Erbe ICC-BP Service guideline

Page 1
Title: Erbe ICC Bipolar Date: 2-10-2013 By: Erbe File = Erbe-ICC-BP.doc
1. General Applies to model “ICC BIPOLAR” Specifications: See Appendix
2. Reference Documents Erbotom Operating Instructions. Erbotom Service Manual (includes models ICC 50 and 80)
3. Tools / Fixtures / Accessories
Electrosurgical Analyzer: ex: Dempsey 443 or better. Bipolar Cable/Forceps (optional). Safety Tester
4. Performance Verifications
4.1. Initial Inspection
Check unit for any damage; i.e. connectors, power cord, the footswitch and its cable.
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND WITH POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE. NEITHER PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE.
4.2. With no accessories plugged in, turn on unit. Verify that
all segments of the displays briefly illuminate, along with several test tones.
4.3. Setup
Connect analyzer to ESU using a banana plug test leads. Before activating the unit, check that the up and down buttons can control each display between 0 (---) and 50 without noticeable contact bounce.
Note: The analyzer that is shown has load settings of only 500 and 125 ohms. Refer to load curves in Appendix for reference.
4.4. Cut Mode
Set the analyzer to 500 ohm load. When pressing the cut pedal, output power should easily track the cut display from 1 to 50 for either EFF1 or EFF2 selection.
.
4.5. Coag Mode
Set analyzer load to 500 ohms; Coag mode to EFF2. Using the footswitch, key the unit and note the distinct audible tone. Gradually turn power to max dial and note power into analyzer. Refer to charts in appendix for acceptable results, which depend on load.
4.6. Other Safety Checks
Using the safety tester, check that line leakage with ground open is less than 50 uA. Using the safety tester, check that ground lead resistance is less than 0.15 ohms. Resistance between case and either bipolar output contact: >10 megohms.
DEMPSEY 443 ANALYZER
Page 2
APPENDIX
CUT and COAD LOAD CURVES
Dial/Display =25
10
12
14
16
18
20
22
24
26
28
0 100 200 300 400 500 600 700
LOAD (Ohms)
Output Power (Watts)
CUT
COAG
OUTPUT VS DISPLAY
0
10
20
30
40
50
60
0 10 20 30 40 50
DISPLAY
OUTPUT (watts)
CUT (150 ohms)
COAG EFF 1 (150 ohms)
COAG EFF 2 (150 ohms)
COAG EFF 1 (500 ohms)
COAG EFF 2 (500 ohms)

Erbe ICC Bipolar Specifications

POWER OUTPUT RELATIONSHIPS
Bipolar Cut, coagulate (both with 2 effect
Max. energy output ( at rated ohms) Up to 50 Watt Max voltage, vp-p 380 effect 1; 1,600 effect 2 WEIGHT, kg (lb) 2.8 (6.2) GENERATOR Solid-state Frequency, Khz 350 Visual Yes Audible 4 tones Volume control Yes Line Voltage 110/120/230/240
COOLING SYSTEM Convection OTHER SPECIFICATIONS Low constant voltage; automatic
modes)
power adjustment; defined bipolar sine wave. Meets requirements of AAMI, HF 18, CSA, EN 46001, IEC 60601-1 and 60601-2-2, ISO 9001,SEV, and UL.