Page 5
TIDALFLUX 4300 F
1.1 Intended use
The TIDALFLUX 4300 F has been designed for measuring the flow of conductive fluids, even in
partially filled pipes. It can be combined with the IFC 300 electromagnetic flow converter.
1.2 Safety instructions from the manufacturer
1.2.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to German copyright. Contributions from
third parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
SAFETY INSTRUCTIONS 1
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.2.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect, incidental, punitive and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
5
Page 6
1 SAFETY INSTRUCTIONS
1.2.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.2.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
TIDALFLUX 4300 F
6
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 7
TIDALFLUX 4300 F
1.2.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
SAFETY INSTRUCTIONS 1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.3 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
This flowmeter can measure the flow of conductive liquids, even in partially filled pipes. To be
able to do this, a capacitive height measurement has been integrated into a normal
electromagnetic flowmeter. If both the filled fraction and the velocity of the fluid are known, it is
easy to calculate the amount of fluid running through the pipe.
2.3 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
Figure 2-2: Example of nameplate
1 Logo and address of manufacturer
2 Type designation
3 GK/GKL values (measuring sensor constants); size (mm/inches); field frequency
4 Materials of wetted parts; protection category
8
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 9
TIDALFLUX 4300 F
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in its original packing.
INSTALLATION 3
3.3 Transport
Figure 3-1: Transport
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
9
Page 10
3 INSTALLATION
3.4 Installation conditions
3.4.1 Inlet and outlet
Figure 3-2: Recommended inlet and outlet sections, top view
1 ≥ 5 DN
2 ≥ 3 DN
3.4.2 Mounting position
TIDALFLUX 4300 F
CAUTION!
Only install the flow sensor in the shown position to keep the electrodes under water. Limit the
±2°
rotation to
Figure 3-3: Mounting position
to maintain the accuracy.
10
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 11
TIDALFLUX 4300 F
3.4.3 Flange deviation
CAUTION!
Max. permissible deviation of pipe flange faces:
- L
L
max
Figure 3-4: Flange deviation
1 L
max
2 L
min
≤ 0.5 mm / 0.02"
min
INSTALLATION 3
3.4.4 Vibration
Figure 3-5: Avoid vibrations
3.4.5 Magnetic field
Figure 3-6: Avoid magnetic fields
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
11
Page 12
3 INSTALLATION
3.4.6 Control valve
Figure 3-7: Installation before control valve
3.4.7 Slope
CAUTION!
The accuracy is influenced by the slope. Stay within
measurements!
TIDALFLUX 4300 F
±
1% to get the most accurate
Figure 3-8: Recommended slope
3.4.8 Mounting advice for difficult situations
If you can not meet the installation conditions install the flowmeter between two containers. The
inlet to the flowmeter must be higher than the outlet of the fluid. In this way you will have a calm
flow into the flowmeter, resulting in a highly accurate measurement. The sizes of the containers
must be proportional to the size of the flowmeter.
Figure 3-9: Installing in difficult situations
1 Use a container 2 if the Inlet pipe has a slope > 1%. Make sure that the outlet level of this pipe is below the inlet to the
flowmeter.
2 Inlet container
3 Inlet section of 10 DN
4 Outlet section of 5 DN
5 Outlet container advisable if outlet pipe has a slope > 1%.
12
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 13
TIDALFLUX 4300 F
3.4.9 Cleaning of flow sensor
The TIDALFLUX flow sensor is highly resistant against dirt and the measurement will rarely be
influenced by anything. However, it is advisable to create a possiblity for cleaning just before or
after the sensor.
Figure 3-10: Option for cleaning of flow sensor
1 Opening for cleaning
3.4.10 Temperatures
INSTALLATION 3
Temperature rangeProcess [°C]Ambient [°C]Process [°F]Ambient [°F]
min.max.min.max.min.max.min.max.
All versions-560-256023140-13140
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
13
Page 14
3 INSTALLATION
3.5 Mounting
3.5.1 Mounting grounding rings
CAUTION!
In order to get a reliable height measurement it is absolutely necessary
connecting pipeline is electrically conductive and connected to ground. If not, tailor-made
grounding rings with a cylindrical part can be delivered. Please contact your local agency in case
of doubt.
Figure 3-11: Grounding with grounding rings
1 Existing pipeline
2 Grounding rings, custom made to inner diameter of pipeline
3 TIDALFLUX
4 Insert the cylindrical part of the grounding ring into the pipeline. Use an appropiate gasket between the grounding ring
and the flange.
TIDALFLUX 4300 F
absolutely necessary that the inner side of the
absolutely necessaryabsolutely necessary
INFORMATION!
Sizes of the grounding rings are diameter dependent and available on request.
3.5.2 Torques and pressures
Figure 3-12: Tightening of bolts
Tightening of bolts
1 Step 1: Apply approx. 50% of max. torque given in table.
2 Step 2: Apply approx. 80% of max. torque given in table.
3 Step 3: Apply 100% of max. torque given in table.
14
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 15
TIDALFLUX 4300 F
INFORMATION!
Tighten the bolts uniformely in diagonally opposite sequence.
INSTALLATION 3
Nominal size
DN [mm]
Nominal size
[inch]
Pressure
BoltsMax. torque [Nm]
r a t i n g
200PN 108 × M 2068
250PN 1012 × M 2065
300PN 1012 × M 2076
350PN 1016 × M 2075
400PN 1016 × M 24104
500PN 1020 × M 24107
600PN 1020 × M 27138
700PN 1020 × M 27163
800PN 1024 × M 30219
900PN 1028 × M 30205
1000PN 1028 × M 35261
Flange class
BoltsMax. torque [Nm]
[lb]
81508 × 3/4"69
1015012 × 7/8"79
1215012 × 7/8"104
1415012 × 1"93
1615016 × 1"91
1815016 × 1 1/8"143
2015020 × 1 1/8"127
2415020 × 1 1/4"180
2815028 × 1 1/4"161
3215028 × 1 1/2"259
3615032 × 1 1/2"269
4015036 × 1 1/2"269
INFORMATION!
Information for bigger sizes is available on request.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
15
Page 16
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
TIDALFLUX 4300 F
4.2 Important notes on electrical connection
DANGER!
Electrical connection is carried out in conformity with the VDE 0100 directive "Regulations for
electrical power installations with line voltages up to 1000 V" or equivalent national regulations.
CAUTION!
•
Use suitable cable entries for the various electrical cables.
•
The sensor and converter are configured together in the factory. For this reason, please
connect the devices in pairs. Ensure that the sensor constant GK (see type plates) are
identically set
16
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 17
TIDALFLUX 4300 F
4.3 Connection of cables
Figure 4-1: Electrical connection
1 Unscrew the cover to reach the connectors
2 Unscrew the cover to reach the connectors
3 Field current cable
4 Interface cable
5 Signal cable (DS or BTS)
ELECTRICAL CONNECTIONS 4
Connection diagram
Figure 4-2: Connection diagram
1 Protective Earth connection (PE)
2 Mains power neutral (N)
3 Mains power live (L)
4 Field current cable
5 Interface cable
6 Signal cable. Shown is the BTS cable. In case of DS cable, do not use connectors 20 and 30.
7 Connect housing to PE
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
17
Page 18
4 ELECTRICAL CONNECTIONS
Flow sensors with protection class IP 68 can not be opened anymore. The cables are factory
connected and labeled as follows.
Figure 4-3: Labeled cables for IP 68 versions
1 Mains power (10 = blank, 11 = blue, 12 = black)
2 Field current (7 = white, 8 = green)
3 Data interface (black wires, C = marked "1", D = marked "2", E = marked "3")
4 Electrodes (1 = blank, 2 = white, 3 = red)
TIDALFLUX 4300 F
4.4 Cable lengths
CAUTION!
The maximum allowed distance between the flow sensor and the converter is determined by the
shortest cable length.
Interface cable:
Interface cable: maximum length is 600 m.
Interface cable:Interface cable:
Type B (BTS) signal cable:
Type B (BTS) signal cable: maximum length is 600 m.
Type B (BTS) signal cable:Type B (BTS) signal cable:
Type A (DS) signal cable:
Type A (DS) signal cable: the maximum length depends on the conductivity of the fluid:
Type A (DS) signal cable:Type A (DS) signal cable:
Electrical conductivity
[µS/cm]
50120
100200
200400
≥400600
Field current cable:
Field current cable: The cross section of the cable determines the maximum length:
Field current cable:Field current cable:
Cross sectionMaximum length
2 x 0.75 mm
2 x 1.5 mm
2 x 2.5 mm
Maximum length
[m]
[m]
2
2
2
150
300
600
18
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 19
TIDALFLUX 4300 F
ELECTRICAL CONNECTIONS 4
4.5 Signal cable A (type DS 300), construction
• Signal cable A is a double-shielded cable for signal transmission between the measuring
sensor and signal converter.
• Bending radius: ≥ 50 mm / 2"
Figure 4-4: Construction of signal cable A
2
1 Stranded drain wire (1) for the inner shield (10), 1.0 mm
2 Insulated wire (2), 0.5 mm
3 Insulated wire (3), 0.5 mm
4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60)
2
Cu / AWG 20
2
Cu / AWG 20
Cu / AWG 17 (not insulated, bare)
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
19
Page 20
4 ELECTRICAL CONNECTIONS
4.6 Prepare signal cable A, connect to measuring sensor
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Required materials
• PVC insulation tubing, Ø2.0...2.5 mm / 0.08...0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the twisted stranded drain wires (1) and (6)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)
TIDALFLUX 4300 F
Figure 4-5: Prepare signal cable A, connect to measuring sensor
a = 50 mm / 2"
1 Strip the conductor to dimension a.
2 Cut the outer shields (60) and (10). Make sure not to damage the stranded drain wires (1) and
(6).
3 Twist the stranded drain wires (6) of the outer shield and the drain wire (1) of the inner shield
(10).
4 Slide an insulating tube over the stranded drain wires (1) and (6).
5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wires (1) and (6).
6 Pull the heat-shrinkable tubing over the prepared signal cable.
20
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 21
TIDALFLUX 4300 F
ELECTRICAL CONNECTIONS 4
4.7 Signal cable B (type BTS 300), construction
• Signal cable B is a triple-shielded cable for signal transmission between the measuring
sensor and signal converter.
• Bending radius: ≥ 50 mm / 2"
Figure 4-6: Construction of signal cable B
2
1 Stranded drain wire for the inner shield (10), 1.0 mm
2 Insulated wire (2), 0.5 mm
3 Insulated wire (3), 0.5 mm
4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60), 0.5 mm
2
Cu / AWG 20 with stranded drain wire (20) of shield
2
Cu / AWG 20 with stranded drain wire (30) of shield
Cu / AWG 17 (not insulated, bare)
2
Cu / AWG 20 (not insulated, bare)
4.8 Preparing signal cable B, connection to measuring sensor
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Required materials
• PVC insulation tubing, Ø2.0...2.5 mm / 0.08...0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the twisted stranded drain wires (1) and (6)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
21
Page 22
4 ELECTRICAL CONNECTIONS
Figure 4-7: Preparing signal cable B, connection to measuring sensor
a = 50 mm / 2"
TIDALFLUX 4300 F
1 Strip the conductor to dimension a.
2 Cut the outer shields (60), (10), the shields around the insulated conductors (2, 3) and the
stranded drain wires (20, 30) . Make sure not to damage the stranded drain wires (1) and (6).
3 Twist the stranded drain wires (6) of the outer shield and the drain wire (1) of the inner shield
(10).
4 Slide an insulating tube over the stranded drain wires (1) and (6).
5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wires (1) en (6).
6 Pull the heat-shrinkable tubing over the prepared signal cable.
22
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 23
TIDALFLUX 4300 F
ELECTRICAL CONNECTIONS 4
4.9 Preparing field current cable C, connection to measuring sensor
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
• The field current cable is not part of the scope of supply.
• The shield is connected in the terminal compartment of the converter directly via the shield
and a clip.
• The shield is connected in the sensor via the special cable gland.
• Bending radius: ≥ 50 mm / 2"
Required materials
• Shielded 2-wire insulated copper cable
• Insulating tubing, size according to the cable being used
• Heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used
Figure 4-8: Preparation of field current cable C
a = 50 mm / 2"
b = 10 mm / 0.4"
1 Strip the conductor to dimension a.
2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
4 Pull a shrinkable tube over the prepared cable.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
23
Page 24
4 ELECTRICAL CONNECTIONS
At flow converter side:
Connecting shielding under clamp in connection box of converter
Figure 4-9: Clamping of shields
1 Field current cable
2 Signal cable
At flow sensor side:
Connecting shielding via special cable gland
TIDALFLUX 4300 F
24
Figure 4-10: Connecting the shield within the cable gland
1 Wires
2 Isolation
3 Shielding
4 Isolation
5 Feed cable through dome nut and clamping insert and fold shielding over clamping insert. Make sure that the braided
shield overlaps the O-ring by 2 mm / 3/32".
6 Push clamping insert into body.
7 Tighten the dome nut.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 25
TIDALFLUX 4300 F
4.10 Interface cable
The data interface cable is a shielded, 3 x 1.5 mm2 Liycy cable. Standard is 10 m included in the
delivery. Connect the shielding at both sides of the cable via the special cable gland.
Connecting shielding via special cable gland
ELECTRICAL CONNECTIONS 4
Figure 4-11: Connecting the shield within the cable gland
1 Wires
2 Isolation
3 Shielding
4 Isolation
5 Feed cable through dome nut and clamping insert and fold shielding over clamping insert. Make sure that the braided
shield overlaps the O-ring by 2 mm / 3/32".
6 Push clamping insert into body.
7 Tighten the dome nut.
4.11 Grounding
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
CAUTION!
In order to get a reliable height measurement it is absolutely necessary
connecting pipeline is electrically conductive and connected to ground. If not, tailor-made
grounding rings with a cylindrical part can be delivered. Please contact your local agency in case
of doubt.
absolutely necessary that the inner side of the
absolutely necessaryabsolutely necessary
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
25
Page 26
5 START-UP
5.1 Switching on the power
Before connecting to power, please check that the system has been correctly installed.
This includes:
• The device must be mechanically mounted safely in compliance with the regulations.
• The power connections must be in compliance with the regulations.
• Make sure that all electrical connections are made and that the covers of the terminal
compartments are closed.
• Check that the electrical operating data of the power supply are correct.
• Switch on the power.
INFORMATION!
The sensor can not be programmed or changed in any way. All settable functions are included in
the converter. Please see the relevant documentation of the converter for more information.
TIDALFLUX 4300 F
26
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 27
TIDALFLUX 4300 F
6.1 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are under normal operating conditions subjects
to wear and tear.
6.2 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, technical support and training.
INFORMATION!
For more precise information, please contact your local representative.
6.3 Returning the device to the manufacturer
SERVICE 6
6.3.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
27
Page 28
6 SERVICE
6.3.2 Form (for copying) to accompany a returned device
Company:Address:
Department:Name:
Tel. no.:Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
TIDALFLUX 4300 F
This medium is:water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such
substances.
We have flushed out and neutralized all cavities in the
device.
We hereby confirm that there is no risk to persons or the environment through any residual media
contained in the device when it is returned.
Date:Signature:
Stamp:
6.4 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
28
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 29
TIDALFLUX 4300 F
7.1 Measuring principle
The TIDALFLUX 4000 is an electromagnetic flow sensor with an integrated capacitive level
measurement system, designed for electrically conductive process liquids. The flow rate Q(t)
through the tube is:
Q(t) = v(t) x A(t)
Q(t) = v(t) x A(t), in which
Q(t) = v(t) x A(t)Q(t) = v(t) x A(t)
v(t) = Flow velocity of liquid product
A(t) = Wetted area of tube section.
The flow velocity is determined on basis of the known electromagnetic measurement principle.
The two measuring electrodes are located in the lower part of the measuring tube, on a level of
approx. 10% of the inner diameter of the pipe in order to get a reliable measurement to a level of
10%.
An electrically conductive fluid flows inside an electrically insulating pipe through a magnetic
field. This magnetic field is generated by a current, flowing through a pair of field coils. Inside of
the fluid, a voltage U is generated:
U = v * k * B * D
U = v * k * B * D
U = v * k * B * DU = v * k * B * D
in which:
v = mean flow velocity
k = factor correcting for geometry
B = magnetic field strength
D = inner diameter of flow meter
TECHNICAL DATA 7
The signal voltage U is picked off by electrodes and is proportional to the mean flow velocity v
and thus the flow rate q. The signal voltage is quite small (typically 1 mV at v = 3 m/s / 10 ft/s and
field coil power of 1 W). Finally, a signal converter is used to amplify the signal voltage, filter it
(separate from noise) and convert it into signals for totalising, recording and output processing.
Figure 7-1: Measuring principle TIDALFLUX
1 Electrodes
2 Induced voltage (proportional to flow velocity)
3 Capacitive plates in liner for height measurement
4 Magnetic field
5 Field coils
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
29
Page 30
7 TECHNICAL DATA
The wetted area A is computed from the known inside diameter of the pipe by the patented
capacitive level measurement system that is built into the measuring tube liner. The required
electronics unit is accomodated in a compact housing that is mounted on top of the measuring
sensor. This electronics is connected to the remote IFC 300 F converter by means of a digital
communication line.
TIDALFLUX 4300 F
30
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 31
TIDALFLUX 4300 F
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principleFaraday's law
Application rangeElectrically conductive fluids
Measured value
Measured value
Measured valueMeasured value
Primary measured valueFlow velocity
Secondary measured valueVolume flow
Design
FeaturesFlange version with full bore flow tube
Modular constructionThe measurement system consists of a flow sensor and a signal
Remote versionIn field (F) version with IFC 300 converter: TIDALFLUX 4300 F.
Nominal diameterDN200...1600 / 8...64"
Measurement range-12...+12 m/s / -40...+40 ft/s
TECHNICAL DATA 7
Level
Standard as well as higher pressure ratings
Broad range of nominal sizes
converter. It is available as remote version. More information about
the signal converter can be found in the documentation of the signal
converter.
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
31
Page 32
7 TECHNICAL DATA
Measuring accuracy
Reference conditionsSlope: 0%
Maximum measuring errorFor detailed information on the measuring accuracy, see chapter
Field current cableShielded cable must be used, no part of delivery.
Signal cableDS 300 (type A)
Data interface cableFor transmission of measured level to IFC 300 F.
Cable entries2x M20 x 1.5
TIDALFLUX 4300 F
directive "Regulations for electrical power installations with line
voltages up to 1000 V" or equivalent national specifications.
settable by switch
Option: 24 VAC, 50/60 Hz
DS 300 (type A)
DS 300 (type A)DS 300 (type A)
Max. length: 600 m / 1950 ft (dependent on electrical conductivity).
BTS 300 (type B)
BTS 300 (type B)
BTS 300 (type B)BTS 300 (type B)
Max. length: 600 m / 1950 ft
Shielded Liycy cable, 3 x 0.75 mm
1x M20 x 1.5 EMC type
1x PG9 EMC type
2
Approvals and certificates
CE
CE
CECE
This device fulfills the statutory requirements of the EC directives.
The manufacturer certifies successful testing of the product by
applying the CE mark.
Other approvals and standardsOther approvals and standards
Protection category acc. to
IEC 529 / EN 60529
Vibration resistanceIEC 68-2-6
Random vibration testIEC 68-2-34
Shock testIEC 68-2-27
Standard: IP 66/67 (NEMA 4/4X/6)
Option: IP 68 (NEMA 6P)
34
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 35
TIDALFLUX 4300 F
7.3 Sizing
INFORMATION!
These tables state the flowrate in a partially filled pipe, dependant on slope, filling level, inner
diameter and friction factors of the pipe. The values are calculated according the ManningStrickler equation. The results do not consider: deposits, profile distortion, backwater, friction of
air and the slightly smaller inner diameter of the TIDALFLUX.
Flow and velocity at 100% level, gravity fed, metric values
TECHNICAL DATA 7
DNSteel / PVCSmooth concreteConcrete
Nominal
size
slope 0.5%
2000.961090.86970.5866
2501.111961.001770.67118
3001.263211.132880.75191
3501.394811.254330.84291
4001.526881.376200.91412
5001.7712511.5911241.06749
6002.0020361.818321.21221
7002.2130621.9927571.331843
8002.4243792.1839451.452624
9002.6260002.3553821.573596
10002.8179452.5371531.684750
12003.17129062.85116031.97736
14003.51194513.16175112.1111693
16003.84277943.45249712.316647
slope 1.0%
2001.361541.221380.8192
2501.572771.422510.94166
3001.784531.64071.07272
3501.976821.776131.18409
4002.159731.948781.29584
5002.517672.2515901.51060
6002.8228702.5425851.691720
7003.1343362.8239071.882605
8003.4261893.0855732.053709
9003.784743.3376262.225084
10003.97112253.57100942.386729
12004.48182404.03164082.6910952
14004.97275424.47247712.9816514
16005.43393024.89353943.2623596
v [m/s]Q [m3/h]v [m/s]Q [m3/h]v [m/s]Q [m3/h]
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
35
Page 36
7 TECHNICAL DATA
TIDALFLUX 4300 F
Figure 7-2: Relative flow and velocity versus relative filling level
1 Flow Q
2 Velocity v
INFORMATION!
v
max
and Q
are the values of v and Q if the pipe is completely filled
max
EXAMPLE:
Consider a maximum flow to be measured of 1200 m3/h. The material of the pipeline is steel and
the slope 1.0%.
From the table a maximum flow has to be selected that is bigger than the flow to be measured.
Selection: DN500, Q
= 1767 m3/h and v
max
To determine the level in the pipe at 1200 m
See above figure and read out the ratio's of H/DN and v/v
= 2.5 m/s.
max
3
/h, calculated the ratio Q / Q
:
max
= 1200 / 1767 = 0.68.
max
1 H/DN = 0.6 or 0.6 x 500 mm = 300 mm,
2 At H/DN = 0.6, find v/v
= 1.05, so v = 1.05 x 2.5 = 2.63 m/s.
max
INFORMATION!
A sizing tool for various slopes is available at the website of the manufacturer.
36
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 37
TIDALFLUX 4300 F
7.4 Dimensions and weights
The inner pipe diameter should match the inner diameter of the flowmeter. Since the inner
diameter is not a standard DN size, choose the inner pipe diameter to be just a little bit bigger
than the flow meter diameter. If a lot of sediment or fat is expected the optimal solution is to
produce a diameter compensation ring on both sides to have smooth transits.
TECHNICAL DATA 7
IP 67 versions:
IP 67 versions:
IP 67 versions:IP 67 versions:
k = 232 mm / 9.1"
m = 110 mm / 4.3"
n = 202 mm / 7.95"
IP 68 versions:
IP 68 versions:
IP 68 versions:IP 68 versions:
k = 344 mm / 13.54"
m = 155 mm / 6.1"
n = 284 mm / 11.18"
EN 1092-1
Nominal sizeDimensions [mm]Approx.
DNPNabØcdjØDØD
IP 67IP 68
2001035047353229114617734018940
2501040052157933116620539523154
3001050057162938119123544528166
3501050062368242821430650531695
40010600681739483242386565365115
50010600784843585293386670467145
60010600894952694347386780567180
7001070010101069812406455895666265
80010800112511849224615351015768350
900109001246130510645326251115863425
10001010001338139611325666951230965520
12006120015291588134067085414051169659
140061400173217911521761103416301367835
1600616001932199117218611234183015491659
Weight
i
[kg]
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
DiameterVacuum load in mbar abs. at a process temperature of
[mm]40°C60°C
DN200...1600500600
DiameterVacuum load in psia at a process temperature of
[inches]104°F140°F
8...64"7.38.7
38
www.krohne.com06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en
Page 39
TIDALFLUX 4300 F
7.6 Measuring accuracy
The measuring accuracy for partly filled pipes and completely filled pipes are different. In these
graphs it is assumed that the velocity at full scale value is at least 1 m/s (is also the standard
value for calibration, since it will result in the most accurate measurements).
Fully filled pipes
TECHNICAL DATA 7
Figure 7-3: Maximum measuring error of measured value.
Partly filled pipes
Figure 7-4: Maximum measuring error of measured value.
1 Advised working area