IDEAL 61-610 User Manual

Page 1
#61-610
Analog Multimeter
Instruction Manual
WARNING
Read First: Safety Information
Please read and follow operating instructions carefully. Misuse or abuse of this multimeter cannot be prevented by any printed word. Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.
WARNING
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use if meter appears damaged. Visually inspect the meter to ensure case is not cracked and back case is securely in place.
• Inspect and replace leads if insulation is damaged, metal is exposed, or probes are cracked. Pay particular attention to the insulation surrounding the connectors.
• Do not use meter if it operates abnormally as protection maybe impaired.
• Do not use during electrical storms or in wet weather.
• Do not use around explosive gas, dust, or vapor.
• Do not apply more than the rated voltage to the meter.
• Do not use without the battery and the back case properly installed.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 1
Page 2
• Before each use:
- Zero out the test leads in the resistance mode to verify the functionality of
the battery and test leads.
- Verify meter operation by measuring a known voltage. Apply meter to
circuit under test. And, test on a known live voltage again to ensure proper operation.
• Remove test leads from input jacks prior to removing back of case.
• Only replace the fuse with specified volt, amp, and rupture speed ratings.
CAUTION
To protect yourself, think "Safety First":
• Voltages exceeding 30VAC or 60VDC pose a shock hazard so use caution.
• Use appropriate personal protective equipment such as safety glasses, face shields, insulating gloves, insulating boots, and/or insulating mats.
• Use the 3 Point Safety Method.
• Never ground yourself when taking electrical measurements.
• Connect the black common lead to ground before applying the red test lead to voltage. Disconnect the red test lead from the voltage first.
• Always work with a partner.
• When using the probes, keep fingers as far behind the probe tips as possible.
• Do not expose meter to mechanical shock, vibration, extreme temperature, or strong magnetic fields.
Features:
• Measures AC/DC Voltage, DC Current, Resistance, and Battery Check
• Mirrored scale to reduce parallax error
• Fuse protected
• Diode protected meter movement
• Pocket-sized, lightweight, high-impact thermoplastic case
• Recessed input terminals for added safety.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 2
Page 3
WARNING: Before taking any measurements, always carefully set the
rotary switch to the highest range in the measurement function desired and then switch to lower ranges until the pointer reads in the upper half of the scale.
MEASUREMENT PROCEDURE:
This high quality analog multimeter provides a long field life by following the steps outlined below:
• Best Accuracy
- Ensure that the pointer is set exactly on the zero line located at the left
end of the arc using the Zero Dial. Note that changes in atmospheric conditions can cause the pointer to drift daily.
- The pointer should read in the upper half of the scale, preferably in the
fourth quarter of the arc.
• Switch position
1) Prior to use, set the switch position to the highest range of the function
to be measured.
2) Adjust the switch down to lower successive ranges in this function until
the pointer reads in the upper half of the scale. By using this method routinely, the meter is protected against overloads that may "peg the needle" and cause damage to the pointer, particularly in cases where circuit values are not known.
WARNING: When finished with taking measurements of one function,
remove the test leads completely from the circuit, before switching to another function and repeating the measurement procedure above.
To Measure AC Voltage:
• Set the rotary switch to the highest range (500) in the ACV function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the
black test lead into the – input jack on the meter.
• Hold the test leads behind the finger guards and apply the lead tips in
parallel to the circuit under test.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 3
Page 4
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy and
prevent over-ranging of the pointer.
• Read the AC voltage measurement using the AC V labeled red arc (third arc
from the top).
CAUTION: For maximum safety, do not hold the meter in your hands while measuring higher voltages using the 250 and 500 ACV ranges.
To Measure DC Voltage:
• Set the rotary switch to the highest range (500) in the DCV function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the
black test lead into the – input jack on the meter.
• Hold the test leads behind the finger guards and apply the lead tips in
parallel to the circuit under test.
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy and
prevent over-ranging of the pointer.
• Read the DC voltage measurement using the DC labeled black arc (second arc
from the top).
CAUTION: For maximum safety, do not hold the meter in your hands while measuring higher voltages using the 250 and 500 DCV ranges.
CAUTION: Discharge all capacitors and inductors before taking current measurements.
To Measure DC milli-amps:
• Set the rotary switch to the 250m range in the DCA function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the
black test lead into the – input jack on the meter.
• Hold the test leads behind the finger guards and apply the lead tips in series
with the circuit under test.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 4
Page 5
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy.
• Read the DC mA measurement using the mA labeled black arc (second arc
from the top).
WARNING: Verify that the circuit is completely de-energized (dead) by measuring the circuit voltage to ensure that zero volts (AC/DC) is present before taking any resistance measurements.
Exposing live voltage to the Ohms circuit by misuse or abuse may result in injury to the operator and damage to the meter and equip­ment under test. The extent of injury and damage will depend on the level of voltage and current in the circuit.
To Measure Resistance (Ohms):
• Verify the circuit is completely de-energized.
• Set the rotary switch to the highest range (Rx100) in the OHM function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the black test lead into the – input jack on the meter.
• Short the red and black test lead tips together and adjust the Zero Dial until the pointer is directly over the zero position on the right end of the Ω labeled green arc (top arc on the display). Note: If the pointer cannot be zeroed, then the battery should be replaced.
• Apply the test leads to the component or dead circuit under test.
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy.
• Read the ohms measurement using the Ω labeled green arc (top arc on the display).
To Perform a Battery Check:
• Set the rotary switch to the 1.5V or 9V range on the B.TEST function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the black test lead into the – input jack on the meter.
• Apply the red test lead tip to the positive (+) battery contact and the black test lead tip to the negative (-) battery contact.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 5
Page 6
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy.
• Read the pointer position on the Bat labeled arc at the bottom of the display to determine if the battery is Good or needs to be Replaced. Note: The meter may not be able to check some 1.5VDC button cell batteries.
To Measure Decibels (dB):
• Set the rotary switch to the 10 range in the ACV function.
• Insert the red test lead fully into the + input jack on the meter and fully seat the black test lead into the – input jack on the meter.
• Hold the test leads behind the finger guards and apply the lead tips to the cir­cuit under test.
• Follow the MEASUREMENT PROCEDURE on page 4 to maximize accuracy and prevent over-ranging of the pointer.
• Read the decibels measurement using the dB labeled red arc (fourth arc from the top).
• For measurements above the basic dB range, set the switch to next highest range on the ACV function and add dB correction as shown on the table in the lower right of the scale plate on the meter. Note: To remove the DC components in the circuit under test, connect a 0.1µF capacitor, or larger, in series with one of the test leads before testing decibels.
MAINTENANCE Battery Replacement:
• Remove test leads from meter.
• Remove the screw from the back of the case.
• Remove the back of the case.
• Replace battery with a 1.5V AA type battery.
• Assemble the back case on the meter and re­tighten the screw.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 6
Fuse Replacement:
• Remove test leads from meter.
• Remove the screw from the back of the case.
• Remove the back of the case.
• Replace fuse with same fuse type, size, and ratings (IDEAL #F-16).
• Assemble the back case on the meter and re-tighten the screw.
Replacement test leads: IDEAL #TL-34.
Maintenance:
• Clean the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Service, and Replacement Parts:
For replacement parts or to inquire about service information contact IDEAL INDUSTRIES, INC. at 1-877-201-9005 or visit our website www.testersandmeters.com.
Specifications:
Sensitivity: 2kΩ/V AC/DC Fuse: (1) 0.25A, 250V, 5mm x 20mm, fast-acting fuse
(IDEAL #F-16) Battery: (1) 1.5V AA type battery. Operating: 32 F to 122 F (0 C to 50 C) Environment Maximum Relative Humidity <80% @ 31 C /
Minimum 50% @ 40 C Altitude: 6561.7 ft. (2000m) Dimensions: 4"H x2.9"W x0.9"D (100mm H x63mm W x34mm D) Weight: 3.6 oz (102g) Safety: Cat II-500V Accessories Test Leads (#TL-34), (1) 1.5V AA type battery, operating included: instructions.
Page 7
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 7
Ranges & Accuracies:
AC Voltage: 10/50/250/500 ACV +/-4% FS DC Voltage: 10/50/250/500 DCV +/-3% FS DC Current: 250 mADC +/-4% FS Resistance: 5/50/500 Ω Mid-scale +/-3% of Arc Length
1k/10k/100k Ω Full-scale +/-3% of Arc Length Decibels: -20 to +56 dB on ACV ranges Battery Check: 1.5V / 9V ranges
Double Insulation
Instrument has been evaluated and complies with insulation category II (over­voltage category II). Pollution degree 2 in accordance with IEC-644. Indoor use.
Warranty Statement:
This tester is warranted to the original purchaser against defects in material and work­manship for five years. During this warranty period, IDEAL INDUSTRIES, INC. will, at it’s option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument.
Any implied warranties arising out of the sale of an IDEAL product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.
State laws vary, so the above limitations or exclusions mat not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary form state to state.
Page 8
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 8
Page 9
#61-610
Multímetro analógico
Manual de Instrucciones
ADVERTENCIA
Lea primero: Información de seguridad
Por favor, lea y siga cuidadosamente las i nstrucciones de operación. Ninguna palabra impresa puede prevenir el mal uso o abuso de este multímetro. Utilice el medidor sólo según se especifica en este manual; de lo contrario, la protección que él proporciona podría deteriorarse.
ADVERTENCIA
• Siga estas instrucciones para evitar una posible descarga eléctrica o lesiones personales:
• No utilice el medidor si se lo ve dañado. Inspeccione visualmente el medidor para asegurarse de que la caja no esté agrietada y la caja posterior esté colo­cada firmemente.
• Inspeccione y reemplace los cables si el aislamiento está dañado, el metal está al descubierto o las sondas están agrietadas. Preste atención en particular al aislamiento que rodea los conectores.
• No utilice el medidor si funciona de forma anormal, ya que la protección puede haberse dañado.
• No lo utilice durante tormentas eléctricas o con tiempo húmedo.
• No lo utilice cerca de gases, polvo o vapores explosivos.
• No aplique al medidor un voltaje mayor que el voltaje nominal.
ND 1722-3 61-610 INSTRUCTONS 1/9/07 9:47 AM Page 9
Loading...
+ 19 hidden pages