Leica DM LSP Instructions Manual

Leica DMLSP
Instructions · Bedienungsanleitung
Mode d’emploi
MICROSYSTEMS
UK
This manual is a supplement to the main DMLS manual, which is also supplied with every DMLSP polarizing microscope in English, French and German. The manual DM LS also contains all the technical data and safety infor­mation.
A special brief instruction manual for the DMLSP is also available in several languages.
Information about the range of objectives is given on the separate “OPTICS” data sheet.
D
Zusätzlich ist eine spezielle Kurzanleitung DMLSP in verschiedenen Sprachen verfügbar.
Über das Objektivprogramm informiert das ge­trennte Datenblatt „OPTIK“.
F
Cette notice est un complément du mode d’em­ploi complet Leica DMLS en français, allemand et anglais qui est également fourni avec chaque microscope polarisant Leica DMLSP. Le mode d’emploi du Leica DMLS contient des données techniques, des consignes de sécurité et des in­formations générales qui se rapportent au mi­croscope de base.
Diese Anleitung ist eine Ergänzung zur ausführli­chen Anleitung DMLS, die in Deutsch, Englisch und Französisch jedem Polarisationsmikroskop DMLSP ebenfalls beigefügt ist. Die Anleitung DMLS enthält darüber hinaus sämtliche techni­sche Daten, Sicherheitshinweise und Grundla­gen zum Mikroskop.
3
edition, issued in 2000 by/
3. Auflage, herausgegeben 2000 von/ Edtion 1998 par:
Leica Microsystems Wetzlar GmbH
Le mode d’emploi condensé est lui aussi dispo­nible en différentes langues.
Informations sur le programme objectif : voir feuilles de données séparées «OPTIQUE».
Ernst-Leitz-Strasse D-35578 Wetzlar (Germany)
Responsible for contents/ Verantwortlich für den Inhalt/ Département responsable du contenu: Marketing MQM, Product Management
Phone/Tel./Tél. +49 (0) 64 41-2922 80 Fax +49 (0) 6441-29 2255
Leica DMLSP
Instructions
MICROSYSTEMS
4
Contents
Pol components .................................................... 5
Adjustment.............................................................. 7
Operation of objectives (Pol) .............................. 8
Operation of transmitted light polarization ...... 9
Evaluation of conoscopy...................................... 14
Possible errors ...................................................... 15
Incident light techniques .................................... 15
EU conformity declaration .................................. 16
Text Symbols and their meaning:
Caution! Operation errors can dam­age the microscope and/or its acces­sories.
Not part of all configurations/option.
Numbers with an arrow, e.g. p. 20, refer to a particular page in this man­ual.
Numbers in brackets, e.g. (1.2), refer to illustrations, in this example Fig. 1, item 2.
*
p. 20
(1.2)
5
Pol components
Components
The microscope model DMLSP (Fig. 43) differs in the following points from the DMLS model described on pages 5 – 51. However, all points of the DMLS manual apply to the DM LSP as well as long as it is equipped with the relevant com­ponents. The following components are the only ones that cannot be used.
Filter magazine (Fig. 11)
Filter holder for 2 filters (Fig. 12). For combina-
tion with IC/P polarizer (43.10) only: the similar filter holder (43.11), although for one filter only, should be used instead.
Heating stages
Pol tubes
Polarization tubes have a groove on the under­side into which a guide pin in the Pol microscope stand and the Pol module engages, so that the cross line in the Pol eyepiece is oriented. For this purpose, the right-hand eyepiece mount has a snap-in groove and a mechanical compensa­tion to make sure that the cross line in the eye­piece remains oriented when the interpupillary distance is adjusted p. 28. The following tubes are available, on the outside they are only slight­ly different from the normal tubes illustrated in Fig. 35 and 37: explanation of abbreviations p. 25.
LMP--7 Monocular Pol tube
HC LBP 0/3/4 Binocular Pol tube
HC L1TP 4/5/7 Trinocular Pol tube with
1 beamsplitter switching position (intensity ratio of binocular viewing port: monocular photo/TV exit = 50 %:50 %).
HC L3TP 4/5/7 Trinocular Pol tube with 3 switching positions (intensity ration bin: ver­tical exit 100% :0 %/50:50%/0 :100%).
Ordinary tubes can also be used for polarization microscopy, although the orientation of the cross line in the eyepiece is not then guaranteed. In this case the disturbing bearing ring with latch in the Pol eyepiece can be removed with a very small screwdriver. If using additional graticules, e.g. for photomicrography, you should order a third eyepiece, as 2 different graticules in the left and right-hand eyepieces are irritating.
Pol eyepieces
On polarized light microscopes like the DMLSP, the (right-hand) eyepiece has a cross line which is aligned by a latch and corresponding groove in the right-hand tube. If aligned in a horizontal/ vertical position (Fig. 41/42) the crossline in­dicates the vibration direction of the polarizer (east – west) and the analyser (north – south). If the crossline is set at an angle of 45°, it indicates the vibration direction in a birefringent specimen when it is rotated to a diagonal position (= max. light intensity). There is also a scale for length measurement p. 42.
Pol module
The Pol module (intermediate tube, 43.3; 43.4) also has an orientation (see above). It consists of:
a switchable analyser (43.4), orientation north – south.
a switchable (43.3) and centerable (43.3a) Bertrand lens for conoscopy p. 10, with coupled pinhole diaphragm for conoscopic isolation of small specimen areas.
6
Quartz plate. This has a depolarizing effect. If the tube is mounted directly to the micro­scope, i.e. without the Pol module and there­fore without the quartz plate, anomalous inter­ference colours (pseudodichroism) may occur when the analyser is disengaged.
Centering nosepiece
All objectives of the quadruple centering nose­piece are centrable to the axis of rotation of the stage with the two centering keys (1.3; 43.15), p. 5.
Pol objectives
Objectives with the additional engraving P or POL are manufactured to be particularly low in strain. However, they may show considerable signs of strain if subjected to rapid changes in temperature or mechanical damage.
Pol rotary stage
Together with the centered objectives, the ball­bearing precision stage enables exact angle measurements.
Pol condensers
The condensers (43.8) also have to be strain-free for all examinations in polarized light. Therefore, only the special Pol condensers CLP/PH or UCLP
0.85 (identified by the code letter P and the aper-
ture value 0.85) can be used.
Polarizer
First the switchable polarizer (43.10), rotatable through 360° and with slot for λ or λ/4 compen­sator above (8.2) or, as a simple solution, the fil­ter holder (28.4) with slot-in polarizer can be used.
Analyser mount TL L 1/25 (option)
(not illustrated) In case of alternative outfits without Pol module,
the modular analyser mount TL L 1/25 (11505 121) can be directly interposed between stand and tube. This allows the use of fixed and rotatable analysers (180°) in slider.
Assembly
First read the general assembly instructions on page 5 – 21. The polarizer (43.10) is screwed onto the underside of the condenser holder at the left.
Caution!
When assembling the intermediate tube (Pol module, tubes) and the (right-hand) Pol eye­piece, remember to watch out for the orientation aids (pin and groove) at the joints.
Operation of transmitted light pol
Basically all information on pages 21 – 52 is valid. In addition, please observe the following specific measures for polarization.
7
Method II (Fig. 42)
Move the prominent point on the specimen (42a) to the centre of the crossline M. Rotate the stage until the point on the specimen is furthest away from the centre of the crosslines M (position A, Fig. 42b). Point A (= maximum distance of the specimen point from the centre) may even be outside the field of view. Turning the centering keys, adjust the image until the specimen point A is midway (= pos. B) between pos. A and the centre of the crosslines M (42c). Move point A to M and check that A stays at M when the stage is rotated (42d). Repeat the centering process if necessary.
Each objective must be centered separately. If an objective is screwed out of the nosepiece, e.g. for cleaning, and screwed back in the same place, is centration is more or less retained.
The centering keys are then stored in the two receptacles on the stage bracket (43.15).
Adjustment
Objective centration
Objectives are centered by adjusting (43.14) them with two Allen keys (1.3; 43.15) until the optical axis of the objective (and therefore the centre of the image) matches the axis of rotation of the specimen stage (41 and 42). When an objective is centered correctly, a focused speci­men area does not drift out of the field of view when the stage is rotated. A specimen point in the center of the crossline therefore does not change positon during a complete rotation of the stage. It is a good idea to use a specimen with plenty of contrast and detail for the objective centration. Disengage the analyser and the Bertrand lens (43.3; 43.4). Greatly narrow the aperture diaphragm (43.7). Insert the two objective cen­tering keys above the objective (43.14) you want to center. Focus the specimen. There are two similar ways of centering objectives:
Method I (Fig. 41)
Rotate the stage and note the point on the speci­men that remains stationary. This point corre­sponds to the mechanical axis of rotation of the stage. Now move this prominent point of the specimen to the centre of the crosslines with the two cen­tering keys. Rotate the stage and fine-adjust the centration if necessary.
M
MM
AA
A
B
M
B
abc
d
Fig. 41
Centration method I
Fig. 42
Centration method II
8
Crossing the polarizers
Focus on an empty area of the specimen or remove the specimen from the light path. Remove compensators (43.10; 43.20; 28.2), Ber­trand lens (43.3) from the light path. Turn the condenser disc of the UCLP* condens­er. In case of condenser CLP/PH* (43.8) pull out the light ring slider to the BF = brightfield posi­tion. Engage the analyser (43.4).
Watching the empty field of view, rotate the polarizer (43.10) until you obtain the optimum extinction position. The setting may be inaccu­rate if the specimen, condenser lenses or polar­izers are dirty, so clean them beforehand if nec­essary!
Fig. 43 Controls of the polarizing microscope Leica DMLSP 1 Eyepiece, with adjustable eyelens* and anti-glare protec-
tion, removable, 2 Interpupillary distance setting, 3 Bertrand lens, off/on, 3a Centration
+
of Bertrand lens, 4 Analyser off/on, 5 Objective, objective nosepiece, 6 Object guide* and specimen, 7 Aperture diaphragm, 8 CLP/PH* condenser, 9 Condenser centration
+
, on left and right and condenser fixing screw on the right, 10 Polarizer with clamp screw for rotation, hinged, with slot for
λ or λ/4 compensator, 11 Filter
holder* with filter*, 12 Field diaphragm, 13 Brightness adjust­ment for transmitted light
+
(left, hidden in the illustration), 14 Objective centration, 15 Holes for keeping centering keys when not in use, 16 Condenser height adjustment
+
,
17 Condenser height stop, 18 Focusing
+
, fine and coarse,
19 Mains switch, 20 Slot for compensators
* not part of all configurations
+
control also situated on right of microscope
Operation of objectives (Pol)
1
19
2
3 4
5 6
7 8
9
11 12
13
10
3a
+
20 14
15
16
17
18
9
A particularly exact way of crossing the polar­izers is to use the built-in Bertrand lens (43.3) or the auxiliary telescope (25.1) as follows: Use an objective with a fairly high magnification, e.g. 40x, 50x, 63x. Open the aperture diaphragm (43.7) (PH position). For the auxiliary telescope: foxus until the some­what brighter circle in the middle of the field of view is sharply defined. When the polarizer is not quite adjusted, 2 dark stripes are visible which close to form a cross when the polarizers are exactly crossed (44a). This cross usually does not close completely in the case of objectives and condensers without the P engraving.
The following section is only intended to give a rough survey of the examination methods. Further details can be found in textbooks on polarization microscopy.
Examinations: One polarizer only
If specimens are to be examined with other transmitted light methods such as brightfield, phase contrast and darkfield instead of with crossed polarizers, it is usually sufficient to dis­engage either the analyser or the polarizer. If the image is not bright enough, both the polarizer and the analyser should be disengaged.
Birefringent specimens with inherent colours may exhibit changes in brightness and/or colour when the stage or polarizer is rotated (with analyser disengaged). This so-called dichroism or pleochroism is a key indication in crystal examinations. However, this effect can be simu­lated on non-polarizing microscopes which have no depolarizing quartz plate, or also if an inci­dent light reflector has been left in the light path when transmitted light is switched on.
Fig. 44 Crossing the polarizers, viewing with a Bertrand lens and a high-aperture objective, without a specimen a exactly crossed, b not exactly crossed Often, Pos. a cannot be set at all if there is strain in the con­denser or objective
Operation of transmitted light polarization
a
b
10
Crossed polarizers
The DIN and ISO standard vibration directions are indicated on the microscope (sticker).
If the specimen contains many non-birefringent or opaque particles, the polarizer is frequently turned out of the crossed position by a few degrees so that these particles show up at least faintly (they remain dark when the polarizers are exactly crossed). It is not customary to examine specimens with the polarizers parallel, as this method of identifying birefringence is not sensi­tive enough.
Change in brightness when birefringent objects are rotated
When the stage is rotated, the brightness of birefringent (anisotropic) objects changes peri­odically. During a full rotation the object dis­appears four times after each 90° interval. The four dark positions are called extinction or nor- mal positions. Exactly between each of these extinction positions the object can be observed with maximum light intensity. These are the four diagonal or 45° positions. In the extinction posi­tions the object vibration directions run parallel to the transmission directions of the polarizers, at maximum intensity the object vibration direc­tions represent the angle bisectors of the polar­izer directions. The crosslines in the (right-hand) eyepiece of polarized light microscopes can either be aligned at N – S/E – W, i.e. in the polar­izer directions, or at 45° angles, i. e. correspond­ing to the object vibration directions in the diagonal position.
λ and λ/4 compensator
Depending on the microscope model, the quar­ter- and whole-wave compensators are either slotted in the compensator slot (43.20) or above the polarizer (43.10; 28.2) or in the light ring slot (CLP/PH) or integrated in the condenser disc (UCLP) (9.6). When a compensator is engaged, the phase difference is increased or decreased (see Fig. 45). The vibration direction γ (i.e. corre­sponding to the greater refractive index n
γ
) can
be determined from the colour changes.
Fig. 45 Interference colours in relation to phase difference, or to thickness and colour change for the addition and subtraction position of a whole-wave and a quarter-wave compensator
Black Lavender gray
200
1st order
400
600
800
2nd order
1000
1200
Phase difference
1400
3rd order
1600
λ
+ λ
Gray blue Yellowish-white
Vivid yellow
Red-orange Deep red
Indigo Sky blue Greenish blue
Light green Pure yellow
Orange red Dark violet red
Indigo Greenish blue
Sea green Greenish yellow
Flesh color Crimson
Matt purple
λ
– – 4
λ
+ – 4
11
Quartz wedge
The quartz wedge (46.6) is inserted into the com­pensator slot (43.20). It allows phase shifts from 0 to about 4 λ (orders).
Circular polarization
Birefringent objects exhibit four extinction posi­tions for one stage rotation. Particularly when scanning a large area of the specimen, some of the birefringent objects will always happen to be at the extinction position. Circular polarization is used for simultaneous observation of the inter­ference colours of all objects: Remove the specimen from the light path or find an empty area of the specimen. Cross the polar­izers exactly – they must also be exactly at the N – S/E – W positions. Insert quarter-wave compensator (46.5) in the compensator slot (43.20). Push the quarter-wave compensator (46.1) into the slot above the po­larizer (43.10) and rotate until the empty field of view appears at its darkest position (first cross polarizers exactly!). The quarter wave compen­sator that can be integrated in the condenser disc (9.6) is not suitable for circular polarization.
Compensators for quantitative measurements
Only in conjunction with polarized light micro­scopes in transmitted light. Adjustable compen­sators are used for exact measurements of phase differences. For a known specimen thick­ness d and the measured phase difference gamma (Γ) the birefringence ∆n can be worked out using the following formula:
d
Γ = d x n [nm] or n =
Γ
Fig. 46 Compensators 1, 2
λ/4 and λ compensator in holder Ø 32 mm. Only for polarized light microscopes: 3 λ/4 and
λ compensator for condenser disk, 4, 5 λ/4 and λ compensators for compensator slot (43.20),
6 Quartz wedge, 7 Tilting compensator, 8 Brace-Koehler compensator
1
2
3
4
5
6
78
12
To perform the measurement, the compensator is introduced into the tube slot and adjusted until the object to be measured is in its maximum extinction position. For this purpose the object has to be moved into a certain diagonal position. Further details are given in the instructions for the use of the compensators. The following compensators are available:
Elliptical Brace-Koehler compensator (46.8)
Rotary compensator with compensator plate of about λ/10 phase difference. Measurement is carried out in white or in monochromatic light. Measurement range up to approx. 50 nm.
Tilting compensator B (Berek compensator) measuring up to 5 orders
Compensator (46.7) with MgF
2
plate for meas­urements in monochromatic or white light of up to about 5 orders phase difference. The phase difference can be read directly from the sum of the two angles of compensation produced when the compensator plate is tilted in both directions, from a supplied calibration chart.
Tilting compensator K, measuring up to 30 orders (like 46.7)
For the measurement of phase differences in white or monochromatic light up to the maximum phase difference mentioned above. The com­pensator plate is made of calcite; evaluation is based on simple calculation by means of enclosed tables and the stated calibration con­stant. A programmed computer can be used for evaluation of measurements taken with tilting compensators. The necessary formulae and parameters are given in: Kornder, F. and W. J. Patzelt: The use of mini­computers to evaluate polarization-optic com­pensator measurements. – Leitz Scientific and Technical Information IX/1, 30 – 32, 1986.
13
Conoscopy of crystals
Birefringent crystals cause interference patterns (Fig. 47) in the exit pupil of the objective (i.e. inside the objective). These are also called conoscopic images. The shape of these interfer­ence patterns and the way they change when compensators are used supply information on the number of crystal axes (uniaxial or biaxial crystals), the orientation of these axes and the plus or minus sign of the birefringence (positive or negative birefringent crystal).
As these interference patterns occur in the pupil, they are not normally visible during normal microscopic observation (orthoscopy). Their ob­servation can be improvised by removing one of the eyepieces and looking into the tube with one eye from a distance of a few centimetres. Observation is better with the auxiliary telescope for phase contrast (Fig. 25.1). However, other crystals in the field of view disturb the interfer­ence patterns of a crystal in the centre, so that it needs to be isolated. This can only be done with the Bertrand lens (43.3) in the Pol module, as here isolation is performed by a fitted dia­phragm. The diameter of the isolated object field is about 55 µm for a 40x objective, about 36 µm for a 63x objective and 23 µm for a 100x objective.
Setting the microscope for conoscopy
The most suitable object areas for conoscopy are those that show the lowest possible phase differences (chart in Fig. 45). Exact centration of the strain-free Pol objectives and exactly crossed polarizers are essential for perfect conoscopic observation. Turn an objective with as high an aperture as possible (e.g. 40x, 50x or 63x) into the light path. Open the aperture diaphragm (43.7). Move the crystal you want to examine as near to the cen­tre of the field of view as possible. Narrow the field diaphragm (43.12) as well, if necessary. Push in the Bertrand lens (43.3). To improve the quality of the image, especially for small crystals, lower the stage by about 0.2 mm.
The insert the centering keys (1.3) into the two openings (43.3a) one after the other and adjust until the circular area (objective pupil) is aligned to the centre of the crosslines.
14
Fig. 47 Determination of the optical character of uniaxial structures. Positive and negative uniaxial crystal, cut vertically to the optical axis. Biaxial positive and negative crystal, cut vertically to the acute bisectrix.
Determination of optical character
Uniaxial crystals (Fig. 47)
Uniaxial crystals observed in the conoscopic (divergent) beam show a dark cross, whose cen­tre indicates the position of the optical axis. The cross is surrounded by coloured interference fringes*.
For the determination of the optical character, cutting directions where the optical axis of the crystal is slightly inclined to the direction of observation are also suitable. The optical char­acter can mostly be determined even when the centre of the cross is outside the field of view.
Biaxial crystals (Fig. 47)
Cutting directions where the bisectrix of the two optical axes is parallel to the viewing direction (section vertical to the acute bisectrix) are par­ticularly suitable for determining the optical character.
In the divergent beam a dark cross will be seen which opens up into the two brances of a hyper­bola, the so-called isogyres, when the stage is being rotated. The cross and the branches of the hyperbola are surrounded by interference fringes. According to Fig. 47 or the rule men­tioned below the optical character can be deter­mined from the displacement direction of these fringes after operation of the compensator. The symmetry plane of the isogyres (axial plane) must be vertical to the γ direction of the compen- sator.
Evaluation of conoscopy
Uniaxial
Without compensator
With λ compensator* vibration direction λ
Displacement of the stripes with compensators
With thin specimens or specimens with low birefringence, only the cross is visible. * with the λ/4 compensator, black dots will occur instead of the black arcs
Biaxial
Loading...
+ 32 hidden pages