UFC 030
UFC 030 ULTRASONIC FLOW CONVERTER
The converter has a backlit local dispaly with three push buttons. All operation can be done by
using the push buttons or with the aid of a hand-held bar magnet, without opening the converter
housing. The compact converter (UFC 030 K) is mounted directly on the flow sensor, while the
separate version (UFC 030 F) is supplied with a bracket for wall or pipe mounting.
The converter is fully digitised. Measured values are obtained by use of DSP (Digital Signal
Processing) ensuring stable readings, even under severe conditio ns. The converter has standard
HART® communication and is optionally available with Profibus.
Ultrasonic Flow Converter UFC 030
# Standard * Optional
General:
The converter has a backlit local display with three push buttons. All operation can be done by
push button or using a hand-held bar magnet, without opening the converter housing.
Applied Materials
Converter housing: # Die-cast aluminum
* Stainless steel 316 L(1.4404)
Finish: # Polyurethane coated
* Offshore paint system, silver
Protection category
to IEC 529 # IP 67 eq. NEMA 6
Overall functionality/
Measurements available
and actual volume total
# HART® communication
# Flow direction (forward or reverse)
# Velocity of Sound (VOS)
# Signal strength
# Self diagnostics
# Simple single stage batching
* Corrected or standardized volume flow rate to API
2540 or customer specified
* Customer specified mass flow (requires customer
data)
*Profibus PA
* Thermal energy measurement
Local display
Operation: With cover removed, all display operations incl.
changing settings and parameters can be done using
the push buttons. With cover in place the measured
# Continuous measurement of actual volume flow rate

values and (error) messages can be viewed. Error
resetting is still possible; in this case only with the aid
of a hand-held bar magnet.
3-field LCD: The converter has a backlit local display with 3 push
buttons.
1
display and symbols for key acknowledgement.
2
3
st
line 8 character 7 segment alphanumeric
nd
line 10 character, 14 segment text display.
rd
line 5 markers to identify display in measuring
mode.
Parameters: # Actual volume flow rate in m³, barrels, liters, US
gallons or user defined volume unit (positive,
negative, and sum totals), minimum 1 year overflow
time.
# Actual volume total in m³, barrels, lites, US gallons
or user defined volume unit (positive, negative, and
sum totals), minimum 1 year overflow time.
* Velocity of sound in m/s or ft/s
* Errors (flashing display and error code)
* Signal strength (in dB)
# Corrected standard volume flow rate in m³, barrels,
liters, US Gallons or user defined volume unit per
hour, minute, second or user-defined time unit.
* Calculated mass flow rate in user defined mass unit.
* Corrected standard volume total in m³, barrels, liters,
US Gallons or user defined corrected volume unit,
minimum 1 year overflow time.
* Calculated mass total in suer defined unit, minimum
1 year overflow time.
* Analog input in ˚C, ˚F bar or psig.
* Thermal power
* Thermal energy totalized