21Leica M844 – Leica M820 / Ref. 10 713 294 / Version I
Preparation for operation
Checklist: Before the operation
Cleaning the optical accessories
➩Select the eyepieces, objective and the documentation
accessories (if used) and check them for cleanliness.
➩Remove dust and dirt (see page 58).
Checking accessories
➩Lock the swing arm.
➩Outfit the microscope with all accessories for use
(see pages 22-24).
➩Turn 0°-assistant’s attachment to the desired side
(Leica M844, see page 25) or install attachment for second
observer on the desired side (Leica M820).
Balancing
➩Release and balance swing arm
(see pages 33 and 34).
Function check
➩Switch the microscope on.
Warning 1
Motors return to their home positions
➩Before switching on the microscope, ensure that the
travel paths of the X- and Y-axes and the focus motor
are free of obstructions.
➩Check the Main Light 1, Main Light 2 and OttoFlex
TM
II
illuminators. Replace defective bulbs before the
operation begins.
➩Test all handswitch and footswitch functions.
➩Check the brake function using both the handles and
the remote brake release (see page 33).
Positioning at the OP table
➩Position the surgical microscope at the operating table as
desired and secure the brakes on the floor stand.
(see page 36).
Checking tube settings
➩Check the tube and eyepiece setting for the selected user
(see page 24).
➩Treat the eyepieces with an antifogging compound if
necessary.
Sterility
➩Fit sterile components and sterile drape if used
(see page 59).
Warning 2
Danger of fatal electric shock
➩Operate the system only with all equipment in its
proper position (all covers fitted, doors closed).