CHAPTER 12
General Technical Data
Power connection
Rated supply voltage 100 V to 240 V (± 10 %)
Rated supply frequency 50 / 60 Hz
Line current Max. 2.0 A
Power input in standby mode < 15 watts
Power input with max. HF output 190 watts / 200 VA
Terminal for grounding (potential equalization) Ye s
12 • General Technical Data
Power fuses T 4 A
Operating mode
Intermittent operation Duty cycle 25% (10 s ON / 30 s OFF)
Dimensions and weight
80113-114
02 / 2010
Width x height x depth 280 x 135 x 300 mm / 11.0" x 5.3" x 11.8"
Weig ht 4.0 kg / 8 lbs. 13 oz.
Ambient conditions for transport and storage of unit
Temperature -40 °C to +70 °C / -40 °F to +158 °F
Relative humidity 10% – 95%
Ambient conditions for operation of unit
Temperature +10 °C to +40 °C / +50 °F to +104 °F
Relative humidity 15% – 80%, noncondensing
Acclimatizing
If the unit has been stored or transported at temperatures below +10 °C (+50 °F) or above +40 °C (+104 °F), the unit will
require approx. 3
hours to acclimatize at room temperature.
75 / 86

12 • General Technical Data
Standards
Protective class according to UL 60601-1 I
Classification according to EC Directive 93/42/EEC II b
Protection class as per EN 60 601-1 I
Type as per EN 60 601-1 CF
80113-114
02 / 2010
76 / 86