Emergency Care • OR/ Anesthesia • Critical Care • Perinatal Care • Home Care
Revision 0
6013.053
MS14879
Because you care
Copyright by Dräger Medical AG & Co. KGaA, Lübeck, Germany.
No reproduction allowed for commercial purposes.
Read and understand the Instructions for Use/Operator’s Manual.
This Technical Documentation does not replace the Instructions for Use/Operator’s
Manual.
The warranty and liability conditions of the general terms and conditions for business
transactions of Dräger Medical AG & Co. KGaA are not extended by this Technical
Documentation.
Observe all applicable technical laws and regulations.
Insofar as reference is made to laws, regulations or standards, these are based on the
legal system of the Federal Republic of Germany. Observe the laws and regulations
applicable in your country.
Contents
General
1Advisory 3
2Important information 3
2.1Symbols and Definitions ......................................................................................................... 3
1.10 Patient-Related Data Not Retained or Monitor Fails to Compute Trends ............................. 63
Annex
Spare parts list
Test List
Problem Report
All rights reserved. Copyright reserved.
IV
K6013053IECIVZ.fm 03.12.04
Dräger Medical AG & Co. KGaA Contents
General
1
2
Gamma / Gamma XL Patient MonitorGeneral
General
1AdvisoryThis document corresponds to the version/revision level effective at the time
of system delivery. Revisions to hardcopy documentation are not automatically distributed.
The installation and service of equipment described herein is to be performed
by qualified personnel who are employed by Dräger Medical or one of its affiliates or who are otherwise authorized by Dräger Medical or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly
affiliated with or authorized by Dräger Medical or one of its affiliates are
directed to contact one of the local offices of Dräger Medical or one of its affiliates before attempting installation or service procedures.
2Important informa-
tion
This Technical Documentation/Service Manual conforms to the International
Standard IEC 60601-1.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation/
Service Manual, the equipment may operate improperly or unsafely, or the
equipment could be damaged.
The maintenance procedures described in this Technical Documentation/
Service Manual may be performed by qualified service personnel only. These
maintenance procedures do not replace inspections and servicing by Dräger
Medical.
Strictly follow the Instructions for Use/Operating Instructions! This
Technical Documentation does not replace the Instructions for
Use/Operating Instructions. Any use of the product requires full
understanding and strict observation of the product-specific Instructions for Use/Operating Instructions.
Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany.
2.1Symbols and Definitions
This symbol is used to provide important information that, if ignored,
could lead directly to a patient’s or operator’s injury. It is also used to
provide important information that, if ignored, could lead directly to
equipment damage and, indirectly, to a patient’s injury.
The following three alert levels are used in this documentation to indicate a
hazardous situation and how to avoid it.
All rights reserved. Copyright reserved.
Gamma_Gamma_XL_SM_General.fm 03.12.04
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
3
GeneralGamma / Gamma XL Patient Monitor
DangerDanger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
WarningWarning indicates a potentially hazardous situation
which, if not avoided, could result in death or serious
injury.
CautionCaution indicates a hazardous situation which, if not
avoided, may result in minor or moderate injury. Caution
may also be used to alert against unsafe practices.
Danger
Danger indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
Note
This symbol is used to provide additional information, operating tips, or
maintenance suggestions.
Definitions according to German standard DIN 31051:
Inspection= examination of actual condition
Servicing= measures to maintain specified condition
Repair= measures to restore specified condition
Maintenance= inspection, servicing, and repair
3IntroductionIn keeping with the service strategy for the Infinity Gamma Patient Monitor
and the Infinity Gamma XL Patient Monitor, this technical manual provides
the necessary information required to maintain a Gamma/Gamma XL Patient
Monitor in the field. The Gamma and Gamma XL are both stationary and portable monitors designed to monitor patient vital signs (refer to user’s guide for
monitoring options). For stationary operation near a bedside, the monitor is
connected to an AC/DC power adapter or placed on a specially designed
docking station attached to a shelf, wall, or rolling stand that securely locks it
into place. While on the docking station, the monitor is powered by an IDS
power supply. When the monitor is detached from an IDS, it is powered by a
lead acid battery or by an optional Lithium ion battery. The monitor is reattached to the AC/DC Power Adapter or placed back on an IDS to recharge
the battery.
3.1Service StrategyThe monitor has been designed for high reliability, with an estimated MTBF of
50,000 hours (5.7 yrs.) of continuous operation.Therefore, the service strategy is based on few failures in the field, a clear definition of failure analysis by
field service personnel, and a quick repair turnaround. The field repair philosophy is based on the distributed and approved spare parts list.
4
This manual is intended to serve as a source of technical information, for
qualified field service personnel to use in maintaining a Gamma/Gamma XL
patient monitor in accordance with the Dräger Medical Service Strategy. Field
service is expected to be successful “First-Time Every Time.”
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
Gamma_Gamma_XL_SM_General.fm 03.12.04
All rights reserved. Copyright reserved.
Gamma / Gamma XL Patient MonitorGeneral
4Product OverviewGamma and Gamma XL Patient Monitors are light-weight, battery-equipped,
hand-held or semi-permanently mounted devices for general purpose monitoring of a preconfigured set of physiological parameters. When not connected to a hospital’s main ac power, they use a battery with approxi-mately
1¼ hours (3 hrs. for Li option battery) of operating time. A power adapter or
IDS, which also charges the battery, can be used to operate the monitor from
the hospital’s main ac power circuit.
4.1Monitored Patient
Parameters
4.2Gamma/Gamma XL
Monitor Controls
The Gamma/Gamma XL monitors the following physiological parameters:
•ECG (three-lead, five-lead, or six-lead pod)
•Respiration
•Pulse Oximetry (SpO2 and PLS)
•Temperature
•NBP
•IBP1, IBP2 (locked option)
•etCO2 via PodComm Port (locked option)
•Arrhythmia
•OCRG (locked option)
•Dual Lead S-T Segment Analysis (locked option)
•Anesthetic Gas Monitoring (Gamma XL only, locked option)
All functions are controlled by a 16-position rotary knob and nine front panel
fixed keys - Alarm Silence, Record, Alarm Limits, NBP Start/Stop, All Alarms
Off, Fast Access, Main Screen, Menu, and ON/OFF. Turning the rotary knob
locates different menu items, and pressing the knob in selects the item.
Depending on the item selected, pressing the knob in may either bring up
another menu or initiate an action. See Section 5 Technical Data. For detailed
operating instructions, consult the Gamma/Gamma XL Patient Monitor User
Guide applicable to the installed software.
4.3TFT-LCD DisplayThe Gamma Patient Monitor has a 6.5 inch (16.5cm), 3-channel (optional 4th
channel) color TFT-LCD display. The Gamma XL Patient Monitor has an 8.3
inch (21cm), 4-channel color TFT-LCD display. Waveforms display in Erase
Bar mode at 25 ±20% mm/s (except for respiration and etCO2 waveforms
which display at 6.25 ±20% mm/s). All displays for a given parameter (label,
unit of measure, and waveform) are in the same color. If a waveform is not
displayed for a parameter, its label is gray.
4.4AlarmsAlarm limits can be set either on a user-definable setup table, or automatic-
ally based on current parameter values. Three alarm grades, each with a distinct alarm tone, announce alarm situations of varying severity, as follows:
•life-threatening (asystole or ventricular fibrillation - red)
•serious (parameter limit alarms - yellow)
•advisory (technical alarms - white)
The message field background and parameter field of the parameter in alarm,
and alarm LED, are displayed in the color associated with the alarm grade as
given above.
All rights reserved. Copyright reserved.
Gamma_Gamma_XL_SM_General.fm 03.12.04
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
5
GeneralGamma / Gamma XL Patient Monitor
4.5Monitor/Software
Tracking
Each monitor has a unique ID chip installed in its rear housing for diagnostic
and tracking purposes, and un/locking optional software features.
5Technical DataTechnical Data included in this Section is as of publication date of this Man-
ual. Changes are reported in User Guide applicable to installed SW.
5.1GeneralTable 1General Specifications
ParameterSpecification
Power Requirements100-250 VAC through AC power adapter
Mains Frequency50/60 Hz
AC Power Consumption60 VA AC
Battery TypeLead-acid: PANASONIC LC-T121R8PU or equivalent
Lithium-ion: Dräger Medical Li+ Battery Pack
DC Input11 - 14 V; 32 W continuous, 49 W peak
Battery Operating Time (means running with NBP measurement every 15 min @ 25°C temperature, no etCO2
running
Battery Recharging TimeLead-acid: 5 ½ hours, typical
Lead-acid: 75 mins
Lithium-ion: 180 mins
Lithium-ion: 8 hours, typical
Battery Charge/Discharge/Charge:Lithium-ion only (operating as defined above): 2 hours,
charging for 2 hours, operating 2 hours
Patient Leakage Current<10 µA @ 110 V and 60 Hz (per UL 544)
<10 µA @ 220 V and 50 Hz (per IEC 601-1)
Chassis Leakage Current with battery eliminator<100 µA @ 110 V and 60 Hz (per UL 544)
Sweep Speedsfixed 25 mm/s ±20% for ECG, SpO2, and IBP curves
fixed 6.25 mm/s ±20% for Resp and etCO2 curves
fixed 1.0 mm/s ±20% for optional OCRG curve
Display ModeErase bar (updates waveforms from left to right)
All rights reserved. Copyright reserved.
Gamma_Gamma_XL_SM_General.fm 03.12.04
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
7
GeneralGamma / Gamma XL Patient Monitor
5.4OutputsTable 4Output Specifications
ParameterSpecification
QRS Synchronization:
Timing:
Output Pulse:
Alarm Output12 V Open collector output for external alarm indicator
RecorderUART interface w/ recorder through interface plate or
Debug PortUART interface w/ a PC to retrieve diagnostic information
External VGAVideo signals sent to external VGA display for remote
Export ProtocolUART interface w/ external devices using proprietary
NetworkSerial connection to Infinity Network through Infinity LAN
For heart rates from 30 to 250 [1/min], with QRS widths
from 40 to 120 msec and QRS amplitudes from 0.5 to 5
mV, a sync pulse is delayed no more than 35 msec from
peak of R-wave for each valid QRS complex.
+12 V, 100 ms duration
docking station connector
through interface plate or docking station connector
viewing of Gamma screen. -- not available when Infinity
Serial Hub interface plate in use.
export protocol. -- not available when Infinity Serial Hub
interface plate in use.
or docking station connector, or with a wireless PC card
in an Infinity Wireless Network.
5.5ConnectorsTable 5Connector Specifications
ParameterSpecification
DC InputDräger Medical 2-pin power connector
Docking StationDräger Medical 28-pin connector to provide Alarm Output,
Recorder, Debug Port, Network, External VGA and Power
For patient parameter specifications, refer to User Guide applicable to
installed software version.
8
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
Gamma_Gamma_XL_SM_General.fm 03.12.04
All rights reserved. Copyright reserved.
Gamma / Gamma XL Patient MonitorGeneral
6Monitor ControlsThe rotary knob in the lower right corner of the front panel is a pointing and
selecting device. Turn the knob to select a screen area or menu item or to
change a default value, and press the knob in to confirm your selection and to
set a default value. Press Main Screen key to return to the MAIN screen.
Note
Instructions in this chapter are intended to provide only a cursory overview
of basic monitor controls for accessing and performing service-related
functions. Refer to the User Guide for the installed software version for
complete operating information.
6.1Main Screen KeyPressing the Main Screen key exits the current menu or screen and displays
the home screen.
6.2Menu Key-- provides access to Main menu. In general, functions of direct concern to
the FSE or Biomed are accessed via Monitor Setup → Biomed on Main
menu. Only authorized personnel should perform password-protected service-related functions. Use Biomed password (375) to access the following:
•Save Setups - Confirm or Cancel
•Locked Options - four locks into which monitor-specific 2-digit codes must
be entered to enable locked options
•Diagnostic Logs
•Units
a) Temperature - °C or °F
b) etCO2 - mmHg, kPa, Vol %
c) Pressure - mmHg or kPa
d) ST - mm, mV
•Service - requires Service password. (The password is given on the Service Setup Instructions for the installed software version.)
a) Update Software Load
b) Test Pulse (Confirm or Cancel - one-shot test pulses for ECG (1mV
spike) and Temp (-5°C and +50°C, respectively). An additional test is
performed for IBP, Resp Pulse, and SpO2. Test indication is reported
in trend table.)
c) Monitor Setup Language
d) SCIO Port X5
e) USB
f)Data Collection OFF
g) SpO2
h) NBP
i)Line Frequency 50
j)60
k) Service
l)Network SetupNetwork Config SSID
m) Central Yes
All rights reserved. Copyright reserved.
Gamma_Gamma_XL_SM_General.fm 03.12.04
n) Station No
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
9
GeneralGamma / Gamma XL Patient Monitor
o) Keep Bed Yes
p) Label No
q) Network WEP Transmit Key
r)Key 1
s) Key 2
t)Key 3
u) Key 4
v) Network Config
w) Network Setup
x) Network IP Address
y) Info Network Setup
z) Service
•Exit
Note
Set the line frequency equal to the ac mains line according to local conditions (50 or 60 Hz). An incorrect setting of line frequency can cause artifact
or excessive waveform noise on the ECG waveform.
6.3Alarm Limits Key-- calls up a setup table for alarms.
1. Turn rotary knob to select desired parameter field and limits, and press
knob in to activate your selection.
2. The number representing the limit value turns black on a blue background, indicating that you can change it. Turn knob to change value.
3. When desired setting is displayed, press knob in to set value.
4. Press MAIN Screen key to return to MAIN screen.
6.4Alarm Silence Key-- silences an active alarm tone for 1 minute ±5 seconds, and turns active
blinking parameter areas into active steady parameter areas
6.5All Alarms Off Key-- suspends alarms for a fixed 3-minute ±5 second period.
6.6NBP Start/Stop Key-- starts and stops non-invasive blood pressure measurement.
6.7Fast Access Key-- allows access to the monitor’s bottom channel menu as well as tabular
trends, graphical trends and Event recall screen.
6.8Record Key-- starts a manual, timed recording on a connected R50™ recorder or on a
networked postscript laser printer in an Infinity Network.
10
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
Gamma_Gamma_XL_SM_General.fm 03.12.04
All rights reserved. Copyright reserved.
Gamma / Gamma XL Patient MonitorGeneral
Note
If a recorder or networked postscript laser printer is not available, pressing
the Record fixed key writes 15 seconds of waveform and vital signs information to internal memory. Monitors can store up to ten recordings, which
are automatically printed as soon as the recorder or networked postscript
laser printer is available.
All rights reserved. Copyright reserved.
Gamma_Gamma_XL_SM_General.fm 03.12.04
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
11
GeneralGamma / Gamma XL Patient Monitor
12
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
1OverviewInfinity Gamma and Gamma XL Patient Monitors are configured monitors
running on one processor, an MPC823E Power PC processor, which attends
to all monitoring functions, controls all graphics functions, generates video
and timing signals for the LCD screen, and interfaces with the PCMCIA card
and USB port. It also performs several peripheral control functions, such as
NIBP control, audio volume control, and timing generation for the front end.
See Figure 1.
2Parameter InputsThe data acquisition front end acquires and digitizes signals derived from a
three-, five-, or six-electrode ECG patient lead set, a Nellcor® or Masimo®
SpO2 transducer, an Impedance respiration measurement system, a thermistor-based Temperature transducer, and two strain-gauge IBP transducers
(IBP2 = locked option). The NIBP main transducer signal is digitized together
with the rest of the front end parameters. See Section 4 Front End and Section 5 Physiological Parameter Data Acquisition for more detailed information.
3Main PC BoardThe Main MPC823E Power PC processor not only attends to monitoring
functions, but also controls all graphics functions, generates the video and
timing signals for the LCD screen, interfaces with the PCMCIA card and USB
All rights reserved. Copyright reserved.
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
15
Function DescriptionGamma / Gamma XL Patient Monitor
port, and controls the network link. In addition, it performs a host of peripheral
control functions, such as NIBP control, audio volume control, and timing
generation for the front end.
3.1LCD ControlA set of buffer/drivers are used to drive the 6.5” screen in Gamma monitors or
the 8.3” screen in the Gamma XL monitors. In parallel, a triple video DAC
generates analog RGB signals for an external VGA monitor (typically a CRT
or LCD).
3.2Network InterfaceThe Infinity Gamma / Gamma XL Monitor interfaces with the physical inter-
face device (e.g., IDS) automatically, when connection to the device has been
detected. Connections to Infinity network services are established and maintained by software components resident on both the Gamma and Gamma XL
Monitor and the physical interface device.
3.3Front Panel CircuitryThe front panel circuit processes the audio information, drives the fluorescent
tubes on the LCD, implements a secondary alarm in case the unit resets or
turns off, and routes the video and timing signals to the screen. It also routes
the UART signal coming from the Pod interface to the main board Quad
UART.
3.4Pod InterfaceThe Pod Interface generates an isolated voltage to power the pod and also
converts the Pod Comm protocol from the pod into a UART stream that can
be interpreted by the microprocessor.
3.5Battery Control and
ON/OFF Control
3.6BOOT Process,
Flash Memory, and
DRAM
The Pb-acid or Lithium battery charging and discharging cycles are controlled
by a special charger circuit. The circuit initiates a charge cycle when commanded by the microcontroller. The charge cycle for a Pb-acid consists of a
bulk charge period in which the battery is being supplied a constant current of
~400mA, a constant voltage period in which the battery voltage is held constant at ~14.8V and the current is allowed to diminish as the charge
approaches 100%, and a float cycle in which the voltage is maintained at
~13.7V. For Lithium batteries, the charger circuit acts as a constant voltage
source of 16.8V. The battery is charged from a switching supply controlled by
the charger chip. The microcontroller also reads the front panel keys and the
rotary knob, encodes the information coming from them, and routes it to the
main processor. When the On/Stdby key is pressed, it turns the monitor on
and off. In addition, the microcontroller controls the NIBP safety timer.
The BOOT EEPROM contains the boot code and is preprogrammed at the
factory. It can be reprogrammed in the field by means of a special PCMCIA
card, if required. (Contact your local Dräger Medical service representative.)
The executable software normally resides as compressed operational code in
Flash memory. When the 68HC912D60A microprocessor senses that the
on/off switch on the front panel has been pressed, it turns on (or turns off) the
3.3V and 5V supplies. As the 3.3V supply turns on, it wakes up the MPC823E
main processor, which begins execution from the BOOT PROM. During boot
initialization, the main processor attempts to read the Memory Card to detect
authorized software. If a authorized software memory card is present, the
software is loaded from the card. Otherwise, the main processor loads software from the Flash to the main processor DRAM, from which it completes
initialization and enters operational mode. DRAM contains expanded operational code, and data space variables and stacks.
16
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
3.7SRAMThe 512K x 8 SRAM is battery backed up and is used for error logs, trends,
recordings and other non-volatile memory uses.
3.868HC912D60A
Microcontroller
On/Off controlWhen the ON/OFF push button (either local or remote is pressed), the micro-
NBP Valve modulationWhen directed by the main processor, the microcontroller supplies modula-
NBP Safety TimerWhen the pump or the valve V2 are turned on, the microcontroller initiates a
Battery ChargerThe microcontroller initiates a battery charge when needed, and stops the
The 68HC912D60A microcontroller, with 64K of EEPROM and 2048 bytes of
RAM, is powered as long as there is a main supply plugged into the system
or when the user presses the ON/OFF button. The code is stored in its internal flash memory, but can be downloaded from the MPC823E. The microcontroller performs the following functions:
controller activates the 3.3V and 5V supplies, which wakes up the MPC823E
through a power-on reset. In addition, the microcontroller has control over a
flyback supply, which comes on any time the unit is plugged into AC power (in
order to charge the battery) or is turned on.
The microcontroller also reads the front panel keys and the rotary knob,
encodes the information coming from them, and routes it to the main processor.
tion signals for the two NBP manifold valves.
128 sec. timer (90 sec. or 60 sec. for neonates) which, if exceeded, produces
an NBP fault and results in cut off of main 12V power to the NBP manifold.
charging process when the battery reaches full capacity. It can recognize
whether a Pb or Lithium battery is connected into the system, and directs the
battery controller chip to charge to different levels depending on the battery
type. See Section 3.5 Battery Control and ON/OFF Control. The microcontrol-
ler also acquires the battery voltage and current for monitoring purposes.
Recorder PowerThe microcontroller controls power applied to a stand-alone R50 Recorder.
Main Audio GeneratorThe microcontroller generates the fundamental audio frequency of the unit’s
tone generator, as directed by the main microprocessor.
USB PowerWhen directed by the MPC823E, the 68HC912D60A microcontroller turns
power ON/OFF on USB buss and determines the transaction speed.
4Front EndAll physiological signals (except etCO2) are digitized through a high speed
multiplexing system and a common 16 bit ADC. The data is then transferred
through the isolation barrier to an HDLC port in the main processor, where it
is digitally filtered and processed.
4.1NBP ControlThe NBP main transducer signal is digitized together with the rest of the front
end parameters. However, the redundant (overpressure) transducer is processed separately on the grounded end of the board. The pump on/off signal
and valve enable signals are generated off of the MPC 821 microprocessor.
The PWM signals for the valve flow control and the redundant safety timer
are implemented in a separate microcontroller (MC68HC912D60A).
4.2Safety•Patient isolation withstands 5kV during defib.
•Leakage currents are limited to safe values normally and during single
fault conditions.
•Patient is protected against electrosurgical burns at the electrodes.
All rights reserved. Copyright reserved.
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
17
Function DescriptionGamma / Gamma XL Patient Monitor
•Defibrillation protection does not drain excessive current away from the
patient.
•Specially shielded connectors and cables are used to provide excellent
immunity up to 1000MHz and can not be touched by the patient even
when disconnected.
•Single cable from MultiMed Pod to main Gamma / Gamma XL Monitor
unit reduces clutter between bed and monitor.
Figure 2Front End
5Physiological
Parameter Data
Acquisition
Transducers gather physiological data at the patient and feed them into the
small MultiMed Pod at the bed. The MultiMed Pod in turn is connected via a
3-meter cable to the front end in the main unit where analog ECG, Respiration, Temperature, and SpO2 signals are converted to digital form and sent
through isolators for processing.
5.1ECG/RespThe MultiMed Pod located close to the patient accepts a set of 3, 5 or 6
shielded ECG electrode leads, an SpO2 (Nellcor) cable adapter, and a temperature sensor. The ECG section contains RF filters, and overvoltage
clamps that include 1k series resistors to limit shunting of defibrillator current.
The SpO2 and temperature sections also contain RF filters. Impedance respi-
18
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
ration is sensed through the ECG electodes. Void-free potting and internal
shielding enable compact containment of high voltage defibrillator and electrosurgery pulses. The small interconnecting cable to the main assembly is
captive at the MultiMed POD but plugs into the MultiMed front end via a specially shielded connector.
The front end accepts physiological signals from the MultiMed POD connector and feeds temperature, respiration, and ECG signals via RF filters, configuration multiplexers, and pre-amplifiers to a high-speed multiplexer driving a
16-bit analog-to-digital (A/D) converter. The data stream is sent to the Main
Processor board via an opto-isolator. Control commands from the Processor
are sent out to the front end on a similar isolating link. Isolated DC power is
also provided.
The ECG signals are conductively coupled to the isolated circuits via currentlimiting series resistors, whereas the SpO2 signals are optically isolated at
the transducer. Temperature signals are doubly insulated at the patient by
disposable boots on the sensors. AC (40kHz) excitation currents for respiration monitoring are dc-isolated by high-voltage ceramic capacitors.
The A/D samples the following parameters:
Table 1Parameter Sampling Table
Parameter# of Channels
ECG4
Pace2
SpO2 Red1
SpO2 IR1
NBP1
Resp1
Te mp2
The hardware pace detector monitors the ECG signal in two of the four channels (those not connected to the chest leads). All other signals are decimated
and filtered using digital signal processing in the MPC823E. High oversampling rate is required to minimize the requirements (and size) of the analog
anti alias filters. Superior rejection to ESU and other types of interference is
achieved with this type of design.
5.1.1ECG•Bandwidth is set flexibly by software filters.
•Reconfigurable neutral selector can drive any electrode.
•Lead-on detection functions with even poor electrodes.
•Calibration voltages can be superimposed on patient wave-forms or onto
flat baselines.
See Figure 3. Composite electrocardiographic (ECG) signals generated by
the heart and by a pacemaker are filtered to reduce RF interference from
impedance respiration and electrosurgery and then injected with dc lead-off
detection currents. Over-voltage clamps protect the semiconductors from the
surges passing the spark gaps in the MultiMed Pod and also reduce the dc
All rights reserved. Copyright reserved.
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
current applied to the patient due to a component fault.
19
Function DescriptionGamma / Gamma XL Patient Monitor
Figure 3Lead-Forming Network
5.1.2Lead SelectionA lead-forming network following the RF filter generates the necessary refer-
ence points for electrocardiographic measurements. Both normal leads (I, II,
III, V1 and V2) and augmented leads (aVL, avR, and avF) can be obtained.
Four differential channels generate the main axes I, II, V1 and V2. The
remaining leads are derived mathematically as indicated in the vector diagram of Figure 3.
5.1.3Lead-Off DetectionLead-off detection is accomplished by introducing a very small current into
each patient electrode, which would drive the corresponding input high if it
were disconnected. A set of five comparators detects a lead-off condition.
5.1.4Low-Pass Filtering and
Common Mode
Enhancement
The ECG preamplifier has a flat frequency response of 0.5 - 40Hz, with a
software notch filter at 50/60 Hz. A 180° combined signal drives the neutral
electrode to increase the CMMR.
5.2RespirationImpedance respiration is monitored by injecting a 40 kHz square wave of cur-
rent into the RA electrode. The resulting 40 kHz voltage drop between the RA
+ LL electrodes is proportional to the impedance. Especially balanced true
current sources do not load the ECG electrodes or distort the ECG morphology. The returning 40 kHz differential voltage is amplified, synchronously
demodulated, and low-pass filtered. An AC-coupled stage with an “autobloc”
DC restorer feeds the input to the A/D converter with a nominal output of 60
mV per Ohm.
The pulse oximeter circuit uses a Nellcor® sensor to detect the oxygen saturation level in arterial blood flow. Determination of the concentration of oxygen in the blood is based upon the principle that the absorption of red (R)
light depends on the degree of oxygenation of the blood, whereas the absorption of infrared (IR) radiation is relatively independent of oxygenation and
causes only constant attenuation. See Figure 4. In the SpO2 sensor, R and
IR light emitting diodes (LEDs) are alternately pulsed ON at a 25% duty cycle.
The light is transmitted through a well-perfused part of the body, such as a
fingertip or an ear lobe. The intensity of light (including ambient) transmitted
through or scattered by the blood is converted to a current by a photodiode in
the sensor. The current that appears when both LEDs are OFF depends
mainly on the ambient light, which is later subtracted to leave only the R or IR
signal levels. The large dynamic range of the light intensities requires constant automatic monitoring and adjustment.
The intensities of the R and IR sources are independently controlled by two
digital-to-analog converters (DACs) attenuating the 2.5 V reference. These
levels or zero are sequentially selected by a multiplexer, and converted to a
driving current which is further guided or inverted by an output multiplexer to
the LEDs in the sensor.
5.3.1SpO2 Front EndThe primary purpose of the SpO2 front end is to convert the sensor’s analog
signal into individual digitized signals for the red and infrared analog signals
for processing by the microprocessor. See Figure 5. Circuitry in the front end
first eliminates the non-pulsatile component in the input signal, then demultiplexes the resulting pulsatile signal to separate the R and IR signal components, and finally converts the demultiplexed R and IR analog signals into
serial digital data streams.
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
21
Function DescriptionGamma / Gamma XL Patient Monitor
Figure 5Sensor LED Timing Diagram
A sequence of light pulses, driven from the chopped current source in the
sensor LEDs, are passed through a finger or an earlobe to a photodiode. The
sensor LEDs are connected in an anti-parallel fashion on one pair of wires. A
timing generator controls the sensor LEDs and signal multiplexing/ demultiplexing (see Figure 5) by means of three control signals:
•IRONL (infrared LED)
•AMBONL (LEDS not lit)
•REDONL (red LED)
5.3.2Input StageA preamplifier converts the photocurrent to an equivalent voltage, and applies
it to a 20 Hz high-pass filter that removes the non-pulsatile component. The
output of the preamplifier is fed to a saturation detector.
5.3.3Brightness ControlIf the output of the preamplifier is in saturation, the gate array provides a signal to the digital-to-analog converters (DACs), which controls the drive current to increase or decrease the brightness of the LEDs.
Controlling LED brightness extends the system dynamic range. For a very
transparent subject it may not be possible to reduce the gain to prevent saturation. In that event, the brightness must be reduced. An additional purpose is
to equalize the received amplitude of each wavelength. If both LEDs are
turned ON to maximum brightness, and the software finds an extraordinary
difference between the two, the microprocessor tends to reduce that difference by equalizing the R or IR brightness signals.
22
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
The ambient rejection amplifier is a synchronous detector. The signal applied
to its inverting input is a composite of R, IR, and ambient signals. The noninverting input is the same signal gated by the timing generator. This synchronously multiplexes the IR, ambient, and R analog signals.
Figure 6 IBP Functional Block Diagram
The IBP circuit has been designed to be used with a strain gauge pressure
transducer. See Figure 6. The analog portion of the IBP circuit provides excitation voltages for resistance bridge transducers. These voltages are derived
from a reference which is also used to derive the A/D converter reference
voltage. At the circuit input, a resistor divider network provides for transducer
unplugged detection. R-C filtering and protection diodes limit the effects produced during electrosurgery, defibrillation, and other such procedures. A
selector multiplexer allows for the insertion of calibration signals into the
amplifier stage. The multiplexor feeds the pressure signal to a buffer amplifier, which in turn feeds the AD converter analog input. This allows the monitor to measure pressure signals in a range greater than ±700 mmHg with a
resolution of approximately .02mmHg/LSB.
When no pressure transducer is plugged into the monitor, the resistor divider
network puts a negative signal into the instrumentation amplifier, which propagates through the system to indicate the unplugged condition.
All rights reserved. Copyright reserved.
Dräger Medical AG & Co. KGaA6013.053 Revision 0 Released
23
Loading...
+ 66 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.