Leica CM3050 S User Manual

Page 1
Instruction Manual
Leica CM3050 S – Cryostat English V1.2– 11/2006 Always keep this manual near the instrument. Read carefully prior to operating the instrument.
Leica CM3050 S
Cryostat
Page 2
Page 3
Important note
Serial No.: .............. ......................................................
Year of Manufacture: ...............................................
Country of Origin: Federal Republic of Germany
The information contained in this document represents state-of-the-art technology as well as the current state of knowledge.
Leica will not assume liability for errors that might be contained in this manual, nor for accidental damage or damage arising from the delivery, performance or use of this manual.
Therefore, no claims can be made based on the text and illustrations in this instruction manual.
Leica Microsystems Nussloch GmbH reserves the right to change technical specifications without notice since each of our products is subject to a policy of continuous improvement.
This document is protected under copyright laws. Leica Microsystems Nussloch GmbH retains all rights related to this documentation. Any reproduction of text and illustrations - or any parts thereof - in form of printing, photocopying, microfiches, or other methods, including electronic systems, requires the prior written permission of Leica Microsystems Nussloch GmbH.
The serial number and the year of manufacture are specified on the nameplate at the back of the instrument.
Issued by:
Leica Microsystems Nussloch GmbH Heidelberger Str. 17 - 19 D-69226 Nussloch Germany Phone: +49 6224 143-0 Fax: +49 6224 143-200 Internet: http://www.histo-solutions.com
© Leica Microsystems Nussloch GmbH
3
Page 4
Table of Contents
1. Important information ................................................................................................................ 7
1.1 Symbols used in this manual and .................................................................................................... 7
1.2 Designated use ................................................................................................................................... 8
1.3 Qualification of personnel .......................................................................................................... 8
2. Sicherheit .................................................................................................................................... 9
2.1 General information on instrument design and safe handling .................................................... 9
2.2 Integrated safety devices ................................................................................................................. 9
2.2.1 Locking the handwheel ................................................................................................................... 10
2.2.2 Centering the handwheel grip ........................................................................................................ 10
2.2.3 Emergency stop function (instruments with sectioningmotor only) ........................................ 11
2.2.4 Knife guard ........................................................................................................................................ 11
2.3 Safety instructions: handling and operation of the instrument................................................. 12
2.3.1 Transport ........................................................................................................................................... 12
2.3.2 Site requirements ............................................................................................................................. 12
2.3.3 Electrical connections ..................................................................................................................... 12
2.3.4 Handling microtome knives/blades ............................................................................................... 13
2.3.5 Knife guard/handwheel lock........................................................................................................... 13
2.3.6 Motorized sectioning ....................................................................................................................... 13
2.3.7 Defrosting/Handling frozen tissue ................................................................................................. 13
2.3.8 Frozen parts of the instrument and frozen accessories ............................................................ 13
2.3.9 Infectious/radioactive material ...................................................................................................... 14
2.3.10 Disinfection and cleaning ............................................................................................................... 14
2.3.11 Removing/reinstalling the microtome ........................................................................................... 15
2.3.12 Display message ‘Dry microtome’ ................................................................................................. 15
2.3.13 Maintenance .............................................................................................................................. 15
3. Installation................................................................................................................................. 16
3.1 Site requirements ............................................................................................................................. 16
3.1.1 General site requirements .............................................................................................................. 16
3.1.2 Electrical connections ..................................................................................................................... 16
3.2 Unpacking and installation ............................................................................................................. 16
3.2.1 Repacking .......................................................................................................................................... 16
3.3 Standard delivery ............................................................................................................................. 17
3.4 Installing/inserting accessories ..................................................................................................... 19
3.4.1 Installing the handwheel ................................................................................................................. 19
3.4.2 Inserting the accessories ............................................................................................................... 19
3.4.3 The footswitch .................................................................................................................................. 20
3.5 Prior to switching on the instrument ........................................................................................ 20
4
Instruction Manual V1.2 - 11/2006
Page 5
Table of Contents
4. Instrument Properties .............................................................................................................. 21
4.1 Overview ............................................................................................................................................ 21
4.2 Technical Data........................................................................................................................... 23
5. Operation ................................................................................................................................... 25
5.1 Operating the instrument for the first time ................................................................................... 25
5.2 Switching on/fuses ........................................................................................................................... 26
5.3 Control panel 1 .................................................................................................................................. 27
5.3.1 Key functions in control panel 1..................................................................................................... 27
5.3.2 Display functions in control panel 1 .............................................................................................. 27
5.3.3 Display indications when switching on ........................................................................................ 28
5.3.4 Menu functions: setting parameters of refrigeration, time and preset counter .................... 29
5.3.5 Status display .................................................................................................................................... 32
5.3.6 Indication of refrigeration state, section thickness setting and counters .............................. 33
5.4 Control panel 2 .................................................................................................................................. 34
5.4.1 Selecting a sectioning mode / ........................................................................................................ 35
5.4.2 Setting the sectioning window ....................................................................................................... 36
5.4.3 Selecting sectioning speed ............................................................................................................ 37
5.4.4 Emergency stop ................................................................................................................................ 37
5.4.5 Coarse feed keys .............................................................................................................................. 38
5.4.6 Trimming and sectioning functions ............................................................................................... 39
5.4.7 Setting the section thickness ......................................................................................................... 39
5.5 Daily operation of the instrument .................................................................................................. 40
5.5.1 Freezing specimens onto specimen .............................................................................................. 40
5.5.2 Freezing specimens onto specimen .............................................................................................. 41
5.5.3 Optional accessories for freezing .................................................................................................. 41
5.5.4 Inserting specimen discs into the .................................................................................................. 42
5.5.5 Specimen orientation ....................................................................................................................... 42
5.5.6 Trimming / sectioning ...................................................................................................................... 43
5.6 Finishing work ................................................................................................................................... 44
5.6.1 Daily routine steps............................................................................................................................ 44
5.6.2 Switching off the instrument for a ................................................................................................. 44
5.7 Defrosting .......................................................................................................................................... 45
5.7.1 Automatic chamber defrost cycle ................................................................................................. 45
5.7.2 Manual chamber defrost cycle ...................................................................................................... 46
5.7.3 Defrosting the specimen head ................................................................................................ 46
Leica CM3050 S – Cryostat
5
Page 6
Table of contents
6. Troubleshooting, applications tips ....................................................................................... 47
6.1 Displayed error messages .............................................................................................................. 47
6.1.1 Other error indications .................................................................................................................... 47
6.2 Potential problems - causes and remedies ................................................................................. 48
6.3 Reference chart for temperature settings (in minus °C) .................................................... 52
7. Disinfection, cleaning and maintenance ............................................................................ 53
7.1 Safety instructions ........................................................................................................................... 53
7.2 Spray disinfection with Leica Cryofect ......................................................................................... 53
7.3 Disinfection with conventional ....................................................................................................... 54
7.4 Cleaning ............................................................................................................................................. 54
7.5 Removing / reinstalling the microtome ......................................................................................... 55
7.5.1 Removing the microtome ................................................................................................................ 55
7.5.2 Disassembling the specimen head ................................................................................................ 55
7.5.3 Reinstalling the microtome ............................................................................................................. 56
7.6 Exchanging the fluorescent light lamp................................................................................... 57
8. Warranty and service .............................................................................................................. 58
9. Appendix .................................................................................................................................... 59
1. Electrohydraulic height adjustment .............................................................................................. 59
2. Ordering information - accessories ....................................................................................... 62
10. EC Declaration of Conformity ..................................................................................... 63
11. Decontamination Certificate (master) .................................................................................... 64
6
Instruction Manual V1.2 - 11/2006
Page 7
1. Important information
The chapters of the Leica CM3050 S cryostat instruction manual:
Chapter 1 Structure of the manual:
Table of contents
Important information on this manual
Chapter 2 Safety
Read this chapter before operating the instrument!
Chapter 3 Installation
Unpacking and installation
Standard delivery, information on accessories
Chapter 4 Instrument properties
Overview
Technical data
Chapter 5 Operation
Controls
Setup, daily use and shut down
Chapter 6 Trouble shooting
and applications tips
Operating errors
Display error messages
Temperature reference chart
Chapter 7 Disinfection, cleaning
and maintenance
Chapter 8 Warranty and service
Chapter 9 Electrohydraulic height adjustment
Ordering information accessories EC Declaration of Conformity
1.1 Symbols used in this manual and their meaning
Warnings appear in a grey box and are marked by a warning triangle .
Notes i.e. important user information appears in a grey box and is marked by an information symbol.
(5)
(Fig.5)
Instrument type:
All information given in this instruction manual applies only to the instrument type indicated on the title page.
A name plate, indicating the instrument serial number, is attached to the back of the instru­ment.
Required in­formation for all inquiries:
For any inquir­ies please specify:
Instrument
type
Serial-
number
Figures in brackets refer to item numbers in drawings or to the drawings themselves.
7
Page 8
1. Important information
General
This instruction manual includes important instructions and information related to the operating safety and maintenance of the instrument.
The instruction manual is an important part of the product. It must be read carefully before using the instrument for the first time and must always be kept with the instrument.
If additional requirements, which exceed the scope of this manual, are imposed by regula-tions and/or laws on accident prevention and environmental protection in the country of operation, appropriate instructions for compliance with such requirements must be added to this manual.
Read this instruction manual carefully before attempting to use or operate the instrument.
Please pay particular attention to chapter 2 (safety features, safety instructions). - Please read this information, even if you are already familiar with the operation and use of other Leica products.
1.2 Designated use
The Leica CM3050S is a powerful cryostat for routine as well as research applications in biology, medicine and industry.
The instrument has been designed for rapid freezing and sectioning of tissue samples.
The instrument has not been designed for un­attended storage of tissue material.
The instrument may only be operated within the scope of its designated use as described above and as per the instructions given in this manual.
Any other use of the instrument is considered improper.
1.3 Qualification of personnel
The Leica CM3050S may only be operated by trained laboratory personnel.
All laboratory personnel designated to operate the instrument must carefully read the present instruction manual prior to starting work with the instrument.
8
Instruction Manual V1.2 - 11/2006
Page 9
2. Sicherheit
2.1 General information on instrument design and safe handling
This instrument has been built and tested in ac­cordance with the following safety regulations on electrical measuring, control, regulating and laboratory devices:
DIN EN 292
DIN EN 61010-1
EN 50082-1
EN 55011
IEC 1000-4
as well as according to the international quality standard
DIN ISO 9001
In order to maintain this condition and to ensure safe operation, the operator must observe the instructions and warnings contained in this in­struction manual.
2.2 Integrated safety devices
The instrument is equipped with the following safety devices:
Handwheel lock
Handwheel grip centering
Emergency stop function (instruments with sectioning motor only)
Knife holder equipped with knife guard
The safety devices installed by the manufactu­rer of the instrument only constitute the basis of accident prevention.
Mainly responsible for accident-free operation is above all the institution which owns the in­strument and, in addition, the designated per­sonnel who operate, service or repair the instrument.
Leica CM3050 S – Cryostat
9
Page 10
2. Safety
2.2.1 Locking the handwheel
Always cover the cutting edge with the knife guard and lock the hand­wheel:
• Prior to doing any work on knife and/or specimen.
• Prior to exchanging specimens.
• During work breaks.
The handwheel can be locked in 2 positions:
with the grip in the uppermost position (left),
with the grip in the lowest position (right).
2
Instruments with sectioning motor:
The sectioning motor is now blocked.
All instruments:
The message ‘LOCKED’ in the display of control panel 1 indicates that the handwheel has been locked:
CT-30°C OT-35°C
LOCK ED
Unlocking:
To unlock, push locking pin (2) to the left into position (2a).
Display indication ‘LOCKED’ disappears.
Instruments with sectioning motor:
The sectioning motor can now be activated again.
2
Locking:
Rotate handwheel, until grip (1) is in upper or lower position.
To lock, press pin (2) to the right into position (2b).
--> The upper locking position for pin (2) is marked by a black dot (4).
10
2.2.2 Centering the handwheel grip
During motorized sectioning, for safety reasons always center the handwheel grip!
To center grip (1), pull outwards and pivot into center of handwheel.
When released, the grip locks into position.
Instruction Manual V1.2 - 11/2006
Page 11
2. Safety
2.2.3 Emergency stop function (instruments with sectioning motor only)
The emergency stop is activated via the red emergency stop button in control panel 2 or via the footswitch
Control panel 2
Footswitch
Activating the emergency stop function
2.2.4 Knife guard
All knife holders are equipped with a knife guard (--> see separate instruction manuals on knife holders).
Always cover the cutting edge with the knife guard:
• Prior to doing any work on knife and/ or specimen.
• Prior to exchanging specimens.
• During work breaks.
Press emergency stop button or step on footswitch forcefully.
-->
STOP (red) lights up.
As soon as the emergency stop function is acti­vated, the sectioning motor stops.
Deactivating the emergency stop
To deactivate, rotate emergency stop button in direction of arrow.
If the emergency stop function has been activated by the footswitch, unlocking is not necessary (func-tion is unlocked as soon as the footswitch is released).
Leica CM3050 S – Cryostat
11
Page 12
2. Safety
2.3 Safety instructions: handling and operation of the instrument
2.3.1 Transport
To avoid severe damage to the instrument by running it while the compressor oil is dis­placed from its regular position:
Do not tilt the instrument,
only transport in an upright position.
After transport, wait at least 4 hours be-
fore turning the instrument on.
-->The compressor oil may have been displaced during transport and must settle to its original position be­fore switchingthe instrument on. Otherwise, the instrument may be severely damaged.
2.3.2 Site requirements
Do not operate the instrument in rooms with explosion hazard.
To ensure proper instrument function main­tain a minimum distance of 10 cm between walls and/or furniture and all sides of the in­strument!
2.3.3 Electrical connections
Do not use extension cords for connecting the instrument to mains.
--> Fire hazard!
--> Instrument malfunctions caused by
voltage drop.
During the start-up phase of the compressor, the nominal voltage must not drop below the values specified in chapter 4.2 ‘Technical data’!
--> The compressor needs a start-up current
at between 25 and 35 A (see chapter 4.2 ‘Technical data’)
Ensure uniform current supply according to specifications.
--> Electrical power supply deviating from
specifications damages the instrument.
Therefore, arrange for the electrical in-
stallations on site to be checked by a trained professional and make sure any necessary upgrades are installed!
Have the circuit protected by a fuse of its
own!
Do not connect any other consumers to
the same circuit.
Prior to connecting the instrument to
mains, make sure the electrical power supply in your laboratory corresponds to the values indicated on the instrument nameplate (located at the rear of the in­strument)!
12
Instruction Manual V1.2 - 11/2006
Page 13
2. Safety
2.3.4 Handling microtome knives/blades
Danger!
-->Microtome knives and disposable blades have extremely sharp cutting edges and can cause serious injuries. Therefore:
Handle knives / blades with utmost care.
Never leave any knives / blades in unpro­tected places!
Never place a knife, no matter where, with the cutting edge facing upwards.
Never try to catch a falling knife.
Always insert the specimen inserting the knife.
before
2.3.5 Knife guard/handwheel lock
Always cover the cutting edge with the knife guard and lock the handwheel:
Prior to doing any work on knife and / or
specimen.
Prior to exchanging specimens.
During work breaks.
2.3.6 Motorized sectioning
Do not interrupt sectioning / trimming by set­ting the sliding potentiometer to zero speed!
-->This does not really switch the sectioning
function off - it only operates at ‘0’ speed. If the sliding potentiometer is accidental­ly moved, the instrument will resume sec­tioning immediately (risk of injury)!
2.3.7 Defrosting/Handling frozen tissue
Never leave specimens unattended in the cryochamber over an extended period of time!
-->In case of power failure or instrument failure, or during the automatic defrost cy­cle, tissue material can be destroyed.
During the defrost cycle the cryochamber is partially warmed. - Therefore:
Remove sensitive specimens from the
chamber prior to defrosting.
If automatic defrosting is programmed to
take place during the night, remember to remove all specimens from the cryo­chamber prior to leaving work.
2.3.8 Frozen parts of the instrument and
frozen accessories
Prolonged contact of bare skin to frozen surfaces of the instrument or to frozen accessories (specimen discs, knife holder, shelves etc.) can cause frostbite.
Wear protective gloves.
During motorized sectioning, always center the handwheel grip!
Leica CM3050 S – Cryostat
13
Page 14
2. Safety
2.3.9 Infectious/radioactive material
Danger! Use caution when working with potentially infectious specimens:
-->Risk of infection!
When working with potentially infectious / radioactive specimens:
Wear protective clothes (gloves, protec-
tive boots, mask, lab coat), in compliance with radiation safety regulations and/or in-house regulations on handling infec­tious / radioactive material.
When working with radioactive specimens:
Comply with applicable radiation safety
regulations!
Dispose of radioactive specimen waste ac-
cording to applicable regulations.
2.3.10 Disinfection and cleaning
Prior to disinfection, switch the instrument off and unplug it from mains.
For removal of the microtome from the cryo­chamber see chapter 2.3.11 ‘ Removing the microtome’.
For disinfection, wear protective gear: (gloves, mask, lab coat etc.)!
For disinfection, only use alcohol-based dis­infectants!
Do not use solvents (xylene, acetone etc.) for cleaning or disinfection!
Do not spray disinfectants into the evapora­tor!
--> Risk of icing!
Explosion hazard when working with alcohol: Make sure the premises are appropriately ventilated!
When using disinfectants and detergents, comply with all safety instructions supplied by the manufacturer of the product!
14
Dispose of waste liquids from disinfection/ cleaning as well as of sectioning waste ac­cording to applicable regulations on disposal of special category waste!
Disinfected accessories must be thoroughly dry when reinserting them into the chamber!
--> Risk of icing!
Make sure the chamber is completely dry before switching the instrument back on:
--> Explosion hazard through alcohol vapors!
--> Risk of icing!
Instruction Manual V1.2 - 11/2006
Page 15
2. Safety
2.3.11 Removing/reinstalling the microtome
Before removing the microtome:
Switch instrument off.
Unplug from mains.
Place handwheel grip in lowest position and
lock.
-->When removing the microtome, the
specimen head must always be lok­ked in the lowest position. Otherwise the upper part of the slot cover might be bent and consequently damaged!
When removing the microtome:
Wear gloves when removing the micro­tome while it is still frozen.
-->Risk of frost bite!
On instruments with specimen cooling: do not distort the refrigerating tube! If distorted it might break, causing ex­tremely cold refrigerant to escape. - ­> Risk of frost bite!
Before reinstalling the microtome:
Microtome must be completely dry. -­>Humidity in the interior of the micro­tome freezes and causes microtome malfunctions and/or damage to the microtome.
2.3.12 Display message ‘Dry microtome’
If the error message ‘Dry Microtome’ is dis­played in control panel 1, the following has happened:
Cryochamber refrigeration has been in-
terrupted for an extended period of time (e.g. power failure), causing the chamber temperature to rise into the positive digits.
• If this message appears,
the instrument but remove the microtome from the chamber, disinfect, if necessary, and
dry thoroughly before reinstalling it
into the chamber (see chapter 7.1 to 7.5).
do not switch on
2.3.13 Maintenance
Only technical service engineers authorized by Leica may access the internal compo­nents of the instrument for service and re­pair.
The fluorescent light lamp (chamber illumi­nation), unless broken or splintered, can be replaced by the user:
Switch off mains switch!
Unplug the instrument from mains!
If the lamp is broken or splintered:
Have lamp replaced by Leica Technical Service!
-->Risk of injury!
All accessories/tools removed from the cryochamber must be thoroughly dry be­fore putting them back into the chamber!
-->Risk of icing!
Leica CM3050 S – Cryostat
Use only those replacement lamps that cor­respond to technical specification (see chapter 4.2 ‘Technical Data’).
15
Page 16
3. Installation
3.1 Site requirements
Make sure to read and follow all safety instructions in chapter 2.3.2 ‘Site requirements’!
3.1.1 General site requirements
No direct sunlight.
Electrical power supply within distance (length of power cord = approx. 4 meters ­do not use extension cords!).
No draft (caused by air conditioning etc.).
Even floor surface.
Practically vibration-free floor.
Handwheel easily accessible.
Room temperature constantly below +22 °C.
Relative humidity of air maximum 60 %.
High ambient temperature and/or high air humidity negatively affect instrument cooling performance!
3.2 Unpacking and installation
Unpacking instructions are always located in a transparent protective envelope on the outside of the instrument shipping crate.
Make sure to read and follow all safety instructions provided in chapter 2.3.1 ‘Transport’ and on the unpacking instructions!
3.2.1 Repacking
We recommend to keep the original shipping crate and the unpacking instructions for the Leica CM3050 S.
For repacking, proceed as per unpacking in­structions, in reverse order.
Available models
Basic instrument
- with sectioning motor
- without specimen cooling
3.1.2 Electrical connections
Make sure to read and follow all safety instructions in chapter 2.3.3 ‘Electrical connections’!
16
Basic instrument
- without sectioning motor
- with specimen cooling
Basic instrument
- with sectioning motor
- with specimen cooling
Instruction Manual V1.2 - 11/2006
Page 17
3.3 Standard delivery
3. Installation
1 Cryostat with microtome .............................................................................0460 32037
1 Handwheel with marking, antibacterial ...................................................0471 42558
1 Heat extractor, stationary ........................................................................... 0369 11197
1 Low-temperature stabilizer for heatextractor ........................................ 0452 27919
1 90° prism for direct specimen freezing on specimen head
with clamping screw to facilitate specimen trimming .......................... 0443 25949
1 Set of specimen discs ................................................................................. 0470 43550
- 4 Specimen discs, 25 mm ......................................................................... 0416 19275
- 4 Specimen discs, 30 mm ......................................................................... 0370 08587
1 Storage shelf, right ...................................................................................... 0443 25723
1 Storage shelf, left ......................................................................................... 0443 33549
1 Section waste tray .......................................................................................0400 26817
1 Foot switch with protective guard ............................................................ 0502 29977
1 Rubber mat .................................................................................................... 0443 25732
1 Freezing shelf cover .................................................................................... 0443 30783
1 Brush shelf .................................................................................................... 0398 13088
1 Tool set ........................................................................................................... 0436 43463
-1 Brush, fine .................................................................................................0183 28642
-1 Leica-brush ............................................................................................... 0183 30751
-1 Allen key, size 1.5 .....................................................................................0222 10050
-1 Allen key, size 2.5 .....................................................................................0222 04137
-1 Allen key, size 3.0 .....................................................................................0222 04138
-1 Allen key, size 4.0 .....................................................................................0222 04139
-1 Allen key with spherical head, size 4.0 ................................................ 0222 32131
-1 Allen key, size 5.0 .....................................................................................0222 04140
-1 Allen key with handle, size 5.0 ............................................................... 0194 04760
-1 Allen key, size 6.0 .....................................................................................0222 04141
-1 Single-head wrench, size 13.................................................................. 0330 33149
-1 Single-head wrench, size 16.................................................................. 0330 18595
1 Bottle of OCT-Compound, mounting medium
for cryosectioning, 125 ml ........................................................................... 0201 08926
1 Bottle of cryostat oil, type 407, 50 ml ........................................................ 0336 06098
1 Pair of cut resistant gloves, size S ............................................................ 0340 40859
1 Instruction manual Leica CM3050 S - G/E/F/S ........................................ 0708 37109
In addition to the above:
Instruments with specimen cooling:
1 90° Prism
for low temperature quick freezing
Configured instruments:
1 Knife holder base 1 Knife holder with accessories
Instruments with sectioning motor:
1 Footswitch with
protective guard
17
Page 18
3. Installation
Further accessories:
In addition to the above:
Instruments with specimen cooling: 1 90° Prism for low temperature quick freezing
Configured instruments: 1 Knife holder base 1 Knife holder with accessories
Instruments with sectioning motor: 1 Footswitch with protective guard
Further accessories which you ordered will be included in the box containing the standard delivery items.
Knife holders are delivered with anti-roll guide, knife guard, and a separate instruction manual. In case of non-configured instruments, the knife holder is not a part of standard delivery but must be ordered separately.
18
Check all delivered parts against the packing list and against your order to verify whether the delivery is complete! If there is any difference, contact your local Leica office immediately.
Instruction Manual V1.2 - 11/2006
Page 19
3. Installation
3.4 Installing/inserting accessories
3.4.1 Installing the handwheel
Insert pin (1) of the handwheel shaft into hole (2).
Place spring washer (3) onto screw (4) as shown.
Tighten screw (4) with an Allen key.
Cover with selfadhesive disc (5).
For purposes of transport (e.g. narrow doors), the handwheel can be removed.
3.4.2 Inserting the accessories
Place the rubber mat on top of the housing.
Insert the storage shelves into the cryo­chamber.
Install the stationary heat extractor into the quick-freeze shelf (see also drawing in chapter 5.5.1 ‘Freezing specimens onto specimen discs’).
Insert the low temperature stabilizer into the quick freeze shelf (it must be located in the pivoting range of the heat extractor - see also drawing in chapter 5.5.1 ‘Freezing spec­imens onto specimen discs’).
Insert section waste tray and brush shelf.
Install knife holder base onto microtome base plate and clamp.
Install knife holder and clamp (see knife holder instruction manual for details).
Place knife case with knife into chamber to precool.
Place all tools needed for section prepara­tion into the chamber.
Close the sliding window.
To remove the handwheel, proceed as de­scribed above but in reverse order.
For a complete overview of all individual parts:
--> see Chapter 4.1 ‘Overview’.
19
Page 20
3. Installation
3.4.3 The footswitch
Function
The footswitch performs the same task as the RUN/STOP and RUN/ENABLE keys (activating/ deactivating motorized sectioning / trimming). In addition, the footswitch can be used to acti­vate the emergency stop function.
Models with footswitch:
All instruments with sectioning motor.
Important note: In all instrument models that are delivered with footswitch, the footswitch must be installed! ­Otherwise the instruments are not functional.
Connecting the footswitch
Insert footswitch into port (1) and secure.
1
3.5 Prior to switching on the instrument
After transport, observe a waiting period of at least 4 hours before turning the instrument on! - See also safety instructions in chapter 2.3.1 ‘Transport’.
Have you observed all safety instructions in chapters 2.3.2 ‘Site requirements’ and 2.3.3 ‘Electrical connections’? If not:
--> Please read chapters 2.3.2 and
2.3.3!
Insert mains plug into wall outlet.
Continue with chapter 5.1 ‘Setting up the in­strument’.
20
Instruction Manual V1.2 - 11/2006
Page 21
4. Instrument Properties
4.1 Overview
2
3
9
10
11
8
12
1 Cryostat CM3050 S 2 Control panel 1 3 Control panel 2
4
5
6
1
7
22
4 Storage shelf, left 5 Storage shelf, right 6 Rubber mat 7 Mains switch /
Automatic cutout for instrument, Automatic cutout for sectioning motor, Footswitch port
8 Quick-freeze shelf 9 Stationary
heat extractor
10 Mobile heat extractor 11 Specimen disc 12 Thermoblock (optional) 13 Section waste tray 14 Brush shelf 15 Specimen head w/o
specimen cooling
16 Specimen head with
specimen cooling (Option)
17 90° Prism
(instruments with specimen cooling only)
18 Knife holder base 19 Knife holder CE 20 Knife holder CN 21 Knife holder CS 22 Footswitch with
protective guard
Leica CM3050 S – Cryostat
21
Page 22
4. Instrument Properties
15
17
16
11
11
13
14
21
20
22
19
18
Instruction Manual V1.2 - 11/2006
Page 23
4. Instrument properties
4.2 Technical Data
Operating temperature range (ambient temperature): + 18°C to + 40°C. All specifications related to tempe­rature are valid only up to an ambient temperature of + 22°C and relative air humidity of less than 60 %!
Type CM3050S-10 CM3050S-1 CM3050S-8 CM3050S-9
Marks of conformity - - - ­Nominal voltage 100 V AC ±10% 230 V AC ±10% 240 V AC ±10% 100 V AC ±10% Nominal frequency 50 Hz 50 Hz 50 Hz 60 Hz Power draw 1800 VA 1800 VA 1800 VA 1800 VA Max. start-up current for 5 sec 35 A eff. 25 A eff. 25 A eff. 30 A eff. Protective fuse I I I I Automatic cutout T15A M3 T10A T1 T10A T1 T15A M3 Pollution degree 2222 Overvoltage installation category II II II II Heat emission (max.) 1800 J/s 1800 J/s 1800 J/s 1800 J/s
Refrigeration system 50 Hz 60 Hz
Cryochamber
Temperature selection range 0° C to -40°C (+ 2 K / - 0 K) 0° C to -40°C (+2 K / - 0 K) Defrosting Automatic hot-gas defrost cycle, Automatic hot-gas defrost cycle,
Refrigeration capacity 690 W 690 W Safety factor 3 3 Refrigerant 280 g ±5g refrigerant R 404A * 280 g ±5g refrigerant R 404A* Compressor oil 0.6 l EMKARATE RL-22S, ICI * 0.6 l EMKARATE RL-22S, ICI *
programmable in 15-min increments; programmable in 15-min. increments; 1 automatic defrost cycle per 24 hours. 1 automatic defrost cycle per 24 hours. Duration: 6 - 12 minutes Duration: 6 - 12 minutes Manual defrost cycle Manual defrost cycle
Quick-freeze shelf
Maximum temperature - 43°C (+ 0 K / - 2 K) - 43°C (+ 0 K / - 2 K) Number of quick-freeze stations 10 10
Specimen cooling (Option)
Temperature range -10° C to -50°C ±3K -10° C to -50°C ±3K Defrosting Manual defrost cycle (electric) Manual defrost cycle (electric) Refrigeration capacity 320 W 320 W Safety factor 3 3 Refrigerant 185 g ±5g refrigerant R 404A * 185 g ±5g refrigerant R 404A * Compressor oil 0.4 l alpha 22, Kyodo * 0.4 l alpha 22, Kyodo *
or RENISO E22, Fuchs * or RENISO E22, Fuchs *
Leica CM3050 S – Cryostat
23
Page 24
4. Instrument properties
*) Refrigerant and compressor oil to be replaced only by trained and authorized service personnel!
CM3050S-3 CM3050S-6 CM3050S-7
c-UL - ­120 V AC ±10% 208 V AC ±10% 230 V AC ±10% 60 Hz 60 Hz 60 Hz 1800 VA 1800 VA 1800 VA 35 A eff. 30 A eff. 25 A eff. II I T15A T1 T12A T1 T10A T1 22 2 II II II 1800 J/s 1800 J/s 1800 J/s
Fluorescent light lamp (cryochamber illumination):
50 Hz version: Osram Dulux S 11 W/21
Light color ‘LUMILUX hellweiß’
(brilliant white)
60 Hz version: Osram Dulux S 13 W/21
Light color ‘LUMILUX hellweiß’
(brilliant white)
Cryocabinet
Microtome
Rotary microtome Section thickness setting 0.5 - 300 μm Overall specimen feed 25 mm Vertical stroke 59 mm Specimen retraction 50 μm Maximum specimen size 40 x 55 mm Specimen orientation 8° (x-, y-, z-axis) Trimming (6 thickness settings) 5 - 150 μm Motorized coarse feed
- slow 500 μm/s
- fast 1,000 μm/s
Sectioning motor (Option):
Sectioning speed
- slow min. 0.1 mm/s max. 100 mm/s v
max
210 mm/s
- fast min. 0.1 mm/s max. 170 mm/s v
max
210 mm/s
Dimensions
Width (w/o handwheel) 800 mm Width (including handwheel) 882 mm Depth 766 mm Height (arm rest level) 840 mm Overall height 1040 mm
Weight
(incl. microtome, w/o specimen cooling) ca. 180 kg
24
according to IEC-1010, UL 3101
1
2
according to CECOMAF: Liquid temperature 45°C Evaporation temperature -25°C
Instruction Manual V1.2 - 11/2006
Page 25
5. Operation
5.1 Operating the instrument for the first time
Prior to setting up the instrument, make sure to read the instruction manual carefully and in full.
When setting up the instrument, please follow the individual chapters in the following order:
First, study chapters 1, 2, and 3 - install the instrument as described in chapter 3. The overview in chapter 4.1 is helpful during the unpacking and instal­lation procedure in order to identify individual parts and place them in the correct location.
Go on to chapter 5. Read chapters 5.1, 5.2, 5.3 and 5.4 to learn all about the controls.
Next, start to work with the instrument (chapter 5.5). Follow the instruc­tions in chapter 5.5. Chapter 5.5.2 applies to instruments with specimen cooling only. The temperature chart in chapter 6.3 is useful when it comes to select­ing appropriate sectioning temperatures for different types of speci­mens.
Leica CM3050 S – Cryostat
When disinfecting / cleaning the instrument and/or removing / rein­stalling the microtome, please follow the instructions in chapter 7.
In case of problems and malfunctions, please refer to chapters 6.1 and
6.2. These chapters list the most common types of problems which many times have simple causes that can be eliminated by the actual user. If you find ‘Call Technical Service’ in the ‘Remedy’ column, please do so. Never try to solve these problems yourself!
25
Page 26
5. Operation
5.2 Switching on/fuses
2 (Automatic cutout
for sectioning motor)
1 (Combination Mains switch /
automatic cutout for the instrument)
Switching on
To switch on the instrument, place switch (1) in upper lock-in position.
Mains switch / automatic cutout for the instrument:
Automatic cutout for sectioning motor:
The instrument is equipped with an automatic cutout (2) for the sectioning motor.
If the sectioning motor is overloaded (e.g. when sectioning extremely hard specimens etc.) the automatic cutout (2) will be triggered.
To make the instrument functional again after the automatic cutout (2) has been triggered:
Push automatic cutout (2) inwards until it locks into position.
In addition to its function as ON/OFF switch, switch (1) is also the automatic cutout for the in­strument.
If the automatic cutout has been triggered (e.g. caused by overcurrent):
Put switch (1) back in upper lock-in position.
26
Instruction Manual V1.2 - 11/2006
Page 27
5. Operation
5.3 Control panel 1
5.3.1 Key functions in control panel 1
ON/OFF key for cryochamber illumination.
To select the menu items for setting / changing instrument parameters.
In the status indication, pressing the ar­row keys displays the individual current parameters. In the menu, pressing the arrow keys changes set values. As soon as a value is displayed, it is au­tomatically stored.
Activates the manual defrost cycle (cry­ochamber). This key is functional only when one of the menu items ‘SET TEMP CT ... °C’ or ‘SET TEMP OT ... °C’ (last one only applies to instruments with specimen cooling) has been selected first.
Lock / unlock key:
- Locks the display to protect against accidental changes of set parameters.
- Activates / deactivates a number of functions.
5.3.2 Display functions in control panel 1
The display in control panel 1 combines several different functions:
1. When switching the instrument on:
A series of standard indications is displayed, during which two basic parameters can be se­lected (with the arrow keys): Language (of display indications) and speed level of sectioning motor. See chapter 5.3.3 ‘Display indications when
switching on’.
2. Menu function (activated via menu key):
Via the menu key a number of parameters can be selected, then be activated /deactivated via the lock/unlock key and/or changed via the ar­row keys. See chapter 5.3.4 ’Menu functions: Setting param-
eters of refrigeration, time and preset counter’.
3. Status indication function:
The actual status of selected parameters is in­dicated in the display screen. Some of those parameters are displayed auto­matically (e.g. sectioning stroke phase, coarse feed position, handwheel locked). Others can be displayed by pressing the arrow keys; e.g. selected section thickness, preset counter status etc. See chapter 5.3.5 ‘Status indication’ and chap­ter 5.3.6 ‘Displaying state of refrigeration, sec­tion thickness and counters’.
27
Page 28
5.3.3 Display indications when switching on Meaning of indication:
Initialization phase 1
8 888 8 8888888 8888 8 888 8 8888888 8888
5. Operation
Operator activity:
---
Initialization phase 2
             
Software version
CM3 0 5 0S Vs . X X . XX
Selected display language
CUT T I NG MOTOR
Sectioning motor speed level ‘slow’ / (fast)
yes
Temperature upon switching on 0 °C
yes
no
CT . . ..°C OT . . . . ° C *)
DR Y MI CRO T OME
ENGL I S H
SLOW
Temperature prior to switching off < 0 °C
---
---
Select language via
keys
Select speed level via
keys
no
CT....°C OT....° C *)
POWER FA I L URE
Acknowledge indication by pressing
once briefly.
CT . . ..°C OT . . . . ° C *)
12 : 0 0
If this indication is displayed: Do not switch on the instrument! Prior to switching on, both microtome and accessories must be completely dry!
Standard indication ‘locked’ is displayed: (Display background dark) To unlock: press ground lights up:
until the display back-
Remove microtome/accessories, disinfect if applicable, then dry thoroughly and reinstall. (--> See chapters 7.1 - 7.5).
*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.
28
CT....°C OT....°C

HOME ❄❄
Instruction Manual V1.2 - 11/2006
Page 29
5. Operation
5.3.4 Menu functions: setting parameters of refrigeration, time and preset counter
Once the instrument has gone through the initial sequence of indications, you can go on to the menu functions to select and/or change instru­ment parameters.
If display is locked: Press key until display background lights up.
Next, press the menu key. With each key stroke the following parameters will be displayed (dis­play sequence as listed):
1. Specimen temperature (in instruments with specimen cooling only) (-10 - -50 °C)
2. Cryochamber temperature (0 - -40 °C)
3. Selected number of sections - preset counter (ON/OFF and setting selection from 1 to 999)
4. Maximum specimen temperature (ON/OFF)
5. Real time (in 1-minute steps)
6. Defrost time (starting time of automatic cryochamber defrosting) (in 15-minute steps)
7. Duration of defrost cycle (cryochamber defrosting) (6 to 12 minutes duration)
1. Specimen temperature: (instruments with specimen cooling only)
Press key until menu item ‘Set specimen temperature’ is displayed:
SET TEMP OT - - - °C
--
Press key briefly. The line in the specimen temperature field is replaced by an asterisk.
SET TEMP OT- --°C
-
From this point on the refrigeration sys­tem is activated and the desired set temperature can be selected with one of the arrow keys. Press arrow key and release upon reaching the desired value.
CT . . ..°C OT - 35 °C

HOME -
Note on the arrow key function:
Each time one of the arrow keys is pressed briefly, the value displayed in the menu will be changed one step up or down. If an arrow key is pressed and held, the value will change continually (auto-repeat function). When holding the key, scrolling speed will in­crease after a while. As soon as a value is displayed, it is at the same time automatically stored.
Menu items 5 and 6 displayed on 24 hour clock.
29
Page 30
5. Operation
2. Cryochamber temperature:
Press key until menu item ‘Set chamber temperature’ is displayed:
--> Continue as described for specimen temperature.
3. Preset counter (instruments with sectioning motor only):
Press key until menu item ‘Preset counter’ is displayed:
PRE S ET COUNTER
dn- --
= Preset counter activated
PRE S ET COUNTER
dn056
= Preset counter deactivated
Preset counter function:
The preset counter allows the user to carry out a predefined number of sections in the ‘Continu­ous stroke’ sectioning mode. Once the preselected number of sections has been carried out (= section counter reading ‘0’), the specimen head will stop automatically in the lower limit position.
The reference value (i.e. the value from which the preset counter counts downward) can be selected in the corresponding menu item.
In order to be able to set a reference value, the preset counter function must be active.
If the preset counter is deactivated (see diagrams above), press
to activate.
Now there are two options:
1. Increase or decrease the actual reference value by pressing
2. Press reload the former reference value (‘0’ when
setting up the instrument) and then press ei­ther the desired reference value. Depending on
where you start from and where you want to go, method 1 or method 2 is quicker (see ex­amples).
Example 1:
Actual preset counter reading = 50 Former reference value = 300 New reference value of 65 to be selected.
Method 1 is most convenient:
Increase from 50 to 65 by pressing
Example 2:
Actual preset counter reading = 50 Former reference value = 300 New reference value of 320 to be selected.
Method 2 is most convenient:
Press reload old reference value (300):
• Increase from 300 to 320 by pressing
For further information on the preset counter, please refer to chapter 5.3.6, Item 4 ‘Preset counter’.
and simultaneously in order to
or individually in order to select
and simultaneously to
or .
.
.
30
Instruction Manual V1.2 - 11/2006
Page 31
5. Operation
4. Quick-freezing at maximum specimen temperature* (instruments with specimen cooling only)
Press key until menu item ‘Fast object freezing’ is displayed:
FAST OBJ FREEZIN
OF F
or
FAST OBJ FREEZ IN
ON
Press lock/unlock key to activate or de­activate (ON/OFF).
5. Real time
Press key until menu item ‘Time’ is dis­played:
TIME 10:23
Use arrow keys (
or ) to set real time. Setting minutes as well as hours in 1-minute steps. Holding arrow button in­creases scrolling speed!
6. Starting time of automatic defrost cycle:
Press key until menu item ‘Defrost time’ is displayed:
DEFR T IME 4 : 30
(Time on display = start time of defrost cycle)
Use arrow keys ( or ) to select the de­sired defrost start time.
The automatic defrost cycle takes place once every 24 hours at the selected defrost time.
7. Duration of defrost cycle:
Press key until menu item ‘Defrost period’ is displayed:
DEFROST PER I OD CHAMBER : 0 8 MIN
Use defrost keys ( frost time.
(Only if necessary - see below).
Important information for setting the defrost parameters Starting time of defrost cycle: We recommend to program the automatic defrost cycle for a time out­side regular working hours as a period of 4 hours should pass after the defrost cycle before work with the instrument is resumed (see also chapter 5.7.1 ‘Automatic chamber defrost cycle’).
Defrost cycle duration: The duration of the chamber defrost cycle is set to 8 minutes in factory. This setting should only be changed, if the evaporator is not completely de-iced during the 8-minute cycle. In such cases, increase the defrost cycle duration gradually until the the evaporator is ice-free after defrosting. The selected defrost cycle duration is valid for both automatic and manual chamber defrost cycle.
or ) to adjust de-
*) with 90° prism - see also chapter 5.5.2.
Leica CM3050 S – Cryostat
31
Page 32
5. Operation
5.3.5 Status display
Each individual parameter is displayed in a certain area of the display screen.
Upper line:
In the menu (see chapter 5.3.4): indication of selected menu item. Standard indication: Chamber temperature (instruments with specimen cooling: chamber and specimen temperature)
Standard indication upper line: Instruments with chamber and specimen cooling
CT-30°C OT-35°C

HOME 
Instruments with chamber cooling only
CT-30°C

HOME  -
Upper line
Lower line
1 2 3 4 5 6 7 8 9 10111213141516
1 2 3 4 5 6 7 8 9 10111213141516
Digits 1 and 2
indicate sectioning stroke position:
Sectioning stroke (outside sectioning window)
: Retraction stroke
[ ]: Sectioning stroke
(inside sectioning window)
Sectioning
○○○
○○
○○○○○○○○○
stroke
Sectioning window [
○○
]

○○○○
○○
Retraction stroke
Digits 3 - 6
indicate coarse feed position:
• Rear limit position: HOME
• Front limit position: STOP
Indication of ‘malfunction’ or ‘handwheel locked’:
• Microtome plug loose,
implausible values etc.: ????
• Handwheel is locked: LOCKED
Indication ‘LOCKED’ completely overlays digits 3 through 16, i.e. while ‘LOCKED’ is displayed, no other parameters will be displayed in digits 3 - 16 of the lower line.
• Further indications: see chapter 6.1 ‘Error messages indicated in the display screen’
Digits 7 - 16
indicate the following parameters:
• Refrigeration status: Digit 15 = cryochamber Digit 16 = specimen Meaning of refrigeration symbols:
-
Refrigeration OFF
Refrigeration ON
Defrosting active
• Section thickness (set value)
• Section counter (actual value)
• Indication of total section thickness (actual value)
• Preset counter
Digits 7 - 16:
Scroll via arrow buttons to display individual parameters
32
Instruction Manual V1.2 - 11/2006
Page 33
5. Operation
5.3.6 Indication of refrigeration state, section thickness setting and counters
These parameters can be displayed via the ar­row keys. When pressing
, the parameters are dis­played in the sequence listed on this page; when pressing
, they are displayed in re­verse order. The display can be scrolled contin­uously, i.e. after the last parameter the display will automatically start again with the first.
1. Refrigeration state
CT- 3 0 °C OT- 35 °C

HOME 
Meaning of the refrigeration symbols (asterisks):
--> see chapter 5.3.5 ‘Status indication’.
2. Section thickness
CT- 3 0 °C OT - 35°C
μ010.0
The sectioning thickness is set with the
or
key (see chapter 5.4.7 ‘Setting the section
thickness’).
3. Section counter
CT- 3 0 °C OT- 35 °C
∑∑
n000
∑∑
Once displayed (via arrow keys), the section counter can be set to ‘0’ at any time.
To set to ‘0’ press
.
If the section counter is not set to ‘0’ by the user at some point, sections are counted up to ‘999’ at which point the counter automatically switches back to ‘0’.
4. Total section thickness counter
CT- 3 0 °C OT - 35°C
∑∑
μ0000.0
∑∑
Can also be set to ‘0’ at any time (via
key). If the total section thickness indication is not set to ‘0’ manually, the instrument counts up to maximum
9999.5 μm at which point the counter will automati­cally switch back to ‘0’. This indication shows the total thickness of all sections carried out since the counter was last set to ‘0’. (
All sections, in sectioning and trim-
ming mode, are counted).
5. Preset counter
CT-30°C OT-35°C
dn056
Setting preset counter to ‘0’:
Press
key.
can also be pressed while the continuous sectioning mode is in motion; in this case the preset counter is set to ‘0’ and sectioning stops upon reaching the lower limit position. After it has been set to ‘0’, the preset counter is inactive.
Reloading the former reference value:
Press
and simultaneously.
Reloading the former reference value is possible also while the continuous sectioning mode is in motion: Upon pressing both arrow keys, the instrument continues sectioning without interruption, start­ing to count down from the old reference value.
33
Page 34
5. Operation
5.4 Control panel 2
Instruments without sectioning motor Instruments with sectioning motor
In instruments without sec­tioning motor, control panel 2 only consists of functional area B (see below)
In instruments with section­ing motor, the control panel is equipped with three func­tional areas: A, B and C (see large diagram).
A
Functional area A:
• Sectioning modes
• Sectioning window
• Sectioning speed
• Emergency stop function
B
Functional area B:
• Coarse feed
• Trimming
• Section thickness selection
34
B
Functional area B:
• Coarse feed
• Trimming
• Section thickness selection
C
Functional area C:
• Start / Stop sectioning
Instruction Manual V1.2 - 11/2006
Page 35
5. Operation
Functional areas A and C - Instruments with sectioning motor only
5.4.1 Selecting a sectioning mode / start and stop sectioning
Three different sectioning modes can be select­ed for motorized sectioning:
Sectioning mode 1 - Intermittent stroke: The specimen head travels up and down as long as the RUN/STOP and RUN/ENABLE keys are pressed simulta­neously or as long as the footswitch is depressed.
Sectioning mode 2 - Continuous stroke: The specimen head travels up and down until it is stopped by pressing RUN/STOP or RUN/ENABLE or by depressing the footswitch.
Sectioning mode 3 - Single stroke: The specimen head carries out one sec­tioning stroke and then stops in the low­er limit position.
Activating a sectioning mode
Press one of the three sectioning mode keys de­scribed above. While a sectioning mode is ac­tive, the LED in the corresponding key will be il­luminated.
Start / stop motorized sectioning
Sectioning mode 1 - Intermittent stroke:
Simultaneously press and hold RUN/STOP and RUN/ENABLE or depress and hold footswitch.
To stop the sectioning stroke, release the keys or the footswitch: The specimen head stops immediately.
Sectioning mode 2 - Continuous stroke:
Simultaneously press and release RUN/STOP and RUN/ENABLE or depress and release footswitch.
To stop sectioning: Press and release RUN/STOP or RUN/EN-
ABLE or depress and release the footswitch: The specimen head stops upon reaching the
lower limit position.
Sectioning mode 3 - Single stroke:
Simultaneously press and release RUN/STOP and RUN/ENABLE or depress and release footswitch.
The instrument carries out one complete section / retraction stroke and the specimen head stops upon reaching the lower limit po­sition.
While sectioning is in motion, LED ‘RUN’ is illu­minated. While the sectioning motor is deactivated, LED ‘STOP’ is illuminated. Both LEDs are located in functional area C.
35
Page 36
5. Operation
Functional areas A and C - Instruments with sectioning motor only
5.4.2 Setting the sectioning window
The term ‘Sectioning window’ stands for a dif­ferent (slower) speed setting in that part of the sectioning stroke where the specimen is actual­ly cut, while the remaining sectioning stroke (above and below the specimen) as well as the retraction (return) stroke are carried out at a higher speed level (time-saving feature).
If a sectioning window is set, speed inside the window is as set on the control panel. Outside the sectioning window the specimen head trav­els at a proportionately higher speed.
Always set the sectioning window while in the sectioning stroke (speci­men moves downward). Setting the window in retraction stroke (speci­men travels upward) is not possible.
Position lower edge of specimen approx. 3 mm above the knife edge (rotate hand­wheel clockwise).
The value has been stored when the LED in the sectioning window key lights up briefly.
If the LED flashes, the value has not yet been stored, because the user has tried to set a win­dow margin outside the possible window range (the area in which the sectioning window can be set is slightly smaller than the total section stroke).
• In that case, repeat the step in question.
.
The sectioning window cannot be deleted. If need be, it can be set to maximum size (i.e. almost the en­tire sectioning stroke will be car­ried out at one and the same speed).
Every time after the instrument is disconnected from mains, the sec­tioning window is set to maximum size upon switching back on. This also applies when the instrument is switched on for the first time.
Press sectioning window key.
Position upper edge of specimen just below the knife edge (continue to rotate the hand­wheel clockwise).
Press sectioning window key once more.
36
Instruction Manual V1.2 - 11/2006
Page 37
5. Operation
Functional areas A and C - Instruments with sectioning motor only
5.4.3 Selecting sectioning speed
Sectioning speed is selected via the sliding
potentiometer. Speed can be changed while sectioning is in motion.
Caution when working in continuous mode:
If the slider is set to ‘0’, the sectioning motion stops, however, the actual sectioning mode is still active - just operating at ‘0’ speed.
If the slider is accidentally moved, sectioning will be resumed! - Risk of injury!
Therefore:
Do not interrupt sectioning / trim­ming by setting the sliding potenti­ometer to zero speed! See also safety instruction in Chap­ter 2.3.6 ‘Motorized sectioning’.
Vmax key
5.4.4 Emergency stop
Activate by:
either pressing the emergency stop button forcefully (must lock in position),
or depressing the footswitch force­fully.
Either way, sectioning is interrupted at once.
In both cases, the red area A lights up, indicating that the emergency stop function has been activated. Simultaneously all 3 sectioning modes are deactivated (= all 3 LEDs off).
To deactivate the emergency stop function, rotate the button in the direction of the ar­row until it unlocks and slips upward into its original position.
As soon as the emergency stop button has been unlocked, the red
If the emergency stop function has been ac­tivated with the footswitch, unlocking is not necessary.
LED in functional
LED is extinguished.
As long as the button is pressed, the sectioning motion is carried out at maxi­mum speed (210 mm/s).
While Vmax is active, the sectioning window is inactive.
When the Vmax key is released, the sectioning window becomes active again.
As soon as the footswitch is released, the red
LED is extinguished.
To continue work, select one of sectioning modes 1 - 3 and restart the sectioning motion.
See also chapter 2.2.3 ‘Safety - Emergency stop function’ for further information on the emer­gency stop.
37
Page 38
Functional area B - all models
5. Operation
5.4.5 Coarse feed keys
The four coarse feed keys in functional area B function as follows:
Coarse feed ‘fast backward’
- Travelling speed 1,000 μm/s
Press Specimen head travels backward
(= away from knife):
a. until it reaches the rear limit position,
b. until it is stopped by pressing one of the
coarse feed keys (to stop the function, any of the four coarse feed keys can be pressed).
To start, press Specimen head travels backward (= away
from knife) as long as the key is held.
key to activate.
Coarse feed ‘slow backward’
- Travelling speed 500 μm/s
and hold.
LED ‘HOME’ (located in key ‘Coarse feed fast backward’)
Coarse feed ‘fast forward’
- Travelling speed 1,000 μm/s
To activate the function, press
and hold.
The specimen head travels forward (= to­wards the knife) as long as the key is held.
Coarse feed ‘slow forward’
- Travelling speed 500 μm/s
To activate the function, press
and hold.
The specimen head travels forward (= towards the knife) as long as the key is held.
‘STOP’-LED (located in key ‘Coarse feed fast forward’)
As long as the forward coarse feed (both fast and slow) is in motion, the ‘STOP’ LED is illumi­nated.
Upon reaching the front limit position, the ‘STOP’ LED starts flashing and the display in control panel 1 reads ‘STOP’.
As long as the backward coarse feed (both fast and slow) is in motion, the ‘HOME’ LED is illuminated.
Upon reaching the rear limit position, the ‘HOME’ LED starts flashing and the display in control panel 1 reads ‘HOME’.
38
Instruction Manual V1.2 - 11/2006
Page 39
5. Operation
Functional area B - all models
5.4.6 Trimming and sectioning functions Switching between trimming and sectioning
Press this key to alternate between the two functions.
illuminated: Trimming mode is active
off: Sectioning mode is active.
Selecting trimming thickness
Activate trimming mode
Select trimming thickness (= sectioning thickness at which trimming is carried out) by pressing
The selected thickness is displayed in the cor­responding LED
or .
:
Motorized trimming
5.4.7 Setting the section thickness
Activate sectioning mode ( off).
CT....°C OT....°C
μ010.0
Press Pressing one of these keys once makes the display in control panel 1 switch to section thickness indication (thickness value not yet
being altered). Pressing starts to change the actual section thick­ness setting.
Selectable section thickness settings:
0 - 2 μm in 0.5-μm steps,
2 - 10 μm in 1-μm steps, 10 - 20 μm in 2-μm steps, 20 - 60 μm in 5-μm steps, 60 - 100 μm in 10-μm steps,
100 - 300 μm in 50-μm steps.
or to select section thickness.
or further
Select ‘Continuous stroke’ in functional area A.
Activate trimming by pressing RUN/STOP and RUN/ENABLE (located in functional area C) simultaneously or by depressing the footswitch.
To stop trimming press RUN/STOP or RUN/ ENABLE or depress the footswitch again.
Trimming thickness can be changed while trim­ming is in motion by pressing
or .
39
Page 40
5. Operation
5.5 Daily operation of the instrument
Select chamber temperature and, in instru­ments with specimen cooling, specimen temperature according to the type of speci­men to be sectioned (see also chapter 6.3 ‘Reference chart for temperature settings’). Do not continue with step 5.5.1 before the
• actual chamber / (specimen) temperature corresponds to the set value. This may take several hours (e.g. when operating the in­strument for the first time or when the instru­ment had been switched off).
Prior to operating the instrument , please read and observe the fol­lowing safety instructions:
• ‘Integrated safety devices’ (Chapter 2.2) .
• ‘Handling microtome knives / blades’ (Chapter 2.3.4)
• ‘Knife guard / Handwheel lock’ (Chapter 2.3.5)
5.5.1 Freezing specimens onto specimen discs with quick-freeze shelf and stationary heat extractor
Size the specimen.
Apply sufficient amount of embedding medi-
um onto ambient temperature specimen disc.
Inside the cryochamber, place the specimen
onto the disc and orient.
Place specimen disc with specimen into one
of the 10 openings in the quick-freeze shelf (see below) and freeze.
12
3
40
• ’Motorized sectioning’ (Chapter 2.3.6)
• ‘Defrosting / handling frozen tissue’ (Chapter 2.3.7)
• ‘Frozen instrument parts and accessories’ (Chapter 2.3.8)
• ‘Infectious / radioactive material’ (Chapter 2.3.9)
The freezing process can be sped up further by using the stationary heat extractor (1). Next to the low temperature stabilizer (2) (‘Parking station’), 3 specimen discs (3) can be placed within its pivoting range.
Place the heat extractor (1) directly onto the specimen surface. After approx. 30 seconds the specimen is completely frozen.
Instruction Manual V1.2 - 11/2006
Page 41
5. Operation
5.5.2 Freezing specimens onto specimen discs with 90 ° prism
(Instruments w/specimen cooling only)
In instruments with specimen cooling, the 90 ° prism, which is inserted into the specimen head, is an additional option to freeze specimens at maximum temperature .
4
Installing the 90 ° prism
Loosen screw (3) at specimen head.
Insert pin (1) of prism completely into recep-
tacle (2) in the specimen head. Important: the entire rear surface of the
prism must fit tightly against the specimen head surface!
Retighten screw (3) at specimen head.
The prism now forms a horizontal surface for quick specimen freezing.
Freezing specimens with the 90 ° prism
Apply sufficient amount of embedding medi-
um onto ambient temperature specimen disc, add specimen and orient.
Insert specimen disc with specimen into the
receptacle (4) of the prism.
Press
Freezing’ and activate function by pressing
to activate menu item ‘Fast Object
The specimen head is now automatically cooled down to the lowest temperature (-50 °C).
Once the specimen is frozen, remove the disc from the prism and store it in the quick­freeze shelf until inserting it into the speci­men head.
Press Object Freezing’ and deactivate the function by pressing
5.3.4 ‘Menu functions’ item 4).
The specimen head temperature now automati­cally returns to the previously selected set val­ue.
Loosen screw (3) at specimen head and re­move prism (1).
Store prism inside the cryochamber.
again to activate menu item ‘Fast
once again (see chapter
5.5.3 Optional accessories for freezing
specimens onto specimen discs and for removing frozen specimens
Mobile heat extractor
In addition to the stationary heat extractor (part of standard delivery), a mobile heat extractor is available as an optional accessory - see Ap­pendix ‘Ordering information - accessories ):
Store heat extractor in chamber to precool.
Place onto specimen in quick-freeze shelf.
Remove as soon as specimen is frozen.
(see chapter 5.3.4 ‘Menu functions’ item 4).
41
Page 42
5. Operation
Thermoblock
The thermoblock makes it easy to remove fro­zen specimens from the specimen discs.
Do not store the thermoblock inside the cryochamber - store at room temperature and only place into the chamber for removing specimens from discs.
5.5.4 Inserting specimen discs into the specimen head
Lock handwheel in upper position.
If a knife has already been inserted into the knife holder, cover knife edge with knife guard!
Place cap (9) on one of the two ends so that the correct size opening (7) for the disc to be inserted is exposed.
Insert pin (6) of specimen disc (5) into the opening (7).
After about 20 seconds remove the frozen specimen from the specimen disc (use tweezers).
If cap (9) fits too loosely, readjust with screw (10). Do not tighten screw too much!
Once the specimen has been detached from the disc, remove the thermoblock from the cryochamber.
42
Loosen clamping screw (3).
Insert specimen disc in opening (2) (rear surface of specimen disc must fit tightly against surface of specimen head).
Tighten clamping screw (3).
5.5.5 Specimen orientation
Open clamping lever (4) --> (4a).
Orient specimen via orienting screws (1) and (5).
Retighten clamping lever (4).
Instruction Manual V1.2 - 11/2006
Page 43
5. Operation
5.5.6 Trimming / sectioning
Inserting the knife
Insert the knife / disposable blade into the knife holder.
Select appropriate clearance angle at knife holder (settings between 4° and 6° are suit­able for the majority of applications).
For details on how to insert and ad­just the knives, please refer to the separate instruction manual(s) for your knife holder(s). See also chapter 2.2 ‘Integrated safety devices’ incl. subchapter
2.2.4 ‘Knife guard’.
Manual trimming
Adjust knife / knife holder relative to speci­men (see ‘Knife holder’ instruction manual).
Unlock handwheel.
Use motorized coarse feed to bring speci­men close to knife (see chapter 5.4.5 ‘Coarse feed functions’).
Select desired section thickness (see chap­ter 5.4.7 ‘Setting the section thickness’).
Move anti-roll plate away from knife.
Rotate handwheel to trim specimen down to desired sectioning plane.
Use motorized coarse feed to bring speci­men close to knife (see chapter 5.4.5 ‘Coarse feed functions’).
Move anti-roll plate away from knife.
--> Next steps: see chapter 5.4.6 ‘Trimming’.
Sectioning
Adjust knife / knife holder relative to speci­men (see ‘Knife holder’ instruction manual).
Place anti-roll guide onto knife.
Orient specimen, if necessary (see chapter
5.5.5 ‘Specimen orientation’).
Select desired section thickness (see chap­ter 5.4.7 ‘Setting the section thickness’).
Start sectioning at approx. 20 μm, reduce section thickness gradually until reaching the appropriate thickness for the type of specimen to be sectioned.
Discard the first 2 to 3 sections cut at the de­sired section thickness.
When sectioning manually, rotate the hand­wheel evenly and at uniform speed.
For all details on motorized sectioning, please refer to chapters 5.4.1 to 5.4.4:
‘Selecting a sectioning mode / Start and stop sectioning’, ‘Setting the sectioning window’, ‘Selecting sectioning speed’, ‘Emergency stop’.
Motorized trimming
Adjust knife / knife holder relative to speci­men (see ‘Knife holder’ instruction manual).
Unlock handwheel.
Center handwheel grip (See chapter 2.2.2).
For advice on appropriate Sectioning tempera­tures for different types of tissue, please refer to
chapter 6.3 ‘Reference chart for temperature settings’.
43
Page 44
5. Operation
5.6 Finishing work
5.6.1 Daily routine steps
Remove all specimens from the cry­ochamber and store safely! See also corresponding safety instruc­tions in chapter 2.3.7 ‘Defrosting / handling frozen tissue’!
Lock handwheel.
Remove knife from knife holder, disinfect, dry (see chapter 7 ‘Disinfection, cleaning and maintenance’) and store in the knife case.
Use a cold brush to sweep frozen sectioning waste into the section waste tray.
Empty section waste tray (dispose of sec­tioning waste according to the applicable regulations on disposal of special category waste!).
If necessary, carry out a complete disinfection procedure (see chapter 7 ‘Disinfection, cleaning and maintenance’).
Remove all specimens from the chamber and store safely.
Close sliding window.
Switch off cryochamber illumination.
Use lock/unlock key to lock display in control panel 1 (Display background dark = locked):
5.6.2 Switching off the instrument for a longer period of time
If you do not plan to use the instrument for an extended period of time, it can be switched off.
However, since this involves a lot of work (see below), it should be considered carefully whether switching off the instrument is really the better option.
If the instrument is switched off, the microtome must be removed. Then the microtome, cryo­chamber and all accessories must be disinfect­ed and dried!
For that purpose, follow the instructions in
chapters 7.1 through 7.5.3.
44
CT. . ..°C OT . . . . °C
12 : 00
Do
not switch off mains switch / automatic cutout (see chapter 5.2 ‘Switching on / fuses’)!
Instruction Manual V1.2 - 11/2006
Page 45
5. Operation
5.7 Defrosting
The Leica CM3050 S has 3 (instruments without specimen cooling: 2) different defrosting options.
- Automatic chamber defrost cycle
- Manual chamber defrost cycle
- Specimen head defrosting (manual) - instru­ments with specimen cooling only.
Read the safety instructions in chap­ter 2.3.7 ‘Defrosting / frozen speci­mens’ !
Make sure to remove sensitive speci­mens from the cryochamber before the defrost cycle starts!
5.7.1 Automatic chamber defrost cycle
The automatic defrost cycle is carried out once every 24 hours at the preselected time. (For details see chapter 5.3.4 ‘Menu functions’ ­items 6 and 7).
During the defrost cycle, the corresponding as­terisk is displayed inversely:
The specimen head refrigeration is automatical­ly switched back on when:
at least 4 hours have passed since the de­frost cycle was completed;
and the actual chamber temperature devi­ates less than 5 K from set value.
After the specimen head refrigeration has been switched back on, the specimen head tempera­ture rises briefly to +10 °C and then decreases until it reaches the set value.
Switching specimen cooling back on immedi­ately after defrosting:
As soon as the chamber defrost cycle is com­pleted, specimen cooling can be switched back on immediately:
Press
until menu item ‘Set specimen
temperature’ is displayed:
SET TEMP OT - 35°C
-
Press
.
SET TEMP OT - 35 °C

Specimen cooling is switched back on.
CT-30°C OT-35°C

HOME
Once the defrost cycle is finished, the instrument automatically switches back to refrigeration.
Instruments with specimen cooling:
When defrosting starts, to avoid icing the speci­men head is automatically adjusted to a temper­ature between -10 °C and -5 °C, i.e. the set tem­perature value is deactivated temporarily.
Despite the possibility of switching the specimen cooling system back on immediately after the automatic chamber defrost cycle, we recom­mend the automatic defrost cycle be set at a point in time when the entire procedure can be controlled automatically. (See also information on menu items 6 and 7 in chapter 5.3.4 ‘Menu functions’).
45
Page 46
5. Operation
5.7.2 Manual chamber defrost cycle
In addition to the automatic defrost cycle, a manual chamber defrost cycle can be activat­ed when necessary.
Press
until menu item ‘Set chamber
temperature’ is displayed:
SET TEMP CT - 30 °C
-
Press
. A brief sound signals that the
manual chamber defrost cycle has started:
SET TEMP CT - 30 °C
 -
The manual defrost cycle ends either when the selected defrost cycle duration is over or when a chamber temperature of -5 °C is reached. To abort the manual defrost cycle:
Go back to menu item ‘Set chamber tem­perature’ and press
again.
Instruments with specimen cooling:
5.7.3 Defrosting the specimen head (instruments with specimen cooling only)
To activate specimen head defrosting:
Press
temperature’ is displayed:
Press
manual chamber defrost cycle has started:
Specimen head temperature increases to be­tween +20 °C and +30 °C for a duration of 10 minutes, after which it returns to the set value.
To abort specimen head defrosting:
Press
temperature’ is displayed and press once more.
until menu item ‘Set specimen
SET TEMP OT - 35°C

. A brief sound signals that the
SET TEMP OT - 35 °C
until menu item ‘Set specimen
When defrosting starts, to avoid icing the spec­imen head is automatically adjusted to a tem­perature between -10 °C and -5 °C, i.e. the set temperature value is deactivated temporarily. (same as with automatic defrost cycle - see chapter 5.7.1 ‘Automatic chamber defrost cy­cle’)
10 seconds after the manual chamber defrost cy­cle has been completed, the specimen head re­frigeration is switched back on, cooling down to the set value.
46
Instruction Manual V1.2 - 11/2006
Page 47
6. Troubleshooting, applications tips
6.1 Displayed error messages
The autodiagnosis function of the CM3050 S displays error messages in case of certain mal­functions.
1. Power failure
This error message is displayed after the instru­ment had been disconnected from mains, either due to power failure or because the mains switch had been turned off.
CT- 3 0 °C OT - 35 °C
POWE R FA I L U RE
To acknowledge the message, press about 5 seconds.
for
2. Coarse feed
Coarse feed malfunction has occurred.
CT- 3 0 °C OT - 35 °C
????
This message is also displayed, if the coarse feed plug has not been reconnected properly after reinstallation of the microtome.
3. Temperature indications
CT-##°C OT-35°C
or
CT- 3 0 °C OT - ## °C
If the # symbol is displayed instead of chamber or specimen temperature indication, the corre­sponding temperature sensor is broken or a short circuit has occurred.
4. Service
CT-30°C OT-35°C
SERV ICE
This message is displayed after a certain number of operating hours to remind the user to have a preven­tive maintenance done by Leica Technical Service. This message is displayed only, if the instrument is not regularly ser­viced by Leica Technical Service.
We recommend a maintenance contract once the warranty period has expired!
Contact your local Leica sales office for details!
If one of the error messages listed under 2, 3 and 4 is displayed, please contact Leica Technical Service.
6.1.1 Other error indications
(red ) in control panel 2 is illuminated:
Emergency stop button accidentally activat­ed.
--> Unlock.
Footswitch not properly connected.
--> Check connection - connect properly, if applicable.
47
Page 48
6.2 Potential problems - causes and remedies
6. Troubleshooting, applications tips
Problem
Frost on chamber walls and microtome
Sections smear
Sections splinter Sections not properly flattened
Causes
- Cryostat is exposed to draft (open windows, doors, air­conditioning).
- Sliding window has been open for a long period of time at a very low chamber temperature.
- Frost build-up caused by breathing into the cryocham­ber.
- Specimen not cold enough.
- Knife and/or anti-roll plate not yet cold enough - sections melt.
- Specimen too cold.
- Static electricity / draft.
- Specimen not cold enough.
- Large surface specimen.
- Anti-roll plate poorly adjusted.
- Anti-roll plate poorly aligned relative to knife edge.
- Wrong clearance angle selected.
- Knife blunt or damaged.
Remedies
- Eliminate draft sources or change place of installation of the cryostat.
- Close sliding window.
- Consider wearing a mask.
- Select lower temperature.
- Wait until knife and/or anti-roll plate have reached chamber temperature.
- Select higher temperature.
- Eliminate cause.
- Select lower temperature.
- Trim specimen parallel; in­crease section thickness.
- Readjust anti-roll plate.
- Align correctly.
- Set correct clearance angle.
- Use different part of knife or replace knife.
Sections not properly flattened despite correct temperature and correctly aligned anti-roll plate
48
- Knife and/or anti-roll plate dirty.
- Edge of anti-roll plate dam­aged.
- Blunt knife.
- Clean with dry cloth or brush.
- Replace anti-roll plate.
- Use different part of knife or replace knife.
Instruction Manual V1.1 02/2000
Page 49
6. Troubleshooting, applications tips
Problem
Sections curl on anti-roll plate.
Scraping noise during section­ing stroke and specimen return stroke.
Ondulated sections.
Chatter marks form during sectioning.
Causes
- Anti-roll plate does not pro­trude far enough beyond the knife edge.
- Anti-roll plate protrudes too far beyond the knife edge and is scraping against the specimen.
- Knife damaged.
- Edge of anti-roll plate damaged.
- Specimen insufficiently frozen onto specimen disc.
- Specimen disc not clamped tightly enough.
- Specimen holder ball joint not clamped.
- Knife not clamped tightly enough.
- Selected section thickness too thick - specimen has come off the specimen disc.
- Specimen very hard and inho­mogeneous.
- Blunt knife.
- Knife profile inappropriate for the type of specimen to be sectioned.
- Wrong clearance angle.
Remedies
- Readjust correctly.
- Readjust correctly.
- Use different part of the knife or replace knife.
- Replace anti-roll plate.
- Refreeze specimen onto disc.
- Check disc clamping.
- Check ball joint clamping.
- Check knife clamping.
- Refreeze specimen onto disc.
- Increase section thickness; if possible, reduce specimen surface.
- Use different part of knife or replace knife.
- Use knife with a different pro­file.
- Readjust clearance angle set­ting.
Condensation forms on anti-roll plate and knife during cleaning.
Leica CM3050 S – Cryostat
- Brush, forceps, cloth or other cleaning item too warm.
- Accessories and cleaning material to be used pre-cooled only. - Store on storage shelves inside the cryochamber.
49
Page 50
6. Troubleshooting, applications tips
Problem
Anti-roll plate damaged after adjustment.
Thick/thin sections.
Causes
- Anti-roll plate protrudes too far beyond the knife edge. Adjust­ment was done in the direction of the knife.
- Wrong temperature for the tissue to be sectioned.
- Knife profile inappropriate for the specimen to be sectioned.
- Ice build-up on the knife back.
- Handwheel rotation not uni­form or inappropriate rotation speed.
- Knife not clamped tightly enough.
- Specimen disc not clamped correctly.
- Embedding medium poured onto cold specimen disc; spec­imen comes off the disc once frozen.
- Blunt knife.
- Incorrect clearance angle.
- Specimen dried out.
Remedies
- Replace anti-roll plate; in future, lift away from the knife when aligning.
- Handle anti-roll plate more carefully.
- Select correct temperature.
- Wait until correct temperature has been reached.
- Use knife with different profile (c or d) or possible switch to disposable blade system.
- Remove ice.
- Adapt speed.
- Check knife clamping.
- Check disc clamping.
- Place embedding medium on luke-warm specimen disc, inside the cryochamber place the specimen onto the disc and freeze in quick-freeze shelf.
- Use different part of knife or replace knife.
- Set correct clearance angle.
- Prepare new specimen.
Tissue sticks to the anti-roll plate.
Flattened sections curl up as soon as anti-roll plate is lifted from the knife.
50
- Anti-roll plate too warm or incorrectly adjusted.
- Static electricity.
- Fat on corner or edge of anti­roll plate.
- Knife rusty.
- Static electricity or draft.
- Anti-roll plate too warm.
- Cool or readjust anti-roll plate.
- Eliminate cause.
- Remove fat (alcohol).
- Remove rust.
- Eliminate cause.
- Cool anti-roll plate.
Instruction Manual V1.1 02/2000
Page 51
6. Troubleshooting, applications tips
Problem
Sections tear.
Inconsistent or insufficient specimen feed.
Specimen disc cannot be removed.
Causes
- Temperature too low for the tissue to be sectioned.
- Knife blunt, dirty, dusty, frosted or rusty.
- Upper edge of anti-roll plate damaged.
- Hard particles in the tissue.
- Rear face of knife dirty.
- Microtome was not completely dry when refrigeration was switched on; ice build-up inside the micrometer mechanism.
- Microtome defective.
- Specimen disc stuck to quick­freeze shelf or specimen head due to moisture on underside of specimen disc.
Remedies
- Select higher temperature and wait for the new temperature to be reached.
- Eliminate cause (--> see details on individual problems).
- Exchange anti-roll plate.
- If application permits, section at lower plane.
- Clean.
- Remove microtome from cham­ber and dry - possibly with hair­dryer (hairdryer only to be used outside the cryochamber!) or at ambient temperature.
- Call Technical Service.
- Apply strong alcohol to contact point or heat the specimen head.
Cryostat inoperational.
Sectioning motor inoperational.
No or insufficient refrigerating performance.
Leica CM3050 S – Cryostat
- Mains plug not properly con­nected.
- Automatic cut-out has been triggered.
- Footswitch not connected.
- Automatic cut-out of sectioning motor has been trig­gered.
- Motor defective.
- Stopper not placed properly into drain in bottom of cryo­chamber.
- Compressor defective.
- Leak in cooling system.
- Check mains plug connection.
- Reconnect automatic cut-out (switch instrument back on).
- Connect footswitch.
- Reconnect automatic cut-out (push inwards until it locks in place).
- Call Technical Service.
- Close drain with stopper.
- Call Technical Service.
- Call Technical Service.
51
Page 52
6. Troubleshooting, applications tips
6.3 Reference chart for temperature settings (in minus °C)
Tissue 10 - 15 15 - 25 25 - 50
Adrenal Bone marrow Brain Bladder Breast - fatty Breast - little fat Cartilage  Cervical Fatty Heart and vascular Intestinal Kidney Laryngeal Lip  Liver Lung Lymphoid Muscular Nose Pancreatic Prostate Ovarian Rectal Skin with fat Skin without fat Spleenal or bloody tissue  Testicular  Thyroid Tongue Uterus curettage
Please note: The recommended temperature settings are based on long-term experience. However, they can only be
guidelines for general orientation, as each individual specimen may require particular adjustments.
52
Instruction Manual V1.1 02/2000
Page 53
7. Disinfection, cleaning and maintenance
7.1 Safety instructions on disinfection and cleaning
When disinfecting the instrument, make sure to observe the following safety instructions:
• ‘Disinfection and Cleaning’ (Chapter 2.3.10)
• ‘Handling microtome knives / blades’ (Chapter 2.3.4)
• ‘Defrosting / handling frozen tissue’ (Chapter 2.3.7)
• ‘Frozen instrument parts and accessories’ (Chapter 2.3.8)
• ‘Infectious / radioactive material’ (Chapter 2.3.9)
7.2 Spray disinfection with Leica Cryofect
For practical spray disinfection, we recommend Leica Cryofect which is fully effective at tem­peratures in the negative range down to -20 °C.
Set chamber temperature -20 °C or warmer.
Remove specimens from the chamber and store safely.
Carefully remove knife / blade from knife holder.
Remove section waste tray from cryocham­ber and empty.
Remove remaining section waste from cryo­chamber.
Working accessories (brushes, specimen holders etc.) can be disinfected inside the cryochamber.
Once the selected chamber temperature has been reached, spray Leica Cryofect on the surfaces and accessories to be disinfected.
Allow to soak for 15 minutes.
Use paper towels or another type of disposable cleaning cloths to wipe up disinfectant and remaining section waste.
Dispose of paper towels together with sec­tion waste.
If previously altered, set chamber tempera­ture back to its former value.
Put specimens back into the cryochamber.
53
Page 54
7 Disinfection, cleaning and maintenance
7.3 Disinfection with conventional disinfectants
Switch the instrument off.
Carefully remove knife / blade as well as spec-
imens and accessories from the chamber.
Remove section waste from the chamber
and dispose of it.
Remove sliding window (see chapter 7.6 ‘Ex-
changing the fluorescent light lamp’).
• Remove microtome (see chapter 7.5.1
‘Removing the microtome’).
Allow the microtome to thaw, disinfect and
allow to dry thoroughly.
Drying the microtome:
• Use a hairdryer: dry and before reinstalling into the chamber allow the microtome to cool down to room temperature.
• Allow to dry at room temperature (at least over night - longer, if necessary).
Allow cryochamber to dry thoroughly.
Reinstall microtome (see chapter 7.5.3).
Only when they are completely dry, reinsert the disinfected accessories / tools into the cryochamber.
Reinsert the sliding window.
Switch the instrument back on.
Once a suitable chamber temperature has been reached, return the specimens to the cryochamber.
7.4 Cleaning
When necessary, the instrument surfaces can be cleaned with mild commercial house­hold cleaners.
Surfaces which are potentially contaminated
• (infectious germs) should always be cleaned with alcohol disinfectants!
• Disinfect and dry disassembled parts (accessories and tools).
Disinfect chamber walls with commercial alcohol disinfectant.
Place an adequate container underneath the drain tube at the right side of the instrument.
Pull stopper out of chamber bottom and drain off disinfecting solution into the con­tainer.
Put stopper back into drain in cryochamber.
54
Instruction Manual V1.2 - 11/2006
Page 55
7. Disinfection, cleaning and maintenance
7.5 Removing / reinstalling the microtome
7.5.1 Removing the microtome
When removing the microtome make sure to observe the following safety instructions:
• ‘Removing / reinstalling the microtome’ (Chapter 2.3.11)
• ‘Defrosting / handling frozen tissue’ (Chapter 2.3.7)
• ‘Infectious / radioactive material’ (Chapter 2.3.9)
• ‘Disinfection and cleaning’ (Chapter 2.3.10)
Prior to removing the microtome, carry out preparation steps as per chapter 7.3 (switch instrument off, remove accessories etc.)
7.5.2 Disassembling the specimen head (instruments w/specimen cooling only)
Use Allen key size 5 to remove screw (2)
from the center of the specimen head (1).
Do not twist the refrigerating tube (see
chapter 2.3.11 ‘Removing / reinstalling the microtome’).
Hold hook (4) (part of standard delivery) in
one of your hands and with your other hand rest refrigerating tube (3) on hook.
Insert hook (4) with refrigerating tube (3) into
the opening above mark ().
55
Page 56
7. Disinfection, cleaning and maintenance
Remove plug of temperature sensor (7).
Remove plug of coarse feed motor: Pull bushing backward (arrow 1) and hold. Place your thumb between plug (6) and socket at the microtome (arrow 2). To remove plug (6) from microtome, pull it away from the microtome while at the same time carrying out a slight jogging movement (arrows 3).
7
9
6
8
2
3
3
1
Loosen screws (8) with Allen key.
Lift microtome a little and pull slightly to the left.
This loosens the plastic coupling (11) which connects the two axles.
Remove the microtome from the chamber.
7.5.3 Reinstalling the microtome
When reinstalling the microtome make sure to observe the following safety instructions:
• ‘Removing / reinstalling the microtome’ (Chapter 2.3.11)
• Display message ‘Dry Microtome’ (Chapter 2.3.12)
Lock handwheel in lower position.
Place microtome in chamber.
Put one drop of cryostat oil per surface on the contact surfaces of plastic coupling (11) and handwheel coupling (12).
Place plastic coupling (11) on handwheel coupling (12).
Push microtome slightly towards the right until axle (9) fits tightly into coupling (11).
Unlock handwheel.
Adjust the parts into each other by rotating the handwheel slightly backward and forth.
Lock handwheel again in lower position.
Tighten screws (8).
Reconnect plugs of coarse feed motor (6) and temperature sensor (7).
Mount specimen head (in reverse order as described on previous page).
Insert accessories (storage shelf, section waste tray etc.).
Insert sliding window.
Connect instrument to mains and switch back on.
56
Instruction Manual V1.2 - 11/2006
Page 57
7. Disinfection, cleaning and maintenance
7.6 Exchanging the fluorescent light lamp
Prior to exchanging the lamp, please read and observe the fol­lowing safety instructions:
• ‘Maintenance’ (Chapter 2.3.13)
Removing the sliding window
Hold sliding window (1) at grip bar (2), lift slightly and pull out towards yourself.
Removal
The lamp (3) is located behind the glare shield (4).
Carefully hold lamp with both of your hands. Slightly pull the lamp downward at the left side to remove it from fastening clip (5).
Pull the lamp towards the left to remove it from its socket (6).
Installing the replacement lamp
Hold the lamp in correct position (see draw­ing below) and push towards the right until it locks into the socket.
Push the lamp slightly upwards until it locks into the clip.
Put the sliding window back in place.
Reconnect the instrument to mains and switch it back on.
57
Page 58
8. Warranty and service
Warranty
Leica Microsystems Nussloch GmbH guarantees that the contractual prod­uct delivered has been subjected to a comprehensive quality control procedure based on the Leica in-house testing standards, and that the prod­uct is faultless and complies with all technical specifications and/or agreed characteristics warranted.
The scope of the warranty is based on the content of the concluded agree­ment. The warranty terms of your Leica sales organization or the organiza­tion from which you have purchased the contractual product shall apply exclusively.
Technical service information
If you require technical service or replacement parts, please contact your Leica sales representative or dealer who sold the product.
Please provide the following information:
Model name and serial number of the instrument.
Location of the instrument and name of the person to contact.
Reason for the service call.
Date of delivery.
Decommissioning and disposal
The instrument or parts of the instrument must be disposed of in compliance with the local laws.
58
Instruction Manual V1.2 - 11/2006
Page 59
9. Appendix
1.3 Safety instructions1. Electrohydraulic height adjustment
An electrohydraulic lifting device for individual working height adjustment is available as an optional accessory for the CM3050 S cryostat series.
1.1 Designated use
The electrohydraulic height adjustment device has been designed exclusively for adjustment of the instrument working level to the individual height of the user. It has been designed for use with the Leica cry­ostat series CM3050 S only and may under no circumstances be used to lift other instruments or persons. Damage resulting from improper use of the product and/or faulty operation invalidates the warranty. Leica will not assume liability for any such dam­age.
1.2 Standard delivery
Prior to operating the height adjustment device, make sure to read the present instruction manu­al. Pay special attention to the following safety instructions:
Prior to connecting the height ad­justment device to mains, check to make sure that the nominal voltage and frequency indications on the device name plate correspond to the actual mains voltage in your lab!
Do not move the height adjustment device with the Leica CM3050 S in­stalled on the frame. If the place of installation needs to be changed, first take the Leica CM3050 S down from the height adjustment device and then move both devices sepa­rately.
Electrohydraulic height adjustment device for cryostat series CM3050 S, complete outfit, con­sisting of a U-shaped frame on four height-ad­justable feet and a two-key control unit.
The height adjustment device is available in 2 voltages:
1. 230 – 240 V / 50 Hz – order no. 0443 26147
2. 230 – 240 V / 50 Hz – order no. 0443 26147
Never place your feet underneath the U-frame when lowering the height adjustment device! Risk of injury!
Only authorized Leica technical ser­vice engineers may carry out re­pairs on both the Leica CM3050 S and the height adjustment device. This applies especially to all work carried out on the hydraulic system of the height adjustment device.
59
Page 60
9. Appendix
1.4 Installation
Open the shipping box and remove all packing material including the Sty­rofoam pieces.
Take the height adjustment device out of the box.
At the place of installation, place the U-frame on the floor with the open side facing yourself. Adjust the U-frame parallel to the wall, maintaining a distance of approx. 10 cm between wall and U-frame.
Push the (ab) button to lower the height adjustment device to the lowest possible position.
In order to be able to move the instrument, use a fork wrench and unscrew the rollers on which the Leica CM3050 S moves by about 2 cm.
Roll the cryostat onto the U-frame as far as it will go. The rear of the cryostat housing has to fit closely to the rear bar of the U-frame.
60
When placing the instrument onto the U-frame, make sure the power cord does not get caught and thus crushed between U-frame and cry­ostat housing.
Roll the cryostat onto the U-frame as far as it will go. The rear of the cryostat housing has to fit closely to the rear bar of the U-frame.
Plug the height adjustment device into the mains socket. The unit is now ready for use.
Instruction Manual V1.2 - 11/2006
Page 61
1.5 Control keys 1.7 Troubleshooting
1. Slight and uniform loss of height of the de­vice in the course of several weeks is nor­mal. Remedy:
Correct the loss of height by pressing the button until the instrument is lifted back up to the desired working level.
2. The cryostat is in an inclined position due to an uneven upward movement of the lifting pistons. Difference of level among the four legs > 5 mm. Remedy: Have the instrument travel once up and down over the entire height adjustment dis-
until the instrument reach-
The control keys are located on top of the con­trol unit (see photograph).
To lift the instrument, press
To lower the instrument, press
As long as a button is held, the instrument will go up or down until reaching the upper or lower limit.
.
.
tance: press es the upper limit position and then
the instrument reaches the lowest possible position. Then lift the instrument back to reg­ular working height.Should the problem re­occur several times, call Leica Technical Service.
3. The height adjustment device does not reach the maximum upper level or the instru­ment lowers itself by a relatively large dis­tance (> 5 cm) within a short time. Remedy: Call Leica Technical Service.
1.8 Technical Data
9. Appendix
until
1.6 Maintenance and care
Protect the lifting pistons in the four legs of the height adjustment device against dust and water.
Approximately every six months lubricate the legs with a thin coat of resin-free oil.
Once every year have the height adjustment device checked by an authorized Leica ser­vice engineer.
Nominal voltage: 230 - 240 V 100 - 115 V Nomi nal frequency: 50 Hz 60 Hz Type of actuation: Elektrohydraulik Total height of stroke: 230 mm Carrying power: 200 kp Dimensions: Width: 880 mm (
1030 mm (with control unit) Depth: 555 mm Weigth: approx. 30 kg
without control unit)
61
Page 62
9. Appendix
2. Ordering information - accessories
2.1 Knife holders
Knife-holder CN for reusable microtome knives´w/anti-roll guide,
adjustable knife guard and knife support bars for slim and wide knives ............................................. 0419 18649
Knife support bar for knife holder CN, for slim knives .............................................................................. 0419 19426
Knife support bar for knife holder CN, for wide knives ............................................................................ 0419 19427
Knife-holder CE w/ glass anti-roll guide, for low profile disposable blades,
with lateral displacement feature ................................................................................................................ 0419 30383
Knife-holder CE w/ glass anti-roll guide, for high profile disposable blades
with lateral displacement feature ................................................................................................................ 0419 30390
Knife-holder CS for reusable knives, with one knife support
(mounted optionally on the left or right), anti-roll guide and knife guard ............................................. 0419 19787
Auxiliary clamping device for knife holder CS
(to be used instead of knife support, for increased stability) ................................................................. 0419 20971
Knife support bar for knife holder CS .......................................................................................................... 0419 25681
Knife holder base for knife holders CN, CE and CS................................................................................... 0419 26140
Knife rest with section collecting tray for knife holder CS ...................................................................... 0419 18996
2.2 Specimen discs
Specimen disc, 20 mm ø ................................................................................................................................ 0370 08636
Specimen disc, 25 mm ø ................................................................................................................................ 0416 19275
Specimen disc, 30 mm ø ................................................................................................................................ 0370 08587
Specimen disc, 40 mm ø ................................................................................................................................ 0370 08637
Specimen disc, 55 mm ø ................................................................................................................................ 0419 26491
Specimen disc, 50 x 90 mm ............................................................................................................................ 0419 26750
other sizes on request
2.2 Accessories
90° prism ........................................................................................................................................................... 0443 25949
Mobile heat extractor ..................................................................................................................................... 0443 26836
Thermoblock .................................................................................................................................................... 0398 18542
Glass anti-roll guide, complete assy., for knife holder CE ....................................................................... 0419 30402
Glass anti-roll guide upgrade kit,
for retrofitting knife holders CN, CS (for C-profile knives) ...................................................................... 0419 30403
2.4 Consumables
OCT embedding compound for freezing tissue specimens onto specimen discs,
bottle of 125 ml ................................................................................................................................................. 0201 08926
Cryostat oil No. 407, bottle of 100 ml ............................................................................................................ 0336 06099
62
Instruction Manual V1.2 - 11/2006
Page 63
10. EC Declaration of Conformity
EC Declaration of Conformity
We herewith declare, in exclusive responsibility, that the
Leica CM3050 S – Cryostat
was developed, designed and manufactured to conform with the
Council Directive 73/23/EEC (Low Voltage),
Council Directive 89/336/EEC, Appendix I (Electromagnetic Compatibility) and
European council Directive 98/79/EG (IVDD).
The following harmonized standards were applied:
• EN 61010-1: 2001
Safety requirements for electrical equipment for measurement, control and laboratory use ­Part 1: General requirements
• EN 61010-2-101: 2002
Safety requirements for electrical equipment for measurement, control and laboratory use ­Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
• EN 61326-1: 1997 + A1: 1998 + A2: 2001 + A3: 2003
Electrical equipment for measurement, control and laboratory use - EMC requirements ­Part 1: General requirements
• EN 61000-3-2: 2000
Electromagnetic compatibility (EMC) Part 3-2: Limits - Limits for harmonic current emissions
• EN 61000-3-3: 1995 + A1: 2001
Part 3: Limits ­Section 3: Limitation of voltage fluctuations and flicker in low-voltage supply systems for equipment with rated current £ 16 A Electromagnetic compatibility (EMC)
• EN 418: 1992 • EN 1037: 1995
Emergency stop equipment. Safety of machinery. Principles for design. Prevention of unexpected start-up.
In addition, the following in-house standards were applied:
• DIN EN ISO 9001: 2000.
Leica Microsystems Nussloch GmbH Postfach 1120 D-69222 Nussloch
January 12, 2004
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dag Graupner
Managing Director
Page 64
11. Decontamination Certificate (master)
Dear Customer, any product which is be returned to Leica Microsystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. This certificate, duly completed, must be placed in the unit prior to packaging or handed over to the service engineer. Packages will not be opend nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. NB: Microtome knives must be in boxes.
Description
Name/Model Fabr. No.
KAT No. Quantity
Tick Box A if applicable. Otherwise please complete all parts of B, providing further information as requested or appropriate.
Yes
A B
Yes No Please provide further detail here:
Yes No Please provide further detail here:
This equipment has not been in contact with unfixed biological samples.
1
This equipment has been exposed internally or externally to hazardous materials as indicated below:
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
2
This equipment has been cleaned and decontaminated:
If yes, give details of the method:
If no*, please indicate why not:
* Such equipment must not be returned without the written agreement of Leica Microsystems.
3
Yes No
64
The equipment has been prepared to ensure safe handling/transportation. Whenever possible, please use the original transportation case/box.
Instruction Manual V1.2 - 11/2006
Page 65
11. Decontamination Certificate (master)
Important - to avoid refusal of shipment:
Place one copy in the unit prior to packaging or hand it over to the service engineer. Customer as­sumes all responsibility for the immediate return shipment of articles sent to Leica without proper decontamination documentation.
If you have any further question, please call your local Leica organisation.
Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number: Job Sheet No.: ____________________
SU Return Goods Authorisation: ____________________ BU Return Authorisation Number: ____________________
Signature/Date
Name
Position
eMail
Leica Microsystems Nussloch GmbH Heidelberger Str. 17-19 69226 Nussloch, Germany
Phone:++49 (0) 6224 143 0 Fax: ++49 (0) 6224 143 200 www.leica-microsystems.com
Institute
Department
Adress
Phone Facsimile
65
Page 66
66
Instruction Manual V1.2 - 11/2006
Loading...