Jenway 4510 User Manual

Page 1
Model 4510
Conductivity/TDS Meter
Operating Manual
451 050/REV A/03-03
Page 2
SAFETY Please read this information carefully prior to installing or using this equipment.
1. The unit described in this manual is designed to be operated only by trained personnel. Any adjustments, maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware of the hazards involved.
2. It is essential that both operating and service personnel employ a safe system of work, in addition to the detailed instructions specified in this manual.
3. References should always be made to the Health and Safety data supplied with any chemicals used. Generally accepted laboratory procedures for safe handling of chemicals should be employed.
4. If it is suspected that safety protection has been impaired in any way, the unit must be made inoperative and secured against any intended operation. The fault condition should immediately be reported to the appropriate servicing authority.
451 050/REV A/03-03
Page 3
Conductivity/TDS Meter
Operating Manual
Contents
Section 1 Introduction
Instrument Description 1.1 Instrument Specification 1.2
Section 2 Installation
Unpacking 2.1 Installation 2.2 Displays 2.3 Controls 2.4 Inputs/Outputs 2.5
Section 3 Operation
Good Practice Guidelines 3.1 Instrument Set up 3.2 Preparation of Standards 3.3 Calibration: With Known Cell Constant 3.4 With Standard Solution 3.5 Sample Measurement 3.6
Model 4510
Section 4 Maintenance
General 4.1
Section 5 Optional Accessories
Optional Accessories 5.1 Spares 5.2
Section 6 Interfacing
Analogue 6.1 RS232 6.2 Keypad Emulation 6.3 Printing 6.4
Section 7 Troubleshooting
k Troubleshooting 7.1
451 050/REV A/03-03
Page 4
Section 1
Introduction
1.1 Instrument Description
A general purpose Conductivity/TDS/temperature bench meter for use in laboratory situations. The instrument includes a 32 reading memory facility. Features include: automatic range selection, calibration on standard solutions or direct cell constant entry, automatic temperature compensation, analogue output and RS232 serial interface.
1.2 Instrument Specifications
Conductivity
Ranges 0 to 19.99µS / 0 to 199.9µS / 0 to 1999µS /
0 to 19.99mS / 0 to 199.9mS / 0 to 1999mS
Resolution 0.01µS / 0.1µS / 1µS / 0.01mS / 0.1mS / 1mS*
*only with cell constant >5
Accuracy ±0.5%±2 digits
TDS
Ranges 0 to 19.99mg/l / 0 to 199.9mg/l / 0 to 1999mg/l /
0 to 19.99g/l / 0 to 199.9g/l / 0 to 1999g/l
Resolution 0.01mg/l / 0.1mg/l / 1mg/l / 0.01g/l / 0.1g/l / 1g/l*
*only with cell constant >5
Accuracy ±0.5%±2 digits
Temperature
Range -10 to +105°C (14 to 221°F) Resolution 0.1°C (1°F) Accuracy ±0.5°C (±1°F) ATC Range 0 to 100°C (32 to 212°F) Manual Temp. Comp. Range 0 to 100°C (32 to 212°F) Display 3½ digit LCD Power 9V power supply Size 210x250x55mm Weight 850g
1
451 050/REV A/03-03
Page 5
Section 2
Installation
2.1 Unpacking
Remove the Model 4510 from the packaging and ensure the following items are included:
1. Model 4510 Conductivity/TDS Meter
2. Glass bodied conductivity cell with ATC K=1 (027 013)
3. Electrode holder
4. Power Supply (as specified at time of ordering the product)
5. Condensed operating instructions (451 051)
6. Operating Manual (451 050) The electrode stand requires minimal assembly (refer to the diagram below) Any shortages or damage should be reported immediately to the manufacturer or your local
distributor.
2.2 Installation
The Model 4510 is supplied ready to use. Connect the conductivity cell to the rear panel DIN socket. The electrode stand requires minimal assembly (refer to the diagram below).
Fig. 2.2.1 Electrode Holder Assembly
4
1
2
3
451 050/REV A/03-032
Page 6
2.3 Displays
1. Standard selection – indicates which type of standard is being used. Also indicates when a calibration is being performed.
2. Symbol – displayed during set-up of instrument parameters.
3. Primary display – 4½ digit. Provides direct readout in Conductivity or TDS of samples and standards.
4. Mode annunciators – shows selected measurement mode; Conductivity (µS or mS) and TDS (mg/l or g/l).
5. Endpoint symbol
6. Secondary display – 6 digit display. Provides direct readout of automatic or manual temperature. Scrolls and displays selected parameter information in set-up mode.
7. Status display – 2½ digit.
8. Mode annunciators – indicates temperature in °C or °F and whether the measurements are manually or automatically temperature compensated.
9. Mode tags – Each mode tag is highlighted when selected; SETUP, MODE or RESULTS.
3
451 050/REV A/03-03
Page 7
2.4 Keypad
1. ESC used to switch the instrument on and to place into standby mode (only if power supply lead remains connected to the instrument). Also used to escape/exit a mode.
2. CAL / CLR used to select and perform a calibration sequence. This key is also used to clear readings from Memory.
3. Print key used to initiate a print.
4. Up Arrow used for adjustment during set up, to scroll results and to toggle between Conductivity (µS or mS) and TDS (mg/l or g/l) modes.
5. Down Arrow used for adjustment during set up, to scroll results and to toggle between Conductivity µS or mS) and TDS (mg/l or g/l) modes.
6. Left Arrow used for adjustment during set up and to move between mode tags.
7. Right Arrow used for adjustment during set up and to move between mode tags.
8. STO used to accept an entered value in set-up mode and to instigate a stored reading. This key can also be used as a CAL key during calibration.
4
451 050/REV A/03-03
Page 8
2.5 Inputs/Outputs
2.5.1 Rear panel layout
1. Conductivity Socket 7 pin DIN socket which allows the conductivity cell to be connected.
2. Analog Out 2 x 4mm sockets. Analogue output (buffered electrode potential).
3. Output Socket 9 way socket for RS232 connection.
4. Power In AC 9V I/P socket. 2.1 x 5.5mm socket allowing the power supply to be
connected to the instrument.
451 050/REV A/03-035
Page 9
Section 3
Operation
3.1 Good Practice Guidelines
1. For greatest accuracy ensure no particulate matter is suspended in the solution under test. If
necessary, filter or allow the particles to settle prior to use. Do not allow the cell to come into contact with any sediment which may be present.
2. Ensure no air bubbles are trapped in the cell area between the plates.
3. Ensure the cell plates are completely immersed in the solution under test.
4. Thoroughly rinse the cell plates in deionised water after use, and for short term storage
immerse the cell in deionised water. Although it is not essential to store the conductivity cell with the plates in a wetted condition, if they are allowed to dry out completely, initial stability on re-use may be impaired until the cell plates become re-wetted.
5. Do not attempt to clean the cell plates as this will remove the black platinization and alter the
calibration and accuracy of the probe. If the cell plates become worn or damaged the plates should be re-platinised and the cell constant recalculated.
6. Ensure no salt deposits or particulate matter are allowed to build up around the cell plates or
on the probe body as this may produce a conductivity path lower than that through the solution. It is recommended that such deposits be removed by soaking the cell in deionised water. No attempt should be made to wipe off these deposits as this may cause damage to the cell plates.
7. Ensure the correct reference temperature is selected for the operating procedures being
used.
451 050/REV A/03-036
Page 10
3.2 Instrument Set-Up
The following section details the set-up modes available to the user. These are: Cell constant, Coefficient, Reference Temperature, TDS EC Ratio, Temperature units and Baud Rate. These can be set in sequence as detailed in this section or, by entering the SET UP mode and using the STO key to select a specific parameter at any time.
To exit the set up menus at any time press the ESC key. This will return the instrument to the MODE menu. Any parameters not saved will remain as defaults or previous setting.
3.2.1 Cell Constant (K) entry
Select SETUP mode on the display using the Left arrow key.
The secondary display will show CELL K and will then scroll CELL CONSTANT after 10 seconds. Symbol will flash while adjustment is being made. The first digit of the displayed value will be flashing. Adjustment of the flashing digit can be made
using the Up/Down arrow keys. The Left/Right arrow keys are used to select the next adjusting digit which will flash when selected. Once the full reading has been adjusted to the required value, press the STO key. The symbol will
disappear and the secondary display will momentarily show SAVED. The instrument display will update and then show the Coefficient set up menu.
7
451 050/REV A/03-03
Page 11
3.2.2 Temperature Coefficient set up
The display will show:
The secondary display will show COEFF and will then scroll TEMPERATURE COEFFICIENT after 10 seconds.
Symbol will flash while adjustment is being made. The first digit of the displayed value will be flashing. Adjustment of the flashing digit can be made
using the Up/Down arrow keys. The Left/Right arrow keys are used to select the next adjusting digit which will flash when selected. Once the full reading has been adjusted to the required value, press the STO key. The symbol will
disappear and the secondary display will momentarily show SAVED. The instrument display will update and then show the Reference Temperature set up menu.
3.2.3 Reference Temperature selection
The display will show:
The secondary display will show REF T and will then scroll REFERENCE TEMPERATURE after 10 seconds. Select the reference temperature from the options available (18, 20 or 25°C) by scrolling through the list using the Up/Down arrows.
Symbol will flash while adjustment is being made. The Left/Right arrow keys and the CAL/CLR key have no function during this set up. To accept the correct reference temperature value, press the STO key. The symbol will disappear
and the secondary display will momentarily show SAVED. The instrument display will update and then show the TDS EC ratio set up menu.
8
451 050/REV A/03-03
Page 12
3.2.4 TDS EC Ratio
The display will show:
The secondary display will show TDS EC and will then scroll TDS EC RATIO after 10 seconds. Symbol will flash while adjustment is being made. The first digit of the displayed value will be flashing. Adjustment of the flashing digit can be made
using the Up/Down arrow keys. The Left/Right arrow keys are used to select the next adjusting digit which will flash when selected. Once the full reading has been adjusted to the required value, press the STO key. The symbol will
disappear and the secondary display will momentarily show SAVED. The instrument display will update and then show the Temperature Units set up menu.
3.2.5 Temperature Units selection
The display will show:
The secondary display will show UNITS and will then scroll TEMPERATURE UNITS after 10 seconds. Select the preferred temperature unit from the options available (°C or °F). The Up/Down arrows toggle between the units.
Symbol will flash while adjustment is being made. The Left/Right arrow keys and the CAL/CLR key have no function during this set up. To accept the preferred temperature unit, press the STO key. The symbol will disappear and the
secondary display will momentarily show SAVED. The instrument display will update and then show the Baud Rate set up menu.
9
451 050/REV A/03-03
Page 13
3.2.6 Baud Rate
For full details relating to serial protocol refer to Section 6 of this manual. The display will show:
The secondary display will show BAUD and will then scroll SERIAL PORT BAUD RATE after 10 seconds.
Select the required baud rate (9600 or 1200) using the Up/Down keys which toggle between the values. The data bits and parity will automatically adjust as the baud rate is selected.
The Left/Right arrow keys and the CAL/CLR key have no function during this set up. Symbol will flash while adjustment is being made. To accept the selected baud rate, press the STO key. The symbol will disappear and the secondary
display will momentarily show SAVED and the instrument will return to the MODE menu.
10
451 050/REV A/03-03
Page 14
3.3 Preparation of Conductivity Standards
Suitable conductivity standards are available commercially or these can be made up as required from analytical grade reagents with reference to relevant physical tables.
Method for general purpose Conductivity Standard
Accurately weigh out 0.746 grammes of dried A.R. grade Potassium Chloride (KCl) and dissolve in 1 litre of good quality water. This produces a 0.01N solution with a conductivity of 1413µS at 25°C.
Storage
This solution must be stored in a plastic container and the air space above the solution should be kept to an absolute minimum. The shelf life of 1 week can be increased by storing below 4°C, but where any doubt exists about the viability of stored solution a fresh batch should be prepared.
3.4 Calibration with Known Cell Constant
1. Connect a standard pre-calibrated cell to the unit.
2. Select SETUP mode on the display using the Left arrow key.
3. Set the cell constant value as indicated on the body of the conductivity cell (refer Section 3.2.1) The standard X1.0 cell has a cell constant range from 0.8 to 1.20 and is generally calibrated to
2 decimal places. The display should be set to indicate this figure exactly. The X10 cell has a cell constant range from 8.0 to 12.0 and is generally calibrated to 2 decimal
places. The display should be set to indicate this figure.
4. Set the Temperature Coefficient value (refer Section 3.2.2).
5. Select the required reference temperature (refer Section 3.2.3).
3.5 Calibration with Standard Solution
Calibration of the unit and cell with standard solutions will only be necessary if:
1) The cell constant is unknown.
2) The cell constant has changed due to replatinising, wear or damage to the plates.
3) An ATC slope other than 2%/°C is required (refer Special Calibration).
General Calibration
1) Immerse the conductivity cell into the prepared standard.
2) Select the MODE menu using the keypad.
3) Press the CAL/CLR key. The unit will calibrate to the nearest standard.
11 451 050/REV A/03-03
Page 15
3.6 Sample Measurement
General
Conductivity is a temperature dependent measurement. All substances have a conductivity coefficient which varies from 1% per °C to 3% per °C for most commonly occurring substances. The automatic temperature compensation on the 4510 defaults to 1.91% per °C, this being adequate for most routine determinations. Conductivity readings varying with temperature may be due to the substances under test having a coefficient other than the typical value of 1.91% per °C. To eliminate this variation it is necessary to maintain all samples at the reference temperature by use of a thermostatic water bath or equivalent. Adjustment may be made by entering the SETUP menu and selecting COEFF (refer Section 3.2.2). The reading can then be adjusted to the required value (0.00 to 4.00) by using the keypad.
Sample Measurement
After calibration the measurement of samples is carried out by immersing the cell in the samples, allowing the readout to stabilise, and recording the result. The cell should be rinsed in deionised water between each sample to avoid contamination, shaken to remove internal droplets, and the outside wiped prior to immersion in the next sample. On completion of sample measurement the cell should be thoroughly rinsed in deionised water.
Storage Short Term the cell should be immersed in deionised water to keep the plates wetted.
Long Term the cell should be thoroughly rinsed in deionised water, the exterior body
wiped and then stored dry.
NOTE: When preparing the cell for storage the plate area MUST NOT be wiped dry.
When using a dry cell initial stability on re-use may be impaired until the cell plates become re-wetted.
To obtain optimum performance refer to Section 3.1 - Good Practice Guidelines.
12
451 050/REV A/03-03
Page 16
3.7 Results storage and display
To store the current displayed result press the STO key. The instrument display will momentarily show STORED on the secondary display. The memory location will be given on the status display.
Up to 32 results can be stored. Each result will be stored in the next available memory location. The instrument will store: Primary Conductivity or TDS readings
Temperature readings and the unit of measurement (°C or °F) MAN if manual temperature measurement is used Endpoint symbol
Selecting the RESULTS menu displays the reading stored at the displayed memory location.
The Up arrow increments the memory index, the Down arrow decrements the memory index. The Left arrow exits to the main measurement display. The Right arrow and STO keys do not function. CAL/CLR deletes the current memory location. The secondary display scrolls DELETED. To exit this mode press the ESC key. If the memory is full the secondary display will show FULL.
13
451 050/REV A/03-03
Page 17
Section 4
Maintenance
4.1 General
The Model 4510 is designed to give optimum performance with minimum maintenance. It is only necessary to keep the external surfaces clean and free from dust. To give added protection when the unit is not in use the unit should be disconnected from the mains supply and covered with the optional dust cover (060 406). For longer term storage or re-shipment it is recommended that the unit be returned to the original packing case. All conductivity cells should be thoroughly rinsed after use and stored in deionised water. Do not attempt to clean the surface of the plates as this could result in damage to the black platinised surface.
Section 5
Optional Accessories
5.1 Optional Accessories 060 406 Dust cover
037 701 Printer supplied with roll of thermal paper, serial connection lead, power supply,
power connection lead (UK) and pouch
037 801 Interface cable kit 050 002 Serial communication software (3½” disk)
Conductivity Cells 027 113 Conductivity Cell K=0.1 027 114 Conductivity Cell K=1 (plastic covered)
NOTE: The cells listed below are glass free and epoxy bodied. 027 211 Conductivity Cell with ATC K=0.1
027 212 Conductivity Cell with ATC K=1 027 213 Conductivity Cell with ATC K=10
Calibration Standards 025 138 1413µS Standard (500ml) 025 156 12.88mS Standard (500ml) 025 164 84µS Standard (500ml) 025 139 10µS Standard (500ml) 025 165 1382ppm TDS Standard (500ml)
5.2 Spares 027 013 Conductivity Cell K=1
037 702 Paper roll, thermal 021 030 UK 230V power supply 021 031 European 230V power supply 021 032 US 115V power supply 021 033 230V leaded power supply
14
451 050/REV A/03-03
Page 18
Section 6
Interfacing
6.1 Analogue
All units are provided with 2 x 4mm sockets, marked as ANALOG OUT, on the rear panel. An analogue output voltage of 1mV per least significant displayed digit is available from these sockets.
6.2 RS232
The Bi-directional RS232 interface is available on the rear panel 9 way D type connector. The connections are as follows:
DCD 1 - LINKED TO DTR AND DSR RXD 2 - INPUT TO 4510 TXD 3 - OUTPUT FROM 4510 DTR 4 - LINKED TO DCD AND DSR GND 5 DSR 6 - LINKED TO DCD AND DTR RTS 7 - OUTPUT FROM 4510 CTS 8 - INPUT TO 4510
Suggested interconnections are detailed below:
4510 IBM PC XT (25 way “D”)
1 DCD DCD 8 2 RXD RXD 3 3 TXD TXD 2 4 DTR DTR 20 5 GND GND 7 6 DSR DSR 6 7 RTS RTS 4 8 CTS CTS 5 9
4510 IBM PC XT (9 way “D”)
1 DCD 1 DCD 2 RXD 2 RXD 3 TXD 3 TXD 4 DTR 4 DTR 5 GND 5 GND 6 DSR 6 DSR 7 RTS 7 RTS 8 CTS 8 CTS 9 9
NOTE: Interface Cable (Order Code: 013 203) is required.
15
451 050/REV A/03-03
Page 19
Interfacing (continued)
The RS232 communications parameters on the computer or printer need to be set to match those of the Model 4510, as detailed below:
1200 Baud 9600 Baud 7 Data Bits 8 data bits Odd Parity OR No parity
1 Stop Bit 1 stop bit Setting of these options is detailed in Section 3.6.6. The Model 4510 supports both hardware (CTS/RTS) flow control and software XON/XOFF flow control. Pressing the PRINT key outputs from the RS232 interface. Sending an ASCII “D” to the 4510 causes a printout of the current displayed reading plus sample
number. Sending an ASCII “C” causes a printout of the last calibration parameters. Sending an ASCII "P" causes a printout of the stored readings.
6.3 Keypad Emulation
Keypad remote control using RS232 interface:
7 - Instrument On / Standby / Escape 1 - Calibrate / Memory Clear 9 - Print 8 - Up Arrow 2 - Down Arrow 4 - Left Arrow 6 - Right Arrow 3 or 5 - Enter / Store
16
451 050/REV A/03-03
Page 20
6.4 Printing
A 32 column serial printer (037 701) is available for use with the Model 4510. Connect the printer via the cable supplied with the printer to the 9 way socket located on the rear
panel of the instrument. To intiate a print out of data press the print key. When the first print is performed a header section will be printed showing:
Instrument name Spacing for entry of Operator & User ID Most recent calibration information (Cell Constant) Reference temperature Temperature coefficient TDS EC ratio
This will be followed by results data in either µS / mS (Conductivity) or mg/l / g/l (TDS) dependent on mode selected. Details will also be given on temperature.
An asterisk (*) after the sample number indicates that an endpoint has been reached. Each reading will be numbered sequentially. A calibration will reset the data number to 0001 and the header information will be re-printed. To obtain a print out of stored readings, enter the RESULTS MODE and press the print key. A print
out of all 32 print locations will then be generated.
17 451 050/REV A/03-03
Page 21
6.4.1 Example Printout
4510 Header printout 4510 Results printout
18
451 050/REV A/03-03
Page 22
Section 7
Troubleshooting
7.1Troubleshooting
Fault Possible Cause Action
No display Check power supply Check that correct 9V ac power supply
is connected and switched on.
Erratic display Check power supply Unit must be used with supplied 9V
acpower supply. Usage of other units will
cause the 4510 not to operate. Unstable display Conductivity cell defective Replace conductivity cell. Display permanently Intermittent or no connection Check cell connection to 4510.
under or over range Conductivity cell defective Replace conductivity cell
Contaminated solutions Replace solutions
Intermittent display Conductivity cell not fitted Check connections.
correctly
Reading drifts Conductivity cell stored dry Allow to soak for 2 hours
CO2 absorbtion by sample Noticeable for low conductivity - do not
allow sample to stand in unstoppered
bottles. Non linear readings Conductivity cell not zeroed Zero conductivity cell. Poor reproducibility Carryover between solutions Rinse cell in distilled water between
measurements. Will not print Connection 4510 requires the RS232 cable to
connect the printer.
Paper out The feed light on the printer will flash if
the unit requires paper.
Battery flat Connect ac power supply.
19 451 050/REV A/03-03
Page 23
EC Declaration of Conformity
JENWAY Model 4510 Conductivity/TDS Meter complies with the following European Standards: EN 50081-1:1992 Electromagnetic compatibility - Generic emission standard EN 50082-1:1992 Electromagnetic compatibility - Generic immunity standard (Performance crite-
rion B) EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control and
laboratory use Following the provision of: EMC Directive - 89/336/EEC and Low Voltage Directive - 73/23/EEC
Martyn J. Fall Managing Director, Jenway Gransmore Green, Felsted, Dunmow, Essex, CM6 3LB, England
20
451 050/REV A/03-03
Loading...