IDEAL 61-300 User Manual

Field Test for 61-310, 61-312, 61-340 and 61-342
Most meters function in the same way. They take an electrical or physical measurement and convert it to a small electrical voltage or current to be interpreted via an Analog to digital converter and then to some kind of output device like a display or interface.
So let’s do a quick test of the meter. Before beginning, replace the battery and install the test leads of the Unit Under Test {UUT} Since most meters convert inputs to a DC voltage for the Analog to Digital converter a DC Voltage test
Turn the meter to DC Volts ( As an example on the IDEAL 61-310 Select the 20 V DC range.
Measuring a new 9V battery. The unit should read 9.2 to 9.9 Volts.
Turn the meter to AC Volts (
Measure an AC outlet. The unit should read between 108 and 132 Volts.
Turn the meter to Ohms. ( Ideal 61-310 this would be the 200 range
Short the Test leads together. The meter should read between 0 and no more that 5 counts. i. Example 0.3
If the meter does not read properly check the Leads.
If the ohms function is protected via a fuse like the 61-310 series or the 61-340 series
Check the Fuse.
If all three tests are working, you have tested about 80% of the base functions of the meter.
Now lets look at Temperture. Most meters use a K type Thermocouple {TC} which is made of 2 Dissimilar metals bonded at one end.
When 2 dissimilar metals are bonded together they cause a small electrical charge that is proportional to the environmental temperature.
It is easy to check a TC by using the ohms Function.
Place the meter into the ohms funtion and measure the imdepance of the probe.
You should expect some impedance due to the small electrical charge caused by the 2 dissimilar
metals. The reading should change as you heat or cool the TC
If the meter reads open then the TC should be replaced or repaired.
To Check the TC function select the C or F function
Us the test leads and short them together.
If you see a temperature reading, the meter is working OK if you are getting an Open {OL or
“1.___ “} display, the input protection or the meter is at fault.
Most meters TC function is protected by a Fuse.
Check the input protection Fuse.
), if you have a manual ranging unit select the proper range
), if you have a manual range meter select 200 or greater range
). If you have a manual range meter select the lowest Ohms range. On the
Loading...