Read this operating manual completely before using the device for the first time. Please also note the
operating instructions for the accessories, if applicable.
This operating manual is part of the product. Thus, it must always be easily accessible.
Enclose this operating manual when transferring the device to third parties.
If this manual is lost, please request another one. You will find the current version on our webpage
www.eppendorf.com/worldwide.
1.2Danger symbols and danger levels
The safety instructions in this manual appear with the following danger symbols and danger levels:
5
1.2.1Danger symbols
BiohazardExplosion
Electric shockCrushing
Hazard pointMaterial damage
1.2.2Danger levels
DANGERWill lead to severe injuries or death.
WARNINGMay lead to severe injuries or death.
CAUTIONMay lead to light to moderate injuries.
NOTICEMay lead to material damage.
1.3Symbols used
Symbol Meaning
Handling
1.
2.
• List
Text Name of fields in the software
Actions in the specified order
Useful information
Page 6
Operating instructions
Centrifuge 5418 / 5418 R
6
English (EN)
1.4Abbreviations used
PCR
Polymerase chain reaction
RZB/rcf
Relative centrifugal force – g-force in m/s
rpm
Revolutions per minute – in rpm
UV
Ultraviolet radiation
2
Page 7
2Product description
2.1Main illustration
Abb. 2-1: Front and rear view of the Centrifuge 5418
1
2
3
4
5
Fig. 2-1:Front and rear view of the Centrifuge 5418
Abb. 2-2: Front and rear view of the Centrifuge 5418 R
Product description
Centrifuge 5418 / 5418 R
7
English (EN)
1
2
6
7
8
9
1
2
3
4
10
10
Fig. 2-2:Front and rear view of the Centrifuge 5418 R
1 Centrifuge lid
2 Monitoring glass
6 Name plate
7 Mains/power switch
Visual control for rotor stop or speed control
option using stroboscope
3 Display
Depiction of the current centrifuging parameters
8 Mains connection
and device settings.
1
2
5
6
9
7
8
Switch for switching the device on and off.
Switch position 0: The device is switched off.
Switch position I: The device is switched on.
Connection socket for the mains cable supplied.
4 Control panel
For operating the centrifuge.
5 Emergency lid release
9 Fuse holder
10 Condensation water tray (only Centrifuge 5418
R)
Page 8
Product description
Centrifuge 5418 / 5418 R
8
English (EN)
2.2Delivery package
2.2.1Centrifuge 5418
Quantity Order no.
(International)
1--See chapter Ordering Information for corresponding
15425 351.0030226681882.5 A T (230 V), set of 2
or5425 353.0060226682265 A T (120 V), UL, set of 2
15416 301.001022634305Standard
1--
15401 900.0225401900022Languages: EN, DE, FR, ES, IT, PT
15401 900.0305401900030Languages: DA, FI, EL, NL, SV (230 V devices only)
Order no.
(North America)
Description
Centrifuge 5418
device version, equipment and order number
Fuse
Rotor key
Power supply device
Operating manual Centrifuge 5418/5418 R
2.2.2Centrifuge 5418 R
Quantity Order no.
(International)
1--See chapter Ordering Information for corresponding
15425 351.0030226681882.5 A T (230 V), set of 2
or5426 355.1000226682006.25 AT (100 V/120 V), set of 2
15416 301.001022634305Standard
15401 850.0765401850076
1--
15401 900.0225401900022Languages: EN, DE, FR, ES, IT, PT
15401 900.0305401900030Languages: DA, FI, EL, NL, SV (230 V devices only)
Order no.
(North America)
Description
Centrifuge 5418 R
device version, equipment and order number
Fuse
Rotor key
Tray for condensation water
Power supply device
Operating manual Centrifuge 5418/5418 R
Page 9
Product description
Centrifuge 5418 / 5418 R
English (EN)
2.3Features
The compact and easy-to-use Centrifuge 5418 / 5418 R has a capacity of 18 x 2 ml and reaches a maximum
of 16,873 x g / 14,000 rpm. The microcentrifuge is equipped with an aerosol-tight standard rotor for
centrifugation of the following tubes:
• Micro test tubes (0.2 to 2.0 ml)
• Microtainers (0.6 ml)
• Spin columns (1.5/2.0 ml)
The Centrifuge 5418 R has an additional temperature control function for centrifugation between 0°C and
+40°C. The fast temp function can be used to start a temperature control run without samples to adjust the
rotor chamber incl. rotor and adapters quickly to the set target temperature.
2.4Rotor FA-45-18-11
Before using tubes, observe the manufacturer's recommended specifications on resistance to
centrifugation (max. g-force).
9
Max. capacityMax.
g-force (rcf)
or
Max. load
per rotor
(1)
bore
Notes
speed (rpm)
without
adapter
Acceleration/deceleration
(2)
time
Rotor
FA-45-18-11
18 micro test tubes of 1.5/
2.0 ml each or spin columns.
With adapters:
• 0.2 mL PCR tubes
16,873 x g /
3.75 g
14,000 rpm
5418: 16 s / 18 s
5418 R: 13 s / 13 s
• Aerosol-tight
rotor lid
(aluminum).
(3)
• 0.4 ml/0.5 ml micro test
tubes
• 0.6 mL Microtainers
(1) Maximum load per rotor bore for adapter + tube + contents.
(2) According to DIN 58 970 (device version: 230 V, 50 to 60 Hz).
(3) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health
Protection Agency, Porton Down (UK) (see certificates at the end of this operating manual).
Rotor code:
All Eppendorf® rotors are identified
using a simple, alphanumeric
format that represents the technical
specifications in a uniform series of
letters and numbers.
Fixed-angle
rotor
F
Aerosol-tight
version
Angle of
bores
Ø of
bores (mm)
A453011
Number of
bores
Swing-bucket
rotor
A
481
Number of
buckets
Ø of
buckets (mm)
Page 10
10
Product description
Centrifuge 5418 / 5418 R
English (EN)
2.4.1rcf display and calculation
Use the rpm/rcf key to switch the display of centrifugation speed between rcf and rpm.
Only 5418 R: For speeds ≤ 800 rpm only the lowest adjustable g-force (100 x g) is
displayed upon switching. The exact g-force (rcf) can be determined using the formula
given below.
Ensure that the g-force (rcf) displayed upon switching is standardized to suit the rotor
without an adapter. You can achieve the following maximum g-forces (rcf) at maximum
speed (rpm) when adapters are used:
AdapterMax. centrifugation
radius r
max
[cm]
Max. g-force (rcf)
Without adapter7.716,873
for 0.2 mL PCR tubes5.612,271
for 0.4 mL micro test tubes7.716,873
for 0.5 mL micro test tubes6.614,462
for 0.6 mL Microtainers7.716,873
To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970
with the following formula:
-5
rcf = 1.118 · 10
· n2 · r
max
n: Revolutions per minute (rpm)
r
: Max. centrifugation radius in cm
max
Example:
The 0.2 ml adapter has a maximum radius of 5.6 cm. At 5,000 rpm a maximum g-force of 1,565 x g is
reached.
Page 11
Safety
Centrifuge 5418 / 5418 R
English (EN)
3Safety
3.1Intended use
The Centrifuge 5418 / 5418 R is intended exclusively for indoor use and for separating aqueous solutions
and suspensions of various densities in approved test tubes.
3.2User profile
This device may only be operated by trained specialist staff. They must have carefully read the operating
manual and be familiar with the function of the device.
3.3Application limits
3.3.1Declaration concerning the ATEX directive (94/9/EC)
11
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device for processing any substances which could generate an explosive
atmosphere.
Due to its design and the environmental conditions inside the device, the Centrifuge 5418 / 5418 R is not
suitable for use in a potentially explosive atmosphere.
The device only must be used in a safe environment, such as the open environment of a ventilated
laboratory or fume hood. The use of substances which could create a potentially explosive atmosphere is
not permitted. The final decision on the risks associated with the use of these types of substances is the
responsibility of the user.
3.3.2Maximum service life for accessories
WARNING! Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can lead to serious internal material amage.
Protect all accessory parts from mechanical damage.
Inspect the accessories for damage before each use. Replace any damaged accessories.
Do not use rotors or rotor lids with signs of corrosion or mechanical damage (e.g.
deformations).
Do not use any accessories which are past their use-by date.
Page 12
12
Safety
Centrifuge 5418 / 5418 R
English (EN)
AccessoriesMaximum service from the first
commissioning onward
Rotor lid3 years
Requirement: The "click" is still
audible when tightening the rotor lid
screw.
Plastic adapters1 year
For the rotor described there is no limit on the service life as long as the following prerequisites are met:
• Proper use,
• Recommended maintenance
• Undamaged condition
The date of manufacture is stamped on the rotors in the format 03/10 (= March 2010). This is for
information only and does not have any reference to the service life.
To ensure aerosol tightness, the following applies:
• Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.
• Replace the seal of QuickLock rotor lids after 50 autoclaving cycles.
3.4Information on product liability
In the following cases, the designated protection of the device may be compromised. Liability for any
resulting property damage or personal injury is then transferred to the operator:
• The device is not used in accordance with the operating manual.
• The device is used outside of its intended use.
• The device is used with accessories or consumables which are not recommended by Eppendorf.
• The device is maintained or repaired by people not authorized by Eppendorf.
• The user makes unauthorized changes to the device.
3.5Warnings for intended use
Read the operating manual and observe the following general safety instructions before using the
Centrifuge 5418 / 5418 R.
3.5.1Personal injury or damage to the equipment
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.
Only use devices that have been properly installed or repaired.
In case of danger, disconnect the device from the mains supply by pulling the power plug
from the device or the mains socket or, by using the isolating device intended for this
purpose (e.g., emergency stop switch in the laboratory).
Page 13
Safety
Centrifuge 5418 / 5418 R
English (EN)
WARNING! Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside the device
can be contacted by accident.
Do not remove the housing of the device.
Do not allow any liquids to penetrate the inside of the housing.
Do not allow the device to be opened by anyone except service personnel who have been
specifically authorized by Eppendorf.
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources that match the electrical requirements listed on
the name plate.
Only use sockets with a protective earth (PE) conductor and suitable power cable.
WARNING! Damages to health due to infectious liquids and pathogenic germs.
When handling infectious liquids and pathogenic germs, observe the national regulations,
the biological security level of your laboratory, the material safety data sheets, and the
manufacturer's application notes.
Use aerosol tight sealing systems for the centrifugation of these substances.
When working with pathogenic germs belonging to a higher risk group, more than one
aerosol-tight bioseal must be used.
Wear personal protective equipment.
Consult the "Laboratory Biosafety Manual" (Source: World Health Organization,
Laboratory Biosafety Manual, as amended) for comprehensive regulations on the handling
of risk group II germs or biological materials).
13
WARNING! Crushing of the fingers with the centrifuge lid.
When opening or closing the device lid, do not reach between the lid and device or into the
latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or
accept any liability for damage resulting from the use of incorrect or non-recommended
accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
Page 14
14
Safety
Centrifuge 5418 / 5418 R
English (EN)
NOTICE! Damage to device due to spilled liquids.
1. Switch the device off.
2. Disconnect the device from the power supply.
3. Carefully clean the device and the accessories in accordance with the cleaning and
disinfection instructions in the operating manual.
4. If a different cleaning and disinfecting method is to be used, contact Eppendorf AG to
ensure that the intended method will not damage the device.
NOTICE! Damage to electronic components due to condensation.
Condensation can form inside the device after the device has been moved from a cool to a
warmer environment.
5418: Wait for at least 3 hours before connecting the device to the power supply.
Only 5418: Alternative: Let the device heat up for 30 minutes right before a brief transport.
5418 R: Wait for at least 4 hours before connecting the device to the power supply.
3.5.2Incorrect handling of the centrifuge
NOTICE! Damage from knocking against or moving the device during operation.
If the rotor bangs against the rotor chamber wall, it will cause considerable damage to the
device and rotor.
Do not move or knock against the device during operation.
3.5.3Incorrect handling of the rotors
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes.
Only load adapters with suitable tubes.
Always use the same type of tubes (weight, material/density and volume).
Check symmetric loading by balancing the adapters and tubes used with scales.
CAUTION! Risk of injury from overloaded rotor.
The Centrifuge 5418 / 5418 R is designed for the centrifugation of centrifugation material with
a max. density of 1.2 g/mL at maximum speed and volume.
Please note the information on each rotor on the maximum load (adapter, tube and
contents) per rotor bore and do not exceed it.
Page 15
NOTICE! Damage to rotors from aggressive chemicals.
Rotors are high-quality components which withstand extreme stresses. This stability can be
impaired by aggressive chemicals.
Avoid the use of aggressive chemicals, including strong and weak alkali, strong acids,
solutions with mercury, copper and other heavy metal ions, halogenated hydrocarbons,
concentrated saline solutions and phenol.
Due to the manufacturing process, color variations may occur on rotors marked "coated".
These color variations do not effect service life or resistance to chemicals.
3.5.4Extreme strain on the centrifuging tubes
CAUTION! Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.
Only use tubes which are approved by the manufacturer for the required rcf.
Safety
Centrifuge 5418 / 5418 R
English (EN)
15
NOTICE! Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the
accessories and loss of the samples.
Before use, visually check all of the tubes for damage.
NOTICE! Risk from open tube lids.
Open tube lids can break off during centrifugation and damage both the rotor and the
centrifuge.
Carefully seal all tube lids before centrifuging.
NOTICE! Hazard to plastic tubes from organic solvents.
The density of plastic tubes is reduced when organic solvents (e.g., phenol, chloroform) are
used, i.e. the tubes could become damaged.
Note the manufacturer’s information on the chemical resistance of the tubes.
NOTICE! Sample tubes heat up.
In uncooled centrifuges, the temperature in the rotor chamber, rotor and sample can increase
to above 40 °C, based on the run time, g-force (rcf)/speed and ambient temperature.
Note that this can reduce the centrifugation resistance of the sample tubes.
Please note the temperature resistance of the samples.
Page 16
16
Safety
Centrifuge 5418 / 5418 R
English (EN)
3.5.5Aerosol-tight centrifugation
WARNING! Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid
combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this
purpose are used. For fixed-angle rotors the labeling always begins with FA, swing-bucket
rotors are labeled with AT (aerosol tight).
The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red
ring on the rotor and a red rotor lid screw.
For aerosol-tight centrifugation, always simultaneously use rotors and rotor lids which are
marked as aerosol-tight in the centrifuge intended for the corresponding purpose. The
details specifying in which centrifuge you may use the aerosol-tight rotors and rotor lids
can be found on the rotor and, beginning from production date of October 2003, on the
upper side of the rotor lid.
Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor
lid.
WARNING! Health hazard from limited aerosol-tightness due to incorrect use.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive
solvents can impair the aerosol-tightness of the rotors and rotor lid.
Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.
Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.
Thinly brush the threads of the rotor lid screw with pivot grease (order no. Int.
5810 350.050, North America 022634330). Do not apply the pivot grease to the seals.
Replace aerosol-tight rotor lids and caps after 50autoclaving cycles.
For QuickLock rotor lids, the seal must be replaced after 50 autoclaving cycles.
Never store aerosol-tight rotors or buckets closed.
3.6Safety instructions located on the device
DisplayMeaningLocation
Follow the instructions in the operating manual. 5418: Rear of the device
CAUTION
Always tighten the rotor using the supplied rotor
key.
CAUTION
Close all tubes and use a rotor lid.
5418 R: Right side of the
device
Top of device, below the
centrifuge lid
Top of device, below the
centrifuge lid
Page 17
4Installation
4.1Selecting the location
NOTICE! If an error occurs, the objects in the immediate proximity of the device will be
damaged.
In accordance with recommendations in EN 61010-2-020, leave a safety clearance of 30
cm around the device during operation.
Please remove all materials and objects from this area.
NOTICE! Damage from overheating.
Do not install the device near to any heat sources (e.g., heating, drying cabinet).
Do not expose the device to direct sunlight.
Ensure unobstructed air circulation. Keep free a clearance of at least 33 cm around all
ventilation grilles.
Installation
Centrifuge 5418 / 5418 R
English (EN)
17
Select the location for the device according to the following criteria:
• Suitable power connection as per the name plate (230 V/120 V/100 V).
• Stable, horizontal and resonance-free lab bench. Weight of the device: 7.7 kg (5418) or 22 kg (5418 R).
• A well ventilated environment which is protected from direct sunlight to prevent the device from heating
up more.
4.2Preparing installation
Prerequisites
The weight of the Centrifuge 5418 is 7.7 kg or the Centrifuge 5418 R 22 kg. For unpacking and installing
the Centrifuge 5418 R, you require the assistance of another person.
Perform the following steps in the sequence described.
1. Open the packaging board.
2. Remove the accessories.
3. Lift the Centrifuge 5418 / 5418 R out of the carton. Lift the Centrifuge 5418 R out of the carton with the
help of another person.
4. Remove the front and back transport protection pads.
5. Place the device on a suitable lab bench.
6. Remove the plastic sleeve.
Page 18
18
Installation
Centrifuge 5418 / 5418 R
English (EN)
4.3Installing the instrument
NOTICE! Damage to electronic components due to condensation.
Condensation can form inside the device after the device has been moved from a cool to a
warmer environment.
5418: Wait for at least 3 hours before connecting the device to the power supply.
Only 5418: Alternative: Let the device heat up for 30 minutes right before a brief transport.
5418 R: Wait for at least 4 hours before connecting the device to the power supply.
NOTICE! Centrifuge 5418 R: compressor damage after improper transport.
Only switch on the centrifuge 4 hours after installation.
Perform the following steps in the sequence described.
1. Let the device warm up to ambient temperature.
2. Check that the mains voltage and frequency match the requirements on the device type plate.
3. Connect the centrifuge to the mains and switch it on using the mains/power switch.
• Display is active.
• Lid opens automatically
4. Loosen and remove the rotor using the supplied rotor key.
5. Remove the transport protection pad from the motor shaft.
6. Use the details included in the scope of delivery to check that the delivery is complete.
7. Check all parts for any transport damage. Contact your dealer if any damage is found.
8. Only 5418 R: Insert the condensation water tray at the side of the device into the holder provided.
Retain the packaging material and the transport protection device for subsequent transport or
storage. See also the instructions relating to transport.
Page 19
Operation
Centrifuge 5418 / 5418 R
19
English (EN)
5Operation
5.1Overview of operating controls
Familiarize yourself with the operating controls and the display before using the Centrifuge 5418 / 5418 R
for the first time.
Abb. 5-1: Control panel of Centrifuge 5418 and Centrifuge 5418 R.
1
5418
2
45
3
9
8
4 (9)
5418 R
8
Fig. 5-1:Control panel of Centrifuge 5418 and Centrifuge 5418 R.
1 Switch the displayed speed of centrifugation
(rpm/rcf)
2 Display
3 Release lid
4 Set centrifugation time
6 Adjust temperature (only 5418 R)
7 Start the temperature control run fast temp
(only 5418 R)
8 Short-spin centrifugation
9 Start and stop centrifugation
Only 5418 R: Press the time dial to start and stop
centrifugation.
25 (1)
6
7
3
6
5 Set speed of centrifugation
Only 5418 R: Press the speed dial to switch the
displayed centrifugation speed (rpm/rcf).
Page 20
20
Operation
Centrifuge 5418 / 5418 R
English (EN)
Abb. 5-2: Display of Centrifuge 5418 and Centrifuge 5418 R.
12
5418
3
7
5418 R
Fig. 5-2:Display of Centrifuge 5418 and Centrifuge 5418 R.
1 Centrifugation time
2 g-force (rcf) / speed (rpm)
5418 R: Set value x 1000
3 Status of centrifuge
: Centrifuge lid unlocked.
: Centrifuge lid locked.
(Flashing): Centrifuging in progress.
1. Switch on the centrifuge using the mains switch.
The centrifuge lid opens automatically after switching on using the mains switch.
2. Open the closed centrifuge lid by pressing the open key.
The parameter settings of the last run are displayed.
5.2.2Inserting the rotor
NOTICE! If handled incorrectly, the rotor can fall over.
The rotor lid screw may become loose if it is used to retain the rotor.
Always grasp the rotor with both hands for holding or transport.
1. Place the rotor on the motor shaft.
2. Insert the delivered rotor key in the rotor nut.
3. Turn rotor key clockwise until the rotor nut is firmly tightened.
Page 21
Centrifuge 5418 / 5418 R
5.2.3Loading the rotor
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes.
Only load adapters with suitable tubes.
Always use the same type of tubes (weight, material/density and volume).
Check symmetric loading by balancing the adapters and tubes used with scales.
CAUTION! Risk from damaged or overloaded tubes.
When loading the rotor, observe the safety precautions on dangers as a result of
overloaded or damaged tubes (see Warnings for intended use on p. 12).
To load the rotor, proceed as follows:
1. Check the maximum load (adapter, tube and contents) per rotor bore.
The maximum load is 3.75 g per rotor bore. This information can also be found on the rotor.
Operation
21
English (EN)
2. Load rotors and adapters only with the tubes intended for them.
3. Insert tubes opposite each other in pairs into the rotor bores. For symmetric loading, tubes that face
each other must be the same type and have the same filling quantity.
To keep the weight differences between the filled tubes low, we recommend taring with a balance. This
is gentle on the drive and reduces the operating noises.
5.2.4Closing the rotor lid
With the rotor FA-45-18-11 centrifugation is also possible without a rotor lid.
Please also note:
• The tube lids must be closed securely.
• Without the rotor lid, the rotor is not aerosol-tight.
• The centrifugation is slightly louder.
• Spin columns must always be centrifuged with a rotor lid.
Page 22
22
Operation
Centrifuge 5418 / 5418 R
English (EN)
1. Check to see if the outer sealing ring is correctly positioned in the
groove.
2. Fit the rotor lid vertically onto the rotor.
3. Lock the rotor by turning the red rotor lid screw clockwise beyond
an audible click until it can be turned no further.
The rotor is correctly locked after the audible "click" is heard!
5.2.5Closing the centrifuge lid
WARNING! Crushing of the fingers with the centrifuge lid.
When opening or closing the device lid, do not reach between the lid and device or into the
latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
1. Check correct attachment of rotor and rotor lid.
2. Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.
The centrifuge will close automatically.
The display shows the symbol.
Only 5418: The open key lights blue.
5.3Cooling (only 5418 R)
5.3.1Temperature adjustment
Set the temperature using the temp arrow keys from 0°C to +40°C.
5.3.2Temperature display
If the rotor is stopped:Set temperature
During centrifugation:Actual temperature
5.3.3Temperature monitoring
After the set temperature has been reached the centrifuge responds as follows to temperature fluctuations
during centrifugation:
Deviation from the target valueAction
ΔT > 3°CTemperature display flashes.
ΔT > 5°CPeriodic warning tone and display Error 18.
Centrifugation is stopped automatically.
Page 23
Operation
Centrifuge 5418 / 5418 R
English (EN)
5.3.4FastTemp
This function can be used to start a temperature control run directly without samples with a rotor and
temperature-specific speed in order to quickly adjust the rotor chamber, incl. rotor and adapters, up to the
set target temperature.
Prerequisites
• The centrifuge is switched on.
• Rotor and rotor lid are correctly mounted.
• The centrifuge lid is closed.
• The temperature and g-force (rcf) or speed (rpm) are set for the subsequent centrifugation (see
Centrifuging on p. 24).
1. Press the fast temp key.
The display shows FT as well as the current temperature and g-force (rcf) or speed (rpm).
The cooling time from room temperature (≈23 °C) to 4 °C takes Approx. 16 min.
The temperature control cycle ends automatically when the target temperature has been reached. A
periodic signal tone sounds.
23
2. Press the start/stop key to end the temperature control run early.
After the set temperature has been reached and the temperature control run is complete, the centrifuge
keeps the rotor chamber, with the centrifuge lid closed, at the set temperature, if the temperature is below
the ambient temperature. Irrespective of the target temperature, however, this continuous cooling does not
go below 4°C to prevent the rotor chamber from freezing.
The centrifuge stops the run automatically when the rotor has reached the set temperature.
Therefore, there may be a delay between the display of the achieved set temperature and the
automatic end of the temperature control run.
5.3.5Continuous cooling
When the rotor is stopped the rotor chamber is kept at the target temperature when the following
prerequisites are met:
• The centrifuge is switched on.
• The centrifuge lid is closed.
• The target temperature is below the ambient temperature.
During continuous cooling the following applies:
• The target temperature is displayed.
• Irrespective of the target temperature, continuous cooling does not go below 4°C to prevent the rotor
chamber and the samples from freezing and increased condensation in the device.
• The temperature adjustment takes longer because the rotor is not rotating.
Open the centrifuge lid to end continuous cooling.
If the centrifuge is not used for more than 8 hours, the continuous cooling is switched off automatically.
This protects against ice formation in the rotor chamber and the tubes, and against increased condensation
in the device.
The display shows the actual temperature of the rotor chamber.
With fast temp you can quickly reach the desired temperature again (see p. 23).
Page 24
24
Operation
Centrifuge 5418 / 5418 R
English (EN)
5.4Centrifuging
CAUTION! Risk from incorrectly-loaded rotors and damaged/overloaded tubes!
Before commencing centrifugation, follow the safety instructions relating to risks from
asymmetrically loaded and/or overloaded rotors and from overloaded, damaged and/or
open tubes (see Warnings for intended use on p. 12).
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
Before using the Centrifuge 5418 / 5418 R for the first time, familiarize yourself with the operating controls
and the display (see Overview of operating controls on p. 19).
Each of the centrifuging variants described here must be preceded by the preparation described above (see
Preparing for centrifugation on p. 20).
Only 5418 R: Please also note the instructions on cooling (see p. 22).
5.4.1Centrifugation with time setting
Perform the following steps in the sequence described.
1. Use time to set the run time.
2. Only 5418 R: Use temp to set the temperature.
3. Use speed to set the g-force (rcf) or speed (rpm).
4. Press start/stop to start centrifuging.
During centrifugation:
• In the display flashes while the rotor is running.
• Only 5418 R: The current temperature will be displayed.
• The fast temp (only 5418 R), open and short keys are blocked during centrifugation.
• During the run you can modify the total run time, the temperature (only 5418 R), the speed and the
rpm/rcf display. Only 5418 R: To change the centrifugation parameters, press the short key first.
The values flash in the display during the change. The new parameters are adopted immediately.
When the time is changed during a run, the time which has already elapsed is taken into account.
Please note that the shortest new total run time that can be set is the elapsed time plus 2 minutes.
• You can also terminate the centrifugation before the set run time has elapsed by pressing the start/
stop key.
Page 25
Operation
Centrifuge 5418 / 5418 R
English (EN)
End of centrifugation
• The centrifuge automatically stops after the set time has elapsed. The elapsed centrifugation will be
shown in a blinking display during the braking process. A signal tone sounds when the rotor is at a
standstill.
• Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
• Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You can open it
by pressing the open key.
5. Remove the material for centrifuging.
5.4.2Centrifuging in continuous operation
Perform the following steps in the sequence described.
1. Use time to set the continuous run.
The continuous run function can be set above 9:59 h or under 30 s (5418) or above 99 min or under
0.5 min (5418 R). The timer shows oo to indicate continuous operation.
25
2. Only 5418 R: Use the temp arrow keys to adjust the temperature.
3. Use speed to set the g-force (rcf) or speed (rpm).
4. Press start/stop to start centrifuging.
In the display flashes while the rotor is running.
Time is counted upwards, first in 30-second increments and then in minute increments from ten
minutes.
5. Press start/stop to end centrifuging after the desired time period.
• The centrifugation time will be shown in a blinking display during the braking process.
• A signal tone sounds when the rotor is at a standstill.
• Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
• Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You can open it
by pressing the open key.
6. Remove the material for centrifuging.
5.4.3Short-spin centrifugation
5418: You can carry out a short run with the maximum g-force (rcf) / speed (rpm).
5418 R: You can carry out a short run with the currently set or with the maximum g-force (rcf) /
speed (rpm). Set this short spin mode as described in the following section.
5.4.3.1 Selecting short spin mode (only 5418 R)
1. Press the short key while the centrifuge lid is open.
The current mode is displayed:
• Display 1 – 14 t (run at preselected speed)
• Display 14 t (run at maximum speed of 14.000 rpm)
2. When the centrifuge lid is open, press the short key for longer than 2 s to switch between these modes.
Page 26
26
Operation
Centrifuge 5418 / 5418 R
English (EN)
5.4.3.2 Performing short spin centrifugation
1. Only 5418 R: A short run at preselected g-force (rcf) or speed (rpm) can be set directly using the speed
dial.
2. Only 5418 R: Use the temp arrow keys to adjust the temperature.
3. Start short run: Hold down the short key.
• In the display flashes while the rotor is running.
• The time is counted upwards in seconds.
• During short run centrifuging all other keys are blocked.
4. End short run: Release the short key.
• The centrifugation time will be shown in a blinking display during the braking process.
• Only 5418: The centrifuge lid opens automatically. The display shows the symbol .
• Only 5418 R: The centrifuge lid remains closed to maintain the sample temperature. You can open it
by pressing the open key.
5. Remove the material for centrifuging.
During the braking process, you can restart the centrifugation up to two times by pressing the
short key again.
5.4.4Removing the rotor
1. Turn the rotor nut counterclockwise using the supplied rotor key.
2. Remove the rotor vertically in an upward motion.
3. Only 5418 R: Switch off the centrifuge after use and empty the condensation water tray (pull out from
the left or right side of the device). Leave the centrifuge lid fully open and secure it against closing.
5.5Standby mode (only 5418)
If the centrifuge has not been used for 15 min, it switches to standby mode. The EP logo then appears in
the display. When a button or knob is used or the centrifuge lid is closed, the centrifuge is reactivated and
ready for operation.
5.6Special functions (only 5418 R)
FunctionStatus of centrifuge lidPress > 2 s keyDisplay
Modify parameter during
the cycle.
Enable/disable signal
tone.
Enable/disable key lock.
At set rpm
closedshort keyFlashes 5 s
open
open
open
open
short + open
time
/
/
Page 27
Maintenance
Centrifuge 5418 / 5418 R
English (EN)
6Maintenance
6.1Prepare cleaning/disinfection
The procedure described in the following chapter applies to the cleaning as well as to the disinfection or
decontamination. The table below describes the steps required on top of this:
CleaningDisinfecting/decontamination
1. Use a mild cleaning fluid to clean the accessible
surfaces of the device and the accessories.
2. Carry out the cleaning as described in the
following chapter.
1. Choose the disinfection method which
corresponds to the legal regulations and
guidelines in place for your range of application.
For example, use alcohol (ethanol, isopropanol)
or alcohol-based disinfectants.
2. Carry out the disinfection or decontamination as
described in the following chapter.
3. Then clean the device and the accessories.
27
If you have any further questions regarding the cleaning and disinfection or decontamination
or regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The
contact details are provided on the back of this manual.
6.2Cleaning/disinfection
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, immediately clean it by
means of a mild cleaning agent.
NOTICE! Corrosion from aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.
Do not incubate the accessories in aggressive cleaning agents or disinfectants for a longer
period of time.
Page 28
28
Maintenance
Centrifuge 5418 / 5418 R
English (EN)
NOTICE! Damage from UV and other high-energy radiation.
Do not use UV, beta, gamma, or any other high-energy radiation for disinfecting.
Avoid storage in areas with strong UV radiation
Autoclaving
Rotors, rotor lids and adapters can be autoclaved (121 °C, 20 min).
Replace the sealing ring in the lid groove of the aerosol-tight rotor lid after a maximum of 50
autoclaving cycles.
Aerosol tightness
Check that the seals are intact before use.
Only QuickLock rotor lid: Replace the sealing ring in the lid groove when it becomes worn.
The sealing rings require regular care to protect the rotors.
Aerosol-tight rotors should never be stored with lids screwed on!
In order to prevent damage, lightly lubricate the lid thread of the aerosol-tight rotors with
pivot grease (order no. Int.: 5810 350.050/North America: 022634330).
6.2.1Cleaning and disinfecting the device
1. Open the lid. Switch off the device with the mains/power switch. Disconnect the power plug from the
power supply.
2. Loosen the rotor nut by turning the rotor key counterclockwise.
3. Remove the rotor.
4. Clean and disinfect all accessible surfaces of the device, including the power cable, using a damp cloth
and the recommended cleaning agents.
5. Thoroughly clean the rubber seals of the rotor chamber with water.
6. Rub the dry rubber seals with glycerine or talcum powder to prevent them from becoming brittle. Other
components of the device, such as the lid latch, motor shaft and rotor cone, must not be lubricated.
7. Clean the motor shaft with a soft, dry and lint-free cloth. Do not lubricate the motor shaft.
8. Check the motor shaft for damage.
9. Inspect the device for corrosion and damage.
10.Leave the centrifuge lid open when the device is not being used.
11.Only connect the device to the power supply if it is fully dry inside and out.
6.2.2Cleaning and disinfecting the rotor
1. Inspect the rotor and accessories for damage and corrosion. Do not use any damaged rotors or
accessories.
2. Clean and disinfect the rotors and accessories with the recommended cleaning agents.
3. Use a bottle brush to clean and disinfect the rotor bores.
4. Remove the sealing ring to clean and disinfect the rotor lid. Clean the groove below it and the sealing
ring.
Page 29
Maintenance
Centrifuge 5418 / 5418 R
English (EN)
5. Rinse the rotors and accessories thoroughly with distilled water. Rinse the rotor bores of fixed-angle
rotors particularly thoroughly.
Do not immerse the rotor in liquid as liquid can get trapped inside the cavities.
29
6. Place rotors and accessories on a cloth to dry. Place fixed-angle rotors with the rotor bores facing
downwards to allow the bores to also dry.
7. Correctly reinsert the rotor lid sealing ring in the clean and dry groove.
8. Clean the rotor cone with a soft, dry and lint-free cloth. Do not lubricate the rotor cone.
9. Inspect the rotor cone for damage.
10.Place the dry rotor onto the motor shaft.
11.Tighten the rotor nut firmly by turning it clockwise with the rotor key.
12.Leave the rotor lid open when the rotor is not being used.
6.3Additional service instructions for Centrifuge 5418 R
Empty and clean the condensation water tray regularly and especially after liquid spillage in the rotor
chamber. Pull out the condensation water tray from the left or right side of the centrifuge.
Regularly free the rotor chamber ice formations via thawing, by leaving the centrifuge lid open or
carrying out a short temperature control run at approx. 30 °C.
Leave the centrifuge lid open when not in use for a long period.
Wipe up condensate in the rotor chamber using a soft, absorbent cloth.
Remove dust deposits from the ventilation slits of the centrifuge using a brush or swab at the latest
every six months. First switch off the device and remove the power plug.
6.4Glass breakage
When using glass tubes there is a risk of glass breakage in the rotor chamber. The resulting glass splinters
are swirled around in the rotor chamber during centrifugation and have a sandblasting effect on the rotor
and accessories. The smallest glass particles become lodged in the rubber parts (e.g., the motor guide, the
rotor chamber gasket, and the rubber mats of adapters).
Page 30
30
Maintenance
Centrifuge 5418 / 5418 R
English (EN)
NOTICE! Glass breakage in the rotor chamber
Glass tubes in the rotor chamber may break if the g-force is too high. Broken glass can
damage the rotor, accessories and samples.
Please note the manufacturer's information on the recommended centrifugation
parameters (load and speed).
Effects of glass breakage in the rotor chamber:
• Fine black metal abrasion in the rotor chamber (in metal rotor chambers)
• The surfaces of the rotor chamber and accessories are scratched.
• The chemical resistance of the rotor chamber is reduced.
• Contamination of samples
• Wear on rubber parts
How to proceed in case of glass breakage
1. Remove all splinters and glass powder from the rotor chamber and accessories.
2. Thoroughly clean the rotor and rotor chamber. Thoroughly clean the bores of the fixed-angle rotors, in
particular.
3. If required, replace the adapters to prevent any further damage.
4. Regularly check the rotor bores for deposits and damage.
6.5Fuses
The fuse holder is located under the mains power socket (5418) or on the left next to the mains switch
(5418 R).
1. Disconnect the mains plug.
2. Remove the fuse holder.
Both fuses are now accessible and can be replaced.
6.6Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer
for disposal please note the following:
WARNING! Risk to health from contaminated device
1. Observe the notes on the decontamination certificate. You find it as a PDF file on our
website (
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
www.eppendorf.com/decontamination).
Page 31
Troubleshooting
Centrifuge 5418 / 5418 R
English (EN)
7Troubleshooting
If you cannot remedy an error with the recommended measures, please contact your local Eppendorf
partner. The contact address can be found online at:
www.eppendorf.com/worldwide.
7.1General errors
Symptom/message CauseRemedy
No display.No power connection.
No display.Power failure.
Lid of the device
cannot be opened.
Lid of the device
cannot be opened.
Device cannot be
started.
Device shakes
when it starts up.
Centrifuge brakes
during a short run
centrifugation,
although the short
key is pressed.
Temperature
display flashes.
(only for 5418 R)
Rotor is still running.
Power failure.1. Check the mains fuse of the device
Lid of the device is not closed.
Rotor loaded unsymmetrically.1. Stop the device and load
The short key was released briefly more
than twice (protective function for the
drive).
Temperature deviation from the set
value: ±3 °C.
Check mains/power connection.
Check the mains fuse of the device
(see Fuses on p. 30).
Check the mains fuse of the
laboratory.
Wait for rotor to stop.
(see Fuses on p. 30).
2. Check the mains fuse of the
laboratory.
3. Activate the emergency lid release
(see p. 33).
Close device lid.
symmetrically.
2. Restart device.
Press the short key continuously
during a short run centrifugation.
Check the settings.
Wait until the set temperature has
been reached.
Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
31
Page 32
32
Troubleshooting
Centrifuge 5418 / 5418 R
English (EN)
7.2Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.
Symptom/message CauseRemedy
LID ERROR (5418)/
LID (5418 R)
LID ERROR (5418)/
LID (5418 R)
LID ERROR (5418)/
LID (5418 R)
INTMains/power failure during a run.
NO RPM (5418)/
Error 3 (5418 R)
Error 5
(only for 5418 R)
Err 6 (5418)/
Error 6 (5418 R)
Error 6
(only for 5418 R)
Err 7 (5418)/
Error 7 (5418 R)
Err 8 (5418)/
Error 8 (5418 R)
Err 9 to 17 (5418)/
Error 9 to 17 (5418
R)
Error 18
(only for 5418 R)
Centrifuge lid cannot be locked.
Centrifuge lid cannot be released.1. Switch the centrifuge off and back on.
Centrifuge lid must not be released
during a run.
Try again to close centrifuge lid.
2. Press the open key.
If the error occurs again:
1. Switch off centrifuge.
2. Activate the emergency lid release
(see Emergency lid release on p. 33).
Wait for rotor to stop.
Check the mains connection.
Error in the speed measuring system or
drive overheated.
Prohibited opening of lid or lid switch is
defective during a run.
Drive fault.
Leave the device switched on until the
error message disappears (10 s or 6
min).
1. Wait for rotor to stop.
2. Open and close again the lid of the
device.
3. Repeat the run.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on
again after > 20 s.
Drive overheated.
Major deviation in the speed control.1. Wait for rotor to stop.
Drive fault.1. Wait for rotor to stop.
Electronics error.
Too high temperature deviation from set
value in the rotor chamber.
Allow the drive to cool down for at
least 15 min.
2. Tighten rotor.
2. Repeat the run.
Switch centrifuge off and back on
again after > 20 s.
Check the settings.
Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
Page 33
Troubleshooting
Centrifuge 5418 / 5418 R
Symptom/message CauseRemedy
Error 19
(only for 5418 R
Error 20
(only for 5418 R)
Error 21
(only for 5418 R)
Error 24
(only for 5418 R)
Cooling circuit is overheated.
Temperature sensor in rotor chamber is
faulty.
Temperature sensor on condenser is
faulty.
Cooling unit fault, e.g. overheated.
Check unhindered air circulation
through the air slots and allow device
to cool down.
Switch centrifuge off and back on
again after > 20 s.
Switch centrifuge off and back on
again after > 20 s.
Allow the centrifuge to cool down and
repeat the run.
7.3Emergency lid release
If the centrifuge lid cannot be opened, the emergency lid release can be operated manually.
33
English (EN)
WARNING! Risk of injury from rotating rotor.
Wait until the rotor has come to a standstill before you operate the emergency release.
To check, look through the inspection glass in the centrifuge lid.
7.3.1Centrifuge 5418
1. Disconnect the mains plug.
2. Open the plastic cover of the emergency lid release.
This is located behind the front left device foot in the base plate (Fig. 2-1 on p. 7).
3. Pull the cord vertically downwards.
The centrifuge lid is unlocked.
4. Before closing the centrifuge lid again: push the cord completely into the housing and insert the plastic
cover in the base plate.
7.3.2Centrifuge 5418 R
1. Disconnect the mains plug.
2. Turn the plastic cover of the emergency lid release anti-clockwise by 90° with a suitable tool (e.g.
screwdriver), and remove it.
The plastic cover is located on the right side of the device (Fig. 2-2 on p. 7).
3. Insert the centrifuge rotor key into the hexagonal opening at rear until some resistance can be felt.
4. Press the rotor key slightly down, and turn it clockwise.
This will release the centrifuge lid.
5. Open the centrifuge lid.
6. Remove the rotor key and put the plastic cover back on rotating it 90° clockwise.
Page 34
34
Troubleshooting
Centrifuge 5418 / 5418 R
English (EN)
Page 35
Transport, storage and disposal
Centrifuge 5418 / 5418 R
English (EN)
8Transport, storage and disposal
8.1Transport
CAUTION! Bodily injury due to lifting and carrying heavy loads
The device is heavy. Lifting and carrying the device can lead to back injuries.
The device must be transported by least two people.
Use a transport aid (e.g., dolly) to transport the device longer distances.
Remove the rotor from the centrifuge before transport.
Use the original packaging for transport.
Air temperatureRelative humidityAtmospheric pressure
General transport-25 °C – 60 °C10 % – 75 %30 kPa – 106 kPa
Air freight-20 °C – 55 °C10 % – 75 %30 kPa – 106 kPa
35
8.2Storage
Air temperatureRelative humidityAtmospheric pressure
In transport packaging-25 °C – 55 °C10 % – 75 %70 kPa – 106 kPa
Without transport
packaging
-5 °C – 45 °C10 % – 75 %70 kPa – 106 kPa
8.3Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations
based on EU Directive 2002/96/EC pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business
sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To
document this, they have been marked with the following identification:
Because disposal regulations may differ from one country to another within the EU, please contact your
supplier if necessary.
Page 36
36
Transport, storage and disposal
Centrifuge 5418 / 5418 R
English (EN)
Page 37
Centrifuge 5418 / 5418 R
9Technical data
9.1Power supply
Centrifuge 5418
Mains connection230 V, 50 Hz – 60 Hz
120 V, 50 Hz – 60 Hz
100 V, 50 Hz – 60 Hz
Current consumption1.4 A (230 V )
2.8 A (120 V )
3.4 A (100 V )
Power consumptionMaximum 170 W
EMC: Interference emission. (radio interference)EN 61326 – Class B (230 V)
FCC15 - Class A (120 V/100 V)
EMC: Noise immunityEN 61326
Overvoltage categoryII
Fuses2.5 AT (230 V)
5.0 AT (120 V / 100 V)
Degree of pollution2
Technical data
37
English (EN)
Centrifuge 5418 R
Mains connection230 V, 50 Hz – 60 Hz
120 V, 50 Hz – 60 Hz
100 V, 50 Hz – 60 Hz
Current consumption1.4 A (230 V)
2.8 A (120 V)
3.0 A (100 V)
Power consumptionMaximum 320 W
EMC: Interference emission. (radio interference)EN 61326 – Class B (230 V)
FCC15 - Class A (120 V/100 V)
EMC: Noise immunityEN 61326
Overvoltage categoryII
Fuses2.5 AT (230 V)
6.25 AT (120 V / 100 V)
Degree of pollution2
Page 38
38
Technical data
Centrifuge 5418 / 5418 R
English (EN)
9.2Ambient conditions
Environment:Only for use indoors.
Ambient temperature: Centrifuge 5418: 2 °C to 40 °C
Centrifuge 5418 R: 15 °C to 35 °C
Max. relative humidity:75 %, non-condensing humidity.
Atmospheric pressure:Use up to an altitude of 2 000 m above MSL.
Degree of pollution:2
9.3Weight/dimensions
Tab. 9-1: Centrifuge 5418
Dimensions:Width: 208 cm (8.19 in)
Depth: 300 cm (11.8 in)
Height: 210 cm (8.27 in)
Weight without rotor:7.7 kg
Noise level:< 51 dB (A) *
Tab. 9-2: Centrifuge 5418 R
Dimensions:Width: 298 cm (11.73 in)
Depth: 463 cm (18.23 in)
Height: 463 cm (18.23 in)
Weight without rotor:22 kg
Noise level:< 57 dB (A) *
*) The noise level was measured according to DIN EN ISO 3745 frontally in a sound measuring room with
accuracy class 1 at a distance of 1 m from the device and at lab bench height.
Page 39
9.4Application parameters
Start and stop times according to DIN 58 970
Rotor54185418 R
VoltageAcceleration
times
FA-45-18-11230 V
120 V
100 V
These values were calculated at 23°C.
16 s
16 s
18 s
Deceleration
times
18 s
18 s
18 s
VoltageAcceleration
230 V
120 V
100 V
Technical data
Centrifuge 5418 / 5418 R
English (EN)
Deceleration
times
13 s
13 s
13 s
times
13 s
13 s
13 s
39
run time
(5418)
run time
(5418 R)
Temperature
(only 5418 R)
Relative centrifugal force (RCF)1 to 16,873 x g
RPM100 to 14,000 rpm
Maximum load18 micro test tubes of 2.0 mL each.
Max. kinetic energy2,600 Nm
Test log mandatory:No
Allowable density of the centrifuge material (at
maximum g-force (rcf)/speed (rpm) and maximum
load)
30 s to 9:59 h, as well as infinity (oo),
Adjustable to 10 min in 30 s increments, then
increments of 1 min
0.5 min to 99 min, as well as infinity (oo),
Adjustable to 10 min in 0.5 min increments, then
increments of 1 min
0°C to 40°C
adjustable to 5000 × g in 100 x g increments, then
Abb. 1-1: Vista anterior y posterior de la Centrífuga 5418
1
2
3
4
5
Imag. 1-1: Vista anterior y posterior de la Centrífuga 5418
Abb. 1-2: Vista anterior y posterior de la Centrífuga 5418 R
Descripción del producto
Centrifuge 5418 / 5418 R
45
Español (ES)
1
2
6
7
8
9
1
2
1
2
5
3
4
10
6
9
10
7
8
Imag. 1-2: Vista anterior y posterior de la Centrífuga 5418 R
1 Tapa de la centrifuga
2 Tubito de control
Control visual de la parada del rotor y/o
posibilidad de controlar las revoluciones
mediante estroboscopio.
6 Placa de características
7 Interruptor de red
Interrumptor para activar y desactivar el aparato.
Posición 0: el aparato está desconectado.
Posición I: el aparato está conectado.
3 Indicador
Representación de los parámetros de
centrifugación y ajustes de aparato actuales .
4 Panel de mando
Para el manejo de la centrífuga .
5 Desbloqueo de emergencia de la tapa
8 Alimentación eléctrica
Toma hembra para la conexión del cable de red
suministrado.
9 Portafusibles
10 Recipiente colector de agua condensada (sólo
Centrífuga 5418 R)
Page 46
46
Descripción del producto
Centrifuge 5418 / 5418 R
Español (ES)
1.2Rotor FA-45-18-11
Antes de utilizar tubos de muestras, observe las especificaciones respecto a la resistencia al centrifugado
(fuerza de la gravedad máx.) recomendadas por el fabricante.
Capacidad máx.Máx. número
g (rcf) /
velocidad (rp
m) sin adapta-
Carga máx.
por cada
orificio del
(1)
rotor
Indicaciones
dor
Tiempo de arranque / parada
(2)
Rotor
FA-45-18-11
18 tubos de reacción de 1,5/
2,0 ml o Spin-Columns.
Con adaptadores:
• Recipientes PCR de 0,2 ml
16.873 x g /
14.000 rpm
5418: 16 s / 18 s
5418 R: 13 s / 13 s
3,75 g• Tapa de rotor
hermética a los
(3)
aerosoles
(aluminio).
• Tubos de reacción de 0,4
ml/0,5 ml
• Tubos Microtainer de 0,6
ml
(1) Carga máxima por cada orificio del rotor para adaptador + tubo + contenido.
(2) Según DIN 58 970 (variante de aparato: 230 V, 50 hasta 60 Hz).
(3) Hermético a los aerosoles comprobado y certificado por el Centre of Emergency Preparedness and
Response, Health Protection Agency, Porton Down (UK), (véanse los certificado al final de este manual de
instrucciones).
Rotor code:
All Eppendorf® rotors are identified
using a simple, alphanumeric
format that represents the technical
specifications in a uniform series of
letters and numbers.
Fixed-angle
rotor
F
Aerosol-tight
Angle of
bores
bores (mm)
A453011
version
Number of
bores
Ø of
Swing-bucket
rotor
A
481
Number of
buckets
Ø of
buckets (mm)
Page 47
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
2Instrucciones generales de seguridad
2.1Utilización de estas instrucciones
Lea detenidamente este manual de instrucciones antes de poner en funcionamiento el equipo por
primera vez. Tenga en cuenta, en caso necesario, los manuales de instrucciones de los accesorios.
Encontrará una descripción detallada del equipo en la versión en inglés y en alemán del manual de
instrucciones.
Igualmente, puede encontrar las características técnicas e información de pedidos en la versión inglesa
y alemana.
Este manual de instrucciones es parte del producto. Consérvelo en un lugar accesible.
Incluya siempre este manual de instrucciones cuando entregue el equipo a terceros.
En caso de pérdida del manual de instrucciones, solicite otro. Encontrará una versión actual de este
manual en nuestra página de internet
www.eppendorf.com/worldwide.
2.2Uso de acuerdo con lo previsto
47
La Centrífuga 5418 / 5418 R está destinada exclusivamente para su uso en interiores y sirve para la
separación de soluciones acuosas y suspensiones de diferente densidad en tubos de muestras autorizados.
2.3Requerimiento para el usuario
Este aparato debe ser manejado exclusivamente por personal técnico debidamente formado. El personal
debe haber leído atentamente el manual de instrucciones y estar familiarizado con el funcionamiento del
aparato.
2.4Límites de aplicación
2.4.1Aclaración de la directriz ATEX (94/9/EG)
¡PELIGRO! Peligro de explosión.
No utilice el equipo en salas en donde se trabaje con sustancias explosivas.
No procese con este equipo sustancias explosivas o que reaccionen bruscamente.
No procese con este equipo sustancias que puedan crear una atmósfera explosiva.
La Centrífuga 5418 / 5418 R no está indicada para su utilización en una atmósfera potencialmente
explosiva debido a su construcción y a las condiciones en el interior del equipo.
El equipo sólo puede utilizarse en un ambiente seguro, es decir en el ambiente abierto de un laboratorio
ventilado o una campana extractora. No está permitido el uso de sustancias que puedan originar una
atmósfera potencialmente explosiva. La decisión definitiva respecto a los riesgos relacionados con el uso
de tales sustancias es responsabilidad del usuario.
Page 48
48
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
2.4.2Máxima vida útil de los accesorios
¡ADVERTENCIA! Peligro de lesión a causa de accesorios dañados química o
mecánicamente.
Tanto arañazos como grietas pequeñas pueden provocar graves daños en los materiales
internos.
Proteja todas las piezas de los accesorios contra los daños mecánicos.
Controle la presencia de daños en los accesorios antes de cada uso. Cambie los accesorios
dañados.
No utilice rotores o tapas de rotor con huellas de corrosión o daños mecánicos (p.ej.
torceduras).
No utilice ningún accesorio, cuya máxima vida útil haya caducado.
AccesoriosMáxima vida útil a partir de la
primera puesta en marcha
Tapa del rotor3 años
Condición: el "clic" al ajustar el
tornillo de la tapa del rotor es todavía
audible.
Adaptador de plástico1 año
El rotor descrito no tiene una vida útil limitada, siempre y cuando se cumplan las condiciones siguientes:
• Uso correcto
• Mantenimiento recomendado
• Estado libre de daños.
La fecha de fabricación está grabada en los rotores como 03/10 (= marzo 2010). Esto sólo es a título
informativo y no es una referencia a la vida útil real.
Para garantizar la hermeticidad a los aerosoles se aplica lo siguiente:
• Cambie las tapas de rotor y cubiertas herméticas a los aerosoles después de 50 ciclos de autoclave.
• Cambie el anillo de obturación de la tapa de rotor QuickLock después de 50 ciclos de autoclave.
2.5Indicaciones relativas a la responsabilidad del producto
Los siguientes casos pueden reducir la protección prevista del aparato. La entidad explotadora asume
entonces la responsabilidad por las lesiones personales y daños materiales producidos:
• El equipo no es utilizado de acuerdo con lo especificado en el manual de instrucciones.
• El equipo no es utilizado de acuerdo con lo prescrito.
• El equipo se utiliza con accesorios o consumibles no recomendados por Eppendorf.
• El equipo es mantenido o reparado por personas que no han sido autorizadas por Eppendorf AG.
• El usuario realiza modificaciones no autorizadas en el equipo.
Page 49
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
2.6Peligros durante el uso previsto
Antes de utilizar la Centrífuga 5418 / 5418 R, lea primero el manual de instrucciones y tenga en cuenta las
siguientes indicaciones generales de seguridad.
2.6.1Daños personales o al equipo
¡ADVERTENCIA! Descarga eléctrica por daños en el equipo o cable de alimentación.
Sólo encienda el equipo si éste y el cable de alimentación no presentan ningún daño.
Utilice únicamente equipos que hayan sido instalados o reparados correctamente.
En caso de peligro, separe el equipo de la red eléctrica extrayendo el enchufe de
alimentación del equipo o con ayuda del dispositivo de separación previsto para ello (p.ej.,
el interruptor de emergencia del laboratorio).
¡ADVERTENCIA! Descargas de tensión mortales en el interior del equipo.
Cerciórese de que la carcasa siempre esté cerrada y no presente daños, de modo que no
sea posible tocar piezas en el interior del equipo por descuido.
No desmonte el revestimiento del equipo.
Ningún líquido debe penetrar en el interior de la carcasa.
Sólo deje que personal de mantenimiento autorizado por Eppendorf abra el equipo.
49
¡ADVERTENCIA! Peligro a causa de un suministro de energía eléctrica equivocado.
Sólo conecte el equipo a fuentes de alimentación que cumplan los requisitos eléctricos
especificados en la placa de características.
Sólo utilice enchufes con toma de tierra y un cable de red adecuado.
¡ADVERTENCIA! Daños para la salud a causa de líquidos infecciosos y gérmenes
patógenos.
Siempre tenga en cuenta las disposiciones nacionales, el nivel de seguridad biológica de
su laboratorio, así como las fichas de datos de seguridad y las instrucciones de uso del
fabricante cuando maneje líquidos infecciosos y gérmenes patógenos.
Utilice sistemas de cierre hermético a los aerosoles al centrifugar estas sustancias.
Cuando trabaje con gérmenes patógenos que pertenezcan a un grupo de riesgo superior,
debe disponer de más de una impermeabilización biológica hermética a los aerosoles.
Póngase su equipo de protección personal.
Unas prescripciones amplias respecto al manejo de gérmenes o material biológico del
grupo de riesgo II o superior se encuentran en el "Laboratory Biosafety Manual" (fuente:
World Health Organisation, Laboratory Biosafety Manual, en la versión actualmente
vigente).
¡ADVERTENCIA! Peligro de aplastamiento de los dedos con la tapa de la centrífuga.
Al abrir y cerrar la tapa del equipo, no ponga la mano entre la tapa y el equipo o en el
mecanismo de cierre de la tapa.
Abra siempre la tapa de la centrífuga por completo para que ésta no pueda cerrarse de
golpe.
Page 50
50
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
¡ATENCIÓN! Riesgos de seguridad debido a accesorios y piezas de recambio equivocados.
Los accesorios y piezas de recambio no recomendados por Eppendorf merman la seguridad,
el funcionamiento y la precisión del equipo. Por daños producidos por accesorios y piezas de
recambio no recomendados por Eppendorf o por un uso incorrecto, Eppendorf queda eximido
de cualquier responsabilidad o garantía.
Utilice exclusivamente accesorios y piezas de recambio recomendados por Eppendorf.
¡AVISO! Daños en el equipo a causa de líquidos derramados.
1. Apague el equipo.
2. Desconecte el equipo del suministro de corriente.
3. Lleve a cabo una limpieza cuidadosa del equipo y sus accesorios según las indicaciones de
limpieza y desinfección del manual de instrucciones.
4. Si debe utilizarse otro método de limpieza o desinfección, consulte a Eppendorf AG para
asegurarse de que el método previsto no dañe el equipo.
¡AVISO! Daños en los componentes electrónicos debido a la formación de condensación.
Después de transportar el equipo de un ambiente frío a un ambiente caliente se puede formar
líquido de condensación en el equipo.
5418: espere por lo menos 3 horas hasta que conecte el equipo al suministro de corriente.
Sólo 5418: Alternativa: Encienda el equipo y deje que se caliente durante una media hora
directamente antes de un transporte de corta distancia.
5418 R: espere por lo menos 4 horas hasta que conecte el equipo al suministro de
corriente.
2.6.2Manipulación errónea de la centrífuga
¡AVISO! Daños por golpes o movimientos del equipo en funcionamiento.
Un rotor que golpea contra la pared de la cámara produce daños considerables en el equipo y
en el rotor.
No mueva o golpee el equipo mientras éste está en funcionamiento.
Page 51
2.6.3Manipulación errónea de los rotores
¡ADVERTENCIA! Peligro de lesión por rotores y tapa del rotor incorrectamente fijados.
Centrifugue sólo cuando el rotor y la tapa del rotor estén bien fijados.
Si al arrancar la centrífuga se producen ruidos no habituales, puede que el rotor o la tapa
del rotor no estén fijados correctamente. Finalice el centrifugado de inmediato pulsando la
tecla Start/stop.
¡ATENCIÓN! Peligro de lesión por carga asimétrica de un rotor.
Equipe los rotores siempre simétricamente con los mismos tubos.
Cargue los adaptadores sólo con los tubos adecuados.
Utilice siempre tubos del mismo tipo (peso, material/densidad y volumen).
Compruebe la carga simétrica tarando o equilibrando los tubos y adaptador utilizados con
una báscula.
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
51
¡ATENCIÓN! Peligro de lesión por sobrecarga del rotor.
La Centrífuga 5418 / 5418 R está diseñada para el centrifugado de sustancias con una
densidad máxima de 1,2 g/mL con un número de revoluciones máximo y un volumen de
llenado o carga máximo.
Observe las indicaciones en cada rotor respecto a la carga máxima (adaptador, tubo y
contenido) por cada orificio del rotor y no las sobrepase.
¡AVISO! Deterioro de los rotores a causa de productos químicos agresivos.
Los rotores son componentes de alta calidad que resisten cargas extremas. Esta estabilidad
puede verse perjudicada por productos químicos agresivos.
Evite el uso de productos químicos agresivos; a esta clase de productos pertenecen, entre
otros, álcalis fuertes y débiles, ácidos fuertes, soluciones con iones de mercurio, cobre u
otros metales pesados, hidrocarburos halogenados, soluciones salinas concentradas y
fenol.
En los rotores identificados con "revestido" ("coated") pueden surgir cambios de color
debido al proceso de producción. Éstos no tienen ninguna repercusión sobre la
durabilidad o la resistencia a los productos químicos.
2.6.4Utilización extrema de los tubos de centrifugado
¡ATENCIÓN! Peligro de lesión por tubos sobrecargados.
Observe los valores límite especificados por el fabricante de los tubos sobre la capacidad
de carga de los mismos.
Utilice exclusivamente tubos autorizados por el fabricante para la fuerza de la gravedad
(FCR) deseada.
Page 52
52
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
¡AVISO! Peligro por tubos dañados.
No deben utilizarse tubos dañados. La consecuencia pueden ser daños adicionales en el
equipo y en los accesorios así como la pérdida de muestras.
Inspeccione visualmente todos los tubos respecto a posibles daños antes de su uso.
¡AVISO! Peligro por tapas de tubo abiertas.
Las tapas de tubo abiertas pueden romperse en el centrifugado y dañar tanto el rotor como la
centrífuga.
Cierre cuidadosamente todas las tapas de tubo antes del centrifugado.
¡AVISO! Daños de los tubos de plástico por disolventes orgánicos.
En caso de utilización de disolventes orgánicos (p.ej. fenol, cloroformo) se reduce la
resistencia de los tubos de plástico, de forma que se pueden dañar los tubos.
Observe las indicaciones del fabricante sobre la resistencia química de los tubos.
¡AVISO! Los tubos de muestras se calientan.
En las centrífugas no refrigeradas, dependiendo de la duración del ciclo, la fuerza de la
gravedad (FCR)/número de revoluciones y la temperatura ambiente, la temperatura en la
cámara del rotor, el rotor y la muestra puede aumentar por encima de 40 °C.
Tenga en cuenta que en este caso la resistencia al centrifugado de los tubos de muestras
se reduce.
Observe la resistencia a la temperatura de las muestras.
2.6.5Centrifugado hermético a los aerosoles
¡ADVERTENCIA! Riesgos para la salud debido a una hermeticidad a los aerosoles limitada
por una combinación incorrecta de rotor/tapa del rotor.
El centrifugado hermético a los aerosoles sólo está garantizado en caso de utilizar los rotores
y tapas de rotor previstos para ello. En los rotores de ángulo fijo la denominación siempre
comienza con FA, en los rotores basculantes con AT (hermético a los aerosoles).
Los rotores y las tapas de rotor herméticos a los aerosoles de esta centrífuga están marcados
adicionalmente con un anillo rojo en el rotor y un tornillo de fijación rojo para la tapa del rotor.
Para el centrifugado hermético a los aerosoles en la centrífuga prevista, utilice siempre al
mismo tiempo rotores y tapas de rotor con marca de hermético a los aerosoles. La
indicación sobre qué tipo de rotores y tapas de rotor herméticos a los aerosoles pueden
utilizarse en una determinada centrífuga la encontrará en el rotor y a partir de la fecha de
producción de octubre de 2003 en la parte superior de la tapa del rotor.
Utilice tapas de rotor herméticas a los aerosoles exclusivamente en combinación con
rotores que estén indicados en la tapa del rotor.
Page 53
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
¡ADVERTENCIA! Riesgos para la salud debido a una hermeticidad a los aerosoles limitada
por una aplicación incorrecta.
Las autoclaves, las cargas mecánicas y la contaminación por sustancias químicas u otras
soluciones agresivas pueden perjudicar la hermeticidad a los aerosoles de los rotores y las
tapas de rotor.
53
Controle antes de cada uso la integridad de los anillos de obturación de las tapas o
cubiertas herméticas a los aerosoles.
Utilice tapas de rotor o cubiertas herméticas a los aerosoles sólo con anillos de obturación
limpios y en buen estado.
Unte ligeramente la rosca del tornillo de la tapa de rotor después de cada autoclave
correcta (121 °C, 20 min.) con grasa para gorrones (n.° de pedido. int. 5810 350.050,
Norteamérica 022634330). No unte la grasa para gorrones sobre los anillos de obturación.
Cambie las tapas de rotor y cubiertas herméticas a los aerosoles después de 50 ciclos de
autoclave.
Con las tapas de rotor QuickLock, el anillo de obturación debe cambiarse después de 50
ciclos de autoclave.
Nunca almacene cerrados los rotores herméticos a los aerosoles o los frascos.
2.7Indicaciones de seguridad en el equipo
RepresentaciónSignificadoLugar
Observe el manual de instrucciones.5418: Parte posterior del
ATENCIÓN
Apriete siempre el rotor con la llave de rotor
adjunta.
ATENCIÓN
Cierre todos los tubos y utilice una tapa de rotor.
equipo
5418 R: Lado derecho del
equipo
Lado superior del equipo,
debajo de la tapa de la
centrífuga.
Lado superior del equipo,
debajo de la tapa de la
centrífuga.
Page 54
54
Instrucciones generales de seguridad
Centrifuge 5418 / 5418 R
Español (ES)
Page 55
3Instalación
3.1Seleccionar ubicación
¡AVISO! En caso de error se pueden dañar objetos que se encuentran junto al equipo.
Respete las recomendaciones de la norma EN 61010-2-020, dejando durante el
funcionamiento un espacio de seguridad de 30 cm alrededor del equipo.
Retire todos los materiales y objetos que se encuentren en esta área.
¡AVISO! Daños por sobrecalentamiento.
No coloque el equipo cerca de fuentes de calor (p.ej., calefacción, armario de desecación).
No exponga el equipo a la radiación solar directa.
Asegúrese de que la circulación de aire no se obstaculice. Mantenga una distancia mínima
de 30 cm de todas las rendijas de ventilación.
Instalación
Centrifuge 5418 / 5418 R
Español (ES)
55
Determine la ubicación del equipo según los siguientes criterios:
• Alimentación eléctrica apropiada conforme a la placa de características (230 V/120 V/100 V).
• Mesa de laboratorio estable, horizontal y libre de resonancias. Peso del equipo: 7,7 kg (5418) ó 22 kg
(5418 R).
• Entorno bien ventilado y protegido de la radiación solar directa para evitar un calentamiento adicional.
3.2Instalación del aparato
¡AVISO! Daños en los componentes electrónicos debido a la formación de condensación.
Después de transportar el equipo de un ambiente frío a un ambiente caliente se puede formar
líquido de condensación en el equipo.
5418: espere por lo menos 3 horas hasta que conecte el equipo al suministro de corriente.
Sólo 5418: Alternativa: Encienda el equipo y deje que se caliente durante una media hora
directamente antes de un transporte de corta distancia.
5418 R: espere por lo menos 4 horas hasta que conecte el equipo al suministro de
corriente.
¡AVISO! Centrífuga 5418 R: daños en el compresor tras un transporte inadecuado.
No conecte la centrífuga hasta 4 horas después de haberla colocado en su respectivo sitio.
Page 56
56
Instalación
Centrifuge 5418 / 5418 R
Español (ES)
Ejecute los siguientes pasos en el orden indicado:
1. Deje que el equipo se caliente a temperatura ambiente.
2. Verifique que la tensión y frecuencia de red coinciden con los requisitos especificados en la placa de
características del equipo.
3. Conecte la centrífuga a la red y enciéndala con el interruptor de red.
• El indicador está activo.
• La tapa se abre automáticamente.
4. Soltar el rotor con ayuda de la llave de rotor adjunta y extraerlo.
5. Retirar el acolchado de seguridad para el transporte del eje motor.
6. Compruebe todas las piezas respecto a posibles daños de transporte. Póngase en contacto con su
proveedor en caso de que exista algún daño.
7. Sólo 5418 R: Coloque el recipiente colector de agua condensada de un lateral del equipo en el soporte
destinado para ello.
Guarde el material de embalaje y los bloqueadores de transporte para un transporte posterior
o almacenaje. Observe también las indicaciones para el transporte .
Page 57
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4Manejo
4.1Visión general de los elementos de control
Antes de utilizar la Centrífuga 5418 / 5418 R, familiarícese con los elementos de control y el indicador.
Abb. 4-1: Panel de control de la Centrífuga 5418 y Centrífuga 5418 R.
57
1
5418
2
45
3
9
8
4 (9)
5418 R
8
Imag. 4-1: Panel de control de la Centrífuga 5418 y Centrífuga 5418 R.
1 Conmutación de la velocidad de centrifugado
indicada (rpm/FCR)
2 Indicador
3 Desbloquear la tapa
4 Ajustar la duración de centrifugado
6 Ajustar la temperatura (solo 5418 R)
7 Iniciar la marcha de calentamiento fast temp
(solo 5418 R)
8 Centrifugado short spin
9 Iniciar y detener el centrifugado
Pulsar el Sólo 5418 R: botón giratorio time para
iniciar y detener la centrifugación.
25 (1)
6
7
3
6
5 Ajustar la velocidad de centrifugado
Pulsar el Sólo 5418 R: botón giratorio speed para
conmutar la indicación de la velocidad de
centrifugación (rpm/FCR).
Page 58
58
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
Abb. 4-2: Indicador de la Centrífuga 5418 y Centrífuga 5418 R.
12
5418
3
7
5418 R
1
Imag. 4-2: Indicador de la Centrífuga 5418 y Centrífuga 5418 R.
1 Duración del centrifugado
2 Número g (rcf) / velocidad (rpm)
5418 R: valor establecido x 1000
3 Estado de la centrífuga
: tapa de centrífuga desbloqueada.
: tapa de centrífuga bloqueada.
(parpadea): centrifugado en marcha.
4 At set rpm (solo 5418 R)
: inicio del ciclo cuando se alcanza el 95% de
la fuerza de gravedad (FCR)/número de
revoluciones (rpm) predeterminado.
: inicio del ciclo inmediato.
5 Temperatura (solo 5418 R)
6 Estado de bloqueo de tecla (solo 5418 R)
: los parámetros de centrifugado no se pueden
cambiar inintencionadamente.
: el teclado no está bloqueado.
7 Estado del indicador de la velocidad de
centrifugación
rcf: número g (fuerza centrífuga relativa, FCR)
rpm: velocidad (revoluciones por minuto)
8 Estado del altavoz (sólo 5418 R)
: conectado.
Ningún símbolo: desconectado.
6
4
5
8
3
2
7
4.2Preparación del centrifugado
4.2.1Conexión de la centrífuga
1. Conexión de la centrífuga con el interruptor de red.
Después de la conexión con el interruptor de red, la tapa de la centrífuga se abre automáticamente.
2. Abra la tapa de la centrífuga pulsando la tecla open.
A continuación aparecen los ajustes de parámetros del último funcionamiento.
4.2.2Inserción del rotor
¡AVISO! El rotor se puede caer en caso de una manipulación errónea.
Los tornillos de la tapa del rotor se pueden soltar, si los utiliza para sujetar el rotor.
Sujete el rotor para moverlo o transportarlo siempre con las dos manos.
1. Coloque el rotor verticalmente en el eje del motor.
2. Inserte la llave de rotor suministrada en la tuerca del rotor.
3. Gire la llave de rotor en el sentido de las agujas del reloj hasta que la tuerca del rotor esté totalmente
apretada.
Page 59
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4.2.3Carga del rotor
Para cargar el rotor, proceda del siguiente modo:
1. Compruebe la carga máxima (adaptador, tubo y contenido) permisible para cada orificio de rotor.
La máxima carga es de 3,75 g por orificio del rotor. Estos datos los encuentra también en el rotor.
2. Cargue el rotor y el adaptador sólo con los tubos previstos para ello.
3. Introduzca los tubos por parejas en posición opuesta en los orificios del rotor. Para una carga simétrica,
los tubos en posición opuesta tienen que ser del mismo tipo y contener la misma cantidad de sustancia.
Para que la diferencia de peso entre los tubos de muestras llenos sea la mínima posible, es
recomendable pesarlos con una báscula. A través de ello se protege el accionamiento y se reducen los
ruidos de funcionamiento.
4.2.4Cierre de la tapa del rotor
Con el rotor FA-45-18-11 se puede centrifugar también sin tapa.
59
En este caso debe observar lo siguiente:
• Las tapas de los tubos tienen que estar cerradas fijamente.
• Sin la tapa el rotor no es hermético a los aerosoles.
• Durante la centrifugación se produce un poco más ruido.
• Las spin columns (columnas de centrifugación) siempre se tienen que centrifugar con la
tapa de rotor puesta.
1. Compruebe el asiento correcto del anillo de obturación exterior
en la respectiva ranura.
2. Coloque la tapa del rotor perpendicularmente al rotor.
3. Cerrar el rotor ajustando bien los tornillos rojos de la tapa del
rotor en el sentido de las agujas del reloj hasta que se escuche un
claro clic.
Sólo después de haber escuchado el clic, el rotor está cerrado
correctamente.
4.2.5Cierre de la tapa de la centrífuga
1. Compruebe la fijación correcta del rotor y de su tapa.
2. Presione la tapa de la centrífuga hacia abajo hasta que el bloqueo de la tapa enganche y la tapa se cierre
automáticamente.
La centrífuga se cierra automáticamente.
En el indicador aparece el símbolo .
Sólo 5418: la tecla open parpadea en color azul.
4.3Refrigeración (sólo 5418 R)
4.3.1Ajuste de temperatura
Ajustar la temperatura con la tecla de flecha temp de 0 °C hasta +40 °C.
Page 60
60
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4.3.2Indicación de temperatura
En caso de parada del rotor:Temperatura teórica
Durante el centrifugadoTemperatura real
4.3.3Supervisión de temperatura
Después de alcanzar la temperatura teórica, la centrífuga reacciona de la siguiente manera a desviaciones
de temperatura que ocurren durante el centrifugado:
Divergencia del valor teóricoAcción
ΔT > 3 °CLa indicación de temperatura parpadea.
ΔT > 5 °CTono de aviso periódico e indicación de Error 18. El
centrifugado finaliza automáticamente.
4.3.4FastTemp
Con esta función, se inicia directamente la marcha de calentamiento sin pruebas con el rotor y con la
velocidad de temperatura específica, para llevar la cámara del rotor, incl. rotor y adaptador, a la
temperatura teórica.
Requisitos
• La centrífuga está encendida.
• El rotor y la tapa del rotor están montados correctamente.
• La tapa de la centrífuga está cerrada.
• La temperatura y el número g (rcf) / velocidad (rpm) está ajustado para la centrifugación posterior (ver
Centrifugado en pág. 61).
1. Pulse la tecla FastTemp.
En el indicador se muestra FT así como la temperatura actual y el número g (rcf) / velocidad (rpm).
El tiempo de enfriamiento desde la temperatura ambiente (≈23 °C) a 4 °C asciende a aprox. 16 min.
La marcha de calentamiento finaliza automáticamente al alcanzar la temperatura teórica. Un tono de
aviso se escucha periódicamente.
2. Pulse la tecla start/stop para finalizar el ciclo de atemperado antes de tiempo.
Después de alcanzar la temperatura teórica y después de finalizado el ciclo de calentamiento, la centrífuga
mantiene la cámara del rotor a la temperatura teórica estando la tapa cerrada, siempre y cuando ésta sea
inferior a la temperatura ambiente. Independientemente de la temperatura teórica, no se llega a exceder
una temperatura de 4 °C durante esta refrigeración continua para evitar que el compartimiento del rotor se
congele.
La centrífuga termina el ciclo automáticamente, cuando el rotor se haya termostatizado
completamente. Por esta razón es posible que se produzca un retardo entre la indicación de la
temperatura teórica alcanzada y el fin automático de la marcha de calentamiento.
Page 61
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4.3.5Refrigeración continua
Cuando el rotor está detenido, el compartimiento del rotor es mantenido a la temperatura teórica, siempre
y cuando se den las siguientes condiciones:
• La centrífuga está encendida.
• La tapa de la centrífuga está cerrada.
• La temperatura teórica es inferior a la temperatura ambiente.
Durante la refrigeración permanente sirve lo siguiente:
• La temperatura teórica es indicada.
• Independientemente de la temperatura teórica, no se llega a exceder una temperatura de 4 °C para
evitar una congelación de la cámara del rotor y de las pruebas, así como una mayor condensación en el
aparato.
• Como el rotor no gira, la adaptación de la temperatura sucede con más lentitud.
Para terminar la refrigeración continua, abra la tapa de centrifugado.
Si la centrífuga no es utilizada durante más de 8 horas, la refrigeración continua se desconecta
automáticamente. Esto protege del hielo a la cámara del rotor y a los tubos, así como una mayor
condensación en el equipo.
61
En el indicador aparece la temperatura real de la cámara del rotor.
Con fast temp puede volver a la temperatura deseada rápidamente (ver en pág. 60).
4.4Centrifugado
Antes de utilizar la Centrífuga 5418 / 5418 R, familiarícese con los elementos de control y los indicadores
(ver Visión general de los elementos de control en pág. 57).
La condición para cada variante de centrífuga aquí descrita es la preparación mencionada anteriormente
(ver Preparación del centrifugado en pág. 58).
Sólo 5418 R: tenga en cuenta también las advertencias para el enfriamiento (ver en pág. 59).
Page 62
62
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4.4.1Centrifugado con ajuste de tiempo
Ejecute los siguientes pasos en el orden indicado:
1. Ajuste con time la duración del ciclo.
2. Sólo 5418 R: ajustar con temp la temperatura.
3. Ajustar con speed el número g (rcf) / velocidad.
4. Pulse start/stop para iniciar el centrifugado.
Durante el centrifugado
• En la indicación, parpadea mientras el rotor esté en marcha.
• Sólo 5418 R: se indica la temperatura actual.
• Las teclas fast temp (solo 5418 R), open y short están bloqueadas durante la centrifugación.
• Durante la marcha, puede modificar el tiempo de ejecución total, la temperatura (solo 5418 R), la
velocidad y el indicador rpm/rcf. Sólo 5418 R: para cambiar los parámetros de centrifugación, pulse
la tecla short.
Durante las modificaciones, los valores del indicador parpadean. Los parámetros nuevos se aceptan
inmediatamente. En caso de modificar la duración durante una marcha, se tiene en cuenta el tiempo
ya transcurrido. Tenga en cuenta que el tiempo total del ciclo nuevo más breve que se puede ajustar
es el tiempo ya transcurrido más 2 minutos.
• También puede acabar el centrifugado antes de que se termine el tiempo establecido pulsando la
tecla start/stop.
Finalización del centrifugado
• Una vez transcurrido el tiempo ajustado, la centrífuga se detiene automáticamente. Durante el
proceso de frenado se indica el tiempo de duración del centrifugado de forma parpadeante. En
cuanto el rotor se haya detenido por completo, se escucha una señal acústica.
• Sólo 5418: la tapa de la centrífuga se abre automáticamente. En el indicador se muestra el símbolo
.
• Sólo 5418 R: la tapa de la centrífuga permanece cerrada para conservar la temperatura de las
muestras. Puede abrir la tapa pulsando la tecla open.
5. Extraiga el material de centrifugado.
4.4.2Otras posibilidades de centrifugación
4.4.2.1 Ciclo permanente
Ajustar con time la marcha permanente.
En el indicador se muestra oo.
Page 63
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
4.4.2.2 Short Spin
5418: puede realizar un centrifugado de corta duración con el máximo número g (rcf) / velocidad (rpm).
5418 R: puede realizar un centrifugado de corta duración con el número g (rcf) / velocidad (rpm)
establecido actualmente o máximo .
Seleccionar el modo Short Spin (solo 5418 R)
Pulsar la tecla short cuando la tapa de la centrífuga esté abierta.
Se muestra el modo actual:
• Indicador 1 – 14 t (marcha con velocidad previamente seleccionada)
• Indicador 14 t (marcha con revoluciones por minuto máximas 14.000 rpm)
Con la tapa de la centrífuga abierta, pulsar la tecla short durante más de 2 segundos para modificar los
modos.
Realizar la centrifugación Short Spin
Iniciar un centrifugado de corta duración con la tapa cerrada manteniendo pulsada la tecla short. Para
detenerlo, suelte la tecla.
63
4.4.3Extracción del rotor
1. Gire la tuerca del rotor con la llave de rotor suministrada en el sentido contrario a las agujas del reloj.
2. Extraiga el rotor sacándolo verticalmente hacia arriba.
3. Sólo 5418 R: Apague la centrífuga después del uso y vacíe el recipiente colector del agua condensada
(jalar hacia afuera por el lado izquierdo o derecho del aparato). Deje la tapa de la centrífuga
completamente abierta y asegúrese de que ésta no pueda cerrarse sola de golpe.
4.5Modo standby (solo 5418)
Si la centrífuga no se ha utilizado durante 15 minutos, ésta cambia al modo standby. En el indicador
aparece entonces EP. Al pulsar una tecla o un mando giratorio, o cerrando la tapa de centrifugado, la
centrífuga se vuelve a activar y está lista para funcionar.
4.6Funciones especiales (solo 5418 R)
FunciónEstado tapa de
centrifugado
Cambiar parámetro
durante la marcha.
Conectar / desconectar
tono de señal.
Conectar / desconectar
bloqueo de tecla.
At set rpm
cerradoshortparpadea 5 seg
abierto
abierto
abierto
Pulsar la tecla > 2 sIndicación
open
short + open
time
/
/
Page 64
64
Manejo
Centrifuge 5418 / 5418 R
Español (ES)
Page 65
Mantenimiento
Centrifuge 5418 / 5418 R
Español (ES)
5Mantenimiento
5.1Preparación de la limpieza / desinfección
El proceso descrito en el capítulo siguiente sirve tanto para la limpieza como también para la desinfección
o descontaminación. En la siguiente tabla se describen los pasos que se requieren adicionalmente:
LimpiezaDesinfección/descontaminación
1. Utilice un producto de limpieza suave para la
limpieza de las superficies accesibles del aparato
y de los accesorios.
2. Realice la limpieza tal y como se describe en el
capítulo siguiente.
1. Seleccione los métodos de desinfección que
correspondan a las disposiciones legales y
directrices vigentes para su campo de
aplicación. Utilice, p.ej., alcohol (etanol,
isopropanol) o desinfectantes que contengan
alcohol.
2. Realice la desinfección o descontaminación tal y
como se describe en el capítulo siguiente.
3. Limpie a continuación el aparato y los
accesorios.
65
En caso de preguntas sobre la limpieza y la desinfección o descontaminación y sobre los
detergentes utilizables, diríjase a Application Support de Eppendorf AG. Las informaciones de
contacto se encuentran en la parte posterior de estas instrucciones.
5.2Realización de la limpieza / desinfección
¡PELIGRO! Descarga eléctrica debido a la penetración de líquidos.
Apague el equipo y desenchúfelo de la alimentación eléctrica antes de empezar los
trabajos de limpieza o de desinfección.
Ningún líquido debe penetrar en el interior de la carcasa.
No efectúe ninguna limpieza o desinfección por pulverización en la carcasa.
Sólo vuelva a conectar el equipo a la red eléctrica si está completamente seco por dentro y
por fuera.
¡AVISO! Daños por sustancias químicas agresivas.
De ninguna manera utilice sustancias químicas agresivas como, por ejemplo, bases fuertes
o débiles, ácidos fuertes, acetona, formaldehídos, hidrógeno halogenado o fenol con el
equipo y sus accesorios.
Limpie el equipo inmediatamente con un producto de limpieza suave en caso de una
contaminación con una sustancia química agresiva.
¡AVISO! Corrosión producida por productos de limpieza y desinfección agresivos.
No utilice productos de limpieza corrosivos ni disolventes agresivos o abrillantadores.
No incube los accesorios durante un tiempo prolongado en productos de limpieza o
desinfectantes agresivos.
Page 66
66
Mantenimiento
Centrifuge 5418 / 5418 R
Español (ES)
¡AVISO! Daños por radiación UV u otro tipo de radiación rica en energía.
No realice ninguna desinfección con radiación UV, beta o gamma ni con ningún otro tipo
de radiación rica en energía.
Evite el almacenamiento en zonas con fuerte radiación ultravioleta.
Autoclave
Los rotores, tapas de rotor y adaptadores pueden ser autoclavados (121 °C, 20 min).
Cambie el anillo de obturación de la ranura de la tapa del rotor hermético a los aerosoles
después de 50 ciclos de autoclave como máximo.
Hermeticidad a los aerosoles
Asegúrese de la integridad de los anillos de obturación antes de la utilización.
Sólo con la tapa de rotor QuickLock: Sustituya el anillo de obturación de la ranura de la tapa
cuando esté desgastado.
Es necesario el cuidado habitual de los anillos de obturación para proteger a los rotores.
No almacene nunca los rotores herméticos a los aerosoles con la tapa cerrada.
Engrase regularmente la rosca de las tapas de los rotores herméticos a los aerosoles con grasa
para gorrones (n.º de pedido int.: 5810 350.050/Norteamérica: 022634330).
5.2.1Limpieza y desinfección del equipo
1. Abra la tapa. Apague el equipo con el interruptor de red. Desenchufe el enchufe del suministro de
corriente.
2. Suelte la tuerca del rotor girándola con la llave de rotor en sentido contrario a las agujas del reloj.
3. Retire el rotor.
4. Limpie y desinfecte todas las superficies accesibles del equipo, incluido el cable de alimentación, con
un paño húmedo y el producto de limpieza recomendado.
5. Enjuague los anillos de obturación de goma de la cámara del rotor minuciosamente con agua.
6. Frote los anillos de obturación de goma secos con glicerina o talco para evitar que se vuelvan
quebradizos. Otros componentes del equipo como, p.ej., el bloqueo de la tapa, el eje del motor y el cono
del rotor no se pueden engrasar.
7. Limpie el eje del motor con un paño suave, seco y que no deje pelusas. No engrase el eje del motor.
8. Compruebe que el eje del motor no presente daños.
9. Controle que el equipo no presente daños ni corrosión.
10.Deje la tapa de la centrífuga abierta cuando no se utilice el equipo.
11.Sólo vuelva a conectar el equipo a la red eléctrica cuando éste esté completamente seco por dentro y
por fuera.
5.2.2Limpieza y desinfección del rotor
1. Controle que el rotor y los accesorios no presenten daños ni corrosión. No utilice rotores ni accesorios
dañados.
2. Limpie y desinfecte los rotores y accesorios con los productos de limpieza recomendados.
3. Limpie y desinfecte los orificios del rotor con un cepillo para botellas.
4. Para la limpieza y desinfección de la tapa del rotor debe retirar el anillo de obturación. Limpie también
la ranura de debajo y el anillo de obturación.
Page 67
Mantenimiento
Centrifuge 5418 / 5418 R
Español (ES)
5. Aclare los rotores y accesorios cuidadosamente con agua destilada. Aclare cuidadosamente los orificios
de los rotores de ángulo fijo.
No sumerja el rotor, ya que puede entrar líquido.
67
6. Deje secar los rotores y accesorios sobre un paño. Coloque los rotores de ángulo fijo con las
perforaciones hacia abajo para que las perforaciones también se puedan secar.
7. Coloque correctamente el anillo de obturación de la tapa del rotor de nuevo en la ranura limpia y seca.
8. Limpie el cono del rotor con un paño suave, seco y que no deje pelusas. No engrase el cono del rotor.
9. Compruebe que el cono del rotor no presente daños.
10.Coloque el rotor seco en el eje del motor.
11.Apriete la tuerca del rotor girándola con la llave de rotor en el sentido de las agujas del reloj.
12.Deje la tapa del rotor abierta cuando no se utilice el rotor.
5.3Indicaciones de cuidado adicionales para la Centrífuga 5418 R
Vacíe y limpie con regularidad el recipiente colector de agua condensada y especialmente después de
derrames de líquido en la cámara del rotor. Extraiga el recipiente colector de agua condensada que se
encuentra en el lado izquierdo o derecho de la centrífuga.
Retire con regularidad la formación del hielo de la cámara del rotor mediante descongelación, dejando
abierta la tapa de la centrífuga o realizando un ciclo de atemperado breve a aprox. 30 °C.
Deje la tapa de la centrífuga abierta cuando no la vaya a utilizar durante un largo periodo de tiempo.
Limpie el agua de condensación de la cámara del rotor. Utilice para ello un paño absorbente suave.
Elimine el polvo de las ranuras de ventilación de la centrífuga como máximo cada medio año, utilizando
un pincel o una escobilla. Antes de la limpieza, apague la centrífuga y extraiga el enchufe de la toma de
corriente.
5.4Rotura de vidrio
Al utilizar tubos de vidrio se pueden producir roturas en la cámara del rotor. Las astillas de vidrio
producidas se arremolinan durante el centrifugado en la cámara del rotor y afectan al rotor y los accesorios
como un chorro de arena. Las partículas de vidrio más pequeñas se acumulan en las piezas de goma (p.ej.
en el manguito del motor, el anillo de obturación de la cámara del rotor y las esterillas de goma de los
adaptadores).
Page 68
68
Mantenimiento
Centrifuge 5418 / 5418 R
Español (ES)
¡AVISO! Rotura de vidrio en la cámara del rotor
Con una fuerza de la gravedad elevada pueden romperse los tubos de vidrio en la cámara del
rotor. La rotura de vidrio provoca daños en el rotor, los accesorios y las muestras.
Tenga en cuenta las indicaciones del fabricante de los recipientes acerca de los parámetros
de centrifugado recomendados (carga y número de revoluciones).
Consecuencias de la rotura de vidrio en la cámara del rotor:
• Abrasión fina y negra en la cámara del rotor (con recipiente del rotor de metal).
• Las superficies de la cámara del rotor y los accesorios se arañan.
• La resistencia a los agentes químicos de la cámara del rotor se reduce.
• Impurezas de las muestras.
• Abrasión en las piezas de goma.
Comportamiento ante la rotura de vidrio
1. Retire las astillas y el vidrio pulverizado de la cámara del rotor y de los accesorios.
2. Limpie el rotor y la cámara del rotor exhaustivamente. Limpie con especial atención los orificios de los
rotores de ángulo fijo.
3. En caso necesario, sustituya los adaptadores para evitar posteriores deterioros.
4. Compruebe regularmente si hay residuos o daños en los orificios del rotor.
5.5Fusibles
El portafusible se encuentra debajo del conector hembra de la red eléctrica (5418) y/o a la izquierda del
interruptor principal (5418 R).
1. Extraiga el conector de red.
2. Extraiga el portafusible.
Ahora ambos fusibles son accesibles y pueden ser sustituidos.
5.6Descontaminación antes del envío
Cuando envíe el equipo en caso de reparación al servicio técnico autorizado o en el caso de eliminación del
mismo a su concesionario, tenga en cuenta lo siguiente:
¡ADVERTENCIA! Riesgo para la salud debido a la contaminación del equipo.
1. Tenga en cuenta las indicaciones del certificado de descontaminación. Encontrará estas
indicaciones como fichero PDF en nuestra página de Internet (
decontamination).
2. Descontamine todas las piezas que desee enviar.
3. Adjunte al envío el certificado de descontaminación completamente rellenado.
www.eppendorf.com/
Page 69
Solución de problemas
Centrifuge 5418 / 5418 R
Español (ES)
6Solución de problemas
Si no puede solucionar el error con las medidas sugeridas, póngase en contacto con su socio local de
Eppendorf. Encontrará la dirección de contacto en Internet bajo
www.eppendorf.com/worldwide.
6.1Errores generales
Síntoma/mensajeCausaAyuda
Ningún dato.Ninguna conexión a la red.
Ningún dato.Caída de corriente.
La tapa del equipo
no se puede abrir.
La tapa del equipo
no se puede abrir.
El equipo no se
deja arrancar.
El equipo da
sacudidas al
arrancar.
La centrífuga se
detiene durante un
centrifugado de
corta duración,
aunque la tecla
short esté pulsada.
La indicación de
temperatura
parpadea.
(sólo 5418 R)
El rotor todavía gira.
Fallo de la tensión de red.1. Comprobar el fusible de la red
Tapa del equipo sin cerrar.
El rotor se ha cargado asimétricamente.1. Parar el equipo y cargar de forma
La tecla short ha sido soltada más de dos
veces de forma breve (función de
protección para el accionamiento).
Divergencia de la temperatura del valor
teórico: ±3 °C.
Controlar la conexión de la red de
distribución.
Comprobar el fusible de la red
eléctrica del equipo (ver Fusibles en
pág. 68).
Comprobar el fusible de la red
eléctrica del laboratorio .
Esperar hasta que se pare el rotor.
eléctrica del equipo (ver Fusibles en
pág. 68).
2. Comprobar el fusible de la red
eléctrica del laboratorio .
3. Accionar el desbloqueo de
emergencia de la tapa (ver en
pág. 72).
Cierre la tapa del equipo.
simétrica.
2. Iniciar de nuevo el equipo.
Mantenga pulsada la tecla short
durante un centrifugado de corta
duración.
Comprobar el ajuste.
Esperar hasta que se alcance la
temperatura ajustada.
Comprobar que el aire circule
libremente a través de las ranuras de
ventilación.
Descongelar el hielo o apagar el
equipo y dejar que se enfríe.
69
Page 70
70
Solución de problemas
Centrifuge 5418 / 5418 R
Español (ES)
6.2Mensajes de error
Cuando aparecen los siguientes mensajes de error, debe proceder del siguiente modo:
1. Eliminar error (v. ayuda).
2. Si es necesario, repetir el centrifugado.
Síntoma/mensajeCausaAyuda
LID ERROR5418)/
LID (5418 R)
LID ERROR5418)/
LID (5418 R)
LID ERROR5418)/
LID (5418 R)
INTInterrupción de la red de alimentación
NO RPM (5418)/
Error 3 (5418 R)
Error 5
(sólo 5418 R)
Err 6 (5418)/
Error 6 (5418 R)
Error 6
(sólo 5418 R)
Err 7 (5418)/
Error 7 (5418 R)
Err 8 (5418)/
Error 8 (5418 R)
Err 9 hasta 17
(5418)/
Error 9 hasta 17 (5
418 R)
La tapa de la centrífuga no se puede
bloquear.
La tapa de la centrífuga no se puede
desbloquear.
La tapa de la centrífuga no se debe
desbloquear durante un ciclo.
Cerrar de nuevo la tapa de la
centrífuga.
1. Desconectar la centrífuga y
conectarla de nuevo.
2. Pulsar open.
En caso de repetirse el error:
1. Desconectar la centrífuga.
2. Accionar el desbloqueo de
emergencia de la tapa (ver
Desbloqueo de emergencia de la tapa
en pág. 72).
Esperar hasta que se pare el rotor.
Comprobar la conexión de la red de
durante un ciclo.
Error en el sistema medidor de velocidad
o accionamiento sobrecalentado.
Apertura de tapa impermisible o
interruptor de la tapa defectuoso durante
un ciclo.
Error de accionamiento
distribución.
Dejar la centrífuga encendida hasta
que desaparezca el error (10 s o 6
min).
1. Esperar hasta que se pare el rotor.
2. Abrir la tapa de la centrífuga y volver
a cerrarla.
3. Repetir el ciclo.
Repetir el ciclo.
En caso de un nuevo mensaje de la
centrífuga, apagar y volver a
encender después de 20 segundos.
Accionamiento sobrecalentado.
Una desviación considerable en el
control de revoluciones.
Error de accionamiento1. Esperar hasta que se pare el rotor.
Error del sistema electrónico.
Dejar enfriar el accionamiento 15
min. como mínimo.
1. Esperar hasta que se pare el rotor.
2. Apretar el rotor.
2. Repetir el ciclo.
Desconectar la centrífuga y
conectarla de nuevo después de >20
s.
Page 71
Síntoma/mensajeCausaAyuda
Error 18
(sólo 5418 R)
Desviación de la temperatura del valor
teórico demasiado grande en la cámara
del rotor.
Comprobar los ajustes.
Comprobar que el aire circule
libremente a través de las ranuras de
ventilación.
Descongelar el hielo o apagar el
equipo y dejar que se enfríe.
Error 19
(sólo 5418 R
Error 20
(sólo 5418 R)
Error 21
(sólo 5418 R)
Error 24
(sólo 5418 R)
El círculo de refrigeración está
sobrecalentado.
El sensor de temperatura en la cámara
del rotor está defectuoso.
El sensor de temperatura en el
condensador está defectuoso.
Error en el grupo de refrigeración, p.ej.
sobrecalentamiento.
Compruebe que el aire circule
libremente a través de las ranuras de
ventilación y deje que el equipo se
enfríe.
Desconectar la centrífuga y
conectarla de nuevo después de >20
s.
Desconectar la centrífuga y
conectarla de nuevo después de >20
s.
Dejar que la centrífuga se enfríe y
repetir el ciclo.
Solución de problemas
Centrifuge 5418 / 5418 R
Español (ES)
71
Page 72
72
Solución de problemas
Centrifuge 5418 / 5418 R
Español (ES)
6.3Desbloqueo de emergencia de la tapa
Si la tapa de la centrífuga no se deja abrir, usted puede accionar el desbloqueo de emergencia de la tapa
manualmente.
¡ADVERTENCIA! Peligro de lesión debido a la rotación del rotor.
Espere a que el rotor se detenga antes de pulsar el desbloqueo de emergencia.
Como medida de control, mire a través del tubito de control en la tapa de la centrífuga.
6.3.1Centrífuga 5418
1. Extraiga el conector de red.
2. Suelte la cubierta de plástico del desbloqueo de emergencia de la tapa.
Éste se encuentra en la parte trasera izquierda del pie en la placa del fondo (Fig. 1-1 en pág. 45).
3. Tire hacia abajo el cordel perpendicularmente.
La tapa de la centrífuga se desbloquea.
4. Antes de cerrar la centrífuga con la tapa de nuevo, arrastre el cordel por completo en la carcasa y ponga
la tapa de plástico en las placas del fondo.
6.3.2Centrífuga 5418 R
1. Extraiga el conector de red.
2. Gire la cubierta de plástico del desbloqueo de emergencia de la tapa 90° en sentido antihorario
utilizando una herramienta adecuada (p.ej. un destornillador) y extráigala.
La cubierta de plástico se encuentra en el lado derecho del aparato (Fig. 1-2 en pág. 45).
3. Introduzca la llave de rotor de la centrífuga en la abertura hexagonal situada directamente detrás hasta
que perciba una resistencia.
4. Gire la llave de rotor en sentido horario manteniéndola ligeramente presionada.
A través de ello se desbloquea la tapa de la centrífuga.
5. Abra la tapa de la centrífuga.
6. Retire la llave del rotor y vuelva a colocar la cubierta de plástico girándola 90° en sentido horario.
Page 73
Transporte, almacenaje y eliminación
Centrifuge 5418 / 5418 R
Español (ES)
7Transporte, almacenaje y eliminación
7.1Transporte
¡ATENCIÓN! Daños personales al levantar y transportar cargas pesadas
El equipo es pesado. Elevar y cargar el equipo puede causar dolores de espalda.
Transporte el equipo con ayuda de otra persona.
Utilice una ayuda de transporte para transportes de largo recorrido (p.ej., una carretilla).
Saque el rotor de la centrífuga antes del transporte.
Utilice el embalaje original para el transporte.
Temperatura del aireHumedad relativaPresión atmosférica
Temperatura del aireHumedad relativaPresión atmosférica
en embalaje de
transporte
sin embalaje de
transporte
-25 °C – 55 °C10 % – 75 %70 kPa – 106 kPa
-5 °C – 45 °C10 % – 75 %70 kPa – 106 kPa
7.3Eliminación
Observe las disposiciones legales correspondientes al eliminar el producto.
Información para la eliminación de aparatos eléctricos y electrónicos en la Comunidad Europea:
En la Comunidad Europea la eliminación de aparatos eléctricos es regulada por reglamentaciones
nacionales que se basan en la Directiva comunitaria 2002/96/CE relativa a residuos de aparatos eléctricos y
electrónicos (RAEE).
Según esta directiva, todos los aparatos suministrados después del 13/08/2005 en el área de comercio de
empresa a empresa, al que este producto pertenece indudablemente, no pueden eliminarse conjuntamente
con la basura comunal o doméstica. Para documentar esto, estos aparatos han sido identificados con los
siguientes símbolos:
Dado que las normas de eliminación dentro de la UE pueden variar de país a país, le rogamos contactar con
sus proveedores en caso de necesidad.
Page 74
74
Transporte, almacenaje y eliminación
Centrifuge 5418 / 5418 R
Español (ES)
Page 75
Page 76
Page 77
Page 78
Page 79
Page 80
Evaluate your manual
Give us your feedback.
www.eppendorf.com/manualfeedback
Your local distributor: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany
eppendorf@eppendorf.com · www.eppendorf.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.