Please read this information carefully prior to installing or using this equipment.
1. The unit described in this manual is designed to be operated only by trained personnel. Any
adjustments, maintenance and repair must be carried out as defined in this manual, by a person
qualified to be aware of the hazards involved.
2. It is essential that both operating and service personnel employ a safe system of work, in addition to
the detailed instructions specified in this manual.
3. References should always be made to the Health & Safety data supplied with any chemicals used.
Generally accepted laboratory procedures for safe handling of chemicals should be employed.
4. If it is suspected that safety protection has been impaired in any way, the unit must be made
inoperative and secured against any intended operation. The fault condition should immediately be
reported to the appropriate servicing authority .
Theory of pH measurement3.1
pH Measurement3.2
Preparation of Buffer Solution3.3
Solution T emperature V alues3.4
Good Practice Guidelines3.5
Instrument Set-Up3.6
pH Calibration3.7
Error Codes3.8
mV Mode3.9
Performing Measurements3.10
Results Storage and Display3.11
Model 3510
Section 4Maintenance
General4.1
Cleaning/Re-conditioning of Glass Electrodes4.2
The Model 3510 is a general purpose pH/mV/Temperature bench meter used for routine laboratory
analysis. The meter supports 1, 2 or 3 point pH calibration on either manually entered pH buffer
values or automatically temperature compensated buffers to DIN, JIS and NIST standards and
Jenway buffers supplied with the instrument. Up to 3 decimal place resolution is available. The 3510
includes a 32 reading memory facility .
1.2 Instrument Specification
pH (1, 2 or 3 point cal)
Range:-2.000 to 19.999pH
Resolution:0.001 / 0.01 / 0.1pH
Accuracy:±0.003pH
mV (Absolute or Relative)
Range:-1999 to +1999mV
Resolution:0.1mV
Accuracy:±0.2mV
Input Impedance:>10
12
ohms
Temperature Measuring
Ranges:-10 to +105°C / 14 to 221°F
Resolution:0.1°C / 1°F
Accuracy:±0.5°C / ±1°F
ATC Range:0 to 100°C / 32 to 212°F
Manual Temp. Compensation:0 to 100°C / 32 to 212°F
Auto Buffer Selection:Jenway (2.00, 4.00, 7.00, 9.20 and 10.00)
DIN (3.06, 4.65, 6.79, 9.23, 12.75)
NIST (1.68, 4.01, 6.87, 9.18, 12.45)
JIS (1.68, 4.01, 6.87, 9.18, 12.45) or manually entered buffers
Calibration:User selectable 1, 2 or 3 point
Outputs:Analogue 1mV per 0.01pH
RS232
Display:Back lit custom LCD
Power:Power Supply 9V ac
Size:275(l)x240(w)x150(d)mm
Weight:850g
1
351 050/REV B/09-06
Section 2
Installation
2.1 Unpacking
Remove the Model 3510 from the packaging and ensure the following items are included:
4.Calibration point – shows 1, 2 or 3 point symbol depending on level of calibration selected by
the user.
5.Buffer selection – indicates whether the instrument is using manually entered or automatic
selection buffers. Will show which type of buffer is being used.
6.Endpoint symbol – this symbol is displayed when the pH changes by less than 0.005pH
(0.2mV) over a five second period. Once an endpoint has been detected the reading must
change by more than 0.005pH (0.2mV) to clear the endpoint symbol.
7.Secondary display – 6 digit display . Provides direct readout of automatic or manual
temperature. Scrolls and displays selected parameter information in set-up mode.
8.Mode annunciators – indicates temperature in °C or °F and whether the measurements are
manually or automatically temperature compensated.
9.Status display – 2½ digit. Provides information relating to electrode slope value, mV Eo value
at calibration and memory result number.
10.Mode tags – Each mode tag is highlighted when selected; SETUP, MODE or RESULTS.
3
351 050/REV B/09-06
2.4 Keypad
2.4.1 Keypad
1. ESCused to switch the instrument on and to place into standby mode (only if power
supply lead remains connected to the instrument). Also used to escape/exit a
mode.
2. CAL / CLRused to select and perform a calibration sequence. This key is also used to clear
readings from Memory. Used to select Abs/Rel mV in mV mode.
3. Print keyused to initiate a print.
4. Up Arrowused for adjustment during set up, to scroll results and to toggle between mV and
pH modes.
5. Down Arrowused for adjustment during set up, to scroll results and to toggle between mV and
pH modes.
6. Left Arrowused for adjustment during set up and to move between mode tags.
7. Right Arrowused for adjustment during set up and to move between mode tags.
8. STOused to accept an entered value in set-up mode and to instigate a stored
reading.
This key can also be used as a CAL key during calibration.
4
351 050/REV B/09-06
2.5 Inputs/Outputs
Fig. 2.5.1 – Rear panel layout
1. Ref Socket2mm pin socket. Connection socket for separate reference electrode.
When performing measurements with some pH and ion selective
electrodes a separate reference electrode is needed.
2. pH SocketBNC type socket which allows combination pH or redox electrodes to be
used.
3. Temp Socket8 pin mini-DIN socket. This allows the Automatic Temperature
Compensation (A TC) probe (027 500) to be connected.
4. Analog Out2 x 4mm sockets. Analogue output (buffered electrode potential).
5. Output Socket9 way socket for RS232 connection.
6. Power InAC 9V I/P socket. 2.1 x 5.5mm socket allowing the power supply to be
connected to the instrument.
5
351 050/REV B/09-06
Section 3
Operation
3.1 Theory of pH Measurement
pH is a unit of measurement which defines the degree of acidity or alkalinity of a solution. It is usually measured on a scale of 0 to 14. The pH value quantifies the degree of hydrogen ion activity of
an acid or a base in terms of hydrogen ion activity .
The internationally accepted symbol, pH, is derived from “p”, the mathematical symbol of the
negative logarithm and “H”, the chemical symbol for Hydrogen. The pH value is the negative
logarithm of Hydrogen ion activity as shown in the mathematical relationship pH= -log[H+].
The pH value of a substance is directly related to the ratio of the Hydrogen ion [H+] and the Hydroxyl
ion [OH-] concentrations. If the concentration of H+ is greater than OH-, the material is acidic and
has a pH value of less than 7. Conversely , if the concentration of OH- is greater than H+ the material
is basic, with a pH value greater than 7. If the concentrations of H+ and OH- are equal the material is
neutral with a pH value of 7.
It can, therefore, be seen that pH is a measurement of both acidity and alkalinity , even though by
definition it is a selective measurement of hydrogen ion activity . The logarithmic relationship
between hydrogen ion concentration and the pH unit means that a change of one pH unit
represents a ten-fold change in hydrogen ion concentration.
3.2 pH Measurement
pH can be measured by using either pH papers/indicators or a pH meter , dependent on the level of
accuracy required. pH papers or indicators change colour as the pH level varies. These can be
used as a guide to the pH level, but can be limited in accuracy and difficult to interpret correctly in
murky or coloured samples.
For greater accuracy the use of a high impedance pH meter is recommended, together with a pH
measuring electrode and reference electrode.
Each component part of the measurement system can be described as follows:
a) the pH meter – is a high impedance amplifier used to accurately measure the minute electrode
voltages produced. The pH meter will display the results directly in pH units on either an analogue
or digital display. Voltages can also be read for special applications, ORP (Oxidation-Reduction
Potential) measurements or with Ion Selective Electrodes.
b) the pH electrode – is a hydrogen ion sensitive glass bulb, with a millivolt output that varies with
the changes in the relative hydrogen ion concentration inside and outside of the bulb. The pH
electrode has very high internal resistance, making the voltage change with pH difficult to measure.
The input impedance of the pH meter and leakage resistances are therefore important factors.
c) the reference electrode – these cells consist of an internal element, usually a silver/silver
chloride wire, electrolyte (KCl) and a liquid junction. The liquid junction provides a leak path for the
internal electrolyte to “weep” into the sample chamber and provide an electrical contact with the
liquid to be measured. If the liquid junction is inefficient then measurement will be inaccurate. It is
common for the reference electrode to be incorporated into the pH electrode. It is then called a
combination electrode. The Model 3510 meter is supplied with a combination electrode.
6
351 050/REV B/09-06
The voltage developed by each individual pH electrode in the presence of a known hydrogen ion
concentration is theoretically predictable, but in practise deviations from the theoretical value can be
expected. These deviations will change slowly during the life of an electrode. It is therefore essential
to routinely calibrate the system using solutions with a known and constant pH value. These
solutions are called buffers.
3.3 Preparation of Buffer Solutions
Care must be taken in the preparation of all buffer solutions. The correct quantity of distilled or
deionised water should be used when preparing the solutions. For accurate and repeatable results
it is essential to follow the manufacturers instructions carefully.
3.4 Solution Temperature Values
The value of all buffer solutions varies with solution temperature. For accurate calibration of
electrodes using buffer solutions, it is necessary to measure the temperature of the buffer solution
being used. The unit should then be calibrated to the corrected pH value. Manufacturers of buffer
powders and solutions will provide a table of values at varying temperatures for their buffers.
Note:Buffer solutions will contaminate with exposure to air and should be stored
in airtight containers when not in use. Used solution should be discarded and
not returned to the container as this will cause contamination.
For best results fresh solutions should be prepared prior to calibration.
7
351 050/REV B/09-06
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.