ERBE SystemCarrier compact, SystemCarrier compact plus User Manual

Page 1
USER MANUAL
USA
SystemCarrier
compact/
compact plus
80180-054_V24892 2023-04
Page 2
Page 3
USER MANUAL
compact plus
Page 4
Erbe Elektromedizin GmbH owns these trademarks in the United States: APC, APC 300, AUTO CUT, BICLAMP, ENDO CUT, ERBE, ERBECRYO, erbe power your performance., ERBEJET, ERBELIFT, ERBOKRYO, FIAPC, FORCED COAG, HYBRIDAPC, HYBRIDKNIFE, ICC 200, KYRON, NESSY, NESSY Ω,
demarks
Tra
®
PRECISE APC, PULSED APC, REMODE, SOFT COAG, SWIFT COAG, THERMO SEAL, TWIN COAG, VIO.
Erbe USA, Inc. owns these trademarks: CLEVERCAP, ECO2, ERBEFLO, ERBEFLO AeroRinse, THE TRUE BLUE PROBE FOR APC, THE TRUE BLUE PROBE FOR ARGON PLASMA COAGULATION and the color blue for APC probes.
Trademarks
TM
Erbe USA, Inc. uses these trademarks: APC 360° Circumferential Probe, AUTO START, AUTO STOP, BIPOLAR COAG, BIPOLAR CUT, DRY CUT, dryCUT, EIP, ESM, FORCED APC, forcedAPC, highCUT, POWER PEAK SYSTEM PPS, preciseSECT, PULSED COAG, PULSED CUT, SPRAY COAG, sprayCOAG.
For an up-to-date list of all Erbe trademarks, please visit www.erbe-med.com/IP.
REF No. 20180-100, 20180-101
All rights to this User Manual, in particular rights of duplica
t
ion, dissemination and translation, are reserved. No part of this User Manual may be reproduced in any form (by photocopying, microfilming or other methods) or processed, duplicated or dissemi­nated by the use of electronic systems without the written consent of Erbe Elektromedizin GmbH.
The information contained in this User Manual can be changed or
panded without prior notice and without obligation on the
ex
part of Erbe Elektromedizin GmbH.
The illustrations in this User Manual may differ in some details from your product.
Printed by Erbe Elektromedizin
Printed in Germany
Copyright © Erbe Elektromedizin
G
mbH, Tübingen 2023
Page 5
Table of Contents
1 General Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
80180-054_V24892
2023-04
3 Description of the elements of the SystemCarrier compact . . . . . . 17
Table of Contents
Intended use / indications for SystemCarrier compact, compact plus. . . . . . . . . . . . 7
Compatibility of SystemCarrier compact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compatibility of SystemCarrier compact plus. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Qualification of user . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety notations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Meaning of the note. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Who must read this User Manual? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compliance with safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Structure of safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating errors and incorrect installation by persons without training. . . . . .
Risks due to the environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electric shock, only applies to SystemCarrier compact plus . . . . . . . . . . . . . . . .
Burns. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Risk of injury due to installation of incompatible units or accessories . . . . . . .
Risk of injury due to damage to the cart and accessories . . . . . . . . . . . . . . . . .
Risk of infection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 7
. . . . 7
. . . . 8
. . . . 8
. . . . 8
. . . . 9
. . . . 9
. . . 10 . . . 10 . . . 10 . . . 11 . . . 11 . . . 12 . . . 13 . . . 13 . . . 15 . . . 15
4 Description of the elements of the SystemCarrier compact plus
5 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Checking the cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Installation of units, gas bottles, and accessories . . . . . . . . . . . . . . . . . . . . . . . .
Pushing the cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Braking the cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the shelf for two-pedal footswitch, footswitch holders, and cable
holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the SystemCarrier compact storage compartments . . . . . . . . . . . . . . . . .
Using the SystemCarrier compact plus gas bottle compartment . . . . . . . . . . . .
SystemCarrier compact plus power fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 21
. . 25
. . . 25 . . . 25 . . . 26
. . . 27 . . . 27 . . . 27 . . . 28
6 Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Selecting a suitable cleaning and disinfection agent . . . . . . . . . . . . . . . . . . . . .
Supplies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Validated procedure for cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . .
. . . 29 . . . 30 . . . 30 . . . 30 . . . 31
5 / 50
Page 6
Table of Contents
7 Technical data for SystemCarrier compact . . . . . . . . . . . . . . . . . . . . 33
8 Technical data for SystemCarrier compact plus . . . . . . . . . . . . . .
9 Maintenance, Customer Service, Warranty, Disposal . . . . . . . . . .
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 35
. . 37
. . . .37
. . . .38
. . . .39
10 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Accessories for SystemCarrier compact, compact plus . . . . . . . . . . . . . . . . . . . . . . . .41
Compatible power cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . .41
11 Spare parts for SystemCarrier compact . . . . . . . . . . . . . . . . . . . . . . 43
12 Spare parts for SystemCarrier compact plus . . . . . . . . . . . . . . . .
13 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 45
. . 47
80180-054_V24892
2023-04
6 / 50
Page 7
Chapter 1

General Instructions for Use

This User Manual describes the normal use of the cart.
1 General Instructions for Use
IES 3
EIP 2 - -
80180-054_V24892
2023-04
Note: Report serious inciden a user in the European Union, also report incidents to the responsible authority in your Member State.
Please observe the Warnings in Chapter 2.
ts with the product to your local dealer or Erbe. If you are

Intended use / indications for SystemCarrier compact, compact plus

The SystemCarrier compact and compact plus are intended to accommodate Erbe units and accessories.

Compatibility of SystemCarrier compact

The SystemCarrier compact is intended for use with the following units:
VIO C VIO 200 S VIO 300 S VIO 200 D VIO 300 D VIO 3
-
-
Do not connect any other units to the SystemCarrier compact.
-
-
-
-
-
-
-

Compatibility of SystemCarrier compact plus

The SystemCarrier compact plus is intended for use with the following units:
VIO 200 S VIO 300 S VIO 200 D VIO 300 D VIO 3
IES 3
EIP 2 -
APC 2
APC 3 - - - - - - - -
-
-
Do not connect any other units to the SystemCarrier compact plus.
-
-
-
-
-
-
- -
-
7 / 50
Page 8
1 General Instructions for Use

Environment

For the intended use, the cart may only be operated in premises used for medical pur­poses.

Qualification of user

For the intended use, the cart may only be operated by medical professionals who have been trained in the use of the cart on the basis of the User Manual.

Performance characteristics

This is an accessory for a medical device. There are no clinical performance character­istics.
80180-054_V24892
2023-04
8 / 50
Page 9
Chapter 2

Safety Instructions

Safety notations

indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
2 Safety Instructions
DANGER
WARNING
CAUTION
indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
NOTICE
indicates a potentially hazardous situation which, if not avoided, may result in property damage.
80180-054_V24892
2023-04

Meaning of the note

"Note:"
Refers a) to manufacturer's information that relates directly or indirectly to the safety of
people or protection of property. The information does not relate directly to a risk
or dangerous situation.
Refers b) to manufacturer's information that is important or useful for operating or ser
icing the unit.
v

Who must read this User Manual?

Knowledge of the User Manual is absolutely essential for correct operation and instal­lation of the cart.
The User Manual must therefore be read by everyone who works with the cart, installs u
its on it, or attaches accessories to it.
n
Anyone who prepares, disassembles, cleans, or disinfects the cart must also read the User
Please pay particular attention to the safety instructions in each chapter.
Man
ual.
9 / 50
Page 10
2 Safety Instructions

Compliance with safety information

Working with medical units is associated with certain risks to patients, medical per­sonnel and the environment. Risks cannot be en alone.
Safety does not depend solely on the unit. Safety depends to a large extent on the t
aining of medical personnel and correct operation of the unit.
r
The safety instructions in this chapter must be read, understood, and applied by ev­eryone who prepares, disassembles, cleans, or disinfects the cart.
Anyone who installs units on it or attaches accessories to it must also read, under­stand, and apply the safety instructions.
tirely eliminated by design measures

Structure of safety instructions

The safety instructions are structured according to the following risks:
Operating errors and incorrect installation by persons without training
Risks due to the environment
Electric shock, only applies to SystemCarrier compact plus
Burns
Risk of injury due to installation of incompatible units or accessories
Risk of injury due to damage to the cart and accessories
Risk of infection
Notes

Operating errors and incorrect installation by persons without training

WARNING
Incorrect operation of the cart, incorrect installation of units and accessories
Risk to medical staff and to the installer! Risk of damage to property.
The cart must only be used by persons who have been trained on
how to use it properly according to this User Manual.
Units and accessories on the cart must only be installed by em-
ployees of Erbe or by persons trained by Erbe. Both of these groups must comply with the User Manual and the assembly in­structions.
Instruction and training must only be provided by persons who
are suitable on the basis of their knowledge and practical experi­ence.
In the event of uncertainties or if you have any questions, please
contact Erbe USA. You will find the addresses in the address list at the end of this User Manual.
80180-054_V24892
2023-04
10 / 50
Page 11
2 Safety Instructions
NOTICE
WARNING
WARNING

Risks due to the environment

NOTICE
Unsuitable temperature or humidity in transit or storage
If you transport or store the cart at an unsuitable temperature or level of humidity, it may be damaged.
Transport and store the cart at a suitable temperature and level
of humidity. You will find the tolerances for temperature and hu­midity in the Technical Data.
If other ambient conditions have to be observed for transport and
storage of the cart, you will also find them in the Technical Data.
Unsuitable temperature or level of humidity during operation
If you operate the cart and unit at an unsuitable temperature or level of humidity, they may sustain damage, fail, or not perform properly.
Operate the cart and unit at a suitable temperature and level of
humidity. You will find the tolerances for temperature and humid­ity in the Technical Data.
If other ambient conditions have to be observed for operation of
the cart and unit, you will also find them in the Technical Data.
NOTICE
Overheating of a unit or units on the cart due to bad ventilation
If ventilation is bad, units can overheat, sustain damage, and fail.
Install the cart in such a way that there is an unobstructed circu-
lation of air around the unit housings. Installation in confined wall
80180-054_V24892
2023-04
recesses is prohibited.

Electric shock, only applies to SystemCarrier compact plus

Damaged cart or modified cart
Risk of electric shock and other injuries to medical personnel! Risk of damage to property.
Check the cart for damage every time before using it. You must
not use a damaged cart.
If the cart is damaged, please contact Erbe USA.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in­validate liability on the part of Erbe Elektromedizin GmbH.
Defective grounded power outlet, power supply network without proper grounding, inferior-quality power cord, incorrect line volt­age, multiple power outlets, extension cords
Risk of electric shock and other injuries to medical personnel! Risk of damage to property.
11 / 50
Page 12
2 Safety Instructions
1
Connect the unit / cart to a properly installed grounded power
outlet.
Use the compatible Erbe power cord (see the Accessories, Com-
patible power cords chapter).
Check the power cord for damage. You must not use a damaged
power cord.
The supply voltage must match the voltage specified on the unit's
rating plate.
Do not use multiple power outlets.
Do not use extension cords.
WARNING
Connection of unit / cart and power supply during cleaning and disinfection
Risk of electric shock to the medical personnel!
Switch off the unit. Unplug the power plug of the unit / cart.
CAUTION
Incorrect power fuse
The power supply cable may get hot – risk of burning.
Power fuses for adapting the unit configuration on the cart may
only be installed by a qualified technician. Only fuses that have the same rating as the one specified on the rating plate of the cart may be used.
Blown power fuses may only be replaced by a competent techni-
cian. Only use spare fuses with the same rating as specified on the cart rating plate.
The fuse holder (1) with the fuse is shown below.
80180-054_V24892
2023-04
12 / 50

Burns

CAUTION
Dirty casters
Dirt may affect the discharge capacity of the casters. A spark may ig­nite flammable gases.
Risk of injury to the medical personnel! Risk of damage to property.
Clean dirty casters.
Page 13
2 Safety Instructions
WARNING

Risk of injury due to installation of incompatible units or accessories

CAUTION
Units or accessories that are not compatible are installed on the cart. Improper items are placed on or in the cart.
Units, accessories, and items may fall off.
Risk of injury to the medical personnel! Risk of damage to property.
Only install units specified in the section on compatibility.
Only install accessories specified in the section on compatibility.
Only use compatible footswitches for the footswitch holder and
footswitch shelf.
Only load the storage compartments of the SystemCarrier com-
pact with 15 kg max. (33.1 lbs) per compartment.
Only load the gas bottle compartment of the SystemCarrier com-
pact plus with a gas bottle weighing 15 kg (33.1 lbs) max.
CAUTION
Units not listed in the section on Compatibility are connected to the auxiliary power sockets. Units not installed on the cart are connected to the auxiliary power sockets.
The power supply cable may get hot – risk of burning.
For auxiliary power outlets see Figure below.
Only connect units to the auxiliary power outlets that are specified
in the compatibility section and that are installed on the cart.
80180-054_V24892
2023-04

Risk of injury due to damage to the cart and accessories

Damaged cart or modified cart
Risk of electric shock and other injuries to medical personnel! Risk of damage to property.
Check the cart for damage every time before using it. You must
not use a damaged cart.
If the cart is damaged, please contact our customer service.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in­validate liability on the part of Erbe Elektromedizin GmbH.
13 / 50
Page 14
2 Safety Instructions
CAUTION
The handle on the cart is used for lifting and carrying the cart
The handle can break off. The handle is only designed for pushing and pulling the cart.
Risk of injury to the medical personnel! Risk of damage to property.
Do not lift or carry the cart with the handle.
CAUTION
The cart is not pushed or pulled using the handle
Units or accessories may fall off.
Risk of injury to the medical personnel! Risk of damage to property.
Close the door before moving the cart.
Pull and push the cart only with the handle. See Figure below.
Do not push against the units.
CAUTION
The casters are not fixed with the brakes.
The cart may move unintentionally.
Risk of injury to the medical personnel! Risk of damage to property.
Lock the brakes on all four casters.
80180-054_V24892
2023-04
14 / 50
Page 15
2 Safety Instructions
WARNING
CAUTION
The gas bottle was not secured with the fixing strap
The gas bottle could fall over.
Risk of injury to the medical personnel! Risk of damage to property.
Secure the gas bottle with the fixing strap.

Risk of infection

The cart is contaminated
80180-054_V24892
2023-04
Risk of infection for the medical staff.
Make sure no liquid can penetrate the cart.
Do not place vessels containing liquids on top of the cart.
Follow the instructions for cleaning and disinfecting the cart.
WARNING
Alternate use of disinfectant solutions based on different active ingredients
The agents may affect one another. The disinfection effect of the dis­infectant solution may be weakened. Risk of infection for the medical staff.
Do not use these substances alternately.

Notes

Grounding
Note: If necessary, connect the potential equalization pin of the unit or the cart to the
potential equalization terminal of the operating room using a grounding cable.
15 / 50
Page 16
2 Safety Instructions
80180-054_V24892
2023-04
16 / 50
Page 17
3 Description of the elements of the SystemCarrier compact
1
Chapter 3

Description of the elements of the SystemCarrier compact

80180-054_V24892
2023-04
(1) Handle
For pushing or pulling the cart.
17 / 50
Page 18
3 Description of the elements of the SystemCarrier compact
2
3
3
4
5
6
7
Note: Different configurations are possible when mounting the footswitch holders on the cart. The figure shows two possible combinations.
(2) Br
akes
r braking the caster wheels. Each caster wheel is equipped with a brake.
Fo
(3) F
ootswitch holder for one-pedal footswitch of IES 3 or VIO electrosurgical units
r holding one-pedal footswitches of the unit
Fo
s IES 3, VIO 200 S, VIO 300 S, VIO 200 D,
80180-054_V24892
2023-04
VIO 300 D, VIO 3.
(4) Foo
tswitch holder for one-pedal footswitch of EIP 2 or VIO C
r holding one-pedal footswitches of the units EIP 2, VIO C.
Fo
(5) S
helf for two-pedal footswitch
r holding two-pedal footswitches of the units VIO C, VIO 200 S, VIO 300 S, VIO 200 D,
Fo VIO 300 D, V
(6) Ca
r winding up footswitch cable or power cord.
Fo
(7) Pot
r connecting the potential equalization pins of the units and for connecting the cart
Fo to th
IO 3.
ble holder for footswitch cable and power cord
ential equalization terminals
e potential equalization terminal of the operating room.
18 / 50
Page 19
3 Description of the elements of the SystemCarrier compact
8
9
80180-054_V24892
2023-04
(8) Cart door
Closes the storage compartments.
(9) Storage compartments
For
holding suitable items. Maximum load per compartment is 15 kg (33.1 lbs).
19 / 50
Page 20
3 Description of the elements of the SystemCarrier compact
80180-054_V24892
2023-04
20 / 50
Page 21
4 Description of the elements of the SystemCarrier compact plus
1
Chapter 4

Description of the elements of the SystemCarrier compact plus

80180-054_V24892
2023-04
(1) Handle
For pushing or pulling the cart.
21 / 50
Page 22
4 Description of the elements of the SystemCarrier compact plus
2
3
3
4
5
6
7
8
9
10
11
Note: Different configurations are possible when mounting the footswitch holders on the cart. The figure shows two possible combinations.
(2) Br
akes
r braking the caster wheels. Each caster wheel is equipped with a brake.
Fo
(3) F
ootswitch holder for one-pedal footswitch of IES 3 or VIO electrosurgical units
r holding one-pedal footswitches of the unit
Fo VIO 300 D, VIO 3.
(4) Foo
tswitch holder for one-pedal footswitch of EIP 2
r holding one-pedal footswitches of the unit EIP 2.
Fo
(5) S
helf for two-pedal footswitch
r holding two-pedal footswitches of the units VIO 200 S, VIO 300 S, VIO 200 D,
Fo VIO 300 D,
(6) Ca
ble holder for footswitch cable and power cord
r winding up footswitch cable or power cord.
Fo
(7) Pot
ential equalization terminals
r connecting the potential equalization pins of the units and for connecting the cart
Fo to th
e potential equalization terminal of the operating room.
(8) Op
ening
s IES 3, VIO 200 S, VIO 300 S, VIO 200 D,
VIO 3.
80180-054_V24892
2023-04
22 / 50
For feeding through the pressure line of the gas bottle and the high-pressure sensor line.
Page 23
4 Description of the elements of the SystemCarrier compact plus
12
13
(9) Auxiliary power outlets
Only connect units to the auxiliary power outlets that are specified in the compatibility sect
ion and that are installed on the cart. Use the compatible Erbe power cord (see the
Accessories, Compatible power cords section).
(10) Po
wer fuses
own power fuses may only be replaced by a competent technician. Only use spare
Bl fuses
with the same rating as specified on the cart rating plate.
(11) Po
wer connection
nnect the cart to a properly installed, grounded power outlet. Use the compatible
Co Erbe po
wer cord (see the Accessories, Compatible power cords section).
80180-054_V24892
2023-04
(12) Cart door
loses the gas bottle compartment.
C
(13) Gas
For h
bottle compartment
olding a gas bottle.
23 / 50
Page 24
4 Description of the elements of the SystemCarrier compact plus
80180-054_V24892
2023-04
24 / 50
Page 25
Chapter 5

Operation

5 Operation

Checking the cart

WARNING
Damaged cart or modified cart
Risk of electric shock and other injuries to medical personnel! Risk of damage to property.
Check the cart for damage every time before using it. You must
not use a damaged cart.
If the cart is damaged, please contact Erbe USA.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in­validate liability on the part of Erbe Elektromedizin GmbH.

Installation of units, gas bottles, and accessories

The installation is described in the assembly instructions.
80180-054_V24892
2023-04

Pushing the cart

CAUTION
The handle on the cart is used for lifting and carrying the cart
The handle can break off. The handle is only designed for pushing and pulling the cart.
Risk of injury to the medical personnel! Risk of damage to property.
Do not lift or carry the cart with the handle.
CAUTION
The cart is not pushed or pulled using the handle
Units or accessories may fall off.
Risk of injury to the medical personnel! Risk of damage to property.
Close the door before moving the cart.
Pull and push the cart only with the handle. See Figure below.
Do not push against the units.
25 / 50
Page 26
5 Operation
Close the door before moving the cart.Push or pull the cart only with the handle. Do not push against the units.

Braking the cart

CAUTION
The casters are not fixed with the brakes.
The cart may move unintentionally.
Risk of injury to the medical personnel! Risk of damage to property.
Lock the brakes on all four casters.
80180-054_V24892
2023-04
26 / 50
Lock all four caster wheels with the brakes when the cart is in the required posi-
tion.
Page 27
5 Operation

Using the shelf for two-pedal footswitch, footswitch holders, and cable holders

Place the two-pedal footswitch on the shelf.
80180-054_V24892
2023-04
Insert the one-pedal footswitch into the correct footswitch holder. Coil the footswitch cable and power cord around the cable holders.

Using the SystemCarrier compact storage compartments

Use the storage compartments for suitable items. Maximum load per compartment is 15 kg (33.1 lbs) Close the door before moving the cart.

Using the SystemCarrier compact plus gas bottle compartment

The gas bottle compartment can hold gas bottles weighing up to maximum 15 kg (33.1 lbs). The maximum diameter is 172 mm (6.8“); the maximum height is 740 mm (29.1“
).
27 / 50
Page 28
5 Operation
1

SystemCarrier compact plus power fuses

CAUTION
Incorrect power fuse
The power supply cable may get hot – risk of burning.
Power fuses for adapting the unit configuration on the cart may
only be installed by a qualified technician. Only fuses that have the same rating as the one specified on the rating plate of the cart may be used.
Blown power fuses may only be replaced by a competent techni-
cian. Only use spare fuses with the same rating as specified on the cart rating plate.
The fuse holder (1) with the fuse is shown below.
The cart as delivered has 2 T 10 A H / 250 V fuses for connecting units with a voltage between 220 V to 240 V.
If you wish to connect units with a voltage between 100 V to 120 V, a qualified tech­nician must replace the fuses with the two included T 12.5 A H / 250 V fuses. If the f
ses are not changed, they may blow.
u
80180-054_V24892
2023-04
28 / 50
Page 29
Chapter 6
WARNING

Cleaning and disinfection

Safety Instructions

WARNING
Connection of unit / cart and power supply during cleaning and disinfection
Risk of electric shock to the medical personnel!
Switch off the unit. Unplug the power plug of the unit / cart.
WARNING
The cart is contaminated
Risk of infection for the medical staff.
Make sure no liquid can penetrate the cart.
Do not place vessels containing liquids on top of the cart.
Follow the instructions for cleaning and disinfecting the cart.
6 Cleaning and disinfection
Alternate use of disinfectant solutions based on different active
80180-054_V24892
2023-04
ingredients
The agents may affect one another. The disinfection effect of the dis­infectant solution may be weakened. Risk of infection for the medical staff.
Do not use these substances alternately.
CAUTION
Dirty casters
Dirt may affect the discharge capacity of the casters. A spark may ig­nite flammable gases.
Risk of injury to the medical personnel! Risk of damage to property.
Clean dirty casters.

Selecting a suitable cleaning and disinfection agent

The following categories of cleaners and disinfectants should not cause any damage to the surfaces of unit/cart and the mounting bracket when used properly:
Alc
ohol-based agents
Agent
Agents based on active oxygen-based per compounds
Agents based on peracetic acid
Chlorine-based agent with up to 3% bleach
s based on guanidine
29 / 50
Page 30
6 Cleaning and disinfection
Frequent use of the abovementioned agents may cause harmless discoloration to sur­faces.
Do not use a cleaning and disinfection agent containing the following substances:
Sodium hydroxide, strong organic acids
, and strong oxidizing agents
These substances may attack metal parts due to their corrosive effect.
Qu
aternary ammonium cations, phenolic compounds (e.g., phenoxyethanol)
Solvent
s and benzine
These substances cause damage to plastics.

Supplies

Ready-for-use wipes with a cleaning agent or disinfectant solution.
or
Low-lint single-use wipes moistened before
cleaning or disinfection with a suit-
able product.
Use different wipes for cleanin
and disinfecting.
g

Cleaning

Use only cleaning agents that comply with relevant national standards. Observe the manufacturer’s specifications for the cleaning agent, particularly regarding material compatibility. If you use different products for cleaning and disinfection, the products must be compatible with each other.
1. Prepare the cleaning agent solution in a concentration specified by the manufac­tu
er. Moisten low-lint single-use wipes with the cleaning agent solution.
r
2. Or use ready-for-use cleaning wipes with a cleaning agent solution.
3. Wipe the surfaces thoroughly.
4. Ensure that the surfaces are completely wetted.
5. Leave the cleaning agent to work as specified by the manufacturer (exposure time).
6. Keep the surfaces wet throughout the entire exposure time. Use additional wipes if needed.
7. Clean all visible contamination (e.g., blood) off the surfaces. Residual contamina­tion may reduce the effectiveness of the subsequent disinfection.
8. Visually inspect all surfaces after cleaning. If you find any residual contamination, repeat cleaning until all contamination has been removed.
Note: I
f you cannot carry out a full clean, do not carry out a wipe disinfection. Do not
se the unit or the cart. Contact Technical Service.
u

Disinfection

Use only disinfectants that comply with relevant national standards. Comply with the manufacturer’s specifications for the disinfectant, particularly regarding material com­patibility. If you use different products for clea be compatible with each other.
ning and disinfection, the products must
80180-054_V24892
2023-04
30 / 50
Page 31
6 Cleaning and disinfection
1. Mix the disinfectant in a concentration specified by the manufacturer. Moisten low-lint single-use wipes with the disinfectant solution.
2. Or use ready-for-use disinfectant wipes with a disinfectant solution.
3. Wipe the surfaces thoroughly.
4. Ensure that the surfaces are completely wetted.
5. Leave the disinfectant to work as specified by the manufacturer (exposure time).
6. Keep the surfaces wet throughout the entire exposure time. Use additional wipes if needed.
7. Visually inspect the surfaces for any damage after disinfection.
Note: I
f the surfaces are damaged, do not use the unit or the cart. Contact Technical
vice.
Ser

Validated procedure for cleaning and disinfection

Cleaning
Clean with Sani-Cloth Bleach Germicidal Disposable Wipes, PDI, Woodcliff Lake, NJ 07677 USA. Leave the cleaning agent to work for fo
ur minutes, making sure it is visibly wetted. Clean all visible contamination (e.g., blood) off the surfaces. Residual contamination may reduce the effectiveness of the subsequent disinfection.
Disinfection
Disinfect with Sani-Cloth Bleach Germicidal Disposable Wipes, PDI, Woodcliff Lake, NJ 07677 USA. Leave the disinfection solution to wor
k for four minutes, making sure
it is visibly wetted.
Erbe recommends this processing procedure for the unit, the cart, and the mounting brac
et. Other equivalent procedures are possible. The user must ensure that the pro-
k
cedures used are suitable by means of suitable measu
res (e.g., validation, routine
monitoring, check of material compatibility).
80180-054_V24892
2023-04
31 / 50
Page 32
6 Cleaning and disinfection
80180-054_V24892
2023-04
32 / 50
Page 33
7 Technical data for SystemCarrier compact
Chapter 7

Technical data for SystemCarrier compact

Terminal for grounding (potential equalization) Yes (3 unit s )
Caster wheel diameter 100 mm (3.9“)
The caster wheels are conductive.
Width x height x depth 615 x 955 x 615 mm (24.2“ x 37.6“ x 24.2“)
Weight 26 kg (57.3 lbs)
Maximum load 34 kg (75 lbs)
Maximum shelf load 15 kg (33.1 lbs)
Ambient conditions for transport and storage of the cart
Te mp er a tu r e -29 °C to +60 °C (-20.2 °F to +140 °F)
Relative humidity 15%–85%, non-condensing
Ambient conditions for operation of the cart
80180-054_V24892
2023-04
Te mp er a tu r e +10 °C to +40 °C (50 °F to 104 °F)
Relative humidity 15%–80%, non-condensing
Standards
Classification according to Regulation (EU) 2017/745 I
33 / 50
Page 34
7 Technical data for SystemCarrier compact
80180-054_V24892
2023-04
34 / 50
Page 35
8 Technical data for SystemCarrier compact plus
Chapter 8

Technical data for SystemCarrier compact plus

Rated supply voltage 100 - 120 VAC (±10%) /
220 - 240 VAC (±10%)
Mains rated input current 12.5 A /
A
10
Rated supply frequency 50 Hz / 60 Hz
Power fuses for 100 - 120 VAC units, included for replace­ment
Power fuses for 220 - 240 VAC units, installed on delivery 2 x T 10 A H / 250 V
Power inputs / power outputs 1 power input / 3 power outputs
Terminal for grounding (potential equalization) Yes (4 unit s )
Caster wheel diameter 100 mm (3.9“)
The caster wheels are conductive.
Width x height x depth 615 x 955 x 615 mm (24.2“ x 37.6“ x 24.2“)
Weight 26 kg (57.3 lbs)
80180-054_V24892
2023-04
Maximum load 49 kg (108 lbs)
Maximum load of the gas bottle compartment 15 kg (33.1 lbs)
Maximum size of the gas bottle Diameter 172 mm (6.8“), height 740 mm (29.1“)
Ambient conditions for transport and storage of the cart
Te mp er a tu r e -29 °C to +60 °C (-20.2 °F to +140 °F)
2 x T 12.5 A H / 250 V
Relative humidity 15%–85%, non-condensing
Ambient conditions for operation of the cart
Te mp er a tu r e +10 °C to +40 °C (50 °F to 104 °F)
Relative humidity 15%–80%, non-condensing
If you transport, store or operate the cart together with gas bottles, observe the spe­cific ambient conditions for the gas bottles.
35 / 50
Page 36
8 Technical data for SystemCarrier compact plus
Standards
Classification according to Regulation (EU) 2017/745 I
80180-054_V24892
2023-04
36 / 50
Page 37
9 Maintenance, Customer Service, Warranty, Disposal
Chapter 9

Maintenance, Customer Service, Warranty, Disposal

Maintenance

Modifications and repairs
Authorized persons
Modifications and repairs must not impair the s afety of the unit or c art and accessories for the patient, user and the environment. This condition is met when changes to the structural and functional characteristics are not detrimental to safety.
Modifications and repairs may only be undertaken by Erbe or by persons expressly au­thorized by Erbe. Erbe accepts no liability if accessories are made by unauthorized persons. This will also invalidate the warranty.
modifications and repairs to the unit or

Customer Service

If you are interested in a service contract, please contact Erbe.
Servicing
80180-054_V24892
2023-04
Procure service work as follows:
1. Contact the Technical Service Department at Erbe USA, Inc. for a Service Call or to
ob
tain a Return Authorization Number. This Number must appear on all corre-
spondences.
Tel ep ho ne : 770-955-4400 or 1-800-778-3723 E-Mail: technicalservice@erbe-usa.com FAX: 770-955-2577
Address: Erbe USA, Inc. 2225 Northwest Parkway Marietta, GA 30067
Merchandise Return
2. If the product is being returned to Erbe USA, Inc.; the shipment must be insured
as well as safely and securely packed, preferably in the original shipping carton, and should include a letter explaining the problem and make reference to the Return Authorization Number. All transportation and insurance charges and risk of loss are the responsibility of the customer and must be prepaid. A purchase order must be issued to Erbe USA, Inc. to cover all transportation and insurance charges for the return shipment after service.
Authorization from the Erbe USA, Inc. Customer or Technical Service Departments must be obtained prior to the return of any merchandise. To obtain an appropriate Return Authorization Number (R.A. #), please be prepared to give the following infor­mation when contacting Erbe:
1. The customer name and/or number as it appears on the original invoice for the
m
e
rchandise involved.
2. The name and telephone number of the contact person.
37 / 50
Page 38
9 Maintenance, Customer Service, Warranty, Disposal
3. The catalog number, or model number, and serial number of the merchandise.
4. The REASON for your request to return the merchandise.
The R.A. #, and the name and telephone number of the contact person at your location M
ST be clearly indicated on the packing list you send with the merchandise.
U
The user must decontaminate any item being returned that possess a risk of transmit­ting disease.
Erbe USA, Inc. reserves the right to refuse and/or return, transportation charges col­lect, any merchandise that does not have an appropriate R.A. # as issued by the Cus­tomer or Technical Service Department. All shipments must be insured, and safely and s
e
curely packed, preferably in the original shipping carton. All transportation and in­surance charges and risk of loss are the responsibility of the customer and must be prepaid.
Merchandise returned within 30 days of the invoice date may be eligible for full credit, i
f
unused and in resaleable condition, in the original packaging. Merchandise returned after 30 days from the invoice date will be subject to a minimum 15% restocking charge.
Merchandise returned (excluding warranty situations) in the following circum-stances i
NOT eligible for credit:
s
Additional Information needed?
1. More than 90 days has transpired since the invoice date.
2.
pecial order items.
S
3. Discontinued items.
4. Engraved or marked items.
5. Items that have been used, or are not in the original packaging, or are damaged in
any way, or are not complete as originally shipped.
Please contact Erbe with any concerns regarding the equipment or this Manual. Also, as requested and available, copies of scientific publications will be provided.

Warranty

Erbe warrants the material and workmanship of the equipment (i.e., the VIO Cart) for a period of 90 days from the delivery date.
At installation, the equipment should be inspect other deficiency. A report should be prepared with the information being brought to Erbe's attention.
If this product should become inoperable due to a defect in material or workmanship durin
warranty period; Erbe will, at its option, repair or replace the product. This ac-
g tion will be performed by Erbe authorized personnel. This limited warranty does not i
clude replacement or service to repair damage from improper installation, external
n electrical fault, accident, disaster, use for a purpose other than that for which originally designed or indicated in this Manual, negligence, modification, unauthorized service by non Erbe personnel, or normal wear.
d for transportation damage and any
e
80180-054_V24892
2023-04
38 / 50
The sole and exclusive remedy under this warrant
y
shall be repair or replacement. In no event will Erbe be liable for any damages arising out of the loss of use, or any other incidental or consequential damages.
No person, agent, distributor, dealer, or com
is authorized to change or modify
pany
the terms of this warranty.
Page 39
9 Maintenance, Customer Service, Warranty, Disposal

Disposal

Your product bears a crossed-out garbage can icon (see picture). Meaning: In all EU countries this product must be disposed of separately in accordance with the national laws implementing EU Directive 2012/19/EU of 07/04/2012, WEEE.
In non-EU countries the local regulations must be observed.
If you have any questions about disposal of the product, please contact Erbe Elektro­medizin or your local distributor.
80180-054_V24892
2023-04
39 / 50
Page 40
9 Maintenance, Customer Service, Warranty, Disposal
80180-054_V24892
2023-04
40 / 50
Page 41
Chapter 10

Accessories

10 Accessories

Accessories for SystemCarrier compact, compact plus

Use only the following accessories with these REF Nos.:
20180-463 One-pedal footswitch mount VIO footswitch
20180-464 One-pedal footswitch mount EIP 2 / VIO C footswitch
20180-460 Cable holder
20325-002 Hook for irrigation-liquid bag on bracket / VIO CART
20180-150 Fastening set device
20180-151 Fastening set VIO C on SystemCarrier compact
20180-152 Fastening set IES 3 on SystemCarrier compact
20180-153 Fastening set EIP 2 on SystemCarrier compact

Compatible power cords

Use the following Erbe power cords for the mains connection of the cart to the ground­ed power outlet of the operating room.
80180-054_V24892
2023-04
REF no. (supplied with the unit)
51704-037 52704-037 2.5 m (8.2 ft),
51704-038 52704-038 2.5 m (8.2 ft),
51704-039 52704-039 4 m (13.1 ft),
51704-040 52704-040 2.5 m (8.2 ft),
51704-042 52704-042 5 m (16.4 ft),
51704-043 52704-043 5 m (16.4 ft),
REF no. (spare part)
Length, amperage
10
A
10
A
10 A
10
A
10 A
10 A
Power plug Erbe designation
303A E+F
404 G
NEMA 5-15 P B
403 L
303A E+F
404 G
Power plug international designation
51704-045 52704-045 5 m (16.4 ft),
10 A
51704-050 52704-050 4 m (13.1 ft),
10 A
403 L
PRC/3 I
41 / 50
Page 42
10 Accessories
REF no. (supplied with the
REF no. (spare part)
Length, amperage
Power plug Erbe designation
Power plug international designation
unit)
51704-051 52704-051 4 m (13.1 ft),
10 A
51704-052 52704-052 4 m (13.1 ft),
10 A
51704-053 52704-053 4 m (13.1 ft),
10 A
51704-054 52704-054 4 m (13.1 ft),
10 A
51704-057 52704-057 5 m (16.4 ft),
10 A
51704-058 52704-058 5 m (16.4 ft),
10 A
51704-075 52704-075 5 m (16.4 ft),
10 A
Use the following Erbe power cord for the mains connection of the connected unit to the auxiliary outlet of the cart.
RA/3 I
512 J
DK3 K
BR/3 N
VII E+F
IL/3G H
I I
REF no. (supplied with
REF no. (spare part)
Length, amperage
fastening set)
51704-041 51704-041 0.48 m (1.6 ft),
10 A
Power plug Erbe designation
C14W C14
Power plug international designation
80180-054_V24892
2023-04
42 / 50
Page 43
11 Spare parts for SystemCarrier compact
40180-410 (1)
40180-400 (1)
40180-405 (1)
20180-460 (2)
30180-081 (1)
Chapter 11

Spare parts for SystemCarrier compact

80180-054_V24892
2023-04
40180-413 (2)
40180-430 (1)
20180-464 (1)
43 / 50
Page 44
11 Spare parts for SystemCarrier compact
40180-450 (1)
20180-463 (1)
20180-460 (1)
40180-440 (4)
40180-420 (1)
30180-080 (4)
30180-081 (1)
40180-461 (1)
80180-054_V24892
2023-04
44 / 50
Page 45
12 Spare parts for SystemCarrier compact plus
40180-430 (1)
40180-418 (1)
40180-415 (1)
20180-463 (1)
20180-460 (1)
40180-505 (1)
40180-405 (1)
40180-400 (1)
40180-120 (1)
Chapter 12

Spare parts for SystemCarrier compact plus

80180-054_V24892
2023-04
45 / 50
Page 46
12 Spare parts for SystemCarrier compact plus
40180-450 (1)
20180-463 (1)
20180-460 (1)
40180-440 (4)
40180-420 (1)
30180-080 (4)
30180-081 (1)
40180-461 (1)
80180-054_V24892
2023-04
46 / 50
Page 47
Chapter 13

Symbols

Individual details of the symbols in this chapter may deviate from your product. Not all symbols may necessarily appear on your unit or its packaging.
Standard Ref. no. Symbol Title Explanation
13 Symbols
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and
80180-054_V24892
2023-04
information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
5.4.4 Caution Before switching the unit on or perform­ing another action related to the unit, read the safety instructions in the User Manual.
5.4.3 Consult instruc-
tions for use or consult electronic instructions for use.
5.1.6 Catalogue num-
ber
5.1.7 Serial number Indicates the manufacturer’s serial num-
5.1.1 Manufacturer Indicates the medical device manufactu-
5.1.3 Date of manufac-
ture
Indicates the need for the user to consult the instructions for use.
Indicates the manufacturer’s catalogue number so that the medical device can be identified.
ber so that a specific medical device can be identified.
rer.
Indicates the date when the medical device was manufactured.
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
5.3.2 Keep away from
sunlight
5.3.4 Keep dry Indicates a medical device that needs to
5.3.7 Temperature limit Indicates the temperature limits to which
Indicates a medical device that needs protection from light sources.
be protected from moisture.
the medical device can be safely exposed.
47 / 50
Page 48
13 Symbols
Standard Ref. no. Symbol Title Explanation
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 15223-1 Medical devices – Symbols to be used with medi­cal device labels, labelling and information to be supplied
ISO 7010 Graphical symbols ­Safety colours and safety signs
- Registered safety signs
ISO 7010 Graphical symbols ­Safety colours and safety signs
- Registered safety signs
ISO 7010 Graphical symbols ­Safety colours and safety signs
- Registered safety signs
IEC 60417 Graphical Symbols for Use on Equipment
IEC 60417 Graphical Symbols for Use on Equipment
5.3.8 Humidity limita­tion
5.3.9 Atmospheric pressure limita­tion
Indicates the range of humidity to which the medical device can be safely exposed.
Indicates the range of atmospheric pres­sure to which the medical device can be safely exposed.
5.7.7 Medical device Indicates the item is a medical device.
M002 Refer to instruc-
tion manual/
To signify that the instruction manual/ booklet must be read.
booklet
W012 Warning; Elec-
To warn of electricity.
tricity
General warning
To signify a general warning.
sign
5021 Equipotentiality To identify the terminals which, when
connected together, bring the various parts of an equipment or of a system to the same potential, not necessarily being the earth (ground) potential, e.g. for local bonding.
5140 Non-ionizing
electromagnetic radiation
To indicate generally elevated, potentially hazardous, levels of non-ionizing radia­tion, or to indicate equipment or systems e.g.
in the medical electrical area that include RF transmitters or that intention­ally apply RF electromagnetic energy for diagnosis or treatment.
80180-054_V24892
2023-04
IEC 60417 Graphical Symbols for Use on Equipment
IEC 60417 Graphical Symbols for Use on Equipment
IEC 60417 Graphical Symbols for Use on Equipment
48 / 50
5336 Defibrillation-
proof type CF applied part
To identify a defibrillation-proof type CF applied part complying with IEC 60601-1.
Note 1 – C = Cardial. Note 2 – F = Floating applied part.
5988 Computer net-
work
To identify the computer network itself or to indicate the connecting terminals of the computer network.
5034 Input To identify an input terminal when it is
necessary to distinguish between inputs and outputs.
Page 49
Standard Ref. no. Symbol Title Explanation
CE
X
ECB
13 Symbols
IEC 60601-2-2 Medical electri­cal equipment - Part 2-2: Par­ticular requirements for the
safety and essential per-
basic formance of high frequency surgical equipment and high frequency surgical accessories
IEC TR 60878 Graphical sym­bols for electrical equipment in medical practice
IEC TR 60878 Graphical sym­bols for electrical equipment in medical practice
MDD 93/42/EEC, Annex XII European confor-
EN 50419 Marking of electrical and electronic equipment in accordance with Article 11(2) of Directive 2002/96/EC (WEEE)
HF isolated
patient circuit
1853, 5114
5269, 5268
Separate disposal The product must be disposed of sepa-
Foot switch To identifiy a foot switch or the connec-
„OUT“/"IN" posi­tion of a bi-sta­ble push control
mity marking
The danger of leakage currents and therefore the danger of burns is substan­tially reduced for the patient.
tion for a foot switch.
To associate the "OUT" position of a bi-stable push control with the
corr
esponding function.
To associate the "IN" position of a bi-stable push control with the cor-
responding function.
Confirmation from the manufacturer that the product meets the requirements of the applicable European guidelines.
rately.
Quantity (x) Quantity of the product. x pieces of the
product are supplied in one packaging unit.
80180-054_V24892
2023-04
Erbe Communi-
cation Bus
Air inlet Indicates an air inlet.
Air outlet Indicates an air outlet.
Indicates an ECB connection. ECB is a sys­tem for transferring data between Erbe units.
49 / 50
Page 50
13 Symbols
80180-054_V24892
2023-04
50 / 50
Page 51
Kontakt, Contact, Contacto,
联系我们
Manufacturer
Erbe Elektromedizin GmbH Waldhörnlestraße 17 72072 Tübingen, Germany erbe-med.com
Contacts worldwide
België/Belgique/Belgien
Erbe Belgium Bvba Horizon Park – Gebouw 7 Leuvensesteenweg 510 1930 Zaventem, België Tel +32 2 254 88 60 Fax +32 2 254 88 69 info@erbe-belgium.com
中国
爱尔博 (上海)医疗器械有限公司 上海市长宁区延安西路 2201 号 3002 室 电话:+86 21 62758 440 传真:+86 21 62758 874 E-Mail: info@erbechina.com
Service Hotline
Tel +49 7071 755 437 Fax +49 7071 755 189 techservice@erbe-med.com
Italia
Erbe Italia S.r.l. Viale Sarca 336/f 20126 Milano, Italia Tel +39 02 647468 1 Fax +39 02 647468 30 info@erbe-italia.com
LATAM
Erbe do Brasil Ltda. Av. Andrômeda, 885 – Of. 1019 Alphaville Empresarial 06473-000 – Barueri – SP, Brasil Tel +55 11 4191 4146 info@erbe-brasil.com
Россия
ООО«ЭрбеЭлектромедицин» ул.Хамовническийвал,12 119270,Москва,Россия Телефон:+74952560052 Факс:+74952560052 info@erbe‐russia.com
Schweiz/Suisse/Svizzera
Erbe Swiss AG Fröschenweidstrasse 10 8404 Winterthur, Schweiz Tel +41 52 233 37 27 Fax +41 52 233 33 01 info@erbe-swiss.ch
Deutschland
Erbe Elektromedizin GmbH Waldhörnlestraße 17 72072 Tübingen, Germany Tel +49 7071 755 400 Fax +49 7071 755 188 Vertrieb-Deutschland@erbe-med.com
España
Erbe España Soluciones Médicas, S.L. C. Sta. Leonor 65, Edificio B Bajo Derecha 28037 Madrid, España Tel +34 91 954 38 61 info@erbe-spain.com
France
Erbe Médical SARL ZAC Sans Souci 470 Allée des Hêtres 69760 Limonest, France Tél +33 4 78 64 92 55 Fax +33 4 78 66 16 43 erbe@erbe-france.com
80103-161_V24354 2022-03
India
Erbe Medical India Private Ltd. Plot No - 447, Block - B, Sector-19, Dwarka, New Delhi - 110075, India Tel +91 11 4100 9808 erbe@erbe-india.com
Middle-East/Africa
Erbe Elektromedizin GmbH Representative Office, MEA Kaslik Étoile Center, 5th Floor Kaslik roundabout, Zouk Mikael Beirut, Lebanon Tel +961 9 211 277 Fax +961 9 213 713 info@erbe-lebanon.com
Nederland
Erbe Nederland B.V. Velsenstraat 10 4251 LJ Werkendam, Nederland Tel +31 183 509 755 klantenservice@erbe-nederland.com
Österreich
Erbe Elektromedizin Ges.m.b.H. Modecenterstraße 14 / Top B1 1030 Wien, Österreich Tel +43 1 893 24 46 Fax +43 1 893 24 46 3 info@erbe-austria.com
Polska
Erbe Polska Sp. z o.o. Al. Rzeczypospolitej 14 lok. 2.8 02-972 Warszawa, Polska Tel +48 22 642 25 26 Faks +48 22 642 88 99 sales@erbe.pl
South-East Asia
Erbe Singapore Pte. Ltd. 25 International Business Park #02-61/64 German Centre Singapore 609916, Singapore Tel +65 65 6283 45 Fax +65 65 6283 46 info@erbe-singapore.com
대한민국
에르베메디칼코리아 ( 유 ), 서울시 송파구 올림픽로 35 다길 42, 4 층 ( 신천동 , 한국루터회관 ) 05510 Tel +82 2 6475 1883 Fax +82 2 6475 1884 info@erbe-korea.com
United Kingdom
Erbe Medical UK Ltd The Antler Complex 1A Bruntcliffe Way, Morley Leeds LS27 0JG, United Kingdom Tel +44 113 253 0333 Fax +44 113 253 2733 sales@erbe-uk.com
USA
Erbe USA, Inc. 2225 Northwest Parkway Marietta, GA 30067, USA Tel +1 770 955 4400 Fax +1 770 955 2577 Customer Support: +1 800 778 3723 info@erbe-usa.com
Page 52
Loading...