Eutech Instruments ALPHA PH1000 User Manual

0 (0)

Instruction Manual

alpha-pH1000

pH and ORP Controller/Transmitter

Technology Made Easy ...

68X216801 rev 1.1 07/02

Instruction Manual

alpha-pH 1000

Preface

This manual serves to explain the use of the α lpha-pH1000 series pH/ORP controller/transmitter. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the α lpha-pH1000 pH/ORP controller/transmitter. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Eutech Instruments’ Authorised Distributor.

The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Eutech Instruments Pte Ltd.

Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.

Copyright 1997 Eutech Instruments Pte Ltd. Revised in July 2002.

All rights reserved.

Instruction Manual

alpha-pH 1000

Safety Information

The Eutech Controller/ Transmitter shall be installed and operated only in the manner specified in the Instruction manual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument.

Before powering up the unit, make sure that power source it is connected to, is as specified in the top label. Failure to do so may result in a permanent damage to the unit.

The unit has live and exposed parts inside. If it has to be opened, make sure that the power to the unit is off and disconnected.

The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual.

Instruction Manual alpha-pH 1000

TABLE OF CONTENTS

1

 

INTRODUCTION

5

 

1.1

Description of Unit

5

 

1.2

Applications

5

2

 

ASSEMBLY AND INSTALLATION

6

 

2.1

Measurement and Control System

6

 

2.2

Unit Dimensions

6

3

 

ELECTRICAL CONNECTION

7

 

3.1

Connection Diagram

7

 

3.2

Back Panel

9

4

 

OVERVIEW

10

 

4.1

Keypad and Display

10

 

4.2

Function Groups

11

 

4.3

Control Concept

12

5

 

MEASUREMENT

13

 

5.1

Display in Measurement mode

13

 

5.2

Security Codes

13

6

 

CALIBRATION MODE

15

 

6.1

pH Calibration

15

 

6.2

ORP – mV Calibration

16

 

6.3

ORP -% Calibration

17

7

 

ADVANCED SET-UP MODE

18

 

7.1

Electrode Offset (OFS) sub-function

18

 

7.2

Setting temperature (Set oC) sub-function

18

 

7.3

Control Relay A/Control Relay B (SP1/SP2) sub-function

20

 

7.4

Controller (Cntr) sub-function

22

 

7.5

Current Output (rng) sub-function

24

 

7.6

Configuration (ConF) sub-function

25

 

7.7

Calibration (CAL) sub-function

28

8

 

AUTO/MANUAL MODE

29

 

8.1

Auto mode (mode after switch-on)

29

 

8.2

Manual mode

29

9

 

TECHNICAL SPECIFICATIONS

30

10

 

ACCESSORIES

32

11

 

GENERAL INFORMATION

33

 

11.1

Warranty

33

 

11.2

Packaging

33

 

11.3

Return of Goods

33

 

11.4

Guidelines for Returning Unit for Repair

33

12

 

APPENDICES

34

 

12.1

Appendix 1

34

 

12.2

Appendix 2

35

 

12.3

Appendix 3

36

 

12.4

Appendix 4

37

Instruction Manual

alpha-pH 1000

1 INTRODUCTION

1.1Description of Unit

Thank you for purchasing Eutech’s ¼ DIN alpha-1000 series pH/ORP process controllers. This unit is used for measuring either pH or ORP parameter one at a time, and the operational mode is switchable from the menu. You can use this unit to measure pH or ORP with proportional or limit control. This controller has many user-friendly and safety features which include:

Menu-driven program that simplifies set-up

Built-in memory backup to ensure that calibration and other information are not erased if power supply fails

Push-button two-point calibration and electrode offset adjustment from the keypad

Automatic temperature compensation (ATC)

Manual temperature compensation setting without the ATC probe, with independent setting for calibration and process temperature

0 to 2000 second time delay adjustment on all relays – minimise false alarms

Separately adjustable high and low set point hysteresis (dead bands) prevent chattering of relays around the set points

Asymmetrical/symmetrical input for pH/ORP mode of operation

Three control modes: limit, proportional pulse length or proportional pulse frequency

Large dual display LCD for easy reading with clear multiple annunciators, alarm status, operational and error messages

Two switching contacts as set-point triggering relays and an alarm output relay

A third relay that can be set up to perform either Alarm or Wash function

When set to work as Alarm relay, alerts you to set points exceeded for a set-point limits for certain time and if the Pt100/Pt1000 wires are broken or disconnected during the ATC function

Programmable duration and frequency in the Wash mode

Hold function freezes output current (0/4...20mA) and releases control relays

LED indicators signal control activities to monitor controller status from a distance

Protection against electromagnetic interference - galvanically isolated 0/4..20mA output provides safety for data logging and control purposes

Choice of 7 buffers from USA or NIST standards

1.2Applications

Use this controller in panel mounted enclosures for applications such as water treatment and monitoring, galvanic-decontamination, chemical processing food processing, clean or waste water control and neutralisation process.

5

Eutech Instruments ALPHA PH1000 User Manual

Instruction Manual

alpha-pH 1000

2 ASSEMBLY AND INSTALLATION

2.1 Measurement and Control System

A typical measurement system consists of:

a pH/ORP process controller

a pH/ORP combination electrode with integrated or separate temperature sensor Pt 100/1000,

an immersion, flow or process assembly with or without a potential matching pin (PMP)

a final control element such as pump or valve

a recorder and

an appropriate pH or ORP measurement cable.

Flow Assembly

pH Controller

 

 

 

 

 

Chart Recorder

 

 

alpha-pH1000

 

 

MEAS

 

 

 

7.02pH

 

 

25.0

oC

 

 

ATC

 

 

 

ALARM

 

CAL

REL A

 

 

REL B

 

 

 

REL A

 

 

 

 

ENTER

AUTO

 

 

MANU

 

 

 

REL B

 

 

ESC

 

 

 

Process Assembly

 

 

Power Mains

Measurement Cable

(220/110 VAC)

with Electrode

 

 

 

 

2.2Unit Dimensions

Flat Gasket (1mm)

(To be Inserted By Customer)

Note: The Taped Corners Have to Be On Top

92 + 0.5

92

 

96

 

 

 

56

32

92 + 0.5

Mounting Cut-Out

max. 45

max. 175

The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).

6

Instruction Manual

alpha-pH 1000

3 ELECTRICAL CONNECTION

3.1Connection Diagram

Caution: Ensure electrical mains is disconnected before proceeding.

pH

 

 

 

 

 

 

 

 

 

 

 

 

 

V

 

 

 

 

 

S/

S/

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

18

19

20

21

22

 

 

 

 

 

 

 

 

 

BNC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PE/S

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pt 100 Potential Matching Pin (PMP)

Signal Input pH/ORP

 

Power Mains

 

 

 

 

 

Relay A

 

 

Relay B

Alarm/Wash

 

 

 

 

 

 

 

Hold Input

Signal Output

pH

mA

- +

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

 

 

PE/S

AC: L

N

PE

* ) indicated contact positions are for currentless conditions

7

Instruction Manual

 

 

 

 

 

 

 

alpha-pH 1000

 

 

 

 

 

 

Temp.

pH

PE

N

L

S

0 / PA

 

 

 

 

 

 

 

 

 

 

 

 

3

2

1

22

21

20

19

18

S

BNC

Symmetrical high-impedance connections

Measuring Cable

 

 

 

 

 

Potential

 

pH Combination

 

 

Matching Pin (PM)

 

Electrode

 

 

 

 

Asymmetrical connections

8

Instruction Manual

alpha-pH 1000

3.2Back Panel

Caution: Ensure electrical mains is disconnected before proceeding.

The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and the other is the 5-way connector.

Connection for the 17-way screw terminals (from left to right):

1.

AC mains live wire

10.

Alarm/Wash relay working position (NO)

2.

AC mains neutral wire

11.

Alarm/Wash relay common

3.

AC mains protective earth wire

12.

Alarm/Wash relay resting position (NC)

4.

Low set relay resting position (NC)

13.

Hold function switch terminal 1

5.

Low set relay common

14.

Hold function switch terminal 2

6.

Low set relay working position (NO)

15.

No connection

7.

High set relay resting position (NC)

16.

0/4 - 20 mA for -ve connection

8.

High set relay common

17.

0/4 - 20 mA for +ve connection

9.

High set relay working position (NO)

 

 

Connections for the 5-way screw terminals:

18.Pt100/Pt1000 lead 1 terminal

19.Pt100/Pt1000 sense lead terminal

20.Pt100/Pt1000 lead 2 terminal

Note: If using a two-wire RTD, short terminal 19 to terminal 18.

Pt 100/Pt 1000 is selectable via an internal jumper. Factory default is Pt100. See Appendix 1 for directions on switching the RTD type.

21.pH/ORP (potential matching pin)

21. pH/ORP (shield)

Pt100/

0V/

S/

Pt1000

PAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FUSE 250VAC 100mA (F)

L N PE

 

RELAY A

RELAY B

ALARM / WASH HOLD

NC

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-

+

IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts of the relay (11 and 12). In the Wash mode, the relay works on a positive mode. Therefore, the Wash device should be connected to the contacts 10 and 11.

9

Instruction Manual

alpha-pH 1000

4 OVERVIEW

4.1Keypad and Display

4.1.1Keypad

CAL

Perform rapid 2-point calibration

Allows entry to Set up mode

Select individual functions within the function group of Set up mode

Store input data in the Set up mode

Start calibration in the calibration mode

Select various function groups in the Set up mode.

Set parameters and numerical values in sub functions of Set up mode If pressed continuously, the setting speed increases

Control the relays in the MANUAL function

Return to the Measurement mode when both keys are pressed together

Switch between AUTO and MANUAL relay operation using a code

Display limit set-point values for the switch contacts in AUTO relay operation mode

Switch between RELAY A and RELAY B in MANUAL relay operation mode

4.1.2Display

The LCD display features two numerical displays that show status messages and measured values for easy, quick reference. The display provides short-text information for setting parameters and configuration.

 

 

MEAS: Measurement mode

 

 

SETUP: Set-up mode of function groups

 

 

CAL: Calibration mode of pH/ORP and temperature

SETUP

MEAS CAL CON

READY: Comes on after a successful calibration

READY

7.00 pH

HOLD

HOLD: Relay position and current output are frozen

ERR

oC

ATC: Comes on in the ATC mode. Disappears in the

 

25.0 ATC

 

Manual temperature Compensation mode.

 

 

 

“ATC” flashes if the temperature probe is faulty in its ATC

 

 

 

mode

ERR: Error or alarm indicator

Display for RELAY A/B. Green LED indicates measured value within limit while RED LED indicates measured value outside limit.

Alarm display if limit value overshoot or the ATC probe fails; with ERR annunciator appears on the display. Wash display if cleaning cycle is on.

10

Instruction Manual

alpha-pH 1000

4.2Function Groups

The main function and sub-function groups are organised in a matrix format for configuration and selection of parameters. The main function groups are:

1)Offset adjustment (OFS)

2)Temperature Measurement / compensation settings (Set oC)

3)Control relay 1 configuration (SP1)

4)Control relay 2 configuration (SP2)

5)Control type (Cntr)

6)Current output (rng)

7)Configuration (ConF)

8)Calibration (CAL pH)

SETUP

 

 

SETUP

HOLD OFS

 

 

HOLD

 

 

 

 

 

 

SEt

oC

SETUP

 

SETUP

HOLD SP1

 

HOLD SP2

 

 

 

SETUP

HOLD

CAL

pH

 

SETUP

 

SETUP

 

SETUP

 

HOLD ConF

 

HOLD rng

 

HOLD Cntr

 

 

 

 

 

 

 

 

 

 

 

 

The set-up parameters can be viewed or changed by entering a security code. See Section 5.2 for security code information.

4.2.1 How to view operating parameters without access to change them:

 

SETUP

000

a) Press the ENTER key. The display will prompt the user to enter a

 

 

 

HOLD

 

security code (S.Cd). Leave the security code at “000” (do not enter a

 

 

S.Cd

 

security code).

 

 

b)

Press ENTER key again. This allows you only to view (not change) any

 

 

 

 

sub-functions’ settings.

 

 

 

 

 

 

 

c)

Press the ▲ or ▼ key to scroll through the sub-functions.

d) Press the ENTER key at a particular sub-function to view in detail.

e)

Press the ENTER key to return to the sub-function menu.

f)

Press the ▲ and ▼ keys simultaneously (as an Escape key) at any time to return to the Measurement

 

mode.

 

 

 

Note: To simplify operations, the controller will not display parameters that are not relevant to a particular subfunction. For example: If the user set the controller for Limit control, it will not display pulse length/frequency settings.

11

Instruction Manual

alpha-pH 1000

4.3Control Concept

The main function and sub-function groups are organised in a matrix format as shown below. These functions can be accessed via the front keypad for configuration and selection of parameters.

MEASURE

CODE

OFFSET

SET oC

SET-POINT 1

SET-POINT 2

CONTROLLER

 

Set Offset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Switch On/Off ATC

 

 

 

 

Process temp.

 

 

 

 

Calib. temp.

 

 

 

 

 

Calibrate Pt100

 

 

 

 

 

 

 

 

 

 

 

 

 

(MTC)

 

 

 

 

(MTC)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set value 1

 

 

 

MIN / MAX Function

 

 

 

 

Hysteresis 1

 

 

 

 

 

Pick-up delay

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

contact 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set value 2

 

 

 

MIN / MAX Function

 

 

 

 

Hysteresis 2

 

 

 

 

 

Pick-up delay

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

contact 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Controller type

 

 

 

N.O/N.C. Contact

 

 

 

 

Proportional

 

 

 

 

 

Cycle time bzw.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

range XP

 

 

 

 

 

max. frequency

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MEASUREMENT

 

 

 

 

 

Current output

 

 

pH / ORP

 

 

pH / ORP

 

 

 

 

 

 

 

RANGE

 

 

 

 

 

0..20 / 4..20 mA

 

 

at 0/4 mA

 

 

at 20 mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONFIGURATION

 

 

 

 

 

pH / ORP Symm.

 

 

ALr

 

 

Alarm delay

 

 

 

 

Alarm contact

 

 

 

 

 

 

 

 

 

 

 

 

 

/ Asymmetrical

 

 

 

 

 

 

 

cont./wiper contact

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLn

 

 

 

 

 

 

1000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

00

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Intv

 

 

 

 

Dur

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CALIBRATION

 

 

 

 

Calibrate 1st value

 

Calibrate 2nd value

 

Display zero point

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and slope

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dropout delay contact 1

Dropout delay contact 2

Electrode type

 

 

Reset to factory

Glass/Antimony

 

 

default settings

 

 

 

 

 

The controller offers two levels of password protection: (1) for direct access to calibration function and (2) for setting or editing specific controller parameters or functions in the SETUP mode to suit individual requirements.

Note: The passwords are not user-defined and have been set by factory. It is very important to keep these passwords strictly confidential to avoid unauthorised tampering of the system at all times.

Note: If the user reads parameters only, the controller automatically reverts to Measurement mode if no key is pressed for 30 seconds.

12