close to the product's operating location or
directly with the product.
Flowserve products are designed, developed and
manufactured with state-of-the-art technologies in
modern facilities. The unit is produced with great
care and commitment to continuous quality control,
utilizing sophisticated quality techniques, and
safety requirements.
Flowserve is committed to continuous quality
improvement and being at your service for any
further information about the product in its
installation and operation or about its support
products, repair and diagnostic services.
These instructions are intended to facilitate
familiarization with the product and its permitted
use. Operating the product in compliance with
these instructions is important to help ensure
reliability in service and avoid risks. The
instructions may not take into account local
regulations; ensure such regulations are observed
by all, including those installing the product.
Always coordinate repair activity with operations
personnel, and follow all plant safety requirements
and applicable safety and health laws/regulations.
These instructions must be read prior to
stalling, operating, using and maintaining the
equipment in any region worldwide. The
equipment must not be put into service until all
the conditions relating to safety, noted in the
instructions, have been met. Failure to follow
and apply the present user instructions is
considered to be misuse. Personal injury,
product damage, delay or failure caused by
misuse are not covered by the Flowserve
warranty.
1.2 CE Marking and approvals
It is a legal requirement that machinery and
equipment put into service within certain regions of
the world shall conform with the applicable CE
Marking Directives covering Machinery and, where
applicable, Low Voltage Equipment, Electromagnetic
Compatibility (EMC), Pressure Equipment Directive
(PED) and Equipment for Potentially Explosive
Atmospheres (ATEX).
Where applicable, the Directives and any additional
Approvals, cover important safety aspects relating to
machinery and equipment and the satisfactory
provision of technical documents and safety
instructions. Where applicable this document
incorporates information relevant to these Directives
and Approvals.
To confirm the Approvals applying and if the product
is CE marked, check the serial number plate
markings and the Certification. (See Section 8)
1.3 Disclaimer
Information in these User Instructions is believed
to be complete and reliable. However, in spite of
all of the efforts of Flowserve Corporation to
provide comprehensive instructions, good
engineering and safety practice should always be
used.
Flowserve manufactures products to exacting
International Quality Management System Standards
as certified and audited by external Quality Assurance
organizations. Genuine parts and accessories have
been designed, tested and incorporated into the
products to help ensure their continued product
quality and performance in use. As Flowserve cannot
test parts and accessories sourced from other
vendors the incorrect incorporation of such parts and
accessories may adversely affect the performance
and safety features of the products. The failure to
properly select, install or use authorized Flowserve
parts and accessories is considered to be misuse.
Damage or failure caused by misuse is not covered
by the Flowserve warranty. In addition, any
modification of Flowserve products or removal of
original components may impair the safety of these
products in their use.
1.4 Copyright
All rights reserved. No part of these instructions
may be reproduced, stored in a retrieval system or
transmitted in any form or by any means without
prior permission of Flowserve.
1.5 Duty conditions
This product has been selected to meet the
specifications of your purchase order. The
acknowledgement of these conditions has been
sent separately to the Purchaser. A copy should be
kept with these instructions.
The product must not be operated beyond
the parameters specified for the application. If
Page 3 of 20flowserve.com
Page 4
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
there is any doubt as to the suitability of the
product for the application intended, contact
Flowserve for advice.
If the conditions of service on your purchase order are
going to be changed (for example: liquid pumped,
temperature or duty) it is requested that the user
seeks the written agreement of Flowserve before start
up.
1.6 Safety
1.6.1 Summary of safety markings
These User Instructions contain specific safety
markings where non-observance of an instruction
would cause hazards. The specific safety markings
are:
This symbol indicates electrical safety
instructions where non-compliance will involve a
high risk to personal safety or the loss of life.
This symbol indicates safety instructions
where non-compliance would affect personal safety
and could result in loss of life.
This symbol indicates “hazardous and toxic
fluid” safety instructions where non-compliance
would affect personal safety and could result in loss
of life.
This symbol indicates safety
instructions where non-compliance will involve some
risk to safe operation and personal safety and would
damage the equipment or property.
This symbol indicates explosive atmosphere
zone marking according to ATEX. It is used in
safety instructions where non-compliance in the
hazardous area would cause the risk of an
explosion.
This symbol is used in safety instructions to
remind not to rub non-metallic surfaces with a dry cloth; ensure the cloth is damp. It is used in safety
instructions where non-compliance in the
hazardous area would cause the risk of an
explosion.
This sign is not a safety symbol but
indicates an important instruction in the assembly
process.
1.6.2 Personnel qualification and training
All personnel involved in the operation, installation,
inspection and maintenance of the unit must be
qualified to carry out the work involved. If the
personnel in question do not already possess the
necessary knowledge and skill, appropriate training
and instruction must be provided. If required the
operator may commission the
manufacturer/supplier to provide applicable
training.
Always coordinate repair activity with operations
and health and safety personnel, and follow all
plant safety requirements and applicable safety
and health laws and regulations.
1.6.3 Safety action
This is a summary of conditions and actions to
help prevent injury to personnel and damage to
the environment and to equipment. For
products used in potentially explosive
atmospheres section 1.6.4 also applies.
NEVER DO MAINTENANCE WORK
WHEN THE UNIT IS CONNECTED TO POWER
(Lock out.)
HANDLING COMPONENTS
Many precision parts have sharp corners and the
wearing of appropriate safety gloves and
equipment is required when handling these
components. To lift heavy pieces above 25 kg (55
lb.) use a crane appropriate for the mass and in
accordance with current local regulations.
1.6.4 Products used in potentially explosive
atmospheres
Measures are required to:
Avoid excess temperature
Prevent buildup of explosive mixtures
Prevent the generation of sparks
1.7 Specific machine performance
For performance parameters see section 1.5 Duty
conditions. Where performance data has been
supplied separately to the purchaser these should be
obtained and retained with these User Instructions if
required.
1.8 Specifications
1.8.1 SELD-103 Unit Specifications
Operating frequency: 868 MHz or 900 MHz
Channels: (External) 4-GP, 1-Gas, 2-Inst,
2-DI, 1-RS-485
Channels: (Internal)1-Battery Voltage, 1-Onboard Temperature
Ports: 5-Sensor ports / 1-Diagnostics port (RS-
485)
Page 4 of 20flowserve.com
Page 5
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Transmission rates: 5 seconds or higher
Operating temperature: -20°C to +60°C (-40°F to
+185°F)
Humidity: 0-95% noncondensing
Power requirement: 3.6 VDC Battery Pack
(S1-1043)
Base: 316LSS
Base Mounting: Magnetic base, unistrut mount
Sensor Cable Connection: 6 pin amphenol style
1.8.2 Optional add-on sensor specifications
RTD: 1000 or 100 Ohm (Platinum, 2 wire)
Thermocouple: Type: J, K, N, T, etc. (+ Cold
Reference and Polynomial)
Pressure: Vacuum to 10Kpsi (Specific sensor for
each range)
Vibration (Velocity): Single Axis or Tri Axis 0-25
mm/sec (0-1 ips)
Vibration: (Acceleration): 0 - 10G Peak
Discrete: Dry Contact Only
Gas: See test sheet for details
2 TRANSPORT AND STORAGE
Make sure that hazardous substances are
disposed of safely and that the correct personal
protective equipment is used. The safety
specifications must be in accordance with the
current local regulations at all times.
2.1 Consignment receipt and unpacking
Immediately after receipt of the equipment it must
be checked against the delivery/shipping
documents for its completeness and that there has
been no damage in transportation. Any shortage
and/or damage must be reported immediately to
Flowserve and must be received in writing within
ten days of receipt of the equipment. Later claims
cannot be accepted.
Check any crate, boxes or wrappings for any
accessories or spare parts that may be packed
separately with the equipment or attached to side
walls of the box or equipment.
Each product has a unique serial number. Check
that this number corresponds with that advised and
always quote this number in correspondence as
well as when ordering spare parts or further
accessories.
2.1.1 Unpacking
Carefully open package received from
factory and remove protective wrapping from
SELD-103 units and accessories. Inspect all
hardware for damage. Report any damage to
shipping carrier immediately.
Ensure you have received the correct sensors and
mounting accessories for your application. Crosscheck hardware received versus packing slip and
purchase order.
The SELD-103 contains sensitive electronic
components that may be damaged by static
electricity present in work environments. The
following precautions are recommended to reduce
the risk of damage due to electrostatic discharge:
Only wipe exterior of unit with a damp cloth
Use of a grounding wrist strap when removing
outer plastic cover is recommended
2.2 Handling
Boxes, crates, pallets or cartons may be unloaded
using fork lift vehicles or slings dependent on their
size and construction.
2.3 Storage
Store the equipment in a clean, dry
location. The SELD-103 unit should be powereddown prior to storage (see section 4
MAINTENANCE for how to power down unit) to
preserve battery life.
2.3.1 Storage and packaging
All SELD-103 units must be carefully packaged for
transport. Each unit should be boxed individually
with protective packaging (foam, bubble wrap, etc.)
in the box surrounding the unit.
After unpacking, protection will be the responsibility
of the user.
2.4 Recycling and end of product life
At the end of the service life of the product or its
parts, the relevant materials and parts should be
recycled or disposed of using an environmentally
acceptable method and in accordance with local
regulations. If the product contains substances
that are harmful to the environment, these should
be removed and disposed of in accordance with
current local regulations.
Page 5 of 20flowserve.com
Page 6
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Sensor Input
Connection Ports
Diagnostics
Connection
Port
2.5 Disposal Instructions
At the end of the product’s life, do not
dispose of any electronic component or instrument
in the domestic waste. Disposal should be done in
accordance with applicable regulations, which vary
from state to state and country to country.
Battery Disposal Instructions: Disposal should
be done in accordance with applicable regulations,
which vary from country to country. Batteries
should not be incinerated, unless suitable
procedures are followed and qualified handlers
have taken appropriate precautions. Exposure of
these cells to high temperatures or fire can cause
the cells to vent and/or rupture. These cells do not
contain dangerous substances. The reaction
products are inorganic and do not represent
environmental hazards, once the decomposition or
neutralization process has terminated.
Disposal in Europe
Batteries for disposal should not be transported by
air. For road transport of dangerous goods ADR
special provision 636 and packing instruction 903a
apply.
Disposal in US
Hazardous waste of spent batteries can be disposed
after they are first neutralized through an approved
secondary treatment prior to disposal. Disposal of
spent batteries should be performed by authorized,
professional disposal company, which has the
knowledge in the requirements of the Federal, the
State and the Local authorities regarding hazardous
materials, transportation and waste disposal. In
any case it is recommended to contact the local
EPA office.
The SELD-103-X (where “X” designates a specific
hardware or software configuration custom per the
order) model is a wireless transmitting unit
designed to monitor multiple sensor inputs and
wirelessly transmit the data to a receiver. It has the
option of enabling wireless transmission (to
become part of an IPS Wireless system) or
disabling wireless functionality (to perform as a
local alert or data-logging unit).
Figure 1: Sample SELD-103 unit
The SELD-103 is a self-powered unit that has 5
communication ports to measure up to 7 different
sensor inputs (utilizing combination sensors) per
unit (configuration limitations are detailed in section
3.1 Options and model numbers). Each unit also
has a single diagnostics port (on the back of the
base) for programming and troubleshooting the unit
as well as logging data with the VB-103-TAM
accessory (where collected data is stored).
Programming the unit requires a programming
cable and special software utility (PSP).
Figure 2: SELD-103 unit connection ports
The SELD-103 model has a built-in Omnidirectional wireless antenna, allowing for reliable
signal transmission. Data transmittal can occur at
a customizable time interval, ranging from every 5
seconds to once a day (this time interval is usually
pre-configured at the factory). Transmission rates
can also be varied according to sensor signal
values (see section 5 CUSTOM
CONFIGURATION PROGRAMMING), thereby
prolonging battery life. The SELD-103 unit has a
Page 6 of 20flowserve.com
Page 7
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Part Number
Description
M1-0045
Magnetic Mounting Base,
Two-pole, 70 lbs (31.75 kg)
High Temp.
M1-0081
Unistrut Mounting Bracket
for SELD-103 unit, SS
casting base
M1-0085
SS cap for SELD-103
connection ports (single)
301-4024
Programming cable for
SELD-103 Unit: 5’ length,
Serial Port- RS-485
connection
PSP
SELD-103 Unit
Configuration Software
Utility
VB-103-TAM
Take-Away Memory Module
VB-101-DOCK
IPS Dock for communication
with VB-103-TAM TakeAway Memory Module
TAM103
VB-103-TAM data download
and configuration software
VB-103-VIS
Visual Light Alert Module
Parameter
Name
Parameter
Description
Default
Value
Normal
Measurement
Interval
How often data is
read from each of
the sensors by the
5 minutes
maximum data transmittal range of ¾ mile (1.2km)
under ideal conditions, but its effective
transmission range can be extended using a
network of repeaters.
Each SELD-103 can be part of a larger wireless
monitoring network. As such, each unit comes with
a calibration sheet from the factory which contains
unit-specific configuration information describing
the sensors and wireless network the unit is
configured for (parameters such as the wireless
system number, unit transmitter number, data
transmittal/check-in interval, scale factor and offset
to use with specific sensors). As long as your
specific application has been communicated to the
factory, the SELD-103 unit should come preconfigured for your specific use.
3.1 Options and model numbers
The SELD-103 can either have wireless data
transmission enabled or disabled. In both settings,
it is compatible with all SELD-103 accessories.
SELD-103 units with wireless transmission
disabled (designated as SELD-103-VB) are
intended for use as local data collectors with either
the take away memory (VB-103-TAM) or visual
light alert modules (VB-103-VIS).
The SELD-103 is available with the following
sensors as options to connect to each port on the
unit:
1. Vibration & Temperature (RTD) Combination
Sensor
2. Vibration Sensor
3. Temperature Sensor (RTD)
4. Temperature Sensor (Thermocouple)
5. 4-20mA Analog Input Sensor
6. General Pulse Input Sensor
7. Tachometer (TTL)
8. Pressure & Temperature (RTD) Combination
Sensor
9. Discrete Sensor
10. Oil/Water Sensor
11. Vibration Sensor with FFT analysis output
12. Serial Port RS-485 Interface
13. Gas Sensor
Additionally the SELD-103 has the following
options available:
Table 1: Optional parts for SELD-103 unit
Figure 3: Magnetic mount option for SELD-103
unit
The exact configuration of different sensors
attached to each SELD-103 unit is limited to certain
combinations. Contact the factory for a complete
list of the different configurations available for the
SELD-103.
3.2 Default Configuration
The SELD-103 is configured to the following
parameters by default, unless otherwise specified.
Table 2: Parameter default configuration values
for SELD-103 unit
Page 7 of 20flowserve.com
Page 8
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
SELD-103 unit
Measurement Interval
Estimated Battery Life
5 seconds
4 months
10 seconds
8 months
30 seconds
2 years
1 minute
4 years
2 minutes
8 years
5 minutes
8 years
10 minutes
8 years
Measurement Interval
Data
Points
5 s
10 s
30 s
1
min 2 min 5 min
10
min
1
50
day
100
day
10
mo
1.6
yrs
3.3
yrs
4 yrs
4 yrs 2 25
day
50
day 5 mo
10
mo
1.6
yrs
4 yrs
4 yrs 3 16
day
33
day
3.3
mo
6.6
mo
1.1
yrs
2.7
yrs
4 yrs 4 12
day
25
day
2.5
mo 5 mo
10
mo
2 yrs
4 yrs 5 10
day
20
day 2 mo 4 mo 8 mo
1.6
yrs
3.3
yrs
6
8
day
16
day
50
day
100
day
6.6
mo
16.6
mo
2.7
yrs
7
7
day
14
day
42
day
2.8
mo
5.7
mo
14.2
mo
28.5
mo
LED
State
Description
GREEN
Flash every
5 seconds
Normal operation – all sensors are within
acceptable limits
3.3 Performance and operation limits
This product has been selected to meet the
specification of your purchase order.
The following data is included as additional
information to help with your installation. This is
typical information and if required, a definitive
statement for your application can be obtained from
Flowserve.
SELD-103 material compatibility is
the responsibility of the end user.
3.3.1 Sensor specifications
Refer to section 1.8 Specifications for available
sensor specifications.
3.3.2 Battery
Power is provided by a user-
replaceable lithium battery pack S1-1043, located
in the inner portion of the transmitter housing.
Battery life is dependent upon data transmittal
intervals. Battery voltage should normally be
between 2.8 and 3.6 VDC. If it is lower than this,
the battery will need to be replaced. The battery
voltage can be read at the receiver or through the
PSP software when connected directly to the
SELD-103 unit (refer to PSP software user
instructions for details on how to do this). See
table below for approximate battery life times for
the SELD-103.
Table 3: SELD-103 approximate battery life at
various measurement intervals under ideal
conditions
3.4 VB-103-TAM Optional Accessory
The VB-103-TAM is an optional Take Away
Memory module for the SELD-103 unit. When
connected to an SELD-103, it automatically logs
data from all sensors at the preconfigured log
interval. Data from the VB-103-TAM and the log
interval can be reviewed utilizing an IPS Dock
connected to a PC. The VB-103-TAM is set to a
default log interval (how often current data from the
SELD-103 is logged to the VB-103-TAM device) of
5 minutes (300 seconds), unless otherwise
specified. This is independent of the normal
measurement interval of the SELD-103 which
governs how often sensors are read.
The VB-103-TAM has a separate battery
that governs the lifetime of this unit. The expected
life of this battery is 6 years at the default 5 minute
log interval. The battery in the VB-103-TAM is not
replaceable and requires replacing the VB-103TAM unit when it is depleted.
The VB-103-TAM also has a finite amount of data
storage available. Dependent on the amount of
data points being measured and the measurement
interval, this storage capacity is defined as listed in
the table below.
Table 4: VB-103-TAM data storage capacity at
various measurement intervals and numbers of
data parameters
3.5 Visual Alert Optional Accessory
The Visual Alert is an optional accessory for the
SELD-103 unit. When connected to an SELD-103,
it automatically provides a visual light indication if
any sensors on the SELD-103 unit is or have been
in alarm status. To define the alarm limits for each
sensor, refer to Section 5 CUSTOM
CONFIGURATION PROGRAMMING.
The table below describes each of the different
types of indications offered by the Visual Alert
accessory.
Table 5: Visual Alert description
Page 8 of 20flowserve.com
Page 9
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
RED
Flash every
5 seconds
Current alarm – one of the sensor
measurements has exceeded its alarm
limit and has not returned within the
acceptable range
RED &
GREEN
Flash every
5 seconds
Alarm – one of the sensor measurements
previously exceeded its alarm limit, but has
now returned within the acceptable range
Magnetic set
screw for powering
unit on/off
WARNING STATIC HAZARD: Wipe only with
a damp cloth due to electrostatic discharge hazard.
4 MAINTENANCE
The only field maintenance required on the SELD103 unit is replacement of the battery. The SELD103 can also be turned off when not in use for
extended periods of time to conserve battery life.
4.1 Tools required
Allen Wrenches (1/8” and 3/16”)
Tools required to install the various mounting
configurations are conditional based on the
type of mounting and location
Phillips head torque screw driver
4.2 Turning unit on/off
To power off the SELD-103 unit remove the
magnetic set screw at the bottom of the base. This
will stop sensor readings and data transmission.
To power up the SELD-103 unit again, simply
install the magnetic set screw at the bottom of the
base. This will activate sensor readings and data
transmission at the pre-configured intervals.
The substitution of components may
impair intrinsic safety.
Power is provided by a lithium battery pack (S1-
1043), located in the inner portion of the transmitter
housing. Battery life is dependent upon data
transmittal intervals. Battery voltage should
normally be between 2.8 and 3.6 VDC. If it is lower
than this, the battery will need to be replaced. The
battery voltage can be read at the receiver or
through the PSP software when connected directly
to the SELD-103 unit (refer to PSP software user
instructions for details on how to do this).
The battery pack is NOT re-chargeable and
shall only be replaced by contacting your local
Flowserve representative (battery pack part
number S1-1043).
Perform the following steps to replace the battery:
1. Power down unit by removing the magnetic set
screw located at the bottom of the base.
2. Remove sensor cable(s) and any installed
accessories.
3. Take SELD-103 to non-classified area before
performing next step.
4. Loosen the set screw in the side of the tube
and gently unscrew the external housing (tube)
from the transmitter assembly.
Figure 4: Magnetic Set Screw at Bottom of
SELD-103
4.3 Battery Replacement
EXPLOSION HAZARD: The unit must be
powered OFF during the battery installation & must
be in a General Purpose Classified Area to prevent
the possible ignition of a hazardous atmosphere.
Page 9 of 20flowserve.com
Figure 5: Set screw on side of tube keeping
external housing on SELD-103 unit
Page 10
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Nuts holding
battery bracket
Battery bracket
Battery wires
5. The battery pack should be visible on the lower
end of the transmitter assembly. Carefully
disconnect the battery pack connector.
Remove the nuts holding the battery bracket in
place and remove the bracket. Remove the
battery pack.
Figure 6: SELD-103 unit battery with external
housing removed
5 CUSTOM CONFIGURATION
PROGRAMMING
This section is optional. As long as your
specific application was communicated to the
factory prior to shipment, the SELD-103 unit is preconfigured for your specific use.
Errors in programming
configuration can render the SELD-103 unit
inoperable, proceed with caution. Flowserve is
not liable for any damage caused by user
configuration errors.
For instructions on how to change software
configuration options on the SELD-103 unit, refer to
the PSP software user instructions. The list below
provides an overview of the functions that can be
performed using this software utility:
- Reading & saving SELD-103 configuration
- View & modify SELD-103 transmission
intervals
- Setting the time on the SELD-103
- Calibration of individual sensors on SELD103 unit
- Read live data from sensors on the SELD103 unit
- Turn specific SELD-103 sensor channels
on or off
Figure 7: SELD-103 unit with battery bracket
removed and battery pack partially removed
6. Install new battery pack (part number S1-
1043). Reinstall battery bracket and nuts,
reconnect battery pack connector.
7. Install tube back onto unit and tighten set
screw to no more than 0.9 Nm (8 lb.-in US).
8. Reinstall SELD-103 unit in the field.
9. Reconnect sensor cable(s).
10. Install magnetic set screw into the base of the
unit.
11. Install any applicable accessories on the
SELD-103 unit.
12. Dispose of battery pack in accordance with
instructions in section 2.5 Disposal Instructions.
For instructions on how to perform software options
on the VB-103-TAM unit, refer to the IPS Dock user
instructions. The list below provides an overview of
the functions that can be performed using the IPS
Dock and appropriate software utility:
- Reading & saving VB-103-TAM log data
- Clearing the log from the VB-103-TAM
- Setting the time on the VB-103-TAM
- Changing the data logging interval on the
VB-103-TAM
6 INSTALLATION
Always wear appropriate personal
protective equipment during the installation of
SELD-103 units.
6.1 Tools required
Allen Wrenches (1/8” and 3/16”)
Tools required to install the various
mounting configurations are conditional
based on the type of mounting and location
Phillips head torque screw driver
Page 10 of 20flowserve.com
Page 11
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Magnetic mount set
screw installation
location
6.2 Preliminary operational check
After unpacking and before field installation,
perform the following operational check on the unit:
1. Ensure the magnetic set screw (at the bottom
of the base) is removed. This keeps the unit
powered down during the installation process.
2. Place SELD-103 unit in vertical position with a
clear line of sight to and close to (<10ft) the
receiver. If only using the VB-103-TAM or VB103-VIS accessories, install the appropriate
accessory on the SELD-103 unit (VB-103-TAM
on port 6 and VB-103-VIS on port 5).
3. Insert sensor cable(s) into the proper port(s).
4. Install the magnetic set screw (in the bottom of
the base) to power up the unit.
5. Ensure the receiver is receiving data from the
unit. Let the VB-103-TAM log data for a few
minutes and confirm appropriate light operation
of the VB-103-VIS unit as applicable.
6. If applicable, disconnect VB-103-TAM and use
IPS Dock to confirm accurate data has been
logged (refer to IPS Dock user instructions for
how to do this).
7. Remove magnetic set screw (bottom of base)
to power down unit.
8. Disconnect sensor cable(s) from SELD-103
port(s).
Lastly, attach the other end of the mounting
adapter to the pole (using a U-bolt) or unistrut
(using bolting).
Figure 8: Pole or unistrut mounting options
6.3.2 Magnetic Mount Option
Insert set screw into bottom of SELD-103 unit and
tighten.
In case of issues with powering up the unit, refer to
section 9 Trouble-Shooting Guide.
6.3 Mounting
Each unit should be clear of physical obstructions
to the receiver and/or repeater(s) in order to
maintain consistent communications.
The SELD-103 is designed to be mounted in a
vertical position and should only be mounted in a
horizontal position with approval from the factory.
Ensure the SELD-103 is not subjected to excessive
vibration unless a routine maintenance program
includes verification of associated connections.
The SELD-103 unit should be installed near but not
on process equipment being monitored.
Mount the SELD-103 unit before connecting any of
the sensor cables.
6.3.1 Pole or Unistrut-Mounted Option
With this option the SELD-103 unit can be mounted
to either a vertical or horizontal pole or unistrut.
Start by bolting the SELD-103 unit to the mounting
bracket and then bolt the bracket to the appropriate
holes on the mounting adapter (see figure below).
Figure 9: Magnetic mounting option set screw
location
Next, take magnetic mounting base and thread
onto the exposed set screw from the bottom of the
SELD-103 unit. The magnetic mounting base (with
the SELD-103 unit attached) can now be attached
to any metal structure in a vertical position.
Figure 10: Sample SELD-103 unit with
magnetic mount option installed on unistrut
Page 11 of 20flowserve.com
Page 12
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Part
Number
Description
Recommend
Quantity for
2-Year Operational
Spares
S1-1043
Replacement Lithium
Battery Pack
1 (dependent on
transmit interval
selected)
M1-0083
Replacement Outer
Tube. 4” UV Protected
PVC Cap, Red, dwg#
100-2012
0
M1-0085
SS cap for SELD-103
connection ports (single)
6
G0-0047
Viton cap sealing gasket
1
G0-PACK
Desiccant pack
2
6.4 Connecting Sensors and Power-Up
Install all sensors on equipment to be monitored.
Ensure sensor cable connections and SELD-103
sensor ports are free from dust and debris (use
compressed moisture-free air to blow out any
contamination if needed). Then insert the sensor
cable(s) into the proper port(s) on the SELD-103
unit (reference the label on the connector side of
each sensor cable to determine which port it
attaches to). SELD-103 port numbers are labeled
on the stainless steel base and the unit serial
number is printed below the bar code on top of the
red cap. Lastly, install the magnetic set screw (in
the bottom of the base) to power up the SELD-103
unit.
Figure 11: Sample SELD-103 unit with one
sensor connected and port caps (optional item)
on remaining ports
6.5 Connecting 103 Take Away Memory
Install the VB-103-TAM into port 6 (single
amphenol-style connection on the back) of the
SELD-103 unit. The VB-103-TAM will
automatically start logging data from the SELD-103
unit at its preconfigured log interval.
To change the logging interval or download data
from the VB-103-TAM, refer to the IPS Dock User
Instructions.
6.6 Connecting 103 Visual Alert
Disconnect sensor (if any) installed on port 5 of the
SELD-103 unit. Install the VB-103-VIS into port 5
of the SELD-103 unit. Reinstall sensor (if any) on
port 5 of the SELD-103 unit. If no sensor is to be
used on port 5, be sure to install a port cover on
the sensor connection of the VB-103-VIS.
The VB-103-VIS will automatically review all
sensors on the SELD-103 unit for alarm conditions
and respond accordingly.
7 PARTS LIST AND DRAWINGS
7.1 Spare Parts
Available spare parts for the SELD-103 unit are
shown below. No spare parts are needed during
start-up. Recommended quantities of 2-year
operational spares are listed below.
Table 6: SELD-103 Spare Parts
7.2 Drawing
Page 12 of 20flowserve.com
Page 13
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Figure 12: SELD-103 drawing
Page 13 of 20flowserve.com
Page 14
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
0518
II 1 GD
Ex ia
CSA 2010 2273150
IECEx SIR.10.0006
Cl I, Zn 0 A/Ex ia IIC T6
I.S. for Cl I, Div 1 Grps A, B,
C, D;
Cl II, Div 1, Grps E, F, G;
Cl III, Div 1, and provides I.S.
circuits with Entity
Parameters per Install
Manual 701-2000 and 7013000
Tem. Code T6
-20°C ≤ Ta ≤ +60°C
Sira 10ATEX2013
Ex ia IIC T6 Ga
Ex iaD IIIC T85°C DA
8 CERTIFICATION
The following certifications are applicable to the
SELD-103:
Table 7: Model SELD-103 certifications
9 TROUBLE-SHOOTING GUIDE
EXPLOSION HAZARD: The unit must be
powered OFF during the battery installation & must
be in a General Purpose Classified Area before
removing outer plastic tube to prevent the possible
ignition of a hazardous atmosphere.
9.1 Loss of Communications
Loss of communications with SELD-103 unit may
occur for several reasons:
Unit not powered up
Unit not configured correctly
Unit out of range of receiver
Receiver not configured correctly
4. Remove sensor cable(s) and configuration
cable.
5. Ensure SELD-103 is in a non-classified area
before performing next step.
6. Loosen the set screw in the side of the tube
and gently unscrew the external housing (tube)
from the transmitter assembly.
7. The battery pack should be visible on the lower
end of the transmitter assembly. Carefully
disconnect the battery pack.
8. Use a voltmeter to read battery voltage across
pins with red and black wires on connector to
battery pack. If the voltage level is below 2.8V,
replace the battery according to instructions in
section 4 MAINTENANCE.
Figure 13: Measuring battery level on SELD-103
unit (external housing removed) using a
voltmeter
See solutions below to each of the possible
causes:
9.1.1 Unit Not Powered Up
Verify set screw is installed at base of SELD-103
unit (see section 4 MAINTENANCE on how to turn
on unit).
1. Connect diagnostics RS-485 cable to port 6 of
SELD-103 and other end to a PC.
2. Read battery level using PSP software. If the
battery level is above 2.8V, skip to step 8. If
you are able to successfully read the battery
voltage on this screen but the level is below
2.8V, replace the battery according to
instructions in section 4 MAINTENANCE.
3. If communications are still not successful,
power down unit by removing the magnetic set
screw located at the bottom of the base.
Page 14 of 20flowserve.com
Figure 14: Detail view of voltmeter connection
to battery pack cables.
9.1.2 Unit Not configured Correctly
To perform the diagnostic steps for this issue, the
SELD-103 unit will need all sensors connected to it.
Refer to PSP software instructions to reload the
original configuration into the SELD-103 unit.
9.1.3 Unit Out of Range of Receiver
If communications from the SELD-103 unit to the
receiver are still not successful, physically move
the SELD-103 unit next to and with a clear line of
sight to the receiver.
Page 15
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
If data is being read from this location, then ensure
that the repeater network is appropriately located
and functional.
9.1.4 Receiver Not Configured Correctly
Check receiver configuration to ensure it is set up
to receive data from this particular SELD-103 unit
(refer to receiver user instructions).
If none of the above solutions are successful in
getting data reception at the receiver, contact the
factory for additional support.
9.2 Inaccurate or Missing Data
If the data being read from the SELD-103 unit is
either inaccurate or missing, this could have
several causes:
Ensure the sensor cables to and sensors ports on
the SELD-103 unit are free from dust and debris.
Use compressed moisture free air to blow out any
debris if needed.
Reconnect all sensor cables to the SELD-103 unit
to ensure tight connections.
9.2.2 Unit Configuration Incorrect
If the inaccurate/missing data problem is across all
sensors, reload the original configuration for the
SELD-103 unit (see PSP user instructions).
9.2.3 Inaccurate Sensor Calibration
If the data received on certain channels/sensors is
not accurate, recalibrate those sensors (see PSP
user instructions).
9.2.4 Incorrect Receiver Configuration
Refer to receiver instructions to confirm correct
configuration. If necessary, contact factory to
reload receiver configuration settings.
9.3 No Data Recorded on VB-103-TAM
If no data is being stored to the VB-103-TAM unit,
this could have several causes:
Ensure the sensor cables to and sensors ports on
the SELD-103 unit are free from dust and debris.
Use compressed moisture free air to blow out any
debris if needed.
Reconnect all sensor cables to the SELD-103 unit
to ensure tight connections.
9.3.2 Loose VB-103-TAM Connection
Ensure the VB-103-TAM and SELD-103 ports are
free from dust and debris. Use compressed
moisture free air to blow out any debris if needed.
Disconnect and reconnect the VB-103-TAM from
the SELD-103 unit. Ensure that VB-103-TAM is
completely and fully threaded onto the SELD-103.
9.3.3 Low Battery on VB-103-TAM
Check to see if VB-103-TAM unit can be read on
an IPS Dock. If not, replace VB-103-TAM unit (the
VB-103-TAM battery is not user replaceable).
Connect the VB-103-TAM to the IPS Dock and
ensure the log interval is set to the expected value
(see IPS Dock user instructions for additional
details on how to perform this task).
9.4 No Lights on VB-103-VIS
If no lights are flashing on the VB-103-VIS, this
could have several causes:
Loose VB-103-VIS connection
9.4.1 Loose VB-103-VIS Connection
Ensure the VB-103-VIS and SELD-103 ports are
free from dust and debris. Use compressed
moisture free air to blow out any debris if needed.
Disconnect and reconnect the VB-103-VIS from the
SELD-103 unit. Ensure that VB-103-VIS is
completely and fully threaded onto the 103.
If none of the above solutions are successful in
resolving the issue, contact the factory for
additional support.
Page 15 of 20flowserve.com
Page 16
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
EN 61326-1
2006
Electrical equipment for measurement, control and laboratory use EMC requirements - Part 1: General Requirements
EN 55011
2009
Industrial, scientific and medical equipment. Radio-frequency
disturbance characteristics. Limits and methods of measurement
Class A, Group 1
EN 61000-3-2,
EN 61000-3-3,
EN 61000-4-2,
EN 61000-4-3,
EN 61000-4-4,
EN 61000-4-5,
EN 61000-4-6,
EN 61000-4-8,
EN 61000-4-11
Series, Electromagnetic compatibility (EMC).
EN 61010-1
2002
Safety requirements for electrical equipment for measurement, control
and laboratory use. Part 1 General requirements
10 DECLARATION OF CONFORMITY (TYPICAL)
We, Flowserve Corporation
10920 West Sam Houston Parkway North
Suite 950
Houston, Texas 77064
USA
Declare in sole responsibility that the equipment:
ELD-102-X Single Point Transmitters
SELD-103-X Smart Point Transmitters
ELD-FA-403-Receiver, AO/DO and Serial Modbus
S-EGI-NB-4/20-AC/SP, Smart RTU
ELD-FA-525-AC/SP, Intelligent Repeater
Including all options and versions of the base model numbers to which this Declaration refers are in compliance
with the Directives and Norms specified herein.
1.1 EMC Directive 2004/108/EC
Basis for compliance:
The equipment has been assessed using the following standards and is supported by the following
technical documents:
Under these environmental and use conditions:
Industrial Applications, Pollution Degree 2
Technical File Reference Number: 701-5000
Installation and use of this equipment shall comply with the relevant installation and use instructions.
1.2 Low Voltage Directive 2006/95/EC
Basis for compliance:
The Single point transmitters are fully-enclosed, intrinsically-safe apparatus and are considered to meet
the requirements as set forth in the standard below without further consideration:
Under these environmental and use conditions: Industrial Applications, Pollution Degree 2
Page 16 of 20flowserve.com
Page 17
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
EN 60079-0
2006
(+ corr. 1)
Electrical apparatus for explosive gas atmospheres Part 0
General requirements
EN 60079-11
EN 60079-26
2007
2007
Explosive atmospheres Part 11 Equip protection by intrinsic
safety i
Explosive Atmospheres- Part 26: Equipment with equipment
protection level (ELP) Ga
EN 61241-0
2006
Elec App for use in presence of combustible dust Part 0 Gen
requirements
EN 61241-1
2004
(+ Corr. 1
& 2)
Elec app for use in presence of combustible dust Part1
Protection by enclosures tD
EN 61241-11
2005
Elec app for use in presence of combustible dust Part11
Protection by intrinsic safety iD
Declaration of Conformity
1.3 ATEX Directive 94/9/EC + all amendments
Basis for compliance:
The equipment has been assessed using the following standards and is supported by the following
technical documents:
Notified Body: CSA International
Report Number: 1005105
Markings:
Page 17 of 20flowserve.com
Page 18
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
The technical documentation required to demonstrate that the product meets the requirements of the Directives
has been compiled by the signatory below and is available for inspection by the relevant enforcement authorities.
Signed: _____________________________________________ Date: _________________________
Authorized Person, Rick Lawson
General Manager
Page 18 of 20flowserve.com
Page 19
Notes:
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Page 19 of 20
flowserve.com
Page 20
IPS WIRELESS SELD-103 ENGLISH 26999976 11-13
Your Flowserve factory contacts:
Flowserve Corporation
10920 W Sam Houston Parkway N, Suite 950
Houston, TX 77064 USA
Phone: +1 832 375 0807
Your Flowserve sales contact:
Go To: www.flowserve.com
Equipment Monitoring and Control Products
FLOWSERVE REGIONAL
SALES OFFICES:
USA and Canada
Flowserve Corporation
Pump
5215 North O’Connor Blvd.,
Suite 2300
Irving, Texas 75039-5421 USA
Telephone 1 972 443 6500
Fax 1 972 443 6800