Joop Geesinkweg 140
1114 AB Amsterdam, The Netherlands
Australia Sponsor
FUJIFILM SonoSite Australasia Pty Ltd
114 Old Pittwater RD
Brookvale, New South Wales 2100, Australia
Caution:
SonoMB, SonoSite, Steep Needle Profiling, X-Porte, and the SonoSite logo are registered (in some jurisdictions) and unregistered trademarks
owned by FUJIFILM SonoSite, Inc. in various jurisdictions.
DICOM is the registered trademark of the National Electrical Manufacturers Association.
All other trademarks are the property of their respective owners.
Patents: US 8,216,146; US 8,213,467; US 8,147,408; US 8,137,278; US 8,088,071; US 8,066,642; US 8,052,606; US 7,819,807; US 7,804,970;
US 7,740,586; US 7,686,766; US 7,604,596; US 7,591,786; US 7,588,541; US 7,534,211; US 7,449,640; US 7,169,108; US 6,962,566; US
6,648,826; US 6,575,908; US 6,569,101; US 6,471,651; US 6,416,475; US 6,383,139; US 6,364,839; US 6,203,498; US 6,135,961; US
5,893,363 ; US 5,817,024; US 5,782,769; US 5,722,412; AU: 730822; AU: 727381; CA: 2,372,152; CA: 2,371,711; CN 98108973.9; CN:
98106133.8; CN: 97113678.5; DE: 69831698.3; DE: 69830539.6; DE: 69730563.5; DE: 602004027882.3; DE: 602004023816.3; DE:
Before servicing the X-Porte Ultrasound System, please read this manual.
The ultrasound system has multiple configurations and feature sets. All are described in this service manual
but not every option may apply to your system. System features depend on your system configuration,
transducer, and exam type.
Refer to the X-Porte Product Information & Safety Guide (P14646) for additional information regarding
safety, system controls, operation, capabilities, and specifications.
Audience
The intended audience of this manual are properly trained field and in-house service personnel. The
ultrasound system operators are expected to be able to replace all the components of the Desktop version
and the System “Engine” attached to the Stand version. With technical support guidance, a basic knowledge
of hand tools, and the ability to handle about 20 lb (9.1 kg), the system operators are expected to replace
the touch Control Panel and Clinical Monitor display. The Stand Power Management Unit (SPMU) and
batteries in the Stand version will be serviced by qualified field service personnel. The batteries used are
sealed Lithium-ion type which requires no maintenance and are not serviceable. Expected life for these
batteries are between 3-6 years depending on system usage and configuration of the Stand system.
Contact Information
Questions and comments are encouraged. FUJIFILM SonoSite is interested in your feedback regarding the
system and user documentation. Please call FUJIFILM SonoSite at 888-482-9449 in the U.S. Outside the
U.S., call the nearest FUJIFILM SonoSite representative.
For technical support, please contact FUJIFILM SonoSite as follows:
FUJIFILM SonoSite Technical Support
Phone
(U.S. or Canada):
Phone
(Outside U.S. and
Canada):
Fax:425-951-6700
E-mail:ffss-service@fujifilm.com
Web site:www.sonosite.com
Europe
Service
Center:
877-657-8118
425-951-1330
Or call your local representative.
FUJIFILM SonoSite - Amsterdam
Joop Geesinkweg 140
1114 AB Amsterdam
The Netherlands
Tel (Main): +31 20 751 2020
- English support: +44 14 6234 1151
- French support: +33 1 8288 0702
- German support: +49 69 8088 4030
- Italian support: +39 02 9475 3655
- Spanish support: +34 91 123 8451
Chapter 1: Introduction1
FUJIFILM SonoSite Technical Support (Continued)
Asia
Service
Center:
Terms and symbols
The service manual follows these conventions:
•A WARN ING describes precautions necessary to prevent injury or loss of life.
•A Caution describes precautions necessary to protect the products.
•Numbered steps in procedures must be performed in order.
•Items in bulleted lists do not require performance in sequence.
Labeling symbols
Please refer to Chapter 4: Labeling Symbols found in the X-Porte Product Information & Safety Guide
This chapter contains information regarding system specifications and accessory compatibility. The
information applies to the ultrasound system, transducers, accessories, and peripherals.
Specifications
System Specifications
Stand Dimensions:
•Length: 26.4 in. (67.1 cm)
•Width: 21.2 in. (53.8 cm)
•Height (max): 64 in. (162.6 cm)
•Height (min): 42.2 in. (107.2 cm)
•Height Adjustment: 9 in. (22.9 cm) travel
Touch Panel
•Capacitive Display Type
•Diagonal: 12.1 in. (30.7 cm)
•Multi-touch Gestures for System Controls
•Tilt Adjustment: 7.3 in. (18.5 cm)
•Side to Side Turning: +/- 90 degrees from center
Display (HD Monitor)
•Diagonal: 19 in. (48.3 cm)
•Screen Size: 1280 x 800
•Image Size: 800 x 600
•Monitor Tilt: 5° tilting forward from vertical, 20° tilting back from vertical
Electrical specifications
•Power Supply Input (Stand Version): 100-240V~ 6.0A Max, 50-60Hz
• Sony B/W Printer: 100-240V~ 1.5A-0.8A, 50-60Hz
•Power Supply Output 1 (Stand Version): 24 VDC, 11.5 A Max (Output not exceeding 275 watts)
• Power Supply Output 2 (Stand Version): 100-240V~ 3.0A - 1.5A, 50-60Hz
•Batteries (Stand Version): 3 Lithium-ion batteries (385Wh total)
•Battery Use Time: 1.0 hour, 3 days on idle
•Battery Charge Time: 2.5 hours
•Battery Life: 3-6 years
Chapter 2: Specifications3
Environmental limits
Note: The temperature, pressure, and humidity limits apply only to the ultrasound system, transducers,
and battery.
Operating
System and Transducer
10–40°C (50–104°F), 15–95% R.H.
700 to 1060hPa (0.7 to 1.05 ATM)
Shipping and Storage
System and Transducer
-35-65°C (-31-149°F), 15-95% R.H.
500 to 1060hPa (0.5 to 1.05 ATM)
Battery
-20–60°C (-4–140°F), 15–95% R.H. (For storage longer than 30 days, store at or below room temperature.)
500 to 1060hPa (0.5 to 1.05 ATM)
Battery specifications
The X-Porte stand uses three lithium-ion batteries that are charged and controlled by the Stand Power
Management Unit (SPMU) that can be switched off via three rocker type switches. These switches can be
accessed by lifting up from the stand bottom a black plastic door located under the left rear caster leg.
Run time is up to one hour (up to 3days on idle) depending on imaging mode and display brightness. This
chapter contains electrical, and clinical safety information required by regulatory agencies.
Accessories and peripherals
The system supports various accessories and peripherals. See the X-Porte Product Information & Safety Guide (P14646) for a list of compatible products.
4Chapter 2: Specifications
Chapter 3: Safety
This chapter contains electrical and clinical safety information required by regulatory agencies. The
information applies to the ultrasound system, transducers, accessories, and peripherals.
Electrical safety
This system meets EN60601-1, Class I/internally-powered equipment requirements and Type BF
(transducers) and Type CF (ECG leads) isolated patient-applied parts safety requirements.
This system complies with the applicable medical equipment requirements published in the Canadian
Standards Association (CSA), European Norm Harmonized Standards, and Underwriters Laboratories (UL)
safety standards.
For maximum safety observe the following warnings and cautions..
WARNING:
WARNING:
WARNING:
To avoid the risk of injury, do not operate the system in the presence of
flammable gasses or anesthetics. Explosion can result.
To avoid the risk of electrical shock or injury, do not open the system
enclosures. All internal adjustments and replacements, including battery
replacement, must be made by a qualified technician.
To avoid the risk of electrical shock:
• Use only properly grounded equipment. Shock hazards exist if the power
supply is not properly grounded. Grounding reliability can be achieved only
when equipment is connected to a receptacle marked “Hospital Only” or
“Hospital Grade” or equivalent. The grounding wire must not be removed or
defeated.
• This equipment must be connected only to supply mains with protective earth.
• Do not let any part of the system (including the bar code scanner, external
mouse, power supply, or power supply connector), except for the transducer
or ECG leads, touch the patient.
• Do not touch any of the following:
• The ungrounded signal input/output connectors on the back of the
ultrasound system
• The system battery contacts (Stand System Only)
• The system transducer connector when the transducer or Triple
Transducer Connect (TTC) is disconnected
• Any unused TTC transducer connector when the TTC is connected.
• Do not connect the system's power supply to an MPSO or extension cord.
• Before using the transducer, inspect the transducer face, housing, and cable.
Do not use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the
system.
• Do not use any transducer that has been immersed beyond the specified
cleaning or disinfection level. See “Cleaning and Disinfecting X-Porte
Products” P16529
• Use only accessories and peripherals recommended by FUJIFILM SonoSite,
including the power supply. Connection of accessories and peripherals not
recommended by FUJIFILM SonoSite could result in electrical shock. Contact
FUJIFILM SonoSite or your local representative for a list of accessories and
peripherals available from or recommended by FUJIFILM SonoSite.
Chapter 3: Safety5
WARNING:
To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular
basis. Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system
or to mains power must only be used with the power supply, and cannot be
used to connect other devices to mains power.
WARNING:
WARNING:
WARNING:
WARNING:
Caution:
To avoid the risk of electrical shock, burn, or fire, use only AC power supplies
specified by FUJIFILM SonoSite.
To prevent injury to the operator/bystander, the transducer must be removed
from patient contact before the application of a high-voltage defibrillation pulse.
To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment
before clinical use. Connecting additional equipment to the ultrasound system
constitutes configuring a medical system. FUJIFILM SonoSite recommends
verifying that the system, all combinations of equipment, and accessories
connected to the ultrasound system comply with JACHO installation
requirements and/or safety standards such as AAMI-ES1, NFPA 99 OR IEC
Standard 60601-1-1 and electromagnetic compatibility standard IEC 60601-1-2
(Electromagnetic compatibility), and are certified according to IEC Standard
60950 (Information Technology Equipment (ITE) or IEC 60601-1.).
Due to the captive nature of the Stand version system batteries, any shipping
transport, service activities, or extended periods of non-operation with the
system unplugged from the AC wall source, will require that the battery
ON-OFF switch be set to the “•O” (OFF) position. Please refer to Figure 3.1 on
page 6.
Do not use the system if an error message appears on the display monitor:
note the error code; call FUJIFILM SonoSite or your local representative; turn
off the system by pressing and holding the power key until the system powers
down.
Caution:
To avoid increasing the system and transducer connector temperature, do not
block the airflow to the rear fans and ventilation holes around the bottom and
side perimeter of the Engine.
6Chapter 3: Safety
Figure 3.1 Stand System Battery Switches & Transport Positioning
Access to the Stand battery switches can be gained by
lifting the black plastic panel and pulling outward. This
panel location is at the SPMU stand housing area near
the left rear caster leg.
Clinical Monitor display in the down (horizontal)
transport position.
Touch Control Panel in the down (horizontal)
transport position.
Battery
Switches
Stand collapsed
(lowest height)
for transport.
“•O”= OFF
Electrical safety classification
Equipment safety
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning all safety and standards applicable to the X-Porte Ultrasound System
DICOM standard
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning the DICOM standard applicable to the X-Porte Ultrasound System
HIPAA standard
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning HIPAA standard applicable to the X-Porte Ultrasound System
Transporting X-Porte Stand Systems
Please observe the prior Warnings and Cautions outlined in this chapter concerning the stand version
system for transport and servicing. Lower (collapse) the stand to the lowest height position, fold to the
horizontal position the Clinical Monitor display and touch Control Panel, and access the three battery
switches and ensure that all switches are in the “•O” (OFF) position. Many government agencies require
that no electronic equipment be shipped via air transport with live battery power attached to the equipment.
Refer toFigure 3.1 on page 6as to how to position the system stand and disconnect the battery power for
transport and servicing.
Chapter 3: Safety7
8Chapter 3: Safety
Chapter 4: System Overview
About the System
The FUJIFILM SonoSite X-Porte high-resolution ultrasound system is a portable, full featured, general
purpose, software controlled, diagnostic ultrasound system using all digital architecture. The system is used
to acquire and display high-resolution, real-time ultrasound data in 2D, M Mode, Pulsed Wave (PW)
Doppler, Continuous Wave (CW) Doppler, Color Power Doppler (CPD), and color Doppler (Color) or in a
combination of these modes.
The system has an electrocardiography (ECG) display feature and supports a 3-lead ECG cable assembly
to collect data for M Mode and Doppler measurements. The system provides measurement capabilities for
anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The
system has a PW and CW Doppler audio output feature, cine review, image zoom, labeling, biopsy,
measurements and calculations, image storage and review, printing, and recording capabilities.
The system includes optional Digital Imaging and Communications (DICOM) capabilities as well as general
computer communication capabilities to provide the acceptance, transfer, display, storage, and digital
processing of ultrasound images and loops. Security support is also provided to facilitate HIPAA
compliance.
The system displays the current output level in terms of one of two bioeffects indices (“Mechanical Index
[MI]” and “Thermal Index [TI]”) in accordance with the AIUM/NEMA Standard for Real Time Display of
Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.
Theory of Operation
The X-Porte ultrasound systemhas seven (7) major functional groups:
• Transducer
• Acquisition Subsystem
• Processing Subsystem
• Display Subsystem
• Control Subsystem
• User Interface Subsystem
• Power Subsystem
The following block diagram shows the relationship of the functional groups.
Chapter 4: System Overview9
Figure 4.1 Ultrasound System Functional Group
The Transducer elements convert the pulser voltage to acoustic energy during the transmit portion of the
ultrasound acquisition cycle. The elements convert the acoustic echo to voltage in the receive portion of the
acquisition. The voltage developed on the transducer elements is sensed by the acquisition subsystem. The
system transducers have 64 to 256 elements.
The Acquisition Subsystem consists of the beamformer and interface to the transducer. The beamformer
controls the timing of the transmit pulses to focus the acoustic beam. The beamformer amplifies the
low-level received echos and controls the receive focusing. The system beamformer transmits on up to 128
elements and receives on 64 elements.
The Processing Subsystem includes capabilities for interfacing with the beamformer and performing high
speed processing. The processing subsystem demodulates, filters, detects, and compresses the signal
supplied by the beamformer into display information.
The Display Subsystem converts the detected ultrasound data into picture elements (pixels). The software
user interface graphics are combined with the ultrasound information and converted to a video stream. The
external video port supports NTSC and PAL format.
The Control Subsystem consists of the central processing unit, program and video memory, permanent
image storage and retrieval memory, external communication interface ports, and connection to the user
interface keys. The control software includes the acoustic power and intensity software subsystem, power
group monitors, and a beamformer monitor. This software guarantees a level of patient safety by ensuring
the system is operating within acoustic power and intensity limits.
The User Interface Subsystem represents the software interface and form factor. The software interface
is the interaction between the user and the screen layout components. The form factor is the type of physical
buttons, location, and grouping of the buttons and the device size, shape, and weight. Dedicated controls
are for high usage activities and grouped according to the user workflow.
The Power Subsystem provides the system power and protects the hardware from destructive and/or
unsafe conditions by detecting failures in the system through hardware and software monitors. Detection of
a fault results in disabling of the pulser supply, and signaling of an error to the Control Group. The power
subsystem includes the batteries (quantity of three lithium-ion) and battery charging electronics.
Description of Operating Modes
2D Mode2D mode is a two dimensional image of the amplitude of the echo signal. It is used for
location and measurement of anatomical structures and for spatial orientation during
operation of other modes. In 2D, a two-dimensional cross-section of a 3-dimensional
soft tissue structure such as the heart is displayed in real time. Ultrasound echoes of
different intensities are mapped to different gray scale or color values in the display.
The outline of the 2D cross-section may be a rectangle, parallelogram, trapezoid,
sector, or a full circle, depending on the particular transducer used. 2D mode can be
used in combination with any other modes.
MModeM Mode is also known as “T-M mode” or “time-motion” mode. It is used primarily for
cardiac measurements such as valve timing and septal wall thickness when accurate
timing information is required.
Ultrasound echoes of different intensities are mapped to different gray scale values in a
scrolling display. M Mode displays time motion information of the ultrasound data
derived from a stationary beam. Depth is arranged along the vertical axis with time
along the horizontal axis. M Mode can be used alone but is normally used in
conjunction with a 2D image for spatial reference. The 2D image has a graphical line
(M-line) superimposed on the 2D image indicating where the M Mode beam is located.
10Chapter 4: System Overview
Color
Doppler
(Color)
In color Doppler, a real-time, two-dimensional cross-section of blood flow is displayed.
The 2D cross-section may be presented as a rectangle, parallelogram, trapezoid,
sector, or a full circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used
to represent the velocity, both positive and negative, of the blood flow echoes. Often, to
provide spatial orientation, the full color blood flow cross-section is overlaid on top of
the gray scale cross-section of soft tissue structure (2D echo). For each pixel in the
overlay, the decision of whether to display VCD, gray scale (echo) information or a
blended combination is based on the relative strength of echoes from the soft-tissue
structures and from the red blood battery.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly
moving structures. Tissue motion is discriminated from blood flow by assuming that
blood is moving faster than the surrounding tissue, although additional parameters may
also be used to enhance the discrimination. The remaining signal after wall filtering
may be averaged over time (persistence) to present a steady state image of blood flow
distribution. Variance information may also be displayed to provide information when
large variance is observed in the velocity information.
Color Power
Doppler
(CPD)
Continuous
Wave (CW)
Doppler
In CPD, a real-time two-dimensional cross-section of blood flow is displayed. The 2D
cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a
full circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used
to represent the power in blood flow echoes. Often, to provide spatial orientation, the
full color blood flow cross-section is overlaid on top of the gray scale cross-section of
soft tissue structure (2D echo). For each pixel in the overlay, the decision of whether to
display CPD, gray scale (echo) information or a blended combination is based on the
relative strength of echoes from the soft-tissue structures and from the red blood
battery.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly
moving structures. Tissue motion is discriminated from blood flow by assuming that
blood is moving faster than the surrounding tissue, although additional parameters may
also be used to enhance the discrimination. The power in the remaining signal after
wall filtering may be averaged over time (persistence) to present a steady state image
of blood flow distribution.
CW provides a real-time representation of blood flow and is displayed as a
velocity-versus-time sweeping output. Velocity (or frequency) i
s presented as the
vertical axis with time along the horizontal axis. The magnitude of the detected signal is
represented as different gray scale values.
CW Doppler mode provides the clinician with the ability to obtain blood flow velocities
focused about a user specified focal region. A continuous transmit waveform of
ultrasound energy with a known frequency is transmitted and focused by the system;
on the receive side, the transducer receive echoes are continuously amplified, focused
about the focal region and converted to a base band quadrature signal. The signal is
analyzed by a quadrature phase detector that establishes two receive channels to
allow detection of flow direction. These two channels are then analyzed by a fast
complex Fourier transform (FFT) circuit to establish the spectrum of frequencies
present in the echoes. The data are displayed as spectrum frequencies with respect to
time.
CW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located.
Chapter 4: System Overview11
Pulsed Wave
(PW)
Doppler
PW provides a real-time representation of blood flow and is displayed as a
velocity-versus-time sweeping output. Velocity (or frequency) is presented as the
vertical axis with time along the horizontal axis. The magnitude of the detected signal is
represented as different gray scale values. The ultrasound data is derived from a single
area, the sample volume, on a stationary beam.
PW Doppler mode provides the clinician with the ability to obtain blood flow velocities
about a spatial sample volume. A burst of ultrasound with a known spectrum is
transmitted by the system; on the receive side, the transducer receive echoes are
amplified and range gated at the appropriate depth. The signal is analyzed by a
quadrature phase detector that establishes two receive channels to allow detection of
flow direction. These two channels are then analyzed by a fast complex Fourier
transform (FFT) circuit to establish the spectrum of frequencies present in the echoes.
The data are displayed as spectrum frequencies with respect to time.
PW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located. The sample volume position (depth) and
size are also indicated on the D-Line.
Additional System Feature Performances
Broadband ImagingThis ultrasound acquisition system uses high resolution broadband technology
in the transmit pulsers, transducer, and receivers. The receive path can capture
and process signals over a wide spectrum, from below 2.0 MHz to beyond 10
MHz. For each application, the transmit pulse is designed to produce an
appropriate bandwidth. For example, in 2D grayscale imaging, a wide band
pulse is used to support good axial resolution. For Doppler modes, a narrower
band pulse is used, which improves the spectral resolution of the detected
Doppler signal.
In addition to transmit pulse control, programmable digital signal processing is
used in the receive path to further refine the bandwidth used to produce the
final image. Digital filters are applied to the digitized received signal to limit and
shape the spectral bandwidth used to generate the displayed output.
Tissue Specific
Imaging
Biopsy GuidanceThe system can display a pair of biopsy guidelines that represent the
Measurement and
Calculation
Capabilities
In this feature, parameters for signal and image processing are optimized to
maximize the image quality or to obtain the best compromise of resolution and
penetration for different specific clinical applications. These parameters
include: the order of received filters, the bandwidth, the dynamic range, the
compression curve, the gain setting and parameters for compounding
frequency band, etc. For example, different system parameter setups are used
for abdominal or peritoneal scanning. This feature is for ease
operator by automatically setting up system control parameters rather than
manually adjusting settings for best performance.
anticipated path of the biopsy needle. The image of an anatomical target,
biopsy guidelines, a scan plane marker, and a biopsy needle are displayed to
assist in guiding the biopsy needle to the target. The system also provides
needle guidance for vascular access procedures. For additional information,
see the biopsy user guides.
The system offers a variety of measurements and calculations, specific to
exam type and transducer. A list of them, and author references, are in the
system user guide. Measurement accuracy is also discussed.
of use for the
12Chapter 4: System Overview
Continuous Wave
Receive Section
X
T/R Sw
TGC Amplifier
A/DDelay
Weight
Pulser
Waveform
Delay
X
T/R Sw
TGC Amplifier
A/DDelay
Weight
Pulser
Waveform
Delay
Beamformed
Data
Transducer
Transmit Section
Control
Doppler Audio
Output
The system provides for audio output of the CW velocity information. This can
be presented as stereo information, with flow moving towards the transducer
on one channel and flow away on the other, or as a mono output with the single
audio output representing the summation of the flow directions.
Pulsed Wave
Doppler Audio
Output
Electrocardiograph
(ECG) Display
Front End Overview
The Front End is designed to support various imaging modalities such as 2D, M-Mode, Spectral Doppler
and Color Doppler. From the Front End's perspective, all modes can be grouped into a few basic types:
Single mode, simultaneous modes and triggered modes. All these modes are built from similar, basic
transmit and receive sequences controlled within the Front End. A generic top level block diagram of a
typical Front End is shown below in Figure 4.2 on page 13
The system provides for audio output of the PW velocity information. This can
be presented as stereo information, with flow moving towards the transducer
on one channel and flow away on the other, or as a mono output with the single
audio output representing the summation of the flow directions.
ECG is provided to measure the electrical signal generated by the heart. A
three lead interface: Right Arm (RA), Left Arm (LA) and Left Leg (LL), is
provided on the system.
The ECG signal is displayed as an amplitude-versus-time sweeping output.
Amplitude is presented on the vertical axis with time along the horizontal axis.
Figure 4.2 Generic Front End Block Diagram
The transmit section consists of a waveform generator, delay block, and high power high voltage driver to
excite the transducer element. Multiple elements are driven with delays determined by the time of flight in
the medium from the elements to the point in space where the beam is to be focused. The longer the time
of flight is to the focal point the smaller the delay is for a given transmit element to allow all to arrive at the
Chapter 4: System Overview13
focal point at the same time. The number of elements driven is determined by element sensitivity off axis
and depth of field considerations. The waveform is selected to drive the transducer at a certain center
frequency, bandwidth, and power and is optimized for the given mode.
The receive section consists of a transmit/receive switch to protect the receiver from the transmit voltage, a
variable gain receiver to amplify and condition the return echoes, an A/D to digitize the data, a delay block
to focus the return signals and a weight block to scale the return echoes for each channel. All the signals
are then summed together to generate the beam-formed receive data. The analog gain varies with depth
to compensate for signal attenuation through the medium. The delays and weights are independent for
each channel. The delay and weight for the receive channel can typically be changed dynamically to keep
the receive beam in continuous focus. The delay is simply set by the time of flight in the medium from the
point of interest to the element, which starts at skin-line and proceeds to the deepest depth of interest.
The control section drives the data to the various data path elements on a line by line basis, controls the
timing of the transmit and receive sections and controls the tagged information and timing of the data to the
rest of the system.
X-Porte Front End Board
The X-Porte ultrasound system can be divide into two board section groups: Front End and Back End.
Below is the block diagram linking the necessary functional groups of an ultrasound system as they apply
to the X-Porte system. Please refer to Figure 4.3 on page 14
Figure 4.3 X-Porte Front End Board Block Diagram
14Chapter 4: System Overview
PW Doppler Processing
Doppler processing includes both audio processing which presents Doppler signal in the form of stereo
audio and spectral processing which generates data for display of Doppler spectrum in the form of a
scrolling spectrogram. Doppler power spectrum is estimated performing Discrete Fourier Transforms on
short, overlapped segments of wall filtered Doppler signal. Doppler audio data is generated from wall filtered
data by phase shifting the in-phase component. Refer to the PW Doppler Processing Function Block
Diagram Figure 4.4 on page 15
Figure 4.4 PW Doppler Processing Function Block Diagram
Chapter 4: System Overview15
CW Doppler Processing
CW Doppler data will be presented to the signal processor as complex (I/Q) data from the analog front end
of the external DSP. The 20-bit data will be presented as consecutive samples at a data rate varying from
1.5 kHz to 64 kHz for the complex pair. Most of CW processing is similar to that of PW except for the QBP
function. In place of QBP will be a low pass decimating filter that operates on incoming I/Q data.
The Doppler Processing block must allow storage of 256 undetected I/Q pairs in to allow the system to
measure and correct for phase mismatch. Measuring and correcting will need to be accomplished in system
software. Refer to the CW Doppler Processing Function Block Diagram Figure 4.5 on page 16
Figure 4.5 CW Doppler Processing Function Block Diagram
16Chapter 4: System Overview
Back End Overview
The Back End subsystem is responsible for the conversion of raw acquisition data into a raster image ready
for display. The Back End subsystem also contains the video data path that supports generation of video
comprising of the ultrasound image as well as graphics annotation. Video generation of both standard
composite interlaced video and progressive scan video is supported. Most functionality is within the ASIC
but the memory resources for acquisition memory, and display memory are found in external memory
components. The conversion from PC type video to TV type video is also performed externally.
Control is received initially from the CPU to setup each functional block and afterward the hardware is
completely data driven. This control takes the form of programming setup registers inside the blocks and
setting up scan conversion tables. Each block provides temporary storage as required to buffer data and
keep their respective processing pipeline full and operating. Also note that the block diagrams show only
the data path, but each block is responsible for generating any necessary memory addresses for their
respective input data stream.
The X-Porte Back End subsystem is shown in Figure 4.6 on page 17.
Figure 4.6 X-Porte System Back End Subsystem Block Diagram
The Back End Subsystem performs processing encompassing three main data domains, acquisition data,
raster data, and video data. Support for acquisition data includes the input buffer, flash suppression, frame
average, and external ACQ memory. Cine buffer management is performed by the acquisition controller.
Conversion from acquisition data to raster data is performed by the graphics overlay, scan conversion system, sweeping system, and 3D system. Raster data is stored in an external DISPLAY memory. Also supporting raster operations is the graphics support block that provides acceleration hardware for pixel
operations from the CPU and graphics overlay system. The progressive video path includes buffers, priority logic, and LUTs.
Chapter 4: System Overview17
Control Subsystem
LVDS (24 bit)
SDVO 720P
HDMI/DVI
transmitter
AD9889B or
equivalent
Display Control I/F
Display Control I/F
U
S
B
H
o
s
t
x
3
Control FPGA
Interrupts from the Front
End,and FPGA.
Processor
Audio Codec/
Amp
Hard IP Block in FPGA
Soft IP Block in FPGA
Ethernet
W
i
F
i
PCI Express
Video out
(
LVDS)
CH7308A LVDS(24 bit)
SATA
2
Configuration
FLASH/CPLD
Program
Control/Processor Section
Main Clinical display for
POC. If M2/S2 to Dock for
Aux Monitor with Audio
If M2/S2: Main Clinical display
If POC: Touch Screen
JTAG
GPIO
SATA 2
RJ45
Connector
(Debug only)
FE I/F Connector
Hirose:
(Debug only)
JTAG
Mini PCIe
(Debug Only)
PCI Express for WiFi
Video Buffer Ram
DDR3 1 GByte
GPIOs to Dock (TBD)
PCI Express for Dock
Dock Connector
SATA Connector
COMe Type 10
10/100/1G Ethernet
U
SB
H
o
s
t
Backlight enable and brightness
USB Host for Dock
Serial Port
Speakers
HDA
USB Type A
(x3)
SATA Boot
Device
Connector
Speakers
Switch
10
/
100
/
1
G
Ethernet
10/
100/
1G Ethernet
Switch
HDMI/DVI
receiver
TP401A or
equivalent
HDMI in
USB Module
USB
Line In
Audio Line In
24 bit Video to
VGA
ADV7123 or
Equivalent
Some GPIOs
tied to both
FPGA and
CPLD
SATA2
Serial Lite
(
x
4
)
,
WakeFromFE
,
FE
_
Detect
24
bit video
24
bit video
S/PDIF
S/PDIF
Note:
HDMI with
Audio
HeadphonesHeadphones
Serial Port
I
2
C
USB Host for Touch Screen
The X-Porte Control Subsystem is shown in the figure below.
Figure 4.7 Control Subsystem Block Diagram
The core control subsystem contains the processor, the system bus, and the system memory resources of
RAM and solid state hard drives. These two solid state hard drive devices are responsible for the system
operating
system software, ultrasound program software, scanhead operation files, on-board training videos, and
storage for patient data (patient demographic information, images, and video clips)
18Chapter 4: System Overview
The main operational solid state storage drive is a mSATA form factor device totalling 16 GB of storage
capacity. This device is responsible for handling the operational system software, the ultrasound program,
and the on-board training videos.
A secondary solid state storage drive of 80 GB with an 1.8 inch form factor is used to handle the scanhead
data files, other ultrasound system files, and 66 GB of storage capacity reserved for all patient data. It is
with these new solid state devices, the X-Porte architecture is similar to our past products utilizing CF and
SD storage cards, but with much improved data storage and throughput without the use of any mechanical
hard drive devices.
Power Supply and Control
The X-Porte ultrasound system utilizes one of two types of power supply arrangements depending on the
X-Porte system version, which are Desktop or Stand. The Desktop system uses two separate external
power supply (EPS) units. The 150 watt EPS supplies +24VDC to the main Ultrasound System Electronics
(Engine) and a separate 65 watt EPS is utilized to supply +24VDC for the Clinical Display Monitor. No
batteries can be utilized with this version of system. The electrical block diagram for the 150 watt EPS is
shown below in Figure 4.8 on page 19
Figure 4.8 Desktop External Power Supply (EPS)
The Stand version X-Porte system utilizes a power supply/power control module mounted under the stand
base. This System Power Management Unit (SPMU) supplies the +24VDC power to the Engine and Clinical
Display Monitor. This SPMU is also the mounting point for the three lithium-ion batteries, charges the
batteries, and is responsible for switching the system from AC power to battery power when needed. Please
refer to Figure 4.9 on page 20
Chapter 4: System Overview19
Figure 4.9 Stand Power Management Unit (SPMU)
ECG Module
The ECG module allows a representation of the heart electrical activity to be displayed in real time with
ultrasound images acquired and displayed on the system video display.
The ECG module interfaces to the patient through three (3) ECG leads: Right Arm ECG lead (RA), Left Arm
ECG lead (LA), and Left Leg ECG lead (LL). The ECG received signal from the ECG electrodes are isolated,
amplified, and filtered by the ECG module before it is sent to the system for further processing and display.
The ECG module and cable are an integrated assembly. The module receives power from the system.
Patient isolation is provided by the ECG module, allowing the connection and signals to the system to be
system-ground referenced. The isolation between the patient and the system meets the requirements of IEC
601-1 for Type CF equipment.
The ECG function accepts input from an external serial A/D and performs gain, filter, and DC Offset
functions. The resultant data is output at either the 200Hz sample rate or decimates the data by 2 or 4 and
outputs the data into acquisition memory. The data is assumed to be signed. An interrupt is provided that
will interrupt the processor after a set delay from the detected level and slope. A simple block diagram is
shown below in Figure 4.10 on page 21
20Chapter 4: System Overview
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.