System description
ECOFLUX electromagnetic flowmeters are precision measuring instruments designed for the
linear flow measurement of process liquids.
The process liquids must be electrically conductive:≥ 05 µS/cm
≥ 20 µS/cm for demineralized cold water
The full-scale range Q
DN 10 - 150 /
3
/8” - 6” Q
can be set as a function of the meter size:
100%
= 0.1 - 760 m3/hr = 0.4 - 3350 US Gal/min
100%
This is equivalent to a flow velocity of 0.3 - 12 m/s, or 1 - 40 ft/s.
Product liability and warranty
ECOFLUX electromagnetic flowmeters are designed solely for measuring the volumetric flowrate
of electrically conductive, liquid process products.
Flowmeters with the ECOFLUX primary head are not certified for use in hazardous locations.
Other flowmeter series are available for such applications.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.
If ECOFLUX flowmeters need to be returned to Krohne, please note the information given on the
last-but-one page of this manual. Krohne regret that they cannot repair or check your flowmeter(s)
unless accompanied by the completed form sheet.
Standards and approvals
Please refer to the installation and operating instructions for the signal converter.
2
Items included with supply
IFS 1000 F
primary head
• primary head
in the size as ordered
• certificate of calibration data
• installation material as specified
in the following table
• installation instructions
IFM 1010 K and IFM 1080 K
compact flowmeters
• compact flowmeter
in the size as ordered
• certificate of calibration data
• installation material as specified
in the following table
• installation instructions
• installation and operating instruc-
tions for the signal converter
Nominal size
measuring
tube and
pipe
flanges
Pressure
rating or
flange class
of pipe flanges
Max.
allowable
operating
pressure
barpsig
Supply scope S=standard O=option
incl.
grounding
rings
3)
Standard with
centering sleeves
(optionaly with stud
bolts, see below for
type and number
1) For DN 10 and 3/8” sizes, use DN 15 or 1/2” pipe flanges.
2) Ground connecting wires V screwed to housing, supplied without gaskets.
3) DN 10 - 15 and
D2 gaskets between grounding rings and pipe flanges not included with flowmeter, to be provided by customer.
Use Teflon-type gaskets to DIN 2690/ANSI B 16.21, deformable under pressure, 8 - 16 N/mm
DN 25 - 150 and 1” - 6”: grounding ring E (option) supplied loose, ground connecting wires V
screwed to housing, gaskets not included in supply.
For arrangement of gaskets and connection of wires V, see Section 7 “Grounding“.
3
/8” - 1/2”: grounding rings E with inserted D1 gasket, screwed to housing.
2
/1160 - 2320 psi.
3
1 Important information for installation: PLEASE NOTE
•Handling
Do not lift flowmeter by the
signal converter housing or
Do not set flowmeter down on signal
converter housing or terminal box.
the terminal box.
•Use only solventless detergents to clean the signal converter housing (polycarbonate).
•Temperatures
For operating pressure and vacuum load based on flange standards, see Section 10
“Technical data“.
Ambient temperatureProcess temperature
Compact systems-25 to +50 °C (-13 to +122 °F)-25 to +060 °C (-13 to +140 °F)
-25 to +40 °C (-13 to +104 °F)-25 to +120 °C (-13 to +248 °F)
IFS 1000 F (separate)-25 to +60 °C (-13 to +140 °F)-25 to +120 °C (-13 to +248 °F)
In storage-25 to +60 °C (-13 to +140 °F)–
•Location and position as required,
but electrode axis X – • – • – • – X
must be approximately horizontal in a
Y
horizontal pipe run.
Y terminal box or converter housing
X
X
•Measuring tube must be completetly filled at all times.
•Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating
instructions for the signal converter.
•Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
•Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34: below 2.2g for compact flowmeters
in the frequency range of 20-50 Hz with the IFC 010 K
and 20-150 Hz with the IFC 090 K.
•Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.
4
•Strong electromagnetic fields, avoid in vicinity of flowmeter.
•Straight inlet run minimum of 5 ××DN and outlet run minimum of 2 ××DN,
(DN = meter size), measured from the electrode axis.
•Vortex and corkscrew flow: increase length of inlet and outlet runs or install
flow conditioners.
•Mixing different process liquids: install flowmeter upstream of mixing point or at an
adequate distance downstream (minimum of 30 x DN), otherwise display may be unsteady.
•Plastic pipes and internally coated metal pipelines: grounding rings required,
see Section 7 ”Grounding”.
•Insulated pipeline: do not insulate flowmeter.
•Zero setting not necessary. To check, it should be possible to set ”zero” flow velocity
in the completely filled measuring tube. Shutoff valves should therefore be provided either
downsteam of the flowmeter or upsteam and downstream of the flowmeter.
2 Suggestions for installation
To avoid measuring errors due to gas/air inclusion or to pipe runing empty please observe the
following:
Highest point of pipe run
(Air bubbles collect in measuring
tube - faulty measurements!)
Preferred
locations
Downpipe
”Zero” flow
velocity.
Line drained.
Faulty
measurements!
Horizontal pipe run
Install in slightly ascending pipe section. If not
possible, assure adequate velocity to prevent air,
gas or vapor from collecting in
upper part of flow tube.
Open feed or discharge
Install meter in low section of pipe.
Downpipe over 5 m (16 ft) length
Install air valve X downstream of
flowmeter.
Long pipeline
Always install control and shutoff valves
downstream of flowmeter.
Pumps
Never install flowmeter on pump suction side.
> 5m
> 16 ft
open discharge
5