Emerson CMF400, CMF025, CMF010, CMFHC2, CMFHC3 Installation Manual

...
0 (0)
Installation Manual
20002158, Rev DM
January 2018
Micro Motion® ELITE® Coriolis Flow and Density Sensors
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available on the internet at www.emerson.com or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
Contents
Contents
Chapter 1 Planning ...........................................................................................................................1
1.1 Installation checklist .......................................................................................................................1
1.2 Best practices .................................................................................................................................3
1.3 Temperature limits .........................................................................................................................3
1.4 Recommendations for hygienic and self-draining applications ....................................................... 6
Chapter 2 Mounting .........................................................................................................................8
2.1 Recommendations for lifting heavy meters .................................................................................... 8
2.2 Mount the sensor ......................................................................................................................... 10
2.3 Rotate junction box or 800 core processor (optional) ................................................................... 11
2.4 Mount electronics of high-temperature sensors ...........................................................................12
2.5 Mount a CMF010 sensor to a wall or pole ..................................................................................... 16
2.6 Mount a CMFS007, CMFS010 or CMFS015 sensor in a bracket ......................................................17
2.7 Mount a CMFS025, CMFS040 or CMFS050 sensor in a wall mount bracket ................................... 18
2.8 Secure wafer-style process connections ....................................................................................... 19
2.9 Attach extended electronics .........................................................................................................20
Chapter 3 Transmitter power and I/O wiring .................................................................................. 22
3.1 Options for wiring ........................................................................................................................ 22
3.2 Connect 4-wire cable ....................................................................................................................23
3.3 Connect the 9-wire cable ..............................................................................................................27
Chapter 4 Grounding ......................................................................................................................28
Chapter 5 Supplementary information ........................................................................................... 29
5.1 Purge the sensor case ...................................................................................................................29
5.2 Pressure relief ...............................................................................................................................32
Installation Manual i
Contents
ii Micro Motion ELITE
1 Planning
Topics covered in this chapter:
Installation checklist
Best practices
Temperature limits
Recommendations for hygienic and self-draining applications
1.1 Installation checklist
Make sure that the hazardous area specified on the approval tag is suitable for the environment in which the meter will be installed.
Verify that the local ambient and process temperatures are within the limits of the meter.
If your sensor has an integral transmitter, no wiring is required between the sensor and transmitter. Follow the wiring instructions in the transmitter installation manual for signal and power wiring.
If your transmitter has remote-mounted electronics, follow the instructions in this manual for wiring between the sensor and the transmitter, and then follow the instructions in the transmitter installation manual for power and signal wiring.
Planning
Maximum lengths for Micro Motion cableTable 1-1:
Cable type To transmitter Maximum length
Micro Motion 9-wire 9739 MVD transmitter 1000 ft (300 m)
All other MVD transmit­ters
Micro Motion 4-wire All 4-wire MVD transmit-
ters
Maximum lengths for user-supplied 4-wire cableTable 1-2:
Wire function Wire size Maximum length
Power (VDC) 22 AWG (0,35 mm2) 300 ft (90 m)
20 AWG (0,5 mm2) 500 ft (150 m)
18 AWG (0,8 mm2) 1000 ft (300 m)
Signal (RS-485) 22 AWG (0,35 mm2) or larger 1000 ft (300 m)
For optimal performance, install the sensor in the preferred orientation. The sensor will work in any orientation as long as the flow tubes remain full of process fluid.
60 ft (20 m)
- 1000 ft (300 m) without Ex-approval
- 500 ft (150 m) with IIC rated sensors
- 1000 ft (300 m) with IIB rated sensors
Installation Manual 1
Planning
Preferred sensor orientationTable 1-3:
Process
Liquids
Gases
Preferred orienta-
tion
Alternate orientations
Slurries
Install the meter so that the flow direction arrow on the sensor case matches the actual forward flow of the process. (Flow direction is also software-selectable.)
2 Micro Motion ELITE
1.2 Best practices
The following information can help you get the most from your sensor.
There are no pipe run requirements for Micro Motion sensors. Straight runs of pipe
upstream or downstream are unnecessary.
If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward
through the sensor. Gases should flow downward.
Keep the sensor tubes full of process fluid.
For halting flow through the sensor with a single valve, install the valve downstream
from the sensor.
Minimize bending and torsional stress on the meter. Do not use the meter to align
misaligned piping.
The sensor does not require external supports. The flanges will support the sensor in
any orientation. (Some sensor models installed in very small, flexible pipeline have optional installation instructions that allow for external supports.)
Planning
1.3 Temperature limits
Sensors can be used in the process and ambient temperature ranges shown in the temperature limit graphs. For the purposes of selecting electronics options, temperature limit graphs should be used only as a general guide. If your process conditions are close to the gray area, consult with your Micro Motion representative.
Note
In all cases, the electronics cannot be operated where the ambient temperature is
below –40 °F (–40 °C) or above +140 °F (+60 °C). If a sensor is to be used where the ambient temperature is outside of the range permissible for the electronics, the electronics must be remotely located where the ambient temperature is within the permissible range, as indicated by the shaded areas of the temperature limit graphs.
Temperature limits may be further restricted by hazardous area approvals. Refer to the
hazardous area approvals documentation shipped with the sensor or available at
www.emerson.com.
The extended-mount electronics option allows the sensor case to be insulated without
covering the transmitter, core processor, or junction box, but does not affect temperature ratings. When insulating the sensor case at elevated process temperatures (above 140 °F), please ensure electronics are not enclosed in insulation as this may lead to electronics failure.
For the CMFS007 sensor, the difference between the process fluid temperature and the
average temperature of the case must be less than 210°F (99°C)
Installation Manual 3
Planning
Ambient and process temperature limits for ELITE CMFS007, CMFS025–CMFS150 meters
B
113 (45)
Tamb
–40 (–40)
140 (60)
140 (60)
A
B
–148 (–100)
–58 (–50) 400 (204)
T
= Ambient temperature °F (°C)
amb
T
= Process temperature °F (°C)
proc
A = All available electronic options
B= Remote mount electronics only
Ambient and process temperature limits for ELITE CMF***M/L/H/P (excludes special order cryogenic modifications) and CMFS010-015 meters
140 (60)
Tamb
Tproc
A
140 (60)
B
113 (45)
–40 (–40)
B
–148 (–100)
T
= Ambient temperature °F (°C)
amb
T
= Process temperature °F (°C)
proc
A = All available electronic options
B= Remote mount electronics only
4 Micro Motion ELITE
–400
(–240)
Tproc
400
(204)
Planning
Ambient and process temperature limits for special order cryogenic ELITE meters
140 (60)
Tamb
–40 (–40)
A
B
–148 (–100)
–400
(–240)
T
= Ambient temperature °F (°C)
amb
T
= Process temperature °F (°C)
proc
Tproc
A = All available electronic options
B= Remote mount electronics only
Ambient and process temperature limits for high temperature ELITE meters
140 (60)
176
(80)
Tamb
–40 (–40)
–148 (–100)
T
= Ambient temperature °F (°C)
amb
T
= Process temperature °F (°C)
proc
A = All available electronic options
B= Remote mount electronics only
–58
(–50)
A
B
Tproc
662
(350)
Installation Manual 5
Planning
Ambient and process temperature limits for Super Duplex ELITE meters
140 (60)
Tamb
–40 (–40)
140 (60)
A
B
113 (45)
B
–148 (–100)
–40 (–40) 400 (204)
T
= Ambient temperature °F (°C)
amb
T
= Process temperature °F (°C)
proc
A = All available electronic options
B= Remote mount electronics only
Note
For Super Duplex models operating above 177 °C, please consult the factory before purchase.
Tproc
1.4
Recommendations for hygienic and self­draining applications
CMFS sensors are certified EHEDG TYPE EL CLASS I for hygienic applications when installed vertically with the process fitting and gasket combinations listed in the Position Paper of the EHEDG Test Methods Subgroup (available at http://www.ehedg.org). Other process connections/gasket combinations may be used provided they have been evaluated and successfully tested for in-place cleanability according to the latest edition of EHEDG Document 2. Refer to the ELITE Product Data Sheet for further information about fitting options.
For optimal cleanability and drainability:
If possible, install the sensor in a vertical pipeline with the process fluid flowing
upward through the sensor.
If the sensor must be installed in a horizontal pipeline, drainage is accomplished by
air purge evacuation of the pipeline circuit.
For clean-in-place (CIP) applications, Micro Motion recommends using the
generally-accepted flow velocity of at least 1.5 m/s for cleaning the sensor.
The gap between the electronics housing and sensor body should be inspected
periodically. Manually clean this gap when necessary.
6 Micro Motion ELITE
Installation for self-draining applicationsFigure 1-1:
A
B
C
A. Process pipeline B. Direction of normal process flow C. Direction of drainage
Planning
Installation Manual 7
Mounting
2 Mounting
Topics covered in this chapter:
Recommendations for lifting heavy meters
Mount the sensor
Rotate junction box or 800 core processor (optional)
Mount electronics of high-temperature sensors
Mount a CMF010 sensor to a wall or pole
Mount a CMFS007, CMFS010 or CMFS015 sensor in a bracket
Mount a CMFS025, CMFS040 or CMFS050 sensor in a wall mount bracket
Secure wafer-style process connections
Attach extended electronics
2.1 Recommendations for lifting heavy meters
Heavy meters (those over 50 lb. [22 kg]), and even lighter meters that must be installed in elevated or difficult-to-reach places, usually require additional consideration when transporting or lifting them into their installation location.
Safe handling during transportation and installation is the responsibility of the
installer – know and follow all safety practices and regulations for your facility and for any lifting/rigging equipment being used
A professional rigging crew with proper equipment should be used
Typical equipment for handling heavy meters includes the following:
- Fixed hoist boom trucks or cranes
- Continuous web belt slings
- Eye to eye web belt slings
- Two leg wire rope slings
Lift a meter by its case.
Do not lift a meter by its electronics (junction box, transmitter, or any electrical
fittings) or by its purge fittings
It may be useful to identify the meter center of gravity
Protect the sealing surfaces on the process fittings with factory-installed flange
protectors or comparable field-installed protection
8 Micro Motion ELITE
Loading...
+ 26 hidden pages