Druck PTX1230, PTX1235 Data sheet

GE
Sensing
• ±0.25% accuracy
• Excellent stability
• All welded titanium construction
• Backed by five-year corrosion warranty
• FM, CSA intrinsically safe, CE marked
Standard vented cable is polyurethane. Flouropolymer
cable is available on the PTX/PDCR 1280 Series Sensors.
The PTX/PDCR 1230 Series sensors are ideal for use in applications where smaller size is an advantage, such as municipal water supply wells, leachate wells, irrigation projects, etc. The titanium construction also makes the devices suitable for seawater measurement applications.
An advanced micromachined silicon piezoresistive pressure sensor provides excellent performance and resistance to shock and vibration. A tough, polyurethane cable is molded to the transducer body, providing a high integrity, waterproof assembly. The cable is strengthened with Kevlar
®
so that there is no measurable elongation
when the cable is lowered into deep wells.
The fully isolated, titanium design ensures long term reliable measurements in water and wastewater management, industrial, process and marine applications.
PTX/PDCR 1230 Series
Druck Depth and Level Pressure Sensors
Features
PTX/PDCR 1230 Series is a Druck product. Druck has joined other GE high-technology sensing businesses under a new name
__
GE Industrial, Sensing.
g
The PTX/PDCR 1230 Series submersible/depth pressure transducers are specifically designed for depth/level measurements in the groundwater, well water, canals, rivers, and other similar applications. The titanium construction ensures excellent life in the most hostile environments, including corrosive and hazardous chemical applications, as well as environments where icon bacteria may be present. Druck backs its titanium construction with a five year corrosion warranty.
PTX/PDCR 1230 Specifications
General
Operating Ranges
• 1, 2.5 psi (0.75, 1.5 mH2O) gauge
• 5, 10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 500, 900 psi (3.5, 7, 10, 15, 20, 35, 50, 70, 100, 150, 200, 350, 600 mH2O) gauge or absolute
Other pressure units can be specif ied.
Overpressure
• 10X for range 1 psig (0.75 mH
2
O)
• 8X 2.5 psi (1.5 mH2O)
• 6X 5 psi (1.5 to 3.5 mH2O)
• 4X for ranges above 5 psi (3.5 mH2O)
• 2000 psi (1400 mH2O) maximum
Pressure Media
Fluids compatible with titanium and polyurethane
Transduction Principal
Piezoresistive micromachined silicon strain gauge
Combined Error (Non-linearity, Hysteresis, and Repeatability)
±0.25% full scale (FS) base standard line (BSL)
Temperature Effects Over Compensated Range of 30°F to 86°F (-1°C to 30°C)
±2.0% FS total error band (TEB) for ranges below 5 psi (3.5 mH
2
O)
Resolution
Infinite
Insulation Resistance
100 MS @50 VDC
Relative Humidity
0 to 100%
Compensated Temperature Range
30°F to 86°F (-1°C to 30°C)
Operating Temperature Range
-5°F to 140°F (-20°C to 60°C)
Electrical Characteristics
Output
PTX 1230
Two-wire, 4 to 20 mA
PDCR 1230
• 25 mV/V for range 1 psi (0.75 mH
2
O)
• 5 mV/V for range 2.5 psi (1.5 and 3.5 mH2O)
• 10 mV/V for range 10 psi and above (7 mH2O)
Output is ratiometric to power supply (2.5 VDC minimum)
Electrical Connection
Vented polyurethane cable Red
__
Positive supply
Black
__
Negative supply
Shield
__
connected to case
Any other conductors not connected.
Electrical Connection
Vented polyurethans cable Red
__
Positive supply
White
__
Negative supply
Yellow
__
Positive output
Blue
__
Negative output
Shield
__
connected to case
Any other conductors not connected.
Installation Drawings
Ø 0.69 in
(17.5 mm)
7.8 in (198 mm) approximately
M14 x 1.5 in (38 mm) exposed
thread with nose cone removed
5/16 nominal
3.8 in (96.52 mm) approximately
Length to order
M14 x 1.5 in (38 mm) exposed
thread with nose cone removed
5/16 nominal
Ø 0.69 in
(17.5 mm)
GE
Sensing
Length to order
PTX 1230
PDCR 1230
Loading...
+ 2 hidden pages