KROHNE ECOFLUXIFS1000 User Manual

02/97
Electromagnetic
Installation instructions
flowmeters
• Primary heads
• Compact flowmeters
CONTENTS Installation in the pipeline Pages 4-5 and 7-8 Grounding Pages 8-9
IFS 1000 F
ECOFLUX IFS 1000 F
IFM 1010 K IFM 1080 K
IFM 1010 K IFM 1080 K
3.1M65EA1 029721 Order No. DIN size 7.02146.31.00 US size 7.02146.71.00
0.73.00
Contents
System description 2 Product liability and warranty 2 Standards and approvals 2 Items included with supply 3
1 Important information for installation: PLEASE NOTE 4-5 2 Suggestions for installation 5 3 Instrument nameplate 6 4 Flowmeter versions 6 5 Installation in the pipeline 7 6 Torques 8 7 Grounding 8-9 8 Replacement of separate primary head IFS 1000 F 10
9 Spares and order numbers 10 10 Technical data 11 11 Dimensions and weights 12-13
Notes 14 Printed form to accompany flowmeters returned to Krohne 15
System description
ECOFLUX electromagnetic flowmeters are precision measuring instruments designed for the linear flow measurement of process liquids.
The process liquids must be electrically conductive: 05 µS/cm
20 µS/cm for demineralized cold water
The full-scale range Q DN 10 - 150 /
3
/8” - 6” Q
can be set as a function of the meter size:
100%
= 0.1 - 760 m3/hr = 0.4 - 3350 US Gal/min
100%
This is equivalent to a flow velocity of 0.3 - 12 m/s, or 1 - 40 ft/s.
Product liability and warranty
ECOFLUX electromagnetic flowmeters are designed solely for measuring the volumetric flowrate of electrically conductive, liquid process products.
Flowmeters with the ECOFLUX primary head are not certified for use in hazardous locations. Other flowmeter series are available for such applications.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty. In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable. If ECOFLUX flowmeters need to be returned to Krohne, please note the information given on the
last-but-one page of this manual. Krohne regret that they cannot repair or check your flowmeter(s) unless accompanied by the completed form sheet.
Standards and approvals
Please refer to the installation and operating instructions for the signal converter.
2
Items included with supply
IFS 1000 F primary head
primary head
in the size as ordered
certificate of calibration data
installation material as specified
in the following table
installation instructions
IFM 1010 K and IFM 1080 K compact flowmeters
compact flowmeter
in the size as ordered
certificate of calibration data
installation material as specified
in the following table
installation instructions
installation and operating instruc-
tions for the signal converter
Nominal size
measuring tube and pipe flanges
Pressure rating or flange class
of pipe flanges
Max. allowable operating pressure
bar psig
Supply scope S=standard O=option
incl.
grounding rings
3)
Standard with centering sleeves (optionaly with stud bolts, see below for type and number
excl.
grounding rings
2)
. . . DIN 2501 (BS 4504)
DN 10-15 1) PN 16/PN 40 16 230 S 4 xM12 DN 25 PN 16/PN 40 16 230 SO4 xM12 DN 40 PN 16/PN 40 16 230 SO4 xM16 DN 50 PN 16/PN 40 16 230 SO4 xM16 DN 80 PN 16/PN 40 16 230 SO4 xM16 DN 100 PN 16 16 230 SO8 xM16
PN 40 8 xM20
DN 150 PN 16 16 230 SO8 xM20
PN 40 8 xM24
. . . ANSI B 16.5
3
/8”-1/2” 1) 150/300 lb 16 230 S 4 1” 150/300 lb 16 230 SO4 1 1/2” 150/300 lb 16 230 SO4 2” 150 lb 16 230 SO4
300 lb 8
3” 150 lb 16 230 SO4
300 lb 8 4” 150/300 lb 16 230 SO8 6” 150 lb 16 230 SO8
300 lb 12
1
x
/2
1
x
/2
5
x
/8
5
x
/8
5
x
/8
5
x
/8
5
x
/8
5
x
/8
3
x
/4
3
x
/4
. . . JIS
DN 10-15 1) 10 K 07 100 S 4 xM12
20 K 14 200 4 xM12 DN 25 10 K 07 100 SO4 xM12
20 K 14 200 4 xM16 DN 40 10 K 07 100 SO4 xM12
20 K 14 200 4 xM16 DN 50 10 K 07 100 SO4 xM12
20 K 14 200 8 xM16 DN 80 10 K 07 100 SO8 xM12
20 K 14 200 8 xM20 DN 100 10 K 07 100 SO8 xM12
20 K 14 200 8 xM22 DN 150 10 K 07 100 SO8 xM16
20 K 14 200 12 xM22
1) For DN 10 and 3/8” sizes, use DN 15 or 1/2” pipe flanges.
2) Ground connecting wires V screwed to housing, supplied without gaskets.
3) DN 10 - 15 and
D2 gaskets between grounding rings and pipe flanges not included with flowmeter, to be provided by customer. Use Teflon-type gaskets to DIN 2690/ANSI B 16.21, deformable under pressure, 8 - 16 N/mm
DN 25 - 150 and 1” - 6”: grounding ring E (option) supplied loose, ground connecting wires V screwed to housing, gaskets not included in supply.
For arrangement of gaskets and connection of wires V, see Section 7 “Grounding“.
3
/8” - 1/2”: grounding rings E with inserted D1 gasket, screwed to housing.
2
/1160 - 2320 psi.
3
1 Important information for installation: PLEASE NOTE
Handling
Do not lift flowmeter by the signal converter housing or
Do not set flowmeter down on signal converter housing or terminal box.
the terminal box.
Use only solventless detergents to clean the signal converter housing (polycarbonate).
Temperatures
For operating pressure and vacuum load based on flange standards, see Section 10 “Technical data“.
Ambient temperature Process temperature
Compact systems -25 to +50 °C (-13 to +122 °F) -25 to +060 °C (-13 to +140 °F)
-25 to +40 °C (-13 to +104 °F) -25 to +120 °C (-13 to +248 °F) IFS 1000 F (separate) -25 to +60 °C (-13 to +140 °F) -25 to +120 °C (-13 to +248 °F) In storage -25 to +60 °C (-13 to +140 °F)
Location and position as required,
but electrode axis X – • – • – • – X must be approximately horizontal in a
Y
horizontal pipe run. Y terminal box or converter housing
X
X
Measuring tube must be completetly filled at all times.
Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating instructions for the signal converter.
Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34: below 2.2g for compact flowmeters in the frequency range of 20-50 Hz with the IFC 010 K
and 20-150 Hz with the IFC 090 K.
Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.
4
Strong electromagnetic fields, avoid in vicinity of flowmeter.
Straight inlet run minimum of 5 ××DN and outlet run minimum of 2 ××DN,
(DN = meter size), measured from the electrode axis.
Vortex and corkscrew flow: increase length of inlet and outlet runs or install flow conditioners.
Mixing different process liquids: install flowmeter upstream of mixing point or at an adequate distance downstream (minimum of 30 x DN), otherwise display may be unsteady.
Plastic pipes and internally coated metal pipelines: grounding rings required, see Section 7 ”Grounding”.
Insulated pipeline: do not insulate flowmeter.
Zero setting not necessary. To check, it should be possible to set ”zero” flow velocity
in the completely filled measuring tube. Shutoff valves should therefore be provided either downsteam of the flowmeter or upsteam and downstream of the flowmeter.
2 Suggestions for installation
To avoid measuring errors due to gas/air inclusion or to pipe runing empty please observe the following:
Highest point of pipe run
(Air bubbles collect in measuring tube - faulty measurements!)
Preferred locations
Downpipe
”Zero” flow velocity. Line drained. Faulty measurements!
Horizontal pipe run
Install in slightly ascending pipe section. If not possible, assure adequate velocity to prevent air, gas or vapor from collecting in upper part of flow tube.
Open feed or discharge
Install meter in low section of pipe.
Downpipe over 5 m (16 ft) length
Install air valve X downstream of flowmeter.
Long pipeline Always install control and shutoff valves
downstream of flowmeter.
Pumps Never install flowmeter on pump suction side.
> 5m
> 16 ft
open discharge
5