Responsible for contents/
Responsable du contenu rédactionnel/
Verantwortlich für den Inhalt:
Dr. Jasna Röth, Peter Schmitt
(Clinical Microscopy, Product Management)
Holger Grasse
(Safety Officer according to MPG §30/
Sicherheitsbeauftragter nach MPG §30/
Responsable de la sécurité conformément à la loi
relative aux dispositifs médicaux, §30)
In case of questions, please contact the hotline/
Pour toute question, contacter notre service
d’assistance téléphonique/
Bei Fragen wenden Sie sich bitte an die Hotline:
All rights to this documentation are held by
Leica Microsystems Wetzlar GmbH. Reproduction of text or illustrations (in whole or in part) by
print, photocopy, microfilm or other method (including electronic systems) is not allowed without express written permission from Leica
Microsystems Wetzlar GmbH.
The instructions contained in the following
documentation reflect state-of-the-art technology and knowledge standards. We have compiled the texts and illustrations as accurately as
possible. Nevertheless, no liability of any kind
may be assumed for the accuracy of this manual’s contents. Still, we are always grateful for
comments and suggestions regarding potential
mistakes within this documentation.
The information in this manual is subject to modification at any time and without notification.
14.1 Equipping the Condenser Disk ................ 57
15. Index ............................................................ 59
16. EU Declaration of Conformity ................. 60
5
1. Important Notes about this Manual
1. Important Notes about this Manual
Caution!
This operating manual is an essential component of the microscope, and must be read
carefully before the microscope is
assembled, put into operation or used.
Text symbols, pictograms and their meanings:
(1.2)
→
p.20
This operating manual contains important instructions and information for the operational
safety and maintenance of the microscope and
accessories. Therefore, it must be kept and
taken care of.
Numbers in parentheses, such as "(1.2)", correspond to illustrations (in the example, Figure 1,
Item 2).
Numbers with pointer arrows (for example
→ p.20), point to a certain page of this manual.
Caution!
Special safety instructions within this
manual are indicated with the triangle
symbol shown here, and have a gray
background.
Caution! The microscope and accessories can
!
*
6
be damaged when operated incorrectly.
Explanatory note.
Item not contained in all configurations.
2. Intended Purpose of the Microscope
2. Intended Purpose of the Microscope
The Leica DM1000 microscope, to which this
user manual belongs, is designed for biological
routine and research applications. This includes
the examination of samples taken from the human body with a view to provide information on
physiological or pathological states or
congenital abnormalities, or to determine the
safety and compatibility with potential recipients,
or to monitor therapeutic measures.
The above-named microscope complies with
the Council Directive 98/79/EEC concerning in
vitro diagnostics. It also conforms to the Council
Directives 73/23/EEC concerning electrical
apparatus and 89/336/EEC concerning
electromagnetic compatibility for use in an
industrial environment.
The manufacturer assumes no liability for
damage caused by, or any risks arising from
using the microscope for other purposes
than those for which they are intended or
not using them within the specifications of
Leica Microsystems Wetzlar GmbH.
In such cases the conformity declaration
shall cease to be valid.
These (IVD) devices are not intended for use
in the patient environment defined by DIN
VDE 0100-710. Neither are they intended for
combining with medical devices according
to EN 60601-1. If a microscope is electrically
connected to a medical device according to
EN 60601-1, the requirements defined in
EN 60601-1-1 shall apply.
Caution!
Caution!
7
3. Safety Notes
3. Safety Notes
3.1 General Safety Notes
This safety class 1 device is constructed and
tested in accordance with
EN 61010-2-101:2002,
EN 61010-1:2001,
IEC 1010-1:2001,
safety regulations for electrical measuring, control, and laboratory devices.
Caution!
In order to maintain this condition and to ensure safe operation, the user must follow the
instructions and warnings contained in this
operating manual.
Caution!
The devices and accessories described in
this operating manual have been tested for
safety and potential hazards.
The responsible Leica affiliate or the main
plant in Wetzlar, Germany must be consulted
whenever the device is altered, modified or
used in conjunction with non-Leica
components that are outside of the scope of
this manual.
3.2 Electrical Safety
General Specifications
Microscope
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Over voltage category:
Pollution degree:
Caution!
The power plug may only be plugged into an
outlet equipped with a grounding contact.
Do not interfere with the grounding function
by using an extension cord without a ground
wire. Any interruption of the ground wire inside or outside of the device, or release of
the ground wire connection, can cause the
device to become hazardous. Intentional
ground interruption is not permitted!
90-250 V~
50-60 Hz
90 W
F 3,15 A 250 V
15-35°C
max. 80% to 30°C
II
2
Unauthorized alterations to the device or
noncompliant use shall void all rights to any
warranty claims!
8
3. Safety Notes
Caution!
Never use any fuses as replacements other
than those of the types and the current ratings listed here. Using patched fuses or
bridging the fuse holder is not permitted.
The use of incorrect fuses may result in a
fire hazard.
Caution!
The microscope’s electrical accessory components are not protected against water.
Water can cause electric shock.
Caution!
Protect the microscope from excessive temperature fluctuations. Such fluctuations can
lead to the accumulation of condensation,
which can damage the electrical and optical
components.
Ambient temperature: 15-35°C.
Caution!
Before exchanging the fuses or lamps, be
absolutely certain to switch off the main
power switch and remove the power cable.
Integrated halogen illumination
manual adjustment of
• Light Intensity
• Aperture Diaphragm
• Field Diaphragm (only with Koehler kit)
Incident light fluorescence illuminator for up to eyepiece
field number 20 with
•Interchangeable slide with mount for 3 filter systems
•Adjusting lens for lamp
•Light trap for the suppression of extraneous light
• BG38 blue filter and shutter, switchable
optionally with
•Fixed or variable viewing angle
•Up to 3 switching positions
• one or two camera ports
• Ergotube with height-adjustable eye level and camera port
Magnification Changer
(optional)
Objective Turret
X/Y Stage
10
• Manual
•Magnification steps: 1x; 1.5x; 2x
• Manual
• 5-fold for objectives with M25 thread
•With condenser holder
• Coaxial pinion optional telescopable
•Controls mountable on left or right
4. Overview of the Instrument
Specification
Condenser
Focusing
Leica DM1000
• CL/PH 0.90/1.25 OIL condenser with color coding
• CLP/PH 0.85 condenser for polarization
• Achr.apl. A 0.9 (P) condenser with swivelable condenser head
• UCL 0.90/1.25 OIL universal condenser UCLP 0.85 for
polarization with 5-position light ring disk)
• UCL/P pol. universal condenser with interchangeable
condenser head and condenser disk with 6 positions
•Focus wheel for coarse and fine focusing
•Height adjustment
11
4. Overview of the Instrument
6
5
Fig. 1Left side of the Leica DM1000 stand
1 Coarse and fine focusing
2 Condenser height adjustment
3 Brightness control
4 Field diaphragm
5 Aperture diaphragm
6 Condenser
12
4
12 3
4. Overview of the Instrument
1
2
3
4
5
Fig. 2Right side of the Leica DM1000 stand
1 Eyepiece
2 Eyepiece tube
3 Tube
4 Objective turret with objectives
5 Specimen stage with specimen holder
6 Integrated illumination
7 On/Off switch
8 Condenser height adjustment
9 Coarse and fine focusing
10 Coaxial pinion for x/y stage movement
678910
13
5. Unpacking the Microscope
5. Unpacking the Microscope
First, carefully remove all components from the
transportation and packaging materials.
Note:
If at all possible, avoid touching the lens surfaces of the objectives. If fingerprints do appear
on the glass surfaces, remove them with a soft
leather or linen cloth. Even small traces of finger
perspiration can damage the surfaces in a short
time. See the chapter "Care of the Microscope"
p. 54, for additional instructions.
→
Caution!
Do not yet connect the microscope and peripherals to the power supply at this point!
Installation Location
Work with the microscope should be performed
in a dust-free room, which is free of oil vapors
and other chemical vapors, as well as extreme
humidity. At the workplace, large temperature
fluctuations, direct sunlight and vibrations
should be avoided. These conditions can distort
measurements and micrographic images.
Allowable ambient conditions
Temperature15-35°C
Relative humiditymaximum 80% up to 30°C
Microscopes in warm and warm-damp climatic
zones require special care in order to prevent
the build up of fungus.
See the chapter "Care of the Microscope" →
for additional instructions.
Caution!
Electrical components must be placed at least
10 cm away from the wall and away from
flammable substances.
p. 54,
14
Transport
For shipping or transporting the microscope
and its accessory components, the original
packaging should be used.
As a precaution to prevent damage from vibrations, the following components should be disassembled and packaged separately:
• Unscrew the objectives.
• Remove the condenser.
• Remove the coaxial pinion.
• Remove the lamp housings.
• Disassemble the burner of 106 z lamp housing.
• Remove all moving or loose parts.
5. Unpacking the Microscope
15
6. Assembly
6. Assembling the Microscope
The microscope components are logically assembled in this order:
• Stage
• Condenser
•Fluorescence*
• Intermediate systems*
•Tube
• Eyepieces
• Objectives
• Lamp housings with light sources
• Polarization*
Only one commonly used screwdriver is
necessary for assembly, which is included in the
delivery package.
The tool can be stored on a magnetic retainer on
the underside of the stage at the right.
When using intermediate systems and optical
accessories, the sequence may vary.
In this case, read Chapter
"6.9 Optional Accessories" → p. 25.
6.1 Stage
!
Caution:
Before completing the stage, make sure no objectives are installed!
Specimen Holder
• Place the specimen holder on the stage and
fasten it with the two screws (3.1).
Coaxial Pinion
Note:
The coaxial pinion can be mounted on the leftor right-hand side.
16
Fig. 3Specimen stage with specimen holder
1 Lock screws for specimen holder
1
• First, place the fine focus wheel on the side to
which you intend to mount the coaxial pinion.
The wheel is held in place magnetically (4.1).
Ensure that the button snaps into place.
Attach the other focus knob on the opposite
side.
• Loosen the lock screw (5.1) at the front lefthand side of the stage.
• Slide the stage as far back as possible.
• Attach the coaxial pinion with the screw (6.1).
• Pull the stage forward and retighten the lock
screw.
6. Assembly
Fig. 5Underside of stage
1 Lock screw
Fig. 4Focus wheel
1 Magnetic retainer for fine focus wheel
1
Fig. 6Coaxial pinion installation
1 Mounting screw for coaxial pinion
1
1
17
6. Assembly
6.2 Condenser
• If present, screw the condenser head into the
condenser.
• Using the condenser height adjuster (9.3), turn
the condenser holder (fig. 8) completely
downward.
• Unscrew the clamping screw for the condenser (9.2) far enough so that the condenser
can be inserted from the front.
• From the front, insert the condenser into the
condenser holder as far as it will go. On the
underside of the condenser, there is an orientation pin (7.1), which must be located in the
guiding notch (8.1).
• Pull the condenser’s clamping screw (9.2) so
that the condenser is locked in place.
Fig. 7 Underside of condenser(example CL/PH)
1 Orientation pin
2 Auxiliary condenser lens LS
Note:
The condenser must be centered before using
the microscope.
For fluorescence applications, install the
fluorescence illuminator first → p. 21.
If available, the analyzer* (10.1) must be inserted
into the stand. This requires that the guide key
engages in the guide pin (10.2).
To mount the analyzer, the analyzer mount TL*
20 mm or 60 mm can also be placed between
stand and tube.
An intermediate tube pole* with a switchable
analyzer (on/off) and Bertrand lens is also
available as an option.
The tube is mounted to the stand either directly or
with the use of intermediate modules.
• Loosen the clamping screw (11.1) on the
stand.
• Insert the tube in the circular receptacle
(dovetail ring).
• Retighten the clamping screw (11.1).
• The eyepieces are inserted into the eyepiece
tubes on the tube.
6.4 Objectives
Always only use Leica objectives of tube length
× (infinity)! The standard thread is M25. The
objectives should be arranged so that the
magnification increases when the objective
nosepiece is rotated counter clockwise.
!
Attention:
Lower the specimen stage as far as possible
before assembling the objectives. Close vacant
threads in the nosepiece with dust protection
caps!
Be sure that the lamp housing is disconnected from the power supply. Unplug the
power plug and the power supply during assembly.
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
Replacing the Lamp of the Integrated
Illumination
The transmitted light illumination with a lowvoltage tungsten halogen lamp (Fig. 12) is
integrated in the base of the microscope and is
accessible from the right-hand side.
• Remove the insert (12.2).
Caution!
The lamp may still be hot!
• Remove the lamp.
Caution!
Do not remove the new lamp’s dust cover
until you have installed the lamp. Avoid
fingerprints on the lamp.
• Insert the new lamp with the dust cover
straight into the socket until it stops. Be sure
that the lamp is inserted straight.
Fig. 12 Transmitted-light illumination in microscope base
1 Tungsten halogen lamp
2 Insert
12
20
• Remove the lamp’s dust cover.
• Replace the insert (12.2).
6.6 Components for Fluorescence Applications
6.6.1 Fluorescence illuminator
6. Assembly
Caution!
The fluorescence illuminator is mounted in front
of the tube. It is fastened in place with the side
clamping screw (13.1).
6.6.2 106 z Lamp Housing
Caution!
Light sources pose a potential irradiation
risk (glare, UV-radiation, IR-radiation).
Therefore, lamps have to be operated in
closed housings.
During assembly, always unplug the power
supply unit of the 106 z lamp housing from its
socket.
During assembly work on xenon burners, always wear the supplied protective gloves
and face protection (Fig. 14) (risk of explosion).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path
(blinding hazard).
Make sure to follow the instructions and
safety notes of the lamp supplier.
Before changing lamps allow at least 30 mins
for cooling down!
Inserting the Ggas Discharge Lamps (Hg and
Xe) into the 106z Lamp Housing
Hg and Xe lamps are powered by separate
supply units.
Read the separate instruction manual provided
with these supply units.
Fig. 13 Assembly of fluorescence illuminator
1 Clamping screw
1
Fig. 14
Protective gloves and mask
This lamp housing is used with various gas
discharge lamps.
21
6. Assembly
The following gas discharge lamps may be used
and require different supply units and lamp
mounts (Fig. 16):
TypeTypical Bulb Life*
50 W high-pressure mercury burner (alternating current)100 hours.
100 W high-pressure mercury burner (direct current)200 hours.
100 W high-pressure mercury burner (direct current, type 103 W/2)300 hours.
75 W high-pressure xenon burner (direct current)400 hours.
* Please regard the data sheets for the burners.
•To open the 106 z lamp housing, unscrew the
fastening screws (15.8) on the cover.
• Remove the transport anchorage (red plastic
rod in place of the burner) in the lamp mount.
To do so, remove the lower clamp (16.1). Pull
up the cooling element (16.3) and turn it to the
side. Detach the lower clamp system (16.2)
and remove the transport anchorage.
• Install the burner in reverse order.
Caution!
Fig. 15 106 z lamp housing (on the side, open)
1Cover raised
2Collector
3Gas discharge lamp in mount
4Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8Fastening screw for lamp mount
9Socket for contact plug
Hg 50 Burner:
After installation, the labeling must be
upright.
If a glass melt nipple is present (16a.4), position
it by turning the burner so that the nipple does
not come in the way of the beam path later, but
instead is positioned
sideways.
Xe 75 Burner:
Remove the burner’s dust cover (16b.5) after
you have installed the burner.
22
1
2
4
5
3
6
7
898
6. Assembly
• Insert the lamp mount, with the burner installed, into the lamp housing and tighten it
with the screws (15.8).
• Close the lamp housing and retighten the
screws.
• Place the lamp housing in the incident light
lamp housing receptacle (17.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the external
power supply.
Fig. 16 a-c Lamp mounts for gas discharge lamps
1 Upper clamping system
2 Lower clamping system
3 Cooling element
4 Nipple of the mercury 50 burner,
5 Dust cover of the mercury 75 burner
Fig. 17 Mounting the 106 z lamp housing
1 Lamp housing receptacle
1
Hg 50
1
4
a
3
2
Xe 75
b
Hg 100
3
1
1
c
3
5
2
2
23
6. Assembly
6.7 Analyzer and Polarizer*
Analyzer
If the analyzer was inserted into the tube mount
before the tube assembly: (→ p. 19), no additional assembly step is required.
If an intermediate tube pole* or analyzer mount
TL* is used:
• Remove the plug cap on the left side.
• Insert the analyzer into the receptacle until it
latches in place.
Polarizer
• Raise the condenser to its upper stop position.
• Remove the DLF filter magazine from the base
if present.
• Press the polarizer holder in place (Fig. 18).
• Push the polarizer with the labeled side
upward into the lower opening.
Alternative:
• Attach the polarizer holder to the underside of
the condenser holder with the left clamp
screw (19.1). Remove the flip-out blue filter if
required.
• Push the polarizer with the labeled side
upward into the lower opening.
6.8 Lambda Plate Compensator*
• Raise the condenser to its upper stop position.
• Remove the DLF filter magazine from the base
if present.
• Attach the lambda plate compensator to the
base.
Fig. 18 Filter holder
with 2 positions
24
Fig. 19 Assembly of polarizer holder
1 Clamping screw
1
6. Assembly
6.9 Optional Accessories
Camera
A camera can be connected via an adapter.
• Attach the adapter to the top port of the tube
and fasten it tightly with the side clamping
screw.
• Screw on the camera.
Note:
The size of the camera chip and the mounting
system (B-mount, C-mount, etc.) must be
considered when choosing an adapter.
See table.
Calculation of the magnification on the monitor
The magnification M
on the monitor can be
TV
calculated with the following formula or
measured with a stage micrometer and a cm
scale:
= Objective magnification x
M
TV
factor of magnification changer*x
TV adapter magnification x
monitor diameter
chip diameter of camera
Recorded picture diagonal in mm for
1 inch2/3 inch1/2 inch1/3 inch
cameracameracameracamera
Without Zoom Magnification, Only for 1-Chip-Cameras:
C-mount adapter 1 x HC161186
C-mount adapter 0.63 x HC-17.512.79.5
C-mount adapter 0.5 x HC--1612
C-mount adapter 0.35 x HC---17.1
With Zoom Magnification (Vario TV Adapter) for 1-3 Chip-Cameras:
+)
C-mount, 0.32-1.6 x HC--19
-518-3.8
B-mount (ENG), 0.5-2.4 x HC (1/2-inch)--16-3.3-
+)
from zoom factor 0.42 x only!
Without Zoom Magnification, for 1-3 Chip-Cameras:
C-mount adapter 1 x--1612
B-mount adapter 1 x--1612
B-mount adapter 1.25 x-17.5-F-mount adapter 1 x--1612
F-mount adapter 1.25 x-17.5-Plus (essential requirement): TV optics 0.5 x HC
25
6. Assembly
Ergomodule
For raising the eye level of the tube opening, the
30 mm or 60 mm ergomodule may be used.
It is fastened in place with the side clamping
screw.
Ergolift
A base for the stand featuring adjuster wheels
for the base’s height and angle is available to
ensure an optimal working position.
Magnification Changer
Optionally, a magnification changer (fig. 20) can
be used, which is manually operated. On the
knurled ring, the following magnification factors
can be set:
1x; 1.5x; 2x
Viewing Attachments
Viewing attachments featuring illuminated
pointers are available for groups of up to 20
viewers.
The support (21.3) must be aligned precisely.
The fade-in arrow can be moved in x and y
direction (move the lever vertically or pull out/
push in) (21.1) If this lever is rotated, the color of
the arrow can be changed (red/yellow). Use the
brightness control (21.2) to adjust the brightness
of the arrow.
Fig. 20 Magnification
changer
26
Fig. 21 Viewing attachment
1 Movement of light pointer in x and y direction,
and switchover of color filter
2 Brightness control
3 Adjustment of arm support
The external power supply (illuminated arrow) is not
illustrated.
12
3
6. Assembly
Tracing Device
The tracing device L3/20 (fig. 22) allows an
optical overlay of large objects (next to the
microscope) on the microscope image. This
makes it easy to draw specimens by tracing
their outlines or superimposing scales.
6.10 Connection to the Power Supply
• After completing the assembly work, connect
the stand to the power supply using the power
cable supplied (fig. 23).
•When using the lamp housing or the
external power supply unit, connect them to
the power supply, too.
Fig. 22 Tracing device
1 Shutter
Fig. 23 Back of the stand
1 Power supply connection
1
1
27
7. Start-up
7. Start-up
7.1 Switching on the Microscope
•Switch on the microscope with the on/off
switch (24.1).
Caution!
After turning on the gas discharge lamp, the
burner must be immediately adjusted. Therefore, do not turn on the power supply unit
yet. First, work in transmitted light in order to
familiarize yourself with the microscope’s
controls.
The condenser is also pre-adjusted in the
factory.
However, it may be necessary to re-adjust the
condenser in some cases. Therefore, check the
condenser centering.
The following procedure is provided for the
transmitted light brightfield illumination.
• If present click the condenser disk* into the
BF position.
• If present pull the light ring slide* out of the
condenser.
• Select an objective with moderate
magnification (10x-20x).
For condensers with movable condenser
heads:
Swing in the condenser top.
(The condenser top is swung out for objective
magnifications < 10x.)
28
• Insert the specimen into the stage’s specimen
holder.
• Focus on the specimen using the focus wheel
(24.2).
• Set the light intensity using the brightness
control (25.2).
• Close the field diaphragm (25.3) until the edge
of the diaphragm appears in the specimen
plane (26a).
213
7. Start-up
• Using the condenser height adjuster (25.1),
adjust the condenser until the edge of the
field diaphragm appears in sharp relief(26b).
• If the image does not appear in the middle of
the field of view (26c), the condenser must be
moved into the middle of the field of view with
the help of the two centering bolts (25.4). The
tool required for this purpose is magnetically
attached to the underside of the stage.
• Open the field diaphragm just enough for it to
disappear from the field of view (26d).
Note:
The condenser height adjustment depends on
the thickness of the specimen. It may be
adjusted for different specimens.
Fig. 25
1 Condenser height adjuster
2 Brightness control
3 Field diaphragm
4 Condenser centering
7.3 Checking Phase Contrast Rings
If your microscope is equipped for the use of
phase contrast, the light rings that fit the objectives are built into the condenser.
The light rings are already centered in the
factory. However, the centering should be
rechecked.
Note:
A light ring slide which is inserted into the side
of the condenser is used for condensers without
condenser disks. Centering is not required in
this case.
Note:
When swivelling in a suitable objective for
phase contrast, the corresponding light ring
must be chosen.
The objective engraving (e.g. PH 1) indicates the
corresponding light ring (e.g. 1).
Fig. 26 Köhler Illumination
a Field diaphragm not focused, not centered
b Field diaphragm focused, but not centered
c Field diaphragm
Diameter is too small, however
d Field diameter (light) = Field diameter (view)
(Köhler Illumination)
123
4
a
b
cd
29
7. Start-up
• In the place of an eyepiece, insert the focusing telescope (Fig. 27) into the observation
tube.
•Swivel in the phase contrast objective with
the lowest magnification.
• Focus on the specimen with the focus wheel.
• Focus the ring structure (29.a) by slightly loosening the clamping ring (27.2) and moving the
eye lens (27.1).
• Retighten the clamping ring.
• Select the corresponding ring diaphragm
(light ring) in the condenser.
• If the light ring and the phase ring are not
shown as arranged in Fig. 29.c, the light ring
must be centered.
• Insert the centering screws into the openings
provided at the rear of the condenser (28.1).
•Turn the centering screws until the dark ring
(phase ring in the objective) is congruent with
the slightly narrower bright ring (light ring in
condenser) (29 c).
• Repeat the process for all other light rings.
• Remove the centering keys after the centering
procedure.