Eppendorf 5418 R, 5418 User Manual

Page 1
5401 900.022-00/0909
Centrifuge 5418/5418 R
Operating manual
Page 2
Copyright
©
2009 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of
the copyright owner.
eppendorf® is a registered trademark of Eppendorf AG, Hamburg, Germany. Microtainer® is a registered trademark of Becton Dickinson, Franklin Lakes, NJ, USA.
Trademarks are not marked in all cases with ™ or
®
in this manual.
5401 900.022-02/0710
Page 3
You will find a detailed description of these figures in your language in Chapters 2.1 and 4.1/5.1.
1
3 4
2
1
6
7 8 9
2
5
1
3
4
10
2
1
5
6
9
7
10
8
2
Abb. 1: Front and rear view of the Ce ntrifuge 5418

Fig. 1: Front and rear view of the Centrifuge 5418

Abb. 2: Front and rear view of the Ce ntrifuge 5418 R

Fig. 2: Front and rear view of the Centrifuge 5418 R

1 Centrifuge lid 2 Monitoring glass
3 Display 4 Control panel
5 Emergency lid release 6 ID plate
7 Mains switch 8 Mains connection
9 Fuse holder 10 Condensation water tray
Page 4
Abb. 3: Control panel of the Centr ifuge 5418 and Ce ntrifuge 5418 R.

Fig. 3: Control panel of the Centrifuge 5418 and Centrifuge 5418 R.

1 Switch the displayed centrifuging speed (rpm/rcf) 2 Display
3 Release lid 4 Adjust the centrifuging duration
Only 5418 R: Press the time dial to start and stop
centrifugation.
5 Set the speed of centrifugation
6 Adjust the temperature (only 5418 R)
Only 5418 R: Press the speed dial to switch the
displayed centrifugation speed (rpm/rcf).
7 Start the temperature control run fast temp
8 Short spin centrifugation
(only 5418 R)
9 Start and stop centrifugation
Abb. 4: Display of the Centrifug e 5418 and Centrifug e 5418 R.

Fig. 4: Display of the Centrifuge 5418 and Centrifuge 5418 R.

1 Centrifuging duration 2 g-force (rcf) or speed (rpm)
3 Status of the centrifuge 4 At set rpm (only 5418 R)
5 Temperature (only 5418 R) 6 Status of the key lock (only 5418 R)
7 Status of the centrifugation speed display 8 Status of the loudspeaker (only 5418 R)
Page 5
Table of contents
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur

Operating manual

Table of contents
1 User instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operating manual
1.1 Using this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Warning signs and hazard icons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Hazard symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Degrees of danger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Abbreviations used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Product description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Main illustration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Delivery package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Centrifuge 5418 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Centrifuge 5418 R. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 Rotor FA-45-18-11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4.1 Rcf display and calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Application limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Declaration concerning the ATEX directive (94/9/EC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.2 Maximum service life for accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.4 Information on product liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Warnings for intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5.1 Personal injury or damage to the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5.2 Incorrect handling of the centrifuge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.3 Incorrect handling of the rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.4 Extreme strain on the centrifuging tubes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.5.5 Aerosol-tight centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.6 Safety instructions on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1 Selecting the location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Installing the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Overview of operating controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.2 Preparing for centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.2.1 Switching on the centrifuge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.2.2 Inserting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.2.3 Loading the rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.2.4 Closing the rotor lid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.2.5 Closing the centrifuge lid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.3 Cooling (only 5418 R). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3.1 Temperature adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3.2 Temperature display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3.3 Temperature monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3.4 Fast Temp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3.5 Continuous cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.4 Centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.4.1 Centrifuging with time preset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.4.2 Centrifuging in continuous operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.4.3 Short spin centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.4.4 Removing the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.5 Standby mode (only 5418) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.6 Special functions (only 5418 R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
EN
Operating manual
5
Page 6
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
Table of contents
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.1 Prepare cleaning / disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.2 Perform cleaning / disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.2.1 Clean/disinfect device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.2.2 Clean/disinfect rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.3 Additional service instructions for Centrifuge 5418 R. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.4 Glass breakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.5 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.6 Decontamination before shipping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.2 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.3 Emergency lid release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.3.1 Centrifuge 5418 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.3.2 Centrifuge 5418 R. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8 Transport, storage and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.2 Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
9 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9.1 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9.3 Weight / dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9.4 Application parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
10 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.1 Centrifuge 5418 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.2 Centrifuge 5418 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.3 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.4 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.4.1 Fuses for Centrifuge 5418. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
10.4.2 Fuses for Centrifuge 5418 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6
Page 7
Centrifuge 5418 / 5418 R — Operating manual
1 User instructions

1 User instructions

1.1 Using this manual

1.2 Warning signs and hazard icons

1.2.1 Hazard symbols
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
Please read this operating manual completely before using the device for the first time.
  Please view this operating manual as part of the product and keep it somewhere easily
accessible.
When passing the device on to third parties, be sure to include this operating manual.If this manual is lost, please request another one. The current version can be found on our
website www.eppendorf.com.
Biological hazard Danger of explosion
finale Korrektur
EN
Operating manual
Danger of electric shock Danger of crushing
General danger Material damage
1.2.2 Degrees of danger
The degree of danger is a part of a safety note and distinguishes the possible results of non-observance from each other.
DANGER Will lead to severe injuries or death. WARNING Can lead to severe injuries or death. CAUTION Can lead to light to moderate injuries. NOTICE Can lead to material damages.

1.3 Symbols used

Depiction Meaning
You are requested to perform an action.
1.
2.
Perform these actions in the sequence described.
List.
Text Terms and key names from the software.
References useful information.

1.4 Abbreviations used

PCR Polymerase chain reaction
rcf Relative centrifugal force
rpm Revolutions per minute
UV Ultraviolet radiation
7
Page 8
1
3 4
2
1
6
7 8 9
2
5
1
3
4
10
2
1
5
6
9
7
10
8
2
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
2 Product description

2 Product description

2.1 Main illustration

Abb. 1: Front and rear view of the Ce ntrifuge 5418
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
The depiction of the Centrifuge 5418 / 5418 R can also be found on the front fold-out page (see Fig. 1 + 2).
finale Korrektur
Fig. 1: Front and rear view of the Centrifuge 5418
Abb. 2: Front and rear view of the Ce ntrifuge 5418 R
Fig. 2: Front and rear view of the Centrifuge 5418 R
1 Centrifuge lid 2 Monitoring glass
Visual control for rotor stop or option for speed control via stroboscope.
3Display
Depiction of the current centrifuging parameters and
4 Control panel
For operating the centrifuge (see Fig. 3 on p. 17).
device settings (see Fig. 4 on p. 17).
5 Emergency lid release
(see Emergency lid release on p. 29)
7 Mains switch
Switch for switching the device on and off. Switch position 0: The device is switched off. Switch position I: The device is switched on.
9 Fuse holder 10 Condensation water tray (only 5418 R)
6ID plate
8 Mains connection
Connection socket for the mains cable supplied.
8
Page 9
Centrifuge 5418 / 5418 R — Operating manual
2 Product description

2.2 Delivery package

finale Korrektur
EN
2.2.1 Centrifuge 5418
Quantity Order No.
(International)
1- -
1 or
1 5416 301.001 022634305
1 5703 350.102 022639609 1- - Mains cable
1 1
2.2.2 Centrifuge 5418 R
Quantity Order No.
1- -
1 or
1 5416 301.001 022634305 1 5401 850.076 5401850076 Tray for condensation water 1- - Mains cable
1 1
5425 351.003 5425 353.006
5401 900.022 5401 900.030
(International)
5425 351.003 5426 355.100
5401 900.022 5401 900.030
Order No. (North America)
022668188 022668226
5401900022 5401900030
Order No. (North America)
022668188 022668200
5401900022 5401900030
Description
Centrifuge 5418
See the Ordering Information chapter for corresponding device version, equipment and order number
Fuses
2.5 AT (230 V), 2 pieces 5 AT UL (100 V/120 V), 2 pieces
Rotor key
Standard
Captain Eppi rotor key holder
1 piece
Operating manual Centrifuge 5418/5418 R
Languages: EN, DE, FR, ES, IT, PT Languages: DA, FI, EL, NL, SV (230 V devices only)
Description
Centrifuge 5418 R
See the Ordering Information chapter for corresponding device version, equipment and order number
Fuses
2.5 AT (230 V), 2 pieces
6.25 AT (100 V/120 V), 2 pieces
Rotor key
Standard
Operating manual Centrifuge 5418/5418 R
Languages: EN, DE, FR, ES, IT, PT Languages: DA, FI, EL, NL, SV (230 V devices only)
Operating manual

2.3 Features

The compact and easy-to-use Centrifuge 5418 / 5418 R has a capacity of 18 x 2 mL and reaches a maximum of 16,873 x g / 14,000 rpm. The microcentrifuge is equipped with an aerosol-tight standard rotor for centrifugation of the following tubes:
Micro test tubes (0.2 to 2.0 mL)
Microtainers (0.6 mL)
Spin columns (1.5/2.0 mL)
The Centrifuge 5418 R has an additional temperature control function for centrifugation between 0°C and +40°C. The fast temp function can be used to start a temperature control run without samples to adjust the rotor chamber incl. rotor and adapters quickly to the set target temperature.
9
Page 10
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
2 Product description

2.4 Rotor FA-45-18-11

Before use of sample tubes, please note the manufacturer's specifications with regard to centrifugation resistance (max. rcf).
Operating manual
Rotor FA-45-18-11
Max. capacity Max. g-force (rcf)
or speed (rpm) without adapter
Acceleration / deceleration
(2)
time
18 micro test tubes of 1.5/2.0 mL each or spin columns.
With adapters:
0.2 mL PCR tubes
0.4 mL/0.5 mL micro test tubes
16,873 x g / 14,000 rpm
5418: 16 s / 18 s 5418 R: 13 s / 13 s
0.6 mL Microtainers
(1) Maximum load per rotor bore for adapter + tube + content. (2) According to DIN 58 970 (device version: 230 V, 50 to 60 Hz). (3) Aerosol tightness tested and certified by the Centre for Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK) (see certificate at the end of this operating manual).
Max. load per rotor
(1)
bore
3.75 g
Notes
Aerosol-tight
(aluminum).
(3)
rotor lid
2.4.1 Rcf display and calculation
Use the rpm/rcf key to switch the display of centrifugation speed between rcf and rpm. Only 5418 R: For speeds 800 rpm only the lowest adjustable g-force (100 x g) is displayed upon
switching. The exact g-force (rcf) can be determined using the formula given below. Ensure that the g-force (rcf) displayed upon switching is standardized to suit the rotor without an
adapter. When adapters are used, you can achieve the following maximum g-forces (rcf) at maximum speed (rpm):
Adapters Max. centrifugation
without adapter 7.7 16,873 for 0.2 mL PCR tubes 5.6 12,271 for 0.4 mL micro test tubes 7.7 16,873 for 0.5 mL micro test tubes 6.6 14,462 for 0.6 mL Microtainers 7.7 16,873
To determine the g-force (rcf) for a specific adapter, you can calculate per DIN 58 970 using the following formula:
rcf = 1.118 · 10 n: revolutions per minute (rpm) r
: max. centrifuging radius in cm
max
Example
The 0.2 mL adapter has a maximum radius of 5.6 cm. At 5,000 rpm a maximum g-force of 1,565 x g is reached.
-5
· n2 · r
max
radius r
max
Max. g-force (rcf)
[cm]
10
Page 11
3 Safety
3Safety

3.1 Intended use

CAUTION!

3.2 User profile

Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
The Centrifuge 5418 / 5418 R is intended exclusively for indoor use and for separating aqueous solutions and suspensions of various densities in approved test tubes.
Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may impair the safety, function and precision of the device. Eppendorf cannot be held liable or accept any liability for damage resulting from the use of incorrect or non-recommended accessories and spare parts or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
This device may only be operated by correspondingly trained specialist staff. This staff must have carefully read the operating manual and be familiar with the function of the device.
EN
Operating manual

3.3 Application limits

3.3.1 Declaration concerning the ATEX directive (94/9/EC)
Danger of explosion!
Do not operate the device in areas where work is completed with explosive substances.Do not use this device to process any explosive, radioactive or highly reactive substances.
DANGER!
3.3.2 Maximum service life for accessories
Do not use this device to process any substances which could create an explosive
atmosphere.
Due to its design and the environmental conditions on the inside of the device, the Centrifuge 5418 / 5418 R is not suitable for use in a potentially explosive atmosphere.
The device must therefore only be used in a safe environment, such as in the open environment of a ventilated laboratory or an extractor hood. The use of substances which may contribute to a potentially explosive atmosphere is not permitted. The final decision with regard to the risks connected with the use of such substances is the responsibility of the user.
Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can result in serious internal material damage.
Protect all parts from mechanical damage.
CAUTION!
Check accessories regularly.Do not use rotors or buckets with signs of corrosion or mechanical damage (e.g.
deformations).
Do not use accessories whose maximum service life has been exceeded.
Accessory Maximum service life from first commissioning
Rotor lid 3 years
Requirement: The "click" when turning the rotor lid screw is still audible. Plastic adapters 1 year Outer sealing ring of the aerosol-tight rotor 50 autoclave cycles
For the rotor described here there is no limit for its service life, as long as the following conditions are met: proper use, recommended maintenance and undamaged condition.
The date of manufacture is stamped on the rotors in the format 03/07 (= March 2007). This is for information only and does not have any reference to the service life.
11
Page 12
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
3 Safety

3.4 Information on product liability

In the following cases, the protection provided by the device may be impaired. The liability for the function of the device passes to the operator if:
The device is not used in accordance with the operating manual.
The device is used outside of the range of application described in the preceding chapters.
The device is used with accessories or consumables (e.g., tubes) which are not
recommended by Eppendorf.
The device is maintained or repaired by persons not authorized by Eppendorf.
The owner has made unauthorized modifications to the device.
Operating manual

3.5 Warnings for intended use

Read the operating manual first and observe the following general safety instructions before using the Centrifuge 5418 / 5418 R.
3.5.1 Personal injury or damage to the equipment
WARNING!
WARNING!
WARNING!
WARNING!
Electric shock from damage to device or mains cable.
Only switch on the device if the device and the mains cable are undamaged.Only use devices that have been properly installed or repaired.
Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside the device
can be contacted by accident.
Do not remove the housing of the device.Do not allow any liquids to enter the inside of the housing.Do not allow the device to be opened by anyone except service personnel who have been
specifically authorized by Eppendorf.
Risk from incorrect supply voltage
Only connect the device to voltage supplies which correspond with the electrical requirements
on the nameplate.
Damage to health due to handling infectious liquids and pathogenic germs.
Observe the national regulations for handling these substances, the biological security level
of your laboratory, the material safety data sheets and the manufacturer's application notes.
For the centrifugation of these substances, use suitable aerosol-tight closure systems.When working with pathogenic germs belonging to a higher risk group, more than one
aerosol-tight bioseal must be used.
Wear your personal protective equipment (PPE).Follow the instructions regarding hygiene, cleaning and decontamination.For complete instructions regarding the handling of germs or biological material of risk group
II or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health Organization, current edition of the Laboratory Biosafety Manual).
12
WARNING!
Centrifuge lid can crush. Keep hands clear.
When opening or closing the device lid, do not reach between the lid and device or into the
latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
Page 13
3 Safety
NOTICE!
NOTICE!
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
Damage to device by spilling liquids in the rotor or rotor chamber
1. Switch the device off.
2. Disconnect the device from the power supply.
3. Clean the device and the accessories carefully in accordance with the cleaning and disinfection instructions in the operating manual.
4. If a different cleaning and disinfecting method is to be used, contact Eppendorf to ensure that the intended method will not damage the device.
Damage to electronic components through formation of condensation.
After the device has been moved from a cool to a warmer environment, condensation can form inside the device.
Wait at least three hours before connecting it to the power supply.Only 5418: Alternative: let the device run for half an hour just before a short transport.
EN
Operating manual
3.5.2 Incorrect handling of the centrifuge
Damage from knocking against or moving the device during operation.
A rotor banging against the rotor chamber wall can cause considerable damage to the device and rotor.
NOTICE!
3.5.3 Incorrect handling of the rotors
Do not move or knock against the device during operation.
Risk of injury from improperly attached rotors and rotor lids.
Centrifuge only with the rotor and rotor lid firmly tightened.If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
CAUTION!
properly secured. Stop centrifugation immediately by pressing the start/stop key.
Risk of injury from asymmetric loading of rotors.
Load rotors symmetrically with identical tubes.Only load adapters with suitable tubes.
CAUTION!
Always use tubes of the same type (weight, material/density and volume).Check symmetric loading by balancing the adapters and tubes used with scales.
Risk of injury from overloaded rotor.
The Centrifuge 5418 / 5418 R is designed for the centrifugation of material with a max. density of
1.2 g/mL at maximum speed and volume and/or load.
CAUTION!
Observe the information on each rotor relating to maximum load (adapter, tube and contents)
per rotor bore and make sure it is not exceeded.
NOTICE!
Damage to rotors from aggressive chemicals.
Rotors are high-quality components which withstand extreme stresses. This stability can be impaired by aggressive chemicals.
Avoid the use of aggressive chemicals, including strong and weak alkali, strong acids,
solutions with mercury, copper and other heavy metal ions, halogenated hydrocarbons, concentrated saline solutions and phenol.
If the rotor is contaminated by aggressive chemicals, clean it immediately using a neutral
cleaning agent. This applies to the rotor bores in particular.
13
Page 14
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
3 Safety
3.5.4 Extreme strain on the centrifuging tubes
Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.
CAUTION!
NOTICE!
NOTICE!
Only use tubes which are approved by the manufacturer for the required rcf.
Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the accessories and loss of the samples.
Before use, carry out a visual check of all tubes for any damage.
Risk from open tube lids.
Open tube lids can break off during centrifugation and cause damage to the rotor as well as to the centrifuge.
Carefully close all tube lids before centrifuging.
finale Korrektur
Hazard to plastic tubes from organic solvents.
When using organic solvents (e.g. phenol, chloroform) the density of plastic tubes is reduced, i.e., the tubes could get damaged.
NOTICE!
NOTICE!
Follow the manufacturer's information about the chemical resistance of tubes.
Sample tubes heat up.
In uncooled centrifuges the temperature in the rotor chamber, rotor and sample can rise to above 40°C dependent on the run time, g-force (rcf) or speed (rpm) and ambient temperature.
Note that this can reduce the centrifugation resistance of the sample tubes.Please note the temperature resistance of the samples.
3.5.5 Aerosol-tight centrifugation
Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this purpose are used. These are always indicated by the prefix FA.
WARNING!
Always use rotors and rotor lids marked aerosol-tight together for aerosol-tight centrifugation.Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor lid.
Risk to health due to limited aerosol tightness in the event of incorrect use.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive solvents can impair the aerosol-tightness of the rotors and rotor lid.
WARNING!
Regularly check that the seals of aerosol-tight rotor lids are undamaged.Only use aerosol-tight rotor lids with undamaged and clean gaskets.Thinly brush the threads of the rotor lid screw with pivot grease (order no. Int.: 5810 350.050 /
North America: 022634330) after every proper autoclaving (121°C, 20 min.). Do not apply the pivot grease to the gaskets.
Replace the outer sealing ring of the rotor after 50 autoclave cycles.Aerosol-tight rotors should never be stored with rotor lids screwed on.
14
Page 15
Centrifuge 5418 / 5418 R — Operating manual
3 Safety

3.6 Safety instructions on the device

finale Korrektur
EN
Depiction Meaning Location
WARNING
General hazard point. Follow the operating manual.
CAUTION
Always tighten up the rotor using the rotor key supplied.
CAUTION
Close all tubes and use a rotor lid.
5418: Rear of the device 5418 R: Right side of the device
Top of device, below the centrifuge lid.
Top of device, below the centrifuge lid.
Operating manual
15
Page 16
EN
Operating manual
4 Installation
4Installation

4.1 Selecting the location

NOTICE!
NOTICE!
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
If a fault occurs, objects in the immediate vicinity of the devices could get damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety distance of
30 cm clear around the device during operation.
Damage from overheating.
Do not place the device close to sources of heat (e.g., radiator, drying cabinet).Do not expose the device to direct sunlight.Ensure free circulation of air by maintaining a distance of at least 30 cm on all sides of the
device from adjacent devices or the wall and keep the underside of the device clear.
Make sure that the vents in the device are always free of obstruction.
Select the location for the device according to the following criteria:
Suitable power connection as per the name plate (230 V/120 V/100 V).
Stable, horizontal and resonance-free lab bench. Weight of the device: 7.7 kg (5418) or 22 kg
(5418 R).
A well ventilated environment which is protected from direct sunlight to prevent the device
from heating up more.
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur

4.2 Installing the instrument

Centrifuge 5418 R: Compressor damage after improper transport.
Only switch on the centrifuge 4 hours after installation.
NOTICE!
Perform the following steps in the sequence described.
1. Place the device on a suitable lab bench.
2. Allow the device to warm up for at least 3 hours (5418) or 4 hours (5418 R) to the ambient temperature to prevent damage to the electronic components from condensation and damage to the compressor (only 5418 R).
3. Check that the mains voltage and frequency match the requirements on the device type plate.
4. Connect the centrifuge to the mains and switch it on using the mains power switch on the right side of the device.
The display is active.
Lid opens automatically
5. Use the details included in the scope of delivery to check that the delivery is complete.
6. Check all parts for any transport damage. Contact your dealer if any damage is found.
7. Only 5418 R: Insert the condensation water tray at the side of the device into the holder provided (see Fig. 2 on p. 8).
Retain the packaging material and the transport protection device for subsequent transport or
Hint!
storage. See also the instructions relating to transport (see p. 30).
16
Page 17
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
5 Operation
5Operation

5.1 Overview of operating controls

Abb. 3: Control panel of the Centr ifuge 5418 and Ce ntrifuge 5418 R.
Fig. 3: Control panel of the Centrifuge 5418 and Centrifuge 5418 R.
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
Before using the Centrifuge 5418 / 5418 R for the first time, familiarize yourself with the operating controls and the display.
EN
Operating manual
1 Switch the displayed centrifuging speed (rpm/rcf)
2Display
(see Rcf display and calculation on p. 10)
3 Release lid 4 Adjust the centrifuging duration
Only 5418 R: Press the time dial to start and stop
centrifugation.
5 Set the speed of centrifugation
6 Adjust the temperature (only 5418 R)
Only 5418 R: Press the speed dial to switch the
displayed centrifugation speed (rpm/rcf).
7 Start the temperature control run fast temp
(only 5418 R)
8 Short spin centrifugation
(see Short spin centrifugation on p. 23)
9 Start and stop centrifugation
Abb. 4: Display of the Centrifug e 5418 and Centrifug e 5418 R.
Fig. 4: Display of the Centrifuge 5418 and Centrifuge 5418 R.
1 Centrifuging duration 2 g-force (rcf) or speed (rpm)
5418 R: Set value x 1000
3 Status of the centrifuge
: Centrifuge lid unlocked. : Centrifuge lid locked. (Flashing): centrifuging in progress.
4 At set rpm (only 5418 R)
: Start of run time when reaching 95% of the preset
g-force (rcf) or speed (rpm).
: Start of run time immediately.
5 Temperature (only 5418 R) 6 Status of the key lock (only 5418 R)
: Centrifuging parameters cannot be modified
unintentionally.
7 Status of the centrifugation speed display
rcf: g-force (relative centrifugal force) rpm: revolutions per minute
: No key lock.
8 Status of the alarm (only 5418 R)
: Switched on.
No symbol: switched off.
17
Page 18
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
5 Operation

5.2 Preparing for centrifugation

5.2.1 Switching on the centrifuge
1. Switch the centrifuge on, using the mains switch. After switching on at the mains power switch, the centrifuge lid opens automatically.
2. Open the closed centrifuge lid by pressing the open key. The parameter settings of the last run are displayed.
5.2.2 Inserting the rotor
If handled incorrectly, the rotor can fall over.
The rotor lid screw may become loose if it is used to retain the rotor.
NOTICE!
Always grasp the rotor with both hands for holding or transport.
finale Korrektur
1. Place rotor vertically onto the motor shaft.
2. Insert the rotor key supplied into the rotor nut.
3. Turn rotor key clockwise until the rotor nut is firmly tightened.
5.2.3 Loading the rotor
Risk of injury from asymmetric loading of rotors.
Load rotors symmetrically with identical tubes.
CAUTION!
Only load adapters with suitable tubes.Always use tubes of the same type (weight, material/density and volume).Check symmetric loading by balancing the adapters and tubes used with scales.
Risk from damaged or overloaded tubes!
When loading the rotor, note the safety instructions with regard to hazards from overloaded or
CAUTION!
To load the rotor, proceed as follows:
1. Check the maximum load (adapter, tube and contents) per rotor bore.
2. Load rotor and adapters only with the tubes intended for them.
3. Insert tubes opposite each other in pairs into the rotor bores. For symmetrical loading, tubes
damaged tubes (see Warnings for intended use on p. 12)
The maximum load is 3.75 g per rotor bore. This information can also be found on the rotor.
that are opposite each other must be of the same type and contain the same filling quantity.
18
In order to minimize weight differences between filled sample tubes, we recommend taring with a scale. This will reduce wear on the drive and cut running noise.
Page 19
5 Operation
5.2.4 Closing the rotor lid
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
With the rotor FA-45-18-11 centrifugation is also possible without a rotor lid.
Hint!
Please also note:
Operating manual
The tube lids must be closed securely.
Without the rotor lid, the rotor is not aerosol-tight.
The centrifugation is slightly louder.
Spin columns must always be centrifuged with a rotor lid.
1. Check that the outer sealing ring is fitted properly in the groove.
2. Fit the rotor lid vertically on the rotor.
3. Lock the rotor by turning the red rotor lid screw clockwise beyond an audible click until it can be turned no further.
The rotor is not properly closed until an audible click is heard!
5.2.5 Closing the centrifuge lid
Centrifuge lid can crush. Keep hands clear.
When opening or closing the device lid, do not reach between the lid and device or into the
latching mechanism of the lid.
WARNING!
Always open the centrifuge lid completely to prevent it from falling.
1. Check correct attachment of rotor and rotor lid.
2. Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.
The centrifuge will close automatically. The display shows the symbol . Only 5418: The open key lights blue.
19
Page 20
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
5 Operation

5.3 Cooling (only 5418 R)

5.3.1 Temperature adjustment
Set the temperature using the temp arrow keys from 0°C to +40°C.
5.3.2 Temperature display
If the rotor is stopped: Set temperature During centrifugation: Actual temperature
Operating manual
5.3.3 Temperature monitoring
After the set temperature has been reached the centrifuge responds as follows to temperature fluctuation during centrifugation:
Deviation from the target value Action
ΔT > 3°C Temperature display flashes. ΔT > 5°C Periodic warning sound and visual indication
Error 18. Centrifugation is stopped automatically.
5.3.4 Fast Temp
This function can be used to start a temperature control run directly without samples with a rotor and temperature-specific speed in order to quickly adjust the rotor chamber, incl. rotor and adapters, up to the set target temperature.
Requirement
The centrifuge is switched on.
The rotor and rotor lid are properly attached.
The centrifuge lid is closed.
The temperature and g-force (rcf) or speed (rpm) are set for the subsequent centrifugation
(see Centrifugation on p. 21).
1. Press the fast temp key. The display shows FT as well as the current temperature and g-force (rcf) or speed (rpm).
The temperature control run ends automatically when the target temperature has been reached. A periodic signal tone sounds.
The cooling time from room temperature (~ 23 °C) to 4°C takes approx. 16 min.
2. Press the start/stop key to terminate the temperature control run early.
After the set temperature has been reached and the temperature control run is complete, the centrifuge keeps the rotor chamber, with the centrifuge lid closed, at the set temperature, if the temperature is below the ambient temperature. Irrespective of the target temperature, however, this continuous cooling does not go below 4°C to prevent the rotor chamber from freezing.
20
The centrifuge stops the run automatically when the rotor has reached the set temperature.
Hint!
Therefore, there may be a delay between the display of the achieved set temperature and the automatic end of the temperature control run.
Page 21
5 Operation
5.3.5 Continuous cooling
When the rotor is stopped the rotor chamber is kept at the target temperature when the following prerequisites are met:
The centrifuge is switched on.
The centrifuge lid is closed.
The target temperature is below the ambient temperature.
During continuous cooling the following applies:
The target temperature is displayed.
Irrespective of the target temperature, continuous cooling does not go below 4°C to prevent
Because the rotor does not rotate during this process the temperature adjustment is slower.
To end continuous cooling, open the centrifuge lid. If the centrifuge is not used for more than 8 hours, the continuous cooling is switched off
automatically. This protects against ice formation in the rotor chamber and in the tubes as well as increased condensation in the device.
The display shows the set temperature of the rotor chamber. With fast temp you can quickly reach the desired temperature again (see p. 20).
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
Operating manual
the rotor chamber and the samples from freezing and increased condensation in the device.

5.4 Centrifugation

Risk from incorrectly-loaded rotors and damaged/overloaded tubes!
Before commencing centrifugation, follow the safety instructions relating to hazards from
asymmetrically loaded and/or overloaded rotors and from overloaded, damaged and/or open
CAUTION!
tubes (see Warnings for intended use on p. 12).
Risk of injury from improperly attached rotors and rotor lids.
Centrifuge only with the rotor and rotor lid firmly tightened.
CAUTION!
5.4.1 Centrifuging with time preset
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Stop centrifugation immediately by pressing the start/stop key.
Before using the Centrifuge 5418 / 5418 R for the first time, familiarize yourself with the operating controls and the display (see Overview of operating controls on p. 17).
Each of the centrifuging variants described here must be preceded by the preparation described above (see Preparing for centrifugation on p. 18).
Only 5418 R: Please also note the instructions on cooling (see p. 20).
Perform the following steps in the sequence described.
1. Use time to set the run time.
2. Only 5418 R: Use temp to set the temperature.
3. Use speed to set the g-force (rcf) or speed (rpm).
4. Press start/stop to start centrifuging.
21
Page 22
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
5 Operation
During centrifugation
In the display flashes while the rotor is running.
The remaining run time is displayed in minutes, in 30s increments below ten minutes. The
last minute is counted down in seconds.
Only 5418 R: The actual temperature is displayed.
The current g-force (rcf) or rotor speed (rpm) is displayed.
The fast temp (only 5418 R), open and short keys are blocked during centrifugation.
During the run you can modify the total run time, the temperature (only 5418 R), the speed
and the rpm/rcf display. Only 5418 R: To change the centrifugation parameters, press the short key first.
The values flash in the display during the change. The new parameters are adopted immediately. When the time is changed during a run, the time which has already elapsed is taken into account. Note that the shortest new total run time which can be set is the time which has already elapsed plus 2 minutes.
You can also terminate the centrifugation before the set run time has elapsed by pressing
the start/stop key.
End of centrifugation
After completion of the set time, the centrifuge stops automatically. During braking the
elapsed centrifugation time is displayed flashing. When the rotor stops a signal tone is sounded.
Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You
can open it by pressing the open key.
5. Remove centrifuge contents.
5.4.2 Centrifuging in continuous operation
Perform the following steps in the sequence described.
1. Use time to set the continuous run. The continuous run function can be set above 9:59 h or under 30 s (5418) or above 99 min or
under 0.5 min (5418 R). The timer shows oo to indicate continuous operation.
2. Only 5418 R: Use the temp arrow keys to adjust the temperature.
3. Use speed to set the g-force (rcf) or speed (rpm).
4. Press start/stop to start centrifuging. In the display flashes while the rotor is running.
Time is counted upwards, first in 30-second increments and then in minute increments from ten minutes.
5. Press start/stop to end centrifuging after the desired time period.
During the braking process, centrifuging duration flashes in the display.
When the rotor stops a signal tone is sounded.
Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You
can open it by pressing the open key.
6. Remove centrifuge contents.
22
Page 23
5 Operation
5.4.3 Short spin centrifugation
5418: You can carry out a short spin run with the maximum g-force (rcf) or speed (rpm). 5418 R: You can carry out a short spin run with the currently set, or maximum, g-force (rcf) or
speed (rpm). Set this short spin mode as described in the following section.
Selecting short spin mode (only 5418 R)
1. Press the short key while the centrifuge lid is open. The current mode is displayed:
Display 1 – 14 t (run at preselected speed)
Display 14 t (run at maximum speed of 14,000 rpm)
2. When the centrifuge lid is open, press the short key for longer than 2 s to switch between these modes.
Performing short spin centrifugation
1. Only 5418 R: A short run at preselected g-force (rcf) or speed (rpm) can be set directly using the speed dial.
2. Only 5418 R: Use the temp arrow keys to adjust the temperature.
3. Start short spin run: Hold down the short key.
In the display flashes while the rotor is running.
The time is counted upwards in seconds.
During short run centrifuging all other keys are blocked.
4. End short spin run: Release the short key.
During the braking process, centrifuging duration flashes in the display.
Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You
can open it by pressing the open key.
5. Remove centrifuge contents.
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
Operating manual
During the braking process, centrifuging can be restarted up to two more times by pressing the
Hint!
5.4.4 Removing the rotor
short key again.
1. Turn the rotor nut counterclockwise using the rotor key supplied.
2. Remove rotor by lifting vertically.
3. Only 5418 R: Switch off the centrifuge after use and empty the condensation water tray (pull out from the left or right side of the device). Leave centrifuge lid fully opened and protect it against closing.

5.5 Standby mode (only 5418)

If the centrifuge has not been used for 15 min, it switches to standby mode. The EP logo then appears in the display. When a button or knob is used or the centrifuge lid is closed, the centrifuge is reactivated and ready for operation.

5.6 Special functions (only 5418 R)

Function Status of centrifuge lid Press > 2 s key Display
Modify parameter during the cycle. closed short Flashes 5 s Enable/disable signal tone.
Enable/disable key lock. At set rpm (Fig. 4 on p. 17)
open
open open
open
short + open time
/
/
23
Page 24
EN
Centrifuge 5418 / 5418 R — Operating manual
6 Maintenance

6 Maintenance

6.1 Prepare cleaning / disinfection

Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
Clean at least once a week and clean if the accessible surfaces of the device and its accessories are acutely contaminated.
Clean the rotor regularly. This way the rotor is protected and the durability is prolonged. Furthermore, observe the notes on decontamination (see Decontamination before shipping on
p. 26) when the device is sent to the authorized Technical Service for repairs. The procedure described in the following chapter applies for the cleaning as well as for the
disinfection or decontamination. The additional required steps are described in the following table:
finale Korrektur
Operating manual
Cleaning Disinfecting / decontamination
1. For cleaning the accessible surfaces of the device and the accessories use a mild cleaning fluid.
2. Carry out the cleaning as described in the following chapter.
If you have any further questions regarding the cleaning and disinfection or decontamination or
Hint!
regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The contact details are provided on the back of this manual.

6.2 Perform cleaning / disinfection

Electric shock as a result of penetration of liquid.
Switch off the device and disconnect it from the power supply before starting cleaning or
disinfecting.
DANGER!
Do not allow any liquids to enter the inside of the housing.Do not complete a spray disinfection on the housing.Only reconnect the device to the power supply once it is completely dry.
1. Choose the disinfection method which corresponds to the legal regulations and guidelines in place for your range of application. For example, use alcohol (ethanol, isopropanol) or alcohol-based disinfectants.
2. Carry out the disinfection or decontamination as described in the following chapter.
3. Then clean the device and the accessories.
24
NOTICE!
NOTICE!
NOTICE!
Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it immediately with a
mild cleaning agent.
Corrosion from aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.Do not incubate the accessories in aggressive cleaning agents or disinfectants for prolonged
periods.
Damage from UV and other high-energy radiation.
Do not use UV, beta, or gamma radiation or any other high-energy radiation source for
disinfecting.
Page 25
6 Maintenance
Autoclave
Hint!
Hint!
Rotor, rotor lid and adapter can be autoclaved (121°C, 20 min). After a maximum of 50 autoclave cycles, replace the sealing ring in the lid groove of the
aerosol-tight rotor lid.
Aerosol-tightness
Check that the seals are intact before use. Replace the sealing ring in the lid groove of the aerosol-tight rotor lid when worn. Replace the
rotor lids of aerosol-tight rotors when the sealing ring on the lid screw is worn. The sealing rings require regular care to protect the rotors.
Aerosol-tight rotors should never be stored with lids screwed on! In order to prevent damage, lightly lubricate the lid thread of the aerosol-tight rotors with pivot
grease (order no. Int.: 5810 350.050 / North America: 022634330) on a regular basis.
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
Operating manual
6.2.1 Clean/disinfect device
1. Switch off the device from the mains power switch while the lid is opened and remove the mains power switch from the power supply.
2. Unscrew the rotor nut by turning it counterclockwise with the rotor key.
3. Remove rotor.
4. Use the agents specified above (see p. 24) to clean and disinfect the device and the rotor chamber.
5. Wipe all accessible surfaces of the device and the accessories, including the power cable, with a moist cloth.
6. Thoroughly clean the rubber seals of the rotor chamber with water.
7. Rub the dry rubber seals with glycerine or talcum powder to prevent them from becoming brittle. Other components of the device, such as the lid latch, must not be lubricated.
8. Clean the motor shaft with a soft, dry and lint-free cloth.
9. Check device and accessories for corrosion and damage.
6.2.2 Clean/disinfect rotor
1. Clean rotor, rotor lid and adapters with the aforementioned agents and disinfect them (see p. 24).
2. Use a bottle brush to clean and disinfect the rotor bores.
3. In order to clean and disinfect the rotor lid, remove the sealing ring. Clean the groove and the sealing ring below it as well.
4. Rinse rotor, rotor lid and adapters thoroughly with water. Pay particular attention to the rotor bores of the fixed-angle rotors.
5. Place rotor and accessories onto a cloth to dry. Place fixed angle rotors with the rotor bores facing downwards to allow the bores to dry as well.
6. Correctly reinsert the sealing ring of the rotor lid in the clean and dry groove.
7. Clean the rotor cone with a soft, dry and lint-free cloth and inspect for damage.
8. Place the dry rotor onto the motor shaft.
9. Tighten the rotor nut firmly by turning it clockwise with the rotor key.
25
Page 26
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
6 Maintenance

6.3 Additional service instructions for Centrifuge 5418 R

Empty and clean the condensation water tray regularly and especially after liquid spillage in
the rotor chamber. Pull out the condensation water tray from the left or right side of the centrifuge.
Clear the rotor chamber regularly of ice formations by thawing, either by leaving the centrifuge
lid open or carrying out a brief temperature control run at approx. 30 °C.
Wipe up condensate in the rotor chamber. To do so, use a soft absorbent cloth.Remove dust deposits from the ventilation slits of the centrifuge using a brush or swab at the
latest every six months. First switch off the device and remove the power plug.
Operating manual

6.4 Glass breakage

Note when using glass tubes that the danger of breakage of glass increases with an increasing
Hint!
g-force (rcf) or speed (rpm). Please note the manufacturer's information on the recommended centrifugation parameters (loading and speed).
Broken glass scratches the surfaces of the rotor chamber and the accessories (rotor, rotor lid and adapter) so that their chemical resistance is reduced. Therefore, a fine, black metal abrasion develops in the rotor chamber due to the air turbulence which, in addition to causing damage to the rotor chamber and the accessories, can also contaminate the samples.
Carefully remove all splinters and glass powder from the rotor chamber and accessories
If required, replace the adapters to prevent any further damage.Check rotor bores regularly for residues or damage.

6.5 Fuses

The fuse holder is located under the mains power socket (5418) or on the left next to the mains switch (5418 R).
1. Disconnect the mains plug.
2. Remove the fuse holder.
(rotor, rotor lid and adapter) when breakage of glass occurs.
Both fuses are now accessible and can be replaced.

6.6 Decontamination before shipping

If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following:
Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. It is available in PDF format on our
WARNING!
26
homepage (www.eppendorf.com/decontamination)
2. Decontaminate all the parts you want to dispatch.
3. Enclose the fully-completed decontamination certificate for returned goods (including the serial number of the device) with the dispatch.
.
Page 27
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
7 Troubleshooting

7 Troubleshooting

7.1 General errors

Symptom / message Cause Remedy
No display. No mains connection. Check the mains power connection. No display. Power failure.
Lid of the device cannot be opened.
Lid of the device cannot be opened.
Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
If the suggested measures fail repeatedly, please contact Technical Service. You can find the contact addresses at the end of this operating manual or on the Internet under www.eppendorf.com
.
Check the mains fuse for the device
(see Fuses on p. 26).
Check the mains fuse for the laboratory.
Rotor is still running.
Power failure. 1. Check the mains fuse for the device
Wait for the rotor to stop.
(see Fuses on p. 26).
2. Check the mains fuse for the laboratory.
3. Activate the emergency lid release (see p. 29).
EN
Operating manual
Device cannot be started. Lid of the device is not closed. Device shakes when it
starts up.
Centrifuge brakes during a short run centrifugation, although the short key is pressed.
Temperature display flashes.
(only 5418 R)
Rotor is unsymmetrically loaded. 1. Stop the device and load symmetrically.
The short key was released briefly more than twice (protective function for the drive).
Temperature variation from nominal value: ΔT > 3°C.
Close the lid of the device.
2. Restart the device.
Press the short key continuously during a
short run centrifugation.
Check the settings.Check unhindered air circulation through
the vents.
Thaw ice or switch off device and allow it
to cool down.

7.2 Error messages

If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.
Symptom / message Cause Remedy LID ERROR (5418) /
LID (5418 R) LID ERROR (5418) /
LID (5418 R)
Centrifuge lid cannot be locked.
Centrifuge lid cannot be released. 1. Switch the device off and back on.
Try to close the centrifuge lid again.
2. Press the open key.
If the error occurs again:
LID ERROR (5418) / LID (5418 R)
Centrifuge lid must not be released during a run.
1. Switch off the centrifuge.
2. Activate the emergency lid release (see Emergency lid release on p. 29).
Wait for the rotor to stop.
27
Page 28
EN
Operating manual
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
7 Troubleshooting
Symptom / message Cause Remedy INT Mains power failure during a run. Check the mains connection. NO RPM (5418) /
Error 3 (5418 R) Error 5
(only 5418 R)
Err 6 (5418) / Error 6 (5418 R)
Error in speed measuring system or drive overheated.
Prohibited opening of lid or lid switch is defective during a run.
Drive error.
Leave the device switched on until the
error message disappears (10 s or 6 min).
1. Wait for the rotor to stop.
2. Open and close again the lid of the device.
3. Repeat the run.
Repeat the run.If this error message appears again,
switch centrifuge off and back on again after > 20 s.
Error 6
(only 5418 R)
Err 7 (5418) / Error 7 (5418 R)
Err 8 (5418) / Error 8 (5418 R)
Err 9 to 17 (5418) / Error 9 to 17 (5418 R)
Error 18
(only 5418 R)
Error 19
(only 5418 R)
Error 20
(only 5418 R)
Error 21
(only 5418 R)
Error 24
(only 5418 R)
Drive overheated.
Major deviation in the speed control. 1. Wait for the rotor to stop.
Drive error. 1. Wait for the rotor to stop.
Electronics error.
Too high temperature variation from nominal value in the rotor chamber.
Allow the drive to cool down for at least 15
min.
2. Tighten the rotor.
2. Repeat the run.
Switch the centrifuge off and back on
again after > 20 s.
Check the settings.Check unhindered air circulation through
the vents.
Thaw ice or switch off device and allow it
to cool down.
Cooling circuit is overheated.
Temperature sensor in rotor chamber faulty.
Temperature sensor at capacitor faulty.
Cooling unit fault, e.g., overheating.
Check unhindered air circulation through
the vents and allow device to cool down.
Switch the centrifuge off and back on
again after > 20 s.
Switch the centrifuge off and back on
again after > 20 s.
Allow the centrifuge to cool down and
repeat the run.
28
Page 29
7 Troubleshooting

7.3 Emergency lid release

If the centrifuge lid cannot be opened, you can activate the emergency lid release manually.
Risk of injury from rotating rotor.
Wait for the rotor to stop before activating the emergency lid release.
WARNING!
7.3.1 Centrifuge 5418
To check, look through the inspection glass in the centrifuge lid.
1. Disconnect the mains plug.
2. Loosen the plastic cover for the emergency lid release. This is located behind the front left device foot in the base plate (Fig. 1 on p. 8).
3. Pull the cord vertically downwards. The centrifuge lid is unlocked.
4. Before closing the centrifuge lid again: push the cord completely into the housing and insert the plastic cover in the base plate.
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
Operating manual
7.3.2 Centrifuge 5418 R
1. Disconnect the mains plug.
2. Rotate the plastic cover of the emergency lid release 45° counterclockwise with a suitable
3. Enter the centrifuge rotor key into the hexagonal opening at rear until some resistance can be
4. Slightly press and turn the rotor key clockwise.
5. Open the centrifuge lid.
6. Remove the rotor key and put the plastic cover back on rotating it 45° clockwise.
tool (e.g. screwdriver) and remove it. The plastic cover is located on the right side of the device (Fig. 2 on p. 8).
felt.
This will release the centrifuge lid.
29
Page 30
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
8 Transport, storage and disposal

8 Transport, storage and disp osal

8.1 Transport

Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
Only transport the device in the original packaging.
  Use a transport aid for transporting over longer distances.
Air temperature Rel. humidity Air pressure
General transportation -25 to 60 °C 10 to 75% 30 to 106 kPa Air freight -20 to 55 °C 10 to 75% 30 to 106 kPa
Operating manual

8.2 Storage

8.3 Disposal

Air temperature Rel. humidity Air pressure
In transport packaging -25 to 55 °C 10 to 75% 70 to 106 kPa Without transport
packaging
In case the product is to be disposed of, the relevant legal regulations must be observed.
Information on the disposal of electrical and electronic devices in the European Community:
The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2002/96/EC pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005 on a business-to-business basis, to which this product is assigned, may no longer be disposed of in municipal or household waste. To document this they have been marked with the following identification:
-5 to 45 °C 10 to 75% 70 to 106 kPa
30
Because disposal regulations may differ from one country to another within the EU please contact your supplier if necessary.
Page 31
9 Technical data

9 Technical data

9.1 Power supply

Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
Tab. 1: Centrifuge 5418
Mains power connection 230 V, 50 to 60 Hz
Current consumption: 1.4 A (230 V)
Power consumption: max. 170 W EMC: Interference emission (radio
interference) EMC: Noise immunity EN 61326 - performance characteristic B Overvoltage category: II Fuses: 2.5 AT (230 V)
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
EN
Operating manual
120 V, 50 to 60 Hz 100 V, 50 to 60 Hz
2.8 A (120 V)
3.4 A (100 V)
EN 61326 - category B
5.0 AT (120 V / 100 V)
Tab. 2: Centrifuge 5418 R
Mains power connection 230 V, 50 to 60 Hz
Current consumption: 1.4 A (230 V)
Power consumption: max. 320 W EMC: Interference emission (radio
interference) EMC: Noise immunity EN 61326 - performance characteristic B Overvoltage category: II Fuses: 2.5 AT (230 V)

9.2 Ambient conditions

Environment: For indoor use only. Ambient temperature: Centrifuge 5418: 2 to 40°C
Max. relative humidity: 75 %, non-condensing humidity.
120 V, 50 to 60 Hz 100 V, 50 to 60 Hz
2.8 A (120 V)
3.0 A (100 V)
EN 61326 - category B
6.25 AT (120 V / 100 V)
Centrifuge 5418 R: 15 to 35°C
Atmospheric pressure: Use up to an altitude of 2000 m above MSL. Degree of contamination: 2
31
Page 32
EN
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
9 Technical data

9.3 Weight / dimensions

Tab. 3: Centrifuge 5418
Dimensions: Width: 208 mm (8.19 in.)
Depth: 300 mm (11.8 in.)
Height: 210 mm (8.27 in.) Weight without rotor: 7.7 kg Noise level: < 51 dB(A) *
Operating manual
Tab. 4: Centrifuge 5418 R
Dimensions: Width: 298 mm (11.73 in.)
Weight without rotor: 22 kg Noise level: < 57 dB(A) *
*) The noise level was measured frontally in a sound measuring room with accuracy class 1 at a distance of 1 m from the device and at lab bench height.

9.4 Application parameters

Rotor Acceleration and deceleration times
FA-45-18-11 5418:
These values were calculated at 23°C.
Run time: (5418)
Depth: 463 mm (18.23 in.)
Height: 250 mm (9.84 in.)
according to DIN 58 970
5418 R: 230 V: 16 s / 18 s 120 V: 16 s / 18 s 100 V: 18 s / 18 s
30 s to 9:59 h, as well as infinity (oo), adjustable up to 10 min in 30 s increments,
thereafter in 1 min increments.
230 V: 13 s / 13 s
120 V: 13 s / 13 s
100 V: 13 s / 13 s
32
Run time: (5418 R)
Temperature: (only 5418 R)
Relative centrifugal force (RZB or rcf): 1 to 16,873 x g,
Speed: 100 to 14,000 rpm,
Max. load: 18 micro test tubes of 2.0 mL each. Max. kinetic energy: 2,600 Nm Test log mandatory: No Permitted density of the material for
centrifuging (at max. g-force (rcf) or speed (rpm) and max. load):
0.5 min to 99 min, as well as infinity (oo), adjustable up to 10 min. in 0.5 min.
increments, thereafter in 1 min. increments. 0°C to 40°C
adjustable in 100 x g increments.
adjustable in 100 rpm increments.
1.2 g/mL
Page 33
Centrifuge 5418 / 5418 R — Operating manual
10 Ordering information

10 Ordering information

10.1 Centrifuge 5418

Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
finale Korrektur
EN
Order No. (International)
5418 000.017
-
Order No. (North America)
022620321 022620304
Description
Centrifuge 5418
with rotor FA-45-18-11 incl. rotor lid 230 V / 50 - 60 Hz 120 V / 50 - 60 Hz, with US-plug

10.2 Centrifuge 5418 R

Order No. (International)
5401 000.013
-
Order No. (North America)
5401000013 5401000137
Description
Centrifuge 5418 R
with rotor FA-45-18-11 incl. rotor lid 230 V / 50 Hz 120 V / 50 - 60 Hz, with US-plug

10.3 Accessories

Order No. (International)
5418 707.005 022652061
5418 708.001 022652087
5418 709.008 022652109
5425 715.005 5425 717.008 5425 716.001
5416 301.001 022634305
5703 350.102 022639609 5401 850.076 5401850076 Tray for condensation water (only 5418 R)
*) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK).
Order No. (North America)
022636260 022636243 022636227
Description
Rotor FA-45-18-11
aerosol-tight*, angle 45°, 18 places, max. tube diameter 11 mm, incl. rotor lid (aluminum)
Rotor lid for FA-45-18-11
aerosol-tight*, aluminum
Seal for rotor lid FA-45-18-11
5 pieces
Adapter
used in FA-45-18-11 for 0.2 mL PCR tubes, set of 6 for 0.4 mL tubes, set of 6 for 0.5 mL tubes and 0.6 mL Microtainer, set of 6
Rotor key
Standard
Captain Eppi rotor key holder
1 piece
Operating manual

10.4 Fuses

10.4.1 Fuses for Centrifuge 5418 Order No.
(International)
5425 351.003 5425 353.006
10.4.2 Fuses for Centrifuge 5418 R Order No.
(International)
5425 351.003 5426 355.100
Order No. (North America)
022668188 022668226
Order No. (North America)
022668188 022668200
Description
Fuses
2.5 AT (230 V), 2 pieces 5 AT UL (100 V/120 V), 2 pieces
Description
Fuses
2.5 AT (230 V), 2 pieces
6.25 AT (100 V/120 V), 2 pieces
33
Page 34
Centrifuge 5418 / 5418 R — Operating manual
EG-Konformitätserklärung
EC Conformity Declaration
Das bezeichnete Produkt entspricht den einschlägigen grundlegenden Anforderungen der
aufgeführten EG-Richtlinien und Normen. Bei einer nicht mit uns abgestimmten Änderung des
Produktes oder einer nicht bestimmungsgemäßen Anwendung verliert diese Erklärung ihre Gültigkeit.
The product named below fulfills the relevant fundamental requirements of
the EC directives and standards listed. In the case of unauthorized modifications to the product
or an unintended use this declaration becomes invalid.
Produktbezeichnung, Product name:
Produkttyp, Product type:
Einschlägige EG-Richtlinien/Normen, Relevant EC directives/standards:
Vorstand, Board of Management: Projektmanagement, Project Management:
Hamburg, Date:
Eppendorf AG · Barkhausenweg 1 · 22339 Hamburg · Germany
0015 033.509-02
Laborzentrifuge / Laboratory Centrifuge
Centrifuge 5418 / 5418 R
einschließlich Zubehör / including accessories
2006/95/EG, EN 61010-1, EN 61010-2-20
2004/108/EG, EN 55011/B, EN 61000-6-1, EN 61000-3-2, EN 61000-3-3, EN 61326-1
98/79/EG, EN 14971, EN 61010-2-101, EN 61326-2-6
08.10.2009
5418 900.301-03
finale Korrektur
11 Declarations and Certificates

8 Declarations and Certificates

Operating manual — Centrifuge 5418 / 5418 Rseeon p.Fig.Tab.p.
125
Page 35
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
11 Declarations and Certificates
126
Page 36
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
11 Declarations and Certificates
127
Page 37
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
11 Declarations and Certificates
&HUWLILFDWH1XPEHU 5HSRUW5HIHUHQFH ,VVXH'DWH
2009 October 23

This is to certify that
representative samples of

Standard(s) for Safety:
Additional Information:
20091023-E215059 E215059, 2001 April 5
Issued to:
EPPENDORF A G
BARKHAUSENWEG 1 22339 HAMBURG GERMANY
LABORATORY USE ELECTRICAL EQUIPMENT
5415R, 5418R
Have been investigated by Underwriters Laboratories Inc.® (UL) or any authorized licensee of UL in accordance with the Standard(s) indicated on this Certificate.
UL 3101-1, First Edition UL 3101-2-20, First Edition CAN/CSA-C22.2 No. 1010.1-92 CAN/CSA-C22.2 No.1010.2.020-92, CAN/CSA-C22.2 No. 1010.2.020A-97
See UL On-line Certification Directory at WWW.UL.COM for additional information
3DJHRI
128
Only those products bearing the UL Listing Mark for the US and Canada should be considered as being covered by UL's Listing and Follow-Up Service meeting the appropriate requirements for US and Canada.
The UL Listing Mark for the US and Canada generally includes: the UL in a circle symbol with “C” and “US” identifiers:
the word “LISTED”; a control number (may be alphanumeric) assigned by UL; and the product category name (product
identifier) as indicated in the appropriate UL Directory.
Look for the UL Listing Mark on the product
Issued by: Reviewed by:
Thomas Huber, Engineer Walter Hofmair, Staff Engineer
Underwriters Laboratories Inc. Underwriters Laboratories Inc.
Page 38
Centrifuge 5418 / 5418 R — Operating manual
finale Korrektur
11 Declarations and Certificates
129
Page 39
Eppendorf offices
AUSTRALIA & NEW ZEALAND
Eppendorf South Pacific Pty. Ltd. Phone: +61 2 9889 5000 Fax: +61 2 9889 5111 E-mail: Info@eppendorf.com.au Internet: www.eppendorf.com.au
finale Korrektur
Eppendorf offices
AUSTRIA
Eppendorf Austria GmbH Phone: +43 (0) 1 890 13 64 - 0 Fax: +43 (0) 1 890 13 64 - 20 E-mail: office@eppendorf.at Internet: www.eppendorf.at
BRAZIL
Eppendorf do Brasil Ltda. Phone: +55 11 30 95 93 44 Fax: +55 11 30 95 93 40 E-mail: eppendorf@eppendorf.com.br Internet: www.eppendorf.com.br
CANADA
Eppendorf Canada Ltd. Phone: +1 905 826 5525 Fax: +1 905 826 5424 E-mail: canada@eppendorf.com Internet: www.eppendorfna.com
FRANCE
Eppendorf France S.A.R.L. Phone: +33 1 30 15 67 40 Fax: +33 1 30 15 67 45 E-mail: eppendorf@eppendorf.fr Internet: www.eppendorf.fr
ITALY
Eppendorf s.r.l. Phone: +390 2 55 404 1 Fax: +390 2 58 013 438 E-mail: eppendorf@eppendorf.it Internet: www.eppendorf.it
CHINA
Eppendorf China Ltd. Phone: +86 21 38560500 Fax: +86 21 38560555 E-mail: market.info@eppendorf.cn Internet: www.eppendorf.cn
GERMANY
Eppendorf Vertrieb Deutschland GmbH Phone: +49 2232 418-0 Fax: +49 2232 418-155 E-mail: vertrieb@eppendorf.de Internet: www.eppendorf.de
JAPAN
Eppendorf Co. Ltd. Phone: +81 3 5825 2363 Fax: +81 3 5825 2365 E-mail: info@eppendorf.jp Internet: www.eppendorf.jp
CZECH REP. & SLOVAKIA
Eppendorf Czech & Slovakia s.r.o. Phone: +420 323 605 454 Fax: +420 323 605 454 E-mail: eppendorf@eppendorf.cz Internet: www.eppendorf.cz / www.eppendorf.sk
INDIA
Eppendorf India Limited Phone: +91 44 42 11 13 14 Fax: +91 44 42 18 74 05 E-mail: info@eppendorf.co.in Internet: www.eppendorf.co.in
NORDIC
Eppendorf Nordic Aps Phone: +45 70 22 2970 Fax: +45 45 76 7370 E-mail: nordic@eppendorf.dk Internet: www.eppendorf.dk
SOUTH & SOUTHEAST ASIA
Eppendorf Asia Pacific Sdn. Bhd. Phone: +60 3 8023 2769 Fax: +60 3 8023 3720 E-mail: eppendorf@eppendorf.com.my Internet: www.eppendorf.com.my
UNITED KINGDOM
Eppendorf UK Limited Phone: +44 1223 200 440 Fax: +44 1223 200 441 E-mail: sales@eppendorf.co.uk Internet: www.eppendorf.co.uk
SPAIN
Eppendorf Ibérica S.L.U. Phone: +34 91 651 76 94 Fax: +34 91 651 81 44 E-mail: iberica@eppendorf.es Internet: www.eppendorf.es
USA
Eppendorf North America, Inc. Phone: +1 516 334 7500 Fax: +1 516 334 7506 E-mail: info@eppendorf.com Internet: www.eppendorfna.com
SWITZERLAND
Vaudaux-Eppendorf AG Phone: +41 61 482 1414 Fax: +41 61 482 1419 E-mail: vaudaux@vaudaux.ch Internet: www.eppendorf.ch
OTHER COUNTRIES
Internet: www.eppendorf.com/ worldwide
Page 40
finale Korrektur
Your local distributor: www.eppendorf.com/worldwide
Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556 · E-Mail: eppendorf@eppendorf.com
Eppendorf North America, Inc. · 102 Motor Parkway, Suite 410 · Hauppauge, NY 11788-5178 · USA
Tel: +1 516 334 7500 · Toll free phone: +1 800 645 3050 · Fax: +1 516 334 7506 · E-Mail: info@eppendorf.com
Application Support
Europe, International: Tel: +49 1803 666 789 · E-Mail: support@eppendorf.com
North America: Tel: +1 800 645 3050 ext. 2258 · E-Mail: techserv@eppendorf.com
Asia Pacific: Tel: +60 3 8023 6869 · E-Mail: support_asiapacific@eppendorf.com
Loading...