The Service Information provides the following information:
Precautions and safety information
Specifications
Basic maintenance (cleaning, replacing the battery and fuses)
Performance test procedures
Calibration and calibration adjustment procedures
To avoid shock or injury, do not perform the verification tests or calibration procedures
described in the manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
Caution
The 61-320 serials contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES, INC. and its products, and services, visit
IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the Meter only as described in the Service Manual. If you do not do so, the protection provided by the
Meter may be impaired. Read the “Safety Information” page before servicing this product.
In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution
identifies conditions and actions that may damage the Meter or the test instruments.
The Symbols
The symbols used on the Meter and in this manual are explained in Table A.
Table A. The Symbols
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or reinforced
insulation
Battery
Earth
AC measurement
Conforms to EU directives
1
Form Number TM61320-2-4 Rev 3 April 2004
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION: These statements identify conditions or practices that could result in damage to the
equipment or other property.
WARNING: These statements identify conditions or practices that could result in personal injury or
loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
2
Form Number TM61320-2-4 Rev 3 April 2004
SPECIFICATIONS
All specifications are warranted unless noted typical and apply to the 61-320 & 61-322 & 61-324
Stated accuracies are at 23°C±5°C at less than 80% relative humidity and without the battery indicator
displayed.
General specifications
Characteristics Description
Display count 3 3/4
Numeric update rate 1.5 times / sec
Polarity display Automatic
Overrange display “OL” is display
Low voltage indicator is indicated
Automatic power-off time Automatic backslit off = 30 minutes
Power source
9V×1 battery for 61-322 & 61-324;
1.5V×2 batteries for 61-320
Maximum input voltage 1000V CAT III between V and COM
Maximum floating voltage
1000V CAT
III
between any terminal and earth ground
Maximum input current 400mA between mA and COM
Maximum open circuit Voltage (current inputs)600V between mA and COM
Overload protection mA connector 1A (600V) fast blow fuse.
V connector V, V, Ω, , , , Hz, µA
Temperature Coefficient 0.2×(Spec. Accuracy) / °C, <18°C or >28°C
Battery Life
Alkaline 9V 200 hours for 61-322 & 61-324
Alkaline 1.5V×2 AA size 600 hours for 61-320
3
Form Number TM61320-2-4 Rev 3 April 2004
Measurement Characteristics
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 80% R.H.
(1) DC Volts
Range Resolution Accuracy
Over voltage
protection
400.0mV 100µV
4.000V 1mV
40.00V 10mV
±(0.5% reading + 2 digit) 1000V rms
400.0V 100mV
1000V 1V
Input Impedance: 10M (over 1000M in 400mV range).
(2) AC Volts
Range Resolution Accuracy
400.0mV 100µV
4.000V 1mV
±(1.3% reading + 5 digits) *
Unspecified
40.00V 10mV
400.0V 100mV
750V
1V
±(1.2% reading + 5 digits) *
1
2,3
Over voltage
protection
1000V rms
Input Impedance: 10M // less than 100pF.
1
*
Frequency Response: 50Hz ~ 300Hz
2
Frequency Response: 50Hz ~ 500Hz
*
CMRR / NMRR: (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
: CMRR > 60dB at DC, 50Hz / 60Hz
AC
V
: CMRR > 100dB at DC, 50Hz / 60Hz
DC
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type: 61-320: Average sensing rms indication calibrated to the sine wave input.
61-322 / 61-324: AC conversions are ac-coupled, true rms responding, calibrated to
the sine wave input.
3
*
61-322 / 61-324 The specified accuracy is for sine wave at full scale and non-sine wave at half scale
with crest factor up to 2.
Crest Factor: C.F. = Peak/RMS
+1.5% addition error for C.F. from 1.4 to 3
+3.0% addition error for C.F. from 3 to 4.
Form Number TM61320-2-4 Rev 3 April 2004
4
(3) DC Current
Range Resolution Accuracy Voltage Burden
400.0µA 0.1µA 5mV
4000µA 1µA 2V max.
±(1.0% reading + 2 digits)
40.00mA* 10µA 5mV
400.0mA* 0.1mA
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse. (61-322 / 61-324)
µA Input : 600V rms.
* For 61-322 / 61-324 only.
(4) AC Current (61-322 / 61-324)
Range Resolution Accuracy Voltage Burden
40.00mA 10µA 5mV
400.0mA 100µA
±(1.5% reading + 5 digits) *
50Hz ~ 500Hz
1
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse.
AC Conversion Type: AC conversions are ac-coupled, true rms responding, calibrated to the sine wave
input.
1
*
The specified accuracy is for sine wave at full scale and non-sine wave at half scale with crest factor
up to 2.
(5) Resistance
Range Resolution Accuracy
400.0 *
1
0.1±(1.0% reading + 5 digits)
Over voltage
protection
4.000K 1
40.00K 10
±(0.7% reading + 2 digits)
600V rms
400.0K 100
4.000M 1K
40.00M *
2
10K±(1.5% reading + 5 digits)
Open circuit Voltage: -1.3V approx.
1
< 5 digit of reading rolling.
*
2
*
< 2% of reading rolling.
(6) Diode Check and Continuity
Range Resolution Accuracy
Max. Test
Current
Max. Open
Circuit Voltage
1mV ±(1.5% reading + 5 digits) * 1.5mA 3V
* For 0.4V ~ 0.8V
Overload Protection: 600V rms.
Continuity: Built-in buzzer sounds when resistance is less than approximately 450 Response time
is approximately 100 msec.
5
Form Number TM61320-2-4 Rev 3 April 2004
(7) Frequency
Range Resolution Sensitivity Accuracy
4.000KHz 1Hz
40.00KHz 10Hz
400.0KHz 100Hz
150mV rms *
Frequency:
(± 0.01% + 1 digit)
4.000MHz 1KHz 300mV rms
40.00MHz 10KHz 1V rms
* Less than 20Hz, the sensitivity is 1.5V rms.
(8) Capacitance
Overload
protection
600V rms
Range Resolution Accuracy
Over voltage
Protection
4.000nF 1pF
±(3% reading + 10 digits)
40.00nF 10pF
400.0nF 100pF
4.000µF 1nF
±(2% reading + 8 digits)
600V rms
40.00µF 10nF
400.0µF 100nF
4.000mF *
40.00mF *
1
1
1µF
10µF
±(5% reading + 20 digits) *
2
*1 In this range the reading maybe rolling within specification.
2
Specify reading < half full scale of range.
*
(9) Auto Power Off (APO)
If the Meter idles for more than 30 minutes, the Meter automatically turns the power off.
(10) Peak Hold
Function Range Accuracy
400mV Unspecified
4.000V ±(1.5% reading + 300 digits) *
ACV
40.00V
400.0V
±(1.5% reading + 60 digits)
750V
ACA
(61-322/61-324)
40.00mA *
400.0mA *
3
3
±(3% reading + 60 digits)
Note:
1
With zero calibrated before measurement.
2
*
4V range specifies readings above 10% of full scale of range.
3
Amp range specify reading <90% of full scale of range.
*
4
In the noise generating field, may affect intervals.
6
Form Number TM61320-2-4 Rev 3 April 2004
2
Physical and Environmental Characteristics
Characteristics Description
Dimensions (H×W×D)
Weight (with battery) 0.3Kg
With holster 0.5Kg
Environmental characteristics Description
Temperature operating 0 to +50°C
Non-Operating -20 to +60°C
Humidity (operating) <80% R.H.
Altitude Operating 2,000M (6560 ft.)
Non-Operating 12,300M (40354 ft.)
Vibration & shock Operating
Indoor Use Indoor Use
Certifications and compliances
158mm×76mm×38mm
164mm×82mm×44mm (with holster)
MIL-T-28800E TYPE II Class 5
2.66gRMS, 5 to 500Hz, 3axes
(10 minutes each)
Designed to ICE 1010-1, UL3111-1 and CSA specifications
Safety
Input rating
Over voltage category
Pollution Degree 2
EC Declaration
of Conformity
V / /µA: Category III 1000 Volts
V / /mA: Category
CAT III: Distribution level mains, fixed installation.
CAT II: Local level mains, appliances, portable equipment
CAT I: Signal level, special equipment or parts of equipment,
telecommunication, electronics.
Do not operate in environments where conductive
Pollutants may be present.
Meets the intent of Directive 89/336/EEC for Electtromagnetic
Compatibility and Low Voltage Directive 73/23/EEC for product
safety. Compliance was demonstarted to the following specifica-
tions as listed in the official Journal of the European Communites:
En 55011 Class A: Radiated and Conducted
En 50082-1 Immunity: IEC 801-2 Electrostatic Discharge
En 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use.
III 600 Volts
Emissions.
IEC 801-3 RF Radiated
7
Form Number TM61320-2-4 Rev 3 April 2004
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
AC Voltage Range: 0
Accuracy: ±0.07% (Basic)
Frequency Range: 40
Accuracy: ±2%
DC Voltage Range: 0
Accuracy: ±0.006% (Basic)
Current Range: 0
Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ±0.02% (Basic)
Frequency Source: 5.00Hz
Accuracy: ±0.001%
Amplitude: 0.5V p-p
Accuracy: ±5%
Resistance Range: 1
Accuracy: ±0.03% (Basic)
Capacitance Range: 1pF
Accuracy: ±0.10% (Basic)
~ 750V AC
~ 1KHz
~ 1000V DC
~ 10A
~ 100MHz
~ 1.0V p-p (square wave)
~ 100M
~ 10mF
Fluke 5500 or Wavetek
9100 Calibrator or
equipment
Form Number TM61320-2-4 Rev 3 April 2004
8
Basic Maintenance
Warning
To avoid shock, remove the test leads and any input signals before opening the case or replacing the
battery or fuses.
Opening the Meter Case
Caution
To avoid unintentional shock circuit, always place the uncovered Meter assembly on a protective surface.
When the case of the Meter is open, circuit connections are exposed.
To open the Meter case, refer Figure 1 and do the following:
1. Disconnect test leads from any live source, turn the rotary switch to OFF, and remove the test leads
from front terminals.
2. Remove the battery door by using a flat-blade screwdriver to turn the battery door screws turn
counter-clockwise.
3. The case bottom is secured to the case top by four screws and two internal snaps (at the LCD end).
Using a Phillips-had screwdriver, remove the four screws.
Replacing the Battery
The Meter is powered 1.5V x 2 batteries for 61-320 and a single 9V battery for 61-322 / 61-324.
To replace the battery, refer to Figure
Form Number TM61320-2-4 Rev 3 April 2004
9
Testing Fuses
To test the internal fuses of the meter.
1. Turn the rotary selector switch to the Ω position.
2. To test FS1, plug a test lead into VΩHz input terminal, and touch the probe to the mA input terminal.
The display should indicate between 0.0 to 0.2Ω. FS1 (1A 600V) (Bussmann BBS-1 recommended). If
display reads higher than 0.2Ω, replace the fuse.
Fuse Replacement
Refer to the following figure to replace fuse:
Use only a fuse with the amperage, interrupt, voltage, and speed rating specified.
Fuse rating: 1A, 600V, Fast
Replacing Fuses
Warning
To avoid electrical shock, remove the test leads and any input signals before replacing the battery
or fuses. To prevent damage or injury, INSTALL ONLY quick acting fuses with the following