Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Rosemount Analytical
representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to
the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Look-alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent electrical shock
and personal injury.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc.
Alconox is a registered trademark of Alconox, Inc.
SNOOP® is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc.
6565P Davis Industrial Parkway
Solon, OH 44139
T 440 914 1261
T 800 433 6076
F 440 914 1262
E gas.csc@emerson.com
Instruction Manual
748214-V
Model 951C
June 2009
HIGHLIGHTS OF CHANGES
EFFECTIVE JUNE 10, 2009
PAGE SUMMARY
Cover Updated photo, revision number, and release date
P-6 Removed Compliance section
1-1 Added Mid Range
1-2 Added Mid Range
3-1 Added Mid Range, Backlight over range feature, and a note, using the MID
range, set the pressure regulator to 2.5 psig.
3-2 Added a photo of the Signal Board, Added a photo of the Signal Board Test
Points, removed the signal board diagnostic jumpers (Figure 3-4)
3-3 Removed the Hi/Lo Range Select Jumper (Figure 3-5), Updated the PPM Range
selection switch photo and table, removed the decimal point selection jumpers
(Figure 3-7), added Configuration Jumper Settings (Figure 3-7)
3-6 Added step 3 to the Zero Calibration Section (In the Mid Range Configuration,
Set the Sample Pressure Regulator to 2.5 psig), Added step 3 to the span calibration section (In the Mid Range Configuration, Set the Sample Pressure Regulator to 2.5 psig), Added step 6 to the span calibration section (Use R25 and R43
for fine adjustments to the display and recorder output.)
4-3 Updated the Signal Conditioning and Display section to reflect the current PCB’s,
updated the Circuit function section.
4-5 Updated the Analyzer Signal Conditioning Block Diagram
6-1 Updated the MATRIX
6-3 Updated the Replacement Parts List
List of Drawings Added Drawing 1-1, 1-2, and1-3
Back Cover Updated Company addresses and contact information
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Figure 6-4. Case Heater Temperature Control Assembly........................................................... 6-6
Model 951C
LIST OF ILLUSTRATIONS
LIST OF TABLES
Table 3-1. Resistance of Converter Temperature Sensor vs. Temperature .....................................3-7
LIST OF DRAWINGS
Drawing 1-1 Panel Cutout / Installation Drawing
Drawing 1-2 Flow Diagram, Lo/Mid Range
Drawing 1-3 Flow Diagram, Hi Range
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951C
Instruction Manual
748214-V
June 2009
Rosemount Analytical Inc. A Division of Emerson Process Management Contents v
Instruction Manual
Model 951C
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the 951C NOX Analyzer.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
748214-V
June 2009
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748214-V
June 2009
Model 951C
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be
impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the this equipment should be
thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified personnel.
This instrument was shipped from factory set up to operate on 115 volt 50/60 Hz. For operation on
230 volt 50/60 Hz, refer to Section 2-3.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power.
WARNING
INTERNAL ULTRAVIOLET LIGHT HAZARD
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at
the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 951C
WARNING
TOXIC CHEMICAL HAZARD
This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and
lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent
odor.
The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and
may contain other constituents of the sample gas which may be toxic. Such gases include various
oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion
reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of
consciousness, and death.
Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and
exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust
products at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. See Section 2-8 for Leak Test Procedure.
Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon
line. Check vent line and connections for leakage.
748214-V
June 2009
WARNING .
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
WARNING.
HIGH PRESSURE GAS CYLINDERS
This instrument requires periodic calibration with a known standard gas. See Sections 2-5 and 3-3.
See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748214-V
June 2009
WARNING.
TOXIC AND OXIDIZING GAS HAZARD
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure.
Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled.
HANDLE LAMP ASSEMBLY WITH EXTREME CARE
If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
•
Immediately clean up and dispose of the mercury spill and lamp residue as follows:
•
Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump
and aspirator bottle with long capillary tube. Alternatively, a commercially available mercury
spill clean-up kit, such as J. T. Baker product No. 4439-01, is recommended.
•
Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all
mercury, lamp residue and debris into a plastic bottle which can be tightly capped. Label and
return to hazardous material reclamation center.
Model 951C
•
Do not place in trash, incinerate or flush down sewer.
•
Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and
sulfur dust.
WARNING.
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left
side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise)
orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the
latch; however, further extraction must be done very carefully to insure the chassis does not fall
out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it
must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
748214-V
Model 951C
June 2009
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather
and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or
placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be
permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748214-V
June 2009
Model 951C
DOCUMENTATION
The following 951C instruction materials are available. Contact Customer Service Center or the local representative to order.
748214 Instruction Manual (this document)
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951C
CONDENSED STARTUP AND CALIBRATION PROCEDURE
Instruction Manual
748214-V
June 2009
The following summarized instructions on
startup and calibration are intended for operators already familiar with the analyzer.
For initial startup, refer to detailed instructions
provided in Section 3.
1. Review the Purchase Order from the 951C
Analyzer. Make a note of the Range that
was purchase (Low Range, Mid Range, or
High Range).
2. Set Range switch on the Signal Conditioning
Board to position 4, 250ppm, 500ppm, or
2500 ppm (see Figure 3-2, page 3-2).
3. On the Signal Conditioning Board, Verify that
Hi/Mid/Lo Range Configuration Jumpers are
installed for the Range that was purchase
(Figure 3-7, page 3-3)
4. Apply power to the analyzer. The analyzer
will now require approximately one to two
hours for temperature equilibrium before being ready for calibration.
5. Verify that the pressure regulator on the cylinder of zero gas (nitrogen or air) or sample
gas is set for supply pressure of 10 to 17
psig.
8. Establish correct pressure of zero gas:
a. Supply zero gas to rear panel SAM-
PLE inlet and set to 15 psig.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in
Step 6b. If not, adjust output pressure regulator on the zero gas cylinder as required.
7. Establish correct pressure of upscale
standard gas:
a. Supply upscale standard gas to rear
panel SAMPLE inlet.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in
Step 7b. If not, adjust output regulator on cylinder of upscale standard
gas as required.
NOTE
Supply pressure for sample, upscale
standard gas and zero air must be the
same. If not, the readout will be in error.
8. Zero Calibration:
6. Verify that the pressure regulator on the cylinder of air (ozonator supply) is set for supply
pressure of 20 to 25 psig.
7. Establish correct pressure of sample gas:
a. Supply sample gas to rear-panel SAM-
PLE inlet at 10 to 17 psig (normally 15
psig).
b. Adjust SAMPLE Back Pressure Regula-
tor so that SAMPLE Pressure Gauge indicates the value appropriate to the
desired operating range (normal operating pressure is 3 to 5 psig). Using the
Mid range setting, set the sample pressure regulator to 2.5 psig See Figure
3-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7
a. Set PPM RANGE Switch (Figure
3-6, page 3-3) for range to be used
for sample analysis. Set the front
panel RANGE1 Control at normal
operating setting, if known, or at
about middle or the range if normal
setting is not known.
b. Supply zero gas to rear panel SAM-
PLE inlet.
c. Adjust front panel ZERO Control for
reading of zero on meter or recorder.
9. Upscale Calibration:
a. Set PPM RANGE Switch (Figure
3-6, page 3-3) at setting appropriate
to the particular span gas.
Instruction Manual
748214-V
June 2009
Model 951C
b. Supply upscale standard gas of accu-
rately known NOX content to rear panel
SAMPLE inlet.
c. Adjust front panel RANGE1 Control so
that reading on meter or recorder is
equal to the know parts-per-million concentration of NOX in the span gas.
d. Adjust R25 on the signal board so that
the Display value and the recorder output are equal.
NOTE
It is the responsibility of the user to
measure efficiency of the NO2-to-NO
converter during initial startup, and
thereafter at intervals appropriate to the
application, normally once a month.
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 951C
Instruction Manual
748214-V
June 2009
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 OVERVIEW
The Model 951C NOX Analyzer is designed to
measure NOX using one of three sets of ranges
designated as Hi, Mid, or Lo. The Hi Range set
consists of spans with ranges of 0-100, 0-250,
0-1000, and 0-2500 ppm NOX. The Mid Range
set consists of spans with ranges of 0-20, 0-50,
0-200, and 0-500 ppm NOX. The Lo Range set
consists of spans with ranges of 0-10, 0-25,
0-100, and 0-250 ppm NOX.
The NOX analyzer continuously analyzes a flowing gas sample for NOX [nitric oxide (NO) plus
nitrogen dioxide (NO2)]. The sum of the concentrations is continuously reported as NOX.
The analyzer is based on the chemiluminescence method of NO detection. The sample is
continuously passed through a heated bed of
vitreous carbon, in which NO2 is reduced to NO.
Any NO initially present in the sample passes
through the converter unchanged, and any NO
is converted to an approximately equivalent
(95%) amount of NO.
The case heater assembly of the Model
951C maintains the internal temperature at
approximately 50oC (122oF).
1-2 TYPICAL APPLICATIONS
The Model 951C Analyzer has specific applications in the following areas:
•Oxides of nitrogen (NOX)
emissions from the combustion
of fossil fuels in:
• Vehicle engine exhaust
• Incinerators
• Boilers
• Gas appliances
2
• Nitric acid plant emissions
• Turbine exhaust
The NO is quantitatively converted to NO2 by
gas-phase oxidation with molecular ozone produced within the analyzer from air supplied by
an external cylinder. During this reaction, approximately 10% of the NO2 molecules are elevated to an electronically excited state, followed
by immediate decay to the non-excited state,
accompanied by emission of photons. These
photons are detected by a photomultiplier tube,
which in turn generates a DC current proportional to the concentration of NOX in the sample
stream. The current is then amplified and used
to drive a front panel display and to provide potentiometric and isolated current outputs.
To minimize system response time, an internal
sample-bypass feature provides high-velocity
sample flow through the analyzer.
The Backlight blinks when the analyzer is 10%
or more over-range. Selecting a less sensitive
(higher) range restores the display function.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 1-1
•Ammonia in pollution control
equipment (with converter)
•Nitric oxide emissions from
decaying organic material (i.e.,
landfills).
Instruction Manual
748214-V
June 2009
1-3 SPECIFICATIONS
Ranges
LO.........................................
MID.......................................
HI.........................................
Repeatability.................................. within 0.1 ppm or ±1% of fullscale, whichever is greater
Zero/Span Drift .............................. less than ±0.1 ppm or ±1% of fullscale, whichever is greater, in 24
Response Time
(Electronic + Flow) ..............
Sensitivity ...................................... less than 0.1 ppm or 1% of fullscale, whichever is greater
Detector Operating Pressure......... atmospheric
Total Sample Flow Rate ................ 1 Liter per minute at 20 psig
Power Requirements ..................... 115/230 VAC ±10%, 50/60 ±3 Hz, 570 W maximum
Enclosure....................................... General purpose for installation in weather-protected areas
Dimensions.................................... 8.7 in. x 19.0 x 19.0 in. (H x W x D)
Weight ........................................... 22.2 kg (49 lbs) approximate
Model 951C
0 to 10, 0 to 25, 0 to 100, 0 to 250 ppm NO
0 to 20, 0 to 50, 0 to 200, 0 to 500 ppm NO
0 to 100, 0 to 250, 0 to 1000, 0 to 2500 ppm NOX
hours at constant temperature
less than ±0.2 ppm or ±2% of fullscale, whichever is greater, over
any 10°C interval from 4 to 40°C (for rate change of 10°C or less
per hour)
90% of fullscale in less than 1 minute
22.0 cm x 48.3 cm x 48.3 cm (H x W x D)
X
X
Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
1-2
Model 951C
Instruction Manual
748214-V
June 2009
SECTION 2
INSTALLATION
2-1 UNPACKING
Carefully examine the shipping carton and contents for signs of damage. Immediately notify
the shipping carrier if the carton or its contents
are damaged. Retain the carton and packing
material until the instrument is operational.
2-2 LOCATION
See drawing 654063 for Outline and Mounting
dimensions.
Install analyzer in a clean area, free from moisture and excessive vibration, at a stable temperature within 4 to 40°C.
The analyzer should be mounted near the sample source to minimize sample-transport time.
A temperature control system maintains the internal temperature of analyzer at 50°C (122°F)
to ensure proper operation over an ambient
temperature range of 4°C to 40°C (40°F to
104°F). Temperatures outside these limits necessitate use of special temperature-controlling
equipment or environmental protection. Also,
the ambient temperature should not change at a
rate exceeding 10°C/hr.
The cylinders of air and span gas should be located in an area of constant ambient temperature (±10°C).
2-3 VOLTAGE REQUIREMENTS
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance this
instrument must be connected to a properly
grounded three-wire source of power.
This instrument was shipped from the factory
set up to operate on 115 VAC, 50/60 Hz electric
power. For operation on 230 VAC, 50/60 Hz, on
the Power Supply Board (Figure 3-9, page 3-5)
position voltage select switches S1, S2, S3
(Figure 2-1, page 2-2) and S3 (located on the
Temperature Control Board (Figure 2-2, page 2-
2) in the 230 VAC position.
Refer to Figure 2-4, page 2-3. On the rear of the
analyzer, remove the 6.25 A fuse (P/N 902413)
and replace with the 3.15 A fuse (P/N 898587)
provided in the shipping kit.
2-4 ELECTRICAL CONNECTIONS
The power and output (recorder and current)
cable glands are supplied loose in the shipping
kit to allow cable installation to connectors or
terminal strips.
Cable Gland Part No.
Power 899330
Recorder 899329
Remove rear cover to access terminals. Route
each cable through the cable gland and connect
to the appropriate connector or terminal strip,
tighten the gland.
a. Line Power Connections
Refer to Figure 2-3 (page 2-3), Figure 2-4
(page 2-3), Figure 2-5 (page 2-4) and drawing 654063. If this instrument is located on
a bench or table top or is installed in a protected rack, panel or cabinet, power may be
connected via a 3-wire flexible power cord,
minimum 18 AWG (max. O.D. 0.480", min.
O./D. 0.270"), through the hole labeled
POWER, utilizing connector gland (P/N
899330) provided.
Route the power cable through the cable
gland and connect the leads to TB1 on the
rear panel. Tighten the cable gland adequately to prevent rotation or slippage of the
power cable. Since the rear terminals do not
slide out with the chassis, no excess power
cable slack is necessary.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748214-V
June 2009
Model 951C
The following power cord and/or support
feet (for bench top use) are available:
•
Power Cord 634061
•
North American power cord set
(10 foot)
•
Enclosure Support Kit 634958
•
Enclosure support feet (4)
•
Power Cord/Enclosure Support
Kit 654008
•
North American power cord set
(10 foot)
•
Enclosure support feet (4)
If the instrument is permanently mounted in
an open panel or rack, use electrical metal
tubing or conduit.
Set switch window for voltage required.
Figure 2-1. Power Supply Board Voltage Select Switches
S3
Set switch window for voltage required.
Figure 2-2. Temperature Control Board
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.