Operating Instructions EcoTestr EC High Waterproof Pocket Tester

Before First Use: Soak sensor for 2 minutes in alcohol to remove stains/oils. Rinse in clean water.

Testing:

1.Remove cap and press to turn on the tester

2.Dip sensor fully into the test solution

3.Stir once and let the reading stabilize. Note the measured value

4.To hold reading, press . Screen flashes HO once, then displays measurement with blinking unit (mS) to indicate that tester is in the hold mode. Press again to cancel hold mode ( HC)

5.Press to shut the tester o

Note: Tester automatically shuts o after 8.5 minutes of non-use to conserve batteries.

Calibration: The tester is factory calibrated. However, to ensure accuracy, regular calibration is recommended.

1.Soak sensor for 10 minutes in alcohol to remove stains/oils. Rinse in clean water and blot sensor dry

2.Press to turn on the tester

3.Dip sensor fully into standard solution and wait for the value to stabilize

4.Press to enter the calibration sequence. Display flashes CAL momentarily and then shows a flashing default reading

5.Press to scroll to the standard solution value

6.Release and wait for the auto confirmation

7.To abort calibration, press to escape ( ESC)

8.You have successfully calibrated the instrument

Maintenance: Clean the stainless steel sensor by periodically rinsing them in alcohol for 10 - 15 minutes. Rinse sensor in clean water after each use.

Useful Notes: To avoid cross contamination, rinse between samples and calibration standards with de-ionized water.

Battery

cover

 

Step 2

 

Lift and push up

 

Figure 1:

Step 1

Removing battery cover

Lift and hold

to change batteries

 

Changing Batteries:

1.To remove batteries, lift up the front battery cover to release the front catch. Push up and hold in that position

2.Next, lift the pocket clip to release the back catch and slide upwards to remove the battery cover (Figure 1)

3.Replace old batteries with fresh ones. Note polarity as shown in battery compartment

4.No re-calibration is required after battery change. The tester’s non-volatile memory function retains calibration settings

Error Messages:

1.bA – Batteries are weak and need replacement

2.Er – Calibration is performed in solution value below 5 % of full-scale range.

Use higher value calibration standard

Faulty temperature sensor

3.Or – Over range signal for measurements performed above the full-scale range

Reset Option: Reset option allows you to restore the calibration back to factory default settings.

1.Press to turn o tester

2.Press and hold , then press to turn the tester on. The display shows a flashing rS(reset)

3.Press again to confirm reset ( En) or press to escape ( ESC) reset option

Warranty: This tester is supplied with a warranty against manufacturing defects for a period of ONE year from date of purchase.

Accessories:

Eutech

Oakton

Description

Order Code

Order Code

 

ECCON2764BT

00653-20

2764 µS/cm KCl calibration solution

ECCON1288BT

00606-10

12.88 mS/cm KCl calibration solution

 

 

 