Fairbanks Series II User Manual

Installation / Operator /Service Manual
Bench Scales
© 2005-2010 by Fairbanks Scales Inc.
All rights reserved
50777
Revision 6 -- 07/10
07/10 2 50777 - Rev. 6
Amendment Record
Manufactured by Fairbanks Scales Inc.
821 Locust
Kansas City, Missouri 64106
Issue 1 01/05 New Product
Issue 2 06/06 Add programmable division size
Revision 3 08/07 Updated Parts List and model specifications
Revision 4 10/08 Updated Parts List
Revision 5 04/10 Added recharger information
Revision 6 07/10 Added 150 lb capacity scale Measurement Canada exception
Disclaimer
Every effort has been made to provide complete and accurate information in this manual. However, although this manual may include a specifically identified warranty notice for the product, Fairbanks Scales makes no representations or warranties with respect to the contents of this manual, and reserves the right to make changes to this manual without notice when and as improvements are made.
07/10 3 50777 - Rev. 6
Table of Contents
Section 1: General Information
A. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
B. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 4
Section 2: Installation
A. General Service Policy . . . . . . . . . . . . . . . . . . 6
B. Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
C. Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
D. Instrument Location . . . . . . . . . . . . . . . . . . . . 7
E. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
F. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Section 3: Programming
A. Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . 8
B. Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Section 4: Calibration
A. Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . 9
B. Calibration Procedure . . . . . . . . . . . . . . . . . . . 10
Section 5: Operation
A. Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . 11
B. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Section 6: Service and Maintenance
A. Error Messages . . . . . . . . . . . . . . . . . . . . . . . 13
Section 7: Parts
A. Replacement Parts List. . . . . . . . . . . . . . . . . . 14
07/10 4 50777 - Rev. 6
A. Introduction
The Fairbanks Series II bench scales are high-accuracy, portable scales suited for both comercial and non-commercial applications. These scales are designed to be used in a variety of weighing applications including parts counting, inventory, material handling, and general manufacturing.
B. Specifications
1. Instrument Specifications
Model Series II
Product No. Capacities Platform size
25825 60 lb x .02 12.5” x 16.5” (320 x 420 mm) 25826 100 lb x .05 12.5” x 16.5” (320 x 420 mm) 25827 150 lb x .05 12.5” x 16.5” (320 x 420 mm) 25828 200 lb x .1 12.5” x 16.5” (320 x 420 mm) 25829 250 lb x .1 12.5” x 16.5” (320 x 420 mm) 25830 300 lb x .1 12.5” x 16.5” (320 x 420 mm)
25831 200 lb x .1 15.75” x 18.9” (400 x 480 mm) 25832 250 lb x .1 15.75” x 18.9” (400 x 480 mm) 25833 300 lb x .1 15.75” x 18.9” (400 x 480 mm) 25834 400 lb x .2 15.75” x 18.9” (400 x 480 mm) 25835 500 lb x .2 15.75” x 18.9” (400 x 480 mm) 25836 600 lb x .2 15.75” x 18.9” (400 x 480 mm)
Indicator Rating IP 54
Display LCD, height 1” (25 mm); 6 digit; backlit
Units Lbs, Kgs
Zero Range 2 or 100%
Auto Zero Tracking None, 0.6d, 1.0d, and 2.0d
Counting Sample sizes of 10, 20, 50, 100, or 200
Platform Construction 304 Stainless Steel
Base Construction Welded tubular and cast
Overload Safe = 150%
Maximum = 300%
Creep 0.02% (30min)
Bridge Resistance 350 ohms nominal
Section 1: General Information
07/10 5 50777 - Rev. 6
2. Environmental
Power Supply AC/ DC adapter 110-120VAC - 9Vdc @ 500 mA,
Built-in 6Vdc rechargeable lead-acid battery Up to 20hrs continuous use. Recharge time: Minimum of 12 hours for fully discharged battery
All Series II Bench Scales after serial number 505159000150 will have an internal rechargeable battery.
Operating Temperature -14 to 104 F (-10 to 40 C) 85%RH
3. Approvals
NTEP CC#: 05-009 MC #: AM-5618 *Note: 150 lb capacity models are not Measurement Canada approved.