GE
Sensing & Inspection Technologies
MG101
Panametrics Hygrometer
Calibration System
The MG101 hygrometer calibration system is used
to generate precise, repeatable levels of water
vapor in a carrier gas stream. This primary dew/
frost point generation system achieves an accuracy
of ±1.8°F (±1°C) within a continuously adjustable
range.
The MG101 employs the elementary principle of gas
dilution. Dry gas is piped into the system, where it is
divided into two streams. One stream is saturated
with water vapor at a known temperature, while
the other stream remains dry (see ow schematic
on next page). The two streams are then mixed. The
saturated stream is diluted with varying amounts of
the dry gas to produce the desired gas/water vapor
mixture.
The MG101 employs either single stage or two
stage dilution to generate gas with a dew/frost
point range of -103°F (–75°C) to 18°F (10°C) at 20°C
ambient temperature. The MG101 may be used to
verify a wide variety of instruments, including GE
Sensing aluminum oxide moisture probes. When
used with a chilled mirror reference, the system
provides an economical calibration system.
Features
• Generates precise, repeatable levels of
water vapor in a carrier gas
• Generation range is -103°F to 50°F (–75°C
to 10°C) frost/dew point (at typical ambient
temperatures)
• Accurate eld calibration/verication of
moisture sensors
• No power required to operate
• Requires dry nitrogen source gas
Applications
Portable eld calibration system for verifying
moisture and humidity sensors
![](/html/c6/c68d/c68dd5d560e52b21d620aec77b84796f2c620ee21d0ed7b560fea1a68e148607/bg2.png)
MG101 Specications
Generated Dew/Frost Point Temperature Range
–103°F to 50°F (–75°C up to 10°C) below ambient
temperature (equivalent to 1.2 ppmv to 12,120 ppmv
in gases at one atm at 68°F (20°C) ambient temperature)
Accuracy
±1.8°F (±1°C) dew/frost point temperature
Dry gas,
rst dilution
owmeter
Dry gas
inlet
Wet gas
owmeter
Vent valve
Dry down
valve
5μ lter
Saturator
bottle
Temp
Gas outlet
Dry gas,
second
dilution
owmeter
Wet gas,
diluted,
owmeter
Inlet Gas Supply
Nitrogen preferred, supply gas must be 45°F (25°C) drier
than driest dew/frost point temperature to be generated
Maximum Flow Rate
19.5 SCFH (9.2 Liters/minute)
Inert Gas Supply Pressure
55 psig to 60 psig (4 bar to 5 bar)
Outlet Pressure
Ambient to 10 psig (1 bar), adjustable
Power Requirements
None
Dimensions (w x h x d)
12 in x 18 in x 6 in
(304.8 mm x 457.2 mm x 152.4 mm)
Weight (Dry)
25 lb (11.3 kg)
Pressure
regulator
MG101 with Moisture Image® Series 1 analyzer to demonstrate
Flow schematic
moisture sensor calibration
Inlet and Outlet Connections
1/4 in Swagelok® tube tting