4.4.7Activating the Interfering Gas Suppression (IGS) 4-10
4.4.8Setting up a user-definable gas 4-10
4.5Settings Sub-menu 4-12
4.5.1Vacuum & Access Menu 4-12
4.5.2Audio Functions 4-14
4.5.3Display Settings 4-16
4.5.4Setting-up / editing an I•Guide Program 4-18
4.5.5Miscellaneous Settings 4-21
4.6Interfaces 4-23
4.6.1Control location 4-23
4.6.2Recorder outputs 4-23
4.6.3RS232 Protocol 4-26
4.6.4Baud rate & end sign 4-27
4.6.5Select PLC Inputs 4-27
4.6.6Select PLC Outputs 4-28
4.6.7ECO-Check (only available in Advanced Mode) 4-29
4.7The Info Menu 4-30
4.8History & Maintenance 4-34
4.9Monitoring Sensitivity 4-39
0-4Content
5 Ecotec E3000 Messages 6-1
5.1Error Messages and Warnings 6-1
6 Detecting Specific Gases 5-1
6.1R134a / Cyclopentane 5-1
6.2R134a / R245fa 5-1
6.3R600a / Cyclopentane / Isopentane 5-1
6.4Helium 5-1
6.5Hydrogen / Forming gas 5-2
6.6Methan 5-2
kina22eIVZ.fmTechnical Handbook(1206)
7 Equipment Connections 7-1
7.1I/O Port (Control Inputs and Outputs) 7-1
7.1.1Ground connectors 7-2
7.1.224V Output 7-2
7.1.3PLC Inputs 7-2
7.1.4PLC Outputs 7-3
7.1.4.1Relay outputs 7-4
7.1.4.2Recorder Outputs 7-5
7.1.5How to perform a calibration 7-6
7.2RS232 interface 7-7
8 Maintenance 8-1
8.1Maintenance schedule 8-1
8.2Adjusting the IGS function 8-2
8.3Exchanging the air filter 8-3
8.4Replacing the Operation Fluid Reservoir 8-5
8.5Exchanging the mains fuses 8-8
8.6Replacing filters in the sniffer line 8-9
8.6.1Replacing the felt discs of the capillary filter 8-9
8.6.2Replacing the felt discs (with water protection tip) 8-11
8.6.3Checking / replacing the sinter filter 8-11
8.7Switching the capillary filter 8-12
8.7.1Switching from metal to plastic capillary filter 8-13
8.7.2Switching from plastic to metal capillary filter 8-14
8.8Replacing the gas reservoir of the
ECO-Check 8-14
9 Annex 9-1
9.1The gas library 9-1
9.2Declaration of Conformity 9-10
kina22eIVZ.fmTechnical Handbook(1206)
Content0-5
Important Safety Precautions
Warning
Caution
Warning
Indicates procedures that must be strictly observed to prevent hazards to persons.
Indicates procedures that must strictly be observed to prevent damage to or
destruction of the Ecotec E3000 leak detector.
NoticeIndicates special requirements the user must comply with.
The INFICON Ecotec E3000 leak detector has been designed for safe and efficient
operation when used properly and in accordance with this Technical Handbook.
It is the responsibility of the user to carefully read and strictly observe all safety
precautions described in this chapter and throughout this Technical Handbook.
The Ecotec E3000 must only be operated in the proper condition and under the
conditions described in this Technical Handbook.
It must be operated and maintained by trained personal only. Consult local, state,
and national agencies regarding specific requirements and regulations.
Adress any further safety, operation and / or maintenance questions to our nearest
office.
Presumed Risk:
If it is to be assumed that safe operation is no longer possible, the device is to be
taken out of service and secured against unsupervised operation.
Notice This can e. g. be the case if:
•the unit has visible damage,
•fluid has gotten into the device,
•the device is no longer working,
•the device has been stored for an extended period of time under unfavorable
conditions or
•after significant moving or transportation stresses.
0-6Important Safety Precautions
kina22eIVZ.fmTechnical Handbook(1206)
Failure to observe the following precautions could result in serious personal
Warning
Warning
Warning
Warning
Warning
Warning
Warning
injury:
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Ecotec E3000 with the ground conductor unconnected is not
permissible.
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
close distance as this may cause permanent damage to the eye.
Danger of electric shock.
Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not
be exceeded or reached to ground or ground equipment conductors.
According to the type of in- or outputs lower voltages had to be accepted. For this,
please refer to the information given in the responding chapters.
For all maintenance on the Ecotec E3000, the Ecotec E3000 must be disconnected
from power.
Before exchanging the air filter the Ecotec E3000 must be disconnected from
power.
Before exchanging the fuses the Ecotec E3000 must be disconnected from power.
kina22eIVZ.fmTechnical Handbook(1206)
Important Safety Precautions0-7
Warning
Before exchanging the lubricant reservoir the Ecotec E3000 must be disconnected
Warning
Warning
Warning
Caution
Caution
from power.
Dangerous gases pollute the machine.
So you must not use the machine for detecting toxical, acidity, microbiological,
explosive, radioactive or other noxious matters.
Caution: Danger of explosion
Hydrogen forms a highly explosive gas mixture with air.
Great caution is necessary when using hydrogen! No smoking, no naked flames,
avoid sparks.
Danger of explosion!
To use the Ecotec E3000 in explosion hazard areas could cause ignition of
flammable mixtures.
The Ecotec E3000 must only be operated outside of explosion hazard areas.
Failure to observe the following precautions could result in damage to the
equipment:
The Ecotec E3000 must not be operated while standing in water or when exposed
to drip water. The same applies to all other kinds of liquids.
This Ecotec E3000 should only be used in rooms.
Avoid contact of the Ecotec E3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
0-8Important Safety Precautions
kina22eIVZ.fmTechnical Handbook(1206)
Caution
Ensure sufficient air cooling (see also Section 1.2)
Caution
Caution
Caution
Caution
Caution
Caution
Caution
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Ecotec E3000, a space of at least
20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear
must be no less than 10 cm (4 in.). Moreover, the Ecotec E3000 handles for
carrying the leak detector at the sides of the main unit must not be covered at any
time as these acts as air inlet and outlet. Avoid the presence of heat sources in the
vicinity of the Ecotec E3000.
Before connecting the Ecotec E3000 to the mains you must make sure that the
mains voltage rating of the Ecotec E3000 coincides with the locally available mains
voltage.
Do not suck in any liquids.
Permissible maximum input voltage PLC 28 V.
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
kina22eIVZ.fmTechnical Handbook(1206)
Important Safety Precautions0-9
Caution
The air filter should be checked for contamination at least every 6 months and
Caution
Caution
should be definitely exchanged after 2 years.
The lubricant reservoir may contain toxic substances from the pumped media.
Please dispose of lubricant reservoir as required by local regulations. A Safety Data
sheet for the lubricant is available on request.
Caution: Abrupt movements.
Abrupt movements can damage the running turbo pump.
Avoid abrupt movement and vibration of the instrument (e.g. running over cables,
door sills) during operation and up to 4 minutes after switching off since the turbo
pump can be damaged.
0-10Important Safety Precautions
kina22eIVZ.fmTechnical Handbook(1206)
1General Information
Caution
Caution
Caution
The Ecotec E3000 refrigerant leak detector is supplied ready for operation. However,
we recommend that you carefully read the Technical Handbook to ensure optimum
operating conditions right from the start. This handbook contains important
information on functions, installation, start-up and operation of the Ecotec E3000.
1.1Introduction
1.1.1Intended Use
The Ecotec E3000 is a refrigerant leak detector for sniffer applications. It may be
used to localise and quantify leaks in test samples if there is a refrigerant under an
overpressure within the test sample and when searching the test sample with a
sniffer probe from the outside (sniffer method). The use of this sniffer probe is
mandatory for proper operation and it is available as an accessory (Cat. No. 525-001
to 525-004).
The Ecotec E3000 must not be operated while standing in water or when exposed
to drip water. The same applies to all other kinds of liquids.
This Ecotec E3000 should only be used in rooms.
Avoid contact of the Ecotec E3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
Proper use includes:
•Conformance with the technical specifications
•Use of standard and original accessories
•Observance of the instructions and guidelines contained in this document
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-1
1.2Support by INFICON
INFICON Service
If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health or whether
it is contaminated.
If it is contaminated also indicate the nature of the hazard.
INFICON must return any equipment without a Declaration of Contamination to the
sender’s address. You will find an appropriate form at the next page.
General
We reserve the right to alter the design or any data given in this handbook.
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has
been submitted. Non-completion will result in delay.
This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Description of product
Type
Article Number
Serial Number
The product is free of any substances which are damaging to
health yes
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which
Trade/product name
Chemical nam e
(or symbol)
Reason for return
Operating fluid(s) used
(Must be drained before shipping.)
Process related contamination of product:
toxic no 1) yes
caustic no 1) yes
biological hazard no yes 2)
explosive no yes 2)
radi oactive no yes 2)
other harmful substances no 1) yes
1) or not containing any amount
of hazardous residues that
exceed the permissible ex posure limits
the product
Precauti ons associ ated
with substance
may have come into contact with:
2) Products thus contami nated will not be ac cepted without written
evidence of decontami nation!
Action if human contact
Legally binding declaration:
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may
arise. The contaminated p roduct will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
This form c an be downloaded
from our websit e.
www.inficon.com leakdetection.service@inficon.com
1-4General Information
Copies:
Original for addres see - 1 copy for acco mpanying docum ents - 1 copy for file of sen der
South America except Brazilinfoqro@meisa.comInficon Inc.
MEISA S.a. de C.V.
Querètaro
Phone: +52 442 225 42 80
Fax: +52 442 225 41 57
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +45 741 236 36
Fax: +45 744 336 46
Phone: +48 60 23 15 212
Fax: +48 60 23 15 212
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +27 73 15 78 355Agramkow
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +65.890.6250
Fax: +65.890.6266
Phone +420 734 331 758
Fax: +420 604 203 037
Phone: +386 15 63 91 50
Fax: +386 17 22 04 51
Phone: +52 44 22 12 36 15
Fax: +52 44 22 12 19 40
Leybold Optics Ibérica
Barcelona
Agramkow
Sonderborg
INFICON GmbH
Cologne
INFICON Company Limited
Chupei City, HsinChu Hsien
INFICON GmbH
Cologne
Sonderborg
INFICON GmbH
Cologne
INFICON
Blackburn
INFICON GmbH
Cologne
Inficon Inc.
East Syracuse, NY
San Jose, CA
Inficon Inc.
Austin, TX
Phone: +34 93 66 60 778
Fax: +34 93 66 64 612
Phone: +45 741 236 36
Fax: +45 744 336 46
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +886.3.5525.828
Fax: +886.3.5525.829
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +45 741 236 36
Fax: +45 744 336 46
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +44 1254 678 250
Fax: +44 1254 698 577
Phone: +49 221 56788-112
Fax: +49 221 56788-9112
Phone: +1.315.434.1167
Fax: +1.315.434.2551
Phone: +1.408.361.1200
Fax: +1.408.362.1556
Phone: +1.512.448.0488
Fax: +1.512.448.0398
1-6General Information
kina22e chapter 1.fmtechnical handbook(1206)
1.3Unpacking
Unpack the Ecotec E3000 leak detector immediately after it has been received even
if it is to be put into operation at some later date. Examine the shipping container for
any external damage. Completely remove all packaging materials.
Notice:Retain the shipping container and the packaging materials in the event of
possible complaints concerning any damages.
Check if the Ecotec E3000 leak detector is complete (see Section 1.3.1) and carefully
subject it to a visual inspection. If any damage is discovered please immediately
inform the forwarding agent and the insurers. If it is required to exchange the
damaged part please contact our orders department.
1.3.1Supplied Equipment
The Ecotec E3000 leak detector is ready for operation. Before installation please
read Section 1.5. Included with the leak detector are the following items:
•Ecotec E3000 (main unit)
•Mains cord, 3m long
•Set of fuses (3 x 10 pcs.)
•Spare air filter
•8 mm hexagonal wrench
•19 mm ring wrench
•Documentation
– Operating Instructions (kima22e1)
– Technical Handbook Ecotec E3000 (kina22e1)
– Spare Parts List Ecotec E3000 (kiua22d2)
– Interface Description E3000 (kins22e1)
Notice:The sniffer line is available in different configurations and needs to be
ordered separately in the desired length. The sniffer line is not part of the
Ecotec E3000 shipment. (see Section 1.3.3 Accessories)
Notice:The ECO-Check reference leak is an accessory (see Section 1.3.3
Accessories) and needs to be ordered separately
Notice:For the Ecotec E3000RC version the display unit and the connecting cable
are not part of the standard Ecotec E3000 shipment and need to be
ordered separately (see Section 1.3.3)
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-7
1.3.2Technical Data
Physical Data
Smallest detectable leak rate 0.05 g/a for refrigerants
Max. rel. humiditymax. 80% for temperatures up to +31°C,
decreasing linearly to 50% at +40%
Contamination levelII
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”)
Max. altitude above sea level2000m
kina22e chapter 1.fmtechnical handbook(1206)
1-8General Information
1.3.3Accessories
Sniffer line for Ecotec E3000Cat. No. / Ref. No.
SL3000-3, 3 m length525-001
SL3000-5, 5 m length525-002
SL3000-10, 10 m length525-003
SL3000-15, 15 m length525-004
Sniffer line for system integration (robot application)525-015
Sniffer tips
ST 312, 120 mm long, rigid122 13
FT 312, 120 mm long, flexible122 14
FT 200, 200 mm long, rigid122 18
FT 250, 250 mm long, flexible122 66
ST 385, 385 mm long, rigid122 15
FT 385, 385 mm long, flexible122 16
FT 600, 600 mm long, flexible122 09
ST 500, 500 mm long, rigid, 45° angled122 72
Water protection tip for sniffer122 46
Holder for SL3000 sniffer line525-006
ECO-Check reference leak for R134a531-001
Calibrated sniffer leaks for single refrigerants, 2 - 5 g/a
(16 g/a also available)
R134a122 20
R600a122 21
R404A122 22
R502a122 23
R22122 25
R23122 26
R152a122 27
R407C122 28
R410A122 29
R401A122 30
R290122 31
R744 (CO2)122 32
10% Hydrogen (E-5 mbarl/s)122 33
Halon 1301 (R13B1)122 34
HFO-1234yf122 35
Calibrated sniffer leak for helium
S-TL 4, approx.1 x 10
S-TL 5, range 10
kina22e chapter 1.fmtechnical handbook(1206)
S-TL 6, range 10
-4
mbar l/s122 37
-5
mbar l/s122 38
-6
mbar l/s122 39
General Information1-9
Calibrated leaks for forming gas (hydrogen)
10% hydrogen / 90% helium, range 10
-5
mbar l/s122 33
(Calibrated leaks for other refrigerants on request)
External display unit for Ecotec E3000RC
for bench top use551-100
for rack mounting551-101
Connecting cable for external display unit
for Ecotec E3000RC, 5m551-102
for Ecotec E3000RC, 1 m551-103
1-10General Information
kina22e chapter 1.fmtechnical handbook(1206)
1.4Notes on How to Use This Handbook
1.4.1Numbering of Figures
The references to diagrams. e.g. (2-1/6) consist of the Section No., Fig. No. and the
Item No. in that order. For example (2-1/6) means: Section 2, Fig. 1and Item No. 6
(here: mains switch).
1.4.2Symbols of Vacuum Technology
In the following some important symbols of vacuum technology as used in this
handbook are shown:
Vacuum pump in generalDiaphragm pump
Turbo molecular pumpVacuum gauge
1.4.3Definition of Terms
Main menu
This menu is shown first after operating the Menu push-button.
Sub-menus
Comprise all menus which may be accessed from the main menu. Unauthorized
changes to many of these sub-menus may be prevented by a password (see also
Section 4.5.1).
Menu item
A single menu line.
Default condition
Status of the Ecotec E3000 when supplied from the factory.
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-11
Service menu
Comprises the menu lines in the “Service” sub-menu. The service menu is accessed
by scrolling in the basic menu using the navigation push-buttons (see also Section
3.2).
Autozero
Determination and compensation of the refrigerant background. With this function,
the internal ZERO level of the leak rate signal is determined in order to avoid a
readout of the internal refrigerant background and mistaking it as a properly
measured value. If subsequently negative leak rates are obtained due to this
correction, the stored offset values are changed so that ZERO will be the lowest
value which can be obtained. In this way the values adapt automatically to a
decaying background (adaptive background correction).
Internal background
The existing partial pressure in the measurement system. The level of the internal
background is measured all the time and subtracted from the measured signal.
I•Guide Mode
Unit under test
Display limit
In the I•Guide Mode different testing plans can be pre-programmed. During testing
the operator is then constantly prompted for the next action and thus guided through
the testing plan.
Test object that needs to be leak checked.
Limits the measurement data displayed depending on the unit of measurement and
the operator settings.
1-12General Information
kina22e chapter 1.fmtechnical handbook(1206)
1.5Instrument Views of the Ecotec E3000
Caution
Fig. 1-2 Instrument views of Ecotec E3000
Pos. DescriptionPos. Description
1Main display4Lemo Connector for sniffer line
2Handle for carrying the Ecotec E30005Speaker
3ECO-Check reference leak
1.6Installation
1.6.1Set up
The weight of the Ecotec E3000 exceeds 25 kg.
It therefore should not be carried by one single person.
The Ecotec E3000 may tip off its base and injure people.
Place the Ecotec E3000 on a stable base.
How to remove the transportation lock:
The transportation lock is located on the bottom side of the Ecotec E3000 main unit
and consists of a yellow knurled screws. Please remove this screw before startingup the Ecotec E3000. The Ecotec E3000 is supplied ready for operation. Initial startup is described in Section 3.1.
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-13
Fig. 1-3 Removing the transportation lock before starting
Caution
Caution
Caution
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Ecotec E3000, a space of at least
20 cm (8 in.) must be kept unobstructed to the sides. The clearance at the rear must
be no less than 10 cm (4 in.). Moreover, the Ecotec E3000 handles for carrying the
leak detector at the sides of the main unit must not be covered at any time as these
acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity of the
Ecotec E3000.
Make sure that you can always reach the mains plug.
1-14General Information
kina22e chapter 1.fmtechnical handbook(1206)
1.6.2Mechanical Connections
ECO-Check
Reference leak (optional)
Please insert the ECO-Check reference leak into the opening in the housing of the
main unit. Make sure that the Sub-D plug is properly connected with the ECO-Check
leak.
Notice:When properly inserted, the ECO-Check reference leak will still protrude by
approx. 10 mm.
On first usage of your ECO-Check you need to initialize the use of this reference leak
in the Ecotec E3000 software menu.
Please perform the following steps:
1 Insert the ECO-Check into the appropriate opening of the Ecotec E3000
2 In the software menu go to HISTORY & MAINTENANCE / REPLACEECO-CHECK.
Notice:The Ecotec E3000 must be set to
submenu (See section 4.5.3). Go to
ADVANCEDMODE for access to this
SETTINGS / DISPLAY / USERMODE)
3 On the certificate, which is delivered with the ECO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the second line and press OK.
Notice:The ECO-Check reference leak must be installed in the Ecotec E3000
when pressing
Fig. 1-4 Initializing the ECO-Check reference leak
OK.
SL3000 Sniffer line
In order to operate the Ecotec E3000 it is essential for the sniffer line to be
connected. The connection for the sniffer line is located at the front of the Ecotec
E3000 left of the ECO-Check reference leak.
Insert the plug into the opening with the red dot on the plug and the slot in the front
cover aligned until the connector engages. To disconnect the plug, retract the
coupling and remove the probe’s line.
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-15
Holder for SL3000 sniffer line (optional)
An optional holder for the SL3000 sniffer line is available as cat.-no. 525-006. The
holder may be installed on the right or left side of the main unit (for right- or left
handed operators) as shown in Fig. 1-6.
Fig. 1-5 Usage of sniffer line holder
The installation is described in Fig. 1-6. There are two little slots on the front side at
the very top area of the blue square front of the main unit.
Hold the holder horizontally and then insert the two little hooks of the holder into the
two slots (either on the right or the left side).
With the hooks still inserted, let the holder flap down. It will automatically attach to
the metal front by the magnet on the backside of the holder.
Now insert the sniffer probe grip into the opening of the holder and let it sink down
until it rests in the holder.
Fig. 1-6 Installation of sniffer line holder
1-16General Information
kina22e chapter 1.fmtechnical handbook(1206)
Water protection tip (optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to
condensation after performance testing), we strongly recommend to use a water
protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and
2 install the water protection tip instead.
Notice:Please do not forget to re-install the little rubber seal when switching to the
water protection tip.
Fig. 1-7 Installing water protection tip
ECO-Check Reference leak (optional),
for helium and hydrogen only
If detecting helium or hydrogen, a ECO-Check reference leak may be used for
internal calibration. The ECO-Check reference leak has to be inserted to the Ecotec
E3000 (see section 2.3.3).
On first usage of your ECO-Check you need to initialize the use of this reference leak
in the Ecotec E3000 menu.
Please perform the following steps:
1 Insert the ECO-Check into the appropriate opening of the Ecotec E3000
2 In the software menu go to HISTORY & MAINTENANCE /REPLACEECO-CHECK.
Notice:The Ecotec E3000 must be set to
submenu (See section 4.5.3). Go to
3 On the certificate, which is delivered with the ECO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the second line and press OK.
USERMODE „ADVANCED“ for access to this
SETTINGS / DISPLAY / USERMODE.
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-17
For Ecotec E3000RC only
Warning
1
23456
The Ecotec E3000RC has no built-in display unit but a connectors plate is mounted
instead. Please connect the external display unit with the 5 m connecting table (Cat.no. 551-102).
Fig. 1-8 Ecotec E3000RC with external display unit for: (a) bench top use (left side),
(b) rack mounting (right side)
1.6.3Electrical Connections
Fig. 1-9 Electrical connections
Pos. DescriptionPos. Description
1Headphone port4Name plate
2I/O Port5Power switch
3RS232 interface6Power connector
Notice:The local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Ecotec E3000
can be read off from the name plate left of the power switch. The mains
voltage setting of the Ecotec E3000 is fixed and can not be changed. A
separate fuse for each of the mains conductors has been integrated into
the mains socket (Fig. 2-1/6).
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Ecotec E3000 with the ground conductor unconnected is not
permissible.
kina22e chapter 1.fmtechnical handbook(1206)
1-18General Information
The mains voltage is applied to the Ecotec E3000 via the detachable mains cable
Caution
which is supplied with the Ecotec E3000. A main power socket is available for this
purpose at the rear of the Ecotec E3000.
Before connecting the Ecotec E3000 to the mains you must make sure that the
mains voltage rating of the Ecotec E3000 coincides with the locally available mains
voltage.
1.6.4RS232 Interface
The Ecotec E3000 is equipped with a RS232 interface which is located on the rear
right side of the main unit. This interface is of the DCE type (Data Communications
Equipment) and allows the connection of a PC for monitoring and data logging. The
connection is provided through a commercially available Sub-D plug. For further
information see “Interface Description Ecotec E3000” (kins22e1).
1.6.5I/O Port
The I/O part allows communication with external equipment via analog data. For
details see Section 7.2.
Through this connection some functions of the Ecotec E3000 can be controlled
externally or measurement data or the Ecotec E3000 status may be communicated
to external equipment.
Through relay changeover contacts the trigger levels as well as the operating mode
(Ready) of the Ecotec E3000 may be monitored.
kina22e chapter 1.fmtechnical handbook(1206)
General Information1-19
1-20General Information
kina22e chapter 1.fmtechnical handbook(1206)