KROHNE H250-M8 User Manual

Handbook
Handbook
H250
H250
H250H250
Variable area flowmeter
HandbookHandbook
© KROHNE 08/2012 - 4000269302 MA H250-R02 en
: IMPRINT :::::::::::::::::::::::::::::::::::::::
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Subject to change without notice.
Copyright 2011 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
CONTENTS
1 Safety instructions 5
1.1 Intended use ..................................................................................................................... 5
1.2 Certifications .................................................................................................................... 5
1.3 Safety instructions from the manufacturer ..................................................................... 6
1.3.1 Copyright and data protection ................................................................................................ 6
1.3.2 Disclaimer ............................................................................................................................... 6
1.3.3 Product liability and warranty ................................................................................................ 7
1.3.4 Information concerning the documentation........................................................................... 7
1.3.5 Warnings and symbols used................................................................................................... 8
1.4 Safety instructions for the operator................................................................................. 8
2 Device description 9
2.1 Scope of delivery............................................................................................................... 9
2.2 Device version................................................................................................................. 10
2.2.1 Float damping ....................................................................................................................... 12
2.2.2 Pointer damping.................................................................................................................... 12
2.3 Nameplate ...................................................................................................................... 13
2.4 Description code............................................................................................................. 14
3 Installation 15
3.1 Notes on installation ......................................................................................................15
3.2 Storage ........................................................................................................................... 15
3.3 Installation conditions ....................................................................................................16
3.3.1 Tightening torques................................................................................................................ 17
3.3.2 Magnetic filters ..................................................................................................................... 18
3.3.3 Heat insulation ...................................................................................................................... 19
4 Electrical connections 20
4.1 Safety instructions.......................................................................................................... 20
4.2 Electrical connection indicator M8................................................................................. 21
4.2.1 Indicator M8M - limit switches ............................................................................................. 21
4.2.2 Indicator M8E - current output............................................................................................. 21
4.3 Electrical connection indicator M9................................................................................. 24
4.3.1 Indicator M9 - limit switches ................................................................................................ 24
4.3.2 Indicator M9 - current output ESK2A ................................................................................... 27
4.3.3 Indicator M9 - Profibus PA (ESK3-PA) ................................................................................. 30
4.3.4 Indicator M9 - totalizer (ESK-Z)............................................................................................ 31
4.4 Electrical connection indicator M10............................................................................... 34
4.4.1 Indicator M10 ........................................................................................................................ 34
4.4.2 Power supply - current output ............................................................................................. 34
4.4.3 Switching outputs B1 and B2................................................................................................ 37
4.4.4 Switching output B2 as a pulse output ................................................................................. 39
4.4.5 Connection reset input R ...................................................................................................... 40
4.5 Grounding connections................................................................................................... 40
4.6 Protection category ........................................................................................................41
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
3
CONTENTS
H250
5 Start-up 42
5.1 Standard device .............................................................................................................. 42
5.2 Indicator M10 .................................................................................................................. 42
6 Operation 43
6.1 Operating elements........................................................................................................ 43
6.2 Basic principles of operation.......................................................................................... 44
6.2.1 Functional description of the keys........................................................................................ 44
6.2.2 Navigation within the menu structure.................................................................................. 44
6.2.3 Changing the settings in the menu....................................................................................... 45
6.2.4 Measures in the event of faulty indications.......................................................................... 45
6.3 Overview of the most important functions and indicators............................................. 46
6.4 Error messages.............................................................................................................. 47
6.5 Menu indicator M10 ........................................................................................................ 49
6.5.1 Factory settings .................................................................................................................... 49
6.5.2 Menu structure ..................................................................................................................... 50
6.5.3 Menu explanations................................................................................................................ 51
7 Service 55
7.1 Maintenance ................................................................................................................... 55
7.2 Replacement and retrofitting......................................................................................... 55
7.2.1 Replacing floats .................................................................................................................... 55
7.2.2 Retrofitting float damping..................................................................................................... 56
7.2.3 Retrofitting pointer damping ................................................................................................56
7.2.4 Retrofitting limit switch ........................................................................................................ 57
7.2.5 Replacement - Retrofitting ESK2A ....................................................................................... 58
7.2.6 Totalizer ................................................................................................................................ 59
7.3 Spare parts availability...................................................................................................60
7.3.1 List of spare parts................................................................................................................. 60
7.4 Availability of services .................................................................................................... 62
7.5 Returning the device to the manufacturer..................................................................... 62
7.5.1 General information.............................................................................................................. 62
7.5.2 Form (for copying) to accompany a returned device............................................................ 63
7.6 Disposal .......................................................................................................................... 63
8 Technical data 64
8.1 Operating principle......................................................................................................... 64
8.2 Technical data................................................................................................................. 65
8.3 Dimensions and weights ................................................................................................ 76
8.4 Measuring ranges...........................................................................................................80
4
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
1.1 Intended use
The variable area flowmeters are suitable for measuring gases, vapours and liquids.
The devices are particularly suitable for the measurement of:
Water
Hydrocarbons
Corrosion protection agents and lubricants
Chemicals and additives
Solvents
Superheated steam
Food, beverages and tobacco
Air
Industrial gases
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
SAFETY INSTRUCTIONS 1
WARNING!
The operator himself bears the sole responsibility for the intended use of the device regarding the suitability and the corrosion resistance of the used materials against the measured fluid. The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose. Do not use any abrasive media containing solid particles or highly viscous media.
1.2 Certifications
CE marking
The device fulfils all applicable statutory requirements of the following EC directives:
Pressure Equipment Directive 97/23/EC
For devices with electrical installations: EMC Directive 2004/108/EC
Devices for use in hazardous areas: ATEX Directive 94/9/EC
as well as
NAMUR recommendations NE 21 and NE 43
The manufacturer certifies successful testing of the product by applying the CE marking.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
5
1 SAFETY INSTRUCTIONS
1.3 Safety instructions from the manufacturer
1.3.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
H250
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.3.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
6
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
1.3.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.3.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
7
1 SAFETY INSTRUCTIONS
1.3.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
H250
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.4 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
8
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-1: Scope of delivery
1 Measuring device in ordered version 2 For indicator M10 - bar magnet 3 For indicator M10 - key 4 Documentation 5 Certificates, calibration report (supplied to order only)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
9
2 DEVICE DESCRIPTION
2.2 Device version
H250 with indicator M9
H250 with indicator M10
H250 with indicator M8
H250
Figure 2-2: Versions M9 and M10
1 H250/RR/M9
Local indication without auxiliary power supply
max. 2 limit switches, type NAMUR, NAMUR safety-oriented or transistor (3-wire)
®
2-wire current output 4…20 mA, HART
or Profibus-PA communication
6-digit flow counter (non Ex)
Limit switches and signal output - optional intrinsically safe
2 H250/RR/M10
Explosion proof enclosure
2 digital adjustable limit switches, 2-wire open collector or type NAMUR
®
2-wire current output 420 mA, HART
Pulse output up to 10 Hz (also for electro-mechanical counters)
12-digit flow counter with external resetting (batch operation)
The following designs are available as options:
H250 with indicator M9 as high-temperature version HT
H250 with indicator M9 with added impact and corrosion protection (special paint finish)
H250 with indicator M9 in Stainless Steel housing
communication
10
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
Figure 2-3: Version M8
1 H250/RR/M8EG
Electronic bar graph indicator
2-wire current output 420 mA, HART
®
communication
DEVICE DESCRIPTION 2
2 H250/RR/M8MG
Local indication without auxiliary power supply
2 limit switches, 2-wire, type NAMUR or NAMUR safety-oriented
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
11
2 DEVICE DESCRIPTION
2.2.1 Float damping
Float damping is characterised by high standstill times and self-centering. The damping sleeve is made of high performance ceramic or PEEK, depending on the medium and the application. Float damping can also be retrofitted for the user (see Service).
Use of damping
Generally when CIV and DIV floats are used for gas measurement.
For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
Nominal size acc. to Operating primary pressure
EN 1092-1 ASME B16.5 [bar] [psig]
DN 50 ½" 0.3 4.4
DN25 1" 0.3 4.4 DN50 2" 0.2 2.9 DN80 3" 0.2 2.9
DN 100 4" 0.2 2.9
H250
2.2.2 Pointer damping
The pointer system with its magnetic system basically contains pointer damping. An additional eddy current brake is advantageous for fluctuating or pulsing flows. The eddy current brake magnets surround the pointer vane 1 without touching it, damping its movement. The result is a pointer position that is considerably calmer, and no distortion of the measured value. A clamp screw holds it in place securely. The eddy current brake can be retrofitted without having to recalibrate and while in operation (see Service).
12
1 Eddy current brake 2 Pointer vane 3 Bracket 4 Pointer cylinder
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
2.3 Nameplate
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-4: Nameplates on the indicator
1 Device type 2 Manufacturer 3 Notified ATEX & PED body 4 Rating data: temperature & pressure rating 5 PED data 6 Ex data 7 Electrical connection data 8 Internet site
Additional markings on the indicator
SN - serial number
SO - sales order / item
PA - order
Vx - product configurator code
AC - article code
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
13
2 DEVICE DESCRIPTION
2.4 Description code
The description code* consists of the following elements:
1 Device type H250 - standard version H250H - horizontal flow direction H250U - flow direction from top to bottom
2 Materials / versions RR - Stainless Steel C - PTFE or PTFE/ceramics HC - Hastelloy Ti - Titanium F - aseptic version (food)
3 Heating jacket version B - with heating jacket
H250
4 Series of indicators M8 - Indicator M8 M9 - Indicator M9 standard indicator M9S - Indicator with added impact and corrosion protection M9R - Indicator in Stainless Steel housing M10 - Indicator or signal converter M10
5 Design of indicator M8 MG - Mechanical indicator EG - Electronic indicator with signal output 4...20 mA
6 High-temperature version HT - Version with HT extension
7 Electrical signal output ESK - Current output or Profibus-PA ESK-Z - Current output and totalizer
8 Limit switch K1 - One limit switch K2 - Two limit switches S1 - One SIL2 Limit switch acc. to IEC 61508 S2 - Two SIL2 Limit switches acc. to IEC 61508
9 Explosion protection Ex - Explosion-protected equipment
10 SIL SK - SIL2 compliance of limit switches acc. to IEC 61508
14
* positions which are not needed are omitted (no blank positions)
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry and dust-free location.
Avoid lasting direct exposure to the sun.
Store the device in its original packing.
The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
INSTALLATION 3
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
15
3 INSTALLATION
3.3 Installation conditions
CAUTION!
When installing the device in the piping, the following points must be observed:
The variable area flowmeter must be installed vertically (measuring principle). Flow direction from bottom to top. For installation recommendations please refer also to VDI/VDE 3513 Sheet 3.
H250Hs are installed horizontally and H250U devices are installed vertically with the flow direction from top to bottom.
A straight unimpeded inlet run of 5x DN upstream of the device and a straight outlet run of 3x DN downstream of the device are recommended.
Screws, bolts and gaskets are to be provided by the customer and must be selected in accordance with the pressure rating of the connection or the operating pressure.
The inside diameter of the flange deviates from the standard dimensions. Flange seal standard DIN 2690 can be applied without any limitation.
Align the gaskets. Tighten the nuts with the tightening torques of the appropriate pressure rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, see chapter "Tightening torques".
Control devices are to be positioned downstream of the measuring device.
Shutoff devices are preferably to be positioned upstream of the measuring device.
Before connecting, blow or flush out the pipes leading to the device.
Pipes for gas flow need to be dried before the device is installed.
Use connectors suitable for the particular device version.
Align the pipes axially with the connections on the measuring device so they are free of stresses.
If necessary, the piping has to be supported to prevent vibrations being transmitted to the measuring device.
Do not lay signal cables directly next to cables for the power supply.
H250
16
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
INSTALLATION 3
Take special note of the installation position for the H250H with horizontal flow direction:
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for horizontal installation are to be installed in the pipeline so that the display is located on the side of the measuring tube. The maximum medium and ambient temperatures indicated as well as the measuring accuracy are based on lateral installation of the display.
3.3.1 Tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal size acc. to Bolts Max. torque
EN 1092-1 ASME B16.5 EN
DN PN Inches lb 150 lb 300 lb Nm ft*lbf Nm ft*lbf
15 40 ½" 150/300 4 x M 12 4 x ½" 4 x ½" 9.8 7.1 5.2 3.8 25 40 1" 150/300 4 x M 12 4 x ½" 4 x 5/8" 21 15 10 7.2
50 40 2" 150/300 4 x M 16 4 x 5/8" 8 x 5/8" 57 41 41 30 80 16 3" 150/300 8 x M 16 4 x 5/8" 8 x ¾" 47 34 70 51
100 16 4" 150/300 8 x M 16 8 x 5/8" 8 x ¾" 67 48 50 36
ASME EN 1092-1 ASME 150 lb
1092-1
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
17
3 INSTALLATION
3.3.2 Magnetic filters
The use of magnetic filters is recommended when the medium contains particles which can be influenced magnetically. The magnetic filter is to be installed in the flow direction upstream of the flowmeter. Bar magnets are positioned helically in the filter to provide optimal efficiency at low pressure loss. All of the magnets are coated individually with PTFE to protect against corrosion. Material: 1.4571
Magnetic filters
H250
1 Type F - fitting part with flange - overall length 100 mm 2 Type FS - fitting part without flange - overall length 50 mm
18
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
3.3.3 Heat insulation
CAUTION!
The indicator housing may not be heat-insulated. The heat insulation 3 may only reach as far as the housing fastening 4.
Figure 3-1: H250 heat insulation
1 Standard indicator M9 2 Indicator with HT extension
INSTALLATION 3
This applies in the same manner to indicators M8 and M10.
CAUTION!
The heat insulation 1 may only reach to the rear of the housing 2. The area of the cable entries 3 must be freely accessible.
Figure 3-2: Insulation - cross section
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
19
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
H250
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
20
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.2 Electrical connection indicator M8
4.2.1 Indicator M8M - limit switches
The limit switches can be set over the entire measuring range using the limit pointer 1. The set limit values are displayed on the scale. The pointers are set to the desired limit values using a slip coupling along the scale.
ELECTRICAL CONNECTIONS 4
Figure 4-1: Limit switch settings M8MG
1 Maximum pointer, switching point indicator 2 Limit switch 3 Connection board 4 Connection terminal
4.2.2 Indicator M8E - current output
Figure 4-2: Electrical connection M8EG
1 Terminal connection 2 Power supply 14.8...30 VDC 3 Measuring signal 4...20 mA
4 External load, HART
®
communication
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
21
4 ELECTRICAL CONNECTIONS
Power supply M8 with electrical isolation
The circuitry for connection to other devices such as digital evaluator units or process control equipment must be designed with especial care. In some circumstances internal connections in these devices (e.g. GND with PE, ground loops) may lead to impermissible voltage potentials, which can compromise the function of the device itself or a connected device. In such cases a protected extra-low voltage (PELV) is recommended.
H250
Figure 4-3: Electrical connection M8EG with electrical isolation
1 Terminal connection 2 Converter supply isolator with electrical isolation 3 Power supply (see supply isolator information) 4 Measuring signal 4...20 mA
5 External load, HART
®
communication
Power supply
INFORMATION!
The supply voltage has to be between 14.8 VDC and 30 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
22 mA + 14.8 V
L
where U
= the minimum supply voltage and
ext.
= the total measuring loop resistance is.
R
L
22
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
HART® communication
When HART® communication is carried out with the M8E display, the analogue measured data transmission (4...20 mA) is not impaired in any way.
®
Exception for multidrop mode. In multidrop mode, a maximum of 15 devices with HART can be operated in parallel, whereby their current outputs are switched inactive (I approx. 4 mA per device).
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
communication a load of at least 230 ohm is required.
function
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit).
Configuration
The M8E electronic indicator can be configured via HART® communication. DD (Device Descriptions) for AMS 6.x and PDM 5.2 as well as a DTM (Device Type Manager) are available for configuration. They can be downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART counter can be configured. Two limit values can be set and monitored. The limit values are assigned either to flow values or to the counter overflow. The limit values are not depicted on the display.
communication. A flow
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
23
4 ELECTRICAL CONNECTIONS
4.3 Electrical connection indicator M9
4.3.1 Indicator M9 - limit switches
The M9 indicator can be equipped with a maximum of two electronic limit switches. The limit switch functions as a slot sensor which is operated inductively through the semicircular metal vane belonging to the measuring pointer. The switching points are set using the contact pointers. The position of the contact pointer is indicated on the scale.
Limit switch module
H250
1 Min. contact 2 Max. contact 3 Locking screw 4 Maximum pointer 5 Connection terminal
The connecting terminals have a pluggable design and can be removed in order to connect the cables. The built-in limit switch types are shown on the indicator.
Electrical connection of the limit switches
Contact MIN MAX
Terminal no. 1 2 3 4 5 6
Connection 2-wire NAMUR - + - +
Connection 3-wire + - + -
24
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Limit switch connection terminals
1 2-wire limit switch NAMUR 2 3-wire limit switch 3 Terminal connection min contact 4 Terminal connection max contact 5 3-wire load 6 NAMUR isolated switching amplifier 7 3-wire power supply
Limit setting
Figure 4-4: Limit switch settings
1 Contact pointer MAX 2 Contact pointer MIN 3 Locking screw
Setting is carried out directly via contact pointers 1 and 2:
Slide the scale away
Loosen the locking screw 3 slightly
Slide the scale back to the latching point
Set contact pointers 1 and 2 to the desired switching point
After setting has been carried out: Fix the contact pointers with the locking screw 3.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
25
4 ELECTRICAL CONNECTIONS
Switch contact definition
1 MIN contact 2 MAX contact 3 Pointer vane with switching vane
If the pointer vane enters the slot, an alarm is triggered. If the pointer vane lies outside the slot sensor, a wire break also causes the alarm to be triggered.
The 3-wire limit switch does not have any wire break detection.
Definition MinMin - MaxMax
H250
1 MIN 2 contact or MAX 1 contact 2 MIN 1 contact or MAX 2 contact
Current consumption in the position shown:
Contact Type Current
MIN 1 NAMUR 1 mA MIN 2 NAMUR 1 mA MAX 1 NAMUR 3 mA MAX 2 NAMUR 3 mA
26
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.3.2 Indicator M9 - current output ESK2A
The connecting terminals of the ESK2A have a pluggable design and can be removed in order to connect the cables.
1 ESK2A current transmitter 2 Power supply 12...30VDC 3 Measurement signal 4...20 mA
4 External load, HART
®
communication
ELECTRICAL CONNECTIONS 4
Power supply M9 with electrical isolation
The circuitry for connection to other devices such as digital evaluator units or process control equipment must be designed with especial care. In some circumstances internal connections in these devices (e.g. GND with PE, ground loops) may lead to impermissible voltage potentials, which can compromise the function of the device itself or a connected device. In such cases a protected extra-low voltage (PELV) is recommended.
1 Terminal connection 2 Converter supply isolator with electrical isolation 3 Power supply (see supply isolator information) 4 Measurement signal 4...20mA
5 External load, HART
®
communication
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
27
4 ELECTRICAL CONNECTIONS
Power supply
INFORMATION!
The supply voltage has to be between 12 VDC and 30 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
where U
ext.
R
= the total measuring loop resistance is.
L
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
22 mA + 12 V
L
= the minimum supply voltage and
H250
HART® communication
When HART® communication is carried out with the ESK, the analogue measured data transmission (4...20 mA) is not impaired in any way.
Exception for multidrop mode. In multidrop mode, a maximum of 15 devices with HART can be operated in parallel, whereby their current outputs are switched inactive (I approx. 4 mA per device).
®
function
28
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit).
communication a load of at least 230 ohm is required.
Configuration
The ESK can be configured via HART® communication. DD (Device Descriptions) for AMS 6.x and PDM 5.2 as well as a DTM (Device Type Manager) are available for configuration. They can be downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter overflow.
communication. A flow
Self monitoring - Diagnostics
During both start-up and operation, a wide variety of diagnostic functions are performed cyclically in the ESK2A, in order to guarantee function reliability. When an error is detected, a failure signal (high) is activated (current > 21 mA) via the analogue output. In addition, more
®
detailed information can be requested via HART for information and warnings.
(CMD#48). The failure signal is not activated
Diagnostic functions (Monitoring):
Plausibility of FRAM data
Plausibility of ROM data
Working range of internal reference voltages
Signal detection of the measuring range of the internal sensors
Temperature compensation of the internal sensors
Calibration corresponding the application
Plausibility of counting value
Plausibility of physical unit, system and selected unit
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
29
4 ELECTRICAL CONNECTIONS
4.3.3 Indicator M9 - Profibus PA (ESK3-PA)
Bus cable
Shielding and grounding
Shielding and grounding
Shielding and groundingShielding and grounding
The statements of the FISCO model only apply if the bus cable used meets the required specifications. For specifications, see the chapter "Technical data" ESK3-PA.
In order to ensure optimum electromagnetic compatibility of systems it is important that the system components, and in particular the bus cables, are shielded. These shields must have as few gaps as possible.
Connection
H250
Figure 4-5: ESK3-PA connection
1 Signal connection 2 Shielding and grounding
Polarity reversal has no effect on the function. The cable shield should be connected with minimum length to the functional ground FE.
30
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.3.4 Indicator M9 - totalizer (ESK-Z)
The totalizer only works in conjunction with the ESK2A current output. A 6-digit display shows the totalised flow value. It can be changed over to the instantaneous flow value in 0...100%.
A data backup is carried out automatically in the event of a power failure.
The counter is factory-set to the measuring range of the indicator. The total value can be read directly.
Supply 11/12 and measured signals S+ and S- are not electrically isolated. If the measured signal is not needed externally, a short-circuit jumper has to be connected to terminals S+ and S-.
Pulse outputs P+ and P- are electrically isolated. A pulse is generated for each counter advance. If the pulse output is not required, its terminals can remain unused.
ELECTRICAL CONNECTIONS 4
Figure 4-6: Counter connection
1 ESK - measurement signal 4...20 mA 2 Counter module 3 Transfer of the measurement signal or short-circuit jumper 4 Pulse output load 5 Counter power supply 6 Pulse output power supply
A functional extra-low voltage with protective electrical isolation (PELV) in accordance with VDE 0100 Part 410 is required as a power supply. All the instruments (recorder, display, etc.) connected to measuring circuits S+ and S- are connected in series. If this measuring circuit is not needed, then a short-circuit jumper 3 required.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
31
4 ELECTRICAL CONNECTIONS
Settings - display modes
Figure 4-7: Counter display modes
1 Flow rate as % display 2 Flow totalizer display 3 Conversion factor display
The RESET button deletes only the actual totalizer value.
H250
Settings by pressing a button at the moment of switch-on
Figure 4-8: Settings of the counter at the moment of switch-on
RESET button - mA calibration Button 1 - Display test Button 2 - Changing the conversion factor Button 3 - Software hardware version (information)
32
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Conversion factor
The conversion factor is always 10% of the full-scale range. If the measuring range is not know, the conversion factor is factory-set to 1000.
Figure 4-9: Changing the conversion factor
1 Units position 2 Tens position 3 Hundreds and 1000s position
Exit the setting by pressing the RESET button The largest factor that can be set is 1099. Factors with decimal values are not possible.
Counter overflow
Figure 4-10: Depiction of counter overflow
A counter overflow is signaled by all the decimal points lighting up. Reset by pressing the RESET button.
Current input calibration
Current input calibration
Current input calibrationCurrent input calibration During the switching-on process keep the RESET button pressed until three decimal points light up.
Set 4.00 mA
Keep button 1 pressed until the number 0 is displayed
Set 20.00 mA
Keep button 3 pressed until the number 100 is displayed
Exit calibration by pressing button 2
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
33
4 ELECTRICAL CONNECTIONS
4.4 Electrical connection indicator M10
4.4.1 Indicator M10
The display can be removed after the housing lid has been unscrewed. The connection terminals have a spring locking system.
Figure 4-11: Indicator M10 terminal connection
1 Power supply - analog output 2 Switching output B1 3 Switching output B2 or pulse output 4 Reset input R
H250
4.4.2 Power supply - current output
The electrical connection is reverse-polarity protected.
M10 - terminal connection I
1 Terminal connection 2 Power supply 16...32VDC 3 Measurement signal 4...20 mA
4 External load, HART
®
communication
34
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Power supply M10 with electrical isolation
The circuitry to other devices must be designed with special care. In some circumstances, internal connections in these devices (e.g. GND with PE, ground loops) may lead to impermissible voltage potentials, which can compromise the function of the device itself or a connected device. In such cases a protected extra-low voltage (PELV) is recommended.
1 Terminal connection 2 Converter supply isolator with electrical isolation 3 Power supply (see supply isolator information) 4 Measurement signal 4...20mA
5 External load, HART
®
communication
Power supply
INFORMATION!
The supply voltage has to be between 16 VDC and 32 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
22 mA + 16 V
L
where U
= the minimum supply voltage and
ext.
= the total measuring loop resistance is.
R
L
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
35
4 ELECTRICAL CONNECTIONS
HART® communication
When HART® communication is carried out with the M10, this will not in any way impair analogue measured data transmission (4...20 mA).
Exception for multidrop operation. In multidrop operation, a maximum of 15 devices with HART function can be operated in parallel, for which the current outputs are switched to inactive.
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit).
communication a load of at least 230 ohm is required.
H250
®
Configuration
The M10 electronic indicator can be configured via HART® communication. DD (Device Descriptions) for AMS 6.x and PDM 5.2 as well as a DTM (Device Type Manager) are available for configuration. They can be downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter.
communication. The flow
36
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.4.3 Switching outputs B1 and B2
The switching outputs are electrically isolated from each other and from the current output.
CAUTION!
The switching outputs can only be operating if the power supply is applied to terminals I+ and I-.
Switching outputs B1 and B2 can be electrically connected in two ways:
NAMUR switching output - Ri approx. 1 kOhm
OC - (open collector) low-resistance switching output with PNP technology
M10 - switching outputs
ELECTRICAL CONNECTIONS 4
1 NAMUR terminal connection 2 Isolation switching amplifier 3 PNP technology terminal connection 4 Power supply 5 Load
Switching values
NC contact NO contact
NAMUR OC NAMUR OC
Switching value reached 1 mA 1 mA > 3mA max. 100 mA Switching value not reached > 3mA max. 100 mA 1 mA 1 mA
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
37
4 ELECTRICAL CONNECTIONS
Switching capacity of B1 and B2 with PNP technology
Due to the PNP technology and the associated protective elements, there is a voltage drop Uv for the load to be operated.
Switching capacity of B1 and B2
H250
1 Max. switching current I [mA] 2 Minimum load impedance R
3 Power supply U
ext.
L
Power loss of B1 and B2
1 Load impedance RL 100 Ohm 2 Load impedance R 3 Power loss U 4 Power supply U
d
1000 Ohm
L
ext.
[Ohm]
38
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.4.4 Switching output B2 as a pulse output
INFORMATION!
When switching output B2 is used as a pulse output, two separate signal circuits are required. Each signal circuit requires its own power supply. The total resistance 3 must be adapted so that the total current I
Figure 4-12: Electrical pulse output
1 Terminal B2 2 Terminal I 3 Load e.g. counter 4 Flow rate measurement 4...20 mA 5 Pulse output power supply 6 M10 power supply
ELECTRICAL CONNECTIONS 4
does not exceed 100 mA.
tot
Pulse output B2 is a passive "open collector" output which is electrically isolated from the current output and output B1. It can be operated as a low-resistance output or as a NAMUR output.
Figure 4-13: Data pulse output
1 f
= 10 Hz
max
2 t
on
3 t
off
The pulse width t
can be configured from 30...500 ms in the menu of the indicator.
on
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
39
4 ELECTRICAL CONNECTIONS
4.4.5 Connection reset input R
Input R can be used as a reset input for the internal counter.
Figure 4-14: Indicator M10 - reset input
1 Function active HI 2 Function active LO
H250
This reset input can be activated in the menu of indicator M10, and can be configured to ACTIVE HI or ACTIVE LO. See also chapter "Indicator M10 menu explanations".
If the input is set as ACTIVE LO, an interruption causes the counter to be reset.
4.5 Grounding connections
Figure 4-15: Ground connections
1 Indicator M8 2 Indicator M9 3 Indicator M10
40
DANGER!
The grounding wire may not transfer any interference voltage. Do not use this grounding wire to ground any other items of electrical equipment.
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
4.6 Protection category
The measuring device meets all requirements of protection category IP
Indicator Protection category
M9 IP65/67 M8 IP65 M10 IP66/67
DANGER!
After all servicing and maintenance work on the device, the specified protection category has to be ensured again.
Therefore it is essential to observe the following points:
Use only original gaskets. They must be clean and free of any damage. Defective gaskets have to be replaced.
The electrical cables used must be undamaged and must comply with regulations.
The cables must be laid with a loop 3 upstream of the measuring device to prevent water
from getting into the housing.
The cable glands 2 must be tightened.
Close the unused cable glands using blanking plugs 1.
ELECTRICAL CONNECTIONS 4
Figure 4-16: Cable gland
1 Use blanking plugs if no cable is routed through 2 Tighten cable gland firmly 3 Lay the cable in a loop
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
41
5 START-UP
5.1 Standard device
CAUTION!
When starting up the device, the following points must be observed:
Compare the actual operating pressure and the product temperature of the system with the specifications on the nameplate (PS and TS). These specifications may not be exceeded.
Make sure materials are compatible.
Slowly open the shut-off valve.
When measuring liquids, vent the pipes carefully.
When measuring gases, increase pressure slowly.
Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring unit or float.
A minimum operating pressure (primary pressure) is necessary to operate the device:
Medium Pressure loss : operating pressure
H250
Liquids 1 : 2
Gases without float damping 1 : 5
Gases with float damping 1 : 2
5.2 Indicator M10
INFORMATION!
The device is always preset for the user and his application.
Start After the device is switched on, the display shows the following in sequence
"Test",
the device type and
the version number.
Afterwards the device performs a self-test and switches to measurement mode. Here all of the parameters preset for the customer are analysed and checked for plausibility, and the current measured value is displayed.
42
Operation
INFORMATION!
The device is low-maintenance Comply with the application limits with regard to temperature of the medium and ambient temperature.
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
6.1 Operating elements
OPERATION 6
Operation of the device is performed with the cover on the front open, using the mechanical keys or with the cover closed using a bar magnet
bar magnet.
bar magnetbar magnet
keys,
keyskeys
CAUTION!
The switching point of the magnetic sensors is right at the level of the corresponding circle. Only touch the circle vertically and from the front using the bar magnet. Touching it from the side may cause a malfunction.
Figure 6-1: Display and operating elements
1 Enter button (circuit for bar magnet) 2 Up button (circuit for bar magnet) 3 Right button (circuit for bar magnet) 4 Display
The mechanical keys and keys for the bar magnet are identical in their function. In this documentation the keys are represented as symbols to describe the operating functions:
Key Symbol
1 Enter ^ 2 up 3 right
Table 6-1: M10 operation keys
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
43
6 OPERATION
6.2 Basic principles of operation
6.2.1 Functional description of the keys
H250
^
Table 6-2: Description of the opperation buttons
Switch from measurement mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
In change mode:
In change mode: Move the input cursor one position to the right; after the last digit the
In change mode:In change mode: input cursor jumps back to the beginning.
In measuring mode:
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
Change between the menu items within a menu level
In change mode:
In change mode: Changing parameters and settings; running through the available
In change mode:In change mode: characters; shifting the decimal point to the right.
Change one level up at the menu
Return to measuring mode with a query whether the data should be accepted
6.2.2 Navigation within the menu structure
Navigation within the menu is by means of the and ^ buttons. Pressing button takes you one menu level lower, ^ takes you one menu level higher.
If you are already located at the lowest level (function level), you can use the button to go the the change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^ key to exit the menu mode and return to the measuring mode.
Measuring
operation
Table 6-3: Navigation menu structure
Main menu↑→ Sub-menu↑→ Function↑→ Edit
^^^^ ^^^^ ^^^^ ^^^^
→ ↑ ^^^^
44
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
6.2.3 Changing the settings in the menu
Starting operation
Operation is started using the key →.
If a different key is pressed, it is necessary to wait 5 seconds before activating the key →.
If an operation inhibit has been set, the code → → → ^ ^ ^ ↑ ↑ ↑ has to be entered. If no key is pressed within 5 seconds, code input is exited.
Exiting operator input
Operation is exited by pressing the ^ key several times.
If data have been changed:
OPERATION 6
Save Yes Changes are accepted. An update is carried out and the indicator jumps back to
Save No ^ Changes are discarded and the indicator jumps back to measuring operation.
measuring operation.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal plausibility check. If implausible inputs have been made, the indicator remains in the current menu, and the changes are not accepted.
Example: Changing the default parameter from m
3
/h to l/h
Display Display
Example: 7.2
m
1x Fct. 1111.0
OPERATION
2x Fct. 3333.0
INSTALLATION
1x Fct 3.1111
LANGUAGE
12x Fct 3.13
END&UNIT
3
/h
13
1313
1x Fct. 3.13.1111
1x 10,0000
6x 10000
1x ^^^^ Quit Yes
3x ^^^^ 7200
FLOW RATE
3
/h
m
l/h
L/h
6.2.4 Measures in the event of faulty indications
If the indications on the display or the responses to keypad commands are faulty, you have to do a hardware reset. Switch the power supply OFF and ON again.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
45
6 OPERATION
6.3 Overview of the most important functions and indicators
INFORMATION!
For a complete list of all functions and short descriptions refer to Menu explanations on page
. All default parameters and settings are adapted for the specific customer.
51
Level Designation Explanation
1.4 TIME CONST. Time constant, damping value [s]
1.5.2 ERROR Error indicator
Yes: Error messages are deleted
No: Error messages are suppressed.
2.1 4-20mA OUT Check current output
2.2 -2.4 OUTPUT B Check switching outputs and RESET input
3.1 LANGUAGE Select the menu language
3.13.1 FLOW RATE Maximum flow rate
Table 6-4: The most important functions
The value set is represented by a 20 mA analogue current output. If the current value exceeds the preset value, an alarm is indicated.
H250
M10 flow units
M10 flow units
M10 flow unitsM10 flow units
Measured variables
Volume
Mass g/s g/min g/h - Liquids,
Units Measured
products
m3/s m3/min m3/h m3/d
L/s L/min L/h -
ft3/s ft3/min ft3/h ft3/d
gal/s gal/min gal/h gal/d
bb/s bb/min bb/h bb/d
ImpGal/s ImpGal/min ImpGal/h ImpGal/d
kg/s kg/min kg/h kg/d
- t/min t/h t/d
lb/s lb/min lb/h -
- short t/min short t/h short t/d
- - long t/h long t/d
Liquids, vapours, gases
vapours, gases
46
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
6.4 Error messages
Error message Description Category Remedy
OPERATION 6
NOT LINEARIZED Linearization faulty or
NEW LINEARI. TABLE BAD
LINEARIZATIO UNDER CONFIG
UNIT SYSTEM CONFLICT
TOO FEW ENTRIES
NOT MONOTONOUS
FIRST NOT 0 % The first flow value if
LAST NOT 100 % The last flow value if
NO ZERO CAL OF AOThe current output
NO F.SC. CAL OF AOThe current output
NO TEMP. COMPENSATION
OUTPUT NOT LINEARIZED
COUNTER LOST Totalizer value was
not activated = measuring error
Faulty or missing data in the linearization table = measuring error
The device is in linearization mode = measuring error
The unit for the linearization flow is incompatible with the selected flow type (mass/volume)
The linearization table has too few data points
The sequence of the linearization values is not strictly monotonic increasing
the linearization table is not 0%
the linearization table is not 100%
zero point 4.00 mA is not calibrated = poss. measuring error in the process control system
100% = 20.00 mA is not calibrated = poss. measuring error in the process control system
The sensor temperature compensation of the device is faulty or was not carried out. = possible measuring error
Linearization is not activated = measuring error
reset by error/overflow
Error Activate linearization or carry it out again
(HART® communication and linearization software are required; the original calibration values must be known), or send the device back to the manufacturer for linearization.
Error Complete the linearization and activate it
(HART® communication and linearization software are required), or send the device back to the manufacturer for linearization.
Error Correct error, carry out linearization again if
necessary (HART® communication and linearization software are required), or send the device back to the manufacturer for linearization.
Error Carry out linearization at at least 5 points
(HART communication and linearization software are required), or send the device back to the manufacturer for linearization.
Error Check linearization and/or carry it out again
(HART® communication and linearization software are required), or send the device back to the manufacturer for linearization.
Warning Perform calibration using ammeter and menu
3.10 or using standard HART® tools/process control system and poss. external ammeter. Caution: during the calibration, switch the measuring point to manual control.
Warning Perform calibration using ammeter and menu
item 3.11 or using standard HART® tools and external ammeter if necessary. Caution: during calibration, switch the measuring point to manual control.
Error The device, together with an indication of the
error, must be sent back to the manufacturer for checking.
Error Activate linearization or carry it out again
(HART® communication and linearization software are required; the original calibration values must be known), or send the device back to the manufacturer for linearization.
Warning Because the reset time is not known:
Controlled reset of the counter using menu item 1.5.1 or using HART® tools/process control system.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
47
6 OPERATION
Error message Description Category Remedy
H250
FRAM WRITE FAULT
ROM/FLASH ERROR
RESTART OF DEVICE
MULTIDROP MODE
CRYSTAL OSC FAULT
REF VOLTAGE FAULT
SENSOR A FAULT Internal error in
SENSOR B FAULT
MEMORY CORRUPTION
AO FIXED The current output is
AO SATURATED Current output
Internal communication error
Memory error detected during self­test.
A device restart has taken place
The HART® multidrop mode is activated. The current output is set to a fixed value of
4.5 mA.
Internal error in device
Internal error in device
device
Internal error in device
Internal memory error, caused by a hardware or software problem
set to a fixed value.
saturated
Error Check whether the display is plugged in
correctly and restart the device. If the error occurs again: send the device back to the manufacturer with an indication of the error.
Error Restart the device. If the error occurs again:
send the device back to the manufacturer with an indication of the error.
Information The device has been restarted using menu
item 1.5.2 since the last time the error messages were reset.
Information The HART®- multidrop mode is activated by
selecting a polling address not equal to 0 using menu item 3.9. A polling address of 0 reactivates the current output.
Error The device must be sent back to the
manufacturer with an indication of the error.
Error Restart the device; if the error occurs again
the device must be sent back to the manufacturer with an indication of the error.
Information The current output is fixed and does not
reflect the measured value. This is the case in multidrop mode, with current output test/calibration using the menu or HART®
Information The current output is saturated at 20.4 or 22.0
mA (depending on whether the alarm current is activated or deactivated in menu item 3.12), and is no longer coupled with the measured value.
48
Device drivers for HART® tools, process control equipment (e.g. Siemens PDM or AMS)
TM
PACTware
and HART® DTMs are available on the internet site.
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
6.5 Menu indicator M10
6.5.1 Factory settings
Menu Function Setting
1.1.1 Switching value B1 0.0
1.1.2 Hysteresis B1 0.0
1.2.1 Switching value B2 0.0
1.2.2 Hysteresis B2 0.0
1.3 Display Flow rate
1.4 Time constant 3 s
1.5.1 Reset counter NO
1.5.2 Reset error NO
3.1 Language DEUTSCH
3.2 Function B1 INACTIVE
3.3 Contact B1 NC contact
3.4 Function B2 INACTIVE
3.5 Contact B2 NC contact
3.6 Pulse duration 100ms
3.7 Pulse / unit 001 / liter
3.8 Function B3 INACTIVE
3.9 Multidrop polling address 0
3.12 Alarm current OFF
3.13.1 Flow unit see rating plate
3.13.2 Counter unit Derived from the flow unit
3.14 LFC 4% ON 6% OFF
3.15 Input code NO
OPERATION 6
INFORMATION!
The device has been set at the factory in accordance with the customer order. Therefore subsequent configuration via the menu is only necessary if the intended use of the device is changed.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
49
6 OPERATION
6.5.2 Menu structure
Menu Submenu 1 Submenu 2
1111 Operation 1.1 Output B1 1.1.1 Switching value B1
2222 Test & Info 2.1 Output 4...20mA
3333 Installation 3.1 Language
H250
1.1.2 Hysteresis B1
1.2 Output B2 1.2.1 Switching value B2
1.2.2 Hysteresis B2
1.3 Display
1.4 Time constant
1.5 Reset 1.5.1 Reset counter
1.5.2 Reset error
2.2 Output B1
2.3 Output B2
2.4 Input B3
2.5 Serial no.
2.6 Software version
2.7 Tag no.
3.2 Function B1
3.3 Contact B1
3.4 Function B2
3.5 Contact B2
3.6 Pulse duration
3.7 Pulse/unit
3.8 Function B3
3.9 Multidrop
3.10 Calibration 4mA
3.11 Calibration 20mA
3.12 Alarm current
3.13 Upper range value and unit 3.13.1 Flow rate
3.13.2 Counter
3.14 Low Flow Cutoff LFC 3.14.1 Control
3.14.2 Switch-on value
3.14.3 Switch-off value
3.15 Input code
3.16 Basic setting
50
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
6.5.3 Menu explanations
OPERATION 6
Level Designation Selection /
Explanation
Input
1.1.1 OUTPUT B1 INACTIVE
FLOW.VAL B1 Flow value switching point.
COUNTER.VAL B1Counter value switching point.
1.1.2 OUTPUT B1 HYST.B1 Hysteresis setting for the flow value switching point.
The switching point is entered in flow units. If the current flow value exceeds this set switching point, then output B1 is activated.
Note: The function NC or NO can be selected using menu 3.3.
Each positive number can be set here. If the counter exceeds this value, then output B1 is activated.
Note: The function NC or NO can be selected using menu 3.3.
Value range 0...switching point. Example, if a switching value of 200 is set under 1.1.1, then a hysteresis value of 0...200 can be set here. If the value 0 is entered here, then this output does not have hysteresis. If the value 20 is entered here, then the output functions as follows: If the current flow value exceeds the value 200, then the output switches
3. If the current flow value drops below the hysteresis value of 180, then the switching output returns to its normal state 4.
Note: To invert the operating method, use menu 3.3 to set the output from NO 1 to NC 2 or vice versa. This function is not activated on the counter switching point.
1.2.1 OUTPUT B2 INACTIVE
FLOW.VAL B2 see FLOW.VAL B1
COUNTER.VAL B2see COUNTER.VAL B1
PUL. VAL B2 B2 = pulse output
Note: Settings under menu 3.6 pulse duration and
3.7 Pulses/unit
1.2.2 OUTPUT B2 HYST.B2 See HYST. B1
1.3 DISPLAY FLOW RATE
COUNTER
FLOW&COUNT
PERCENT
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
51
6 OPERATION
H250
Level Designation Selection /
Explanation
Input
1.4 TIME CONST. Setting : 1 ... 20 seconds
Note: The settable time constant affects the current output and the displayed current flow rate. It thus enables damped depiction if there is a highly variable flow rate.
If the current flow rate is polled via HART transferred measured value is dependent on the time constant here, too.
1.5.1 RESET COUNTER YES - NO
1.5.2 RESET ERROR YES - NO
2.1 4-20mA OUT The analogue current output can be set to fixed values in 10% steps between 4.00...20.00 mA. This function has no influence on binary switching outputs.
Note: This test function is switched off in multidrop mode. Display: "NOT AVAILABLE".
2.2 OUTPUT B1 OPEN The function assignment in menu 3.2 is not taken into consideration here.
CLOSED
2.3 OUTPUT B2 OPEN The function assignment in menu 3.3 is not taken into consideration here.
CLOSED
2.4 INPUT B3 Here there is a visual depiction of whether or not input B3 has a voltage of
5...30 V. If input B3 is set to ACTIVE HI in menu 3.8, then the display shows "ON" when the switching voltage is applied.
®
communication, then the
Note: NO test function possible when the output is set to INACTIVE in menu 3.8.
3.1 LANGUAGE ENGLISH
DEUTSCH
FRANCAIS
ITALIANO
ESPANOL
CESKY
POLSKI
NEDERLANDS
3.2 FUNCTION B1 INACTIVE Output B1 is switched off.
SWITCHING POINT
COUNTER_LIM Output B1 switches when the counter exceeds the counter limit value.
3.3 CONTACT B1 NC contact Output B1 is normally closed. If an alarm situation occurs, the contact
NO CONTACT Output B1 is normally open. If an alarm situation occurs, the contact
3.4 FUNCTION B2 INACTIVE See FUNCTION B1
SWITCHING POINT
COUNTER_LIM See FUNCTION B1
PULSE OUTPUT Output B2 generates pulses up to 10 Hz depending on the current flow
Output B1 switches at a set value depending on the current flow value.
opens.
closes.
See FUNCTION B1
value.
52
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
OPERATION 6
Level Designation Selection /
Explanation
Input
3.5 CONTACT B2 NC contact See CONTACT B1
NO CONTACT See CONTACT B1
3.6 PULSE
DURATION
3.7 PULSE/UNIT 0.000001 Smallest scaling factor
3.8 FUNCTION B3 INACTIVE
3.9 MULTIDROP 015 Multidrop mode means that the device works continuously in bus mode
3.10 4mA CALIBR. This menu item allows precise calibration of the current output.
3.11 20mA CALIBR. This menu item allows precise calibration of the current output.
3.12 ALARM
CURRENT
3.13 END & UNIT The flow unit and the upper range value can be changed.
30 ms
50 ms
100ms
200 ms
500 ms
Note: In the basic setting, the unit of the pulse output corresponds to the flow unit.
Example: volume flow unit is m
3
or
m
3
/h, so the pulse output is set to pulses /
mass flow unit is kg/h, so the pulse output is set to pulses / kg
999999.0 Largest scaling factor
ACTIVE HI The internal counter is reset to zero when a positive voltage of from
5...30 VDC is applied to terminals R+ and R for at least 100 ms.
ACTIVE LO The internal counter is reset to zero when a positive voltage of from 5...30
VDC applied to terminals R+ and R is interrupted for at least 100ms..
®
via HART The analogue current output is then set to a fixed value of 4.1 mA.
Measured values are transferred via HART
communication (max. 15 devices in parallel).
®
communication. However, the display allows local read-off of the measured values. The polling address can be set to 1...15. Larger integer values are not permitted. If the polling address is set to 0,
then HART
®
bus mode is switched off. The device is working in analogue mode. Current output 4...20 mA is active.
Standard HART
®
communication remains guaranteed.
The device generates a fixed current output of 4.00 mA. If the measured value deviates from the displayed one, then the measured value must be input. When the menu is exited, the corrected value is saved.
The device generates a fixed current output of 20.00 mA. If the measured value deviates from the displayed one, then the measured value must be input. When the menu is exited, the corrected value is saved.
OFF Measured values > 100% are indicated as a current signal up to a
maximum of 22 mA.
ON In the event of an error the current output is set to the fixed value of
22m A.
Note: Changing from volume flow measurement to mass flow measurement is only possible with a new calibration.
3.13.1 FLOW RATE For a units list, see chapter 7.4 of the handbook
3.13.2 COUNTER As standard, the unit for the counter is derived from the unit for the flow measurement. It can also be changed individually.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
53
6 OPERATION
H250
Level Designation Selection /
Explanation
Input
3.14 LFC LFC stands for Low Flow Cutoff.
3.14.1 CONTROL INACTIVE LFC is switched off
ACTIVE LFC is switched on
3.14.2 LFC
ON_VALUE
3.14.3 LFC
OFF_VALUE
119 % Switch-on value 1:
220 % Switch-off value 2:
With variable area flowmeters, the flow range from 0 to 10% is not defined. In order to ensure a stable zero point of the analogue output, the analogue output can be set to a stable value of 4.00mA in a selectable range from 0 to 20%.
The flow is greater than 20%. The current output corresponds to this. If the flow rate falls, then the current output follows it until the ON value. If the flow value continues to fall, the current output is switched to
4.00 mA 3.
The flow rate is 0. The current output is 4.00 mA 3. If the flow rate rises, the current output remains at 4.00 mA 3 until the OFF value, and is switched to the corresponding flow value if the flow value rises further.
3.15 INP. CODE YES The input code is used to prevent authorised adjustment of the
NO
3.16 BASIC
SETTING
YES This menu item can be used to select the calibrated basic setting.
NO
measurement parameters. The input code is not active by default. If YES is selected, the last code entered must be typed in. Factory code: → → → ^ ^ ^ ↑ ↑ ↑ If, after confirmation with YES, the button is also pressed, then a new, individual, nine-element code can be typed in. The display shows the required key combination.
This can be helpful if operating data have been changed a number of times. This menu item cannot be used to reset the calibration.
54
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to the measuring tube and indicator. We advise that inspections be carried out at least once per year. The device must be removed from the piping before cleaning.
CAUTION!
Pressurized pipes must be depressurized before removing the device. Empty pipes as completely as possible. In the case of devices used for measuring aggressive or hazardous media, appropriate safety precautions must be taken with regard to residual liquids in the measuring unit. Always use new gaskets when reinstalling the device in the pipeline. Avoid electrostatic charges when cleaning the surfaces (e.g. sight window)!
7.2 Replacement and retrofitting
SERVICE 7
Some of the variable area flowmeter components can be retrofitted:
Float damping
Indicator M9:
Indicator M9:
Indicator M9:Indicator M9:
Eddy current brake
Limit switch unit
Current output ESK2A
Counter module
The ESK3-PA Profibus can only be retrofitted following recalibration.
7.2.1 Replacing floats
Remove the device from the piping.
Take the upper snap ring out of the measuring unit.
Take the upper float stop and float out of the measuring unit.
Insert the new float into the centre hole of the lower float stop and push into the measuring
unit along with the upper float catcher. While doing this, the float's upper guide rod must be guided through the middle hole of the float stop.
Insert the snap ring into the measuring unit.
Fit the device back into the piping.
CAUTION!
An additional measuring error is to be expected if recalibration is not conducted.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
55
7 SERVICE
7.2.2 Retrofitting float damping
Remove the device from the piping.
Take the upper snap ring 1 out of the measuring unit.
Take the upper float catcher 2 and float 5 out of the measuring unit.
Fasten the snap ring 3 into the lower slot of the float's guide rod.
Slide ceramic sleeve 4 on to the float's guide rod and attach it to the top slot using the span
ring 3.
Insert float into the lower float guide in the measuring unit.
Retrofit the supplied damping cylinder with the integrated float catcher 2 into the measuring
unit.
Insert upper snap ring 1.
H250
1 Snap ring 2 Float catcher 3 Spring washer 4 Ceramic sleeve 5 Float
7.2.3 Retrofitting pointer damping
When retrofitting the pointer damping for indicator M9 with ESK2A current output and limit switches, note that the pointer may move briefly when installing the pointer damping (eddy current brake), which may trigger an error alarm or may change the current output by showing peaks.
The Eddy current brake consists of two parts:
56
1 Turnbuckle 2 Eddy current brake
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
SERVICE 7
The brake with the retainer ring can be clipped onto the pointer cylinder independently of the built-in components (ESK2A, limit switch, counter). When installing the brake, note that the slit between the brake magnets is only about 3 mm and the material thickness of the aluminium pointer vane is 1mm.
Clip on Eddy current brake 1.
Turn brake slightly in a clockwise direction 2.
Check that the pointer vane can be moved between the magnets without touching them.
Screw in turnbuckle 2.
7.2.4 Retrofitting limit switch
Remove counter module (if available).
Loosen the locking screw 2 on the contact pointer.
Merge contact pointer 1 in the middle.
Insert the contact module into the third slot of the bracket until the semi-circle 3 surrounds
the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in order to connect the cables.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
57
7 SERVICE
7.2.5 Replacement - Retrofitting ESK2A
When replacing or retrofitting an ESK2A, the following are required at the time of ordering:
SN - serial number or
SO - sales order
This information can be found on the indicator nameplate
INFORMATION!
The ESK2A is factory calibrated, making it possible to replace it or retrofit it without recalibrating.
De-energise the ESK2A.
Lift and remove the ESK2A with a screwdriver.
H250
Plug-in technology is used to install the ESK2A.
The ESK2A plug-in tongues 1 are plugged in under the two bolts 2 on the baseplate.
Slight pressure is used to press the ESK2A onto the spring pins 3 until it stops, firmly
attaching the ESK2A.
If a change in measuring range, product temperature, product, density, viscosity or pressure is desired, this can be done with the KroVaCal program or with a HART
measuring unit is subject to its own physical limits, which the DroVaCal program correctly calculates, and may thus reject the desired change. If a change is performed using the program, the new data is also transmitted to the ESK2A.
TM
modem. However, each
Program features and possibilities
Device identification
Device address
Serial number
Measuring point designation
Digital measured value query in flow units, % and mA
Test / setting functions
Calibration 4.00 and 20.00 mA
Set current output to any desired value
58
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
7.2.6 Totalizer
The flow counter, in conjunction with the ESK2A electrical current output, can also be retroactively built into the indicator M9.
When ordering the ESK-Z totalizer as a retrofit kit, please indicate the device information (see scale) and the measuring range.
Providing this information will ensure that the new scale supplied with the counter display cutout is ready for installation.
The flow counter is then preset using the conversion factor corresponding to the measuring range.
SERVICE 7
Installation
Push the existing scale out.
Insert the flow counter unit into the centre track of the module carrier.
Insert the new scale into the module carrier.
In so doing, lift the scale slightly until the scale cutout surrounds the counter display.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
59
7 SERVICE
7.3 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions.
7.3.1 List of spare parts
Spare part Order no.
DN 15
DN 15
DN 15DN 15
Float CIV 15, 1.4404 X251041000
Float DIV 15, 1.4404 X251042000
Float TIV 15, 1.4404 X251043000
Float DIVT 15, 1.4404 X251044000
Float TIV 15, Aluminium X251043100
Float TIV 15, Titanium X251043200
Set float stop; standard (1 stop, 1 span ring) X251050100
Set float stop; gas damping (ZrO2) X251050200
Set float stop; gas damping (PEEK) X251050300
Damping bush (7x8) ZrO2 incl. 2 span rings X251053100
Damping bush (7x8) PEEK incl. 2 span rings X251053200
DN 25
DN 25
DN 25DN 25
Float CIV 25, 1.4404 X252041000
Float DIV 25, 1.4404 X252042000
Float TIV 25, 1.4404 X252043000
Float DIVT 25, 1.4404 X252044000
Set float stop; standard (1 stop, 1 span ring) X252050100
Set float stop; gas damping (ZrO2) X252050200
Set float stop; gas damping (PEEK) X252050300
Damping bush (12x8) ZrO2 incl. 2 span rings X252053100
Damping bush (12x8) PEEK incl. 2 span rings X252053200
DN 50
DN 50
DN 50DN 50
Float CIV 55, 1.4404 X253041000
Float DIV 55, 1.4404 X253042000
Float TIV55, 1.4404 X253043000
Float DIVT 55, 1.4404 X253044000
Set float stop; standard (1 stop, 1 span ring) X253050100
Set float stop; gas damping (ZrO2) X253050200
Set float stop; gas damping (PEEK) X253050300
Damping bush (14x10) ZrO2 incl. 2 span rings X253053100
Damping bush (14x10) PEEK incl. 2 span rings X253053200
H250
60
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
SERVICE 7
DN 80
DN 80
DN 80DN 80
Float CIV 85, 1.4404 X254041000
Float DIV 85, 1.4404 X254042000
Float TIV 85, 1.4404 X254043000
Float DIVT 85, 1.4404 X254044000
Set float stop; standard (1 stop, 1 span ring) X254050100
Set float stop; gas damping (ZrO2) X254050200
Set float stop; gas damping (PEEK) X254050300
Damping bush (18x14) ZrO2 incl. 2 span rings X254053100
Damping bush (18x14) PEEK incl. 2 span rings X254053200
DN 100
DN 100
DN 100DN 100
Float CIV 105, 1.4404 X255041000
Float DIV 105, 1.4404 X255042000
Float DIVT 105, 1.4404 X255044000
Set float stop; standard (1 stop, 1 span ring) only for bottom ! X255050100
Set float stop; gas damping (ZrO2) X255050200
Set float stop; gas damping (PEEK) X255050300
Damping bush (18x14) ZrO2 incl. 2 span rings X254053100
Damping bush (18x14) PEEK incl. 2 span rings X254053200
Indicator M9
Indicator M9
Indicator M9Indicator M9
Indicator housing complete no scale X251010000
Indicator complete stainless steel unpainted, no scale X251011000
Cover M9 complete, standard (blue; RAL 5015) X251010100
Cover M9 complete, salt water resistant (grey; RAL 7001) X251010200
Cover M9 complete, no silicone (blue; RAL, 5015) X251010300
Cover M9 complete, stainless steel unpainted X251010400
Inspection glass shatterproof glass X251011100
Inspection glass plastic (Makrolon) X251011200
Cover seal (silicone) X251012100
M9 baseplate standard X251020100
M9 baseplate salt water resistant X251020200
Retrofit kit HT extension X251021000
Module carrier (profile track) X251021100
Set housing attachment parts (1 pair) X251021300
Pointer system, complete X251022100
Eddy current brake X251022200
Printed scale (serial number required) On request
Blank scale X251023200
Printed scale with counter cutout (serial number required) On request
Blank scale with counter cutout X251023400
Other spare parts on request
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
61
7 SERVICE
7.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local representative.
7.5 Returning the device to the manufacturer
7.5.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
H250
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
62
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
7.5.2 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
SERVICE 7
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
7.6 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
63
8 TECHNICAL DATA
8.1 Operating principle
The flowmeter H250 operates on the float measuring principle. The measuring unit consists of a metal cone in which a float can move freely up and down. The medium flows through the flowmeter from bottom to top. The float adjusts itself so that the buoyancy force B, acting on it, the form drag D and its weight W are in equilibrium: W = B + D.
Figure 8-1: Operating principle
1 Indication principle M9 and M8MG 2 Indication principle M10 and M8EG
H250
For indicators M9 and M8MG 1 the flow-dependent height of the float in the measuring unit is transmitted by means of a magnetic coupling and displayed on a scale. For indicators M10 and M8EG 2 the flow-dependent height of the float in the measuring unit is transmitted to the electronic display by magnetic field sensors S1 and S2.
Operating principle of H250H and H250U
Figure 8-2: Operating principle H250H and H250U
1 H250H - horizontal flow direction 2 H250U - flow direction from top to bottom
The flowmeters operate according to a modified float measuring principle. The guided float adjusts itself so that the flow force acting on it is in equilibrium with the opposing spring force. The flow-dependent position of the float in the measuring unit is displayed on a scale by means of a magnetic coupling.
64
INFORMATION!
Flowmeters H250H and H250U only work in conjunction with indicator M9.
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
8.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Application range Flow measurement of liquids, gases and vapors
Operating method / measuring principle Variable area measuring principle
Measured value
Primary measured value Float position
Secondary measured value Operating and standard volumetric flow
Measuring accuracy
Directive VDI / VDE 3513, sheet 2 (qG = 50%)
H250 /RR /HC /F 1.6%
H250/C (Ceramic, PTFE) H250H, H250U, H250 (100 : 1)
TECHNICAL DATA 8
2.5%
Operating conditions
Temperature
Temperature
TemperatureTemperature Max. operating temperature TS -196..+300°C / -321...+572°F
Pressure
Pressure
PressurePressure
Max. operating pressure PS Depending on the version up to 400 bar / 5802 psig
Max. test pressure PT Pressure equipment directive 97/23/EC or AD 2000-HP30
Min. required operating pressure 2 times greater than pressure loss (see measuring ranges)
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:Float damping during gas measurement recommended: DN15...25 / ½"...1" Operating pressure <0.3 bar / 4.4 psig
DN50...100 / 2"...4" Operating pressure <0.2 bar / 2.9 psig
Installation conditions
Inlet run 5 x DN Outlet run 3 x DN
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
65
8 TECHNICAL DATA
Materials
H250
Device Flange / raised face Measur.
tube
H250/RR Stainless Steel
H250/HC Hastelloy
H250/C Ceramic/PTFE
H250/F - Food CrNi-Stahl 1.4435 -
1 CrNi steel 1.4571 on request, for clamp connection CrNi steel 1.4435 2 TFM/PTFE liner (electrically non-conductive)
®
CrNi steel 1.4404 massive 1
CrNi steel 1.4571 with plated Hastelloy
(2.4610) 1
CrNi-Stahl 1.4571 with TFM/PTFE liner 2
®
C4
CrNi steel 1.4404 1 -
Hastelloy® C4 (2.4610)
Float Float stop /
guide
PTFE or Al2O3 with FFKM
gasket
Al2O
3
and PTFE
INFORMATION!
H250/C - DN100 / 4" only PTFE H250/F: wetted surfaces Ra ≤0.8 μm, optional ≤0.6 μm
Other options:
Special materials on request: e.g. SMO 254, titanium, 1.4435
Float damping: ceramic or PEEK
Gasket for devices with female thread as insert: O-ring FPM / FKM
Ring orifice
-
Al2O
3
Temperatures
Temperatures
TemperaturesTemperatures
DANGER!
For devices to be used in hazardous areas, special temperature ranges apply. These can be found in the separate instructions.
Temperatures H250/M9 - mechanical indicator without power supply
Float Liner Product temperature Ambient temperature
[°C] [°F] [°C] [°F]
H250/RR Stainless Steel -196...+300 -321...+572 -40...+120 -40+248 H250/RR screw fitting -20+120 -4+248
H250/HC H250/C PTFE PTFE -196...+70 -321...+158 -40...+70 -40+158
H250/C Ceramic PTFE -196...+150 -321...+302 -40...+70 -40+158 H250/C Ceramic TFM / Ceramic -196...+250 -321...+482 -40...+120 -40+248 H250 H/U Stainless Steel -40...+100 -40+212 -20+90 -4+194
Hastelloy® C4
-196...+300 -321...+572 -40...+120 -40+248
66
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
Temperatures H250/M9 - with electrical components [°C]
Maximum product temperatures T
m
T
< +40°C T
amb.
< +60°C 1
amb.
EN ASME Version with Standard HT Standard HT
DN15, DN25
DN50 2" ESK2A, ESK3-PA +200 +300 +165 +300
DN80, DN100
½", 1" ESK2A, ESK3-PA +200 +300 +180 +300
ESK2A with counter +200 +300 +80 +130
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +130 +295
ESK2A with counter +180 +300 +75 +100
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +120 +195
3", 4" ESK2A, ESK3-PA +200 +300 +150 +250
ESK2A with counter +150 +270 +70 +85
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +190 +300 +110 +160
Temperatures H250/M9 - with electrical components [°F]
Maximum product temperatures T
m
EN ASME Version with Standard HT Standard HT
T
< +104 °F T
amb.
< +104 °F 1
amb.
DN15, DN25
DN50 2" ESK2A, ESK3-PA 392 572 165 572
DN80, DN100
½", 1" ESK2A, ESK3-PA 392 572 356 572
ESK2A with counter 392 572 176 266
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 266 563
ESK2A with counter 356 572 167 212
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 248 383
3", 4" ESK2A, ESK3-PA 392 572 302 482
ESK2A with counter 302 518 158 185
Limit switch NAMUR 392 572 392 572
3-wire limit switch 374 572 230 320
1 if there are no heat insulation measures, a heat-resistant cable is necessary (continuous operating temperature of the cable to be used: +100°C)
Abbreviation
HT High-temperature version ESK2A Current output 2-wire 420 mA
ESK3-PA PROFIBUS PA interface
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
67
8 TECHNICAL DATA
H250
Minimum ambient temperatures T
with ESK and limit switches
amb.
Device [°C] [°F]
Limit switch -25 / -40 -13 / -40
ESK2A - ESK3-PA -40 -40
Temperatures H250 /M8 /M10
[°C] [°F]
M8M
Min. product temperature T
Min. product temperature T
Ambient temperature T
without limit switches -80...+200 -112+392
m
with limit switches -25+200 -13+392
m
amb.
-25+70 -13+158
M8E
Max. product temperature Tm at T
Max. product temperature Tm at T
Max. product temperature Tm at T
Ambient temperature T
amb.
+40°C / +104°F -25...+200 -13…+392
amb.
+50°C / +122°F -25...+185 -13…+365
amb.
+60°C / +140°F -25...+145 -13…+293
amb.
-25+70 -13+158
M10
Max. product temperature Tm at T
Ambient temperature T
amb.
+60°C / +140°F -80...+200 -112…+392
amb.
-40+75 -40+167
68
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
Indicator M8
Indicator M8
Indicator M8Indicator M8
M8M limit switches
TECHNICAL DATA 8
Terminal connection
Limit switch I7S2002-N
Type 2-wire NAMUR 2-wire NAMUR 1 2-wire NAMUR 1
Switch configuration NC contact NC contact NO contact
Nominal voltage U
Pointer vane not read 3 mA 3 mA 1 mA Pointer vane read 1 mA 1 mA 3 mA
1 safety oriented
0
2
2.5 mm
SC2-N0
8VDC 8VDC 8VDC
SJ2-SN SJ2-S1N
M8E current output
Cable gland M16 x 1.5 Cable diameter 810 mm
Terminal connection Measuring signal 4...20 mA = 0100 % flow value in 2-wire technology
Power supply 14.8...30 VDC
Min. power supply for HART
Power supply influence < 0.1%
Dependence on external resistance < 0.1% Temperature influence < 10 µA / K
Max. external resistance / load 640 Ohm (30 VDC)
Min. load for HART
®
®
2
4 mm
20.5 VDC
250 Ohm
M8E HART® configuration
Manufacturer name (code) KROHNE Messtechnik (69)
Model name M8E (230)
HART® protocol revision
Device revision 1
Physical layer FSK
Device category Transmitter
5.1
M8E process variable
M8E process variable flow rate Values [%] Signal output [mA] Over range +102,5 (±1%) 20,24...20,56 Device error identification >106,25 21,00
Maximum 112,5 22
Multidrop operation - 4,5
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
69
8 TECHNICAL DATA
Indicator M9
Indicator M9
Indicator M9Indicator M9 M9 cable glands
Cable gland Material Cable diameter
M 16x1.5 Standard PA 3...7 mm 0.118...0.276"
M20 x 1.5 PA 8...13 mm 0.315...0.512"
M 16x1.5 Nickel-plated brass 5...9 mm 0.197...0.355"
M20 x 1.5 Nickel-plated brass 10...14 mm 0.394...0.552"
M9 limit switches
H250
Terminal connection
Limit switch I7S23,5-N
NAMUR yes yes yes no
Connection type 2-wire 2-wire 2-wire 3-wire
Switching element function
Nominal voltage U
Pointer vane not detected 3 mA 3 mA 1 mA 0.3 VDC Pointer vane detected 1 mA 1 mA 3 mA UB - 3 VDC
Continuous current - - - max. 100 mA
No load current I
1 safety oriented
0
0
2.5 mm
SC3,5-N0
NC contact NC contact NO contact PNP NO contact
8 VDC 8 VDC 8 VDC 10...30 VDC
- - - 15 mA
2
SJ3,5-SN 1 SJ3,5-S1N 1 SB3,5-E2
M9 current output ESK2A
Terminal connection
Power supply 12...30 VDC
Min. power supply for HART
Measuring signal 4.00...20.00 mA = 0...100% flow value in 2-wire technology
Power supply influence <0.1%
Dependence on external resistance <0.1% Temperature influence < 10µA / K
Max. external resistance / load 800 Ohm (30 VDC)
Min. load for HART
Software firmware version 02.15
Ident No. 4000054602
®
®
2.5 mm
18 VDC
250 Ohm
2
70
ESK2A HART ® configuration
Manufacturer name (code) KROHNE Messtechnik (69 = 45h)
Model name ESK2A (226 = E2h)
HART® protocol revision
Device revision 1
Physical layer FSK
Device category Transmitter without galvanic isolation
5.9
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
M9 ESK2A process variable
ESK2A process variable flow rate Values [%] Signal output [mA] Over range +102.5 (±1%) 20.24...20.56
Device error identification > 106.25 >21.00
Maximum 131.25 25
Multidrop operation - 4.5
Min. U
ext.
M9 ESK-Z totalizer
12 VDC
Terminal connection
Power supply 10...30 VDC
R
current loop 0...600 Ohm
ext.
Power consumption max. 2.5 Watt
Indication error < 1% in relation to the value displayed
Max. reset voltage 30 VDC
Min. reset pulse 300 ms
Software firmware version 1.19
Power supply 10...30 VDC
Max. current 50 mA
Max. dissipation 250 mW
T on 80 ms fixed pulse width
T off depending on flow rate U on Ub – 3 VDC
U off 0 VDC
Pulse value
2
2.5 mm
1 pulse = 1 display counter advance (1 litre, 1 m3 ...)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
71
8 TECHNICAL DATA
Indicator M9 ESK3-PA Profibus
H250
Terminal connection Bus cable R´ 15...150 Ohm/km
Bus cable L´ 0.4...1 mH/km Bus cable C´ 80...200 nF/km
2.5 mm
2
M9 ESK3PA Hardware
Hardware acc. to IEC 1158-2 and FISCO model
Supply voltage 9...32 VDC
Base current 12 mA
Starting current < base current
FDE (fault drop electronics) < 18 mA
Accuracy acc. to VDI/ VDE 3513 1.6
Measurement resolution < 0.1% full scale value
Temperature influence < 0.05% / K full scale value
Software firmware version 1.01/000418
Ident No. 3184980200
M9 ESK3PA Software
GSD Devices master file
Device profile Profiles B, V3.0
Function blocks
Flow rate (AI0) Volume or mass
Totalizer (TOT0) Volume totalizer
Default unit: [m
Totalizer (TOT1) Mass totalizer
Default unit: [kg]
Address range 0...126, default 126
SAP`s Service Access Points
DD Device Description
3
]
72
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
Indicator M10
Indicator M10
Indicator M10Indicator M10
M10 cable gland
(Standard) without
M20 x 1.5 On request
M 20x1.5 Ex d On request
M10 current output
Terminal connection Power supply 24 VDC ±30%
Min. power supply for HART
Measuring signal 4.00...20.00 mA = 0...100% flow value in 2-wire technology
Power supply influence < 0.1%
Dependence on external resistance < 0.1% Temperature influence < 5 µA/K Max. external resistance / load 630 Ohm Min. load for HART 250 Ohm
Software firmware version 02.17
Ident No. 3209470500
®
2.5 mm
18 VDC
2
M10 HART
Manufacturer name (code) KROHNE Messtechnik (69 = 45h)
Model name M10 (234 = EA)
HART® protocol revision
Device revision 1
Physical layer FSK
Device category Transmitter
®
5.9
M10 process variable
Values [%] Signal output [mA]
Over range +105 (±1%) 20,64...20.96
Device error identification > 110 > 21.60
Maximum 112.5 22
Multidrop operation - 4.5
Lift-off voltage 12 VDC
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
73
8 TECHNICAL DATA
M10 binary output
Two binary outputs Galvanically isolated
Operating mode Switch output NAMUR or open collector
Configurable as Switch contact or
NAMUR switch output
Power supply 8 VDC
Signal current > 3 mA switching value not
Switch output, open collector
Power supply 8...30 VDC
P
max
I
max
M10 reset input
Binary input Galvanically isolated
Operating mode Reset counter
Configurable as active Hi / active Lo
Voltage level 5...30 VDC Current consumption 1 mA Pulse length (active) 500 ms
pulse output
reached;
500 mW
100 mA
H250
open / closed or max. 10 pulses / s
< 1 mA switching value reached
74
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
Approvals
Standard Indicator Designation
ATEX M9 mechanical II2GD IIC
M9 electrical II2G Ex ia IIC T6
M8 mechanical II2GD IIC
M8 electrical II2G Ex ia IIC T6...T1
M10 II2G Ex d IIC T6...T1
FM M9 IS/I/1/ABCD;T6
M10 XP/I/1/ABCD;T6
CSA M10 XP/I/1/ABCD;T6
Nepsi M9 Ex ia IIC T1-T6
M8 Ex iA IIC T1-T6
M10 Ex d IIC T1-T6
INMETRO M10 II2G EEx d IIC T6...T1
II3GD IIC
II3G Ex nA II T6 II3D IP65 T65°C
II3GD IIC
II3D Ex tD A22 IP66 T65°C
NI/I/2/ABCD;T6 IS/I, II, III/1/A-G NI/II/2/ABCD
NI/I/2/ABCD;T6 XP/I/1/IIC/T6 NI/I/2/IIC/T6 DIP/II,III/1/EFG/T6 S/II,III/2/FG/T6
NI/I/2/ABCD;T6 XP/I/1/IIC/T6 NI/I/2/IIC/T6 DIP/II,III/1/EFG/T6 S/II,III/2/FG/T6
Ex nA II T1-T6
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
75
8 TECHNICAL DATA
8.3 Dimensions and weights
Dimensions H250/M9
Front view Side view with heating High-temperature
H250
a b d h
[mm] ["] [mm] ["] [mm] ["] [mm] ["]
All nominal sizes 138 5.44 250 9.85 181 7.13 150 5.91
ISO 228 300 11.82
H250/C - 3"/300 lb 300 11.82
EN ASME c e Ø f g j
[mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 ½" 110.5 4.35 107 4.22 20 0.79 100 3.94 187 7.37
DN25 1" 110.5 4.35 119 4.69 32 1.26 106 4.18 199 7.84
DN50 2" 123.5 5.22 132 5.20 65 2.56 120 4.73 212 8.35
DN80 3" 123.5 5.22 148 5.83 89 3.51 145 5.71 228 8.98
DN100 4" 123.5 5.22 158 6.22 114 4.49 150 5.91 232 9.14
ISO 228 female thread screwed
ISO 228 female thread welded
H250/F Clamp connection
H250/F Screw connection DIN 11851
76
1
1 Stainless steel 1.4435 - EHEDG tested - wetted surfaces Ra 0.8 / 0.6 µm
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
Dimensions H250/M10 /M8
M10 Front view
M10 Side view
M8 Front view
M8 Side view
Dimensions M10
a b c Ø d e
EN ASME [mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 ½" 147 5.79 83 3.27 118 4.65 132 5.20 55 2.17
DN25 1" 147 5.79 83 3.27 130 5.12 132 5.20 55 2.17
DN50 2" 147 5.79 83 3.27 143 5.63 132 5.20 55 2.17
DN80 3" 147 5.79 83 3.27 160 6.30 132 5.20 55 2.17
DN100 4" 147 5.79 83 3.27 169 6.66 132 5.20 55 2.17
Dimensions M8M Dimensions M8E
f g h f g h
EN ASME [mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 ½" 63 2.48 60 2.36 58.5 2.30 53.5 2.11 66 2.60 52.5 2.07
DN25 1" 75 2.95 60 2.36 58.5 2.30 65.5 2.58 66 2.60 52.5 2.07
DN50 2" 89 3.51 73 2.88 45.5 1.79 79.5 3.13 79 3.11 39.5 1.56
DN80 3" 105 4.14 73 2.88 45.5 1.79 95.5 3.76 79 3.11 39.5 1.56
DN100 4" 114 4.49 73 2.88 45.5 1.79 104 4.12 79 3.11 39.5 1.56
For overall height see devices with indicator M9
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
77
8 TECHNICAL DATA
Weights
Weights
WeightsWeights
Nominal size EN 1092-1 Flange connection Ermeto connection
EN ASME [kg] [lb] [kg] [lb] [kg] [lb]
DN15 ½" 3.5 7.7 5.6 12.6 3.9 8.6
DN25 1" 5 11 7.5 16.5 5.8 12.8
DN50 2" 8.2 18.1 11.2 24.7 9.5 21
DN80 3" 12.2 26.9 14.8 32.6 13.1 28.9
DN100 4" 14 30.9 17.4 38.4 15.7 34.6
Nominal size EN 1092-1 ASME 150 lb ASME 300 lb DIN 11864-1
EN ASME [kg] [lb] [kg] [lb] [kg] [lb] [kg] [lb] DN15 ½" 3.5 7.7 3.2 7.1 3.5 7.7 2 4.4
DN25 1" 5 11 5.2 11.5 6.8 15 3.5 7.7
DN50 2" 10 22.1 10 22.1 11 24.3 5 11
DN80 3" 13 28.7 13 28.7 15 33.1 7.6 16.8
DN100 4" 15 33.1 16 35.3 17 37.5 10.3 22.7
H250
H250 with heating
H250/C [Ceramic / PTFE] Screw connect.
Process connections
Process connections
Process connectionsProcess connections
Standards Conn. dim. Pressure rating
Flanges (H250/RR /HC /C) EN 1092-1 DN15...150 PN16...250
ASME B16.5 ½...6" 150...2500 lb JIS B 2220 15100 10...20K
Clamp connections (H250/RR /F) DlN 32676 DN15...100 10...16 bar
ISO 2852 Size 25...139.7 10...16 bar
Screw connections (H250/RR /HC /F) DIN 11851 DN15...100 25...40 bar
SMS 1146 1...4" 6 bar / 88.2 psig
Female thread welded (H250/RR /HC) ISO 228 G½...G2" 50 bar / 735
ASME B1.20.1 ½…2" NPT
Female thread (H250/RR /HC) with insert, FPM gasket and union nut
Thread connection aseptic (H250/F) DIN 11864 - 1 DN1550 PN40
Flange aseptic (H250/F) DIN 11864 - 2 DN1550 PN40
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:Meters (H250/RR /HC) with heating:
Heating with flange connection EN 1092-1 DN15 PN40
Heating pipe connection for Ermeto - E12 PN40
ISO 228 G½…2" 50 bar / 735 ASME B1.20.1 ½…2" NPT
DN80100 PN 16
DN80DN100 PN 16
ASME B16.5 ½" 150 lb / RF
psig
psig
78
Higher pressure ratings and other connections on request
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
Bolts and tightening torques
For flowmeters with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal sizes EN
Nominal size Bolts Tightening torques
acc. to EN 1092-1 Quantity x size [Nm] [lb-ft]
DN15 PN40 1 4 x M 12 9.8 7.1
DN25 PN40 1 4 x M 12 21 15
DN50 PN40 1 4x M16 57 41
DN80 PN16 1 8x M16 47 34
DN100 PN16 1 8x M16 67 48
1 standard connections; other connection on request
Nominal size ASME
Nominal size Bolts (Quantity x size) Tightening torques
acc. to ASME B 16.5 150 lb 300 lb [Nm] [lb-ft]
½" 150 lb / 300 lb 1 4x ½" 4x ½" 5.2 3.8 1" 150 lb / 300 lb 1 4x ½" 4x 5/8" 10 7.2
2" 150 lb / 300 lb 1 4x 5/8" 8x 5/8" 41 30 3" 150 lb / 300 lb 1 4x 5/8" 8x ¾" 70 51 4" 150 lb / 300 lb 1 8x 5/8" 8x ¾" 50 36
1 standard connections; other connections on request
Low pressure resistance (vacuum) H250/C
Max. process temperature  +70°C (+158°F) +150°C (*302°F) +250°C (+482°F)
Min. operating pressure
Nominal size float lining [mbar
abs.]
DN15...DN100 PTFE PTFE 100 1,45 - - - -
DN15...DN80 ceramic PTFE 100 1,45 250 3,63 - -
DN15...DN80 ceramic TFM /
ceramic
[psia] [mbar
abs.]
100 1,45 100 1,45 100 1,45
[psia] [mbar
abs.]
[psia]
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
79
8 TECHNICAL DATA
8.4 Measuring ranges
H250
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
®
Water Air Max. pressure loss
Nominal
Float TIV CIV DIV TIV
Cone [l/h]
TIV DIV TIV
Alu
[Nm3/h]
TIV CIV DIV
Alu
[mbar]
size
DN15, ½" K 15.1 18 25 - 0.42 0.65 - 12 21 26 -
K 15.2 30 40 - 0.7 1 - 12 21 26 -
K 15.3 55 63 - 1 1.5 - 12 21 26 -
K 15.4 80 100 - 1.7 2.2 - 12 21 26 -
K 15.5 120 160 - 2.5 3.6 - 12 21 26 -
K 15.6 200 250 - 4.2 5.5 - 12 21 26 -
K 15.7 350 400 700 6.7 10 18 1 12 21 28 38
K 15.8 500 630 1000 10 14 28 1 13 22 32 50
K 15.8 - - 1600 2 - - 50 2 - - - 85
DN25, 1" K 25.1 480 630 1000 9.5 14 - 11 24 32 72
K 25.2 820 1000 1600 15 23 - 11 24 33 74
K 25.3 1200 1600 2500 22 35 - 11 25 34 75
K 25.4 1700 2500 4000 37 50 110 1 12 26 38 78
K 25.5 3200 4000 6300 62 95 180 1 13 30 45 103 3
DN50, 2" K 55.1 2700 6300 8400 58 80 230 1 8 13 74 60
K 55.2 3600 10000 14000 77 110 350 1 8 13 77 69
K 55.3 5100 16000 25000 110 150 700 1 9 13 84 104
DN80, 3" K 85.1 12000 25000 37000 245 350 1000 1 8 16 68 95
K 85.2 16000 40000 64000 280 400 1800 1 9 16 89 125
DN100, 4" K105.1 19000 63000 100 000 - 550 2800 1 - - 120 220
1 P >0.5 bar 2 with TR float 3 300 mbar with damping (gas measurement)
80
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
®
Water Air Max. pressure loss
Nominal
Float TIV CIV DIV TIV
Alu
Cone [GPH] [SCFM] [psig]
TIV DIV TIV
Alu
TIV CIV DIV
size
DN15, ½" K 15.1 4.76 6.60 - 0.26 0.43 - 0.18 0.31 0.38 -
K 15.2 7.93 10.6 - 0.43 0.62 - 0.18 0.31 0.38 -
K 15.3 14.5 16.6 - 0.62 0.93 - 0.18 0.31 0.38 -
K 15.4 21.1 26.4 - 1.05 1.36 - 0.18 0.31 0.38 -
K 15.5 31.7 42.3 - 1.55 2.23 - 0.18 0.31 0.38 -
K 15.6 52.8 66.0 - 2.60 3.41 - 0.18 0.31 0.38 -
K 15.7 92.5 106 185 4.15 6.20 11.2 1 0.18 0.31 0.41 0.56
K 15.8 132 166 264 6.20 8.68 17.4 1 0.19 0.32 0.47 0.74
K 15.8 - - 423 2 - - 31.0 2 - - - 1.25
DN25, 1" K 25.1 127 166 264 5.89 8.68 - 0.16 0.35 0.47 1.06
K 25.2 217 264 423 9.30 14.3 - 0.16 0.35 0.49 1.09
K 25.3 317 423 660 13.6 21.7 - 0.16 0.37 0.50 1.10
K 25.4 449 660 1057 22.9 31.0 68.2 1 0.18 0.38 0.56 1.15
K 25.5 845 1057 1664 38.4 58.9 111 1 0.19 0.44 0.66 1.51 3
DN50, 2" K 55.1 713 1664 2219 36.0 49.6 143 1 0.12 0.19 1.09 0.88
K 55.2 951 2642 3698 47.7 68.2 217 1 0.12 0.19 1.13 1.01
K 55.3 1347 4227 6604 68.2 93.0 434 1 0.13 0.19 1.23 1.53
DN80, 3" K 85.1 3170 6604 9774 152 217 620 1 0.12 0.24 1.00 1.40
K 85.2 4227 10567 16907 174 248 1116 1 0.13 0.24 1.31 1.84
DN100, 4" K105.1 5019 16643 26418 - 341 1736 1 - 1.76 3.23
1 P >7.4 psig 2 with TR float 3 4.4 psig with damping (gas measurement)
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
81
8 TECHNICAL DATA
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow rate Max. pressure loss
Water Air Water Air
Liner / Float PTFE Ceramic Ceramic PTFE Ceramic Ceramic
H250
Nominal
Cone [l/h]
[Nm3/h]
[mbar]
size
DN15, ½" E 17.2 25 30 - 65 62 62
E 17.3 40 50 1.8 66 64 64
E 17.4 63 70 2.4 66 66 66
E 17.5 100 130 4 68 68 68
E 17.6 160 200 6.5 72 70 70
E 17.7 250 250 9 86 72 72
E 17.8 400 - - 111 - -
DN25, 1" E 27.1 630 500 18 70 55 55
E 27.2 1000 700 22 80 60 60
E 27.3 1600 1100 30 108 70 70
E 27.4 2500 1600 50 158 82 82
E 27.5 4000 1 2500 75 290 100 100
DN50, 2" E 57.1 4000 4500 140 81 70 70
E 57.2 6300 6300 200 110 80 80
E 57.3 10000 11000 350 170 110 110
E 57.4 16000 1 - - 284 - -
DN80, 3" E 87.1 16000 16000 - 81 70 -
E 87.2 25000 25000 - 95 85 -
E 87.3 40000 1 - - 243 - -
DN100, 4" E 107.1 40000 - - 100 - -
E 107.2 60000 1 - - 225 - -
1 special float
82
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow rate Max. pressure loss
Water Air Water Air
Liner / Float PTFE Ceramic Ceramic PTFE Ceramic Ceramic
Nominal
Cone [GPH] [SCFM] [psig]
size
DN15, ½" E 17.2 6.60 7.93 - 0.96 0.91 0.91
E 17.3 10.6 13.2 1.12 0.97 0.94 0.94
E 17.4 16.6 18.5 1.49 0.97 0.97 0.97
E 17.5 26.4 34.3 2.48 1.00 1.00 1.00
E 17.6 42.3 52.8 4.03 1.06 1.03 1.03
E 17.7 66.0 66.0 5.58 1.26 1.06 1.06
E 17.8 106 - - 1.63 - -
DN25, 1" E 27.1 166 132 11.2 1.03 0.81 0.81
E 27.2 264 185 13.6 1.18 0.88 0.88
E 27.3 423 291 18.6 1.59 1.03 1.03
E 27.4 660 423 31.0 2.32 1.21 1.21
E 27.5 1056 1 660 46.5 4.26 1.47 1.47
DN50, 2" E 57.1 1057 1189 86.8 1.19 1.03 1.03
E 57.2 1664 1664 124 1.62 1.18 1.18
E 57.3 2642 2906 217 2.50 1.62 1.62
E 57.4 4226 1 - - 4.17 - -
DN80, 3" E 87.1 4227 4227 - 1.19 1.03 -
E 87.2 6604 6604 - 1.40 1.25 -
E 87.3 10567 1 - - 3.57 - -
DN100, 4" E 107.1 10567 - - 1.47 - -
E 107.2 15850 1 - - 3.31 - -
1 special float
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
83
8 TECHNICAL DATA
H250H - Horizontal installation position
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
H250
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½ K 15.1 70 1.8 195
K 15.2 120 3 204
K 15.3 180 4.5 195
K 15.4 280 7.5 225
K 15.5 450 12 250
K 15.6 700 18 325
K 15.7 1200 30 590
K 15.8 1600 40 950
K 15.8 2400 60 1600
DN25 1" K 25.1 1300 35 122
K 25.2 2000 50 105
K 25.3 3000 80 116
K 25.4 5000 130 145
K 25.5 8500 220 217
K 25.5 10000 260 336
DN50 2" K 55.1 10000 260 240
K 55.2 16000 420 230
K 55.3 22000 580 220
K 55.3 34000 900 420
DN80 3" K 85.1 25000 650 130
K 85.2 35000 950 130
K 85.2 60000 1600 290
DN100 4" K 105.1 80000 2200 250
K 105.1 120000 3200 340
84
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250H - Horizontal installation position
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
EN ASME Cone Water [GPH] Air [SCFM] Pressure loss
[psig]
DN15 1/2" K 15.1 18.5 1.12 2.87
K 15.2 31.7 1.86 3.00
K 15.3 47.6 2.79 2.87
K 15.4 74.0 4.65 3.31
K 15.5 119 7.44 3.68
K 15.6 185 11.2 4.78
K 15.7 317 18.6 8.68
K 15.8 423 24.8 14.0
K 15.8 634 37.2 23.5
DN25 1" K 25.1 343 21.7 1.79
K 25.2 528 31.0 1.54
K 25.3 793 49.6 1.71
K 25.4 1321 80.6 2.13
K 25.5 2245 136 3.19
K 25.5 2642 161 4.94
DN50 2" K 55.1 2642 161 3.53
K 55.2 4227 260 3.38
K 55.3 5812 360 3.23
K 55.3 8982 558 6.17
DN80 3" K 85.1 6604 403 1.91
K 85.2 9246 589 1.91
K 85.2 15851 992 4.26
DN100 4" K 105.1 21134 1364 3.68
K 105.1 31701 1984 5.00
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
85
8 TECHNICAL DATA
H250U - Vertical installation position
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow direction from top to bottom
H250
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½" K 15.1 65 1.6 175
K 15.2 110 2.5 178
K 15.3 170 4 180
K 15.4 260 6 200
K 15.5 420 10 220
K 15.6 650 16 290
K 15.7 1100 28 520
K 15.8 1500 40 840
DN25 1" K 25.1 1150 30 97
K 25.2 1800 45 85
K 25.3 2700 70 92
K 25.4 4500 120 115
K 25.5 7600 200 172
DN50 2" K 55.1 9000 240 220
K 55.2 15000 400 230
K 55.3 21000 550 240
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
86
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com 04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250U - Vertical installation position
Measuring span: 10 : 1 Declaration of flow: Values = 100% Water: 20°C [68°F] Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow direction from top to bottom
EN ASME Cone Water [GPH] Air [SCFM] Pressure loss
[psig]
DN15 ½" K 15.1 17.2 0.99 2.57
K 15.2 29.1 1.55 2.62
K 15.3 44.9 2.48 2.65
K 15.4 68.7 3.72 2.94
K 15.5 111 6.20 3.23
K 15.6 172 9.92 4.26
K 15.7 291 17.4 7.64
K 15.8 396 24.8 12.3
DN25 1" K 25.1 304 18.6 1.42
K 25.2 476 27.9 1.25
K 25.3 713 43.4 1.35
K 25.4 1189 74.4 1.69
K 25.5 2008 124 2.53
DN50 2" K 55.1 2378 149 3.23
K 55.2 3963 248 3.38
K 55.3 5548 341 3.53
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5 times the pressure loss for gases! The specified pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data (pressure, temperature, density, viscosity) is performed using the calculation method in accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
87
KROHNE product overview
Electromagnetic flowmeters
Variable area flowmeters
Ultrasonic flowmeters
Mass flowmeters
Vortex flowmeters
Flow controllers
Level meters
Temperature meters
Pressure meters
Analysis products
Measuring systems for the oil and gas industry
Measuring systems for sea-going tankers
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 D-47058 Duisburg (Germany) Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de
The current list of all KROHNE contacts and addresses can be found at:
© KROHNE 04/2011 - 4000269302 MA H250-R02 en - Subject to change without notice.
www.krohne.com
Loading...