IDEAL 61-763, 61-765 TECHNICAL MANUAL

Page 1
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763
61-765
The Service Information provides the following information:
• Specifications
• Performance test procedure
• Calibration and calibration adjustment procedure
• Basic maintenance (replacing the battery)
Form number: TM61763-5 Revision: 1.3 Date: September 2010
Form number TM61763-5 Rev 1.3 September 2010
Page 2
TABLE OF CONTENTS
Title
Page
Introduction
1
Precautions and Safety Information
1
Symbol Table
1
TightSight™ Display Notes
1
High Voltage Warning (HI-V)
2
Safety Information
2
Certifications and Compliance
3
Specifications
3
General Specification
3
Voltage Specifications
4
Current Specifications
4
Resistance Specifications
4
Frequency Specification
4
Capacitance Specifications
4
Continuity Specifications
4
Performance Verification
5/6
Calibration
7/8
Replacing the Battery
9
Form number TM61763-5 Rev 1.3 September 2010
Page 3
Symbol
Description
Symbol
Description
Battery
HI-V
High Voltage Indicator >30 V indicator is on ( on VAC,VDC ranges)
Cautionary or important information in manual
Continuity indicator Danger- Risk of electrical shock
Double Insulation- Protection Class II
CAT III
IEC Over-voltage Category III
CAT IV
IEC Over-voltage Category IV
Introduction
Warning
To avoid shock or injury, do not perform the verification
tests or calibration procedures described in this manual
unless you are qualified to do so.
The information provided in this document is for the use
of qualified personnel only.
Caution
The 61-760 series contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
For additional information about IDEAL INDUSTRIES, INC. and its products,
and services, visit IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the meter only as described in the Users Manual. If you do not do so, the protection provided by the meter may be impaired. Read the “Safety Information” page before servicing this product. In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution identifies conditions and actions that may damage the meter or the test instruments.
The Symbols
The symbols used on the meter and in this manual are explained in Table A.
Table A Symbols
Page 1
TightSight™ Display Notes:
Only AC/DC amps units of measure are displayed in the TightSight™ display since primary use is
for viewing current measurements in tight locations. The display will show numerical values only for other functions. The main display is to be used to view units of measure for all other functions.
Form number TM61763-5 Rev 1.3 September 2010
Page 4
Page 2
High Voltage Warning (HI-V):
The meter beeps and lights an LED when >30V AC on ACVoltage function or >30V DC on DC voltage function is present. This enhanced safety feature alerts the user when dangerous voltage is present. Notes: This feature does not work through the clamp head as the clamp is intended to only measure current. Audible indication can be turned off by pressing and holding the MAX/MIN button while turning the function switch from OFF to the desired range position.
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION.
These statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING.
These statements identify conditions or practices that could result in personal injury or loss of life.
Specific precautions
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or damp conditions.
Form number TM61763-5 Rev 1.3 September 2010
Page 5
Certifications and Compliances
Safety
Designed to UL 61010-1 second edition, IEC 61010-2-032 second
edition,IEC61010-031 specifications
Input rating 1000V DC Category III
600V DC Category IV
750V AC Category III
600V AC Category IV
Over voltage category
CAT IV: Outside and service entrance.
CAT III: Distribution level mains, fixed installation.
CAT II: Local level mains, appliances, and portable equipment.
CAT I: Signal level, special equipment or parts of equipment, telecommunication, electronics.
Characteristics
Description
Display
4 Digit LCD display
Display Count
9999 count, maximum reading 9999
Over range Indication
“OL” is displayed
Display up-date rate
2.0 time/second
Operating Relative Humidity
0 C to 50 C (32 F to 122 F) 0 ~ 70% RH
Storage Environment:
-20 C to 60 C (-4 F to 140 F) at <80% relative humidity
Power source:
9V Battery (NEDA 1604)
Battery Life:
100 hours typical (alkaline) {61-763} 100 hours typical (alkaline) {61-765}
Low Battery Indicator:
symbol indicates low battery voltage
Auto power off
Approximately 30 minutes
Dimensions
8.7” H X 3.1 ” W X 1.8 ” D 222mm H X 80mm W X 45.5mm D
Maximum Cable Size
ACA 1¼” (36mm)
Weight:
Approximately 12.4 oz. or 350g including battery
General Specifications
Page 3
RANGES and ACCURACY SPECIFICATION Accuracy: Accuracy specifications at 23°C ±5°C (73.4°F ±9°F) less than 75% RH. Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
from 0 C to 18 C and 28 C to 50 C (32 F to 64 F and 82 F to 122 F) Electrical Specification: Accuracy is ±(reading plus number of digits) at 23°C ±5°C (73.4°F ±9°F) <75% RH
Form number TM61763-5 Rev 1.3 September 2010
Page 6
Page 4
Function / Range
Ranges
Accuracy
AC Voltage (True RMS) (45-400Hz)
0.0V ~ 600.0V, 45Hz - 100Hz
1.0% + 5 digits
600.0V ~ 750.0V, 45Hz - 100Hz
1.5% + 5 digits
0.0V ~ 750.0V, 100Hz - 400Hz
6.0% + 5 digits
DC Voltage
0.0V ~ 600.0V
1.0% + 5 digits
600.0V ~ 999.9V
1.5% + 5 digits
AC Current (True RMS) (45-400Hz)
0.0A ~ 660.0A, 45Hz -60Hz
2.0% + 5 digits
0.0A ~ 660.0A, 60Hz - 400Hz
6.0% + 5 digits Capacitance
0.0µF ~ 999.9uF
5.0% + 10 digits
1
Frequency
20.0Hz ~ 400.0Hz (Auto-ranging)
0.5% + 5 digits
Resistance
0.0 ~ 999.9 , 1000 ~ 9999
1.5% + 5 digits
Continuity
<30 on Continuity
Not specified
Function / Range
Ranges
Accuracy
AC Voltage (True RMS) (20-400Hz)
0.0V ~ 100.0V, 20Hz - 100Hz
1.0% + 5 digits
100.0V ~ 600.0V, 45Hz - 100Hz
1.0% + 5 digits
600.0V ~ 750.0V, 45Hz - 100Hz
1.5% + 5 digits
0.0V ~ 750.0V, 100Hz - 400Hz
6.0% + 5 digits
DC Voltage
0.0V ~ 600.0V
1.0% + 5 digits
600.0V ~ 999.9V
1.5% + 5 digits
AC Current (True RMS) (20-400Hz)
0.0A ~ 660.0A, 20Hz - 100Hz
2.0% + 5 digits
0.0A ~ 660.0A, 100Hz - 400Hz
6.0% + 5 digits DC Current
0 ~ 660.0A
2.0% + 5 digits
Capacitance
0.0µF ~ 999.9uF
5.0% + 15 digits
1
Frequency
20.0Hz ~ 400.0Hz (Auto-ranging)
0.5% + 5 digits
Resistance
0.0 ~ 999.9 , 1000 ~ 9999
1.5% + 5 digits
Continuity
<30 on Continuity
Not specified
61-763
61-765
AC and DC Volts Input Impedance: 1M
AC Converter: 61-763, 61-765 –True RMS sensing Crest factor: 1.5 at full scale, and 3 at half scale
Overload Protection:
AC and DC Voltage: Not to exceed 1000V DC or 750VAC RMS
AC Current: Not to exceed 660A AC DC Current: Not to exceed 660A DC
Resistance: Not to exceed 600V DC or VAC RMS Capacitance, Frequency, Continuity: Not to exceed 600V DC or VAC RMS
1
Frequency Sensitivity: 5V rms on AC VOLTS range
Frequency can be measured through clamp head. ( 5AAC at 20 to 100Hz ) ( 10AAC at 100 to 400Hz ),
Form number TM61763-5 Rev 1.3 September 2010
Page 7
Frequency Overload protection: 1000AAC, 1000VDC or 750VAC rm
Function /Range
Input
Low Limit
High Limit
Model Number
V AC 750V
120V AC @ 45Hz
118.3
121.7
61-763, 61-765
V AC 750V
120V AC @ 400Hz
112.3
127.7
61-763, 61-765
V AC 750V
700V AC @ 45Hz
689.0
711.0
61-763, 61-765
V AC 750V
700V AC @ 400Hz
657.5
742.5
61-763, 61-765
Function /Range
Input
Low Limit
High Limit
Model Number
V DC 1000V
350V
346.0
354.0
61-763, 61-765
V DC 1000V
900V
886.0
914.0
61-763, 61-765
Function /Range
Input
Low Limit
High Limit
Model Number
A AC 660A
10A AC @ 50Hz
9.3
10.7
61-763, 61-765
A AC 660A
10A AC @ 400Hz
8.9
11.1
61-763, 61-765
A AC 660A
500A AC @ 50Hz
489.5
510.5
61-763, 61-765
A AC 660A
500A AC @ 400Hz
469.5
530.5
61-763, 61-765
Function /Range
Input
Low Limit
High Limit
Model Number
A DC 660A
100A DC
97.5
102.5
61-765
A DC 660A
500A DC
489.5
510.5
61-765
Function /Range
Input
Low Limit
High Limit
Model Number
1000
100
98.5
101.5
61-763, 61-765
10000
1K
985
1015
61-763, 61-765
Page 5
PERFORMANCE VERIFICATIONS
Perform the following analysis; if the meter conforms to the limits listed in Table 1 through 7 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed.
Performance Verification Preparation
1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization
should be reached after 30 minutes.
2. Remove battery cover and use a calibrated meter to ensure the battery measures a minimum of 7.5V
DC. If the battery measures under 7.5V DC, replace the battery (see Battery Replacement page 11) before beginning the performance test.
3. Input the values listed in Table 1 through 7.
Table 1 AC Voltage Test
Table 2 DC Voltage Test
Table 3 AC Current Test
Table 4 DC Current Test
Table 5 Resistance Test
Form number TM61763-5 Rev 1.3 September 2010
Page 8
Function /Range
Input
Low Limit
High Limit
Model Number
Hz {through clamp}
50Hz @ 5A
49.70
50.30
61-763, 61-765
Hz {auto} {through clamp}
200Hz @ 10A
198.5
201.5
61-763, 61-765
Hz {V and COM jacks}
60Hz @ 120V
59.2
60.8
61-763, 61-765 MFD 500µF
100µF
94.0
106.0
61-763, 61-765
Function /Range
Test Value
Low limits
High Limit
Model Number
Continuity
20 beep on
61-763, 61-765 40 beep off
Page 6
Table 6 Capacitance and Frequency Test
Table 7 Continuity Check
Form number TM61763-5 Rev 1.3 September 2010
Page 9
Page 7
CALIBRATION
Calibration Preparation
1. Turn on the calibrator, allow calibrator to warm up. Perform calibration at
23±2°C (73.4°F ±3.5°F) at relative humidity of < 70%. Temperature stabilization should be
reached after 30 minutes.
2. Disconnect the test leads and turn the range switch to “OFF”.
3. It is recommended that a new battery be installed before you begin calibration. To replace the battery (see Battery Replacement page 9).
Calibration Procedure
It is recommended that all IDEAL meters undergo the following calibration procedure on an annual basis. The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the accuracy of this instrument.
Auto Calibration:
Set the short circuit 2-pin connector, mounted on PC board, from J2 to J1 position. Replace the case bottom of the meter. It is allowed to do calibration for any one of the following functions that is inaccurate. There is no need to do calibration for all functions.
V AC Calibration
1. Set Function Switch from OFF to ACV position.
2. Short the V and COM inputs on the meter. After a reading in the LCD is settled (about 5 seconds),
press MAX/MIN button until unit display reads 000.0V ± 1 digit. Zero adjustment is complete.
3. Connect the calibrator to the V and COM inputs on the meter.
4. Output AC 50V/50Hz. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 48V ~ 50V. Adjustment is complete.
5. Output AC 500V/50Hz. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 500.0V ± 2 digits. Adjustment is complete.
V DC Calibration
1. Set the function switch to DCV range. After a reading in the LCD is settled (about 5 seconds), press
MAX/MIN button until unit display reads 000.0 ± 1 digit. Zero adjustment is complete.
2. Connect the calibrator to the V and COM inputs on the meter.
3. Output DC +500V. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 500.0V ± 2 digits. Adjustment is complete.
Continuity Calibration
1. Set the function switch to Continuity range. Short the V and COM inputs on the meter. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads
000.0 ± 1 digit. Zero adjustment is complete.
2. Connect the calibrator to the V and COM inputs on the meter.
3. Output 800.0 . After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button
until unit display reads 800.0 ± 2 digits. Adjustment is complete.
Form number TM61763-5 Rev 1.3 September 2010
Page 10
Page 8
Resistance Calibration
1. Set the function switch to range. Short the V and COM inputs on the meter. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 000.0 ± 1 digit. Zero adjustment is complete.
2. Connect the calibrator to the V and COM inputs on the meter.
3. Output 800.0 . After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 800.0 ± 2 digits. Adjustment is complete.
4. Output 8K . After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 8000 ± 2 digits. Adjustment is complete.
Capacitance Calibration
is settled (about 5 seconds), press MAX/MIN button until unit display reads 000.0 ± 1 digit. Zero adjustment is complete.
2. Plug a 390uF standard capacitor to V and COM inputs. Press MAX/MIN button until unit display
reads 390.0 ± 5 digits. Adjustment is complete.
A DC Calibration (61-765 only)
1. Set the function switch to DCA range. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 000.0 ± 1 digit. Zero adjustment is complete.
2. Set output of the DC calibrator for 7.8A DC and connect it to Coil = 50N = 390.0A DC.
3. Clamp the jaws to the coil = 50N.
4. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 390.0A ± 2 digits. Adjustment is complete.
A AC Calibration
1. Set the function switch to ACA range. When a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 000.0 ± 1 digit. Zero adjustment is complete.
2. Set output of the AC calibrator for 1.0A/50Hz and connect it to Coil = 50N = 50.0A AC.
3. Clamp the jaws to the coil = 50N.
4. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads about 48A ~ 50A. Adjustment is complete.
5. Set output of the AC calibrator for 7.8A AC/50Hz and connect it to Coil = 50N = 390.0A AC.
6. Clamp the jaws to the coil = 50N.
7. After a reading in the LCD is settled (about 5 seconds), press MAX/MIN button until unit display reads 390.0A ± 2 digits. Adjustment is complete.
Calibration of the 61-760 series is complete. Remove all leads from the calibrator and equipment. Open the case bottom again, and set the short circuit 2-pin connector on PC board from J1 back to J2 position, then replace the case bottom of the meter. Turn the Function Switch from OFF to any position to restart the meter.
Form number TM61763-5 Rev 1.3 September 2010
Page 11
Battery Replacement (Refer to Figure 1)
1. Disconnect the test leads from any circuit under test and turn off meter.
2. Use a Philips head screwdriver to remove the screws on battery cover.
3. Remove battery from the battery compartment.
4. Install new 9V battery (NEDA #1604). An alkaline type is recommended.
5. Install new battery into compartment using care to install to proper polarity.
6. Reinstall battery cover.
Page 9
Figure 1
Form number TM61763-5 Rev 1.3 September 2010