All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
The variable area flowmeters are suitable for measuring gases, vapours and liquids.
The devices are particularly suitable for the measurement of:
• Water
• Hydrocarbons
• Corrosion protection agents and lubricants
• Chemicals and additives
• Solvents
• Superheated steam
• Food, beverages and tobacco
• Air
• Industrial gases
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
SAFETY INSTRUCTIONS 1
WARNING!
The operator himself bears the sole responsibility for the intended use of the device regarding
the suitability and the corrosion resistance of the used materials against the measured fluid.
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
Do not use any abrasive media containing solid particles or highly viscous media.
1.2 Certifications
CE marking
The device fulfils all applicable statutory requirements of the following EC directives:
• Pressure Equipment Directive 97/23/EC
• For devices with electrical installations: EMC Directive 2004/108/EC
• Devices for use in hazardous areas: ATEX Directive 94/9/EC
as well as
• NAMUR recommendations NE 21 and NE 43
The manufacturer certifies successful testing of the product by applying the CE marking.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
5
1 SAFETY INSTRUCTIONS
1.3 Safety instructions from the manufacturer
1.3.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
H250
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.3.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
6
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
1.3.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.3.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
7
1 SAFETY INSTRUCTIONS
1.3.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
H250
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.4 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
8
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-1: Scope of delivery
1 Measuring device in ordered version
2 For indicator M10 - bar magnet
3 For indicator M10 - key
4 Documentation
5 Certificates, calibration report (supplied to order only)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
9
2 DEVICE DESCRIPTION
2.2 Device version
• H250 with indicator M9
• H250 with indicator M10
• H250 with indicator M8
H250
Figure 2-2: Versions M9 and M10
1 H250/RR/M9
• Local indication without auxiliary power supply
• max. 2 limit switches, type NAMUR, NAMUR safety-oriented or transistor (3-wire)
®
• 2-wire current output 4…20 mA, HART
or Profibus-PA communication
• 6-digit flow counter (non Ex)
• Limit switches and signal output - optional intrinsically safe
2 H250/RR/M10
• Explosion proof enclosure
• 2 digital adjustable limit switches, 2-wire open collector or type NAMUR
®
• 2-wire current output 4…20 mA, HART
• Pulse output up to 10 Hz (also for electro-mechanical counters)
• 12-digit flow counter with external resetting (batch operation)
The following designs are available as options:
• H250 with indicator M9 as high-temperature version HT
• H250 with indicator M9 with added impact and corrosion protection (special paint finish)
• H250 with indicator M9 in Stainless Steel housing
communication
10
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
Figure 2-3: Version M8
1 H250/RR/M8EG
• Electronic bar graph indicator
• 2-wire current output 4…20 mA, HART
®
communication
DEVICE DESCRIPTION 2
2 H250/RR/M8MG
• Local indication without auxiliary power supply
• 2 limit switches, 2-wire, type NAMUR or NAMUR safety-oriented
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
11
2 DEVICE DESCRIPTION
2.2.1 Float damping
Float damping is characterised by high standstill times and self-centering. The damping sleeve
is made of high performance ceramic or PEEK, depending on the medium and the application.
Float damping can also be retrofitted for the user (see Service).
Use of damping
• Generally when CIV and DIV floats are used for gas measurement.
• For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
Nominal size acc. toOperating primary pressure
EN 1092-1ASME B16.5[bar][psig]
DN 50½"≤0.3≤4.4
DN251"≤0.3≤4.4
DN502"≤0.2≤2.9
DN803"≤0.2≤2.9
DN 1004"≤0.2≤2.9
H250
2.2.2 Pointer damping
The pointer system with its magnetic system basically contains pointer damping. An additional
eddy current brake is advantageous for fluctuating or pulsing flows. The eddy current brake
magnets surround the pointer vane 1 without touching it, damping its movement. The result is a
pointer position that is considerably calmer, and no distortion of the measured value. A clamp
screw holds it in place securely. The eddy current brake can be retrofitted without having to
recalibrate and while in operation (see Service).
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-4: Nameplates on the indicator
1 Device type
2 Manufacturer
3 Notified ATEX & PED body
4 Rating data: temperature & pressure rating
5 PED data
6 Ex data
7 Electrical connection data
8 Internet site
Additional markings on the indicator
• SN - serial number
• SO - sales order / item
• PA - order
• Vx - product configurator code
• AC - article code
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
13
2 DEVICE DESCRIPTION
2.4 Description code
The description code* consists of the following elements:
1 Device type
H250 - standard version
H250H - horizontal flow direction
H250U - flow direction from top to bottom
2 Materials / versions
RR - Stainless Steel
C - PTFE or PTFE/ceramics
HC - Hastelloy
Ti - Titanium
F - aseptic version (food)
3 Heating jacket version
B - with heating jacket
H250
4 Series of indicators
M8 - Indicator M8
M9 - Indicator M9 standard indicator
M9S - Indicator with added impact and corrosion protection
M9R - Indicator in Stainless Steel housing
M10 - Indicator or signal converter M10
5 Design of indicator M8
MG - Mechanical indicator
EG - Electronic indicator with signal output 4...20 mA
6 High-temperature version
HT - Version with HT extension
7 Electrical signal output
ESK - Current output or Profibus-PA
ESK-Z - Current output and totalizer
8 Limit switch
K1 - One limit switch
K2 - Two limit switches
S1 - One SIL2 Limit switch acc. to IEC 61508
S2 - Two SIL2 Limit switches acc. to IEC 61508
9 Explosion protection
Ex - Explosion-protected equipment
10 SIL
SK - SIL2 compliance of limit switches acc. to IEC 61508
14
* positions which are not needed are omitted (no blank positions)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in its original packing.
• The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
INSTALLATION 3
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
15
3 INSTALLATION
3.3 Installation conditions
CAUTION!
When installing the device in the piping, the following points must be observed:
•
The variable area flowmeter must be installed vertically (measuring principle). Flow direction
from bottom to top. For installation recommendations please refer also to VDI/VDE 3513
Sheet 3.
H250Hs are installed horizontally and H250U devices are installed vertically with the flow
direction from top to bottom.
•
A straight unimpeded inlet run of ≥ 5x DN upstream of the device and a straight outlet run of
≥ 3x DN downstream of the device are recommended.
•
Screws, bolts and gaskets are to be provided by the customer and must be selected in
accordance with the pressure rating of the connection or the operating pressure.
•
The inside diameter of the flange deviates from the standard dimensions. Flange seal
standard DIN 2690 can be applied without any limitation.
•
Align the gaskets. Tighten the nuts with the tightening torques of the appropriate pressure
rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, see chapter "Tightening
torques".
•
Control devices are to be positioned downstream of the measuring device.
•
Shutoff devices are preferably to be positioned upstream of the measuring device.
•
Before connecting, blow or flush out the pipes leading to the device.
•
Pipes for gas flow need to be dried before the device is installed.
•
Use connectors suitable for the particular device version.
•
Align the pipes axially with the connections on the measuring device so they are free of
stresses.
•
If necessary, the piping has to be supported to prevent vibrations being transmitted to the
measuring device.
•
Do not lay signal cables directly next to cables for the power supply.
H250
16
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
INSTALLATION 3
Take special note of the installation position for the H250H with horizontal flow
direction:
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for
horizontal installation are to be installed in the pipeline so that the display is located on the side
of the measuring tube. The maximum medium and ambient temperatures indicated as well as
the measuring accuracy are based on lateral installation of the display.
3.3.1 Tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange
threads with the following torques:
Nominal size acc. toBoltsMax. torque
EN 1092-1ASME B16.5EN
DNPNIncheslb150 lb300 lbNmft*lbfNmft*lbf
1540½"150/3004 x M 124 x ½"4 x ½"9.87.15.23.8
25401"150/3004 x M 124 x ½"4 x 5/8"2115107.2
50402"150/3004 x M 16 4 x 5/8" 8 x 5/8"57414130
80163"150/3008 x M 16 4 x 5/8"8 x ¾"47347051
100164"150/3008 x M 16 8 x 5/8"8 x ¾"67485036
ASMEEN 1092-1ASME 150 lb
1092-1
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
17
3 INSTALLATION
3.3.2 Magnetic filters
The use of magnetic filters is recommended when the medium contains particles which can be
influenced magnetically. The magnetic filter is to be installed in the flow direction upstream of
the flowmeter. Bar magnets are positioned helically in the filter to provide optimal efficiency at
low pressure loss. All of the magnets are coated individually with PTFE to protect against
corrosion. Material: 1.4571
Magnetic filters
H250
1 Type F - fitting part with flange - overall length 100 mm
2 Type FS - fitting part without flange - overall length 50 mm
18
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
3.3.3 Heat insulation
CAUTION!
The indicator housing may not be heat-insulated.
The heat insulation 3 may only reach as far as the housing fastening 4.
Figure 3-1: H250 heat insulation
1 Standard indicator M9
2 Indicator with HT extension
INSTALLATION 3
This applies in the same manner to indicators M8 and M10.
CAUTION!
The heat insulation 1 may only reach to the rear of the housing 2. The area of the cable entries
3 must be freely accessible.
Figure 3-2: Insulation - cross section
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
19
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
H250
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
20
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.2 Electrical connection indicator M8
4.2.1 Indicator M8M - limit switches
The limit switches can be set over the entire measuring range using the limit pointer 1. The set
limit values are displayed on the scale. The pointers are set to the desired limit values using a
slip coupling along the scale.
ELECTRICAL CONNECTIONS 4
Figure 4-1: Limit switch settings M8MG
1 Maximum pointer, switching point indicator
2 Limit switch
3 Connection board
4 Connection terminal
4.2.2 Indicator M8E - current output
Figure 4-2: Electrical connection M8EG
1 Terminal connection
2 Power supply 14.8...30 VDC
3 Measuring signal 4...20 mA
4 External load, HART
®
communication
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
21
4 ELECTRICAL CONNECTIONS
Power supply M8 with electrical isolation
The circuitry for connection to other devices such as digital evaluator units or process control
equipment must be designed with especial care. In some circumstances internal connections in
these devices (e.g. GND with PE, ground loops) may lead to impermissible voltage potentials,
which can compromise the function of the device itself or a connected device. In such cases a
protected extra-low voltage (PELV) is recommended.
H250
Figure 4-3: Electrical connection M8EG with electrical isolation
1 Terminal connection
2 Converter supply isolator with electrical isolation
3 Power supply (see supply isolator information)
4 Measuring signal 4...20 mA
5 External load, HART
®
communication
Power supply
INFORMATION!
The supply voltage has to be between 14.8 VDC and 30 VDC. This is based on the total resistance
of the measuring loop. To determine this, add up the resistances of each component in the
measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
22 mA + 14.8 V
L
where
U
= the minimum supply voltage and
ext.
= the total measuring loop resistance is.
R
L
22
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
HART® communication
When HART® communication is carried out with the M8E display, the analogue measured data
transmission (4...20 mA) is not impaired in any way.
®
Exception for multidrop mode. In multidrop mode, a maximum of 15 devices with HART
can be operated in parallel, whereby their current outputs are switched inactive
(I approx. 4 mA per device).
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
communication a load of at least 230 ohm is required.
function
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary. The cable shield may only be earthed
(grounded) at one place (on the power supply unit).
Configuration
The M8E electronic indicator can be configured via HART® communication. DD (Device
Descriptions) for AMS 6.x and PDM 5.2 as well as a DTM (Device Type Manager) are available for
configuration. They can be downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART
counter can be configured. Two limit values can be set and monitored. The limit values are
assigned either to flow values or to the counter overflow. The limit values are not depicted on the
display.
communication. A flow
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
23
4 ELECTRICAL CONNECTIONS
4.3 Electrical connection indicator M9
4.3.1 Indicator M9 - limit switches
The M9 indicator can be equipped with a maximum of two electronic limit switches. The limit
switch functions as a slot sensor which is operated inductively through the semicircular metal
vane belonging to the measuring pointer. The switching points are set using the contact pointers.
The position of the contact pointer is indicated on the scale.
Limit switch module
H250
1 Min. contact
2 Max. contact
3 Locking screw
4 Maximum pointer
5 Connection terminal
The connecting terminals have a pluggable design and can be removed in order to connect the
cables. The built-in limit switch types are shown on the indicator.
1 Contact pointer MAX
2 Contact pointer MIN
3 Locking screw
Setting is carried out directly via contact pointers 1 and 2:
• Slide the scale away
• Loosen the locking screw 3 slightly
• Slide the scale back to the latching point
• Set contact pointers 1 and 2 to the desired switching point
After setting has been carried out: Fix the contact pointers with the locking screw 3.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
25
4 ELECTRICAL CONNECTIONS
Switch contact definition
1 MIN contact
2 MAX contact
3 Pointer vane with switching vane
If the pointer vane enters the slot, an alarm is triggered. If the pointer vane lies outside the slot
sensor, a wire break also causes the alarm to be triggered.
The 3-wire limit switch does not have any wire break detection.
Definition MinMin - MaxMax
H250
1 MIN 2 contact or MAX 1 contact
2 MIN 1 contact or MAX 2 contact
Current consumption in the position shown:
ContactTypeCurrent
MIN 1NAMUR≤ 1 mA
MIN 2NAMUR≤ 1 mA
MAX 1NAMUR≥ 3 mA
MAX 2NAMUR≥ 3 mA
26
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.3.2 Indicator M9 - current output ESK2A
The connecting terminals of the ESK2A have a pluggable design and can be removed in order to
connect the cables.
1 ESK2A current transmitter
2 Power supply 12...30VDC
3 Measurement signal 4...20 mA
4 External load, HART
®
communication
ELECTRICAL CONNECTIONS 4
Power supply M9 with electrical isolation
The circuitry for connection to other devices such as digital evaluator units or process control
equipment must be designed with especial care. In some circumstances internal connections in
these devices (e.g. GND with PE, ground loops) may lead to impermissible voltage potentials,
which can compromise the function of the device itself or a connected device. In such cases a
protected extra-low voltage (PELV) is recommended.
1 Terminal connection
2 Converter supply isolator with electrical isolation
3 Power supply (see supply isolator information)
4 Measurement signal 4...20mA
5 External load, HART
®
communication
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
27
4 ELECTRICAL CONNECTIONS
Power supply
INFORMATION!
The supply voltage has to be between 12 VDC and 30 VDC. This is based on the total resistance of
the measuring loop. To determine this, add up the resistances of each component in the
measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
where
U
ext.
R
= the total measuring loop resistance is.
L
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
22 mA + 12 V
L
= the minimum supply voltage and
H250
HART® communication
When HART® communication is carried out with the ESK, the analogue measured data
transmission (4...20 mA) is not impaired in any way.
Exception for multidrop mode. In multidrop mode, a maximum of 15 devices with HART
can be operated in parallel, whereby their current outputs are switched inactive
(I approx. 4 mA per device).
®
function
28
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary. The cable shield may only be earthed
(grounded) at one place (on the power supply unit).
communication a load of at least 230 ohm is required.
Configuration
The ESK can be configured via HART® communication. DD (Device Descriptions) for AMS 6.x and
PDM 5.2 as well as a DTM (Device Type Manager) are available for configuration. They can be
downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART
counter can be configured. Two limit values can be monitored. The limit values are assigned
either to flow values or to the counter overflow.
communication. A flow
Self monitoring - Diagnostics
During both start-up and operation, a wide variety of diagnostic functions are performed
cyclically in the ESK2A, in order to guarantee function reliability. When an error is detected, a
failure signal (high) is activated (current > 21 mA) via the analogue output. In addition, more
®
detailed information can be requested via HART
for information and warnings.
(CMD#48). The failure signal is not activated
Diagnostic functions (Monitoring):
• Plausibility of FRAM data
• Plausibility of ROM data
• Working range of internal reference voltages
• Signal detection of the measuring range of the internal sensors
• Temperature compensation of the internal sensors
• Calibration corresponding the application
• Plausibility of counting value
• Plausibility of physical unit, system and selected unit
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
29
4 ELECTRICAL CONNECTIONS
4.3.3 Indicator M9 - Profibus PA (ESK3-PA)
Bus cable
Shielding and grounding
Shielding and grounding
Shielding and groundingShielding and grounding
The statements of the FISCO model only apply if the bus cable used meets the required
specifications. For specifications, see the chapter "Technical data" ESK3-PA.
In order to ensure optimum electromagnetic compatibility of systems it is important that the
system components, and in particular the bus cables, are shielded. These shields must have as
few gaps as possible.
Connection
H250
Figure 4-5: ESK3-PA connection
1 Signal connection
2 Shielding and grounding
Polarity reversal has no effect on the function. The cable shield should be connected with
minimum length to the functional ground FE.
30
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.3.4 Indicator M9 - totalizer (ESK-Z)
The totalizer only works in conjunction with the ESK2A current output. A 6-digit display shows
the totalised flow value. It can be changed over to the instantaneous flow value in 0...100%.
A data backup is carried out automatically in the event of a power failure.
The counter is factory-set to the measuring range of the indicator. The total value can be read
directly.
Supply 11/12 and measured signals S+ and S- are not electrically isolated.
If the measured signal is not needed externally, a short-circuit jumper has to be connected to
terminals S+ and S-.
Pulse outputs P+ and P- are electrically isolated. A pulse is generated for each counter advance.
If the pulse output is not required, its terminals can remain unused.
ELECTRICAL CONNECTIONS 4
Figure 4-6: Counter connection
1 ESK - measurement signal 4...20 mA
2 Counter module
3 Transfer of the measurement signal or short-circuit jumper
4 Pulse output load
5 Counter power supply
6 Pulse output power supply
A functional extra-low voltage with protective electrical isolation (PELV) in accordance with VDE
0100 Part 410 is required as a power supply. All the instruments (recorder, display, etc.)
connected to measuring circuits S+ and S- are connected in series. If this measuring circuit is
not needed, then a short-circuit jumper 3 required.
The RESET button deletes only the actual totalizer value.
H250
Settings by pressing a button at the moment of switch-on
Figure 4-8: Settings of the counter at the moment of switch-on
RESET button - mA calibration
Button 1 - Display test
Button 2 - Changing the conversion factor
Button 3 - Software hardware version (information)
32
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Conversion factor
The conversion factor is always 10% of the full-scale range.
If the measuring range is not know, the conversion factor is factory-set to 1000.
Figure 4-9: Changing the conversion factor
1 Units position
2 Tens position
3 Hundreds and 1000s position
Exit the setting by pressing the RESET button
The largest factor that can be set is 1099.
Factors with decimal values are not possible.
Counter overflow
Figure 4-10: Depiction of counter overflow
A counter overflow is signaled by all the decimal points lighting up.
Reset by pressing the RESET button.
Current input calibration
Current input calibration
Current input calibrationCurrent input calibration
During the switching-on process keep the RESET button pressed until three decimal points light
up.
• Set 4.00 mA
• Keep button 1 pressed until the number 0 is displayed
• Set 20.00 mA
• Keep button 3 pressed until the number 100 is displayed
• Exit calibration by pressing button 2
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
33
4 ELECTRICAL CONNECTIONS
4.4 Electrical connection indicator M10
4.4.1 Indicator M10
The display can be removed after the housing lid has been unscrewed. The connection terminals
have a spring locking system.
Figure 4-11: Indicator M10 terminal connection
1 Power supply - analog output
2 Switching output B1
3 Switching output B2 or pulse output
4 Reset input R
H250
4.4.2 Power supply - current output
The electrical connection is reverse-polarity protected.
M10 - terminal connection I
1 Terminal connection
2 Power supply 16...32VDC
3 Measurement signal 4...20 mA
4 External load, HART
®
communication
34
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
ELECTRICAL CONNECTIONS 4
Power supply M10 with electrical isolation
The circuitry to other devices must be designed with special care. In some circumstances,
internal connections in these devices (e.g. GND with PE, ground loops) may lead to
impermissible voltage potentials, which can compromise the function of the device itself or a
connected device. In such cases a protected extra-low voltage (PELV) is recommended.
1 Terminal connection
2 Converter supply isolator with electrical isolation
3 Power supply (see supply isolator information)
4 Measurement signal 4...20mA
5 External load, HART
®
communication
Power supply
INFORMATION!
The supply voltage has to be between 16 VDC and 32 VDC. This is based on the total resistance of
the measuring loop. To determine this, add up the resistances of each component in the
measuring loop (not including the device).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
22 mA + 16 V
L
where
U
= the minimum supply voltage and
ext.
= the total measuring loop resistance is.
R
L
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
35
4 ELECTRICAL CONNECTIONS
HART® communication
When HART® communication is carried out with the M10, this will not in any way impair analogue
measured data transmission (4...20 mA).
Exception for multidrop operation. In multidrop operation, a maximum of 15 devices with HART
function can be operated in parallel, for which the current outputs are switched to inactive.
Load for HART® communication
INFORMATION!
®
For HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary. The cable shield may only be earthed
(grounded) at one place (on the power supply unit).
communication a load of at least 230 ohm is required.
H250
®
Configuration
The M10 electronic indicator can be configured via HART® communication. DD (Device
Descriptions) for AMS 6.x and PDM 5.2 as well as a DTM (Device Type Manager) are available for
configuration. They can be downloaded free of charge from our website.
®
The current flow rate can be transmitted using the integrated HART
counter can be configured. Two limit values can be monitored. The limit values are assigned
either to flow values or to the counter.
communication. The flow
36
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.4.3 Switching outputs B1 and B2
The switching outputs are electrically isolated from each other and from the current output.
CAUTION!
The switching outputs can only be operating if the power supply is applied to terminals I+ and I-.
Switching outputs B1 and B2 can be electrically connected in two ways:
Switching value reached≤1 mA≤1 mA> 3mAmax. 100 mA
Switching value not reached> 3mAmax. 100 mA≤1 mA≤1 mA
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
37
4 ELECTRICAL CONNECTIONS
Switching capacity of B1 and B2 with PNP technology
Due to the PNP technology and the associated protective elements, there is a voltage drop Uv for
the load to be operated.
Switching capacity of B1 and B2
H250
1 Max. switching current I [mA]
2 Minimum load impedance R
3 Power supply U
ext.
L
Power loss of B1 and B2
1 Load impedance RL 100 Ohm
2 Load impedance R
3 Power loss U
4 Power supply U
d
1000 Ohm
L
ext.
[Ohm]
38
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.4.4 Switching output B2 as a pulse output
INFORMATION!
When switching output B2 is used as a pulse output, two separate signal circuits are required.
Each signal circuit requires its own power supply.
The total resistance 3 must be adapted so that the total current I
Figure 4-12: Electrical pulse output
1 Terminal B2
2 Terminal I
3 Load e.g. counter
4 Flow rate measurement 4...20 mA
5 Pulse output power supply
6 M10 power supply
ELECTRICAL CONNECTIONS 4
does not exceed 100 mA.
tot
Pulse output B2 is a passive "open collector" output which is electrically isolated from the
current output and output B1. It can be operated as a low-resistance output or as a NAMUR
output.
Figure 4-13: Data pulse output
1 f
= 10 Hz
max
2 t
on
3 t
off
The pulse width t
can be configured from 30...500 ms in the menu of the indicator.
on
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
39
4 ELECTRICAL CONNECTIONS
4.4.5 Connection reset input R
Input R can be used as a reset input for the internal counter.
Figure 4-14: Indicator M10 - reset input
1 Function active HI
2 Function active LO
H250
This reset input can be activated in the menu of indicator M10, and can be configured to ACTIVE
HI or ACTIVE LO. See also chapter "Indicator M10 menu explanations".
If the input is set as ACTIVE LO, an interruption causes the counter to be reset.
4.5 Grounding connections
Figure 4-15: Ground connections
1 Indicator M8
2 Indicator M9
3 Indicator M10
40
DANGER!
The grounding wire may not transfer any interference voltage.
Do not use this grounding wire to ground any other items of electrical equipment.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
4.6 Protection category
The measuring device meets all requirements of protection category IP
IndicatorProtection category
M9IP65/67
M8IP65
M10IP66/67
DANGER!
After all servicing and maintenance work on the device, the specified protection category has to
be ensured again.
Therefore it is essential to observe the following points:
• Use only original gaskets. They must be clean and free of any damage. Defective gaskets have
to be replaced.
• The electrical cables used must be undamaged and must comply with regulations.
• The cables must be laid with a loop 3 upstream of the measuring device to prevent water
from getting into the housing.
• The cable glands 2 must be tightened.
• Close the unused cable glands using blanking plugs 1.
ELECTRICAL CONNECTIONS 4
Figure 4-16: Cable gland
1 Use blanking plugs if no cable is routed through
2 Tighten cable gland firmly
3 Lay the cable in a loop
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
41
5 START-UP
5.1 Standard device
CAUTION!
When starting up the device, the following points must be observed:
•
Compare the actual operating pressure and the product temperature of the system with the
specifications on the nameplate (PS and TS). These specifications may not be exceeded.
•
Make sure materials are compatible.
•
Slowly open the shut-off valve.
•
When measuring liquids, vent the pipes carefully.
•
When measuring gases, increase pressure slowly.
•
Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring
unit or float.
A minimum operating pressure (primary pressure) is necessary to operate the device:
MediumPressure loss : operating pressure
H250
Liquids1 : 2
Gases without float damping1 : 5
Gases with float damping1 : 2
5.2 Indicator M10
INFORMATION!
The device is always preset for the user and his application.
Start
After the device is switched on, the display shows the following in sequence
• "Test",
• the device type and
• the version number.
Afterwards the device performs a self-test and switches to measurement mode. Here all of the
parameters preset for the customer are analysed and checked for plausibility, and the current
measured value is displayed.
42
Operation
INFORMATION!
The device is low-maintenance
Comply with the application limits with regard to temperature of the medium and ambient
temperature.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
6.1 Operating elements
OPERATION 6
Operation of the device is performed with the cover on the front open, using the mechanical keys
or with the cover closed using a bar magnet
bar magnet.
bar magnetbar magnet
keys,
keyskeys
CAUTION!
The switching point of the magnetic sensors is right at the level of the corresponding circle. Only
touch the circle vertically and from the front using the bar magnet. Touching it from the side may
cause a malfunction.
Figure 6-1: Display and operating elements
1 Enter button (circuit for bar magnet)
2 Up button (circuit for bar magnet)
3 Right button (circuit for bar magnet)
4 Display
The mechanical keys and keys for the bar magnet are identical in their function. In this
documentation the keys are represented as symbols to describe the operating functions:
KeySymbol
1Enter^
2up↑
3right→
Table 6-1: M10 operation keys
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
43
6 OPERATION
6.2 Basic principles of operation
6.2.1 Functional description of the keys
H250
→
↑
^
Table 6-2: Description of the opperation buttons
Switch from measurement mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
In change mode:
In change mode: Move the input cursor one position to the right; after the last digit the
In change mode:In change mode:
input cursor jumps back to the beginning.
In measuring mode:
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
Change between the menu items within a menu level
In change mode:
In change mode: Changing parameters and settings; running through the available
In change mode:In change mode:
characters; shifting the decimal point to the right.
Change one level up at the menu
Return to measuring mode with a query whether the data should be accepted
6.2.2 Navigation within the menu structure
Navigation within the menu is by means of the → and ^ buttons. Pressing button → takes you
one menu level lower, ^ takes you one menu level higher.
If you are already located at the lowest level (function level), you can use the button → to go the
the change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^ key to exit the menu mode and
return to the measuring mode.
Measuring
operation
Table 6-3: Navigation menu structure
→Main menu↑→Sub-menu↑→Function↑→Edit
^^^^^^^^^^^^^^^^
→ ↑ ^^^^
44
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
6.2.3 Changing the settings in the menu
Starting operation
Operation is started using the key →.
If a different key is pressed, it is necessary to wait 5 seconds before activating the key →.
If an operation inhibit has been set, the code → → → ^^^ ↑ ↑ ↑ has to be entered. If no key
is pressed within 5 seconds, code input is exited.
Exiting operator input
Operation is exited by pressing the ^ key several times.
If data have been changed:
OPERATION 6
Save Yes→ Changes are accepted. An update is carried out and the indicator jumps back to
Save No^ Changes are discarded and the indicator jumps back to measuring operation.
measuring operation.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal
plausibility check.
If implausible inputs have been made, the indicator remains in the current menu, and the
changes are not accepted.
Example: Changing the default parameter from m
3
/h to l/h
DisplayDisplay
Example:7.2
m
1x →Fct. 1111.0
OPERATION
2x ↑Fct. 3333.0
INSTALLATION
1x →Fct 3.1111
LANGUAGE
12x ↑Fct 3.13
END&UNIT
3
/h
13
1313
1x →Fct. 3.13.1111
1x →10,0000
6x ↑10000
1x ^^^^Quit Yes
3x ^^^^7200
FLOW RATE
3
/h
m
l/h
L/h
6.2.4 Measures in the event of faulty indications
If the indications on the display or the responses to keypad commands are faulty, you have to do
a hardware reset. Switch the power supply OFF and ON again.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
45
6 OPERATION
6.3 Overview of the most important functions and indicators
INFORMATION!
For a complete list of all functions and short descriptions refer to Menu explanations on page
. All default parameters and settings are adapted for the specific customer.
51
LevelDesignationExplanation
1.4TIME CONST.Time constant, damping value [s]
1.5.2ERRORError indicator
Yes: Error messages are deleted
No: Error messages are suppressed.
2.14-20mA OUTCheck current output
2.2 -2.4OUTPUT BCheck switching outputs and RESET input
3.1LANGUAGESelect the menu language
3.13.1FLOW RATEMaximum flow rate
Table 6-4: The most important functions
The value set is represented by a 20 mA analogue current output.
If the current value exceeds the preset value, an alarm is indicated.
H250
M10 flow units
M10 flow units
M10 flow unitsM10 flow units
Measured
variables
Volume
Massg/sg/ming/h-Liquids,
UnitsMeasured
products
m3/sm3/minm3/hm3/d
L/sL/minL/h-
ft3/sft3/minft3/hft3/d
gal/sgal/mingal/hgal/d
bb/sbb/minbb/hbb/d
ImpGal/sImpGal/minImpGal/hImpGal/d
kg/skg/minkg/hkg/d
-t/mint/ht/d
lb/slb/minlb/h-
-short t/minshort t/hshort t/d
--long t/hlong t/d
Liquids,
vapours, gases
vapours, gases
46
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
6.4 Error messages
Error messageDescriptionCategoryRemedy
OPERATION 6
NOT LINEARIZEDLinearization faulty or
NEW LINEARI.
TABLE BAD
LINEARIZATIO
UNDER CONFIG
UNIT SYSTEM
CONFLICT
TOO FEW
ENTRIES
NOT
MONOTONOUS
FIRST NOT 0 %The first flow value if
LAST NOT 100 %The last flow value if
NO ZERO CAL OF AOThe current output
NO F.SC. CAL OF AOThe current output
NO TEMP.
COMPENSATION
OUTPUT NOT
LINEARIZED
COUNTER LOSTTotalizer value was
not activated =
measuring error
Faulty or missing
data in the
linearization table =
measuring error
The device is in
linearization mode =
measuring error
The unit for the
linearization flow is
incompatible with the
selected flow type
(mass/volume)
The linearization
table has too few data
points
The sequence of the
linearization values is
not strictly monotonic
increasing
the linearization table
is not 0%
the linearization table
is not 100%
zero point 4.00 mA is
not calibrated = poss.
measuring error in
the process control
system
100% = 20.00 mA is
not calibrated = poss.
measuring error in
the process control
system
The sensor
temperature
compensation of the
device is faulty or was
not carried out. =
possible measuring
error
Linearization is not
activated = measuring
error
reset by
error/overflow
ErrorActivate linearization or carry it out again
(HART® communication and linearization
software are required; the original calibration
values must be known), or send the device
back to the manufacturer for linearization.
ErrorComplete the linearization and activate it
(HART® communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
ErrorCorrect error, carry out linearization again if
necessary (HART® communication and
linearization software are required), or send
the device back to the manufacturer for
linearization.
ErrorCarry out linearization at at least 5 points
(HART™ communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
ErrorCheck linearization and/or carry it out again
(HART® communication and linearization
software are required), or send the device
back to the manufacturer for linearization.
WarningPerform calibration using ammeter and menu
3.10 or using standard HART® tools/process
control system and poss. external ammeter.
Caution: during the calibration, switch the
measuring point to manual control.
WarningPerform calibration using ammeter and menu
item 3.11 or using standard HART® tools and
external ammeter if necessary. Caution:
during calibration, switch the measuring point
to manual control.
ErrorThe device, together with an indication of the
error, must be sent back to the manufacturer
for checking.
ErrorActivate linearization or carry it out again
(HART® communication and linearization
software are required; the original calibration
values must be known), or send the device
back to the manufacturer for linearization.
WarningBecause the reset time is not known:
Controlled reset of the counter using menu
item 1.5.1 or using HART® tools/process
control system.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
47
6 OPERATION
Error messageDescriptionCategoryRemedy
H250
FRAM WRITE
FAULT
ROM/FLASH
ERROR
RESTART OF
DEVICE
MULTIDROP
MODE
CRYSTAL OSC
FAULT
REF VOLTAGE
FAULT
SENSOR A FAULT Internal error in
SENSOR B
FAULT
MEMORY
CORRUPTION
AO FIXEDThe current output is
AO SATURATEDCurrent output
Internal
communication error
Memory error
detected during selftest.
A device restart has
taken place
The HART® multidrop
mode is activated.
The current output is
set to a fixed value of
4.5 mA.
Internal error in
device
Internal error in
device
device
Internal error in
device
Internal memory
error, caused by a
hardware or software
problem
set to a fixed value.
saturated
ErrorCheck whether the display is plugged in
correctly and restart the device. If the error
occurs again: send the device back to the
manufacturer with an indication of the error.
ErrorRestart the device. If the error occurs again:
send the device back to the manufacturer with
an indication of the error.
InformationThe device has been restarted using menu
item 1.5.2 since the last time the error
messages were reset.
InformationThe HART®- multidrop mode is activated by
selecting a polling address not equal to 0
using menu item 3.9. A polling address of 0
reactivates the current output.
ErrorThe device must be sent back to the
manufacturer with an indication of the error.
ErrorRestart the device; if the error occurs again
the device must be sent back to the
manufacturer with an indication of the error.
InformationThe current output is fixed and does not
reflect the measured value. This is the case in
multidrop mode, with current output
test/calibration using the menu or HART®
InformationThe current output is saturated at 20.4 or 22.0
mA (depending on whether the alarm current
is activated or deactivated in menu item 3.12),
and is no longer coupled with the measured
value.
48
Device drivers for HART® tools, process control equipment (e.g. Siemens PDM or AMS)
TM
PACTware
and HART® DTMs are available on the internet site.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
6.5 Menu indicator M10
6.5.1 Factory settings
MenuFunctionSetting
1.1.1Switching value B10.0
1.1.2Hysteresis B10.0
1.2.1Switching value B20.0
1.2.2Hysteresis B20.0
1.3DisplayFlow rate
1.4Time constant3 s
1.5.1Reset counterNO
1.5.2Reset errorNO
3.1LanguageDEUTSCH
3.2Function B1INACTIVE
3.3Contact B1NC contact
3.4Function B2INACTIVE
3.5Contact B2NC contact
3.6Pulse duration100ms
3.7Pulse / unit001 / liter
3.8Function B3INACTIVE
3.9Multidrop polling address0
3.12Alarm currentOFF
3.13.1Flow unitsee rating plate
3.13.2Counter unitDerived from the flow unit
3.14LFC4% ON 6% OFF
3.15Input codeNO
OPERATION 6
INFORMATION!
The device has been set at the factory in accordance with the customer order.
Therefore subsequent configuration via the menu is only necessary if the intended use of the
device is changed.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
49
6 OPERATION
6.5.2 Menu structure
MenuSubmenu 1Submenu 2
1111 Operation1.1 Output B11.1.1 Switching value B1
2222 Test & Info2.1 Output 4...20mA
3333 Installation3.1 Language
H250
1.1.2 Hysteresis B1
1.2 Output B21.2.1 Switching value B2
1.2.2 Hysteresis B2
1.3 Display
1.4 Time constant
1.5 Reset1.5.1 Reset counter
1.5.2 Reset error
2.2 Output B1
2.3 Output B2
2.4 Input B3
2.5 Serial no.
2.6 Software version
2.7 Tag no.
3.2 Function B1
3.3 Contact B1
3.4 Function B2
3.5 Contact B2
3.6 Pulse duration
3.7 Pulse/unit
3.8 Function B3
3.9 Multidrop
3.10 Calibration 4mA
3.11 Calibration 20mA
3.12 Alarm current
3.13 Upper range value and unit3.13.1 Flow rate
3.13.2 Counter
3.14 Low Flow Cutoff LFC3.14.1 Control
3.14.2 Switch-on value
3.14.3 Switch-off value
3.15 Input code
3.16 Basic setting
50
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
6.5.3 Menu explanations
OPERATION 6
LevelDesignationSelection /
Explanation
Input
1.1.1OUTPUT B1INACTIVE
FLOW.VAL B1Flow value switching point.
COUNTER.VAL B1Counter value switching point.
1.1.2OUTPUT B1HYST.B1Hysteresis setting for the flow value switching point.
The switching point is entered in flow units.
If the current flow value exceeds this set switching point, then output B1
is activated.
Note:
The function NC or NO can be selected using menu 3.3.
Each positive number can be set here.
If the counter exceeds this value, then output B1 is activated.
Note:
The function NC or NO can be selected using menu 3.3.
Value range 0...switching point.
Example, if a switching value of 200 is set under 1.1.1, then a hysteresis
value of 0...200 can be set here.
If the value 0 is entered here, then this output does not have hysteresis.
If the value 20 is entered here, then the output functions as follows:
If the current flow value exceeds the value 200, then the output switches
3.
If the current flow value drops below the hysteresis value of 180, then the
switching output returns to its normal state 4.
Note:
To invert the operating method, use menu 3.3 to set the output from NO
1 to NC 2 or vice versa.
This function is not activated on the counter switching point.
1.2.1OUTPUT B2INACTIVE
FLOW.VAL B2see FLOW.VAL B1
COUNTER.VAL B2see COUNTER.VAL B1
PUL. VAL B2B2 = pulse output
Note:
Settings under menu 3.6 pulse duration and
3.7 Pulses/unit
1.2.2OUTPUT B2HYST.B2See HYST. B1
1.3DISPLAYFLOW RATE
COUNTER
FLOW&COUNT
PERCENT
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
51
6 OPERATION
H250
LevelDesignationSelection /
Explanation
Input
1.4TIME CONST.Setting : 1 ... 20 seconds
Note:
The settable time constant affects the current output and the displayed
current flow rate.
It thus enables damped depiction if there is a highly variable flow rate.
If the current flow rate is polled via HART
transferred measured value is dependent on the time constant here, too.
1.5.1RESETCOUNTERYES - NO
1.5.2RESETERRORYES - NO
2.14-20mA OUTThe analogue current output can be set to fixed values in 10% steps
between 4.00...20.00 mA.
This function has no influence on binary switching outputs.
Note:
This test function is switched off in multidrop mode.
Display: "NOT AVAILABLE".
2.2OUTPUT B1OPENThe function assignment in menu 3.2 is not taken into consideration here.
CLOSED
2.3OUTPUT B2OPENThe function assignment in menu 3.3 is not taken into consideration here.
CLOSED
2.4INPUT B3Here there is a visual depiction of whether or not input B3 has a voltage of
5...30 V.
If input B3 is set to ACTIVE HI in menu 3.8, then the display shows "ON"
when the switching voltage is applied.
®
communication, then the
Note:
NO test function possible when the output is set to INACTIVE in menu 3.8.
3.1LANGUAGEENGLISH
DEUTSCH
FRANCAIS
ITALIANO
ESPANOL
CESKY
POLSKI
NEDERLANDS
3.2FUNCTION B1INACTIVEOutput B1 is switched off.
SWITCHING
POINT
COUNTER_LIMOutput B1 switches when the counter exceeds the counter limit value.
3.3CONTACT B1NC contactOutput B1 is normally closed. If an alarm situation occurs, the contact
NO CONTACTOutput B1 is normally open. If an alarm situation occurs, the contact
3.4FUNCTION B2INACTIVESee FUNCTION B1
SWITCHING
POINT
COUNTER_LIMSee FUNCTION B1
PULSE OUTPUTOutput B2 generates pulses up to 10 Hz depending on the current flow
Output B1 switches at a set value depending on the current flow value.
opens.
closes.
See FUNCTION B1
value.
52
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
OPERATION 6
LevelDesignationSelection /
Explanation
Input
3.5CONTACT B2NC contactSee CONTACT B1
NO CONTACTSee CONTACT B1
3.6PULSE
DURATION
3.7PULSE/UNIT0.000001 Smallest scaling factor
3.8FUNCTION B3INACTIVE
3.9MULTIDROP0…15Multidrop mode means that the device works continuously in bus mode
3.104mA CALIBR.This menu item allows precise calibration of the current output.
3.1120mA CALIBR.This menu item allows precise calibration of the current output.
3.12ALARM
CURRENT
3.13END & UNITThe flow unit and the upper range value can be changed.
30 ms
50 ms
100ms
200 ms
500 ms
Note:
In the basic setting, the unit of the pulse output corresponds to the flow
unit.
Example: volume flow unit is m
3
or
m
3
/h, so the pulse output is set to pulses /
mass flow unit is kg/h, so the pulse output is set to pulses / kg
999999.0Largest scaling factor
ACTIVE HIThe internal counter is reset to zero when a positive voltage of from
5...30 VDC is applied to terminals R+ and R for at least 100 ms.
ACTIVE LOThe internal counter is reset to zero when a positive voltage of from 5...30
VDC applied to terminals R+ and R is interrupted for at least 100ms..
®
via HART
The analogue current output is then set to a fixed value of 4.1 mA.
Measured values are transferred via HART
communication (max. 15 devices in parallel).
®
communication.
However, the display allows local read-off of the measured values.
The polling address can be set to 1...15.
Larger integer values are not permitted. If the polling address is set to 0,
then HART
®
bus mode is switched off.
The device is working in analogue mode. Current output 4...20 mA is
active.
Standard HART
®
communication remains guaranteed.
The device generates a fixed current output of 4.00 mA.
If the measured value deviates from the displayed one, then the
measured value must be input.
When the menu is exited, the corrected value is saved.
The device generates a fixed current output of 20.00 mA.
If the measured value deviates from the displayed one, then the
measured value must be input.
When the menu is exited, the corrected value is saved.
OFFMeasured values > 100% are indicated as a current signal up to a
maximum of 22 mA.
ONIn the event of an error the current output is set to the fixed value of
22m A.
Note:
Changing from volume flow measurement to mass flow measurement is
only possible with a new calibration.
3.13.1FLOW RATEFor a units list, see chapter 7.4 of the handbook
3.13.2COUNTERAs standard, the unit for the counter is derived from the unit for the flow
measurement.
It can also be changed individually.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
53
6 OPERATION
H250
LevelDesignationSelection /
Explanation
Input
3.14LFCLFC stands for Low Flow Cutoff.
3.14.1CONTROLINACTIVELFC is switched off
ACTIVELFC is switched on
3.14.2LFC
ON_VALUE
3.14.3LFC
OFF_VALUE
1…19 %Switch-on value 1:
2…20 %Switch-off value 2:
With variable area flowmeters, the flow range from 0 to 10% is not
defined.
In order to ensure a stable zero point of the analogue output, the
analogue output can be set to a stable value of 4.00mA in a selectable
range from 0 to 20%.
The flow is greater than 20%.
The current output corresponds to this.
If the flow rate falls, then the current output follows it until the ON value.
If the flow value continues to fall, the current output is switched to
4.00 mA 3.
The flow rate is 0. The current output is 4.00 mA 3.
If the flow rate rises, the current output remains at 4.00 mA 3 until the
OFF value, and is switched to the corresponding flow value if the flow
value rises further.
3.15INP. CODEYESThe input code is used to prevent authorised adjustment of the
NO
3.16BASIC
SETTING
YESThis menu item can be used to select the calibrated basic setting.
NO
measurement parameters.
The input code is not active by default.
If YES is selected, the last code entered must be typed in.
Factory code: → → → ^^^ ↑ ↑ ↑
If, after confirmation with YES, the button → is also pressed, then a new,
individual, nine-element code can be typed in.
The display shows the required key combination.
This can be helpful if operating data have been changed a number of
times.
This menu item cannot be used to reset the calibration.
54
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also
be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to
the measuring tube and indicator.
We advise that inspections be carried out at least once per year.
The device must be removed from the piping before cleaning.
CAUTION!
Pressurized pipes must be depressurized before removing the device.
Empty pipes as completely as possible.
In the case of devices used for measuring aggressive or hazardous media, appropriate safety
precautions must be taken with regard to residual liquids in the measuring unit.
Always use new gaskets when reinstalling the device in the pipeline.
Avoid electrostatic charges when cleaning the surfaces (e.g. sight window)!
7.2 Replacement and retrofitting
SERVICE 7
Some of the variable area flowmeter components can be retrofitted:
• Float damping
Indicator M9:
Indicator M9:
Indicator M9:Indicator M9:
• Eddy current brake
• Limit switch unit
• Current output ESK2A
• Counter module
The ESK3-PA Profibus can only be retrofitted following recalibration.
7.2.1 Replacing floats
• Remove the device from the piping.
• Take the upper snap ring out of the measuring unit.
• Take the upper float stop and float out of the measuring unit.
• Insert the new float into the centre hole of the lower float stop and push into the measuring
unit along with the upper float catcher. While doing this, the float's upper guide rod must be
guided through the middle hole of the float stop.
• Insert the snap ring into the measuring unit.
• Fit the device back into the piping.
CAUTION!
An additional measuring error is to be expected if recalibration is not conducted.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
55
7 SERVICE
7.2.2 Retrofitting float damping
• Remove the device from the piping.
• Take the upper snap ring 1 out of the measuring unit.
• Take the upper float catcher 2 and float 5 out of the measuring unit.
• Fasten the snap ring 3 into the lower slot of the float's guide rod.
• Slide ceramic sleeve 4 on to the float's guide rod and attach it to the top slot using the span
ring 3.
• Insert float into the lower float guide in the measuring unit.
• Retrofit the supplied damping cylinder with the integrated float catcher 2 into the measuring
unit.
• Insert upper snap ring 1.
H250
1 Snap ring
2 Float catcher
3 Spring washer
4 Ceramic sleeve
5 Float
7.2.3 Retrofitting pointer damping
When retrofitting the pointer damping for indicator M9 with ESK2A current output and limit
switches, note that the pointer may move briefly when installing the pointer damping (eddy
current brake), which may trigger an error alarm or may change the current output by showing
peaks.
The Eddy current brake consists of two parts:
56
1 Turnbuckle
2 Eddy current brake
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
SERVICE 7
The brake with the retainer ring can be clipped onto the pointer cylinder independently of the
built-in components (ESK2A, limit switch, counter). When installing the brake, note that the slit
between the brake magnets is only about 3 mm and the material thickness of the aluminium
pointer vane is 1mm.
• Clip on Eddy current brake 1.
• Turn brake slightly in a clockwise direction 2.
• Check that the pointer vane can be moved between the magnets without touching them.
• Screw in turnbuckle 2.
7.2.4 Retrofitting limit switch
• Remove counter module (if available).
• Loosen the locking screw 2 on the contact pointer.
• Merge contact pointer 1 in the middle.
• Insert the contact module into the third slot of the bracket until the semi-circle 3 surrounds
the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in
order to connect the cables.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
57
7 SERVICE
7.2.5 Replacement - Retrofitting ESK2A
When replacing or retrofitting an ESK2A, the following are required at the time of
ordering:
• SN - serial number or
• SO - sales order
This information can be found on the indicator nameplate
INFORMATION!
The ESK2A is factory calibrated, making it possible to replace it or retrofit it without
recalibrating.
• De-energise the ESK2A.
• Lift and remove the ESK2A with a screwdriver.
H250
Plug-in technology is used to install the ESK2A.
• The ESK2A plug-in tongues 1 are plugged in under the two bolts 2 on the baseplate.
• Slight pressure is used to press the ESK2A onto the spring pins 3 until it stops, firmly
attaching the ESK2A.
If a change in measuring range, product temperature, product, density, viscosity or pressure is
desired, this can be done with the KroVaCal program or with a HART
measuring unit is subject to its own physical limits, which the DroVaCal program correctly
calculates, and may thus reject the desired change. If a change is performed using the program,
the new data is also transmitted to the ESK2A.
TM
modem. However, each
Program features and possibilities
• Device identification
• Device address
• Serial number
• Measuring point designation
• Digital measured value query in flow units, % and mA
• Test / setting functions
• Calibration 4.00 and 20.00 mA
• Set current output to any desired value
58
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
7.2.6 Totalizer
The flow counter, in conjunction with the ESK2A electrical current output, can also be
retroactively built into the indicator M9.
When ordering the ESK-Z totalizer as a retrofit kit, please indicate the device information (see
scale) and the measuring range.
Providing this information will ensure that the new scale supplied with the counter display cutout
is ready for installation.
The flow counter is then preset using the conversion factor corresponding to the measuring
range.
SERVICE 7
Installation
• Push the existing scale out.
• Insert the flow counter unit into the centre track of the module carrier.
• Insert the new scale into the module carrier.
• In so doing, lift the scale slightly until the scale cutout surrounds the counter display.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
59
7 SERVICE
7.3 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
7.3.1 List of spare parts
Spare partOrder no.
DN 15
DN 15
DN 15DN 15
Float CIV 15, 1.4404X251041000
Float DIV 15, 1.4404X251042000
Float TIV 15, 1.4404X251043000
Float DIVT 15, 1.4404X251044000
Float TIV 15, AluminiumX251043100
Float TIV 15, TitaniumX251043200
Set float stop; standard (1 stop, 1 span ring)X251050100
Set float stop; gas damping (ZrO2)X251050200
Set float stop; gas damping (PEEK)X251050300
Damping bush (7x8) ZrO2 incl. 2 span ringsX251053100
Damping bush (7x8) PEEK incl. 2 span ringsX251053200
DN 25
DN 25
DN 25DN 25
Float CIV 25, 1.4404X252041000
Float DIV 25, 1.4404X252042000
Float TIV 25, 1.4404X252043000
Float DIVT 25, 1.4404X252044000
Set float stop; standard (1 stop, 1 span ring)X252050100
Set float stop; gas damping (ZrO2)X252050200
Set float stop; gas damping (PEEK)X252050300
Damping bush (12x8) ZrO2 incl. 2 span ringsX252053100
Damping bush (12x8) PEEK incl. 2 span ringsX252053200
DN 50
DN 50
DN 50DN 50
Float CIV 55, 1.4404X253041000
Float DIV 55, 1.4404X253042000
Float TIV55, 1.4404X253043000
Float DIVT 55, 1.4404X253044000
Set float stop; standard (1 stop, 1 span ring)X253050100
Set float stop; gas damping (ZrO2)X253050200
Set float stop; gas damping (PEEK)X253050300
Damping bush (14x10) ZrO2 incl. 2 span ringsX253053100
Damping bush (14x10) PEEK incl. 2 span ringsX253053200
H250
60
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
SERVICE 7
DN 80
DN 80
DN 80DN 80
Float CIV 85, 1.4404X254041000
Float DIV 85, 1.4404X254042000
Float TIV 85, 1.4404X254043000
Float DIVT 85, 1.4404X254044000
Set float stop; standard (1 stop, 1 span ring)X254050100
Set float stop; gas damping (ZrO2)X254050200
Set float stop; gas damping (PEEK)X254050300
Damping bush (18x14) ZrO2 incl. 2 span ringsX254053100
Damping bush (18x14) PEEK incl. 2 span ringsX254053200
DN 100
DN 100
DN 100DN 100
Float CIV 105, 1.4404X255041000
Float DIV 105, 1.4404X255042000
Float DIVT 105, 1.4404X255044000
Set float stop; standard (1 stop, 1 span ring) only for bottom !X255050100
Set float stop; gas damping (ZrO2)X255050200
Set float stop; gas damping (PEEK)X255050300
Damping bush (18x14) ZrO2 incl. 2 span ringsX254053100
Damping bush (18x14) PEEK incl. 2 span ringsX254053200
Indicator M9
Indicator M9
Indicator M9Indicator M9
Indicator housing complete no scaleX251010000
Indicator complete stainless steel unpainted, no scaleX251011000
Cover M9 complete, standard (blue; RAL 5015)X251010100
Cover M9 complete, salt water resistant (grey; RAL 7001)X251010200
Cover M9 complete, no silicone (blue; RAL, 5015)X251010300
Printed scale with counter cutout (serial number required)On request
Blank scale with counter cutoutX251023400
Other spare parts on request
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
61
7 SERVICE
7.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local representative.
7.5 Returning the device to the manufacturer
7.5.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
H250
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.
62
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
7.5.2 Form (for copying) to accompany a returned device
Company:Address:
Department:Name:
Tel. no.:Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
SERVICE 7
This medium is:water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such
substances.
We have flushed out and neutralized all cavities in the
device.
We hereby confirm that there is no risk to persons or the environment through any residual media
contained in the device when it is returned.
Date:Signature:
Stamp:
7.6 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
63
8 TECHNICAL DATA
8.1 Operating principle
The flowmeter H250 operates on the float measuring principle. The measuring unit consists of a
metal cone in which a float can move freely up and down. The medium flows through the
flowmeter from bottom to top. The float adjusts itself so that the buoyancy force B, acting on it,
the form drag D and its weight W are in equilibrium: W = B + D.
Figure 8-1: Operating principle
1 Indication principle M9 and M8MG
2 Indication principle M10 and M8EG
H250
For indicators M9 and M8MG 1 the flow-dependent height of the float in the measuring unit is
transmitted by means of a magnetic coupling and displayed on a scale. For indicators M10 and
M8EG 2 the flow-dependent height of the float in the measuring unit is transmitted to the
electronic display by magnetic field sensors S1 and S2.
Operating principle of H250H and H250U
Figure 8-2: Operating principle H250H and H250U
1 H250H - horizontal flow direction
2 H250U - flow direction from top to bottom
The flowmeters operate according to a modified float measuring principle. The guided float
adjusts itself so that the flow force acting on it is in equilibrium with the opposing spring force.
The flow-dependent position of the float in the measuring unit is displayed on a scale by means
of a magnetic coupling.
64
INFORMATION!
Flowmeters H250H and H250U only work in conjunction with indicator M9.
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Application rangeFlow measurement of liquids, gases and vapors
Operating method / measuring principleVariable area measuring principle
Measured value
Primary measured valueFloat position
Secondary measured valueOperating and standard volumetric flow
TemperatureTemperature
Max. operating temperature TS-196..+300°C / -321...+572°F
Pressure
Pressure
PressurePressure
Max. operating pressure PSDepending on the version up to 400 bar / 5802 psig
Max. test pressure PTPressure equipment directive 97/23/EC or AD 2000-HP30
Min. required operating pressure2 times greater than pressure loss (see measuring ranges)
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:Float damping during gas measurement recommended:
DN15...25 / ½"...1"Operating pressure <0.3 bar / 4.4 psig
DN50...100 / 2"...4"Operating pressure <0.2 bar / 2.9 psig
Installation conditions
Inlet run≥ 5 x DN
Outlet run≥ 3 x DN
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
65
8 TECHNICAL DATA
Materials
H250
DeviceFlange / raised face Measur.
tube
H250/RR
Stainless Steel
H250/HC
Hastelloy
H250/C
Ceramic/PTFE
H250/F - FoodCrNi-Stahl 1.4435-
1 CrNi steel 1.4571 on request, for clamp connection CrNi steel 1.4435
2 TFM/PTFE liner (electrically non-conductive)
®
CrNi steel 1.4404
massive 1
CrNi steel 1.4571 with
plated Hastelloy
(2.4610) 1
CrNi-Stahl 1.4571
with TFM/PTFE liner 2
®
C4
CrNi steel 1.4404 1-
Hastelloy® C4 (2.4610)
FloatFloat stop /
guide
PTFE or Al2O3
with FFKM
gasket
Al2O
3
and PTFE
INFORMATION!
H250/C - DN100 / 4" only PTFE
H250/F: wetted surfaces Ra ≤0.8 μm, optional ≤0.6 μm
Other options:
• Special materials on request: e.g. SMO 254, titanium, 1.4435
• Float damping: ceramic or PEEK
• Gasket for devices with female thread as insert: O-ring FPM / FKM
Ring orifice
-
Al2O
3
Temperatures
Temperatures
TemperaturesTemperatures
DANGER!
For devices to be used in hazardous areas, special temperature ranges apply. These can be
found in the separate instructions.
Temperatures H250/M9 - mechanical indicator without power supply
1 P >0.5 bar
2 with TR float
3 300 mbar with damping (gas measurement)
80
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
1 P >7.4 psig
2 with TR float
3 4.4 psig with damping (gas measurement)
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
81
8 TECHNICAL DATA
H250/C - Ceramic/PTFE
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250/C - Ceramic/PTFE
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
83
8 TECHNICAL DATA
H250H - Horizontal installation position
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
H250
ENASMEConeWater [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15½K 15.1701.8195
K 15.21203204
K 15.31804.5195
K 15.42807.5225
K 15.545012250
K 15.670018325
K 15.7120030590
K 15.8160040950
K 15.82400601600
DN251"K 25.1130035122
K 25.2200050105
K 25.3300080116
K 25.45000130145
K 25.58500220217
K 25.510000260336
DN502"K 55.110000260240
K 55.216000420230
K 55.322000580220
K 55.334000900420
DN803"K 85.125000650130
K 85.235000950130
K 85.2600001600290
DN1004"K 105.1800002200250
K 105.11200003200340
84
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250H - Horizontal installation position
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
ENASMEConeWater [GPH]Air [SCFM]Pressure loss
[psig]
DN151/2"K 15.118.51.122.87
K 15.231.71.863.00
K 15.347.62.792.87
K 15.474.04.653.31
K 15.51197.443.68
K 15.618511.24.78
K 15.731718.68.68
K 15.842324.814.0
K 15.863437.223.5
DN251"K 25.134321.71.79
K 25.252831.01.54
K 25.379349.61.71
K 25.4132180.62.13
K 25.522451363.19
K 25.526421614.94
DN502"K 55.126421613.53
K 55.242272603.38
K 55.358123603.23
K 55.389825586.17
DN803"K 85.166044031.91
K 85.292465891.91
K 85.2158519924.26
DN1004"K 105.12113413643.68
K 105.13170119845.00
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
85
8 TECHNICAL DATA
H250U - Vertical installation position
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow directionfrom top to bottom
H250
ENASMEConeWater [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15½"K 15.1651.6175
K 15.21102.5178
K 15.31704180
K 15.42606200
K 15.542010220
K 15.665016290
K 15.7110028520
K 15.8150040840
DN251"K 25.111503097
K 25.218004585
K 25.327007092
K 25.44500120115
K 25.57600200172
DN502"K 55.19000240220
K 55.215000400230
K 55.321000550240
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
86
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
3
Nl/h or Nm
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
www.krohne.com04/2011 - 4000269302 MA H250-R02 en
H250
TECHNICAL DATA 8
H250U - Vertical installation position
Measuring span:10 : 1
Declaration of flow:Values = 100%Water: 20°C [68°F]Air: 20°C [68°F], 1.013 bar abs. [14.7 psia]
Flow directionfrom top to bottom
ENASMEConeWater [GPH]Air [SCFM]Pressure loss
[psig]
DN15½"K 15.117.20.992.57
K 15.229.11.552.62
K 15.344.92.482.65
K 15.468.73.722.94
K 15.51116.203.23
K 15.61729.924.26
K 15.729117.47.64
K 15.839624.812.3
DN251"K 25.130418.61.42
K 25.247627.91.25
K 25.371343.41.35
K 25.4118974.41.69
K 25.520081242.53
DN502"K 55.123781493.23
K 55.239632483.38
K 55.355483413.53
INFORMATION!
The oper. press. should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
The flow measurement of gases are refered to
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)