IDEAL INDUSTRIES, INC.
TECHNICAL MANUAL
MODELS: 61-732
61-736
The Service Information provides the following in formation:
• Precautions and safety information
• Specifications
• Performance test procedure
• Calibration and calibration adjustment procedure
• Basic maintenance (replacing the battery)
HOLD
NCV
Ω
A
OFF
COM
V
V
61-736TRUE RMS
V/Ω
CAT.III
600V
MAX
600V
Form number: TM61732-6
Revision: 5. Date: Oct 2004
Form number TM61732-6 Rev 5 October 2004
Page 2
TABLE OF CONTENTS
Title
Page
Introduction 1
Precautions and Safety Information 1
Symbol Table 1
Safety Information 2
Certifications and Compliance 2
Specifications 3
General Specification 3
Voltage Specifications 4
Current Specifications 4
Resistance Specifications 4
Continuity Specifications 4
Performance Verification 5/6
Calibration 6/7
Calibration Reference Points 8
Replacing the Batteries 9
Form number TM61732-6 Rev 5 October 2004
Page 3
Introduction
Warning
To avoid shock or injury, do not perform the verification
tests or calibration procedures described in this manual
unless you are qualified to do so.
The information provided in this document is for the use
of qualified personnel only.
Caution
The 61-730 series contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
For additional information about IDEAL INDUSTRIES, INC. and its products,
and services, visit IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the meter only as described in the Users Manual. If you do not
do so, the protection provided by the meter may be impaired.
Read the “Safety Information” page before servicing this product.
In this manual, a Warning identifies co nditions and actions that
pose hazard (s) to the user; a Caution identifies conditions and
actions that may damage the meter or the test instruments.
The Symbols
The symbols used on the meter and in this manual are explained
in Table A.
Table A Symbols
Symbol Description Symbol Description
CAT III
Battery
Cautionary or important
information in manual
Danger- Risk of electrical
shock
Double Insulation- Protection
Class II
IEC Over-voltage Category III
NCV
Non-Contact indicator
Continuity indicator
Page 1
Form number TM61732-6 Rev 5 October 2004
Page 4
Page 2
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION.
These statements identify conditions or practices that could result in damage to the equipment or other
property.
WARNING.
These statements identify conditions or practices that could result in perso nal injury or loss of life.
Specific precautions
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
Certifications and compliances
Safety
Input rating
Designed to EN 61010-1, EN 61010-2-032, UL61010B-1,
UL 61010B-2-032 specifications
600V DC Category III
600V AC Category III; Clamp rated at 300V AC Category III only
CAT III: Distribution level mains, fixed installation.
Over voltage category
CAT II: Local level mains, appliances, and portable equipment.
CAT I: Signal level, special equipment or parts of
equipment, telecommunication, electronics.
Form number TM61732-6 Rev 5 October 2004
Page 5
General specifications
Characteristics Description
Display
3¾ Digit LCD display
Display Count 4000 count, maximum reading 3999
Over range Indication “OL” is displayed
Sampling Rate 2.0 times/second
Operating Environment:
Relative Humidity
Storage Environment:
0°C to 50°C (32°F to 122°F)
0 ~ 70% RH
-20°C to 60°C (-4°F to 140°F) at <80% relative humidity
Power source: 1.5V Battery x 2 (R03/Size AAA)
Battery Life: 400 hours typical (alkaline) {61-732}
250 hours typical (alkaline) {61-736}
Low Battery Indicator:
symbol indicates low battery voltage
Auto power off Approximately 10 minutes
Dimensions 7.4” H X 2.6” W X 1.5” D
187mm H X 65mm W X 37mm D
Maximum Cable Size ACA 1¼” (32mm)
Weight: Approximately 6.7 oz. or 190g including battery
RANGES and ACCURACY SPECIFICATION
Accuracy: Accuracy specifications at 23°C ±5°C (73.4°F ±9°F) less than 75% RH.
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F)Electrical Specification: Accuracy are ±(reading plus number of digits) at 23°C ±5°C
(73.4°F ±9°F) <75% RH
AC Converter: 61-732 is Average responding, RMS Calibrated to Sine Wave.
61-736 is True RMS responding.
Overload Protection:
AC and DC Voltage: Not to exceed 600V DC or VAC RMS.
AC Current: Not to exceed 400A AC.
Resistance: Not to exceed 600V DC or VAC RMS
Continuity: Not to exceed 600V DC or VAC RMS
Form number TM61732-6 Rev 5 October 2004
Page 7
Page 5
PERFORMANCE VERIFICATIONS
Perform the following analysis; if the meter conforms to the limits listed in Table 1 through 5 the meter is
functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure
must be performed.
Performance Verification Preparation
1. Turn on th e calibrator, allow calibrator to warm up. Temperature stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the batteries measures a minimum of
2.4VDC. If the batteries measures under 2.4V DC, replace the batteries (see Battery Replacement on page
9) before beginning the performance test.
3. Input the values listed in Table 1 through 5.
Table 1 AC Voltage Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
V AC 400V 350V AC @ 50Hz 345.3 354.7 61-732
V AC 400V 350V AC @ 50Hz 345.0 355.0 61-736
V AC 400V 350V AC @ 500Hz 345.3 354.7 61-732
V AC 400V 350V AC @ 500Hz 345.0 355.0 61-736
V AC 600V 500V AC @ 50Hz 487 513 61-732
V AC 600V 500V AC @ 50Hz 484 516 61-736
V AC 600V 500V AC @ 500Hz 487 513 61-732
V AC 600V 500V AC @ 500Hz 484 516 61-736
Table 2 DC Voltage Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
V DC 400V 350V 348.0 352.0 61-732, 61-736
V DC 600V 500V 495 505 61-732, 61-736
Table 3 AC Current Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
A AC 40A 10A AC @ 50Hz 9.77 10.23 61-732
A AC 40A 10A AC @ 50Hz 9.73 10.27 61-736
A AC 40A 10A AC @ 400Hz 9.64 10.36 61-732
A AC 40A 10A AC @ 400Hz 9.60 10.40 61-736
A AC 400A 100A AC @50Hz 97.7 102.3 61-732
A AC 400A 100A AC @50Hz 97.3 102.7 61-736
A AC 400A 100A AC @400Hz 96.4 103.6 61-732
A AC 400A 100A AC @400Hz 96.0 104.0 61-736
1. Turn on the calibrator, allow calibrator to warm up. Perform calibration at
23±2°C (73.4°F ±3.5°F) at relative humidity of < 70%. Temperature stabilization should be
reached after 30 minutes.
2. Disco nnect the test leads and turn the range switch to “OFF”.
3. Remove the two screws holding the case bottom.
4. Remove the case bottom using care not to damage the battery connector.
5. Using a calibrated meter ensure the batteries measures a minimum of 2.4V DC.
If the batteries measures under 2.4V DC, replace the batteries (see Battery Replacement on page 9).
Calibration Procedure
It is recommended that all IDEAL meters undergo the following calibration procedure on
an annual basis.
The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the
accuracy of this instrument.
Form number TM61732-6 Rev 5 October 2004
Page 9
Page 7
V DC Calibration
61-732 (Refer to Figure 1A), 61-736 (Refer to Figure 2A)
1. Set the function / range to 400mV DC.
2. Connect the calibrator to the V and COM inputs on the meter.
3. Output 390.0mV DC.
Adjust VR1 (VR 1KΩ) until unit display reads 390.0
4. De-energize source and remove test leads.
V AC Zero Calibration
61-736 (Refer to Figure 2A)
1. Set the function / range to 600V AC.
2. Short the V and COM inputs on the meter.
3. Adjust VR4 (VR 220K) until unit display reads 000.
4. De-energize source and remove test leads.
V AC Calibration:
61-736 only (Refer to Figure 2A)
1. Set the function/range to the 400V AC.
2. Connect the calibrator to th e V and COM inputs on the meter.
3. Output 390.0VAC/50Hz
Adjust VR3 (VR 200KΩ) until unit display reads 3900V± 1 digit.
4. De-energize source and remove test leads.
A AC Calibration:
61-732 (Refer to Figure 1B), 61-736 (Refer to Figure 2B) (Adjustments made under front panel label.)
1. Set the function / range to the 400A AC.
2. Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A
AC.
3. Clamp the jaws to the coil = 10N.
4. Adjust VR2 (VR 2KΩ) for a display reading of 100.0 ±1 digit
5. De-energize source and remove test leads.
Calibration of the 61-730 series is complete.
Remove all leads from the calibrator and equipment.
Return unit to proper operating condition.
Form number TM61732-6 Rev 5 October 2004
Page 10
Page 8
Figure 1A (61-732) Figure 1B (61-732)
Figure 2A (61-736) Figure 2B (61-736)
Form number TM61732-6 Rev 5 October 2004
Page 11
Battery Replacement (refer to Figure 3)
1. Disconnect the test leads from any circuit under test and turn off m e ter.
2. Use a Philips head screwdriver to remove the screws on battery cover.
3. Remove batteries from the battery compartment.
4. Install new 1.5V batteries (R03/Size AAA). Alkaline type is recommended.
5. Install new batteries into compartment using care to install to proper polarity.
6. Reinstall battery cover.
Page 9