Eutech instruments ECOTESTR PH 2 Manual

0 (0)
Eutech instruments ECOTESTR PH 2 Manual

Operating Instructions EcoTestr pH 2 Waterproof Pocket Tester

Conditioning Before First Use: Precipitation of the sensor electrolyte may cause white crystals to form around the cap. This is normal and does not a ect the tester’s performance.

1.Fill the transparent cap of the tester with 20 mm of tap water

2.Recap the tester and soak* tester for one hour

3.Rinse the cap after use

4.The sensor is now activated and ready for use

Testing:

1.Remove cap and press to turn on the tester

2.Dip sensor in at least 20 mm of test solution

3.Stir once and let the reading stabilize. Note the measured value

4.To hold reading, press . Screen flashes HO once, then displays measurement with blinking unit (pH) to indicate that tester is in the hold mode. Press again to cancel hold mode ( HC)

5.Press to shut the tester o

Note: Tester automatically shuts o after 8.5 minutes of non-use to conserve batteries.

Calibration: The tester is factory calibrated. However, to ensure accuracy, regular calibration is recommended.

1.This tester allows up to 3-point calibration. Start with pH 7.00 bu er

2.Press to turn on the tester

3.Dip sensor in 20 mm of standard solution, stir gently and wait for the value to stabilize at or near pH 7.00

4.Press to enter the calibration sequence. Display flashes CAL momentarily, then shows a flashing default reading

5.To confirm calibration, press and wait for the auto confirmation ( En)

6.You have successfully completed the first calibration point

7.To obtain your second and third calibration points, proceed to the next bu er value (pH 4.00 or pH 10.00) and repeat step 3 - 6

8.To abort calibration, press to escape ( ESC)

Maintenance:

1.Rinse sensor in tap water after each use

2.In harsh samples, take readings quickly, then rinse sensor with de-ionized water immediately to remove residues and prevent electrode contamination

Useful Notes:

1.To avoid cross contamination, rinse between samples and calibration standards with de-ionized water

2.To maintain tester accuracy, use tester regularly or calibrate at regular intervals as required

3.For long-term storage, fill cap with 20 mm of storage solution (NOT de-ionized water) and replace cap*. This will keep sensor conditioned and ready for next use

* The tester cap is not leak-proof. Please keep tester upright during soaking.

Battery

cover

 

Step 2

 

Lift and push up

 

Figure 1:

Step 1

Removing battery cover

Lift and hold

to change batteries

 

Changing Batteries:

1.To remove batteries, lift up the front battery cover to release the front catch. Push up and hold in that position

2.Next, lift the pocket clip to release the back catch and slide upwards to remove the battery cover (Figure 1)

3.Replace old batteries with fresh ones. Note polarity as shown in battery compartment

4.No re-calibration is required after battery change. The tester’s non-volatile memory function retains calibration settings

Error Messages:

1. bA – Batteries are weak and need replacement

2. Err – Wrong or bad bu er value (out of range) and/or the sensor is failing. Use fresh bu er

3. Or/Ur – Over or Under range signal, or sensor is contaminated/damaged/not fully immersed

Reset Option: Reset option allows you to restore the calibration back to factory default settings.

1.Press to turn o tester

2.Press and hold , then press to turn the tester on. The display shows a flashing rS(reset)

3.Press again to confirm reset ( En) or press to escape ( ESC) reset option

Warranty: This tester is supplied with a warranty against manufacturing defects for a period of ONE year from date of purchase.

Accessories:

Eutech

Oakton

Description

Order Code

Order Code

 

ECBU4BT

00654-00

pH 4.01 bu er solution, 480 ml bottle

ECBU7BT

00654-04

pH 7.00 bu er solution, 480 ml bottle

ECBU10BT

00654-08

pH 10.01 bu er solution, 480 ml bottle

ECRE005

00653-04

Storage solution for pH sensor, 480 ml bottle

ECDPCBT

00653-06

Protein remover solution, 480 ml bottle

 

 

 

Loading...