The Optica Dew Point Analyzer is a General Eastern Instruments product. General Eastern Instruments has
joined other GE high-technology sensing businesses under a new name—GE Sensing.
January 2006
WarrantyEach instrument manufactured by GE Infrastructure Sensing is
warranted to be free from defects in material and workmanship. Liability
under this warranty is limited to restoring the instrument to normal
operation or replacing the instrument, at the sole discretion of GE
Infrastructure Sensing. Fuses and batteries are specifically excluded
from any liability. This warranty is effective from the date of delivery to
the original purchaser. If GE Infrastructure Sensing determines that the
equipment was defective, the warranty period is:
• one year for general electronic failures of the instrument
• one year for mechanical failures of the sensor
If GE Infrastructure Sensing determines that the equipment was damaged
by misuse, improper installation, the use of unauthorized replacement
parts, or operating conditions outside the guidelines specified by GE
Infrastructure Sensing, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of
all other warranties whether statutory, express or implied
(including warranties or merchantability and fitness for a
particular purpose, and warranties arising from course of
dealing or usage or trade).
Return PolicyIf a GE Infrastructure Sensing instrument malfunctions within the
warranty period, the following procedure must be completed:
1. Notify GE Infrastructure Sensing, giving full details of the problem,
and provide the model number and serial number of the instrument. If
the nature of the problem indicates the need for factory service, GE
Infrastructure Sensing will issue a RETURN AUTHORIZATION
number (RA), and shipping instructions for the return of the
instrument to a service center will be provided.
2. If GE Infrastructure Sensing instructs you to send your instrument to
a service center, it must be shipped prepaid to the authorized repair
station indicated in the shipping instructions.
3. Upon receipt, GE Infrastructure Sensing will evaluate the instrument
to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
• If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
• If GE Infrastructure Sensing determines that the damage is not
covered under the terms of the warranty, or if the warranty has
expired, an estimate for the cost of the repairs at standard rates will be
provided. Upon receipt of the owner’s approval to proceed, the
instrument will be repaired and returned.
IntroductionThe GE Infrastructure Sensing Optica is a multi-purpose
chilled-mirror hygrometer, suitable for use in a wide variety of
applications. The Optica can function with any GE Infrastructure
Sensing chilled-mirror sensor to provide the following measurement
ranges (depending on the sensor selected):
• dew/frost point from –80°C to +85°C (–112°F to +185°F)
• relative humidity from 0.002% to 100%
January 2006
• moisture content from 500 ppb
The Optica also measures and displays gas pressure using a GE
Infrastructure Sensing PT-30A or PT-300A pressure sensor, or a
user-supplied 4–20mA or 0–5 VDC pressure sensor.
Note: If the pressure is known to be constant, a fixed pressure can be
programmed, eliminating the need for a pressure sensor.
Using the Optica, you can simultaneously measure and display dew
point, temperature and pressure, with a wide variety of units of
measure. The Optica is Ethernet-ready, so you can access the unit
using the Internet for remote monitoring applications. You can use the
data logging function to record and upload weeks of data.
to over 5.71 × 105 ppm
v
v
Features and Capabilities1-1
January 2006
Electronics EnclosureThe Optica is available in two configurations:
• a benchtop model, with an optional rack-mount adapter available
• a wall-mount unit, housed in a NEMA-4 enclosure, suitable for
industrial environments
See Chapter 2, Installation, for details on how to mount the Optica.
Front PanelThe Optica’s front panel is shown in Figure 1-1 below. The panel
includes: a display, an alphanumeric keypad for data entry, ENTER
and TAB keys, and four softkeys to the right of the display screen.
Two display options are available:
• a high-resolution full-color liquid crystal display (LCD) with data
graphing and on-screen programming capability (referred to as the
“VGA” unit). See Figure 1-1 below.
• a 4-line by 40-character alphanumeric display (referred to as the
“4x40” unit).
Figure 1-1: Optica Benchtop Version with VGA Display
1-2Features and Capabilities
Input/Output CapabilityAvailable inputs and outputs include the following:
• 4-wire Resistance Temperature Detector (RTD) input
• 4–20 mA and 0–5 VDC Pressure Sensor inputs
• two simultaneous analog outputs, each with 4–20 mA and 0–5
VDC capability
• two independent alarm relays (Form C, 5 Amp)
• serial communications port
• Ethernet 10BaseT (on VGA Optica only)
The Optica uses the GE Infrastructure Sensing patented
Programmable Automatic Contaminant Error Reduction (PACER)
system to insure measurement accuracy.
Detailed specifications for the Optica are given in Appendix A.
January 2006
The System
System ComponentsA complete Optica system consists of the following items:
• Electronic monitor• Temperature sensor (optional)
• Dew point sensor• Pressure sensor (optional)
• Interconnecting sensor cable• Maintenance kit
• AC line cord• User’s Manual
• Certification that the unit is traceable to the National Institute
of Standards and Technology (Certificate of Conformance)
System PlanningThe Optica can be used for a wide variety of measurement
applications, including the measurement of dew points of gasses that
are at pressures that exceed the measurement range of GEI’s sensors.
In this situation be sure to check the section Process Pressure on
page 3-4 for installation planning.
SensorsThe Optica can be configured with a chilled-mirror dew point sensor.
The specific sensor is chosen according to the expected dew point
range and the environment in which the dew point is to be
determined. In addition, the Optica can be configured with a
temperature and/or pressure sensor. GE Infrastructure Sensing
provides the following sensors for various applications:
Features and Capabilities1-3
January 2006
Dew Point Sensors• Model 1111H — Single-stage sensor
• Model 1211H — Two-stage sensor; high pressure and temperature
• Model D-2 — Two-stage sensor
• Model SIM-12H — Two-stage heated sensor
• Model 1311DR — Four-stage, liquid or air cooled sensor
• Model 1311XR — Five-stage, liquid cooled sensor
Temperature Sensor• Model T-100E
Pressure Sensor• Model PT-30A or PT-300A
Theory of OperationOptical condensation hygrometry is a precise technique for
determining the water vapor content in gases by directly measuring
dew point or frost temperatures. Using this technique, a metallic
mirror is cooled until it reaches a temperature at which a thin layer of
condensation begins to form on it. The dew layer is detected optically,
and the mirror is held at that temperature. The mirror temperature,
measured with a platinum resistance thermometer, is an accurate
indicator of the dew or frost point. Because these hygrometers are so
accurate, they are widely used as a standard in many of the world’s
metrology laboratories.
Hygrometer FunctionFigure 1-2 on page 1-5 illustrates how GE Infrastructure Sensing
hygrometers detect and measure dew point. The condensate mirror is
illuminated with a high-intensity, solid state, light emitting diode
(LED). A photodetector monitors the LED light reflected from the
mirror. The photodetector is fully illuminated when the mirror is clear
of dew, and it receives less light as dew forms. A separate LED and
photodetector pair are used as a known reference to compensate for
any thermally induced changes in the optical components. The
photodetectors are arranged in an electrical bridge circuit, the output
current of which is proportional to the light reflected from the mirror.
The bridge output controls the electrical current to the thermoelectric
cooler.
A large bridge current develops when the mirror is dry, causing the
mirror to cool toward the dew point. As dew begins to form on the
mirror, less light is reflected, and the bridge output decreases. This, in
turn, causes a decrease in cooling current. A rate feedback loop
within the amplifier ensures critical response, causing the mirror to
stabilize quickly at a temperature that maintains a thin dew or frost
layer on the mirror surface. A precision thermometer element
embedded within the mirror directly monitors this dew point
temperature.
1-4Features and Capabilities
January 2006
Hygrometer CalibrationThe Optica unit can be sent to the National Institute of Standards and
Technology (NIST) in Gaithersburg, Maryland for certification or to
any National Standards lab for calibration against their primary
humidity standards. A calibrated instrument can then be used as a
transfer standard in local laboratories to calibrate lower echelon
instruments.
Caution!
Field calibration is not recommended.
Hygrometers used as calibration standards must have the following
characteristics:
• The mirror thermometer must have suitable long-term accuracy
(such as that obtained with a platinum resistance thermometer).
• A means should be provided for viewing the dew or frost
formation on the mirror.
LED
Regulation
Optical Reference
41.2°F
Gain
Thermoelectric
Heat Pump Power
Dew Point Temperature
(Precision Thermometer)
Figure 1-2: Chilled-Mirror Hygrometer Diagram
Features and Capabilities1-5
January 2006
Other Hygrometer
Applications
Many GE Infrastrucure Sensing Chilled Mirror Hygrometers are used
in industrial applications in addition to metrology. The optical
condensation hygrometer is not readily damaged or contaminated by
industrial process gases that can degrade other secondary
measurement schemes such as saturated salt and polymer-based
sensors. If the sensor or sampling components should become
contaminated with oils, salts, etc., they can be cleaned without harm
to the sensor or impairment to the system accuracy. The performance
of the hygrometer can be checked at any time by heating the mirror
above the dew point, causing the dew deposit to evaporate, then
reclosing the servoloop and checking to see that the system cools and
returns to the same dew point.
The GE Infrastructure Sensing optical condensation sensors cover a
wide range of applications limited only by the heat pumping
capabilities of the thermoelectrically-cooled mirror.
At high dew points (up to 100°C), the sensor is limited by the thermal
properties of the solid state optical components as well as the
thermoelectric heat pump capacity.
In a typical application measuring sub-ambient dew points, a twostage thermoelectrically-cooled mirror can reach a temperature
approximately 65°C lower than an ambient (heat sink) temperature of
+25°C. The thermoelectric cooler pumps heat from the mirror into the
heat sink. By reducing the temperature of the heat sink with a coolant
such as chilled water, or by applying the sensor in a low-temperature
condition such as monitoring of a test chamber, even lower dew
points can be measured. In meteorological applications where the
heat sink temperature is considerably lower, frost points down to –
75°C can be monitored.
Four- and five-stage sensors are available for measuring the lowest
dew/frost points.
The PACER CycleGE Infrastructure Sensing has developed and patented a
compensation technique called PACER (Programmable Automatic
Contaminant Error Reduction) that is very effective in reducing the
Raoult Effect error associated with soluble contaminants, particularly
for near-ambient dew points. The Optica is equipped with the PACER
cycle as well as AUTO balance as found on earlier models. The user
can choose which self-cleaning and balancing routine to run
depending on the severity of contamination.
The PACER cycle, diagrammed in Figure 1-3 on page 1-7, begins
with a coalescence period, during which the mirror is cooled well
below the dew point of the sample gas, condensing out a large amount
of water.
1-6Features and Capabilities
The PACER Cycle (cont.)
January 2006
Figure 1-3: A Typical PACER Cycle
This excess water easily dissolves any water-soluble contaminants.
The mirror is then heated. During the heating phase, the large puddles
of water gradually evaporate, carrying increasingly heavy
concentrations of salts as the puddles become smaller. Finally, when
all the puddles have evaporated, dry “islands” of crystallized salt are
left on the mirror. The area between the islands (80-85% of the mirror
surface) is now clean and shiny, whereas before the PACER cycle it
may have been completely covered. The total amount of
contamination has not been reduced, but instead, redistributed as
shown in Figure 1-4 below, with more clean mirror surface available
for dew formation. The reflected light signal is then electronically
balanced against the reference.
IntroductionThis chapter explains the installation of the benchtop and wall-mount
versions of the Optica, the various sensors used with the system, and
the I/O and power wiring.
Benchtop Installation
Mounting the Benchtop
The Optica benchtop dimensions are shown in Figure 2-1 below. Two
feet on the bottom of the case can be unfolded to raise up the front for
easier viewing, if desired.
Figure 2-1: Optica Benchtop Dimenisons
Installation2-1
January 2006
Using the Rack Mounting
Option
There is an optional kit for mounting the benchtop unit in a standard
19-inch rack (see Figure 2-2 below, Figure 2-3 on page 2-3, and
Figure 2-4 on page 2-4). The two brackets are attached to the front
panel using four No. 8 screws. The Optica is mounted to the brackets
using the eight mounting holes located just in front of and behind the
unit’s feet.
Figure 2-2: Optica Rack-Mount Adapter
2-2Installation
January 2006
1.11
(28.2)
4 places
FH-032-8
1.32
(33.5)
19.00 (482.6)
14.62 (371.3)
12.53 (318.3)
2.69
(68.3)
6.52
(165.6)
3.91
(99.3)
7.07 (179.6)
18.25 (463.6)
13.53 (343.7)
3.23 (82.0)
0.50 (12.7)
2.19 (55.6)
1.32
(33.5)
0.59 (15.0)
2.88
0.09
(2.3)
(73.2)
2.38
(60.5)
3.00
(76.2)
4.00
(101.6)
2.74 (69.6)
0.38
(9.7)
Note: Dimensions are in inches (millimeters).
8.75
(222.3)
Figure 2-3: Optica Rack-Mount Adapter - Front Panel
Installation2-3
January 2006
1.04
1.50
(38.1)
1.13 (28.7)
(26.4)
Note: Dimensions are in inches (millimeters).
11.00 (279.4)
1.00 (25.4)
1.44
(36.6)
0.66 (16.8)
0.75 (19.1)
1.09 (27.7)
0.55 (14.0)
6.00
(152.4)
5.86
(148.8)
3.91
(99.3)
12.41 (315.2)
1.00 (25.4)
1.00 (25.4)
0.80 (20.3)3.03 (77.0)
1.82
(46.2)
0.59 (15.0)
0.98 (24.9)
3.19 (81.0)
Figure 2-4: Optica Rack-Mount Adapter - Bracket
2-4Installation
January 2006
Wiring the Benchtop
Input PowerThe Optica operates with input power from 90 to 126 VAC at 4 amps,
or from 208 to 252 VAC at 2.5 amps. It is designed for a nominal 100,
115 or 230 VAC source. A switch on the rear panel selects the
appropriate voltage range (see Figure 2-5 below):
• The 115 VAC setting operates over a range of 90 to 126 VAC
• The 230 VAC setting operates over a range of 200 to 253 VAC
The Optica’s voltage and frequency rating are listed on the rear panel.
SensorsConnect the dew point sensor cable to the 25-pin connector in Slot B
on the Optica’s rear panel (see Figure 2-5 below). Connect the
optional temperature sensor cable to the 9-pin connector in Slot B on
the rear panel. The optional pressure sensor and other I/O wiring
connects to the terminal block in Slot A.
Optional
LAN
connector*
*
Serial
Port
Figure 2-5: Optica Benchtop Rear Panel
Note: *Depending on the model revision, the LAN connector is
located in one of two places.
Installation2-5
January 2006
Wall-Mount InstallationThe Optica Wall-Mount unit is designed to mount on a flat, vertical
surface, such as a wall or panel. To mount the wall-mount version, see
Figure 2-6 and Figure 2-7 below.
Wiring the Wall-MountAll connections to the wall-mount unit are made through the panel at
the bottom of the case as shown in Figure 2-8 below. Any I/O cabling
is brought into the unit through a gland at the lower left of the case
and connects to the terminal blocks on the left side of the case. Wiring
for these connections is shown in Figure 2-11 on page 2-8. The dew
point sensor and temperature sensor cable connectors are located near
the center of this panel.
Figure 2-8: Wall-Mount Wiring Entrance Locations
Optional Ethernet
LAN Connector
Gland for
I/O Wiring
Serial Port
LAN
IN/OUT
COM
TEMP
DEW POINT
Temperature Sensor
Dew Point Sensor
AC LINE
90/250 VAC 47-63 Hz
4 ASB 250V TYPE T
Gland for
AC Power
Wiring
Figure 2-9: Wall-Mount Bottom Panel (viewed from under the unit)
Installation2-7
January 2006
Input PowerPower wiring enters the case through a gland fitting at the lower right
of the unit and connects to a screw terminal block mounted on the
right side of the case. Wiring of this terminal block is shown in
Figure 2-10 below. The voltage and frequency rating and tolerances,
as well as fusing data, are listed on the bottom of the unit.
Top
FrontL
G
N
of
Case
Figure 2-10: Wall-Mount AC Power Wiring
Output WiringThe benchtop outputs are connected to removable terminal blocks on
the rear panel. Figure 2-5 on page 2-5 shows the location of the
benchtop terminal blocks and Figure 2-11 below shows the Slot A
connections.
ALARM 1
ALARM 2
OUT A
OUT B
PRESSURE
COM
NC
NO
COM
NC
NO
4-20
0-5
4-20
0-5
+V
V in
I in
RTN
The Wall-Mount Optica input/output terminal blocks are located
inside the front door as shown in Figure 2-8 on page 2-7. Cabling is
brought in through the gland on the bottom of the unit and wired to
the terminal blocks shown in Figure 2-11 below.
Note: Output programming is described in Chapters 4, 5 and 6.
Slot A
COM
AL1NC
NO
COM
AL2NC
NO
Front
of
Case
Top
OUT A
OUT B
PRES
4-20
0-5
4-20
0-5
+V
V in
I in
RTN
Benchtop Unit
Wall-Mount Unit
Figure 2-11: Benchtop and Wall-Mount Input/Output Terminal Blocks
2-8Installation
January 2006
Analog OutputsNote: When the Optica is being programmed, the analog outputs
provide 4–20mA and 0–5 VDC signals representing the
designated parameters.
• For 4–20mA output, connect to terminals labelled 4–20 (+) and
RTN (–).
Note: The maximum load allowed for current output is 500 Ohms.
• For 0–5 VDC output, connect to terminals labelled 0–5 (+) and
RTN (–).
Note: The maximum load allowed for voltage output is 5 mA.
Example:Assume a temperature output, scaled to range from 0°C (Tlower) to
100°C (Tupper), with a measured actual temperature of 23°C
(Tactual):
The voltage output is calculated by:
TactualTlower–()
Vout
----------------------------------------------
TupperTlower–()
5×=
yielding an output voltage of 1.15V.
230–()
---------------------
1000–()
The current output is calculated by:
TactualTlower–()
Iout mA()
yielding an output current of 7.68 mA.
----------------------------------------------
TupperTlower–()
230–()
---------------------
1000–()
5×1.15V=
16()×4+7.68mA=
204–()×4+=
Installation2-9
Loading...
+ 120 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.