A 100 Introduction to the Sensistor ISH2000 series
Sensistor ISH2000 can be purchased in one of three versions. A desktop model
(Sensistor ISH2000), a battery operated model (Sensistor ISH2000C), and a
panel mount model (Sensistor ISH2000P).
Figure A-1: Sensistor ISH2000
Sensistor ISH2000 is equipped with a number of powerful functions making it
very easy to integrate in a semi or fully automatic test system. The functions
range from output of all necessary status signals and printer/communication
port to an advanced Active Probe Control system (APC). This makes the detector capable of controlling advanced sample collecting devices down to simple
test fixtures.
Sensistor ISH2000C
Sensistor ISH2000P
Figure A-2: Sensistor ISH2000C
The battery operated model, Sensistor ISH2000C, has all the Sensistor ISH2000
features apart from the APC system. This means that only passive probes (for
example Hand Probe P50) can be used. The battery, a Li-ion battery at 14.8 V,
can not support the current required to operate external probes. Sensistor
ISH2000C will operate for 14 hours on a fully charged battery with screensaver
and mute function. One hour charging will give about one hour of operating
time.
Figure A-3: Sensistor ISH2000P
The panel mount model, Sensistor ISH2000P, has identical features of the Sensistor ISH2000.The difference is that the Sensistor ISH2000 can be installed in
the operator’s panel or any other flat surface. Also it operates on +24 VDC.
Mounting brackets and a panel rubber seal are delivered with the detector.
The Sensistor ISH2000 leak detector is using SENSISTOR sensor technology
inside. SENSISTOR sensor technology uses an extremely sensitive hydrogen gas
sensor based on a microelectronic field effect transistor (MOS-FET).
The gas sensitivity appears when hydrogen absorbs into the sensor through a
metal alloy (metal hydride) layer.
Only hydrogen can diffuse into the metal and this makes the sensors practically
insensitive to other substances that do not contain free hydrogen molecules.
The signals from the sensors are processed by a microprocessor which also controls the sensor temperature with high accuracy, and other sensor diagnostics in
order to ensure perfect functionality. It also automatically compensates for
background gas.
If all you wish to do is to detect the presence of a leak, that is, find out whether
there is a leak or not, then use the Detection Mode (or use the detection bar in
Combined Mode). The definition of Leak/No Leak will then simply be "A leak is
a leak when it can be detected by the detector, set to a specific sensitivity".
To set up:
The operation in Detection Mode is not quantitative. The audio and visual sig-
nal will increase and decrease with the gas concentration. Therefore, there is no
actual calibration to be done, but rather a setting of the sensitivity to a desired
level.
A typical set-up procedure for Detection Mode is:
• Set up a reference leak which corresponds to the smallest leak you wish to
detect.
• Put the probe close to the reference leak and note approximately what reaction you get (no reaction, small, medium, high, full scale) within the first few
seconds.
• Set the sensitivity. This can be done permanently under the menu Detection
Mode Settings or temporarily as a Direct Sensitivity Adjustment on the display
(unless you have set this function to OFF under the Detection Mode Settings
menu).
There is also an Auto ranging function which can be selected under the Detection Mode Settings menu.
Note: If the Detection Mode is used and the alarm function is required to be
activated at a particular calibrated level, then the unit must be calibrated in
accordance with the instructions, see "Calibrate the leak detector" on page 17.
The reason for this is that the alarm is based on the Analysis Mode when the
Detection Mode is displayed.
To Locate Leaks
Note: The Detection Mode (or use the detection bar in Combined Mode) is
used to locate leaks. This mode is semi-quantitative, that is, it gives an audio
and visual signal which increases as a leak is approached (a higher gas concentration) and decreases as you move the probe away from the leak. It does not
display figures. In this mode of operation leaks can easily be detected using a
sensitivity which can be preset.
Leaks can be located very accurately, even when there are other leaks nearby. If,
for example, you are trying to locate a leak on a product and the product has a
major leak, then you will get an audio signal as soon as the probe is placed
close to the product.
When the probe is moved around and over the product, the signal will increase
as the probe approaches the leak. If the signal goes out of scale, simply reduce
the sensitivity setting to bring the signal within the scale. Working with the sensitivity setting this way you will be able to locate multiple leaks that are in close
proximity to each other.
Note: Working inside a confined space such as, for example, a cabinet or a narrow passage on a combustion engine there is a risk that the background concentration accumulates to levels close to the upper detection limit of the
detector. In such case it will not be possible to locate leaks as easily as in open
spaces.
Hint: It is good practice to detect a leak, locate it, and immediately remove the
probe to avoid saturation. The probe is not damaged by the exposure but it will
recover more slowly. After excessive exposure it will be less sensitive for a short
period of time.
The Analysis Mode (or use the analysis figures in Combined Mode) is used for
measuring the size of a leak (or the concentration of a gas sample). To be able
to do this measurement and obtain correct values, the instrument must first be
calibrated using the calibration function.
In the Analysis Mode the detector determines the gas concentration from the
change, as the probe goes from being exposed to background to being
exposed to a certain gas concentration. The detector does not continuously
monitor the gas concentration but takes just one reading instead. Another suitable alternative name for this mode could be Sampling Mode. It is important to
keep this in mind when using the detector in this mode.
In Analysis Mode the probe should be moved directly from a background situation to the test point. The size of the leak in PPM, or any other selected units, is
shown on the display. The probe can and should be removed from the measuring point as the measured value steadies and remains on the display. The period
during which the measured value is displayed can be adjusted in the Analysis
Mode Settings menu.
When pure hydrogen gas is released in air its flammability range spans from 4%
to 75% of hydrogen in air. Below 4% there is insufficient chemical energy available for a flame to occur. Above 75% hydrogen there is not enough oxygen left
to support a flame. When, for example, a mixture of less than 5.5 % hydrogen
in nitrogen mixes with air there is not sufficient energy to support a flame, irrespective of the ratio of air-to-gas. When a mixture of more than 5.5 % hydrogen in nitrogen is released into air there is a region of ratios of air-to-gas where
the mixture is flammable. When, for example, a mixture of 10% hydrogen in
nitrogen mixes with air there is still very little energy available. Only in exceptional circumstances can a flame be self-supporting. However, such mixtures
cannot detonate.
Hydrogen can be stored and transported as either a gas a cryogenic liquid.
The tracer gas is standard welding gas of industry quality, easy to obtain at low
For automatic leak testing of entire products or parts of products. Can also be used
for testing permeability of materials.
Part No: 590-035
AP57 Counter Flow Probe
For fast manual leak location in hard to-reach places. Active probe that sniffs the
sample air past the hydrogen sensor in the probe tip.
Part No: 590-550
For leak detection in enclosed spaces or in environments with a high background
level of tracer gas. An adjustable air flow from the probe tip provides a protective air
curtain against tracer gas in the surroundings.
The normal risks associated with working with all compressed gases must be
considered.
Indicates a potentially hazardous situation which, if not avoided, could result in
property damage.
Indicates a potentially hazardous situation which, if not avoided, could result in
moderate or minor injury. It may also be used to alert against unsafe practices.
Indicates a potentially hazardous situation which, if not avoided, could result in death
or severe injury.
Indicates an imminently hazardous situation that, if not avoided, will result in death or
severe injury (extreme situations).
WARNING!
Pure hydrogen is a flammable gas. Only use readymade Hydrogen Tracer Gas of 5% Hydrogen in
Nitrogen. This is a standard industrial gas mixture
used in various industrial applications.
Note:
Whenever the word Hydrogen is used in
this manual it implies that the hydrogen gas is
safely mixed with Nitrogen in the proportions
- 95% N2.
5% H
2
Leak detector label
WARNING!
Since the tracer gas mix contains no oxygen,
releasing large amounts of gas in a confined space
may lead to asphyxiation.
WARNING!
Compressed gases contain a great deal of stored
energy. Always carefully secure gas bottles before
connecting pressure regulator. Never transport gas
bottle with the pressure regulator fitted.
Before connecting tracer gas: confirm that the connectors or test object is designed
for working at the test pressure.
WARNING!
Pressurising objects at too high pressures can result
in a burst object. This in turn can result in serious
injury or even death.
Never pressurise objects that have not previously
been burst tested or otherwise approved for the
chosen test pressure.
Located at the back of the Sensistor ISH2000, a label indicates the electrical
specifications of the leak detector and its serial number.