached. This is not an error condition.
NO FLOW STOP ERROR
This message appears if the NO-FLOW TIMER has expired, causing the current delivery to be
terminated. This is not an error condition. Entering "0" for this parameter in the GENERAL SET-UP
menu can turn off the NO-FLOW TIMER, however, this results in a system that is less secure.
STOP REQUESTED
This message appears when a STOP command has been sent to the LCR-II system to pause the
delivery. STOP can be activated from the Selector Switch, Lap Pad, or a Terminal. This is not an error
condition.
48
DIAGNOSTIC MESSAGE LIST CONTINUED
DELIVERY END REQST
This message appears when a PRINT command has been sent to the LCR-II system to terminate a
delivery. PRINT can be activated from the Selector Switch, Lap Pad, or a Terminal. This is not an error
condition
POWER FAIL ERROR
This message appears when a delivery was ended due to a loss of input voltage. Check the power
supply and wiring to the LCR-II. Check the in-line fuse. See the troubleshooting section.
PRESET ERROR
This message appears when a flash memory error is detected during a delivery when accessing any of
the fields used by PRESETS. This includes GROSS PRESET amount, NET PRESET amount, S1
CLOSE VALUE, remaining PRESET amounts, the PRESET type, or PRESET ALLOWED fields.
TERMINAL NOT CONNECTED
This message appears when a Terminal or Lap Pad is disconnected during a delivery. Check the
cables and connectors for broken wires and loose connections.
Appendix C - Error Messages
CHECK PRINTER AND CABLE
This message appears when the printer is not responding. This can be the result of a faulty data cable,
power cable, or LCR-II circuit board failure. The most common cause is that a ticket is not in place in
the printer. The Epson Slip printers have a built in photo eye to detect if a ticket is in position.
DATA ACCESS ERROR
This message appears when a flash memory error occurs.
DELIVERY TICKET PENDING
This message appears when a delivery has been terminated but a delivery ticket has not been completely
printed. When this occurs, it is assumed that the delivery is still active. No fields can be changed until
the delivery ticket has been entirely printed. Load a ticket into the printer and print the delivery ticket by
moving the Selector Switch to PRINT, or by pressing PRINT on the Lap Pad.
SHIFT TICKET PENDING
This message appears when a shift ticket is being printed. This is not an error condition.
FLOW ACTIVE
This message appears when a delivery has been initiated. This is not an error condition.
DELIVERY ACTIVE
This message appears when a delivery has been initiated. This is not an error condition.
GROSS PRESET ACTIVE
This message appears when a GROSS PRESET is active. The PRESET type must be set to GROSS
or BOTH and a desired amount entered in the GROSS PRESET field. This is not an error condition.
49
Appendix C - Error Messages
DIAGNOSTIC MESSAGE LIST CONTINUED
NET PRESET ACTIVE
This message appears when a NET PRESET is active. The PRESET type must be set to NET or
BOTH and a desired amount entered in the NET PRESET field. This is not an error condition.
GROSS PRESET STOP
This message appears when the GROSS PRESET amount has been reached. This is not an error
condition.
NET PRESET STOP
This message appears when the NET PRESET amount has been reached. This is not an error condition.
TVC ACTIVE
This message appears when TEMPERATURE COMPENSATION is active on the current product.
This is not an error condition.
S1 CLOSE REACHED
This message appears when the S1 CLOSE value has been reached during a PRESET delivery . This
is not an error condition.
INIT WARNING
This message appears when NO-FLOW TIMER or the ticket required fields return an error. If this
happens, the delivery can continue but the default values of 180 and YES are used.
END OF ERROR LIST
This message appears when the end of the error list has been reached. This is not an error condition.
50
GENERAL START UP
Appendix D - Lap Pad Menu Flow Charts
LectroCount LCR-II SR210v1.00
(c) Copyright Liquid Controls, Inc. 2002
DELIVERY & PRESET
Press Enter or Scroll with ARROW keys.
PRODUCT & SHIFT INFORMATION
Press Enter or Scroll with ARROW keys.
GENERAL SETUP
Press Enter or Scroll with ARROW keys.
SYSTEM CALIBRATION
Press Enter or Scroll with ARROW keys.
Displays for 2 Seconds
See
DELIVERY & SETUP Menu
See
PRODUCT & SHIFT Menu
See
GENERAL SETUP Menu
See
SYSTEM CALIBRATION Menu
PRODUCT CALIBRATION
Press Enter or Scroll with ARROW keys.
DIAGNOSTICS
Press Enter or Scroll with ARROW keys.
SECURITY
Press Enter or Scroll with ARROW keys.
51
See
PRODUCT CALIBRATION Menu
See
DIAGNOSTICS Menu
See
SECURITY Menu
Appendix D - Lap Pad Menu Flow Charts
DELIVERY & PRESET MENU
DELIVERY & PRESET
Press Enter or Scroll with ARROW keys.
ENTER
#^ CODE PRODUCT NAME PROD TYPE^
1
M#M1
GROSS COUNT NET COUNT UNITS^
0.0 0.0 GALLONS
M#M1
NET COUNT PRESET REMAINING UNITS^
0.0 0.0 --------- GALLONS
M#M1
GROSS COUNT PRESET REMAINING UNITS^
0.0 0.0 --------- GALLONS
M#M1
M1
Depending on the position of the CURSOR, the
ARROW Keys will do one of the following:
• Scroll Forward one position
• Scroll Backward one position
• Jump to the first position on the next screen
• Jump to the last position on the previous
screen, (as depicted below).
Cursor
Cursor
PRICE/UNIT TAX/UNIT % TAX
0.0000 0.0000 0.0000
M#M1
SUBTOTAL TAX TOTAL
0.00 0.00 0.00
M#M1
AUX OUT 1^ AUX OUT 2^
ON DURING DELIVERY FLOW DIRECTION
DIAGNOSTICS MESSAGES SUPPLY VOLTAGE
END OF LIST 12.7
M1
M#M1
SOFTWARE LANGUAGE TICKET
SR210v1.00 SL200v1.04 ST200v1.05
M#M1
GROSS COUNT RATE UNITS^ RATE BASE^
0.0 0.0 GALLONS PER MINUTE
M#
M1
SECURITY
FACTORY
PULSER FAULTS
0
M#
DIAGNOSTICS
Press Enter or Scroll with ARROW keys.
ENTER
USER KEY SECURITY^
LOCKED
R100.0 R128.6 RAW ADC RTD SLP RTD OFS
-87.2 2902.9 921 0.024748 2.157
M1
M#
SERIAL ID USER KEY LCR RESET^
8200054 NO
57
Appendix E - Weights & Measures Prover Guide
This guide has been created to clearly convey the method for proving a meter system equipped with a LectroCount
LCR-II electronic register without the use of a Lap Pad or external data entry device. This guide will only cover the
software features of the LCR-II in the calibration/prover mode.
NOTE: The following method provides a means to check the overall meter accuracy as well as adjust the
meter's "k-Factor". The k-Factor moves the entire meter accuracy curve either up or down. If it is believed that
the meter curve itself may have changed (e.g. when a measuring chamber is replaced), and it is desired to use
and/or change multi-point calibration, a Lap Pad or other data entry device must be used per the Set-Up &
Operation Manual.
The LCR-II must initially be set-up and calibrated using a Lap Pad or data entry device. Once this has been accomplished,
the LCR-II can be proved without using such a device. Only products with a valid pulse/unit (k-Factor) can be recalibrated
using this feature.
1. Remove the Switch Plate cover.
2. Turn the Selector Switch to the calibration/prover position. This is accomplished by rotating the Selector Switch
counter-clockwise to the 6:00 position.
3. The display will show "000.000". This represents the delivered volume as recorded by the register. The current
calibration settings will be printed on a ticket.
4. Press the SELECT button one time. The display will show "Prod 1". Press INCREASE to select from the programmed
products (Prod 1 to Prod 4). Only products that have been previously set-up and calibrated will be displayed. If
only one product has been calibrated, only one Prod # will be available.
5. When the desired product number is displayed, press SELECT. The display will advance to Flowrate, "F 0.0".
6. Press SELECT and the display will advance to Temperature, " 42.5 ºF". If a temperature probe is not installed, the
display will show "999.9 ºF".
7. Press SELECT and the display will return to "000.000".
8. To start a delivery, press the INCREASE button one time. This takes the display through an "all segments and
blanks" test and starts a delivery by activating the solenoid valve controls. (All segments and blanks: "8.8:8.8:8.8").
If the register display remains at zero, open the hose nozzle and begin to fill the prover. If the register did not
remain at zero (the hose needed to be packed), leave the pump running and press the INCREASE button to stop
the delivery. Then press the SELECT button until the display digits no longer flash. Press the INCREASE button
on the LCR-II to start the actual delivery with a packed hose. Open the hose nozzle and start to fill the prover.
9. Once the prover is filled to the desired volume, close the hose nozzle and press the INCREASE button on the LCRII to deactivate the solenoid controls and end the delivery.
10. If it is determined that no change to the meter k-Factor is desired, turn the Selector Switch clock-wise to exit the
calibration/prover mode. Proving is complete. If it is determined that a change to the meter k-Factor is desired, use
the INCREASE and SELECT buttons on the LCR-II to adjust the value on the register to match the volume of the
prover. The following page contains an example.
58
Appendix E - Weights & Measures Prover Guide
k-Factor Adjustment Example:
Presume that the LCR-II register displayed "101.235" Units and the prover volume was determined to be "100.593"
units at the end of step 9. ("Units" represents the units of measure as designated on the front cover of the LCR-II.)
When INCREASE is pressed in Step 9 above, the left most digit of the LCR-II register will be flashing on and off,
indicating that it can be changed. To change this value, press INCREASE to scroll through 0-9. The display digits will
wrap around to 0 after 9 has been reached. When the desired value for the first position has been reached, press
SELECT and the next digit to the right will flash on and off. In this example, no adjustment to the first digit is necessary .
The second digit also does not require adjustment, so press SELECT to advance to the third digit. The third digit, "1",
will be flashing on and off and needs to be changed to 0 in order to match the prover volume. Press INCREASE to
scroll through the digits until "0" is displayed. Press SELECT to accept that value and advance to the next digit, "2".
Press INCREASE until the number "5" is displayed. Press SELECT to advance to the next digit, "3". Continue in this
fashion until all the digits on the display are adjusted to the prover volume: "100.593". When this is complete, none of
the digits will be flashing on the display . A new k-Factor is recorded by the LCR-II. Turn the Selector Switch clock-wise
to exit calibration/prover mode.
If a number is entered in this example that results in an unacceptable k-Factor , the display will show "Error". When this
occurs, press SELECT and return to Step 7 and conduct another prover run.
59
LIQUID CONTROLS
105 Albrecht Drive
Lake Bluff, IL 60044
(847) 295-1050