In-Situ Sensor Summary Sheet User Manual

Innovations in
Water Monitoring
TROLL® 9500 Sensor Summary
Operating temp.
Storage temp. -40 to 65° C (-40 to 140° F)
Dimensions & weight
Wetted materials PVC, 316L stainless steel, titanium, Acetal, Viton®, nylon. Cable: Tefzel® or polyurethane
Water tightness rating
Output options RS485/RS232; SDI-12 (optional with SDI-12 adapter); ASCII streaming mode or binary command
Power
TROLL 9500 Logging
Data logging Professional and LTS models only: 16 programmable tests (dened, scheduled to run, or stored). Logging modes: Linear, Linear Average, Event
Memory 4 MB (222,000 data records1)
Standard Sensors Accuracy Range Depth Rating Response Time (T90)
Barometric pressure ±0.3% FS 16.5 psia
Level, Depth, Pressure
TROLL® 9500 Multiparameter Instrument
-5 to 50° C (23 to 122° F)
4.7 cm (1.85 in) OD x 55.25 cm (21.75 in). With twist-lock hanger: 56.52 cm (22.25 in). Restrictor: 8.9 cm (3.5 in) OD x 21 cm (8.25 in) long;
1.9 kg (4.2 lbs)
IP68 with all sensors and cable attached. Battery compartment: IP67 without the battery cover or cable attached
External: 9-16 VDC (optional). Internal: 2 user-replaceable D batteries (use either alkaline or matched pair of Saft lithium batteries)
5
Methodology
±0.1% FS or better Sensor accuracy: -5 to 50° C
Meets highest rating
15, 30, 100, or 300 psi Non-vented
30 psia: 10.90 m (35.76 ft) 100 psia: 60.11 m (197.2 ft) 300 psia: 200.7 m (658.6 ft) Vented 15 psig: 10.55 m (34.61 ft) 30 psig: 21.10 m (69.21 ft) 100 psig: 70.32 m (230.7 ft) 300 psig: 211.0 m (692.1 ft)
<30 sec per 30 m (100 ft) of cable
Instantaneous in thermal equilibrium
Silicon strain gauge
Silicon strain gauge (non-vented or vented)
Conductivity Low: ±0.5% or 2 µS/cm
Dissolved oxygen RDO® sensor
Clark cell electrode
pH (single) or pH/ORP (combo)
Temperature ±0.1° C -5 to 50° C (
Extended Parameter (XP) Sensors
Ammonium (NH
Chloride (Cl-) ±15% 0.35 to 35,500 ppm Cl 70 m (231 ft)
Nitrate (NO
Turbidity
Warranty TROLL 9500 and all sensors (excluding RDO & ISE sensors) come with a 1-year warranty. RDO Sensor: 3-year warranty. ISE sensors: 90-day
Notes
3
-
)
3
High: ±0.5% + 2 µS/cm
±0.1 mg/L ±0.2 mg/L ±10% of reading
±0.2 mg/L
pH: ±0.1 pH unit
6
ORP: ±5.0 mV
+
) ±10% 0.14 to 14,000 ppm N 14 m (46 ft) T98: <60 sec, 1.4 to 14
4
±10% 0.14 to 14,000 ppm N 14 m (46 ft) T98: <60 sec, 1.4 to 14
±5% or 2 NTU/FNU 0 to 2,000 NTU/FNU 105 m (346 ft)
warranty. RuggedCable® System: 2-year warranty.
1
A single data record includes time stamp, temperature, RDO, pH, and conductivity logged in Linear or Linear Average mode. 2Full operating range: 70 to 200,000 µS/cm. 3Full operating range: 0 to 50 mg/L. 4EPA-approved under the Alternate Test Procedure process. 5Time required for a sensor to reach 90% stable. 6pH sensor and pH/ORP sensor temperature range 0 to 50° C.
Low: 5 to 20,000 µS/cm High2: 150 to 112,000 µS/cm
0 to 8 mg/L 8 to 20 mg/L 20 to 50 mg/L
0 to 20 mg/L; saturation
pH: 0 to 12 pH units ORP: ±1400 mV
0 to 200%
23 to 122° F
) Meets highest rating < 30 sec EPA 170.1
Low: Meets highest rating High: Meets highest rating
Meets highest rating
246 m (807 ft)
pH: 211 m (692 ft) pH/ORP: 211 m (692 ft)
Low:
Instantaneous
High:
Instantaneous
T90: <45 sec. T95: <60 sec. T90: <45 sec. T95: <60 sec. T90: <45 sec. T95: <60 sec.
1-mil membrane: 1-2 min @ 25° C 2-mil membrane: 90 sec to 3 min
pH: <15 sec, pH 7 to pH 4 ORP: <15 sec
ppm N
T98: <60 sec, 3.54 to 35.45 ppm Cl
ppm N
Instantaneous (5 sec for rst reading)
Std. Methods 2510, EPA 120.1 Std. Methods 2510, EPA 120.1
EPA-approved In-Situ Methods4 1002-8­2009, 1003-8-2009, 1004-8-2009
Std. Methods 4500-O G, EPA 360.1
pH: Std. Methods 4500­H+, EPA 150.2 ORP: Std. Methods 2580
Std. Methods 4500-NH3 D, EPA 350.3
Std. Methods 4500­Cl- D
Std. Methods 4500­NO3 D
ISO 7027
TROLL® 9500 Sensor Summary
Potential Interferences
pH
Sodium salts
DO (Clark Cell)
CO2, HS-, O
3
DO (Optical)
NaCIO (bleach) >3% Alcohols >5% H2O2 (peroxide) >3% Gaseous SO Gaseous Cl
2
2
Do not use in organic solvents, e.g., acetone, chloroform, or methylene chloride.
ORP
Ions that are stronger reducing agents than hydrogen or platinum, e.g., chromium, vanadium, titanium
Nitrate NO
HCO
3
2-
CO
3
-
CIO
4
I- iodide
-
CIO
3
CN- cyanide Br - bromide
-
NO
nitrite
2
HS- hydrogen sulde CI- chloride H2PO AcO- acetate F- uoride
2-
SO
sulfate
4
*Signicant interference above pH3
-
3
-
hydrogen carbonate* carbonate perchlorate
chlorate
-
hypophosphate
4
Ammonium NH
4
Cs+ cesium K+ potassium TI+ thallium Na+ sodium Li+ lithium H+ pH Ag+ silver
+
Chloride Cl
Complexes with: Bi3+ bismuth Cd2+ cadmium Mn2+ manganese Pb2+ lead Sn2+ tin Tl3+ thallium
Reducing agents: OH- hydroxide NH3 ammonia
2-
S2O
thiosulfate
3
Br - bromide S2- sulde I- iodide CN- cyanide
-
Calibration Solutions
Solution
Quick Cal (pH, ORP, conductivity)
ZoBell’s (ORP)
Low Conductivity (143 µS)
Other Conductivity
pH Buers
Sodium Sulte
Turbidity
Low Chloride
Low Nitrate
Low Ammonium
Other ISE Solutions
pH Fill Solution
Shelf Life After Opening
7-21 days (±10 mV, ±0.05 pH, ±50 µS)—store cool and dark, shake before use
7-21 days (±10 mV)—store cool and dark
hours (±1 µS check before use)
3-6 months
3-6 months
3-6 months
12 months
1 month
1 month
1 month
2 months
6-12 months
For more information contact In-Situ Inc.
221 East Lincoln Avenue, Fort Collins, CO 80524 1-800-446-7488 (toll-free in U.S. & Canada) 1-970-498-1500 (international & domestic)
www.in-situ.com
Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, Pocket-Situ, RDO, RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc. © 2011. All rights reserved.
Tefzel is a registered trademark of E.I. du Pont de Nemours & Co. Viton is a registered trademark of DuPont Performance Elastomers L.L.C.
0032005 rev. 019 09/11