Erbe ICC-BP Service guideline

Title:

Erbe ICC Bipolar

Date: 2-10-2013

By:

Erbe

DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND WITH

File = Erbe-ICC-BP.doc

POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE. NEITHER

PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE.

 

 

1.General

Applies to model “ICC BIPOLAR” Specifications: See Appendix

2.Reference Documents

Erbotom Operating Instructions.

Erbotom Service Manual (includes models ICC 50 and 80)

3.Tools / Fixtures / Accessories

Electrosurgical Analyzer: ex: Dempsey 443 or better.

Bipolar Cable/Forceps (optional).

Safety Tester

4.Performance Verifications

4.1.Initial Inspection

Check unit for any damage; i.e. connectors, power cord, the footswitch and its cable.

4.2.With no accessories plugged in, turn on unit. Verify that all segments of the displays briefly illuminate, along with several test tones.

4.3.Setup

Connect analyzer to ESU using a banana plug test leads. Before activating the unit, check that the up and down buttons can control each display between 0 (---) and 50 without noticeable contact bounce.

Note: The analyzer that is shown has load settings of only 500 and 125 ohms. Refer to load curves in Appendix for reference.

DEMPSEY 443 ANALYZER

4.4.Cut Mode

Set the analyzer to 500 ohm load. When pressing the cut pedal, output power should easily track the cut display from 1 to 50 for either EFF1 or EFF2 selection.

.

4.5.Coag Mode

Set analyzer load to 500 ohms; Coag mode to EFF2. Using the footswitch, key the unit and note the distinct audible tone. Gradually turn power to max dial and note power into analyzer. Refer to charts in appendix for acceptable results, which depend on load.

4.6.Other Safety Checks

Using the safety tester, check that line leakage with ground open is less than 50 uA. Using the safety tester, check that ground lead resistance is less than 0.15 ohms. Resistance between case and either bipolar output contact: >10 megohms.

Erbe ICC-BP Service guideline

APPENDIX

POWER OUTPUT RELATIONSHIPS

 

28

 

 

 

 

 

 

 

60

 

26

 

 

 

 

 

 

 

 

 

24

 

 

 

 

 

 

 

50

Output Power (Watts)

 

 

CUT and COAD LOAD CURVES

 

22

 

 

 

 

 

 

 

 

 

OUTPUT (watts)

40

 

 

 

 

Dial/Display =25

20

 

 

 

 

 

 

 

 

 

 

 

18

 

CUT

 

 

 

 

30

 

 

 

 

 

 

 

 

 

 

 

 

 

16

 

COAG

 

 

 

 

20

 

 

 

 

 

 

 

 

14

 

 

 

 

 

 

 

 

 

12

 

 

 

 

 

 

 

10

 

10

 

 

 

 

 

 

 

 

 

0

100

200

300

400

500

600

700

0

LOAD (Ohms)

0

 

OUTPUT VS DISPLAY

CUT (150 ohms)

COAG EFF1 (150 ohms)

COAG EFF2 (150 ohms)

COAG EFF1 (500 ohms)

COAG EFF2 (500 ohms)

10

20

30

40

50

DISPLAY

Erbe ICC Bipolar Specifications

Bipolar

Cut, coagulate (both with 2 effect

 

modes)

Max. energy output ( at rated ohms)

Up to 50 Watt

Max voltage, vp-p

380 effect 1; 1,600 effect 2

WEIGHT, kg (lb)

2.8 (6.2)

GENERATOR

Solid-state

Frequency, Khz

350

Visual

Yes

Audible

4 tones

Volume control

Yes

Line Voltage

110/120/230/240

COOLING SYSTEM

Convection

OTHER SPECIFICATIONS

Low constant voltage; automatic

 

power adjustment; defined bipolar

 

sine wave. Meets requirements of

 

AAMI, HF 18, CSA, EN 46001, IEC

 

60601-1 and 60601-2-2, ISO

 

9001,SEV, and UL.