IDEAL 61-320, 61-322, 61-324 TECHNICAL MANUAL

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL
MODEL: 61-320 MODEL: 61-322 MODEL: 61-324
Multimeters Service Information
The Service Information provides the following information:
Precautions and safety information Specifications Basic maintenance (cleaning, replacing the battery and fuses) Performance test procedures Calibration and calibration adjustment procedures
Form Number: TM61320-2-4
Revision: 3. Date: Apr 2004
Form Number TM61320-2-4 Rev 3 April 2004
TABLE OF CONTENTS Page
Introduction 1
Precautions and Safety Information 1
Symbols 1
Safety 2
Specifications 3
General Specification 3
Measurement Characteristics 4
Voltage Specification 4
Current Specifications 5
Resistance, Diode Specifications 5
Continuity, Frequency, Capacitance Specification 6
Auto Power Off, Peak Hold 6
Physical and environment characteristics 7
Certification and compliance 7
Required Equipment 8
Basic Maintenance 9
Opening the Meter Case 9
Replacing the Battery 9
Testing Fuses 10
Fuses Replacement 10
Replacing Fuses 10
Cleaning 10
Performance Tests 11
Testing the Display 11
Testing the Voltage Function 12
Testing the Resistance Function 13
Testing the Capacitance Function 13/14
Testing the Diode 14
Testing the milliamp (mA) Function 14/15
Testing the microamp Function 15
Testing the Frequency 15
Testing Frequency Function 15/16
Calibration 17
Calibrating DCV, ACV Functions 17
Calibration Adjustment Points 18/19
i
Form Number TM61320-2-4 Rev 3 April 2004
Introduction
Warning
To avoid shock or injury, do not perform the verification tests or calibration procedures
described in the manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
Caution
The 61-320 serials contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES, INC. and its products, and services, visit
IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the Meter only as described in the Service Manual. If you do not do so, the protection provided by the
Meter may be impaired. Read the “Safety Information” page before servicing this product.
In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution
identifies conditions and actions that may damage the Meter or the test instruments.
The Symbols
The symbols used on the Meter and in this manual are explained in Table A.
Table A. The Symbols
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or reinforced
insulation
Battery
Earth
AC measurement
Conforms to EU directives
1
Form Number TM61320-2-4 Rev 3 April 2004
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION: These statements identify conditions or practices that could result in damage to the
equipment or other property.
WARNING: These statements identify conditions or practices that could result in personal injury or
loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
2
Form Number TM61320-2-4 Rev 3 April 2004
SPECIFICATIONS
All specifications are warranted unless noted typical and apply to the 61-320 & 61-322 & 61-324
Stated accuracies are at 23°C±5°C at less than 80% relative humidity and without the battery indicator
displayed.
General specifications
Characteristics Description
Display count 3 3/4
Numeric update rate 1.5 times / sec
Polarity display Automatic
Overrange display “OL” is display
Low voltage indicator is indicated
Automatic power-off time Automatic backslit off = 30 minutes
Power source
9V×1 battery for 61-322 & 61-324;
1.5V×2 batteries for 61-320
Maximum input voltage 1000V CAT III between V and COM
Maximum floating voltage
1000V CAT
III
between any terminal and earth ground
Maximum input current 400mA between mA and COM
Maximum open circuit Voltage (current inputs) 600V between mA and COM
Overload protection mA connector 1A (600V) fast blow fuse.
V connector V , V , Ω, , , , Hz, µA
Temperature Coefficient 0.2×(Spec. Accuracy) / °C, <18°C or >28°C
Battery Life
Alkaline 9V 200 hours for 61-322 & 61-324
Alkaline 1.5V×2 AA size 600 hours for 61-320
3
Form Number TM61320-2-4 Rev 3 April 2004
Measurement Characteristics
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 80% R.H.
(1) DC Volts
Range Resolution Accuracy
Over voltage
protection
400.0mV 100µV
4.000V 1mV
40.00V 10mV
±(0.5% reading + 2 digit) 1000V rms
400.0V 100mV
1000V 1V
Input Impedance: 10M (over 1000M in 400mV range).
(2) AC Volts
Range Resolution Accuracy
400.0mV 100µV
4.000V 1mV
±(1.3% reading + 5 digits) *
Unspecified
40.00V 10mV
400.0V 100mV
750V
1V
±(1.2% reading + 5 digits) *
1
2,3
Over voltage
protection
1000V rms
Input Impedance: 10M // less than 100pF.
1
*
Frequency Response: 50Hz ~ 300Hz
2
Frequency Response: 50Hz ~ 500Hz
* CMRR / NMRR: (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
: CMRR > 60dB at DC, 50Hz / 60Hz
AC
V
: CMRR > 100dB at DC, 50Hz / 60Hz
DC
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type: 61-320: Average sensing rms indication calibrated to the sine wave input.
61-322 / 61-324: AC conversions are ac-coupled, true rms responding, calibrated to
the sine wave input.
3
*
61-322 / 61-324 The specified accuracy is for sine wave at full scale and non-sine wave at half scale
with crest factor up to 2. Crest Factor: C.F. = Peak/RMS +1.5% addition error for C.F. from 1.4 to 3 +3.0% addition error for C.F. from 3 to 4.
Form Number TM61320-2-4 Rev 3 April 2004
4
(3) DC Current
Range Resolution Accuracy Voltage Burden
400.0µA 0.1µA 5mV
4000µA 1µA 2V max.
±(1.0% reading + 2 digits)
40.00mA* 10µA 5mV
400.0mA* 0.1mA
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse. (61-322 / 61-324)
µA Input : 600V rms.
* For 61-322 / 61-324 only.
(4) AC Current (61-322 / 61-324)
Range Resolution Accuracy Voltage Burden
40.00mA 10µA 5mV
400.0mA 100µA
±(1.5% reading + 5 digits) *
50Hz ~ 500Hz
1
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse. AC Conversion Type: AC conversions are ac-coupled, true rms responding, calibrated to the sine wave
input.
1
*
The specified accuracy is for sine wave at full scale and non-sine wave at half scale with crest factor
up to 2.
(5) Resistance
Range Resolution Accuracy
400.0 *
1
0.1 ±(1.0% reading + 5 digits)
Over voltage
protection
4.000K 1
40.00K 10
±(0.7% reading + 2 digits)
600V rms
400.0K 100
4.000M 1K
40.00M*
2
10K ±(1.5% reading + 5 digits)
Open circuit Voltage: -1.3V approx.
1
< 5 digit of reading rolling.
*
2
*
< 2% of reading rolling.
(6) Diode Check and Continuity
Range Resolution Accuracy
Max. Test
Current
Max. Open
Circuit Voltage
1mV ±(1.5% reading + 5 digits) * 1.5mA 3V
* For 0.4V ~ 0.8V Overload Protection: 600V rms. Continuity: Built-in buzzer sounds when resistance is less than approximately 450 Response time
is approximately 100 msec.
5
Form Number TM61320-2-4 Rev 3 April 2004
(7) Frequency
Range Resolution Sensitivity Accuracy
4.000KHz 1Hz
40.00KHz 10Hz
400.0KHz 100Hz
150mV rms *
Frequency: (± 0.01% + 1 digit)
4.000MHz 1KHz 300mV rms
40.00MHz 10KHz 1V rms
* Less than 20Hz, the sensitivity is 1.5V rms.
(8) Capacitance
Overload
protection
600V rms
Range Resolution Accuracy
Over voltage
Protection
4.000nF 1pF ±(3% reading + 10 digits)
40.00nF 10pF
400.0nF 100pF
4.000µF 1nF
±(2% reading + 8 digits)
600V rms
40.00µF 10nF
400.0µF 100nF
4.000mF *
40.00mF *
1
1
1µF
10µF
±(5% reading + 20 digits) *
2
*1 In this range the reading maybe rolling within specification.
2
Specify reading < half full scale of range.
*
(9) Auto Power Off (APO)
If the Meter idles for more than 30 minutes, the Meter automatically turns the power off.
(10) Peak Hold
Function Range Accuracy
400mV Unspecified
4.000V ±(1.5% reading + 300 digits) *
ACV
40.00V
400.0V
±(1.5% reading + 60 digits)
750V
ACA
(61-322/61-324)
40.00mA *
400.0mA *
3
3
±(3% reading + 60 digits)
Note:
1
With zero calibrated before measurement.
2
*
4V range specifies readings above 10% of full scale of range.
3
Amp range specify reading <90% of full scale of range.
*
4
In the noise generating field, may affect intervals.
6
Form Number TM61320-2-4 Rev 3 April 2004
2
Physical and Environmental Characteristics
Characteristics Description
Dimensions (H×W×D)
Weight (with battery) 0.3Kg
With holster 0.5Kg
Environmental characteristics Description
Temperature operating 0 to +50°C
Non-Operating -20 to +60°C
Humidity (operating) <80% R.H.
Altitude Operating 2,000M (6560 ft.)
Non-Operating 12,300M (40354 ft.)
Vibration & shock Operating
Indoor Use Indoor Use
Certifications and compliances
158mm×76mm×38mm
164mm×82mm×44mm (with holster)
MIL-T-28800E TYPE II Class 5
2.66gRMS, 5 to 500Hz, 3axes
(10 minutes each)
Designed to ICE 1010-1, UL3111-1 and CSA specifications
Safety
Input rating
Over voltage category
Pollution Degree 2
EC Declaration
of Conformity
V / /µA: Category III 1000 Volts
V / /mA: Category
CAT III: Distribution level mains, fixed installation.
CAT II: Local level mains, appliances, portable equipment
CAT I: Signal level, special equipment or parts of equipment,
telecommunication, electronics.
Do not operate in environments where conductive
Pollutants may be present.
Meets the intent of Directive 89/336/EEC for Electtromagnetic
Compatibility and Low Voltage Directive 73/23/EEC for product
safety. Compliance was demonstarted to the following specifica-
tions as listed in the official Journal of the European Communites:
En 55011 Class A: Radiated and Conducted
En 50082-1 Immunity: IEC 801-2 Electrostatic Discharge
En 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use.
III 600 Volts
Emissions.
IEC 801-3 RF Radiated
7
Form Number TM61320-2-4 Rev 3 April 2004
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
AC Voltage Range: 0 Accuracy: ±0.07% (Basic)
Frequency Range: 40 Accuracy: ±2%
DC Voltage Range: 0 Accuracy: ±0.006% (Basic)
Current Range: 0 Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ±0.02% (Basic)
Frequency Source: 5.00Hz Accuracy: ±0.001%
Amplitude: 0.5V p-p Accuracy: ±5%
Resistance Range: 1 Accuracy: ±0.03% (Basic)
Capacitance Range: 1pF Accuracy: ±0.10% (Basic)
~ 750V AC
~ 1KHz
~ 1000V DC
~ 10A
~ 100MHz
~ 1.0V p-p (square wave)
~ 100M
~ 10mF
Fluke 5500 or Wavetek
9100 Calibrator or
equipment
Form Number TM61320-2-4 Rev 3 April 2004
8
Basic Maintenance
Warning
To avoid shock, remove the test leads and any input signals before opening the case or replacing the
battery or fuses.
Opening the Meter Case
Caution
To avoid unintentional shock circuit, always place the uncovered Meter assembly on a protective surface.
When the case of the Meter is open, circuit connections are exposed.
To open the Meter case, refer Figure 1 and do the following:
1. Disconnect test leads from any live source, turn the rotary switch to OFF, and remove the test leads
from front terminals.
2. Remove the battery door by using a flat-blade screwdriver to turn the battery door screws turn
counter-clockwise.
3. The case bottom is secured to the case top by four screws and two internal snaps (at the LCD end).
Using a Phillips-had screwdriver, remove the four screws.
Replacing the Battery
The Meter is powered 1.5V x 2 batteries for 61-320 and a single 9V battery for 61-322 / 61-324.
To replace the battery, refer to Figure
Form Number TM61320-2-4 Rev 3 April 2004
9
Testing Fuses
To test the internal fuses of the meter.
1. Turn the rotary selector switch to the Ω position.
2. To test FS1, plug a test lead into VHz input terminal, and touch the probe to the mA input terminal.
The display should indicate between 0.0 to 0.2. FS1 (1A 600V) (Bussmann BBS-1 recommended). If
display reads higher than 0.2Ω, replace the fuse.
Fuse Replacement
Refer to the following figure to replace fuse:
Use only a fuse with the amperage, interrupt, voltage, and speed rating specified.
Fuse rating: 1A, 600V, Fast
Replacing Fuses
Warning
To avoid electrical shock, remove the test leads and any input signals before replacing the battery
or fuses. To prevent damage or injury, INSTALL ONLY quick acting fuses with the following
Amp/Volt current interrupt rating :
FS1 Fuse : 1A, 600V, FAST. Minimum interrupt rating 10,000A
Cleaning
Warning
To avoid electrical shock or damage to the Meter, never allow water inside the case. To avoid
damaging the Meter’s housing, never apply solvents to the Meter.
10
Form Number TM61320-2-4 Rev 3 April 2004
Performance Tests
The following performance tests verify the complete operability of the Meter and check the accuracy of
each Meter function against the Meter’s specifications.
Accuracy specifications are valid for a period of one year after calibration, when measured at an
operating temperature of 18°C to 28°C and a maximum of 80% relative humidity.
To perform the following tests, it is not necessary to open the case, no Adjustments are necessary,
merely make the required connections, apply the designated inputs, determine if the reading on the
Meter display falls within the acceptable range indicated.
If the Meter fails any of these tests, it needs calibration adjustment or repair.
Testing the Display
Press “HOLD” key while turning the Meter on from the “OFF” position to hold the display in the Display
Test Mode. Compare the display with the example in Figure 2. Turn off the meter to escape the test
mode.
LCD Graphics 61-324
Figure 2 Display Test
Form Number TM61320-2-4 Rev 3 April 2004
11
Testing the Voltage Function
To verify accuracy in the AC and DC voltage ranges, do the following:
1. Turn the rotary switch to “V
” position.
2. Connect the Calibrator to the V and COM inputs on the Meter.
3. Set the Calibrator for the voltage and frequency from step 1 to 8 in Table 1.
4. Compare the reading on the Meter display with the display reading shown in Table 1.
5. If the display reading falls outside of the range shown in Table 1, the Meter does not meet specification.
Table 1 AC Voltage Test:
Step Input Frequency
Reading
1 3.000V 50Hz 2.956 to 3.044
2 3.000V 500Hz 2.956 to 3.044
3 30.00V 50Hz 29.59 to 30.41
4 30.00V 500Hz 29.59 to 30.41
5 300.0V 50Hz 295.9 to 304.1
6 300.0V 500Hz 295.9 to 304.1
7 750V 50Hz 736 to 764
8 750V 500Hz 736 to 764
6. Turn the rotary switch to “V
” position.
7. Set the calibration for the voltage from step 1 to 6 in Table 2.
8. Compare the reading on the Meter display with the display reading shown in Table 2.
9. If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.
Table 2 DC Voltage Test:

Step Input Reading

1 300.0mV 298.3 to 301.7
2 -300.0mV -298.3 to -301.7
3 3.000V 2.983 to 3.017
4 30.00V 29.83 to 30.17
5 300.0V 298.3 to 301.7
6 1000V 993 to 1007
Form Number TM61320-2-4 Rev 3 April 2004
12
Testing the Resistance Function
To verify the accuracy of the resistance function, do the following:
1. Connect the Calibrator to VHz and COM on the Meter.
2. Turn the rotary switch to .
3. Apply the inputs for step 1-6 in Table 3.
4. Compare the Meter display readings to the display readings in Table 3.
5. If the display reading falls outside of the range shown in Table 3, the Meter does not meet specification.
Table 3 Resistance Test:

Step Source Reading

1 300.0 296.5 to 303.5
2 3.000K 2.977 to 3.023
3 30.00K 29.77 to 30.23
4 300.0K 297.7 to 302.3
5 3.000M 2.977 to 3.023
6 30.00M 29.50 to 30.50
Lead resistance on the 400 range is not included in error.
Testing the Capacitance Function
The Meter measures capacitance by charging the capacitor with a known Direct current, measuring the
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an
impedance bridge, a different reading may result. This variance is likely to be Greater at higher
frequencies.
To verify the accuracy of the capacitance measuring function, do the Following :
1. Apply the Capacitor to the VHz and COM inputs on the Meter. For steps 1 through 7 in Table 4.
2. Turn the rotary switch
.
3. Compare the reading on the Meter display to the reading in Table 4.
Note : The meter selects the proper range automatically. Each measurement takes about one second
per range, 5mF takes about 4.5 seconds.
4. If the display reading falls outside of the range shown in Table 4, the Meter does not meet specification.
Form Number TM61320-2-4 Rev 3 April 2004
13
Table 4 Capacitance Test:
Step Source Reading
1 3.000nF 2.900 to 3.100
2 30.00nF 29.00 to 31.00
3 300.0nF 293.2 to 306.8
4 3.000µF 2.932 to 3.068
5 30.00µF 29.32 to 30.68
6 300.0µF 293.2 to 306.8
7 3.000mF 2.830 to 3.170
Checking the Diode Test Function
To check the diode test function, do the following:
1. Connect the Calibrator to the VHz and COM inputs on the Meter.
2. Turn the rotary switch to
.
3. Apply 3.000V DC.
The meter display should read approx. 3.000V dc.
4. Apply a 50 resistor to meter, the built-in beeper alarms.
Testing the milliamp (mA) Function (for 61-322 and 61-324)
To verify the accuracy of AC current measurement functions, do the following:
1.Connect the Calibrator to the mA and COM inputs on the Meter.
2. Turn the rotary switch to mA
3. Apply the inputs for steps 1-4 in Table 5.
4. For each input, compare the reading on the Meter display to the reading for your Meter in Table 5.
5. If the display reading falls outside of the range shown in Table 5, the meter does not meet specification.
Table 5 AC mA Test:
Step Source Frequency Reading
1 30.00mA 50Hz 29.50 to 30.50
2 30.00mA 500Hz 29.50 to 30.50
3 300.0mA 50Hz 295.0 to 305.0
4 300.0mA 500Hz 295.0 to 305.0
Form Number TM61320-2-4 Rev 3 April 2004
14
6. Turn the rotary switch to mA .
7. Set the calibration for the voltage from step 1 - 2 in Table 6.
8. For each input, compare the reading on the Meter display to the reading in Table 2.
9. If the display reading falls outside of the range shown in Table 6, the meter does not meet specification.
Table 6 DC Current Test:
Step Source Reading
1 30.00mA 29.68 to 30.32
2 300.0mA 296.8 to 303.2
Testing the microamp Function
1. Turn the rotary switch to µA
.
3. Apply the inputs for steps 1-2 in Table 7.
4. For each input, compare the reading on the Meter display to the reading for your Meter in Table 8.
5. If the display reading falls outside of the range shown in Table 7, the meter does not meet specification.
Table 7 DC microamp Test:
Step Source Reading
1 300.0µA 296.8 to 303.2
2 3000µA 2968 to 3032
Testing the Frequency Function
To verify the accuracy of the Meter’s frequency function, do the following:
1. Connect the Calibrator to the V and COM inputs on the Meter.
Note: The accuracy of the Calibrator’s frequency function must be appropriate for the specified
accuracy of the Meter.
2. Set the rotary switch to Hz.
3. Set the Function Generator for the square wave voltage and frequency for steps 1-5 of Table 8.
4. Compare the reading on the Meter display with the display reading shown in Table 8.
5. If the display reading falls outside of the range shown in Table 8, the Meter does not meet specification.
Form Number TM61320-2-4 Rev 3 April 2004
15
Table 8 Frequency Test:
Step Soure Level Reading
1 3.000KHz 150mV rms 2.999 to 3.001
2 30.00KHz 150mV rms 29.99 to 30.01
3 300.0KHz 150mV rms 299.9 to 300.1
4 3.000MHz 300mV rms 2.999 to 3.001
5 30.00MHz 1V rms 29.99 to 30.01
Form Number TM61320-2-4 Rev 3 April 2004
16
Calibration Procedure
Recalibrate your meter:
It is recommended that the multimeter be calibrated once each year.
1. Perform calibration at an ambient temperature of 23°C±2°C and a relative humidity of 73% or less.
2. Disconnect the test leads and turn the meter off Remove the test leads from the front terminals.
3. Position the meter face down. Remove the three screws from the case bottom.
4. Lift the end of the case bottom until it gently unsnaps from the case top at the end nearest the LCD.
5. Lift the circuit board from the case top. Do not remove the screws from the circuit board.
(A) DCV Calibration (Adjust VR1)
1. Set the circuit board rotary switch "arrow" to the "V" position of circuit board.
2. Set the output of DC calibrator for 300.0V±0.02% and connect to VHz and COM input terminals on
meter.
3. Using a small flat-tipped screwdriver adjust the potentiometer VR1 until the display reads 299.0 or 300.1.
4. Disconnect the DC calibrator from the meter.
(B) ACV Calibration (Adjust VR2)
1. Set the circuit board rotary switch "arrow" to the "V" position of circuit board.
2. Set the output of AC calibration for 300Vrms 50Hz (61-320) 300Vrms 50Hz (61-322/61-324) and
connect to VHz and COM input terminals on meter.
3. Using a small flat-tipped screwdriver adjust the potentiometer VR2 until the display reads 299.9 or
300.1 (61-320) 299.9 or 300.1 (61-322/61-324).
Form Number TM61320-2-4 Rev 3 April 2004
17
VR1
VR2
Figure 3 61-320 Calibration Adjustment Points
18
Form Number TM61320-2-4 Rev 3 April 2004
VR1
VR2
Figure 4 61-322/324 Calibration Adjustment Points
19
Form Number TM61320-2-4 Rev 3 April 2004