
TABLE
INTRODUCTION
PRECAUTIONS
CONTROLS
TEST
PROCEDURE
CONNECTION
OPERATOR
VISUAL
INDICATORS
AUDIO
INDICATORS
ADDITIONAL
REPLACEMENT
ACCESSORIES
SPECIFICATIONS
Fa
OF
AND
FUNCTIONS
TO A PATIENT
CONTROL
FEATURES
OF
BATTERIES
=
CONTENTS
BUTTONS
PAGE
PAGES
PAGE
PAGE
PAGE
PAGE
PAGE
PAGE
PAGE16
PAGE
Ев
PAGE
0000"
4
6
8
9
11
13
15
17
19
P/N
185040541

f
4
INTRODUCTION
PRECAUTIONS
Nerve stimulation
passing
through
not
by
characteristics
many
other nerve
NS242
deliversa
Skin
resistance,
gel
conductivity
electrodes,
way
to
stimulation.
constant
changing
pulse
must
supramaximal
The
INNERVATOR
truly
constant
at
up
to
the
voltage.
achieve
350
Therefore
are
essential
stimulators,
constantcurrentstimulation
electrode
vary
constant
consistent
The voltage
current
must
impedance,
be
of
sufficient
stimulation.
current
volts.
is
effected
tissue
surface
widely.
current
necessary
vary
and
the
NS242
pulses
by
the
between
and
When
is
supramaximal
intensity
is
up
CURRENT
electrodes,
the
pulse
primary,
the
INNERVATOR
resistance,
using
the
most
to
deliver
automatically
stimulating
to
designed
to
80
milliamperes
to
current
Unlike
pulse.
and
surface
effective
nerve
the
with
current
achieve
deliver
WARNING:
the
Risque
d'anesthesiques
The laws
this
This
allows
This
location.
Dangerously
device
batteries.
Triggering
bycertain
can
noise.
sensitive
life
operating
Possible
presence
device
and
7
οἱ
d’explosion.
and
statutes
to
use
device
should
transthoracic
device
should
Do
not
high
for
up
to
ofthis
electrasurgical
generate
The
NS242
electronic
is
totally
room
explosion
flammable
inflammables.
by,
not
use
voltages
10
nerve
high-energy
environments,
anaesthetics.
Ne
pas
utiliser
of
many
or
onthe
order
be
used
stimulation.
not
be
used
with
needle
may
minutes
after
stimulatar
units
has
been
circuitry
compatible
hazard
if
used
in
en
presence
countries
ina
for
(cauterles),Avhich
random
to
increase
restrict
ofa
physician,
manner
which
internal
nerve
electrodes.
exist
within
removal
of
may
be
caused
frequency
designed
battery
with
“normal”
this
the
with

CONTROLS
KEY
(1)
PROXIMAL — Positive
(2)
FAULT
INDICATOR — Set
(3)
REPEAT
iNDIGATOR — Repetitive
DEMAND
BUTTONS — Deliver a pulse
(4)
SINGLETWITCH = *
(5)
TRAIN
OF
{6}
REPEAT
(7) 1 SECOND
(8)
(8)
(10)
(11)
(12)
(13)
(14)
(15)
{16)
(17)
FOUR
TETANUS
BUTTONS — Activate a repetitive
SINGE
10
SECOND
12
SECOND
POWER
MUTE
CURRENT UP/DOWN
PULSE
LOW
BATTERY
MUTE
VISUAL
DISTAL — Negativa
TWITCH
SINGLE
TRAIN
OF
ON/OFF
BUTTON
BUTTON — Mutes
DELIVERY
INDICATOR
INDICATOR
SYMBOL — Audio
DISPLAY
OF
AND
FUNCTIONS
(Red)
current
TWITCH
FOUR
pulse
delivery
BUTTONS
*
pulsa delivery
SET
OUTPUT
(Black)
not
delivered
pulse
pattern
pattern
pulse
sound
sound muted
CURRENT
on
demand
pattern
operating
©
REPEAT
(3)
INN
FAULT
7
|<
FIGURE
1
om
RV/1TOR
NS
242
(19)
(14) 415)
(2
CURRENT

©
TEST
The
following
each
time
the
instrument
1.
Attach
inserting
banana
DISTAL
Connect
m
initiate a short
Switch
>
OFF
button.
Activate a single
>
by
selecting
(1
SEG
indicators
is
delivered
indicator
Emission
followed
warns
shorted
broken.
Whilst
9
single
to
initiate
instigates
attempt
the
normal
operator
beep
tone.
resistance
circuitry.
6.
Once
ensure
PROCEDURE
test
procedure
INNERVATOR
is
connected
the
lead
the
plugs
into
(black)
the
free
the
INNERVATOR
Set a delivery
the
TW).
should
(every 1 second).
should
of a half
by a continuous 2 secand
that
the
together
still
running a 1
twitch,
disconnect
an
alarm
to
deliver a pulse
audio
should
followed
This
procedure
alarm
the
test
procedure
that
the
must
is
used, and
to a patient.
to
the
colour
output
circuit
appropriate
Both
not
open
hear a half
by a continuous 2 second
INNERVATOR
INNERVATOR
coded
the
PROXIMAL
jacks,
ends
of
the
condition,
on
current
twitch
at 1 second
the
audible and
activate
each
be
shown
second
ends
of
the
properly
or
second
the
circuit
conditions
instead
pulse
delivery
second
tests
and
the
device’s
has
monotone
be
carried
before
right-angled
(red)
lead
together
using
the
of
10mA.
intervals
REPEAT
button
visual
time a puise
The
low
battery
on
the
display.
beep
heep,
lead
are
that
the
lead
repetitive
ends
of
the
condition.
and
at
the
of
hearing
sound,the
monotone
the
high
internal
been
completed
is
switched
out
the
by
and
to
ON/
not
is
lead
This
next
skin
off.
CONNECTION
The
following
connecting
1、
n
>
>
output
s
procedure
the
INNERVATOR
Ensure
that
OFF.
Power
of
the
device
unnecessary
Place
two
{disposable
suitable}
on
course
of
cases
figure
Connect
inserting
the
output
coded
that
Connect
patient
output
connected
attached
distal
electrode.
way
achieved
if
Following
switeh
OFF
until
indicates
stimulation
the
this
2).
the
the
PROXIMAL
jacks.
so
that
correct
the
electrodes.
jackis
jack
with
to
connector
If
then
by a somewhat
the
leads
the
button,
observation
that
TO A PATIENT
should
to a patient:
the
INNERVATOR
should
be
off
the
reusable
of
the
to
be
the
to
the
angled
(red)
and
ends
of
matching
end
The
and
proximal
attached
leads
are
then
increase
of
the
desired
attained.
during
patient
ECG
forearm,
stimulated.
the
are
of
PROXIMAL
the
The
electrode,
lower
on
to
discomfort.
standard
or
the
skin
nerve
is
usually
lead
right
The
colour
connections
other
positive,
is
negative.
the
the
proximal
(black)
the
supramaximal
are
reversed.
the
induction
INNERVATOR
and
the
has
been
be
utilised
is
switched
connection
to
prevent
electrodes
electrodes
along
ulnar
INNERVATOR
banana
connected
twitch
or
In
nerve.
jacks
DISTAL
(black)
lead
are
will
ensure
made.
the
lead
DISTAL
(black)
lead
should
connector
and
to
the
stimulation
current
of
anaesthesia
using
the
the
current
response
supramaximal
©
when
are
the
most
(See
by
into
colour
to
the
(red)
be
(red)
the
distal
in
this
is
than
ON/

İİİ
10
IMPORTANT:
tone
indicator this
exist,
delivered
This
dried
leads.
6.
After
INNERVATOR
Distal
Electrode
(Black)
Proximal
heard
and
to
be
or
faulty
use,
the
If a two
after the
indicates
current
the
patient,
caused
by
electrode(s)
always
is
switched
FIGURE
is
may
/
Z
Electrode
(Red)
second
audible
pulse delivery
that
atarm
selected
excessive
or
disconnected
ensure
OFF.
2
continuous
conditions
is
not
being
impedance,
that
the
OPERATOR
Buttons
are
to
A.
ON/OFF
CURRENT
CURRENT
MUTE
grouped
their
function:
CONFIGURATION
On/Off
Switches
operated
off
These
current
decrease
(mA)
{fthe
the
Similarly
DOWN
the
at
instantaneously.
UP
DOWN
buttons
to
the
steps
CURRENT
display
button
decreases
The
INNERVATOR
current
of
Mute
button.
+
This
buttonactivates
can
be
used
sound.
When
indicated
display.
The
mute
pushes
the
mute
off.
audible
used
to
The
mute
the
CONTROL
into
three
categories
BUTTONS
button.
INNERVATOR
any
time
to
turn
Current
Up
button.
Current
Down
are
used
be
currentsetting
each
quickly
by
holding
to
OMA.
more
to
mute
the
by
the
remains
MUTE
mute
pulse
to
delivered.
time
the
UP
button
advances
dewn
the
set
current
can
not
than
80mA.
the
mute
the
audible
mute
is
mute
symbol
in
effect
button
can
only
delivery
the
audio
BUTTONS
according
on,
and
the
INNERVATOR
button.
set
the
amount
They
increase
by 10
milliampere
button
is
is
held
down,
to
the
CURRENT
display
be
set
to
function,
pulse
in
operation
(2% ) on
until
again
to
switch
be
used
sound,
it
alarm
cannot
sound.
11
<
may
be
CURRENT
of
and
pressed.
then
80mA.
quickly
deliver
a
which
delivery
this
is
the
the
user
the
to
mute
he
©

12
B.
DEMAND
Buttons
which
demand.
pattern
TW
TOF
TET
G.
These
pulse
until
cycle,
patterns
indicator
1
10
12
When
will
Twitch
Delivers a single
Train-of-Four
Delivers a train
pulse
Tetanus
Delivers a tetanus
the
button
REPEAT
are
start/stop
pattern.
the
button
When
below
(4)
SEC
TW 1 Second
Pressing
be
delivered
pulse
button
SEC
TW
Pressing
be
delivered
pulse
button
SEC
TOF
Pressing
pulses
pulse
button
BUTTONS
will
deliver a single
the
be
delivered
button.
is
0.5
seconds
button.
is
BUTTONS
The
pulse
is
pressed
any
one
is
in
is
shown
this
at
pattern
is
pressed
10
Second
this
button
pattern
is
pressed
12
Second
this
to
be
repeated
pattern
is
pressed
button
Is
once
pulse.
button,
of
four
apart.
pulse
depressed.
buttons
which
pattern
again
of
the
operation
on
the
Twitch
button
button
causes a single
one
second
will
play
repeatedly
again
to
Twitch
causes a single
at
10
second
will
play
repeatedly
again
to
Train-of-Four
button
causes
every
will
play
repeatedly
again
to
pulse
pressed
only.
pulses
where
train
for
activate
will
play
to
terminate
three
repeat
the
visual
display.
(142).
(E)
intervals.
halt
the
button
intervals.
halt
the
a
train
12
seconds.
halt
the
pattern
on
the
pulse
each
©
as
long
as
a
preset
repeatedly
the
pulse
REPEAT
pulse
to
This
until
the
function.
(0.1Hz).
pulse
to
This
until
the
function.
button.
of
four
This
until
the
function.
VISUAL
The
INNERVATOR
indicators:
(see
INDICATORS
NS242
has
also
the
Figure 3 overleaf)
following
CURRENT
The
current
to
be
shown
in
mA
symbol.
PULSE
DELIVERY
The
pulse
solid
dot
every
time a pulse
MUTE
SYMBOL «x
This
is a speaker
when
visible
signal
is
REPEAT
SYMBOL
This
is
shown
indicates
patterns 1 SEC
TOF
is
running.
FAULT
SYMBOL
This
is
shown
indicates
„
deliver
the
conditions
activated
impedance,
if
the
patient
LOW
BATTERY
This
is
shown
indicates
and
must
INNERVATOR
normally
not
replaced
become
completely
delivered
milliamps
«delivery
in
the
centre
with a cross
it
indicates
muted.
as a small
that
one-of
TW,
as a small
that
the
INNERVATOR
specified
exist.
The
by
conditions
if
the
leads
electrodes
as a picture
that
the
batteries
be
replaced
will
for a short
eventually
to
and
is
indicated
indicator
of a circle.
is
delivered.
that
triangular
the
three
10
SEC
triangular
current
fault
indicator
are
disconnected
are
dry.
of a battery
are
continue
time.
If
the
the
blank
except
the
patient
is
shown
It
flashes
through
audible
arrow
repeat
TW,
or
12
arrow
is
not
and
may
of
excessive
becoming
shortly.
to
operate
batteries
display
for the
13
visual
mA
is
by
the
©
as
a
it
and
pulse
<
and
pulse
SEC
<
and
able
to
alarm
be
or
and
flat
The
are
will
low

ON
ーー
14
Displayed
battery
symbol.
INNERVATOR
replacing
the
Figure 3 —
By
Once
this
can
only
batteries.
Visual
The
(NNERVATOR
has
occurred
be
operated
、
Indicators
NS242
the
by
AUDIO
The
INNERVATOR
indicators:
PULSE
DELIVERY
This
is
every
time a pulse
AUDIBLE
ALARM
Heard
time a pulse
the
completion
second
conditions
This
is
delivers a pulse
+/-5mA
current.
may
disconnected
The
alarm
only
be
to
deliver a pulse.
The
alarm
automatically
the
alarm
THERE
If
the
alarm
+
do
not
*
do
not
fault
causing
INDICATORS
NS242
heard
as a monotone
as a half
second
is
delivered
of
exist.
every
that
The
conditions
leads
sound
is
each
sound
when
is
repaired.
NO
ALARM
sound
continue
deliver
any further
condition
the
limits
excessive
continuous
heard
tolerance
include
heard
JS
has
the
following
is
delivered.
pulse
beep
time
does
or
dry
NOT
time
the
is
occurring
running a repeat
single
monotone
indicating
each
will
MUTE
is
beep
incorrectly,
train
there
thatalarm
the
INNERVATOR
not
fall
within
side
causing
continuous
an
of
this
impedance,
patient
electrodes.
attempt
be
cancelled
condition
FACILITY
repeatedly:
pulses
repaired.
15
audible
beep
each
AND
at
is a 2
the
the
set
alarm
and
will
is
made
causing
function.
until
the

16
ADDITIONAL
Auto
power
shut-off.
The
INNERVATOR
after
20
minutes
off
the
INNERVATOR
no
buttons
the
Pressing
will
minutes,
This
battery
accidently
are
INNERVATOR
any
cancel
button
the
feature
is
deterioration
switched
will
of
non-use.
will
pressed
within
will
automatically
other
automatic
designed
to
if
on,
or
FEATURES
automatically
Prior
emit
than
shut
down
safeguard
the
is
left
to
power
three
beeps,
the
next
one
switch
the
ON/OFF
for
against
INNERVATOR
turned
switch
another
on.
off
shut-
and
if
minute
off.
button
20
rapid
is
REPLACEMENT
The
INNERVATOR
1.5
volt
AA
Batteries
should
indicator
(É%)
To
replace
top
rear
towards
as
the
should
the
down
Batteries.
polarity
cover.
NS242
batteries.
be
replaced
is
first
batteries,
of
the
the
be
inserted
indicated
OF
BATTERIES
is
powered
when
observed
INNERVATOR
bottom
simply
on
the
on
depress
of
observing
inside
the
the
the
low
LCD
the
by
pushing
back
the
of
the
17
by
three
battery
display.
clip
on
panel.
correct
battery
it
Automatic
is
This
monitoring
minutes.
shutdown
running a repeat
feature
also
has
will
function.
serves
not
been
not
occur
if
to
remind
undertaken
the
the
for
at
operator
user
that
least
20

18
ACCESSORIES
The
following
when
accessories:
NS242
Complete
and
operating
200NS001
200NS002
Nylon,
belt
loop.
200NS003
Adhesive,
pregeiled
domed
stimulator
200NS004
Black
inches).
Connect
200NS007
12m
disposable
ends.
200NS008
1.2m
reusuable
ends.
200NS009
185040541
185040547
part
ordering
Constant
Stimulator
with
manual.
Diagnostic
Carry
shock
resistant,
Colour — grey.
Disposable
dual
element
soft
foam
electrodes).
lead
Reusable
silicone
rubber
Non-adhesive.
using
nerve
Nerve
(48)
with alligator
electrodes
Nerve
(48°)
with
electrodes
Mounting
NS242
NS242
numbers
peripheral
nerve
nerve
Current
Peripheral
stimulator
Batteries
Probe
Case
zippered
Electrodes o ©
perforated
pad.
One
per
pack.
200NS007.
Electrodes
electrode,
Pack
stimulator
Stimulator
clips.
200NS003.
Stimulator
2mm
pin
connectors.
200NS004.
Bracket
Operating
Technical
should
stimulators
lead
not
included.
carry
electrode
dual
element
Connect
İ 7 2
51x38mm
of
10
lead
200NS008.
Lead
For
Colour
Lead
For
Colour
Manual
Manual
be
quoted
and
Nerve
200NS007
case
with
with
(two
using
nerve
(2x1.5
electrodes.
use
with
coded
use with
coded
SPECIFICATIONS
Output
current
Tetanus
frequency
Output
pulse
Output
pulse
Output
pulse
Output
voltage
Set
current
offset
Alarm
tolsrance
Power
requirement
Battery
life
Dimensions
一
一
To
display
oscilloscope
instrument.
small
precision
series
with
connected
across
range
form
width
rise/fall
limits
or
record
directly
To
record
resistor
the
patient.
this
0-80milliamperes,
selectable
steps
50Hz
Square
current,
unidirectional
200
+/- 5 microseconds
time<5
350
<5
+/-
current
Three
batteries
More
continuous
alkaline
Height - 160mm
Width - 73mm
Depth-
Weight — 0.18kg
(without
the
to
the
the
must
The
resistor.
in
wave,
monophasic,
microseconds,
microseconds
volts,
maximum
milliamperes
SmA
of
1.5
volt
than
300
batteries
43mm
batteries)
voltage,
output
jacks
current
be
connected
osciltoscope
19
10mA
constant
the
set
“AA”
hours
use
with
connect
delivered,
of
must
an
the
in
be
a