Ideal 61-361, TEST PRO 360 User Manual

WARNING!
1. DO NOT UNDER ANY CIRCUMSTANCES EXCEED THESE RATINGS:
-Voltage is not to exceed 600V AC or DC.
- Resistance, Capacitance, Logic, and Continuity
functions are not to be performed on circuits capable of delivering greater than 500V AC or DC.
- Current measurements are not to be performed on
2. To avoid electrical shock hazards and/or damage to the meter:
- Do not exceed the voltage ratings for the meter. Use caution when measuring voltage.
- Do not use during electrical storms. AC power sources with inductive loads or electrical storms may result in high voltage. High energy transients can damage meter and present a dangerous shock hazard.
-Turn off the power to the circuit or device being mea­sured before taking resistance and capacitance mea­surements. Fully discharge all capacitors before measuring.
3. Ensure meter is in proper working order before using. Visually inspect meter for damage. Performing a conti­nuity check can verify proper operation. If the meter reading goes from overload to zero, this typically means the meter is in proper working order.
4. Visually inspect leads for damage before using. Replace if insulation is damaged or leads appear suspect.
5. Never ground yourself when taking electrical measure­ments. Do not touch exposed metal pipes, outlets, fix­tures, etc. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any other approved insulating material. Keep your fingers behind the finger guards on the probes. Work with oth­ers.
6. Before beginning all unknown measurements, set meter to the highest range possible.
Page 2
#61-361
IDEAL Test Pro
®
360 Series True RMS Multimeter
Page 4Page 3
WARNING! (cont.)
7. Before breaking a circuit for testing, turn off the power to the circuit. When disconnecting from a circuit, discon­nect the hot lead first, then the common lead.
8. Disconnect the meter from the circuit before turning off any indicator, including motors, transformers,and sole­noids.
Overload Protection
VAC+ V
DC
200m Vrange 500VDC/350VAC
for 15 sec
>200m Vrange 600V AC/DC
A
AC
+ A
DC
mA input 0.5A/250V
10A input 10A/600V Ohms (Ω) 500V AC/DC Diode 500V AC/DC Continuity 500V AC/DC Capacitance
(MFD) 500V AC/DC
Frequency
(Hz) 500V AC/DC
Phase Rotation (RST) 500V AC/DC
Unit of Measure Multipliers
For your reference, the following symbols are often used to make measurement easier: Symbol Verbal Multiplier M mega x1,000,000 k kilo x1,000 m milli ÷1,000 µ micro ÷1,000,000
AC Voltage (VAC)
Function Range Resolution Accuracy
AC Voltage 200.0 mV .1 mV
2000 mV 1 mV
200.0 V .1V 600 V 1 V
To Measure AC Voltage:
1. Plug the test leads into the meter inputs as indicated in the above diagram.
2. Select the proper range to be used within the VAC Function.
3. Connect the meter in parallel with the load or circuit.
4. Measure AC voltage.
VAC
PHASE ROTATI
HZ-MFD-A
2000m
RST
200m
200
600
Black
Red
±(2.0% + 4) 50 to 500Hz
all ranges
PHASE ROTATION-TRUE RMS
HZ-MFD-AUTO OFF
600600
VAC
2000m
200m
RST
Hz
200M
20M
200k
2A
MFD
R/L1
VDC
200
200
20
2000m
200m
ADC
AAC
200
2k
20k
20k
2k
200
MFD
CAT. LLL 600V
COM
V
!
MAX 2A
600V
FUSED
MAX 350V/15 SEC FOR 200mV RANGE
DC Voltage (VDC)
Function Range Resolution Accuracy
DC Voltage 200.0 mV .1 mV
2000 mV 1 mV
20.00 V .01 V
200.0 V .1 V 600 V 1 V
To Measure DC Voltage:
Do not attempt to measure voltage in circuits capable of delivering more than 600V.
1. Plug the leads into meter inputs as indicated in the diagram above.
2. Select the proper range to be used within the VDC Function.
3. Connect the meter in parallel with the load or circuit.
4. Measure DC voltage.
Page 5
Page 6
Resistance (Ohms)
Function Range Resolution Accuracy
Resistance 200.0 .1 ±(1.5% + 4)
2.000k .001k ±(1.5% + 4)
20.00k .01k ±(1.5% + 4)
200.0k .1k ±(1.5% + 4)
20.00M .01M ±(2.5% + 4)
200.0M .1 M ± (5.0% +20)
To Measure Resistance:
Resistance is measured in Ohms.
1. Turn the power off to the circuit or device that is to be mea­sured and discharge all capacitors before attempting a mea­surement.
2. Plug the red test lead into the “V-
” meter input.
3. Plug the black test lead into the “Com” meter input.
4. Select the proper range within the function of the tester.
VDC
ADC
2000m
20
200
600
Black
Red
20M
20k
200k
2k
Hz
200M
200
Black
Red
± (1.2% +1)
all ranges
PHASE ROTATION-TRUE RMS
HZ-MFD-AUTO OFF
600
600
VAC
2000m
200m
RST
Hz
200M
20M
200k
2A
MFD
R/L1
VDC
200
200
20
2000m
ADC
200
2k
20k
20k
2k
200
MFD
CAT. LLL 600V
COM
V
!
MAX 2A
600V
FUSED
MAX 350V/15 SEC FOR 200mV RANGE
+
-
PHASE ROTATION-TRUE RMS
HZ-MFD-AUTO OFF
600600
VAC
2000m
200m
RST
Hz
200M
20M
200k
2A
MFD
R/L1
VDC
200
200
20
2000m
200m
ADC
AAC
200
2k
20k
20k
2k
200
MFD
CAT. LLL 600V
COM
V
!
MAX 2A
600V
FUSED
MAX 350V/15 SEC FOR 200mV RANGE