Created 10/2008
Revision 1 12/2008 Documentation Release
Revision 2 05/2011 Corrected Table of Contents
Disclaimer
Every effort has been made to provide complete and accurate information in this
manual. However, although this manual may include a specifically identified warranty
notice for the product, Fairbanks Scales makes no representations or warranties with
respect to the contents of this manual, and reserves the right to make changes to this
manual without notice when and as improvements are made.
Fairbanks Scales shall not be liable for any loss, damage, cost of repairs, incidental
or consequential damages of any kind, whether or not based on express or implied
warranty, contract, negligence, or strict liability arising in connection with the design,
development, installation, or use of the scale.
This document contains proprietary information protected by copyright. All rights are
reserved; no part of this manual may be reproduced, copied, translated or transmitted
in any form or by any means without prior written permission of the manufacturer.
Table of Contents
1 Safety Information ........................................................................................................ 9
1.1 INTENDED USE ..................................................................................................................... 9
12.5 MECHANICAL DATA .......................................................................................................... 98
12.5.1 Construction 98
12.5.2 Dimensions 98
12.5.3 Weight 98
12.6 USE IN LEGAL-FOR-TRADE MODE .................................................................................. 98
12.6.1 Documentation for Verification on the Enclosed CD 98
12.6.2 Additional Instructions 98
13 Index ............................................................................................................................ 99
05/11 6 51209 Revision 2
Section 1: Safety Information
1 Safety Information
The instrument was in perfect condition with regard to safety features when it left
the factory. To maintain this condition and to ensure safe operation, the operator
must follow the instructions and observe the warnings in this manual.
1.1 Intended Use
The instrument is intended for use as an indicator for weighing functions. Product operation,
commissioning and maintenance must be performed by trained and qualified personnel who are
aware of and able to deal with the related hazards and take suitable measures for self-protection.
The instrument reflects the state of the art. The manufacturer does not accept any liability for
damage caused by other system components or due to incorrect use of the product.
1.2 Initial Inspection
Check the content of the consignment for completeness and inspect it visually for signs of damage
that may have occurred during transport. If there are grounds for rejection of the goods, a claim must
be filed with the carrier immediately and the sales or service organization must be notified.
1.3 Before Commissioning
Visual inspection:
Before commissioning and after and storage or transport, inspect the instrument
visually for signs of mechanical damage.
1.3.1 Installation
The instrument is designed for mounting on standard rails (35 mm, acc. to DIN 46277).
When mounting on the rail, please make sure that the distance from other
instruments left and right of the module is at least 20 mm.
1.3.2 Electrostatically Sensitive Components
This instrument contains electrostatically sensitive components. For this reason, an equipotential
bonding conductor must be connected when working on the open instrument (antistatic protection).
05/11 9 51209 Revision 2
Section 1: Safety Information
1.3.3 Protective Earth
Connection to protective earth must be performed via the mounting rail.
1.3.4 Supply Voltage Connection
The supply voltage is 24V DC +10% / -15%.
Max. power consumption of /00 version: 6.5 W; /01 and /04
version: 8.5 W
For connection to 230 / 115 V AC, an external power supply
(e.g. PR 1624/00 or Phoenix Mini Power) is required.
1.3.5 Failure and Excessive Stress
If there is any reason to assume that safe operation of the instrument is no longer ensured, shut it
down and make sure it cannot be used. Safe operation is no longer ensured if any of the following
is true:
- The instrument is physically damaged
- The instrument does not function
- The instrument has been subjected to stresses beyond the tolerance limits (e.g., during storage
or transport).
1.3.6 Fuse
This instrument does not have a replacable fuse. The load cell supply voltage is protected against
short circuit. In case of failure of the load cell supply voltage, disconnect the instrument from the
supply voltage, determine the cause and take remedial measures. Subsequently, the supply voltage
can be switched on again.
1.3.7 EMC-Compliant Installation
Use only screened data cables
--
Connect screens on both ends with ground
--
Keep unscreened cable ends short
Connect screen rail to cabinet / housing with
-
low impedance
-
Use metal or metalized connector housings
Establish equi-potential bonding between
instruments / system modules (Mandatory for
-
Ex-applications)
-
Use standard reference potential
-
Connect mounting rail to protective earth
Install measure and data cables separately
from power cables
Three PR 5220 series transmitter versions are available; subsequent extension of the version is not
possible.
The version is determined unambiguously by the type number. The front foils are adapted to the
version.
PR 5220/00
PR 5220/01
PR 5220/04
2.1.1 PR 5220/00 Version
This version has digital inputs and outputs as well as an analog output and a LAN adaptor for
configuration and operation of the instrument. Connecting e.g. a remote indicator is possible via the
serial output.
2.1.2 PR 5220/01 Profibus
In addition to PR 5220/00, the instrument is provided with a Profibus port.
2.1.3 PR 5220/04 DeviceNet
In addition to PR 5220/00, the instrument is provided with a DeviceNet port.
05/11 11 51209 Revision 2
Section 2: PR 5220 Ethernet Transmitter Series
2.2 Overview of the Instrument
-
EC test certificate / type approval applied for
-
Accuracy 10,000 e @ 6 samples/sec
-
Internal resolution: 7.5 million counts
-
Linearity: < 0.002%
-
Sampling rate: 6 ... 100/sec selectable
-
Digital filter with selectable characteristic
-
Electrically isolated interfaces
-
3 programmable pairs of limit values
-
24 VDC supply voltage connection
-
Connection using plug-in terminal blocks
-
Socket for LAN adaptor
-
Port for Profibus (/01)
-
The instrument is provided for snap-on mounting on a
standard rail.
-
5 status LEDs für supply voltage, communication, error
detection
Calibration and configuration of the instrument are menu guided
using a PC, or by Profibus commands (PR 5220/01 only).
-
Calibration with weight, using the mV/V method or with load
cell data (“smart calibration”)
-
0/4 ... 20 mA analog output, configurable for gross/net
weight
-
Analog value via Profibus (PR 5220/01 only)
-
3 digital input channels, electrically isolated
-
3 digital output channels, electrically isolated
Communication protocols:
-
RS-422 for remote display
-
RS-485 for SMA protocol
-
Profibus-DP (PR 5220/01 only)
-
DeviceNet (PR 5220/04 only)
2.3 Label on the Housing
A label with the wiring diagram is located on one side of the instrument:
05/11 12 51209 Revision 2
2.4 Housing Dimensions
Section 2: PR 5220 Ethernet Transmitter Series
PR 5220/00 PR 5220/01, -/04
05/11 13 51209 Revision 2
2.5 Operation Using the VNC program
2.5.1 Status Symbols
Section 2: PR 5220 Ethernet Transmitter Series
Gross weight display
(G with NTEP or NSC mode)
Net weight display Stability of the weight value
,
2.5.2 Transmitter Keys
Tare weight or fixtare display
Instrument settings, set-up
Taring, the current gross weight is
stored in the tare memory, provided
that:
- weight value is stable
- indicator not in error status
(function dependent on
configuration)
Gross weight
Tare weight
The weight value is within +/- ¼ d of
zero
Start printing (without function)
Set gross weight to zero, provided
that:
- weight value is stable
- weight within zero setting range
Analog test, weighing function
2.5.3 Navigation Keys, Other Keys
Move cursor right when editing;
selection
Move cursor left when editing;
selection
Scroll down in the menu
Scroll up in the menu
Function key
Information key
Exit from current menu
Enter / confirm
05/11 14 51209 Revision 2
Section 2: PR 5220 Ethernet Transmitter Series
2.5.4 Selection Using the Navigation Keys
Press the down arrow key to scroll down, or the up arrow key to scroll up in a menu. Press
to select a menu item. To choose the desired setting for the selected menu item, press or .
Press the key to exit a menu and continue the operation on the next higher level.
An arrow in front of a menu item indicates that there are menu sublevels. The menu item selected
by pressing is shown inversely.
Info
Show version
Show status
Press the key to select an item.
If the list of menu items is long, a vertical bar graph on the left (black and gray) shows which part of
the list is displayed.
WP A/Calibration
Measuretime
320 ms
Digital filter off
Test mode absolute
W & M
none
Standstill time 0.50 s
Standstill range 1.00 d
Availability of settings options (selectable with or ) is indicated by preceding double arrows .
Measuretime
WP A/Calibration
640 ms Press to select the measuring time.
05/11 15 51209 Revision 2
2.5.5 System Messages during Input
The following types of messages are displayed as confirmation prompts / warnings during input:
Question mark
A question mark indicates that a choice of options (e.g. [Save] for
saving or [Undo] for cancelling) is available.
Stop message
An important indication that an action cannot be executed (e.g., if
saving is not possible because the CAL switch is closed). Read
the description and press [OK] to continue:
Section 2: PR 5220 Ethernet Transmitter Series
Processing
Warning
Informational text
If an action takes a long time (e.g., Max for setting the full scale
deflection), a clock symbol is shown.
A warning is marked by three exclamation points.
An informational text is marked by one exclamation point.
Execution message
Successful execution of an action is indicated by a checkmark.
The graphics are not always included when system messages are depicted in this manual.
05/11 16 51209 Revision 2
Section 3: Installing the Instrument
3 Installing the Instrument
•Before starting work, please read Chapter 1 and follow all instructions.
Further procedures:
• Check the consignment: unpack the components specific to the application.
• Safety check: inspect all components for damage.
• Make sure the on-site installation is correct and complete including cables, e.g. power cable fuse
protection, load cells, cable junction box, data cable, console/cabinet, etc.
• Follow the instructions for installation of the unit relating to application, safety, ventilation, sealing
and environmental influences.
• Connect the cable from cable junction box or platform/load cell.
• If applicable: connect other data cables, network cables, etc.
• Connect the instrument to the supply voltage.
• Check the installation.
05/11 17 51209 Revision 2
Section 3: Installing the Instrument
3.1 Network Port
The network port is built in as standard equipment. The LEDs on the connector indicate whether the
port is functioning.
Remote operation of the instrument from the PC is
possible; install VNC program version 3.3.7 on the PC.
For setting the network address, see Chapter 5.2.2.
Transfer rate: 10 Mbit/sec, 100Mbit/sec,
full/half duplex, auto-detection
Connection
method:
Cable: CAT 5 patch cable, shielded twisted pair
Cable impedance: 150 ohms
Connection: RJ45 socket on top of housing
Point to point
05/11 18 51209 Revision 2
Section 3: Installing the Instrument
3.2 Connections
3.2.1 Load Cell Connection
The cable colors shown in this chapter are applicable to the PR 62XX series load
cells.
Before connecting other types, please, carefully follow the information related to the
assignment of load cell / platform cable colors.
3.2.2 Connection Using 6-Wire Technology
See also label on the housing outside (Chapter 2.3) and manual of the cable junction box.
Terminal
SIGN. M+
SIGN. M-
SENSE S+
SENSE S-
LC SUPPLY
V+
LC SUPPLY
V-
Description
+ signal / LC output
- signal / LC output
+ sense
- sense
+ supply / excitation
- supply / excitation
Recommendations: - Install cable in steel pipe connected to earth potential
- Min. distance to high-voltage cables: 1m
Load cell supply circuit:
The load cell supply voltage is fixed to 12VDC and protected against short circuit
Load resistance of load cells > 75 ohms, e.g. 8 load cells of 650 ohms each
3.2.3 Connection of a Load Cell in 4-Wire Technology
Note that links between SENSE S+ and LC SUPPLY V+ and between SENSE S- and LC SUPPLY
V- directly at the transmitter must be provided.
Terminal
Description
SIGN. M+
SIGN. M-
SENSE S+
SENSE S-
LC SUPPLY
V+
LC SUPPLY
+ signal / LC output
- signal / LC output
+ sense
- sense
+ supply / excitation
- supply / excitation
V-
05/11 19 51209 Revision 2
Section 3: Installing the Instrument
3.2.4 Connecting PR 6221 Load Cells
See PR 6021/08, -/68 operating manual.
3.2.5 External Load Cell Supply
The internal load cell supply voltage of PR5220 (V+, V-) is not connected.
The common line of the symmetrical external supply must be connected to the same terminal of PR
5220 as the shield of the load cell / extension cable.
Specification of external supply: +/- 6 VDC + 5%, -30%; max ripple. 50 mVpp; max. asymmetry +/3%. An external supply voltage smaller than 8 VDC (+/- 4 VDC) must be set under -[Calibration]-
[Param]-[External supply].
3.2.6 Connection via PR 1626/60
Connect the instrument to PR 1626/60 as described below. For additional connections, refer to the
PR 1626/60 manual. The internal load cell supply voltage of PR5220 (V+, V-) is not connected.
If MX8 is closed, [below 8VDC] must
be set under -[Calibration]-[Param]-
[External supply].
05/11 20 51209 Revision 2
3.2.7 RS-422/485 Interface
The interface is intended for connecting a PR 1627 type remote display, a PR 1628 terminal or a
PC for data transmission using the SMA protocol.
Connection method: 4-pin plug-in terminal block
Section 3: Installing the Instrument
Number of channels /
type:
Transfer rate (Bits/s): 300, 600, 1200, 2400, 4800, <9600>,
CE, UL and cUL
Cable: DeviceNet, color: petrol-green
2x2 shielded twisted pair
Cable impedance: 150 ohms
Bus termination: 120 ohms at the cable ends
Electrical isolation: Yes, optocoupler and DC/DC
converter
Connecting diagram for a master with three slaves
05/11 26 51209 Revision 2
Section 4: Controls
4 Controls
4.1 Status Indicator LEDs
The instrument has 5 green LEDs for indication of the operating or error status.
Power supply, bus connection
* The LED for the bus activity (PR 5220/01 a. PR 5220/04) is lit as soon as there is a connection. It
continues being lit, also when there is no communication, or when the physical connection is cut.
The meaning of indicator LEDs and the operating concept are described in Chapter 4.
5.1 Data Backup/Power Failure
The calibration data and parameters as well as all configuration and interface data are stored in a
non-volatile (EAROM) memory. Unauthorized data changing can be prevented by an access code.
Additional write protection is provided for calibration data and parameters (CAL switch, see
Chapter 5.1.1).
5.1.1 CAL Switch
The CAL switch protects the calibration data / parameters against unauthorized access.
When the CAL switch is in 'open' position, the calibration data and parameters can be changed
using the PC program or via the Profibus connection.
With the CAL switch in the 'closed' position, the calibration data (e.g. dead load, Span) and
Parameters (e.g. measure time, zero tracking etc.) cannot be changed.
The CAL switch is located under a cover that can be opened by
means of a knife.
For ’legal-for-trade’ applications, set the CAL switch (4) to the
left position and seal the cover.
5.1.2 Factory Settings
Calibration data <default> Calibration data <default>
Full scale (Max) <3000> <Kg> Measure time (M) <320>ms
Scale interval <1> Measuring rate <160>ms
Dead load <0.000000>mV/V Standstill time <1>M
Span <1.000000>mV/V Standstill range <1.00>d
Tare timeout <8>M
Calibration parameters <default> <Absolute> test mode
Overload (range above Max) <9>d Zero-setting range <50.00>d
* W & M mode <off> Zero-tracking range <0.25>d
Filter <off> Zero-tracking step <0.25>d
Frequency <1.56 Hz> Zerotrack repeat <0>M
* Parameter W&M must be set to 'on' or 'off' prior to input of the calibration data, see Chapter
5.4.20.
.
05/11 28 51209 Revision 2
Section 5: Commissioning
5.2 Switching on the Instrument
The instrument can be put into operation and calibrated using a PC with the VNC program (on the
CD packed with the instrument), an Internet Browser or the ConfigureIt program.
5.2.1 Resetting the Instrument, Activating 'DHCP'
The instrument can be reset using a pin with a diameter of 1.0 mm
(e.g. paper clip).
By actuating during approx. 1 second, the program is re-started.
When actuating during approx. 3 seonds, 'DHCP' is activated in the
instrument (independent of the previous settings). This ensures that a
valid address for identification of the instrument in the network can be
assigned to the instrument by a server.
05/11 29 51209 Revision 2
5.2.2 Searching the Instrument in the Network Using 'IndicatorBrowser'
The address can be determined using the 'IndicatorBrowser' (in a directory on the CD-ROM packed
with the instrument) program.
Install and start the 'IndicatorBrowser'.
When the instrument is connected to a network, it is
“default” in the DHCP mode, i.e. an address is
assigned to it by the server. This address, e.g.
172.24.21.85, can be determined using the
‘IndicatorBrowser’ program.
Section 5: Commissioning
5.3 Operation Using a PC
5.3.1 Operation Using the VNC Program
VNC stands for 'virtual network computing' and is a program for remote operation of computers.
The program distinguishes between the VNC server and VNC client (viewer). The server program is
part of the instrument software. The client program (viewer) must be installed and executed on the
PC to be used for operating the instrument.
For direct operation using the VNC program, the IP
address (extended by :1) must be specified when you
run the program, e.g. 172.24.21.85:1.
The address range of the
controlling PC can be
limited in the instrument;
see Chapter 5.6.5.
The operator interface of
the VNC program appears:
05/11 30 51209 Revision 2
5.3.2 Operation Using Internet Browser
With the Internet browser, the IP address must be filled in.
Example:
Section 5: Commissioning
For instrument operation using the VNC program without additional installation of VNC, see Chapter
5.3.1 .
The weight value is displayed with the unit and status
symbols.
[Configuration Printout]
Can be used for printing the configuration data as a text file.
[Logfiles]
[Screenshot]
Device display for saving the display
[Show error Log]
Display of the error messages
05/11 31 51209 Revision 2
Loading...
+ 70 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.